1. For professional office blood pressure measurement criteria please refer to Chapter 8 e Your blood pressure monitor includes an irregular heartbeat feature Irregular heartbeats can influence the results of the measurement The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated If the measurement results are affected by irregular heartbeats but the result is valid the result is shown together with the irregular heartbeat icon If the irregular heartbeats cause the measurement to be invalid no result is shown If the Irregular heartbeat symbol is shown after you have taken a measurement repeat the measurement If the Irregular heartbeat symbol is shown frequently please make your doctor aware of it 26 3 Using the Unit What is Irregular Heartbeat An irregular heartbeat is a Normal Heartbeat heartbeat rhythm that varies Pulse A AAAA by more than 25 from the average heartbeat rhythm Blood pressure SIASNNA detected while the unit is Irregular Heartbeat measuring the systolic and Short Long diastolic blood pressure Pulse If such an irregular rhythm iS Blood pressure DSIN_SN detected more than twice during measurement the Irregular heartbeat symbol appears on the display with the measurement results What is Arrhythmia A heartbeat is stimulated by electrical signals that cause the heart to contract Arrhythmia is a co
2. SpotArm Type Blood Pressure Monitor Model i Q142 e Instruction Manual e Mode d emploi Gebrauchsanweisung e Manuale di istruzioni e Manual de instrucciones e Gebruiksaanwijzing e PYKOBOJ JCTBO MO SKCNJVATALIMM QD uli Cal I Gis AR All for Healthcare IM HEM 1040 E 06 05 2012 1665009 3F Contents Before using the unit Introduelioniiialaibleidalileaic 3 Important Safety Information ii 4 Te OVETVIEW aaae i a 7 2 PreparatioN suonissa a A a 10 2 1 Connecting the AC Adapter 10 2 2 Setting the Date and TIMA ii 11 Operating instructions 3 Using the Units alia 16 3 1 Howto Sit Correctly When Taking a Measurement 16 3 2 Place Your Arm Through the Arm Cuff sees 18 3 3 Taking a ReadiNdz alia 20 3 4 Instructions for Special Conditions ce eeeeeeeeeeeeeeenees 28 3 5 Using the Memory Function ccccccccecsessseeeeceseeseeeeeeeeeeees 29 Care and maintenance 4 Handling Errors and Problems 36 4 1 ErrorMessages a cilea 36 4 2 TIOUDIESNOOtUNG cnia aotearoa nda rennene 37 5 Maintenance and Storage rin 39 6 OptionalParis i aaa 41 7 TechnicalDataii aan idee 44 8 Some Useful Information about Blood Pressure 46 Introduction Thank you for purchasing the OMRON i Q142 Intellisense SpotA
3. arm cuff has adjusted to the size of your arm adjusts to the size of your arm Mi kg Tea Den w K S gt noo _ die Measurement ends Measurement displayed as the arm cuff deflates Inflates to the required pressure Note The unit may automatically increase inflation if the arm cuff inflation is insufficient Automatic re inflation occurs only once 22 3 Using the Unit To cancel a measurement If you press the O I or START STOP button while the cuff is inflating measurement stops and the arm cuff deflates Note For safety reasons measurement will stop if any button is pressed Except Arm cuff release button Important If inflation does not stop when the O I START STOP or any other button is pressed Press the Emergency deflation switch on the back of the unit 23 3 Using the Unit 6 Check the measurement results The measurement results are SYS automatically stored in the memory LI Refer to Chapter 3 5 o M Note If G Guest was selected LI mo the measurement results are LI not stored in the memory Pulse A A Warning Self diagnosis of measurement results and self treatment are dangerous Always follow the instructions of your doctor If the or A symbols are displayed Please repeat measurement In some cases it may not be possible to take measurement correctly e If 2 is displayed Incorrect posture d
4. moved etc do not place objects writing implements note books the AC adapter etc in the arm cuff This could cause damage to the cuff lining or the unit and prevent the unit from taking correct readings Do not inflate the arm cuff if your arm is not in place Do not take readings if your arm is placed incorrectly upside down or from the opposite side of the unit Do not move the unit unless the arm cuff has been secured Do not hold the unit by the Arm cuff 1 Overview Main Unit Rear view MT A Arm cuff release button I Elbow rest Press this to open as shown above J Arm cuff B Cuff lining K AC adapter jack C Memory button L Emergency deflation switch D Weekly average Morning Press this to stop measurement in an Evening button emergency E User ID selection button C F START STOP button G O I button Power switch H PC printer port 1 Overview Display lt lt HW aay re 900 lt lt am um Lx n N a Es 12888 DO _ Y WEEK A B G LIT ODU VWM M Systolic blood pressure Z Morning hypertension symbol N Diastolic blood pressure Indicates iene out of recommen O Pulse displa i P_Date Ti p P AA Morning average symbol ais ies ney Displayed when viewing morning Q Week display averages using the memory function R Deflation symbol AB Evening average symbol S Posture symbol Dis
