BIOLOG 3000i ™ User Manual - Frank`s Hospital Workshop
Contents
1. 16 2 4 2 Mute the QRS Beeping 1 022 0000000 02200202 16 2 4 3 Transtelephonic 2 6 6600002 0 06000 nnnnen 16 2 4 4 Automatic Device Identification 2 122000000000000 6 9 17 2 4 5 Setting the 17 2 5 SETUJP SCREEN 2 unn penetra ais 18 2 5 f Date Formatz i UR ERRARE RR A 18 2 5 2 Setting the Date 18 2 5 3 Enabling Simultaneous 18 2 5 4 Total Single Lead Recording 19 2 5 5 Selecting the Mains 50 60 Hz Filter 20 VASE PIN S DT EE 20 2 6 SET UP SCREENS ONERE ERES 21 2 6 1 Transmission 5 08 21 1 0000000000 000000 21 2 6 2 Modem 21 2 6 3 Power o ee 21 2 7 SET 5 22 2 7 1 Menu 200 0000000 0 0 0 000 nnns 22 2 7 2 Saving the Set Up 22 RECORDING THE ECG 23 3 1 GENERAL 5 7 44 21 1 nnne nns 23 3 2 THORAX PLACEMENT OPTION eeeee eene 24 3 2 1 Acquiring ECG in Thorax Placement 2
2. 12 lead Biolog Patient cable with the BIOLOG 3000 the Sim option must be enabled in set up menu 2 see section 2 5 3 3 4 1 Recording an ECG with a 6 or 12 Lead Patient Cable e Plug the 6 or 12 lead patient cable into the cable socket Figure 3 5 j Figure 3 5 Connecting Cables to the BIOLOG 3000i RECORDING THE ECG 28 Turn the BIOLOG 3000 on and allow it to cycle to the ECG display screen The BIOLOG 3000i automatically detects the presence of the 6 or 12 lead cable and displays lead I I on the screen Figure 3 6 sonp y gt Ee 1 ADS OFF 5 I HR BIOLOG 3000 Figure 3 6 Monitoring Screen with 12 lead Cable Connected Turn the BIOLOG 3000 off and attach the leads to the patient s chest Figure 3 9 Figure 3 7 Anatomical Positioning of the 6 or 12 lead Cable Electrodes 29 RECORDING THE ECG e Once attached lead will automatically begin monitoring If any lead other than lead I I is desired as the rhythm strip briefly press the MODE key and release This will advance the selected rhythm strip to lead III Any lead may be selected by repeatedly pressing the MODE key Leads available for selection are identified as 1 I1 III BUR RUF for the 6 lead cable and 1 11 BUR 91 02 03 04 0 9 and V6 for the 12 lead cable e Once the lead is selected the BIOLOG 30007 will continue to monitor that lead until you are ready to
3. is incompatible with the 3000i Message When Transferring ECG to PC This can occur if you are touching the back electrodes or are near a source of electrical noise while the download cable is connected It does not interfere with the download operation however to clear the message ensure that the Biolog is resting on an insulated surface and away from power cables transformers or other electrically noisy equipment Unable to take a multi lead ECG with the Biolog Patient cable Make sure that the cable is connected properly Examine the cable s plug for bent pins Ensure that the Biolog is configured for simultaneous cable support on the setup screen Read the instructions for multi lead ECGs carefully If you press the Hold button only a single lead will be stored To acquire a multi lead ECG the Mode button must be pressed and held until acquisition mode is entered Biolog emits a constant sound The Biolog will emit a faint sound when the LCD backlight is on If the backlight is not needed switch it off using the setup screen this will greatly extend the battery life TROUBLESHOOTING 40 Serial Number shown at startup as UUUUU or random value Won t keep correct time and date This is likely to occur if the unit is more than 4 years old Return it to an authorized QRS Diagnostic LLC service center for routine servicing and testing Scrolling lines on the Screen This can occur if holding buttons while switching the B
4. record and save Figure 3 8 12 Lead Biolog Patient Cable RECORDING THE ECG 30 3 3 2 Acquiring and Saving a 6 or 12 Lead ECG e To begin recording the simultaneous 6 or 12 lead samples ensure that all leads are attached reliably then press and hold down the MODE key until a multiple beep is heard Figure 3 11 This will place the BIOLOG 3000 in automatic acquisition mode as indicated by an in the lower right corner of the screen The 6 or 12 leads will then be simultaneously recorded un BIOLOG 2000 Figure 3 9 Starting 12 lead ECG Acquisition The BIOLOG 3000 will automatically save each of the leads when the recording is finished When all leads have been sampled a multiple beep is sounded and a screen will display the status of the acquisition To exit this screen press the MODE key to return to rhythm strip monitoring or REV to print and or review As the BIOLOG 3000 simultaneously records from the 6 or 12 lead cable it checks to ensure that each lead is attached If any detachments are found these will be indicated on the screen at the end of the recording as shown at Figure 3 10 If this message appears the user should check the lead highlighted for faulty electrode connections NOTE if limb leads are detached chest lead detachments may not be detected or the recording may not be able to proceed in the first place Figure 3 10 Leads off message after a 6 or 12 Lead Recording 31 PRINTING amp
