Home

ResMed Flow Generator VPAP Adapt SV User's Manual

image

Contents

1. A m xo 7 e HumidAire HumidAire 2i heated humidifier heated humidifier Humidifier accessory HumidAire only Pa S Medium air tubing amp 21 52 cm Figure 7 Humidifiers compatible with the VPAP Adapt SV When using a humidifier stop the VPAP Adapt SV by using the Start Stop key or by removing the mask to trigger the SmartStop function Do not switch the VPAP Adapt SV off at the power switch or the power source while it is running as this could result in condensation in the motor Please refer to the relevant humidifier user manual for detailed instructions and warnings CAUTIONS Be very careful not to allow water to enter the VPAP Adapt SV as this could affect therapy and or damage the device Do not tip the humidifier while it is connected to the VPAP Adapt SV If water enters the VPAP Adapt SV device turn off the main power switch at the back of the unit and unplug it from the power outlet Return the unit to your eguipment supplier for checking Connecting a Humidifier Note After connecting or removing a humidifier always perform the Air Circuit Learn function see page 21 WARNING When using a humidifier position it so that it is lower than your head will be during sleep and at the same level or lower than the VPAP Adapt SV HumidAire 2i The HumidAire 2i provides heated humidification for the VPAP Adapt SV To attach the HumidAire 2i remove the VPAP Adapt SV front cover and attach the hum
2. when power is restored CAUTION Remove the mask from your face if the power fails High Pressure High Average Pressure Alarms Triggered by High Pressure Alarm the mask pressure exceeds the set trigger level 30 cm H20 for more than 700 milliseconds High Average Pressure Alarm the mask pressure exceeds the set trigger level 22 cm H20 averaged over approximately one minute If a high pressure alarm sounds The treatment will stop Turn power off Check that the air tube and sensor tube are connected properly Turn power back on Remove mask and perform Air Circuit Learn function Try using the flow generator one more time If the high pressure alarm activates repeatedly discontinue use and return to ResMed for servicing If the alarm does not recur then continue to use as normal NOORWDN Note The alarm system is designed to ignore coughing However if a cough is especially intense and prolonged it may trigger the high pressure alarm Will stop when the VPAP Adapt SV is turned off High Leak Alert Triggered by high mask leak greater than 30 L min for more than 20 seconds the mask being removed while SmartStop is disabled If a high leak alert sounds adjust the mask to minimize leak check that the air tube and sensor tube are connected properly Will stop when the mask leak has been rectified Note The High Leak alert can be turned off in the Leak Alert option i
3. Remove the filter cover at the back of the flow generator Remove and discard the old air filter Insert a new filter the blue tinted side facing towards you Note Match the cut corner of the filter with the shape of the casing Replace the air filter cover The blue tinted side of the filter should be facing out Hypoallergenic Air Filter The ResMed hypo allergenic filter is a standard filter to which an electrostatic filter has been bonded to capture very small particles The filter is intended for those users of ResMed flow generators who will benefit from enhanced filtering of the air delivered during treatment Inspect the air filter every month and check if it is blocked by dirt or contains holes The filter should be replaced every three months or more often if the flow generator is operating in a dusty environment Fit as described above in Air Filter Servicing This product VPAP Adapt SV should be inspected by an authorised ResMed Service Centre five years from the date of manufacture Prior to this the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed Applicable ResMed warranty details are provided with the device at the time of original supply Of course as with all electrical devices if any irregularity becomes apparent you should exercise caution and have the device inspected 5
4. enter 1 OPTIONS enter 1 SERVICING enter T ASV MODE Press down arrow on Up Down key to move to next menu Press Left key to enter menu options 1 Press Left key to select change 2 Press Up Down key to reach setting 3 Press Left key to enter setting 4 Press Right key to exit Figure 14 Using the keys to adjust settings Using the VPAP Adapt SV English 19 Menus The options available in the Settings menu will depend on the therapy you are receiving either ASV or CPAP as set by your clinician VPAP ADAPT SV menu SETTINGS enter l ASV MODE CPAP MODE MASK FULL FACE RAMP 20min change 1 exit change 1 LEARN CIRCUIT MASK FULL FACE yes 1 exit change 1 exit SMARTSTOP OFF LEARN CIRCUIT change 1 exit yes t exit SOUND LEVEL LO SMARTSTOP OFF change 1 exit change 1 exit SOUND LEVEL LOW change exit RESULTS enter t MASK FIT good 1 l exit These menus appear l only if at least one l Smart Data option has been enabled by l the clinician l USAGE 4 56hrs T exit SERVICING enter t SN 1234567890123 4567 l exit PCB 123456789012 exit SW SX432 0201 t exit Figure 15 Navigating the Patient menu 20 Settings Menu You will only see the options relevant to your therapy In CPAP Continuous Positive Airway Pr
5. If the pressure sensor tube cannot be dried completely it should be replaced Daily 1 Disconnect the air tubing and pressure sensor tube and hang them in a clean dry place until next use Do not hang the air tubing in direct sunlight as it may harden and crack over time 2 Clean the mask according to the mask user instructions 3 If you are using a humidifier clean it according to the instructions in the manual Weekly 1 Remove the air tubing and pressure sensor tube from the VPAP Adapt unit and the mask 2 Remove the pressure sensor tube and proximal cuff from the air tubing Wash the mask system according to the instructions supplied with it 4 Wash the air tubing and proximal cuff in warm water using mild liguid soap Rinse thoroughly hang and allow to dry 5 Before next use assemble the mask and headgear according to the mask user instructions 6 Reconnect the pressure sensor tube to the air tubing then connect both to the proximal cuff Attach the proximal cuff to the mask o English Air tubing Remove proximal cufff lt Remove the pressure sensor tube from air tubing e Tubing clips Figure 19 Air tubing assembly removed for cleaning CAUTION Do not use bleach chlorine alcohol or aromatic based solutions including all scented oils moisturising or antibacterial soaps to clean the cushion mask air tubing or the VPAP Adapt SV These solutio