5. not be sufficient for an accurate diagnosis Many factors such as physical activity anxiety or the time of day can influence your blood pressure Thus it is best to try and measure your blood pressure at the same time each day to get an accurate indication of any changes in blood pressure Blood pressure is typically low in the morning and increases from afternoon to evening It is lower in the summer and higher in the winter Blood pressure is measured in millimetres of mercury mmHg and measurements are written with the systolic pressure before the diastolic e g A blood pressure written as 140 90 is referred to as 140 over 90 mmHg 46 8 Some Useful Information about Blood Pressure 200 Upper curve systolic blood pressure Lower curve diastolic blood pressure 150 D I E E i N A pep 50 F T T T T 6 12 18 24 Time of day Example fluctuation within a day male 35 years old Classification of Blood Pressure by the World Health Organization The World Health Organization WHO and the International Society of Hypertension ISH developed the Blood Pressure Classification shown in this figure Systolic blood pressure mmHg A Severe hypertension 180 Moderate hypertension 160 Mild hypertension 140 Normal systolic value 196 Normal blood pressure 120 Optimal blood pressure Diastolic target value a blood pressure 80 85 90 100 110 mmHg This classification is based on t
6. software onto a PC To operate the software application please refer to the EN Installation Guide and use the help function in the software application Use only the authorized OMRON PC cable Use of any other cable may cause damage to the unit and will void the user warranty 41 6 Optional Parts 1 Open the PC printer port cover 2 Plug the PC printer port connector plug small connector into the PC printer port on the main unit The circle on the cable plug should face up 3 Plug the USB connector plug large connector into the personal computer 25 USB Connector 42 6 Optional Parts Using the optional printer Please read attached Instruction Manual before using the printer 1 Remove printer cable plug from the printer unit 2 Connect printer cable plug cord to the main unit The circle on the cable plug should face up 43 Note 7 Technical Data Product Description Model Display Measurement Method Measurement Range Accuracy Inflation Deflation Pressure Detection Memory Power Source Operating temperature Humidity Storage temperature Humidity Air Pressure Console Weight Outer Dimensions Arm Circumference Package Content Digital Automatic Blood Pressure Monitor OMRON i Q142 HEM 1040 E LCD Digital Display Oscillometric method Pressure 0 mmHg to 299 mmHg Pulse 40 to 180 min Pressure 3 mmH Pulse 5 of display reading Fuzzy log
7. the display appears as shown below Today s date Current time All items light up Standby display Press the User ID selection button C gt to select the desired user a MEMORY A B G DELDAH 3 Using the Unit The user alternates with each press of the User ID selection button Ca as shown below User A A Notes If G is selected the measurement results are not stored in the memory e If the START STOP button is pressed without selecting a user ID G is displayed and the measurement results are not stored in the memory Check that you are sitting correctly If you are not sitting correctly the Posture symbol gl appears on the display The Posture symbol disappears when you are in the correct posture Start measurement only after checking that you are sitting GE correctly Note Depending on the angle of the arm cuff the Posture symbol amp will appear on the display even if your arm is not in the arm cuff Angle of arm cuff O18 when sitting correctly A 21 3 Using the Unit 5 Press the START STOP button once to start measurement Do not move your arm and remain still until the entire measurement process is completed Flashes ina MEMORY A B G EDA v Display During Measure ment The arm cuff rotating fashion Inflation starts after the
8. Morning Blood Pressure 9 59 Evening Weekly Average This is the average for measurements taken during the evening 19 00 1 59 between Sunday and Saturday An average for each day is calculated for up to three measurements taken within 10 minutes of the last measurement of the evening Measurements of the evening Sunday Monday Tuesday Wednesday Thursday Friday Saturday 19 00 Within 10 minutes Up to 3 measurements Last measurement of Il the evening Evening Blood Pressure 1 59 n n 32 3 Using the Unit To Display Morning and Evening Averages 1 2 Press the O I button to turn the power on Press the User ID selection button C to select the user ID A or B Press the Morning average button or the Evening average button 3 MEMORY A B G The average for the current week THIS WEEK appears on the display The morning and evening average value of the same week can be displayed by pressing the Morning average button or the Evening average button It is important to check both morning and evening average of the same week ur aD DO Continue to press the Morning average button or the Evening average button to display the previous weeks The unit displays 1 WEEK for the previous week to 7 WEEK for the oldest set of averages Note If there are not enough measureme