5. stored in the memory It will remain in memory until a new EGG is recorded and saved NOTE The BIOLOG 3000i is equipped with a small backup battery It will keep its memory when the unit is turned OFF or when the batteries are removed for replacement MM i 65 SL xov BIOLOG 3009 Figure 3 3 Saving an ECG to memory 3 2 3 Reviewing an ECG Stored in Memory To review an ECG stored in memory e Press the REV button for about 1 second until a long beep is heard The ECG saved in memory will begin scrolling across the screen A d will be displayed in the right lower corner of the screen to indicate that the unit is in REVIEW mode Figure 3 4 e To freeze any portion of the ECG for closer examination press the HOLD key once To release the freeze press the HOLD key once again This can be done as often as desired review can be performed as many times as desired on an ECG saved in memory 27 RECORDING THE ECG eoo REV HR 65 SL wer menm BIOLOG 2000 Figure 3 4 Reviewing a stored ECG iz 3 3 Six or Twelve Lead Cable Option The 6 or 12 lead BIOLOG 3000 cable is used to perform simultaneous 6 or 12 lead ECG acquisition and monitoring Alternatively the operator is able to record a single rhythm strip off any one of the six or twelve leads This recording can be between 5 and 80 seconds Refer to Section 2 5 4 NOTE To use the 6 or
6. to Cardioview followed by the date and time To return to the normal mode of operation press any key The Biolog will reset and go through the normal power up sequence NOTE The BIOLOG 3000 should work with any modem 100 compatible with the Hayes command set The modem initialization string can only be edited using the BIOLOG Setup utility provided with CardioView The BIOLOG 3000i is factory pre set to AT amp FXOW2E1 amp KO0 This should work with the majority of modems but in some cases eg GSM mobile phones it will be necessary to simply use AT amp F 35 PRINTING amp TRANSMITTING THE ECG 4 3 Use with CardioView The BIOLOG 3000 can be used with QRS Diagnostic s CardioView software For information please refer to the user manual or consult your nearest QRS Diagnostic distributor HN 4 4 Printing with the Biolog Printer Interface Stored ECGs can be printed using the Biolog Printer Interface For more information please refer to the Biolog Printer Interface user manual or consult your nearest QRS Diagnostic distributor MAINTENANCE 36 CHAPTER 5 MAINTENANCE 1 Batteries Should the BIOLOG 3000 not operate properly when turned ON check the battery polarity Incorrect battery insertion may damage the unit Only alkaline AA LR6 1 5V batteries should be used in the BIOLOG 30007 Do not use Rechargeable Nickel Metal Hydride batteries with the BIOLOG 30001 NOTE Please consult your local suppl
7. to section 2 1 1 Installing the batteries Try adjusting the contrast knob If misadjusted no display will be visible If the unit is more than 4 years old the unit may need servicing Ensure that the previous steps have been taken and if necessary return it to an authorized QRS Diagnostic LLC service center Message Appears in the Display Battery voltage has dropped below an acceptable level The BRTT message will remain on the screen until the batteries have been replaced Complete the procedure and change the batteries as soon as possible 39 MAINTENANCE Appears in the Display This is a normal message indicating the Leads Off condition It will appear if any electrodes fail to make good contact with the skin Check the electrode skin contact until the ECG monitoring mode is established If using a multi lead cable also check any electrode adapters by swapping them If a leads off indication occurs consistently for a particular patient lead after changing the electrode and the adapter it may indicate a fault in the lead in which case please contact your QRS Diagnostic LLC service center NOTE For better direct thorax contact moisten the back electrodes ECG signal appears noisy or shows interference Check that the ECG electrodes are well adhered and applied according to the manufacturer s instructions Distance the BIOLOG 3000 and the patient from other electrical equipment This cable
8. 3000 screen CAUTION Federal law in U S A restricts this device to the sale by or on the order of a physician CAUTION An authorized QRS Diagnostic LLC service agent must service this device Symbols used on the BIOLOG 3000i and patient cables Attention Consult accompanying Patient cable connection documents socket EN Type BF equipment Power 4 BF defibrillator proof applied I Power on n Class II equipment Internally powered device Regulatory compliance IEC60601 1 amendments 1 amp 2 C22 2 No 601 1 M90 amp UL 2601 1 when used as part of the BLY 501 adapter System TABLE OF CONTENTS INTRODUCTION ARE ERR RR RARE 7 1 1 WHAT IS THE BIOLOG 30007 sse 7 1 2 SYSTEM 3000 COMPONENTS sssseeeenee rennen nnn 8 1 3 FEATURE LOCATIONS 55 odere E nent 9 SETTING UP THE BIOLOG 3000 22 10 2 1 SET UP 10 2 1 1 Installing the 0 10 2 1 2 Stam Up s t e e RR RR ER MER MERE SEER RR EUR 11 2 2 CHECKING THE BATTERY 6 12 2 S THE SET PMENU 13 2 3 1 To Access the Set Up 94 13 2 3 2 Navigating through the Set Up 14 2 4 SET UP SCREEN 1 2 16 2 4 1 LCD Backlight
9. 5 3 2 2 Saving a Single Lead ECG to 26 3 2 3 Reviewing an ECG Stored in 26 3 4 SIX OR TWELVE LEAD CABLE 211 een 27 3 4 1 Recording an ECG with a 6 or 12 Lead Patient Cable 27 3 4 2 Acquiring and Saving 6 or 12 Lead 30 PRINTING amp TRANSMITTING THE 31 4 1 TELEPHONE TRANSMISSION 31 4 2 DIGITAL TRANSTELEPHONIC MODE 32 4 3 USE WITH CARDIOVIEW 1 1 eene REEL RER ERR kr nes 35 4 4 PRINTING WITH THE Biolog Printer 35 MAINTENANCE 36 1 36 23 36 3 Cleaning and 37 TROUBLEESHOOTING 38 APPENDIX A SPECIFICATIONS 41 APPENDIX BY NOTIGES 42 7 INTRODUCTION CHAPTER 1 INTRODUCTION 1 1 What is the BIOLOG 3000 BIOLOG 30004 is a hand held battery operated device designed to e Acquire and store in memory 5 to 200 secs of a single lead ECG via thorax placement e Acquire 5 to 80 seconds of simultaneous 6 or 12 Lead ECG via
