Sartorius IB 31000 P Operating instructions

image

Contents

1. Do not expose the scale to extreme moisture over long periods Moisture in the air can condense on the surfaces of a cold scale whenever it is brought to a substantially warmer area If you need to transfer the scale to a warmer area make sure to condition it for about 2 hours at room temperature leaving it unplugged Afterwards ifyou keep the scale connected to AC power the continuous positive difference in temperature between the inside of the scale and the outside will practically ruie out the effects of moisture condensation The components used in the scale are rated to at least Class KSF according to DIN 40040 Your Sartorius scale provides accurate readouts even when it is exposed to unfavorable ambient conditions You can adapt the scale to your requirements simply by changing the code settings in the scale operating menu For more information please read pages 21 through 23 How to Mount the Display Unit The display unit can be mounted as follows base mounted on the front of the scale factory mounted base mounted on the side of the scale see next page for direc tions mounted as a remote display unit see page 11 mounted on a support arm as a raised display unit YDH 01 IB is an Option see page 13 8 Mounting the Display Unit on the Short Side of Scale Base Leave the platform cover on the scale base To mount the display turn the scale upside down so that it rests on the platfo
2. L Stands for the list mode In this mode you can read a menu code setting but you cannot change it Changing a Program Menu Code Change Mode If you wish to change a program menu code you must first unlock the menu access switch to access the program menu To do so remove the protective cover located on the right of the dis play unit and slide the menu access switch 5 in the direction of the arrow The display will indicate C which Stands for the change mode meaning you can now change the menu code settings 21 After you have accessed the menu of the scale operating program the display will show a continuous numerical sequence from 0 to 5 for the page or the first digit of the code in addition to the Status code letter L or C When the first digit of the code you wish to check or change appears press the tare control 6 The page code number now stops in the display and a series of numbers for the 2nd digit or line will begin to cycle Press the tare control again to stop the code number of your choice in the display Next the numbers for the word last digit will cycle in the display Repeat the procedure to enter the last digit of the code The o symbol that appears indicates the actual setting To change any settings C mode press the tare control as soon as the appropriate numeric code is displayed sartorius Ec Ges ee L pav l o Brief display of BUSY and the o s
3. 30 The weight display shows the following special Status messages for your Information BUSY The scale processor is still busy processing a function and will not ac cept any other commands to perform functions at this time STANDBY The display has been turned off with the ON OFF key 8 and the scale is now in the ready to operate mode so that it does not require warmup POWER OFF The scale was disconnected from line power power failure or outage reconnection to line power after scale was unplugged CAL The calibration function has been activated Turning the Display On and Off Press the ON OFF key 8 to turn the display on or off You can also turn it on by pressing the tare control 6 sartorius When the scale is connected to the AC adapter and the power is on the display will go out whenever you turn off the power using the RET ON OFF key All other electronic circuits will remain energized indi cated by STANDBY This means the scale is immediately ready to operate without requiring warmup the next time you turn it on sartorius Self Test After the scale is turned on an automatic self test of the scale s elec tronic circuitry is performed This self test ends with the readout 0 0 g 0 000 kg if the factory code setting 5 1 1 5 1 2 is used Weighing Now place your sample or object on the platform 1 to determine the weight Read off the weight indicated in
4. shows H Load exceeds the capacity of the scale Unload the scale The weight readout changes constantly or the special message BUSY does not go out in the weight display Unstable ambient conditions Too much vibration or there is a draft Sample does not have a stable weight Setup scale in another area Access the menu to adapt scale to the particular type of weighing environment The weight display shows CE A zero readout was not indicated in the weight display when the CAL key 11 was pressed to calibrate The scale is loaded Press tare control 9 and re press the CAL key The code CC in the display does not go out The CAL key 7 has not been pressed when the display read zero The scale is loaded Press the tare control 6 and re press the CAL key Unload the scale The weight readout is obviously wrong Scale has not been calibrated Scale has not been tared before weighing The air bubble of the level indicator 2 is not within the circle only for IB 16000 S Calibrate scale Tare before weighing Level scale POWER OFF blinks in the display The operating voltage is too loo battery discharged Connect the scale or the external rechargeable battery to line voltage using the AC adapter 24 Care and Maintenance Cleaning Please do not use any aggressive cleaning agents solvents or simil