9. and repeat the steps listed under Chapter 3 3 Cancel the measurement and contact your OMRON distributor or Customer Services Contact your OMRON distributor or Customer Services Refer to Chapter 8 Connect the AC adapter Refer to Chapter 2 1 Connect the AC adapter to the power supply Refer to Chapter 2 1 Select the correct user ID Refer to Chapter 3 3 3 4 Handling Errors and Problems 38 Turn the power off then repeat measurement If the problem persists contact your OMRON distributor or Customer Services as mentioned on the package Maintenance and Storage To protect your unit from damage please observe the following e Do not subject the unit to extreme temperatures humidity moisture or direct sunlight Do not carry out repairs of any kind yourself If a defect occurs consult the OMRON distributor or Customer Services as mentioned on the package or seek the advice of your surgical appliance stockist or pharmacist e The OMRON i Q142 is a precision measuring device It is important that the accuracy remains within the specifications It is recommended that you have the unit inspected every two years to ensure it is accurate and functioning correctly Consult the OMRON distributor or Customer Services as mentioned on the package for further details Do not subject the unit to strong shocks or vibrations for example dropping the unit on the floor Do not use volat
10. e Press the Morning average button once to decrease the setting by one year Hold the buttons down to increase decrease the digits rapidly Note The range for the year setting is 2007 to 2030 If the year reaches 2030 it will return to 2007 Press the Memory button to confirm the setting when the desired number appears on the display The year is set and the month digits flash on the display Press the Weekly average buttons to adjust the month setting e Press the Evening average button once to increase the setting by one month Press the Morning average button once to decrease the setting by one month Hold the buttons down to increase decrease the digits rapidly 2 Preparation Press the Memory button to confirm the setting when the desired number appears on the display The month is set and the day digits flash on the display pa October has been set as Press the Weekly average buttons to adjust the day setting e Press the Evening average button once to increase the setting by one day Press the Morning average button once to decrease the setting by one day Hold the buttons down to increase decrease the digits rapidly to confirm the setting when the desired number appears on the display Press the Memory button id CI The day is set and the hour digits flash on the display 13 2 Preparation 14 8 Set the hour and minutes in the same way as
11. e evening average is displayed regardless of the values for the evening average To Delete All Values Stored in the Memory The values stored in the memory are deleted by user ID You cannot partially delete values stored in the memory All values for the user you select will be deleted 1 Press the O I button to turn the power on 2 Press the Arm cuff release button and open the arm cuff 3 Press the User ID selection button C amp D to select the user ID A or B 4 While holding the Memory button press the START STOP button simultaneously for more than 2 seconds and all values will be deleted MEMORY A B G D GI GD Z v v START ER 35 4 Handling Errors and Problems 4 1 Error Messages Movement during measurement EE displayed in the systolic pressure area Are rolled up sleeves squeezing the arm E displayed in the systolic pressure E area Is air leaking from the arm cuff Movement during measurement Arm cuff has been inflated above 299 mmHg A malfunction has occurred Er displayed Note 36 Is the arm placed through the arm cuff correctly Check that the Heartbeat symbol is lit then keep still and repeat measurement Refer to Chapter 3 3 If the EE appears again use again refer to Chapter 3 4 Place your arm correctly Refer to Chapter 3 2 Remove clothing from your upper a