10. BIOLOG 3000 User Manual 6 AND 12 LEAD SIMULTANEOUS ECG RECORDER re 2LO02 co QRS DIAGNOSTIC LLC 14755 27th Ave North Tel 763 559 8492 Plymouth MN 55447 Fax 763 559 2961 USA http www qrsdiagnostic com WARRANTY All instruments sold and supplied by QRS Diagnostic LLC are guaranteed to be free from defects in material and workmanship for a period of 3 years from date of purchase All supplies and accessories carry a 90 day limited warranty This includes ECG lead wires If in the judgement of QRS Diagnostic the instrument is proven to be defective during the warranty period it will be repaired or replaced with no charge for parts or labor An extended hardware warranty is available at additional cost This warranty does not cover any instrument that has been damaged by accident misuse abuse or has been altered or repaired by anyone other than an authorized QRS Diagnostic agent This warranty also does not cover any unit that has had a serial number removed defaced or rendered illegible This warranty is in lieu of all other warranties expressed or implied Including warranties of merchantability and fitness and is hereby limited to repair or replacement of instruments found defective during the warranty period ALL REPAIRS MUST BE MADE BY AN AUTHORIZED QRS DIAGNOSTIC AGENT THE INSTRUMENT MUST BE RETURNED FOR REPAIRS AT THE EXPENSE OF THE PURCHASER IN WARRANTY REPAIRED UNITS WILL BE RETURN
11. ED AT THE EXPENSE OF QRS DIAGNOSTIC OR ITS AUTHORIZED AGENT INSTRUMENTS SENT BY MAIL OR COMMON CARRIER SHOULD BE INSURED AGAINST LOSS OR DAMAGES AS THEY ARE NOT COVERED BY THIS WARRANTY 6000 4182 Rev B SERVICE For service please contact the QRS Diagnostic LLC Service Department at 763 559 8492 Service Representatives are available by phone Monday through Friday from 8 a m to 5 p m Central Standard Time You can also contact Technical Support by email at support QRSdiagnostic com In the event that your problem can not be fixed over the phone or via email you will be issued a Returned Merchandize Authorized number RMA Please have the serial number of your unit available Please re package your unit in the original packing materials if possible and ship it to QRS Diagnostic 14755 27 Ave N Plymouth MN 55447 This document contains proprietary information which is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the written consent of QRS Diagnostic LLC The information contained in this document is subject to change without notice m FOREWORD Thank you for choosing the BIOLOG 3000 portable ECG monitor BIOLOG 3000 is the world s most portable multi functional ECG recorder It which can be used on its own for instant cardiac monitoring or with other System 3000 components to provide an integrated diagnostic monitori
12. IOLOG 3000 monitor when sending a transtelephonic ECG with your BIOLOG 3000 to a reception center equipped with QRS Diagnostic s CardioView software This facility is called Automatic Device Identification or ADI in brief Turn the ADI ON if sending an ANALOG Transtelephonic recording to CardioView Turn ADI OFF if sending an ANALOG Transtelephonic recording to a non QRS Diagnostic receiving station To Select the ADI function press the HOLD key e To change the setting press the REV key 2 4 5 Setting the Time e To change the time move the cursor to the appropriate position HH for hours or MM for minutes by pressing the HOLD key right arrow repeatedly e Press the REV key up arrow to change to the desired time NOTE The time is preset in 24 hour format e g 3 30 PM will be shown as 15 30 The factory setting is Australian Eastern Standard Time SETTING UP THE BIOLOG 3000i 18 2 5 Set Screen 2 EU 127187199841 Sim 9 25 80 OK Flt 58Hz 51680 PrSP 25 25156 SS Figure 2 9 Set Up Screen 2 2 5 1 Date Format To change the date format from DD MM YY European to MM DD YY US move the cursor by pressing the HOLD key right arrow repeatedly until the symbol EU is highlighted To change the date format to US format MM DD YY press briefly the REV key up arrow The symbol US will appear on the screen and the date format will change to MM DD YY The date for
13. TRANSMITTING THE ECG CHAPTER 4 PRINTING amp TRANSMITTING THE ECG The BIOLOG 3000i has a number of methods of printing and transmitting its data after a recording has been made Refer to Section 1 2 4 1 Transtelephonic Transmission This is provided for backward compatibility with old systems and is no longer supported Transmission by digital modem is now recommended see next section To use this option BIOLOG 3000 must be set to TRANSTELEPHONIC MODE Refer to section 2 4 3 Dial the number of the reception station then place the telephone mouthpiece over the Audio Outlet located on the lower right of the unit s face Figure 4 3 Figure 4 3 Setting up BIOLOG 3000i for transmitting an ECG by telephone PRINTING amp TRANSMITTING THE ECG 32 Depress the REV key and hold until the long audio tone is heard Figure 4 4 The recorded ECG will be heard during the transmission Figure 4 4 Transmitting an ECG by telephone During the transmission a high pitched variable audio tone will be heard To ensure a clear transmission avoid moving the BIOLOG 3000 or telephone handset during the transmission 4 2 Digital Transtelephonic Mode The BIOLOG 3000 has the ability to send your ECG by a standard modem or GSM digital mobile phone for interpretation by CardioView This digital transfer allows for high speed accurate transmission of ECG data Figure 4 5 Digital Transtelephonic ECG Data Transf
