Omron Healthcare HEM-711 User's Manual

image

Contents

1. Cuff Dimensions Cuff Circumference Fits arm circumference 9 to 13 Accessories Arm cuff carrying case instruction manual NOTE These specifications are subject to change without notice Optional accessories large cuff fits arms 13 to 17 in circumference and small cuff fits arms 7 to 9 in circumference are sold separately 19
2. Material damage refers to a wide range of damage involving your house household goods domestic animals and pets Examples of signs Matters involving actual prohibitions are indicated by text or pictures in or near O The The icon indicates prohibitions what you cannot do O left icon refers to prohibition to disassemble compulsory actions are indicated by text or pictures in or near The left icon refers to Q The icon indicates something compulsory always follow Matters involving actual general compulsion A Caution Self diagnosis of measured results and treatment are dangerous Please follow the instructions of your doctor Do not use the unit on infants or persons who cannot express one s intention Do not use the unit for any purpose other than measuring blood pressure Do not use a cellular phone near the unit Improper operation may result Do not disassemble repair or remodel the main unit or the arm cuff of the blood pressure monitor The battery liquid may leak and damage the main unit Please observe the following points When you are not going to use the unit for a long period of time approximately three months or more take out the batteries e Replace the worn batteries with new ones immediately e Do not use worn and new batteries together e Do not insert the batteries with their polarities in wrong direction Do not force to bend the arm cuff or the air
3. allow limitations on how long an implied warranty lasts so the above limitation may not apply to you This warranty gives you specific legal rights and you may have other rights which vary from state to state As a condition to operation of your warranty the enclosed registration card must be completed and sent to us within 10 days from the date of purchase FOR CUSTOMER SERVICE CALL TOLL FREE 1 800 634 4350 18 SPECIFICATIONS Model HEM 711 Display LCD Digital Display Measurement Range Pressure 0 to 280 mmHg Pulse 40 to 180 minute Pressure 3 mmHg or 2 of reading Accuracy Calibration Pulse 5 of reading Inflation Automatic by electric pump Deflation Active electronic control valve Rapid Pressure Release Active electronic control valve Pressure Detection Capacitive pressure sensor Measurement Method Oscillometric method Pulse Wave Detection Capacitive pressure sensor Power Source 4 AA batteries or AC adapter sold separately Battery Life Approximately 300 uses Operating Temperature Humidity 50 F to 104 F 10 C to 40 C 30 to 85 RH maximum Storage Temperature Humidity 4 F to 140 F 20 C to 60 C 10 to 95 RH maximum Approximately 12 3 oz 350g not including batteries Approximately 4 1 4 1 x 6 2 3 w x 2 3 5 h 109 mm x 170 mm x 66 mm Approximately 5 1 2 x 19 140 mm x 480 mm Main Unit Weight Main Unit Dimensions
4. brachial artery on the Green Marker inside of the arm Tube should run down center of arm approximately even with middle finger Cuff Tube HOW TO APPLY THE ARM CUFF 5 Pull the cuff so that the top and bottom edges are tightened evenly around your arm When the cuff is positioned correctly press the sewn hook material FIRMLY against the pile side of the cuff Make certain the cuff fits snugly around your arm The cuff should make good contact with your skin You should be able to fit one finger between the cuff and your arm Relax your arm and turn your palm upward Be sure there are no kinks in the air tubing 10 HOW TO TAKE A READING Press the ON OFF button a All display symbols appear for approximately one second b When the monitor becomes ready to measure the Heart Symbol appears on the display Press the START button and remain still As the cuff begins to inflate the monitor automatically determines your ideal inflation level Because this monitor detects the pulse even during inflation do not move your arm and remain still until the entire measurement completes Inflation stops automatically and measurement is started As the cuff slowly deflates decreasing numbers appear on the display and the Heart Symbol 4 flashes at every heartbeat In rare circumstances a higher inflation may be necessary In those cases the monitor reinflates the c
5. OMRON INSTRUCTION MANUAL Asii Automatic Blood Pressure Monitor with IntelliSense For Customer Service Call Toll Free 1 800 634 4350 Model H E M 71 1 Manufactured by OMRON Distributed by OMRON HEALTHCARE INC 300 Lakeview Parkway Vernon Hills Illinois 60061 www omron com ohi Copyright 2001 OMRON HEALTHCARE INC 741 INST TABLE OF CONTENTS igidgere 9 ei 6g e e er tee lt t ritcienree renter tr senna P pra ttree 4 Notes on Saleh cts ts nie ecient eh a rh Mint ese 5 POW YOU pTi e AAE A E E A E 6 Quick Reference Guide cccceeeseceeeeeeeeeeseeeeeeeaeeeeeeseneeeetennees 7 Battery Installation ReplaceMent ccc ccceeceeceeeeeeeeeeeeeeeeeteeeeeeeeeeea 8 How to Apply the Arm Cuff eeceeseeceeeeeeeeeeeeeeeeeeeenaeeeeeeeeeeeeees 9 10 HOW to Take a REAING ccecccccccccececeeeeceeeeeeeeesseseeseetseneaaaes 11 12 How to Use the Memory Function ccceeeeeeeeeeeeeeneeeeeeeeeeeees 13 How to Use the Optional AC adapter cccceceeeeeeeeeeteeeeeeeeeeeeees 14 TOUDIOSHOOUING i repie dete ls ele eran caretaker une a 15 CATON eureuchesvaatvasivenpiasase congeesodes sxarciecet ets a a a a 16 Care and Maintenance cccceeeeeeseneeeeeeeeeeeeeeeeeeeeeeeeeneeeeeeeees 17 Five Year Limited Warranty iis cesccetec cep cecteeee were yceervodcveiinouctavetersunaunes 18 SPECIIGANOMG essenin naknir as anian Naa Nna 19 INTRODUCTION Thank you for purch