12. emory button The average value is shown and the Average symbol appears above the Memory symbol on the display Cee Dn LI no I w 4 Press the Memory button to display the most recent set of individual measurement values on the screen Note After the memory number is shortly indicated the date and time are alternately displayed with the measurement values m Kd Se Da I eg n E n se Do J The values are displayed from the most recent to the oldest by pressing the Memory button Press and hold the Memory button to display the values faster 5 Press the O I button to turn the unit off 3 Using the Unit Morning and Evening Weekly Averages The unit calculates and displays weekly averages for measurements taken in the morning and evening within 8 weeks for each user A and B Note The week begins Sunday at 4 00 31 3 Using the Unit About the Weekly Averages Morning Weekly Average This is the average for measurements taken during the morning 4 00 9 59 between Sunday and Saturday An average for each day is calculated for up to three measurements taken within 10 minutes of the first measurement of the morning Measurements of the morning Sunday Monday Tuesday Wednesday Thursday Friday Saturday 4 00 First measurement of the morning Up to 3 measurements Within 10 minutes
13. ff Remove the cushion sheet before using the unit for the first time Take readings with your bare arm or light clothing Note Remove thick clothing from your upper arm Do not take readings over thick clothes or rolled up sleeves 1 Press the Arm cuff release button to open the arm cuff Note For safety reasons the START STOP button doesn t work without opening the arm cuff The arm cuff can be moved within the range shown to the right 18 3 Using the Unit 2 Place your left arm through the arm cuff Notes Do not remove the cuff lining in the arm cuff e lf the cuff lining becomes detached contact your OMRON distributor or Customer Services as mentioned on the package 3 Place your elbow on the elbow rest Taking measurements on the right arm Refer to steps 2 and 3 above and place your right arm through the arm cuff as shown 4 Check that you are sitting correctly Refer to Chapter 3 1 19 3 Using the Unit 3 3 Taking a Reading Try to take readings at the same time each day within 1 hour after waking up is recommended This will make the results more useful 20 Note If your systolic blood pressure is known to be above 170 mmHg refer to Chapter 3 4 Sit comfortably and relax If you feel tense take a few deep breaths Press the O I button to turn the power on MEMORY A B G v v START STOP The unit is turned on and
14. he blood pressure values measured on people in a sitting position in outpatient departments of hospitals Note There is no universally accepted definition of hypotension However those having the systolic pressure below 100 mmHg are assumed as hypotensive 47 8 Some Useful Information about Blood Pressure Manufacturer OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPAN EU representative OMRON HEALTHCARE EUROPE B V Scorpius 33 2132 LR Hoofddorp THE NETHERLANDS REP www omron healthcare com Production facility fn COs ETD alian OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 ODG U K OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH John Deere Str 81a 68163 Mannheim GERMANY www omron medizintechnik de Subsidiary OMRON SANT FRANCE SAS 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Made in China 48
15. ic controlled by electric pump Automatic pressure release valve Capacitive pressure sensor 84 Measurements with date and time for each user A and B AC DC adapter Included Input 100 240V 50Hz 60Hz Output 6V 9 6W 10 C to 40 C 30 to 85 RH 20 C to 60 C 10 to 95 RH 700 1060 hPa Approximately 2600 g Approximately 286 1 mm x 294 w mm x 271 h mm 22 to 42 cm Instruction manual quick guide AC adapter guarantee card blood pressure pass and instruction manual for optional printer Subject to technical modification without prior notice e This OMRON product is produced under the strict quality system of OMRON Healthcare Co Ltd Japan The Core component for OMRON blood pressure monitors which is the Pressure Sensor is produced in Japan for assembly wee CE0197 This device fulfils the previsions of EC directive 93 42 EEC Medical Device Directive This blood pressure monitor is designed according to the European Standard EN1060 Non invasive sphygmomanometers Part 1 General Requirements and Part 3 Supplementary requirements for electromechanical blood pressure measuring systems 44 7 Technical Data Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible to electromagnetic interference from other devices Electromagnetic interference may resul
