1. Digital Ear Thermometer Model Gentle Temp 520 Instruction Manual ALL for Healthcare pl IM MC 520 E 01 03 2013 9042329 5A Important Safety Information To assure the correct use of the product basic safety measures should always be followed including the precautions listed below A Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury Conducting self diagnosis based on the measurement results and or treatment can be dangerous Please follow the instructions of your doctor Self diagnosis may worsen the symptoms A high or prolonged fever requires medical attention especially for small children Please contact your doctor Please keep still during measurement Do not forcibly insert the probe in the ear If you feel discomfort such as a pain during the measurement stop using the unit immediately It may injure the external auditory canal Do not use this unit if suffering from ear disease such as otitis externa or otitis media It may worsen the condition Do not use this unit when the external auditory canal is wet such as after swimming or taking a bath It may injure the external auditory canal Do not use this unit without attaching a probe cover e Please ensure that the ear canal is clean and free of earwax If the probe cover becomes dirty with earwax or other substances replace it with a new one Do not use a probe cover after someone else has us
2. Niederlassung Consociata Empresa filial Dochteronderneming unuan Yan Kurulus fasih AS ul OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPAN OMRON HEALTHCARE EUROPE B V Scorpius 33 2132 LR Hoofddorp THE NETHERLANDS www omron healthcare com OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 ODG U K OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH Gottlieb Daimler Strasse 10 68165 Mannheim GERMANY www omron medizintechnik de OMRON SANTE FRANCE SAS 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Uniquement pour le march fran ais OMRON Service Apr s Vente N Vert 0 800 91 43 14 consommateurs omron sante fr www omron sante fr contact Made in China Fabriqu en Chine Hergestellt in China Prodotto in Cina Fabricado en China Geproduceerd in China CaenaHo B Kutae Cin de Uretilmistir gale
3. Note When the probe cover is not attached correctly the probe cover symbol gt will flash on the display and a measurement can t be taken E PI C 3 1 Taking a Reading Notes e Make sure the probe cover is attached correctly e It is recommended that you measure 3 times with the same ear If the 3 measurements are different select the EI highest temperature Press the ON MEM button All symbols appear on the display Xe Qh PE ES re kalk Ce Then the display shown at the right will appear with 2 beeps LILI Lt Insert the probe into the ear as far as it comfortably goes in the direction of the eardrum Notes e Gently pull the ear back to straighten the ear canal and position the probe into the ear so it is snug aiming towards the membrane of the eardrum to obtain an accurate reading e Holding the unit too long may cause a higher ambient temperature reading of the probe This could make the body temperature measurement lower than usual 3 Using the Unit Measuring the temperature of an infant Measuring on a Measuring on The earis too small lying baby a sitting baby to insert the probe Lightly support Lightly support the While slightly pulling the child s body child s body and the ear back cover the slightly pull the ear external auditory canal towards the back with the probe without forcibly trying to insert the probe 3 Using the Unit 3 Press the S
4. This may result in incorrect operation of the unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in accordance with EN60601 1 2 2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual Documentation is also available at www omron healthcare com Correct Disposal of This Product Waste Electrical amp Electronic Equipment This marking shown on the product or its literature indicates that it should not be disposed with other household wastes at the end of its working life To prevent possible harm to the environment or human health from uncontrolled waste disposal please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources Household users should contact either the retailer where they purchased this product or their local government office for details of where and how they can take this item for environmentally safe recycling Business users should contact their supplier and check the terms and conditions of the purchase contract This product should not be mixed with other commercial wastes for disposal Pe This product does not contain any hazardous substances Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries 6 S