6. by an authorized ResMed service center 2 n a Lu If you feel the unit is not performing properly see Troubleshooting on page 31 Servicing Australia and New Zealand This device is intended to provide safe and reliable operation when operated and maintained in accordance with the instructions provided by ResMed To ensure that your device continues to provide optimum performance it is recommended that this product VPAP Adapt SV be inspected by an authorised ResMed Service Centre five years from the date of purchase CAUTION 22 Inspection and repair should only be performed by an authorized service agent Under no circumstances should you attempt to service or repair the unit yourself Troubleshooting If there is a problem try the following suggestions If the problem cannot be solved contact your eguipment supplier or ResMed Do not attempt to open the flow generator Problem Possible cause Solution Insufficient air delivered Air filter is dirty Replace air filter from flow generator Air tubing is kinked or Straighten or replace tubing punctured Flow generator operating Water may have entered the Remove the water from the erratically ie switching on air tubing air tubing and off pressure varying at random Pressure rises You are talking coughing Avoid talking with the mask inappropriately voluntarily breath holding on and breathe as normally eg while rolling over in bed as possible or intentionally breathin
7. under the Results menu will be displayed automatically for approximately 20 minutes after your session is completed Control Panel Functions The control panel of the VPAP Adapt SV has a keypad that allows you to start or stop treatment adjust settings on the flow generator mute or stop alarms view treatment data OuickView key LCD display Alarm LEDs Alarm mute key Start Stop Left Up Down Right key key key key Figure 13 The VPAP Adapt SV control panel LCD screen Displays information about flow generator settings alarms and treatment The LCD screen is backlit whenever any key is pressed The lighting will turn off automatically when no button has been pressed for two minutes or can be set to stay on see Options Menu on page 23 An alarm message will overwrite any other message on the screen The original screen will reappear when any key is pressed Start Stop key Press the key once to start or stop treatment Up Down key Moves backwards and forwards within a menu and between menus Increases and decreases values of parameters press once to adjust in single increments hold down to move quickly through the values Using the VPAP Adapt SV English 17 per key green Enters a menu Confirms and applies settings Performs the function indicated by the text above it in the LCD eg enter change Right key red Performs the function indi
8. RESMED VPAP Adapt SV User Guide English 228234 Eng 3 08 10 Contents Introduction o Ak Be AE EEUU Es VPAP Adapt SV User owner responsibility Definitions Medical Information Quick Setup Guide seu eee Es e GWA ee ie FU VPAP Adapt SV System LL I LG ees VPAP Adapt SV Components Assembling the VPAP Adapt SV System Humidifiers ResLink Anti bacterial Filter Using DC Power to Run the VPAP Adapt SV Supplemental Oxygen Using the VPAP Adapt SV LI I LL eee Starting Treatment Stopping Treatment Standby Mode Control Panel Functions Adjusting Settings Menus Treatment Screens A Iarms x secar e ut ane ete e Salt os i s Low Mask Pressure Alarm Power Fail Alarm High Pressure High Average Pressure Alarms High Leak Alert Insufficient Pressure Support Alarm Flow Blocked Alarm Cleaning and Maintenance L L uy Daily Weekly Periodically Air Filter Servicing Troubleshooting LLYYL FL e Glossaty mee ret e e etu S E s Technical Specifications LYL YY eee Limited Warranty soies merai e i aE ER R S uu Index diaaa RR o C9 C9 co Introduction VPAP Adapt SV The VPAP Adapt SV flow generator has been designed to be reliable comfortable and easy to use You can observe your treatment and make adjustments to settings using a keypad and LCD screen CAUTION USA federal law restricts this device to sale by or on the order of a physician User Owner Respon
9. ailable in the Results menu If you want to turn off AUTO APPEAR select enter in the SMART DATA screen by pressing the Left key The AUTO APPEAR screen will be displayed Select change by pressing the Left key and you will switch from ON to OFF similarly you can turn it from OFF to ON Your clinician may also turn off AUTO APPEAR Backlight On Display a permanent backlight for the LCD Auto Backlight turns off if no button has been pressed for two minutes Language Change the language of the LCD display Languages available are English French Spanish and Portuguese Using the VPAP Adapt SV 23 Servicing Menu This menu displays the following information SN the flow generator s serial number PCB the flow generator s PCB number PCB printed circuit board SW rthe version of software currently installed on the flow generator Treatment Screens There are three Treatment screens displayed in the VPAP Adapt SV LCD The LCD will display the Treatment screens when treatment commences and during treatment if no key is pressed for 20 minutes Press the OuickView key at any point within a menu to display Treatment Screen 1 You can return to your original screen if you press the OuickView key again within 20 minutes Press the Right key to move from a Treatment screen to the Standby screen If an alarm is triggered while you are in the Treatment screens the arrow will flash Press the Down key until you reach the alarm mes
10. al and workmanship for the period specified below from the date of purchase by the initial consumer This warranty is not transferable Product Warranty Period ResMed humidifiers ResControl ResLink ResTraxx 1 Year ResMed flow generators 2 Years Accessories mask systems including mask frame cushion 90 Days headgear and tubing Excludes single use devices Note Some models are not available in all regions If the product fails under conditions of normal use ResMed will repair or replace at its option the defective product or any of its components This Limited Warranty does not cover a any damage caused as a result of improper use abuse modification or alteration of the product b repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs c any damage or contamination due to cigarette pipe cigar or other smoke d any damage caused by water being spilled on or into a flow generator Warranty is void on product sold or resold outside the region of original purchase Warranty claims on defective product must be made by the initial consumer at the point of purchase This warranty is in lieu of all other express or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may