16. ile liquids to clean the unit THE UNIT AND THE CUFF SHOULD BE CLEANED WITH A SOFT DRY CLOTH To remove stubborn stains wipe the unit with a damp cloth moistened with mild detergent Do not splash water on the unit or allow liquids to get inside the unit e Do not remove the cuff lining in the arm cuff e Do not wash or moisten the cuff lining GE 39 5 Maintenance and Storage Storage Please observe the following storage conditions Do not subject the unit to extreme temperatures humidity moisture or direct sunlight Do not subject the unit to strong shocks or vibrations or leave it at an angle e Do not store the unit where it will be exposed to chemical or corrosive vapours Storage Precautions Store the unit with the arm cuff closed e When the unit is not in use or if it is moved etc do not place objects writing implements note books the AC adapter etc in the cuff This could cause damage to the cuff lining or the main unit and prevent the unit from taking correct readings 40 6 Optional Parts CD ROM PC cable Printer Blood pressure management software Order No 3094363 0 Order No 1098282 5 A386 Order No 4997582 6 Using the optional software Omron s blood pressure management software allows you to view process and print data about blood pressure measured by the OMRON i Q142 Intellisense Please see attached Installation Guide before installing
17. ious injury General Usage Always consult your doctor Self diagnosis of measurement results and self treatment are dangerous People with severe blood flow problems or blood disorders should consult a doctor before using the unit Cuff inflation can cause internal bleeding Stop using the unit immediately if medication spills on the cuff lining or it otherwise becomes contaminated Continued use could pose a risk of infection Contact your OMRON distributor or customer services in case you want to have the cuff lining replaced Do not take measurements if there is a wound on your arm such as cuts or scratches This could cause the wound to deteriorate or cause an infection through contamination of the cuff lining AC Adapter Usage e Never plug in or unplug the power cord from the electric outlet with wet hands A Caution Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property Important Safety Information General Usage Do not leave the unit unattended with infants or persons who cannot express their consent Do not use the unit for any purpose other than measuring blood pressure Do not disassemble the unit Do not inflate the arm cuff over 299 mmHg If the cuff does not stop inflating or some other abnormality occurs press the Emergency Deflation Switch immediately to p
18. it 2 Insert the power plug of the AC adapter into a power outlet AC adapter AC adapter jack AC adapter plug 10 2 Preparation 2 2 Setting the Date and Time Your blood pressure monitor automatically stores up to 84 individual measurement values with the date and time for each user To make use of the memory and average values functions Set the unit to the correct date and time before taking a measurement for the first time Set the year month day hours and minutes for the current date and time For details on changing the date and time setting refer to Adjusting the Date and Time Setting The Weekly average buttons are used to set the date and time Use the buttons as described below to adjust the date and time setting Evening average button Press this button to increase the value for the date time setting Morning average button Press this button to decrease the value for the date time setting 1 Press the O I button to turn the power on All items on the display will appear briefly then the year digit will flash _ MEMORY A B G ET v v O START STOP Note The illustration to the right shows the display when the unit is used for the first time 5 co 3 11 2 Preparation 12 2 Press the Weekly average buttons to adjust the year setting e Press the Evening average button once to increase the setting by one year
19. its working life To prevent possible harm to the environment or human health from uncontrolled waste disposal please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources Household users should contact either the retailer where they purchased this product or their local government office for details of where and how they can take this item for environmentally safe recycling Business users should contact their supplier and check the terms and conditions of the purchase contract This product should not be mixed with other commercial wastes for disposal This product does not contain any hazardous substances 45 8 Some Useful Information about Blood Pressure What is Blood Pressure Blood pressure is a measure of the force of blood flowing against the walls of the arteries Arterial blood pressure is constantly changing during the course of the heart s cycle The highest pressure in the cycle is called the Systolic Blood Pressure the lowest is the Diastolic Blood Pressure Both pressure readings the Systolic and Diastolic are necessary to enable a doctor to evaluate the status of a patient s blood pressure Why is it a Good Thing to measure Blood Pressure at Home Having your blood pressure measured by a doctor can cause anxiety which is itself a cause of high blood pressure As a variety of conditions affect blood pressure a single measurement may