5. Description Model Sensing Unit Temperature Display Measurement Accuracy Measurement Range Power Supply Power Consumption Battery Life Operating Environment Temp and Humidity Storage Environment Temp and Humidity Protection against electric shock Applied Part Weight 5 Technical Data Digital Ear Thermometer Gentle Temp 520 MC 520 E Thermopile 4 digit F display in 0 1 degree increments 3 digit C display in 0 1 degree increments 0 2 C 0 4 F within 35 5 C to 42 0 C 95 9 F to 107 6 F 0 3 C 0 5 F for other range 34 0 C 93 2 F to 42 2 C 108 0 F 3 0V DC 1 CR2032 Lithium Button Battery 0 015 W With a new battery approx 2 500 measurements or more Ambient environment 25 15 C 50 40 RH 10 C 50 F to 40 C 104 F 0 lt RH lt 85 20 C 4 F to 50 C 122 F 0 lt RH lt 85 Internally powered ME equipment type BF Approx 85g with battery installed 5 Technical Data Outer Dimensions 36 mm w x 161 mm h x 56 mm d Package Content Test Battery Lithium Button Battery CR2032 probe cap 21 probe covers connection ring instruction manual Options Probe cover exclusively for Gentle Temp 520 MC EP2 Notes e The specification may be changed without prior notice e This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co Ltd Japan e This device fulfills the provisions of the EC dir
6. TART button The measurement is complete in 1 second with a long beep The 7 symbol will flash for 5 seconds Note You can take another measurement after 2 beeps Make sure J symbol remains lit 4 Remove the unit from the ear and check the measurement Et PYT result IC Jo Note If your measurement result is over 37 5 C 99 5 F the buzzer will beeps three times after a long beep After each Ear Measurement the unit needs 5 seconds to be ready for next measurement During this 5 seconds waiting time the ear icon will be flashing 5 Press and hold the ON MEM button until OFF appears on the display to turn off the unit The unit automatically stores the measurement in its memory It will automatically turn off after 1 minute 3 2 1 2 10 3 Using the Unit Using the Memory Function This unit automatically stores the results up to 9 sets after each measurement Note If the memory is full the unit will delete the oldest reading Press the ON MEM button to turn on the unit Press the ON MEM button again TE The memory number appears on the display Release the ON MEM button pe The most recent result will appear on the display JE LI Press the ON MEM button repeatedly to view the older results ml Vand 6 nn Press and hold the ON MEM button until OFF appears on the display to turn off the unit It will automatically turn off after 1 minute 4 Troubleshooti
7. e contact the OMRON retail outlet or distributor for having the device checked e The probe is the most delicate part of the unit Use care when cleaning the Infrared sensor to avoid damage Do not store the unit in the following types of places Doing so may damage the unit Wet locations Locations with high heat and humidity or those that are exposed to direct sunlight Areas close to heating equipment dusty locations or environments where there are high salt concentrations in the air Locations where the unit will be subjected to leaning over falling shock or vibration Pharmaceutical storage areas or locations where corrosive gases are present 12 4 Troubleshooting and Maintenance 43 1 Replacing the Battery Battery CR2032 Lithium Button Battery Note To protect the environment discard the used batteries in accordance with the local regulations regarding waste disposal procedure Disposal can be done at your retail store or at appropriate collection sites EI Insert a pointed object into the battery cover pick hole Slide and remove the battery cover with your thumb Remove the battery with a pointed object Note Do not use metal tweezers or a screwdriver Insert the new battery under the metal hook on the left side and press the right side of the battery down until it clicks Note Replace the new battery with the plus side on the top metal hook Replace the battery cover 13 14 Product
8. ective 93 42 EEC Medical Device Directive and the European Standard EN12470 2003 Clinical thermometers Part 5 Performance of infra red ear thermometers with maximum device C 0197 15 16 5 Technical Data OMRON Digital Ear Thermometer Model Gentle Temp 520 MC 520 E Information for accompanying documents in the scope of IEC60601 1 2 2007 Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible to electromagnetic interference from other devices Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to regulate the requirements for EMC Electro Magnetic Compatibility with the aim to prevent unsafe product situations the EN60601 1 2 standard has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufactured by OMRON Healthcare conforms to this EN60601 1 2 2007 standard for both immunity and emissions Nevertheless special precautions need to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device