11. atory and expiratory flow rate The time is measured from the beginning of inspiratory flow to the beginning of pressure rise to IPAP Note The ASV algorithm is designed to treat a patient with essentially normal lungs Hence trigger response is less sensitive and pressure ramp up is more gentle than typical non invasive ventilators Displayed values Value Range Accuracy AR Proximal cuff Pressure 3 to 40 cm H20 0 5cm H20 0 1 cm H20 Mass flow sensor Leak 0 99 liters per minute 12 L min 1 liter per breath Respiratory rate 8 30 breaths per minute 10 1 breath per minute Tidal volume 0 4500 milliliters 10 1 milliliter Minute ventilation 0 40 liters per minute 10 0 1 liters per minute Target MV 0 40 liters per minute 1096 0 1 liters per minute Maximum low impedance flow at stated pressures Pressure cm H50 Flow L min 4 2 217 8 6 218 13 0 219 Read the relevant operating manual before using Glossary of symbols w Type CF equipment H AC switch ol Class II Drip proof IPX1 Technical Specifications English 35 36 Guidance and Manufacturer s Declaration Electromagnetic Emissions and Immunity Guidance and manufacturer s declaration electromagnetic emissions The VPAP Adapt SV is intended for use in the electromagnetic environment specified below The customer or the user of the VPAP Adapt SV should assure that it is used in such an en
12. before operation on DC and it is recommended that you contact ResMed s Technical Services division or see ResMed s website for information on your specific application CAUTION The VPAP Adapt SV uses 30V DC power Only connect the VPAP Adapt SV to DC power using the VPAP Adapt SV DC Converter P N 26928 specified for this application Supplemental Oxygen The VPAP Adapt SV is designed to be compatible with the use of low flow oxygen therapy provided from an external source and supplied at a fixed rate of up to 15 L min At a fixed flow rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on the pressure settings patient breathing pattern mask selection and the leak rate This warning applies to most types of bilevel machines WARNING If oxygen is used with this device the oxygen flow must be turned off when the device is not operating Explanation When the device is not in operation and the oxygen flow is left on oxygen delivered into the ventilator tubing may accumulate within the device enclosure and create a risk of fire Always begin VPAP Adapt SV therapy before the oxygen supply is turned on Always turn the oxygen supply off before stopping VPAP Adapt SV therapy Oxygen supports combustion Oxygen should not be used while smoking or in the presence of an open flame VPAP Adapt SV System English 15 Using the VPAP Adapt SV WARNING J Before starting therapy with a ne
13. cated by the text above it in the LCD eg exit Cancels operations Extended hold exits to the top level of the menu OuickView key Takes you immediately to the Treatment screens with information on your current therapy See Treatment Screens on page 24 Alarm LEDs and Alarm mute The Alarm LEDs are lights that indicate that an alarm has been triggered Alarms can be muted by pressing the Alarm mute key once The Alarm LEDs will remain lit for as long as the alarm is being triggered See Alarms on page 25 Adjusting Settings By using the Left Right and Up Down keys you can adjust some aspects of your treatment The settings can be seen in the LCD display To change a setting press the Up Down key until you arrive at the menu you require Press the Left key to enter the menu then use the Up Down key again to scroll through the options Once you find the option you reguire press the Left key to select the function that is displayed in the bottom left of the screen eg change yes Use the Up Down key to alter the parameters and press the Left key when you have the setting you require Press the Right key to exit from the menu Example You wish to change the setting for the mask you are using VPAP Adapt SV menu SETTINGS enter 1 MASK FULL FACE change exit LEARN CIRCUIT yes 1 exit SMARTSTOP OFF change 1 exit SOUND LEVEL LO change f exit RESULTS
14. eck tube connec ions LCD turns off Power fail alarm Turn off VPAP Adap SV at power switch HIGH PRESSURE Turn power off High pressure alarm Turn off VPAP Adap SV at power switch HIGH AVG PRESS High pressure alarm Turn off VPAP Adap SV at power switch HIGH LEAK Check circuit High mask leak alert Adjust the mask LOW PS Check circuit Insufficient pressure Support alarm Check tubing Run LEARN CIRCUIT FLOW BLOCKED Turn power off Flow blocked alarm Turn off VPAP Adapt SV Check air circuit Turn on VPA P Adapt SV Triggered by Low Mask Pressure Alarm pressure sensor line is blocked or disconnected air pressure at the mask has fallen below a set level mask is removed and SmartStop has been disabled If a low pressure alarm sounds 1 Check that the air tube and sensor tube are connected properly 2 Turn the VPAP Adapt SV off and on again at the power switch If the alarm persists return the unit to ResMed for servicing Alarms 25 26 Will clear when the low pressure condition is fixed treatment is stopped by pressing the Start Stop key Power Fail Alarm Triggered by power failure machine is disconnected or switched off while delivering treatment If a power fail alarm sounds the flow generator stops delivering air pressure Will stop when the Alarm Mute key is pressed or after two minutes or
15. er cords to the VPAP Adapt SV at the same time Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord WARNING Make sure the power cord and plug are in good condition and the eguipment is not damaged 2 Connect air tubing Connect the air tube firmly onto the air outlet at the front of the unit Air tubing connector Pressure sensor tube connector Figure 4 Front view of VPAP Adapt SV To attach the pressure sensor tube to the unit screw the Luer connector onto the socket on the side of the unit Figure 5 Twisting the Luer lock connectors together WARNING Use care when handling the air tubing Be particularly careful when attaching the Luer connector to the VPAP Adapt SV If you kink or excessively twist the sensor tube it may affect the protection offered by the mask pressure alarm system Attach the mask to the air tubing The mask system should already be assembled If it is not please refer to the user instructions supplied with the mask Connect the mask system to the free end of the air tubing Proximal cuff Te J Figure 6 Assembling the VPAP Adapt SV system VPAP Adapt SV System 9 English Humidifiers A humidifier may be required if you are experiencing dryness of the nose throat or mouth The following ResMed humidifiers are compatible with the VPAP Adapt SV