20. ndition where the heartbeat rhythm is abnormal due to flaws in the bio electrical system that drives the heartbeat Typical symptoms are skipped heartbeats premature contraction an abnormally rapid tachycardia or slow bradycardia pulse This can be caused by heart disease aging physical predisposition stress lack of sleep fatigue etc Arrhythmia can only be diagnosed by a doctor through a special examination Whether the appearance of the Irregular heartbeat symbol O in the results indicates arrhythmia or not can only be determined by an examination and diagnosis by your doctor A Warning If the Irregular heartbeat symbol is shown frequently please make your doctor aware of it Conducting self diagnosis and treatment based on measurement results are dangerous Be sure to follow the instructions of your doctor 27 3 Using the Unit 3 4 Instructions for Special Conditions If your systolic pressure is known to be more than 170 mmHg or if you failed to measure blood pressure repeatedly perform the following steps 1 Place your arm through the arm cuff and press the O I button to turn the power on 2 Press the User ID selection button Ca to select your user ID 3 Press and hold the START STOP button until the cuff inflates 30 to 40 mmHg higher than your suspected systolic pressure Note You cannot inflate the cuff above 299 mmHg An error will be displayed if you try to inflate the cuff abo
21. ng a measurement Do not move or talk during measurement Remove tight fitting clothing Correct from your upper arm and any Posture thick clothing such as a sweater Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight Sit upright with your back straight Place the center of the arm cuff so that it is at the same height as your heart The gap between the Place your chair and the top of the elbow here table should be between 25cm to 30cm 16 3 Using the Unit Note You can take a measurement on either your left or right arm The blood pressure can differ between the right arm and the left arm and therefore also the measured blood pressure values can be different Omron recommends to always use the same arm for measurement If the values between the two arms differ substantially please check with your physician which arm to use for your measurement Incorrect Posture If you are leaning forwards this can put pressure on your stomach which could cause the measurement results to be incorrect Readings cannot be taken correctly in the following cases Arched back leaning forwards oo ae If the unit is too far Sitting cross legged away from your e Sitting on a sofa or at a low body table so that you tend to lean forward If the unit is ata level lower than your heart 17 3 Using the Unit 3 2 Place Your Arm Through the Arm Cu
22. nts taken in a week to calculate the average the display will appear as illustrated Press the O I button to turn the unit off 33 3 Using the Unit Notes on Display Combinations In addition to the Morning average and Evening average symbols the unit may also display the Morning hypertension symbol amp if the morning average for that week is above the home blood pressure guideline See Chapter 3 3 Taking a Reading Symbol Information for details Depending on your measurement results these may be displayed in the following combinations Ln II De CI Morning average Morning average from This week Morning average from 7 weeks with Morning average symbol ago with Morning average symbol blinking Heartbeat symbol 4 Morning hypertension symbol 0 I a Ll l m a w Evening average Evening average from This week Evening average from 7 weeks with Evening average symbol ago with Evening average 8 symbol x blinking Heartbeat symbol 4 Morning hypertension symbol A Average within home blood Average above home blood pressure guideline pressure guideline Morning hypertension 3 Using the Unit The Morning hypertension symbol amp appears if the weekly average for morning measurements is above 135 85 mmHg In this case the Morning hypertension symbol 4 is displayed when th
23. played when viewing evening Displayed when sitting incorrectly KG 4 an the ir function T Movement symbol ENGL ee ae Se Displayed if move your body during Displayed when viewing value for the fhe measiremeni last three measurements U Irregular heartbeat symbol AD Memory symbol Displayed when viewing values stored V User select symbol in the memory Indicates to select the user I D or guest W User ID symbol A or B Note If your systolic or diastolic y Indicates user A or user B as selected by the user to take a measurement or when using the memory function Guest symbol Heartbeat symbol 1 Flashes during measurement 2 If flashing after measurement completed or when viewing results stored in the memory indicates blood pressure out of recommended range pressure is outside the standard range above 135 85 mmHg the Heartbeat symbol 4 as well as the Morning hypertension symbol 4 will blink Please refer to Chapter 3 3 Taking a Reading Symbol Information 1 Overview Package Contents AE AC adapter Instruction manual Quick guide e Guarantee card Blood pressure pass Instruction Manual for Optional Printer Note The printer is optional This unit does not include the printer 2 Preparation 2 1 Connecting the AC Adapter A Caution Use only the AC adapter designed for this unit 1 Insert the plug of the AC adapter into the AC adapter jack on the back of the un