5. the display 9 as soon as the weight unit in this case g appears as the stability symbol In addition to grams kilograms this scale gives you a variety of other menu definable international weight unit options Select the weight unit you need from the table of the menu options for the scale operating program Set the appropriate code as described in the section Scale Operating Program Taring If you wish to weigh into a Container or if the weight display does not indicate 0 0 g 0 000 kg or the equivalent with the weight unit of your choice press the tare control to zero the display Auto Zero This scale has an automatic zero tracking function known as Auto Zero can be turned off by menu code see Scale Operating Pro gram Changes off zero x 0 5 of a digit per second are set to zero automati cally 19 20 sartorius Automatic Shutoff Make sure the Automatic Shutoff code is set to ON when using a non rechargeable or rechargeable battery This will increase the amount of time you can operate the scale before having to recharge the battery see Scale Operating Program If you do not press a key or if a weight readout in the display does not change for at least 2 minutes STANDBY will start flashing in the display The scale will shut itself off if it has not been operated after another 2 minutes Calibration This is only possible with an accurate calibration weight 10 kg See
6. 1 21 23 24 25 25 25 26 26 27 About the Product Warranty With this Sartorius Scale you have acquired a high quality electronic Instrument that will ease your daily work load Please read these installation and operating instructions carefully be fore operating your new scale Pursuant to the German Directive for the Implementation of Regulations for Prevention of Accidents Elektrische Anlagen und Betriebsmittel VBG 4 Electrical Installations and Equipment of April 1986 it is hereby certified that the equipment delivered Electronic Precision Scale series IB is manufactured and tested in compliance with the following DIN VDE regulations DIN IEC 348 VDE 0411 Safety requirements for electronic measuring apparatus DIN IEC 380 VDE 0806 Safety of electrically energized Office machines DIN IEC 601 VDE 0750 Safety of medical electrical equipment and with Article 10 of the Low Voltage Directive 73 23 EEC issued on February 19 1973 by the European Community When you use electrical equipment in installations and under ambient conditions requiring higher safety Standards you must comply with the provisions specified in the applicable regulations for installation in your country Do not miss out on the benefits of our full warranty Please complete the warranty card indicating the date of installation and return the card to your Sartorius dealer Unpack the scale Remove the plastic covering and adhesive strip
7. Accessories Unload the scale and press the CAL key 7 for at least 3 seconds un til the calibration weight readout appears in the display Center the calibration weight on the platform Now the weight unit symbol is displayed An acoustic signal indicates the end of the calibration procedure You can block access to the calibration function to find this menu code refer to the Scale Operating Program This function is acces sible when the scale operating program is unlocked using the menu access switch 5 sartorius Scale Operating Program The scale operating program lets you adapt your scale to various am bient conditions and to different weighing requirements and select weight units commonly used in your country At the factory we have set the codes for a Standard program which is protected by a locking function to prevent accidental changes The menu code contains the information of the operating program It consists of three digits known in computerese as the page 1st digit the line 2nd digit and the word 3rd digit Reading a Program Menu Code List Mode How to access the menu of the scale operating program With the display turned off STANDBY state hold down the tare con trol 6 and briefly press the ON OFF key 8 Upon completion of the automatic self test release the tare control as soon as CH5 is dis played The Status of the scale operating program is indicated in the weight display