14. aud Rate BR at which the BIOLOG 3000 will communicate with the modem or GSM digital mobile phone 2 6 3 Power down This is an adjustable time in minutes after which the BIOLOG 3000 will turn off if not in use If this value is set to zero the power down function will be cancelled SETTING UP THE BIOLOG 3000i 22 HERE 2 7 Set Up Screen 4 Menu OK Figure 2 13 Screen 4 2 7 1 Menu Display Selecting this V allows the set up menu to be displayed briefly when turning on the BIOLOG 3000 as mentioned in section 2 3 1 If X is selected the menu will not be displayed on start up however it can still be accessed by pressing the MODE and HOLD keys simultaneously 2 7 2 Saving the Set Up Parameters To save the Set Up Menu parameters press the MODE key OK on the display at any time during the set up Figure 2 12 When OK is selected the new set up parameters will be adopted and any old parameters will be cleared M Sim gt Rec HEC288 OK F1t 58Hz 5168 PrSP 25 25158 mam i EU 12 10 1998 j LI j BIOLOG 2000 Figure 2 14 Saving the set up menu parameters 23 RECORDING THE ECG CHAPTER 3 RECORDING THE ECG HN 3 1 General Preparations Place the patient in an at rest position preferably prone on their back Turn the BIOLOG 3000 on by sliding the power switch on Figure 2 4 To change or review any of the Set Up parameters refer to the S
15. can select between 50 Hz and 60 Hz filters in accordance with the mains current in the country of use The default factory setting is 50 Hz To select a 60 Hz mains filter value move the cursor to the symbol 50 position by pressing the HOLD key repeatedly To change the mains filter to 60 Hz simply press the REV key Figure 2 11 MM EU 12 18 1998 Sint XXI 9 Rec 20 200 OK Fit o Hz EIEEE PrSP 25 25150 AS sz ll BIOLOG 000 Figure 2 11 Selecting Mains Filter 2 5 6 Print Speed The Biolog Printer Interface can be set to either 25mm sec or 50mm sec e select print speed press the Hold key right arrow To set print speed press the REV key to the desired value 25mm sec or 50mm sec 21 SETTING UP THE BIOLOG 3000i HN 2 6 Set Up Screen 3 099 OK BR 9600 Pur 5 Figure 2 12 Screen 3 2 6 1 Transmission Settings When transmitting to the Biolog Printer Interface or Analogue Transtelephonic transmission to CardioView the BIOLOG 3000i may be set to transmit either 3 5 seconds per lead or 10 seconds per lead 2 6 2 Modem selection Selecting this allows the BIOLOG 3000 to communicate with a modem or GSM digital mobile phone refer to section 4 2 This also allows the user to enter the modem telephone number Num to be dialed as well as selecting the B
16. er 33 PRINTING amp TRANSMITTING THE ECG You will need a Biolog to Modem Adapter Cable and a standard serial adapter or cable from the modem to a DB9 female socket To transfer a stored ECG to CardioView by modem enable the modem field in Set up screen 3 refer to section 2 6 Enter the full telephone number of the receiving station and set the baud rate as recommended by your modem manufacturer refer to section 2 6 The telephone dial string may contain the following 0 9 Normal telephone number digits A D Normal telephone digits Hash key E Star key Dial pause for switchboards Tone dialling default Pulse decadic dialling Wait for dial tone Wait for silence Flash On hook and space Ignored may be used to format the dial string Power the BIOLOG 3000 and then connect the BIOLOG 3000 to the modem or digital mobile phone using the cable Press and hold the REV button to start the transfer NOTE receiving station must be ready to receive the data refer to the CardioView user manual for more information The BIOLOG 3000i should make the call and send the data automatically If there is a failure in the connection to the modem it is not switched on or the baud rate is incorrect a higher then a lower long tone will sound and normal review mode will be entered If the modem is connected and operating correctly a screen headed Transfer Mode will appear Below th
17. et Up Menu Instructions If no changes are needed allow the BIOLOG 3000 to proceed to the ECG display screen For best results e For better contact when applying the BIOLOG 3000 directly on the skin surface moisten the electrodes with electrode gel or water When using patient cable accessories e Use with Ag AgCl ECG skin electrodes Attach all electrodes before turning the unit ON Keep the patient and cables as motionless as possible during monitoring Lay the BIOLOG 30007 on a flat surface during monitoring to ensure least disturbance Two options are available to obtain the ECG Thorax placement and 6 lead or 12 lead patient cable acquisition After briefly displaying the set up menu for 2 seconds the BIOLOG 3000 goes into the ECG display screen Figure 3 1 SINGLE LEAD mode is indicated by the symbol SL at the bottom of the ECG display screen RECORDING THE ECG 24 NOTE The gain of the ECG can be switched between 5 and 10 JL mm mV The normal adult setting should be 10 mm mV which is represented by a taller calibration symbol at the bottom of the center ECG Display screen To select 5 mm mV e g for pediatrics applications briefly press the REV key 2000 Mh s H 2 eer HR A SL BIOLOG 566 Figure 3 1 The ECG Display screen 3 2 Option Place the BIOLOG 3000 on the patient s left anterior chest with the electrode fe