9. ed it This can lead to cross infections such as otitis externa Correct measurement result may not be obtained if dirty probe covers are used Proper installation of the probe cover ensures accurate measurements When the infrared sensor becomes dirty wipe it lightly with a soft dry cloth or a cotton swab Do not wipe the infrared sensor with tissue paper or a paper towel Do not use more than one probe cover at a time If there is any temperature difference between the places where the unit is stored and where you are going to measure leave the unit in the room where you are going to use it for more than 30 minutes to allow it to reach room temperature first then measure If the ear is cold wait until the ear is warmed up before taking a temperature measurement The measured result may indicate low when you use an ice bag or an ice pack or immediately after coming in from the outside in winter Do not touch the infrared sensor with a finger or breathe on it Important Safety Information Do not attempt measurements when the unit is wet as inaccurate readings may result Check the symbol on the display before and after the measurement so that the measurement is taken in the appropriate mode Keep the unit out of children s reach Avoid children trying to measure themselves or others as they may damage the ear In an emergency case if a child swallows a battery or a probe cover immediately consult with a doctor Do not throw batter
10. ies into a fire The battery may explode Remove the battery when the unit will not be used for 3 months or more Failure to do so may lead to fluid leakage heat generation or bursting resulting in damage to the unit Do not use the unit in places where strong static electricity or electromagnetic fields are present Doing so may lead to inaccurate readings and may contribute to instrument failure General Precautions Do not use this unit other than for measuring the temperature in the human ear Do not apply a strong shock to drop step on or vibrate the main unit Do not use a mobile phone near the unit The main unit is not waterproof Be careful when handling this unit so that no liquid alcohol water or hot water will get into the main unit When the unit is wet with vapor wait until it dries or wipe it lightly with a soft dry cloth Do not disassemble repair or modify the unit When you inform your doctor of your temperature make sure you state that you measured the temperature in the ear Main unit Infrared sensor Probe Probe cap Probe cover detector Display ON MEM button Battery cover pick hole Battery compartment 9 Q Connection ring Probe cover Display YE Q D ORE a LIL Buzzer Battery icon Memory icon Probe cover icon Ear temperature Temperature mode 2 Preparation 2 1 2 2 N Removing the Insulating Tape Pull the insulating tape out from the battery co
11. mpartment by catching hold of the outer part for the first time fr FO Switching between C and F This unit is set in C as default While the power is off press and hold the START button While holding it down press and hold the Ti F appears on the QO EN ON MEM e Notes e To select the C mode start from step1 e When the unit is switched between C and F all the readings stored in the memory are deleted 2 Preparation 2 3 Setting the Buzzer The buzzer is set on as default 1 Press the ON MEM button to turn on the unit EI 2 Press and hold the ON MEM button for 3 seconds n e 67 C The X symbol flashes on the display TT 3 Release the ON MEM button om e e The NG symbol remains lit and the buzzer is set to off Notes e If the ON MEM button is pressed down for more than 5 seconds after the X flashing the unit turns off without setting the buzzer e To turn the buzzer on start from step1 2 4 1 2 2 Preparation Attaching a Probe Cover Always use a new and undamaged OMRON MC EP2 probe cover Gently twist off the probe cap Note Do not forcibly remove the probe cap Place a new probe cover on the connection ring Place Adhesive Side Upward Probe cover GS Connection ring Note The adhesive side of probe cover should be upward Insert the probe into the probe cover on the connection ring until it clicks
12. ng and Maintenance 4 1 The Icons and Error Messages 29 as Device stabilization in process Wait until 7 stops flashing Ex Replace the battery tnt Battery is low Refer to section 4 3 ES h Probe cover is not attached correctly Attach the probe cover again until gt stops flashing Measurement before device stabilization Wait until 7 stops flashing Er oe Allow the thermometer to rest in a room for at least The device is showing a rapid ambient temperature change SOT Oar temperature 10 C and 40 C The ambient temperature is not within Allow the thermometer to rest in a room for at least the range between 10 C and 40 C 30 minutes at room temperature 10 C and 40 C 50 F 104 F 50 F 104 F eus Remove the battery wait for 1 minute and repower it If Far 500 en Sn Une toning the message reappears please contact the OMRON PAPE retail outlet or distributor for having the device checked 4 Troubleshooting and Maintenance Error Display Temperature taken is higher than 42 2 C Check the integrity of the probe cover and take a new 108 0 F temperature measurement Temperature taken is lower than 34 0 C Make sure the probe cover is clean and take a new 93 2 F temperature measurement Device can not be powered on to the ready Change to a new battery stage Refer to section 4 3 4 2 Maintenance e Please check the device if damaged after it is dropped If unsure pleas