8. CO 98648 000 60 Sartorius Industry IB 16000 S IB 31000 P IB 31 Electronic Precision Scales Installation and Operating Instructions sa rtorius 1 2 3 4 5 Platform cover for scale base Level indicator only for IB 16000S Power receptacle Leveling foot Menu access switch for scale operating program oO N o Tare control CAL key ON OFF key Weight display Manufacturer s label Contents About the Product Warranty Storage and Shipping Conditions Equipment Supplied Installations Instructions Ambient Conditions How to Mount the Display Unit Mounting the Display Unit on the Short Side of the Scale Base Mounting the Remote Display Unit Mounting the Display Unit on the YDH 01 B Support Arm Option Startup Connecting the scale to Line Power Voltage Selection Safety Precautions Connecting Electronic Devices Peripherals Leveling the Scale using the Level Indicator Operation Turning the Display On and Off Self Test Weighing Taring Auto Zero Automatic Shutoff Calibration Scale Operating Program Reading a Program Menu Code List Mode Changing a Program Menu Code Change Mode List of the Programmable Menu Code Settings Troubleshooting Guide Care and Maintenance Cleaning Safety Inspection Accessories Options Interfacing Devices with the Scale RS Interface Specifications Page 13 16 16 16 17 17 17 18 18 19 19 19 19 20 20 21 2
9. Insulation resistance gt 7 megohms measured with a constant voltage of at least 500 V at a500kohmload The duration and number of measurements should be determined by a qualified Sartorius service technician according to the particular ambient conditions and operational conditions of the AC adapter Such inspection should be done at least once a year 25 Accessories Options Data printer with YDP 02 0D date time and statistics functions Print speed approx lines sec 1 5 Printer housing W x D x H in mm 150 x 138 x 43 in inches 5 5x 54x17 Interface YDO 01 IB Data Input incl interface YDI 11 IB Dedicated applications kit for the YDI 11 IB YAK 11 1I 000 amp incl interface Replace amp with the letter code C for parts counting F for formulation A for animal weighing U for the PLUS Performance Package Support arm for raised display unit YDH 01 IB Rechargeable battery pack YRB 02Z external approx 10 hours of operation can be recharged by the AC adapter detailed information for powering the scale is available in our Service Information bulletin no 15 88 Adapter for hook up to a 12 Volt battery YCC 01 0007 Roller track conveyer plate YRT 01 IB 0001 Remote display can be connected to the interface port LCD reflective 737101A for overhead projectors transmissive 7371 02A Calibration weight 1 x 10 kg 7072 18 Dust cover 69 601310
10. Interfacing Devices with the Scale Please note that the interface port is electrically connected to the pro tective grounding conductor of the scale housing The cabling sup plied as accessory components is shielded and electrically connected on both ends to the cases of the connectors This electrical connec tion may result in interference caused by ground loops or by transient currents if you have grounded the housing or connected the protective grounding conductor for AC power If necessary connect an equipotential bonding conductor to the scale Specifications Model IB 16000 S IB 31000 P IB 31 Capacity kg 16 1 4 8 16 31 31 Readability g 0 1 0 1 0 2 0 5 1 1 Maximum overload capacity kg 75 75 75 Tare range by subtraction kg 16 1 31 31 Standard deviation g lt 0 05 0 1 0 1 0 25 1 lt 0 5 Max linearity deviation lt 0 2 lt 0 2 0 2 0 5 1 lt 1 Stabilization time typical S 2 Adaption to ambient conditions and application requirements by selection of one of four digital filter levels Display update rate S Stability range d 0 1 0 8 selectable 0 25 64 selectable Ambient temperature range 0 C 40 C 32 F 104 F Moisture proof rating according to DIN 40040 Class F non condensing Dust and water protection rating in compliance with DIN 40050 IEC 529 IP 65 scale NEMA 4 IP 20 AC adapter Sensitivity drift within 10 30 C l
11. ar agents Instead use a piece of cloth that has been wet with a mild detergent to clean the scale Make sure that no liquid enters the scale housing After cleaning wipe down the scale with a soft dry piece of cloth Safety Inspection If there is any indication that safe Operation of the scale with the AC adapter power supply is no longer warranted turn off the scale and disconnect the equipment from AC power immediately Lock the equipment in a secure place to ensure that it cannot be used for the time being In this case notify the Sartorius Service Center or International Tech nical Support Group based in Goettingen Germany Only service technicians who are authorized by Sartorius and have access to the required manuals are allowed to perform maintenance and repair work on the equipment Safe Operation of the scale with the AC adapter power supply is no longer ensured when there is visible damage to the AC adapter power supply the AC adapter power supply no longer functions properly the AC adapter power supply has been stored for relatively long periods under unfavourable conditions the AC adapter power supply has been exposed to rough handling during shipment We recommend that the scale along with the AC adapter power sup ply be inspected by a qualified Sartorius service technician according to the checklist given below Leakage current 0 05 mA measured by a properly calibrated multimeter