18. et firmly against the skin For ideal placement in Lead II when viewing the BIOLOG 3000i from the front e The lower right electrode foot is the positive electrode It should be placed over the apex of the heart The upper left electrode foot is the negative electrode It should be placed on the sternum as close as possible to the manubrium The other two electrodes act as references with at least one reference electrode in contact with the thorax The BIOLOG 3000 is now ready for monitoring 25 RECORDING THE ECG aT ss Lead I NE II Lead i Y Figure 3 2 Different Leads using Back Electrodes 3 2 1 Acquiring an ECG in Thorax Placement Mode Turn on the BIOLOG 3000 as described above Allow the unit to settle in ECG display mode Place the BIOLOG 3000 on the patient s chest as indicated above The leads off 2indication should vanish and after a few seconds the baseline of the patient s ECG will settle down and appear on the screen Maintain the BIOLOG 3000 in this position for at least as long as the Recording Time selected in the Set up Menu To change or review any of the set up parameters refer to the Set Up Menu Instructions RECORDING THE ECG 26 3 2 2 Saving a Single Lead ECG to Memory e To save an ECG to memory press the HOLD button at the end of the ECG recording Figure 3 3 e wil appear in the lower right of the screen This indicates that the ECG is saved and
19. f automatically Refer to Section 2 2 2 4 2 Mute the QRS Beeping Some patients may be disturbed by the beeping sound made on the detection of every QRS mute this sound press the HOLD key to select the open symbol r e To change the setting press the REV key The symbol will change to X indicating that QRS Beeping is muted 2 4 3 Transtelephonic Mode BIOLOG 3000 is able to transmit a stored ECG recording using acoustic coupling via a telephone to a remote site equipped with a suitable reception system This is provided for backward compatibility with old systems and is no longer supported Transmission by digital modem is now recommended see modem selection 17 SETTING UP THE BIOLOG 3000 The following table highlights the different transtelephonic modes of the BIOLOG 3000i DEFAULT MODE Analog transtelephonic mode disabled Choose this mode for silence during recording or review Analog transtelephonic mode is enabled only for the review or transmission of data The acquisition of data is silent Choose this mode for silence during recording Sound will be emitted during review Analog transtelephonic mode is enabled for the acquisition review and transmission of data This mode is for transtelephonic transmission of the ECG in real time to a receiving center Sound will be emitted during recording and review 2 4 4 Automatic Device Identification ADI You can send the serial number of your B
20. g the patient bed or any conductive material in contact with the patient during defibrillation the delivered energy may be partially discharged through that person Make sure persons stand away from the patient bed and other conductive material before discharging the defibrillator WARNING Batteries and other disposables used in conjunction with this device must be disposed of in accordance with the national requirements in the country of use CAUTION A 6L Biolog Patient cable or 12L Biolog patient cable is required to be used in conjunction with the BIOLOG 30004 in order to provide protection against the effect of the discharge of a cardiac defibrillator and against high frequency burns Conductive parts of electrodes and associated connectors including the neutral electrode should not contact other conductive parts including earth The BIOLOG 3000 is not intended to be sterilized or used in a sterile environment and therefore is not suitable for direct cardiac application The BIOLOG 3000 is not suitable for use with flammable anaesthetics and offers no water ingress protection Inspect the BIOLOG 3000 electrocardiograph and 6L Biolog Patient Cable or 12L Biolog patient cable regularly for signs of mechanical damage particularly around patient leads To verify correct operation of the electrocardiograph connect all patient leads to a patient and check that there is a valid ECG signal on every lead by viewing the signal on the BIOLOG