13. ome Useful Information 6 1 Temperature measurements taken within the ear The Gentle Temp 520 Digital Ear Thermometer detects the infrared heat given off by the eardrum and surrounding tissues and it converts this heat into an equivalent ear temperature The Gentle Temp 520 is less threatening to a child than a rectal thermometer It s faster safer and easier to use EN than an oral thermometer Being digital there s no worry about the hazard of broken glass or mercury ingestion Measurements can even be taken while a child is sleeping For adults the Gentle Temp 520 Digital Ear Thermometer offers fast convenient and accurate readings without the delay of a conventional thermometer Clinical research has shown that the ear is an ideal site for taking body temperature The eardrum shares blood vessels with the hypothalamus the part of the brain that controls body temperature Therefore the ear is an accurate indicator of internal core body temperature An ear temperature unlike an oral temperature is unaffected by factors such as talking drinking and smoking 6 2 Normal and raised temperature We recommend that you practice with the Gentle Temp 520 on yourself and family members This way you can improve your technique and feel more confident of the measurements you take when a family member is ill You will also be able to tell when a reading is higher than normal 17 6 Some Useful Information As ambient temperature
14. sweat or saliva easily affects body temperature measured under the arm or the tongue readings may be lower than the core temperature Tympanic temperature measurement accurately reflects the brain temperature and can lead to a quicker detection of fever Infrared sensor Infrared rays Eardrum Measure point Eardrum External auditory canal In order to make a correct judgement of suspected fever conditions it is important to learn the normal temperatures of family members by measuring their temperatures when they are in good physical condition The temperature measured in the ear is different to that measured rectally Please use the normal as the standard for understanding the temperature difference during fever One speaks of normal body temperature if the measurement value lies within a certain range Body temperature varies however according to age Age Normal ear temperature in C and F Babies 36 4 C 37 5 C 97 5 F 99 5 F Children 36 1 C 37 5 C 97 F 99 5 F Teens Adults 35 9 C 37 5 C 96 6 F 99 5 F Elderly 35 8 C 37 5 C 96 4 F 99 5 F 18 6 Some Useful Information 6 3 Ear temperature compared to other types of body temperature The normal temperature varies according to different locations on the body EI 6 4 Questions and answers How many times can I measure consecutively You can measure consecutively up to three times The main unit will then be warmed up and may not be able
15. to measure correctly If you are going to measure more than three times wait for 10 minutes then measure again ne temperature indicated is rather high The probe cover may be faulty You may have used the thermometer that has been stored in a cool or cold place Measure the temperature after leaving the unit in the room where you are going to use it for more than 30 minutes If you store the unit in the room where you are going to measure the temperature you can promptly use the thermometer 19 20 6 Some Useful Information Is the temperature measured in the right ear different from that measured in the left ear Among healthy people there should be no significant difference in the measurement results Differences may be caused by the following reasons 1 The infrared sensor is not inserted in the same way 2 Measurement is not conducted by inserting the unit in a stable manner in the same angle Try to measure in the ear that consistently shows a higher measurement The temperature shown is rather low 1 2 3 4 5 The probe cover is dirty The infrared sensor is dirty You removed the unit from the ear before the measurement is finished The ear is cold The temperature tends to indicate low when you use an ice bag or an ice pack or immediately after coming in from the cold in winter The thermometer is not inserted deep enough in the ear Manufacturer EU representative EEG Subsidiary Succursale

Download Pdf Manuals

image

Related Search

Related Contents

Société pour la gestion du personnel S E P  Manual de Instruções    NTE INEN 2291: Accesibilidad de las personas con  Philips Forecast Capitola  Owners Manual - MIW Trade Water Coolers (UK)  Tobii Pro T60XL User Manual  取扱説明書(PDF)  Elegant Home Fashions 9HD664 Instructions / Assembly  クレジットシミュレータ/取扱説明書 (PDFファイル)  