12. ed by the Sartorius la bel Use of AC adapters supplied by other manufacturers even if these adapters have a registered approval rating from a national test ing laboratory requires the consent of a certified Sartorius technician Plug the cord of the AC adapter into the power receptacle of the scale Now plug the AC adapter into a wall outlet Voltage Selection You can select the voltage only if you use our portable power supply 69 71172 that has a European type plug rounded prongs Safety Precautions The power supply or AC adapter rated to Class 2 double insulated can be plugged into a wall outlet without taking any additional safety precautions The negative pole ground of the Output voltage protective extra low voltage is connected to the scale housing which can be grounded for Operation The optional interface see Interfacing Devices on page 26 in addi tion is also electrically connected to the scale housing ground Connecting Electronic Devices Peripherals Unplug the AC adapter before you connect or disconnect devices to or from the interface ports Leveling the Scale using the Level Indicator only for model IB 16000 S At the point of use level the scale using the leveling feet 4 so that the air bubble is centered within the circle of the level indicator 2 17 Operation After you have plugged the scale into a wall outlet using the AC 0 adapter allow for at least 30 minutes warmup
13. rm cover Remove the screws from the base plate using the Allen wrench sup plied see arrows and then remove the base plate Unscrew the four screws of the angular cover plate and remove the angular cover plate Remove the fastening screws from the display unit using the Allen wrench to untighten them Remove the cable from the raceway channel Attach the display unit to the short side of the scale base by tightening the two screws Insert the cable in the raceway as shown in the diagram on the left Make sure that the rubber seal see arrow is not damaged Reposition the base plate and fasten it in place by tightening the three M3x8 screws see the large arrows and the two M3x10 screws see the small arrows Afterwards fasten the angular cover plate in place by tightening the SCrews Mounting the Remote Display Unit Leave the platform cover on the scale To mount the display turn the scale upside down so that it rests on the platform cover Remove the screws from the base plate see arrows using the Allen wrench supplied and then remove the base plate Unscrew the four screws of the angular cover plate and remove the angular cover plate Remove the fastening screws from the display unit using the Allen wrench to untighten them Remove the cable from the raceway channel 11 Attach the display unit to the mounting retainers supplied using the two M4x8 screws and washers can al
14. s Place the platform cover 1 on the scale base Storage and Shipping Conditions Allowable storage temperature 40 C 70 C 40 F 158 F The packaging of the scale has been designed to ensure that the scale will not be damaged even if it is dropped from a height of 80 cm about 32 inches After unpacking the scale please check it immediately for any visible damage as a result of rough handling during shipment If this is the case proceed as directed in the section entitled Safety Inspection Save all parts of the packaging and the box because you may need to ship your scale Before you pack your scale to ship it unplug all connected cables to prevent damage Do not expose the scale unnecessarily to extreme temperatures moisture shocks blows or vibrations ved 1 Equipment Supplied The equipment supplied includes the components shown on the left Scale with platform and display unit AG adapter Dust cover 3 Allen wrenches Retainers for mounting the display 2 M4 x8 screws and washers for mounting the Remote Display Unit Installation Instructions Ambient Conditions Unfavourable ambient conditions can influence weighing results Therefore please select a suitable place to set up your scale It should not be exposed to the following heat radiation drafts vibrations aggressive chemical atmospheres the scale must not be used in a hazardous location area
15. so be mounted on a wall for use as a wall console Thread the cable through one of the three openings in the scale hous ing Make sure that the rubber seal see small arrow is not damaged Reposition the base plate and fasten it in place by tightening the three M3x8 screws see the large arrows and the two M3x10 screws see the small arrows Afterwards fasten the angular cover plate in place by tightening the SCrews Mounting the Display Unit on the YDH 011B Support Arm Option Leave the platform cover on the scale To mount the display turn the scale upside down so that it rests on the platform cover Remove the screws from the base plate see arrows using the Allen wrench supplied and then remove the base plate Unscrew the four screws of the angular cover plate and remove the angular cover plate Remove the fastening screws from the display unit using the Allen wrench to untighten them 13 Remove the four screws from the display unit see arrows using the Allen wrench to untighten them Unscrew the two screws from the cable clamp Remove the clamp and then the display holder Remove the cable from the raceway in the bottom of the scale Slide out the narrow cover plate of the support arm in the direction of the arrow Use the Allen wrench to remove the four screws from the rectangular support arm cover plate and remove the cover plate from the support arm base Mount the di