16. essure mode the clinician sets a pressure that will be constant throughout treatment In ASV mode the VPAP Adapt SV continually measures your breathing and regulates the air pressure accordingly Mask Select the type of mask that you will be using you will only need to adjust this if you have more than one mask The following table shows the setting that should be selected for each mask type If your mask type is not mentioned below please contact your clinician Not all mask types can be used with th VPAP Adapt SV English o Option in Settings menu Use setting with Vista Vista Ultra Ultra Mirage Full Face Mirage Full Face Mask Series II Ultra Mirage Full Face Mask Mirage Quattro Full Face Mask Activa Activa Selecting mask options in the Settings menu Learn Circuit The Air Circuit Learn function takes approximately 20 seconds to run It allows the unit to learn what pressure it needs to build in to its calculations based on the components in the system Select LEARN CIRCUIT every time you add or remove a component eg mask humidifier or move to a significantly different altitude To run this function 1 Set up the entire VPAP Adapt SV system flow generator mask humidifier etc as you wish to use it 2 Make sure that the mask is unobstructed so air can flow from the mask to the flow generator 3 Turn on the VPAP Adapt SV at the power switch 4 Move through t
17. g in an unusual manner Severe mask leak or mouth Correct your leak condition leak greater than 30 L min Alarm stays on continuously Blockage in air circuit Ensure there is no barrier to the passage of air Internal malfunction Return the device to your equipment supplier for service Troubleshooting 31 32 Problem No LCD display Possible cause Power not connected or switch at back is not on Solution Ensure the power cable is connected and that the switch at the back of the unit is in the on position FAULT message displayed on LCD Internal malfunction Return the device to your equipment supplier for service System Error 7 message Device poweron has been interrupted by breathing into the mask during warm up Turn the flow generator off and on again Do not place mask on face during warm up 30 seconds Select LEARN CIRCUIT page 21 If the error is not cleared contact your equipment supplier for service Return the device to your equipment supplier for service Internal malfunction Glossary Alarms When more than one alarm is activated at the one time the highest priority alarm will be displayed Backlight The backlight turns off if no button has been pressed for two minutes It can also be kept permanently on by selecting On in the Backlight screen of the Options menu bpm breaths per minute L min liters per minute Mean The average value of the moni
18. g into it The SmartStart function should trigger treatment to commence within the first couple of breaths If not press the Start Stop key once to commence treatment page 16 Note You can observe treatment progress eg mask leak via the Treatment screens page 24 Quick Setup Guide 5 English 7 Remove your mask to stop treatment If SmartStop is enabled treatment will stop automatically If the SmartStop function has been disabled press the Start Stop key once VPAP Adapt SV System VPAP Adapt SV Components The VPAP Adapt SV flow generator is supplied with air tubing and a power cord Masks are supplied separately as you and your clinician have to decide which mask is best for you Some people experience dryness of the nose mouth and throat while using the VPAP Adapt SV system especially during winter Others may experience sneezing and or a runny or blocked nose during the first few weeks of VPAP treatment In many cases these symptoms of nasal irritation can be resolved with a humidifier See page 10 to connect a humidifier to the VPAP Adapt SV LCD screen Control panel Carry handle Removable front cover Pressure sensor tube connector Air tubing connector Figure 1 VPAP Adapt SV flow generator The control panel allows you to adjust settings on the flow generator and the LCD screen displays information about your treatment See Control Panel Functions on page 17 ResLink attachme
19. gen supplemental 15 P power cord connection 8 pressure sensor tube connecting 9 proximal cuff 7 Q quick setup guide 5 R Ramp 21 22 ResLink 14 Results menu 22 S Settings menu 20 Smart data 23 SmartStop 22 Index 41 42 standby mode 16 starting treatment 16 stopping treatment 16 Supplemental oxygen 15 System Error 7 message 32 T Treatment screens 24 U user owner responsibility 3 V VPAP Adapt SV usage 23 VPAP Adapt SV Components 6
20. he menus until you reach LEARN CIRCUIT and select yes by pressing the Left key 5 A message will remind you to check that the airflow from the mask is clear Select start with the Left key 6 A progress screen is displayed while the Air Circuit Learn function is running LEARNING CIRCUIT I Using the VPAP Adapt SV 21 7 When the function is complete a screen will appear saying that the circuit has been learned successfully The next menu item is then displayed Commence treatment as described in Starting Treatment on page 16 If there is a problem with the circuit a screen saying Invalid circuit will be displayed instead of the progress screen Check that all the connections between the pieces of equipment are secure You may need to return to the previous chapter of this manual to check that you have set up the system correctly SmartStop Turn the SmartStop function on or off When SmartStop is on the VPAP Adapt SV will automatically stop therapy when the mask is removed Note The unit will automatically SmartStart whenever you start to breathe into the mask Leak Alert If your clinician has enabled Leak Alert you will hear an alert tone sound when you have a high leak from your mask You can then adjust the mask on your face to remove the leak If Leak Alert is turned off an extra leak alert message will be displayed at the end of a session if high leak occurred The message will be displayed un