21. gure 2 1 Opening the Battery Hatch Load the Batteries with Correct Polarity Match the battery polarity signs with the polarity marked on the interior of the battery compartment casing Figure 2 2 Inserting the batteries Close the battery compartment NOTE Check the batteries were inserted with correct polarity Figure 2 2 11 SETTING UP THE BIOLOG 3000i Align press and slide the battery compartment cover until the locking catch snaps into position Figure 2 3 Figure 2 3 Closing the battery hatch 2 1 2 Start Up Slide the BIOLOG 3000i ON OFF key to the ON position Figure 2 4 The screen will display e The title screen for 2 seconds The Set Up Menu for 2 seconds if configured to and e The ECG display screen If the electrodes are not applied the center of the screen will show the Leads Off symbol Figure 2 4 Turning the BIOLOG 3000i on At any time you can adjust the contrast of the LCD display with the CONTRAST control button situated on the side of the unit this may be necessary to produce a visible display if it has been misadjusted SETTING UP THE BIOLOG 3000i 12 pem 2 2 Checking the Battery Status The power remaining in the batteries is displayed on screen by the battery status bar The battery status bar is located Atthe bottom of the Set Up Menu screen Figure 2 5 Or e Atthe top of the screen during normal operation Figure 2 5 The battery status bar is an indicator of t
22. gy in appearance To disinfect the BIOLOG 30001 wipe the back of the device including the electrodes with a hospital grade disinfectant NOTE Do not saturate the BIOLOG 3000 with any liquid because excess fluid could enter the case and damage the electronics TROUBLESHOOTING 38 CHAPTER 6 TROUBLESHOOTING Symptom Likely Problem and Solution If you should experience any difficulty when using your BIOLOG 3000i please review the suggestions for troubleshooting some minor difficulties presented below If you cannot resolve the difficulty after following these procedures please contact your supplier or call QRS Diagnostic LLC No power the unit is off after being on When not engaged in real time monitoring the BIOLOG 3000 will automatically shut off after the period of time specified section 2 6 3 power down to conserve battery power Move the ON OFF Switch to the OFF position then turn the unit ON again If the unit shuts down during download to PC or printing it is likely that the batteries are flat or are not of the recommended type Replace the batteries if this is occurring Turning off the backlight may give some extra operational time if no fresh batteries are immediately available NOTE Data stored in memory will not be lost No power the unit will not turn on The batteries are flat or they are not properly installed Remove the Battery Compartment cover and check the position of the batteries refer
23. he battery voltage with 8 segments showing the voltage range for usable operation You can visually check at any time how much power remains in the batteries When the voltage approaches the minimum for reliable operation the symbol BAT appears at the bottom of the screen At this stage you should complete your procedure and replace the batteries as some operations such as printing and transferring data to a PC cause a dip in the power levels which can cause the Biolog to shut down without warning Battery longevity is affected by backlighting To preserve battery life you should turn off the backlight If you turn the backlight ON or OFF you will notice that the battery status bar adjusts itself to the new condition a000 d mn 3000 M t i HR 651 01 27 3 XXII RDION OFFIOND 18 56 hh mm BIOLOG 2000 Figure 2 5 Checking Battery Status 13 SETTING UP THE BIOLOG 3000 2 3 Set Up Menu The Set Up Menu displays the following functions and their current settings Screen 1 e The Backlight Mute the QRS beeping e The Transtelephonic Mode e Automatic Device Identification header for transtelephonic transmission e Time Screen 2 e Date format Date Simultaneous Acquisition Enabled Length of recording The 50 60 Hz AC Filter The Print Speed of the printer Screen 3 e Infra red transmission settings e Modem setting
24. ier for further details Without electrode contact through the electrode feet or through the patient cable the BIOLOG 3000 will turn itself off automatically after the period of time specified section 2 6 3 power down to preserve battery longevity In continuous use the batteries will last approximately for more than 10 hours with backlighting and more than 24 hours without backlighting To preserve battery life turn backlighting off when not needed and turn unit off when not in use Remove batteries from the unit if not being used over a long period of time as they may leak and cause damage to the unit 2 Electrodes The contact electrode feet at the back of the BIOLOG 3000 are made of high grade stainless steel and are corrosion resistant Wipe the electrode feet clean of gel or other residue after each use Residue left on the electrodes may affect the BIOLOG 3000 sensitivity 37 MAINTENANCE 3 Cleaning and Disinfection BIOLOG 3000 can be cleaned and disinfected very simply as follows To clean the BIOLOG 3000 plastic case wipe the surfaces of the case with a clean cloth moistened with water only To clean the display gently wipe the plastic display lens with a soft damp cloth moistened with water or a commercial glass cleaner NOTE Do not wipe the case or the display lens with alcohol Alcohol will damage plastic surfaces over time and may cause them to crack In addition the display lens may become fog