24. revent from internal bleeding or damage to peripheral nerves To inflate the cuff manually refer to Chapter 3 4 If the cuff is over inflated it can cause internal bleeding Be sure to remove watches and other jewelry rings bracelets etc that could damage the cuff lining during application Do not take readings if the cuff lining is damaged This could cause injury e Do not use a mobile phone near the unit This could cause a malfunction Do not allow objects to come into contact with the PC printer port of the unit when taking measurements AC Adapter Usage EN e Use only the original AC adapter designed for this unit Use of unsupported adapters may damage and or may be hazardous to the unit e Plug the AC adapter into the appropriate voltage outlet e Do not use the AC adapter if the unit or the power cord is damaged Turn off the power and unplug the power cord immediately General Precautions e Do not apply strong shocks and vibrations to or drop the unit e Do not take measurements after bathing drinking alcohol smoking exercising or eating Important Safety Information Read and follow the Important information regarding Electro Magnetic Compatibility EMC in the Technical Data Section Read and follow the Correct Disposal of This Product in the Technical Data Section when disposing of the device and any used accessories or optional parts e When the unit is not in use or if it is
25. rm Type Upper Arm Blood Pressure Monitor The OMRON i Q142 Intellisense is a fully automatic blood pressure monitor operating on the oscillometric principle It measures your blood pressure and pulse rate simply and quickly For comfortable controlled inflation without the need for pressure pre setting or re inflation the device uses its advanced Intellisense technology The adjustable cuff angle corrects the body posture which tends to be stooped The elbow rest stabilizes the position and keeps the arm at the most appropriate angle By simply placing the arm through the cuff it automatically wraps around your arm to the appropriate size and starts to measure your blood pressure and pulse rate The monitor stores measurement results for two people and features morning and evening averages EJ Omron s blood pressure management software and printer are optional The software allows you to view manage and print data about blood pressure measured by the OMRON i Q142 Intellisense Software is in English only li Please read this instruction manual thoroughly before using the unit For specific information about your own blood pressure CONSULT YOUR DOCTOR Important Safety Information Consult your doctor during pregnancy arrhythmia and arteriosclerosis Please read this section carefully before using the unit A Warning Indicates a potentially hazardous situation which if not avoided could result in death or ser
26. rm and place your arm through the cuff Refer to Chapter 3 2 Contact your OMRON distributor or Customer Services as mentioned on the package Check that the Heartbeat symbol 4 is lit then keep still and repeat measurement Do not inflate the arm cuff above 299 mmHg Refer to Chapter 3 4 Contact your OMRON distributor or Customer Services as mentioned on the package If you change your posture or move during measurement the Posture symbol 72 or Movement symbol A may be displayed during measurement 4 2 Troubleshooting 4 Handling Errors and Problems The reading is extremely high or low Cuff is not inflated Is your arm placed through the arm cuff correctly Talking during the measurement Are rolled up sleeves squeezing the arm Is the cuff lining in the arm cuff damaged Is air leaking from the arm cuff The unit operates normally and measurements appear to be taken but Measurements taken by a doctor are higher or lower e Measurement results are different each time No display appears when O I button is pressed Readings are not stored in the memory The AC adapter was disconnected from the unit The AC adapter was disconnected from the power supply G guest was selected or no user ID was selected Refer to Chapter 3 2 Keep quiet during measurement Carefully read and repeat the steps listed under Chapter 3 3 Carefully read