21. hetics f oxygen is used with the VPAP Adapt SV the oxygen flow should be stopped when the device is not operating If oxygen flow continues when the device is not operating oxygen may accumulate within the device and create a risk of fire Do not use the VPAP Adapt SV if there are obvious external defects or unexplained changes in performance Do not open the VPAP Adapt SV case There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorized service agent CAUTIONS You should report unusual chest pain severe headache or increased breathlessness to your physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during a course of therapy with the VPAP Adapt SV drying of the nose mouth or throat bloating ear or sinus discomfort eye irritation skin rashes chest discomfort Ouick Setup Guide 1 a A WN Attach the air tubing and pressure sensor tube to the VPAP Adapt SV and mask page 9 TH JH TA Attach a humidifier if required page 10 Switch on the VPAP Adapt SV Make sure the mask is unblocked Select the mask setting page 21 Perform the Air Circuit Learn function page 21 at first time use or if you are adding or removing a component of the system eg a humidifier or anti bacterial filter Fit the mask and commence breathin
22. idifier docking station and water chamber No other accessories are required for use Please refer to the HumidAire 2i User s Manual for details Once you have performed the Air Circuit Learn function the VPAP Adapt SV is ready for use See Starting Treatment on page 16 HumidAire 2i VPAP Adapt SV docking station HumidAire 2i water chamber Figure 8 HumidAire 2i attached to the VPAP Adapt SV Q mO EH E Figure 9 HumidAire 2i attached to VPAP Adapt SV and mask WARNING Make sure that the water chamber is empty and thoroughly dried before transporting the HumidAire 2i When using the carry bag always separate the VPAP Adapt SV unit and the Humidaire 2i before placing in their respective compartments VPAP Adapt SV System 11 English HumidAire You will need a medium size 21 52 cm air tube to connect the VPAP Adapt SV unit to the HumidAire see Figure 7 1 Make sure both the HumidAire and the VPAP Adapt SV are turned off Fill the HumidAire with water as described in the humidifier manual Place the filled water chamber inside the HumidAire 2 Gently detach enough of the tube clips to release the end of the pressure sensor tube from the VPAP Adapt SV air tubing Connect the VPAP Adapt SV air tubing to the left connector port on the humidifier and the medium 21 52 cm air tubing to the right connector port Close the Humid Aire lid G 3 Place the VPAP Adapt SV
23. n the Settings menu Insufficient Pressure Support Alarm Triggered by air pressure at the mask has failed to reach an expected level for 3 breaths the Air Circuit Learn function page 21 has not been run after adding a new component eg new mask humidifier to the system If an insufficient pressure support alarm sounds check that the pressure sensor tubing is not kinked check that the pressure sensor tubing is properly connected run LEARN CIRCUIT Will stop when the pressure sensor tubing is unobstructed and properly connected to the flow generator and mask treatment is stopped Start Stop key or SmartStop the VPAP Adapt SV is turned off Flow Blocked Alarm Triggered by blockage in air circuit If a flow blocked alarm sounds The treatment will stop Turn power off Check whether there is a blockage in the air circuit Remove blockage Turn power back on If the flow blocked alarm activates repeatedly discontinue use and return to ResMed for servicing If the alarm does not recur then continue to use as normal aap WN gt Alarms 27 English 28 Will stop when blockage is removed Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this manual CAUTION Do not wash the pressure sensor tube If fluid enters the pressure sensor tube allow it to dry completely by hanging it in a clean place out of direct sunlight
24. nd pressure level 30 dBA tested in accordance with the requirements of ISO 17510 1 Battery type Lithium Technical Specifications 33 34 Alarm specifications High maximum mask pressure alarm Stops air delivery at 30 cm H5O gt 700 msec High mean mask pressure alarm Stops air delivery when mean mask pressure exceeds 22 cm H320 averaged over approx 1 minute Low mask pressure alarm Mask pressure is less than 3 cm H20 for longer than 5 seconds while the patient is using the VPAP Adapt SV Insufficient pressure support alarm Is triggered if for 3 breaths in a row the maximum mask pressure achieved at inspiratory to expiratory transition is either 87596 of the absolute mask pressure targeted at that point or absolute mask pressure targeted minus 1 cm H20 whichever is the lower High leak alarm Leak exceeds 30 L min for more than 20 seconds Flow blocked alarm Flow is less than 6 L min for more than 20 seconds IEC 601 1 classifications Class ll Type CF Class ll medical electrical equipment provides protection against electric shock either by double insulation or reinforced insulation and does not require a protective earth Inspiratory Trigger Characteristics The inspiratory trigger occurs at a flow greater than 0 LPM when tested with EPAP 5 cm H20 IPAP 15 cm H50 sinisoidal breath shape The inspiratory trigger response time is 600 ms when tested with EPAP 5 cm H30 IPAP 15 cm H20 0 67 L sec inspir
25. not apply to you ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from region to region For further information on your warranty rights contact your local ResMed dealer or ResMed office R001 307 2 05 06 Index A air circuit learn function 21 air filter 30 hypoallergenic 30 air pressure monitoring 23 air tubing attaching to mask 9 cleaning 28 connecting 9 air tubing system 6 7 9 alarm flow blocked 27 high leak 27 high pressure 26 low mask pressure 25 power fail 26 alarm LED 18 anti bacterial filter 14 B backlight change setting 23 cC cautions 4 contraindications 3 CPAP mode 21 D DC converter 15 F fault message 31 H HumidAire connection 12 HumidAire 2i 11 humidifier connecting 10 types 10 K key alarm mute 18 25 left 18 right 18 start stop 17 up down 17 English L language change setting 23 LCD display 31 leak alert 22 learn circuit 21 Luer connector 9 M mask connection 9 indication of fit 23 recommended 7 select from menu 21 medical information 3 menus how to use 19 navigation 20 O Options menu 23 oxy
26. ns may cause hardening and reduce the life of the product Do not wash or dry the mask frame at a temperature above 176 F 80 C Exposure to higher temperatures may reduce the life of the product Do not hang the air tubing and pressure sensor tube in direct sunlight as the tubing may harden over time and eventually crack Periodically 1 The mask and air tubing are subject to normal wear and tear Inspect them regularly for damage Remove the power cord Clean the exterior of the flow generator with a damp cloth and mild liguid soap Inspect the air filter to check if it is blocked by dirt or contains holes See full instructions below WARNING Beware of electric shock Do not immerse the flow generator or power cord in water Always unplug the flow generator before cleaning and be sure that it is dry before reconnecting CAUTION Do not attempt to open the VPAP Adapt SV There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorized service agent Cleaning and Maintenance 29 30 Air Filter Inspect the air filter every month and check if it is blocked by dirt or contains holes Air filter cover The filter should be replaced every six months or more often if the flow generator is operating in a dusty environment WARNING Do not wash the air filter once it has become soiled The air filter is not washable or reusable Fitting the filter