6. arantee that interference will not occur in a particular installation If the product does cause harmful interference to radio or television reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the product and receiver e Connect the product into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio TV technician for help POTENTIAL FOR RADIO TELEVISION INTERFERENCE for Canada only This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference causing equipment standard entitled Digital Apparatus ICES 003 of the Canadian Department of Communications Cet appareil num rique respecte les limites de bruits radioel ctriques applicables aux appareils num riques de Classe B prescrites dans la norme sur le mat riel brouilleur Appareils Num riques ICES 003 dict e par le ministre des communications 16 CARE AND MAINTENANCE To protect your monitor from damage please AVOID the following e Subjecting your monitor and cuff to extreme temperatures humidity moisture and direct sunlight e Folding the cuff and tubing tightly e Inflating the monitor over 280 mmHg e Disassembling th
7. asing the Omron Automatic Blood Pressure Monitor with IntelliSense HEM 711 Although your blood pressure fluctuates throughout the day no self adjustment is necessary as the monitor intelligently adjusts to your changing conditions The IntelliSense monitor uses fuzzy logic intelligence to sense both your systolic and diastolic blood pressure values and the result is personalized inflation with every reading Arm discomfort and misreading due to improper cuff inflation are virtually eliminated Your new OMRON monitor uses the oscillometric method of blood pressure measurement This means the monitor detects your blood s movement through your brachial artery and converts the movements into a digital reading At the push of a button the monitor automatically fills the cuff with air starts deflation then displays your systolic diastolic and pulse readings on an extra large display panel Please read this instruction manual thoroughly before using this unit For specific information on your own blood pressure CONTACT YOUR PHYSICIAN NOTES ON SAFETY e The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the risk and the damage to you and others from happening e The icons and meanings are as follow Warning sign Contents Indicates matters in which bodily harm or material damage may arise as a result A Caution of incorrect handling
8. e and automatically deflates the cuff Wait 2 3 minutes before taking another blood pressure measurement You may require more rest time between readings depending on your individual physiological characteristics BATTERY INSTALLATION REPLACEMENT 1 Slide the battery cover off in the direction of the arrow 2 Install or replace 4 AA size batteries so the positive and negative polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover Notes If the Low Battery Indicator appears on the display replace all four batteries Long life alkaline batteries are recommended Remove the batteries if the monitor will not be used for an extended period of time Sit in a chair with your feet flat HOW TO APPLY THE ARM CUFF If the cuff is assembled correctly the hook material will be on the outside of the cuff loop and the metal D ring will not touch your skin If the cuff is not assembled pass the end of the cuff furthest from the tubing through the metal D ring to form a loop The smooth lle side Metorial cloth should be on the inside of Sewn Hook Material the cuff loop a on the floor and place your arm on a table so that the cuff is at the same level as your heart Put your arm through the cuff loop The bottom of the cuff should be approximately 1 2 above the elbow The Green Marker on the cuff should lie over the
9. e monitor e Subjecting the monitor to strong shocks for example dropping the monitor on the floor e Do not clean the monitor with volatile liquids THE MONITOR SHOULD BE CLEANED WITH A SOFT DRY CLOTH Do not clean cuff 17 FIVE YEAR LIMITED WARRANTY Your HEM 711 IntelliSense Automatic Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of five years under normal use The five year warranty excludes the monitor cuff The cuff is warranted for a one year period This warranty extends only to the original retail purchaser Should repair be needed within the warranty period ship the unit prepaid to Omron Healthcare Inc 300 Lakeview Parkway Vernon Hills IL 60061 Attn Repair Dept together with 5 00 for return shipping and insurance Be sure to include the model number of your unit and your phone number on any correspondence We will either repair or replace at our option free of charge any parts necessary to correct defects in the materials or workmanship The above warranty is complete and exclusive The warrantor expressly disclaims liability for incidental special or consequential damages of any nature Some states do not allow the exclusion or limitation of incidental or consequential damages so the above warranty may not apply to you Any implied warranties arising by the operation of law shall be limited in duration to the term of this warranty Some states do not