27. t in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to regulate the requirements for EMC Electro Magnetic Compatibility with the aim to prevent unsafe product situations the EN60601 1 2 standard has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufactured by OMRON Healthcare conforms to this EN60601 1 2 2001 standard for both immunity and emissions Nevertheless special precautions need to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in accordance with EN60601 1 2 2001 is available at OMRON Healthcare Europe at the address mentioned in this instruction manual Documentation is also available at www omron healthcare com Correct Disposal of This Product Waste Electrical amp Electronic Equipment This marking shown on the product or its literature indicates that it should not be disposed of with other household wastes at the end of
28. the date 10 09 settings ATEO I Press the Weekly average a AEG buttons to adjust the digits for the hour and minutes then press the Memory button to confirm the setting When the date and time settings have been set the display changes to that shown to the above Press the O I button to turn the power off Note You can also start taking readings immediately rather than turn the unit off 2 Preparation Adjusting the Date and Time Setting Note If you need to change the date and time for some reason adjust the date and time between 10 00 and 18 59 This will avoid problems with morning and evening weekly averages stored in the memory 1 Press the O I button to turn the power off 2 Turn the unit on then press and hold the Memory button 19 08 for three seconds or dales orta longer when the standby amp AGAGI display shown right appears Standby display When the arm cuff is closed 3 The year setting flashes on the display Refer to Chapter 2 2 to adjust the settings for the date and time 15 3 Using the Unit 3 1 Howto Sit Correctly When Taking a Measurement Correct posture during measurement is necessary to get accurate results Notes e Measurements should be taken in a quiet place and you should be in a relaxed seated position Make sure that the room is not too hot or too cold e Avoid eating drinking alcohol smoking or exercising for at least 30 minutes before taki
29. uring measurement Sit correctly and take the measurement again Refer to Chapter 3 3 4 Note The symbol will disappear in a few seconds but may appear again depending on the cuff angle at the time e If A is displayed Movement during measurement Keep still and repeat the measurement 24 3 Using the Unit If E or EE is displayed Measurement could not be completed successfully Refer to Chapter 4 1 mr Wait 2 3 minutes before taking another measurement Waiting between measurements allows the arteries to return to the condition prior to taking the blood pressure measurement Repeat the measurement after your arm is relaxed again Refer to Chapter 3 3 T Remove your arm and return the arm cuff to its original position The arm cuff will click into place 8 Press the O I button to turn the power off Note If you forget to turn the power off the unit will turn itself off automatically after five minutes 25 3 Using the Unit Symbol Information e If your systolic or diastolic pressure is outside the standard range the Heartbeat symbol 4 will blink when the measurement result is displayed Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home Systolic Blood Pressure Above 135 mmHg Diastolic Blood Pressure Above 85 mmHg This criteria is for home blood pressure measurement
30. ve 299 mmHg 4 Release the START STOP button when the cuff has been inflated to the desired pressure The cuff starts to deflate and measurement starts 5 Therestofthe procedure is the same as for normal measurement Refer to Chapter 3 3 Note Do not apply more pressure than necessary 28 3 Using the Unit 3 5 Using the Memory Function The unit is designed to store the blood pressure and the pulse rate in the memory for two people user A and user B every time a measurement is completed The unit automatically stores up to 84 sets of measurement values blood pressure and pulse rate for each user A and B When 84 sets of measurement values are stored the oldest record is deleted to save the most recent values The unit also stores 8 weeks of morning averages and 8 weeks of evening averages for each user A and B Average Function The unit calculates the average reading based on the three most recent sets of measurement values taken within 10 minutes of the most recent reading Note If two sets of measurement values are stored in the memory for the 10 minutes period the average is based on the two sets of measurement values If one set of measurement values is stored this is displayed as the average To Display the Measurement Values EN 1 Press the O I button to turn the power on 2 Press the User ID selection button C amp to select the user ID A or B 29 3 Using the Unit 30 3 Press the M