27. nt point Air filter Power cord Figure 2 Rear of VPAP Adapt SV See page 14 to connect a ResLink See page 30 to replace the air filter Proximal cuff Tube clip Pressure sensor tube Luer connector BAIE 77 ibm WI mim mim po NM Connect this end Air tubing Connect this end to to the mask the flow generator Figure 3 Air tubing fully assembled All necessary air tubing components are packed with the VPAP Adapt SV a6 6 2 m length of air tubing a pressure sensor tube to measure pressure at the mask with Luer connectors e tube clips to hold the pressure sensor tube onto the air tubing a proximal cuff a special connector that attaches the pressure sensor tube into the main air tubing so you only need to make one attachment to the mask The air tubing is supplied with the proximal cuff connected Attach the tube clips evenly along the air tubing Press the pressure sensor tube into the tube clips and screw the Luer connectors together at the proximal cuff Make sure the pressure sensor tube is not kinked obstructed or twisted Note ResMed recommends the following tubing products for use with the VPAP Adapt SV replacement air hose only PN14986 replacement air delivery system hose sensor line clips proximal cuff PN26923 Masks You will also need a ResMed mask system supplied separately For the latest available masks see www resmed com Please refer
28. on top of the HumidAire Never place the VPAP Adapt SV unit underneath the humidifier This is to avoid water spilling into the unit 4 Connect the free end of the medium air tubing to the VPAP Adapt SV Connect the pressure sensor tube by twisting the Luer lock onto the VPAP Adapt SV see page 9 12 English 5 Connect the mask system to the free end of the long air tubing The final assembly should look like this 6 Switch on the HumidAire and the VPAP Adapt SV WARNING Make sure that the power cords and plugs are in good condition and the equipment is not damaged 7 Perform the Air Circuit Learn function see page 21 For information about filling and maintaining your humidifier refer to the HumidAire User s Manual The VPAP Adapt SV is now ready for use with the HumidAire See Starting Treatment on page 16 CAUTION Setup the HumidAire and the VPAP Adapt SV unit so that the tubing to the mask runs higher than the machines This will prevent condensation build up If condensation appears in the mask turn the humidifier settings down VPAP Adapt SV System 13 ResLink The ResLink is a device that records data onto a SmartMedia card during your therapy The card can be returned to your clinician so that they can observe your progress and adjust flow generator parameters if necessary The ResLink connects to the back of VPAP Adapt SV as shown For full de
29. recent cranial trauma or surgery chronic hypoventilation Below are general warnings and cautions Further specific warnings cautions and notes appear next to the relevant instructions in the manual WARNINGS The VPAP Adapt SV is NOT a life support ventilator The entire manual should be read before using the VPAP Adapt SV Advice contained in this manual should not supersede instructions given by the prescribing physician The VPAP Adapt SV should be used with masks and accessories recommended by ResMed or the prescribing physician Use of incorrect masks and accessories may adversely affect the function of the VPAP Adapt SV The VPAP Adapt SV is designed for use with masks that allow exhaled gases to be flushed out through vent holes Exhaled gases will be rebreathed if the mask is worn with the machine turned off or the vent holes are occluded If this occurs over prolonged periods suffocation may occur n the event of power failure or machine malfunction remove the mask The air flow for breathing produced by this device can be as much as 11 F 6C higher than the temperature of the room Caution should be exercised if the room temperature is warmer than 90 F 32 C The VPAP Adapt SV can be set to deliver pressures up to 25 cm H5O In the unlikely event of certain fault conditions pressures of up to 40 cm H50 for ess than one second are possible The VPAP Adapt SV is not suitable for use in the vicinity of flammable anest
30. s intended for use in an environment in which radiated RF disturbances are controlled The customer or the user of the VPAP Adapt SV can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the VPAP Adapt SV as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz power of transmitter Ww d 1 17 VP d 0 35 VP d 0 7 P 0 01 0 17 0 04 0 07 0 1 0 37 0 11 0 22 1 1 17 0 35 0 70 10 3 69 1 11 2 21 100 11 70 3 50 7 00 For transmitters rated at a din metres m can be de P is the maximum output power rating of the transmit manufacturer NOTE 1 At 80 MHz and 800 MHz the separation dis maximum output power not listed above the recommended separation distance ermined using the equation applicable to the frequency of the transmitter where tter in watts W according to the transmitter ance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Technical Specifications English 39 40 Limited Warranty ResMed warrants that your ResMed product shall be free from defects in materi
31. s or 25 cycles from an uninterruptible power source lt 5 U lt 12V gt 9596dip in Ut gt 95 dip in 240V or 5 sec or 5 sec Power freguency 3 A m 3 A m Power frequency magnetic 50 60 Hz fields should be at levels magnetic field characteristic of a typical IEC 61000 4 8 location in a typical commercial or hospital environment NOTE Ut is the AC mains voltage prior to application of the test level Continued next page Technical Specifications 37 38 Guidance and manufacturer s declaration electromagnetic immunity continued The VPAP Adapt SV is intended for use in the electromagnetic environment specified below The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment Immunity IEC60601 1 2 Compliance Electromagnetic environment test test level e level guidance Portable and mobile RF communications equipment should be used no closer to any part of the VPAP Adapt SV including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms d 1 17 NP IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 10 V m 10 V m d 0 35 VP 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 0 70 NP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d i