10. peat steps listed under How to Take a Reading If EE is displayed again take another measurement by pressing and holding the START button as instruction page 12 under How to Take a Reading instructions for special conditions Cuff over inflated more than 300 mmHg Turn the monitor off carefully read and repeat steps listed under How to Take a Reading Battery voltage is excessively low Replace all four AA batteries with new batteries Problem caused by abnormal memory function Contact OMRON Customer Service Toll Free 1 800 634 4350 If you continue to have difficulty obtaining a reading call OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1 800 634 4350 Please tell the representative you have an OMRON Model HEM 711 15 CAUTION CAUTION Changes or modifications not expressly approved by Omron Healthcare Inc could void the user s warranty NOTE POTENTIAL FOR RADIO TELEVISION INTERFERENCE for U S A only This product has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in a residential installation The product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no gu
11. surement push the ON OFF button The monitor will stop inflating and start deflating i z Value before Value of fourteen rapidly then the monitor will the last reading readings before power off a hopes vena To delete stored readings press the M button and the Start button at once then all the readings will be deleted NOTE You can not delete the stored readings partially RPE Press these two buttons Display of memory deletion at the same time 12 13 HOW TO USE THE OPTIONAL AC ADAPTER ACCESSORY ITEM SOLD SEPARATELY To avoid potential damage to the monitor use only the exclusive AC adapter Model HEM ADPT1 which can be purchased from Omron Healthcare Inc 300 Lakeview Parkway Vernon Hills IL 60061 Toll Free Phone 1 800 634 4350 To connect AC adapter 1 Insert the AC adapter plug Y Plug into the jack on the side of the pe monitor 2 Insert the AC adapter plug into a 120V AC outlet 50 60 cycles Option AC Adapter To remove the AC adapter To remove the AC adapter disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor s jack NOTE The monitor is designed not to draw power from the batteries when the AC adapter is in use 14 TROUBLESHOOTING Error Indicator Cause Correction Cuff under inflated Movement during measurement Turn monitor off wait 2 3 minutes and take another measurement Re
12. tube excessively When removing the air tube pull at the connector Do not apply strong shock or drop the main unit Do not inflate the arm cuff without it being wrapped on the arm KNOW YOUR UNIT Systolic Blood Heart Symbol Pressure Display ON OFF Button START Button a Memory Call Button Air Jack Diastolic Blood Pressure Display Pulse Display Air Plug Connects to the air jack Arm Cuff Air Tube Green Marker Carrying Case QUICK REFERENCE GUIDE 1 Avoid eating smoking and exercising for at least 30 minutes before taking a measurement Remove tight fitting clothing from your upper arm Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff is at the same level as your heart Put your arm through the cuff loop making sure that the bottom edge of the cuff is approximately one half inch above the elbow and that the Green Marker on the cuff is above brachial artery 1 2 inch Green Marker Pull the end of the cuff so that the entire cuff is evenly tightened around your arm and press the hook material firmly against the pile side of the cuff Press the ON OFF button After the Heart Symbol appears on the digital panel press the Start button and remain still until the measurement is complete When measurement is complete the monitor displays your blood pressure and pulse rat
13. uff up to 30 mmHg higher than initial inflation and restarts the measurement When the measurement is complete the arm cuff completely deflates and your blood pressure and pulse rate are displayed NOTE The monitor automatically stores blood pressure and pulse rate into the memory Press the ON OFF button to turn the monitor off NOTE If you forget to turn the monitor off it will automatically shut itself off after five minutes 11 pa Bat Gad Lad ae z lt u Nice ata oo at a gt any a dD Hs 535 Initial LCD Test Ready to Measure gt iis cus Inflation started HOW TO TAKE A READING HOW TO USE THE MEMORY FUNCTION Instructions for special conditions This monitor has a memory capable of storing fourteen sets of readings Every time you complete the measurement the monitor automatically e If your systolic pressure is known stores blood pressure and pulse rate to be more than 220 mmHg push and hold the START button e Push the button to recall stored readings while the until the monitor inflates 30 to 40 Heart Symbol is displayed mmHg higher than your suspected systolic pressure e When the memory is holding fourteen sets of readings the oldest set will be deleted to store a new set NOTE Do not apply more pressure than necessary 1 Last reading CN The monitor will not inflate above 300 mmHg e If you want to stop the mea

Download Pdf Manuals

image

Related Search

Related Contents

Digitaalinen satelliitti-TV-vastaanotin Digital satellit-TV  Etiquette - Plant Products  Prise en compte de la valeur professionnelle par réduction ou  Samsung SGH-Z300 Bruksanvisning    取扱説明書 AL リヤクッション ご使用前に必ずご確認ください  Page 1 Page 2 Page 3 Page 4 Page 5 SETUP WARM/OFF START  KEMEL エアシール AX 型 取扱説明書  