16. splay unit on the support arm using the two retainers and the four screws supplied Route the cable through the cable conduit inside the support arm Mount the angular retainer using the screw and the spring washer so that the bundle of shielded wires contacts the inside of the support arm Insert the cable down through the support arm into opening of the mounting frame Afterwards slide the narrow cover plate into the sup port arm Insert the cable in the raceway as shown in the diagram Make sure that the rubber seal see arrow is not damaged Fasten the support arm to the rear panel of the scale base using the two M5x16 screws supplied and the star lock washers Attach the mounting frame of the support arm to the bottom of the scale using the two M4x8 screws supplied Refasten the rectangular cover plate to the support arm using the four screws Reposition the base plate and fasten it in place by tightening the three M3x8 screws see the large arrows and the two M3x10 screws see the small arrows 15 16 Startup Connecting the Scale to Line Power The scale is powered by an AC adapter Make sure that voltage rating printed on this unit is identical to your local line voltage rating If the line voltage indicated or the plug design of the AC adapter does not match the rating or Standard you use please contact your Sarto rius dealer Important Note Only use original Sartorius AC adapters identifi
17. t 2 5 10 lt 4 10 lt 10 10 Platform size mm 417 x 307 16 4 x 12 1 in Scale base W x D x H mm 417x307 x 113 16 4 x 12 1 x 4 4 in without display unit Net weight of the scale kg 15 33 Ibs Net weight of the display unit kg 12 2 Ibs Power requirements voltage frequency 50 60 Hz 115 or 230 V depending on the AC adapter used Allowable voltage fluctuation 20 15 Power consumption VA 9 typical Interface optional RS 232 C V24 28 RS 423 V10 7 bit parity even mark odd space transmissions rate 150 9600 Baud 27 Sartorius AG amp 37070 G ttingen Cy Weender Landstraf e 94 108 37075 G ttingen 0551 308 0 FAX 0551 308 3289 Internet http www sartorius com Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartorius AG Sartorius AG reserves the right to make change to the technology features specification and design of the equipment without notice
18. ymbol confirms your selection fol DU Ux Qa VI Ru lowed by a return to zero for the 2nd digit or line How to return to the weighing mode Press the tare control each time a 0 appears in the numerical se quence word line page If you have changed a menu code it will be stored as soon as the display returns to the weighing mode Lock the scale operating program using the menu access switch L readout and replace the protective cover 22 List of the Programmable menu Code Settings E l E I Page Line Word Menu of the Balance Operating Program Active Parameters gt vg Code Ambient Conditions Code Program Lock Le tlt Very stable E Cali i OFF C ie Stable Cia ie ON tensa Unstable LH Very unstable Code Acoustic Signal C BP ON Code Stability Range me 2 OFF ce 0 25 digit CES e 0 5 digit Code Weight Units C ie J 1 digit cali i Grams g Le 2 digits L B tie Kilograms kg cCliye 5 4 digits a LiB t 3 Carats ct Lg B 8 digits L amp 1H Pounds Ib Lg i 16 digits t END 5 Ounces Oz t ije a 32 digits L B B Troy ounces ozt Legg 64 digits LIE ti Parts pound o cS i Hong Kong taels tl Code Display Format t EN c Singapore taels tl Cla
19. ys i Last decimal ON m L 5 e Taiwan taels tl tuae Last decimal OFF L 5 g 4 Grains gr t3 3 Last decimal at stability L Blg 5 Pennyweights dwt c Alldecimals at stability L Blg B Mommes o L B g 7 Milligrams o Code Tare parameter BE Austrian carats o CPH i Without stability LH eg At stability a C S e a Call Program Line Code Auto Zero NI Call Program page LS i ON a C eaS OFF Co End of programming Code Calibration Clie i Accessible a t 4 5 g8 Accessible blocked Additional Parameters for the data Output format at the interface port and for calculation programs are available on request Please refer to the Accessories ry Facto Sett ng You can choose any weight unit as long as it can be displayed in the particular weighing range you selected for example do not set the code for kg when you are using a 0 1 mg balance 23 Troubleshooting Guide Problem Causes Remedy No segments are indicated in the weight display 9 Weight display shows L or CH2 No line voltage available The AC adapter is not plugged in The platform cover 1 is not in place An object is pressed against the platform cover or has gotten underneath it Check power supply Plugin the AC adapter Place platform cover on the scale base The platform cover must rest properly on the scale base without any interfering objects in between Weight display