32. s the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol t2 OTE 1 At 80 MHz and 800 MHz the higher frequency range applies OTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and and mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted heoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which he VPAP Adapt SV is used exceeds the applicable RF compliance level above the VPAP Adapt SV should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the VPAP Adapt SV Over the frequency range 150 kHz to 80 MHz field strengths should be less than 10 V m Recommended separation distances between portable and mobile RF communications equipment and the VPAP Adapt SV The VPAP Adapt SV i
33. sage Scroll through the screens in the usual way using the Up Down key Treatment mode Pressure AS 5 0 15 0 INI exit Figure 16 Treatment Screen 1 Leak Respiratory rate LK 18L min RR 12 VT 428 Tidal volume Figure 17 Treatment Screen 2 Minute ventilation Target ventilation TARG 5 8 exit MV 6 2 Figure 18 Treatment Screen 3 Note TARG is not a relevant parameter in CPAP mode Alarms The VPAP Adapt SV unit is fitted with alarms to alert you to changes that will affect your treatment The most common reason for an alarm to sound is because the system has not been properly assembled Check that the air tubing English and pressure sensor tube have been properly attached to the flow generator and mask and humidifier if used You can mute an alarm by pressing the Alarm Mute key once Unless the alarm requires you to turn the power off you can press any key again to remove the alarm message from the LCD screen and to access all menus If the problem is still present the alarm will sound and display again after one minute The Alarm LED will remain lit for as long as the problem is present We recommend that you test the audible alarm once a week To do this hold down the Mute key while the VPAP Adapt is starting up If the alarm is functioning correctly it will beep twice LCD message LOW PRESSURE Check circuit Alarm type Low mask pressure alarm Action Ch
34. sibility The user or owner of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from operation which is not in accordance with the operating instructions supplied maintenance or modifications carried out unless in accordance with authorized instructions and by authorized persons Please read this manual carefully before use Definitions WARNING Alerts you to possible injury CAUTION Explains special measures for the safe and effective use of the device Medical Information Intended Use The VPAP Adapt SV system is intended to provide non invasive ventilatory support to treat adult patients with obstructive sleep apnea OSA and respiratory insufficiency caused by central and or mixed apneas and periodic breathing This device is intended for use in the hospital or home Contraindications The VPAP Adapt SV is not a life support ventilator and may stop operating with power failure or in the unlikely event of certain fault conditions Before using the VPAP Adapt SV tell your doctor if you have any of the following conditions hypotension or significant intravascular volume depletion acute sinusitis or otitis media epistaxis severe nosebleeds causing a risk of pulmonary aspiration conditions predisposing to a risk of vomiting into mask Introduction English impaired ability to clear secretions pneumothorax or pneumomediastinum
35. specified below The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment English Immunity test IEC60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD x6 kV contact x6 kV contact Floors should be wood concrete or ceramic tile If IEC 61000 4 2 8 kV air 8 kV air floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast 2 kV for power 2 kV Mains power quality should transient burst supply lines be that of a typical IEC 61000 4 4 commercial or hospital 1 kV for input Not Applicable environment output lines Surge 1 kV differential 1 kV differential Mains power quality should IEC 61000 4 5 mode mode be that of a typical commercial or hospital 2 kV common 2 kV common environment mode mode Voltage dips short lt 5 U 12V Mains power quality should interruptions and gt 95 dip in Ut gt 95 dip in 240V be that of a typical voltage variations on or 0 5 cycle or 0 5 cycle commercial or hospital power supply input environment lines 40 U 96V If the user of the IEC 61000 4 11 60 dip in Ut 6096 dip in 240V VPAP Adapt SV reguires or 5 cycles or 5 cycles continued operation during power mains interruptions 70 U 168V it is recommended that the 30 dip in Ut 30 dip in 240V VPAP Adapt SV be powered or 25 cycle
36. tails on using the ResLink please refer to the ResLink User s Manual Anti bacterial Filter Your clinician may recommend that you use an anti bacterial filter A filter should be placed in the air tubing between the mask and the flow generator If using a humidifier place the filter between it and the flow generator see Figure 10 below Typically the filter should be replaced daily Check the filter instructions for full details Whenever a filter is added removed or replaced you must run the Air Circuit Learn function see Learn Circuit on page 21 Note Only hydrophobic anti bacterial filters should be used with a Humid Aire 2i Anti bacterial filters Figure 10 Inserting anti bacterial filters Using DC Power to Run the VPAP Adapt SV The VPAP Adapt SV can be run on DC in case of power failure ResMed supplies separately a converter that allows a 12V or 24V battery to be connected to the DC input of the VPAP Adapt SV When using a DC power supply the power is turned on and off via the switch on the converter and not the power switch at the back of the VPAP Adapt SV The DC converter is a separate accessory that can be purchased through your ResMed distributor The humidifier heater will be automatically disabled when the VPAP Adapt SV is operating from the DC supply Power consumption of the VPAP Adapt SV varies with treatment settings Adequate battery type and sizing is required