25. iolog on To cor rect this turn the Biolog off remove the batteries for ten seconds re place the batteries and turn the Biolog back on making sure you do not touch any of the buttons If the problem continues contact your QRS Diagnostic LLC service center 41 APPENDIX APPENDIX A Specifications ELECTRICAL CHARACTERISTICS 6 12 Lead Rear Electrodes Sampling Rate 500 Hz 300 Hz Frequency Response 0 05 175 0 5 100 Hz Hz CMRR gt 60dB gt 80 dB 0 100 Hz Isolation 4kV Requires6 12Lead DISPLAY Sweep Speed ECG 20 mm sec 1 mm sec Display Sensitivity ECG 5 or 10 mm mV x 0 5 mm mV Viewing Area 40 x 48 mm No of Pixels 240 h x 160w Display Type LCD Back Light Electro luminescent MEASURED PARAMETERS Heart Rate HR 20 250 bpm 2 POWER REQUIREMENTS Battery Type 2 x AA 1 5V Alkaline cells recommend Sanyo LR6 Battery Life typical 12 Hours continuous use PHYSICAL CHARACTERISTICS Length 134 mm Width 88 mm Thickness 32 mm Weight 270g ENVIRONMENTAL CHARACTERISTICS TEMPERATURE Operating 0 C 40 Storage 20 C 70 C HUMIDITY Operating 5 95 Storage 5 95 ATMOSPHERIC PRESSURE General 500 hPa 1060 hPa Specifications are subject to change without prior notice APPENDIX 42 APPENDIX B Notices Copyright Notice Copyright 2003 by QRS Diagnostic LLC All rights reserved This document contains pr
26. is the modem s responses to the setup string and dial command will be displayed If the response ERROR is seen then the modem has a problem with the initialization string or the dial string You will need to set the init string to values recommended by your modem instructions and ensure that the dial string does not contain any incorrect characters Note that configuration of the setup string requires the Biolog 3000i Setup utility supplied with CardioView PRINTING amp TRANSMITTING THE ECG 34 There will be no response to the dial string for a time while the call is made and the remote station answers If this takes more than 60 seconds the process will abort there is a counter at the top of the screen showing the progress of the timeout If the line is busy or there is a problem establishing the connection you may see the message 40 CARRIER in which case press any key to reset the Biolog and try again When connection is established a progress indicator DATA gt gt gt gt will extend to the right of the screen until the call is completed Below ths Connected to name of the receiving station defaults to Cardioview will be shown Pressing any key during the data transfer will abort the transmission When the transmission is complete the BIOLOG 3000 will display message Data transfer successful or unsuccessful Below this will be displayed Received by name of the receiving station defaults
27. mat is preset DD MM YY European format 2 5 2 Setting the Date e To change the date move the cursor to the appropriate position MM or DD or YY by pressing the HOLD key right arrow repeatedly To set the date briefly press the REV key up arrow 2 5 3 Enabling Simultaneous Acquisition The Sim selection must be enabled before the BIOLOG 3000 can be used with the 6 or 12 Lead simultaneous cable Even if the Sim selection is changed and saved the ECG in memory will still be stored If the BIOLOG 3000 is turned OFF during Set Up any changes to the Set Up will not be saved however the current saved ECG will be preserved 19 SETTING UP THE BIOLOG 3000 2 5 4 Total Single Lead Recording Time The memory of the BIOLOG 3000i is programmable in 5 second increments between 5 and 200 seconds for a single lead ECG It is also possible to record a single rhythm strip off a 6 or 12 Lead simultaneous cable When the 6 or 12 Lead is enabled you can record between 5 and 80 seconds of single lead ECG To change the Recording Time move the cursor to the Recording Time area Rec by pressing the HOLD key and then increment the Recording Time by pressing the REV key A EU 1271871998 Sin 9 298 OK F1t 58Hz 5168 PrSP 25 25158 ee BIOLOG 2000 Figure 2 10 Changing the single lead recording time SETTING UP THE BIOLOG 3000i 20 2 5 5 Selecting the Mains 50 60 Hz Filter You
28. ng system The BIOLOG 30007 consists of a unit weighing 270 grams and a simultaneous patient cable This allows you to obtain store and analyze patient ECG data quickly and with great confidence Cardiac records are stored by the BIOLOG 30001 which then can be transferred to QRS Diagnostic s patient management software CardioView Alternatively they can be printed to an inkjet printer via the Biolog Printer Interface The BIOLOG 3000 s capability to perform electrocardiograms by direct chest placement makes it especially suited for Emergency Care allowing when necessary the use of real time telephonic monitoring We are sure that this device will enhance your medical practice and allow you to achieve a greater level of freedom without losing the power of diagnosis you have been accustomed to European Representative Advena Ltd P O Box 30 Leominster Herefordshire HR6 0ZQ UK PRECAUTIONS WARNING To prevent the possibility of electrocution ensure that the rear electrodes are not in contact with the patient while device is connected to any peripheral device WARNING The following precautions need to be taken when a defibrillator is used on a patient Do not touch the patient during defibrillation Do not touch the defibrillator s paddle electrode surface when discharging the defibrillator Keep defibrillator electrodes well clear of other electrodes or metal parts in contact with the patient If a person is touchin