37. til a key is pressed Ramp CPAP mode only Ramp time is the length of time taken for the air pressure to build up to the maximum pressure set by your clinician This setting allows you to increase or decrease the ramp time in 5 minute intervals If you find that you are still awake when the air pressure starts to build you may wish to increase the Ramp time Results Menu This menu has a series of screens showing information about your most recent treatment session If enabled by your clinician the following data may be viewed Mask fit Average Pressure Usage Mask fit Gives a rating of how well the mask was fitting during your last session on the English VPAP Adapt SV The stars indicate how good the fit was more stars better fit Star rating Definition EE Excellent EEEF Very good X xL Good Adjust mask Adjust mask HIGH LEAK Adjust mask Average Pressure Provides information on the average air pressure during your most recent session on the VPAP Adapt SV Usage Tells you how long the VPAP Adapt SV was in use during your most recent session Options Menu The Options menu allows you to change Smart Data Auto Appear Backlight Language Smart Data If the SMART DATA AUTO APPEAR option is set to ON the screens available under the Results menu will be displayed automatically for approximately 20 minutes after your session is completed After this time the data will still be av
38. to the table on page 21 to set up your mask correctly for use with the VPAP Adapt SV Your clinician can explain the features of the masks available and will discuss the mask that best suits your needs VPAP Adapt SV System English Assembling the VPAP Adapt SV System Complete the following steps to assemble the VPAP Adapt SV system Make sure the area around the unit is clean dust free and clear of bedding clothes and any other potential blockages To add a humidifier see page 10 WARNING Use only the air tubing system supplied with the VPAP Adapt SV system Connect the power cord There are two sockets at the rear of the flow generator the top one for an AC standard mains electricity and the lower one for a DC power cord see Using DC Power to Run the VPAP Adapt SV on page 15 ResMed recommends using the AC power cord supplied with the unit The customized ResMed power cord can be held in place with the locking clips To insert the locking clip pinch the free ends together and fit the pins into the holes on both sides of the socket Insert the power cord into the socket Push the locking clip down so that the groove holds the power cord in place Plug one end of the power cord into the power socket at the rear of the unit and the other end into a power outlet Power switch AC locking clip AC power cord Plug the free end of the power cord into a power outlet CAUTIONS Do not connect both AC and DC pow
39. tored parameter Minute ventilation A measure of the respiratory rate x tidal volume Proximal cuff A connector between the air tubing and the mask The pressure sensor tub ing attaches to the proximal cuff so that pressure at the mask can be detected and monitored SmartStart A feature that starts the VPAP Adapt SV automatically when you breathe into the mask SmartStop A feature that stops the VPAP Adapt SV automatically when you remove the mask You can turn SmartStop on or off in the Settings menu Tidal volume English The volume of air in liters breathed in or out in one breath Technical Specifications Dimensions H x W x D 5 6 x 9 8 x 11 5 142 mm x 248 mm x 293 mm Weight Flow generator with cover 8 14 lbs 3 7 kg Flow generator with HumidAire 2i empty 9 24 Ibs 4 2 kg Environmental conditions Operating temperature 41 F to 95 F 5 C to 35 C Storage and transport temperature 4 F to 140 F 20 C to 60 C Humidity operating storage and transportation 1096 9596 non condensing Housing construction Injection moulded plastic Air tubing 1 x 66 2m Air filter Synthetic fiber Patient connection port 22 mm taper compatible with ISO 5356 1 2004 Anaesthetic amp Respiratory Equipment Conical Connectors Power supply AC Input range 110 120V and 220 240V 50 60Hz 60VA DC Input range if using DC Converter 12 13 8V DC Performance Operating pressure range 4 25 cm H20 Sou
40. vironment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The V PAP Adapt SV uses RF energy only for its internal function There ore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic Emissions IEC 61000 3 2 Class A Voltage Fluctuations Flicker Emissions IEC 61000 3 3 Complies The V PAP Adapt SV is suitable for use in all establishments including dome direct stic establishments and those y connected to the public low voltage network that supplies buildings used for domestic purposes Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document Warnings The VPAP Adapt SV should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the VPAP Adapt SV should be observed to verify normal operation in the configuration in which it will be used The use of accessories eg Humidifiers other than those specified in this manual is not recommended They may result in increased emissions or immunity of the VPAP Adapt SV Guidance and manufacturer s declaration electromagnetic immunity The VPAP Adapt SV is intended for use in the electromagnetic environment
41. w mask select the correct mask type in the Settings menu page 21 When adding or removing a new component such as a mask humidifier or anti bacterial filter perform the Air Circuit Learn function page 21 Starting Treatment When the system is fully assembled press the power switch at the back of the VPAP Adapt SV to ON The Welcome screen will be displayed while the flow generator is warming up Allow the unit to warm up for approximately 30 seconds before placing the mask on your face Figure 11 Welcome screen Once the VPAP Adapt SV has warmed up breathing into the mask should start therapy immediately If not press the Start Stop key once to commence treatment The VPAP Adapt SV restarts in the mode in which it stopped Stopping Treatment If SmartStop is enabled treatment will stop automatically when you remove your mask If the SmartStop function has been disabled press the Start Stop key Standby Mode VPAP ADAPT SV Figure 12 Standby screen Once you have stopped treatment the VPAP Adapt SV will be in standby mode In standby you can press the power switch at the back to OFF to turn off the flow generator commence treatment again by breathing into the mask or pressing the Start Stop key enter any of the menus by pressing the Left key When in standby mode a gentle non therapeutic flow of air will still be present Note If the SMART DATA AUTO APPEAR option is set to ON the screens available

Download Pdf Manuals

image

Related Search

Related Contents

Ericsson RX8310 User's Manual  Notice de montage  User Manual  DEWW  MANUAL DO USUÁRIO DE REDE    CC-Link System Compact Type Remote I/O Module User`s Manual  

Copyright © All rights reserved.
Failed to retrieve file