29. oprietary information which is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the written consent of QRS Diagnostic LLC The information contained in this document is subject to change without notice and does not represent a commitment on the part of QRS Diagnostic LLC BIOLOG 30001 CardioView amp Biolog Printer Interface are registered trademarks of QRS Diagnostic LLC USA Patents pending The BIOLOG 30004 is year 2000 compliant APPENDIX 43 ELECTROMAGNETIC COMPATIBILITY NOTICE This equipment has been tested and found to comply with the limits for a class B digital device pursuant to part 15 of the FCC Rules AS NZS3200 1 2 1995 and IEC601 1 2 1993 These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio telecommunications In addition radio frequency transmitters such as CB radios and wireless or cellular mobile phones may adversely affect this device There is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception or is itself affected by equipment off and on the user is encouraged to t
30. r changing the Set Up Menu To confirm cancellation of changed items turn the power off and on then review the Set Up Menu e The rightmost column of the setup screen displays symbols to remind you to Press the Up arrow key REV to increase values gt Press Right arrow key HOLD to select values OK Press OK symbol key MODE to save the settings 15 SETTING UP THE BIOLOG 3000 0 ls X xut ADION 9 o X0 Boi XXII OK RDIONHCOFFIOND 18 56 hh nn 1 Figure 2 6 Moving the cursor to the desired position To increase the number press the REV key up arrow until the desired number is shown Hold the key down for fast scrolling Release the key when the desired value is displayed Figure 2 7 o 3 OO S o gt XcKIGI OK RDIONCOFFIOND 56 hh mn Figure 2 7 Selecting a desired value SETTING UP THE BIOLOG 3000i 16 HEX 2 4 Set Up Screen 1 o o 3 Xl gt X OK ADI COFFION 10 56 hh mm a Figure 2 8 Screen 1 2 4 1 LCD Backlight Illumination The LCD Backlight can be turned ON or OFF as desired e First move the cursor to the small circle 2 on the Set Up Menu using the HOLD key e To turn the backlight ON OFF press the REV key When the backlight is turned on the small circle symbol changes to a sun symbol 3 The Battery Status Bar adjusts itsel
31. ry to correct the interference by one or more of the following antenna e Reorient or relocate the receiving antenna e Increase the separation between the interfering devices e Connect the equipment into an outlet on a circuit different from that to which the other equipment is connected e Consult the dealer or an experienced radio TV technician for help
32. s e Power down time Screen 4 e Menu Display 2 3 1 To Access the Set Up Menu There are two ways to access the setup menu e f the setup menu appears briefly after power up press any key while it is displayed The setup menu only appears at power up if the Biolog has been configured this way Or e Press and hold the MODE key then press the HOLD key at the same time The HOLD key must be pressed shortly after holding the MODE key if left too long there will be a timeout and the screen will briefly go blank If this happens release both keys and try again When the setup screen appears both keys can be released SETTING UP THE BIOLOG 3000i 14 2 3 2 Navigating through the Set Up Menu The Hold KEY selects The Hold key selects and highlights the items you wish to change within the Set Up Menu Each time you press the Hold key within the Set Up Menu the cursor moves to the next field or screen There is no button to go backwards if necessary you can repeat the cycle through all of the fields and screens until again reaching the required field The Review REV Key changes The Review key changes the value of the highlighted selected item If there is more than one option you can cycle through all of them until reaching the required selection e The Mode Key saves Press the Mode key to save the new values Cancelling changes The Set Up Menu will revert to previous settings when Mode key is not pressed afte
33. the 6 or 12 Lead patient cable The BIOLOG 3000 incorporates the following features Single lead ECG acquisition e Diagnostic quality 6 and 12 Lead Simultaneous ECG acquisition e High definition LCD screen readable in bright light and with switchable backlighting for low light conditions Heart rate monitoring 20 to 250 beats per Min and display Automatic Device Identification facility for management of recordings e Battery status bar on screen e Infra red linked companion printer interface with 25 and 50 mm second print speeds Transtelephonic transfer of recordings INTRODUCTION 8 EN 1 2 System 3000 Components Biolog Printer Interface BIOLOG CARDIOVIEW 9 INTRODUCTION ER 1 3 Feature Locations AUDIBLE DISPLAY TELEPHONE OUTPUT PATIENT CABLE SOCKET ON OFF SWITCH CONTRAST INFRA CONTROL RED WINDOW MODE SWITCH HOLD SWITCH REVIEW SWITCH Figure 1 1 BIOLOG 30001 View ELECTRODE IDENTIFICATION ewe BATTERY COMPARTMENT COVER Figure 1 2 BIOLOG 3000i Bottom View SETTING UP THE BIOLOG 30007 10 CHAPTER 2 Setting up the BIOLOG 3000 2 1 Set Up Instructions 2 1 1 Installing the Batteries Your BIOLOG 3000 is supplied with two AA LR6 1 5V alkaline batteries Slide the ON OFF switch to OFF position Remove the battery compartment cover by pressing down on the cover while simultaneously sliding it backwards Figure 2 1 Fi