1. This ultrapure water dispensing system has been engineered to consistently produce ultrapure water exceeding every requirement for ASTM Type purity The system is user friendly requires minimal space and maintenance and harmonizes with existing laboratory equipment all at a low initial investment and low operating cost A central dispensing station can supply the high quality water needs of one lab bench or an entire laboratory The system may be wall mounted set on a bench or free standing where bench space is of a premium A unique feature of the Gemini is the Multi Pass UV treatment system in which water passes through a special system of self cleaning Teflon coils While in these coils the water is exposed to the appropriate amount of ultraviolet wavelength to produce biological control This patented technology eliminates the performance deterioration that has been seen over time in other systems and minimizes the maintenance and operating costs compared to other systems The influent water is treated by UV irradiation prior to entering the polishing cartridges phase thereby protecting the system s critical components from bio contamination biofilm buildup and eventual spill over into the product water The polished water is again UV irradiated before leaving the unit The Gemini delivers water of unsurpassed microbiological organic and ionic purity through the dispensing faucet which is continuously exposed to UV
2. Ver 2 1 Thank you for purchasing the Gemini High Purity Water Dispensing System For optimum performance and safety please read and follow these instructions carefully You have a quality world class water treatment system that will provide ultrapure water for many years of trouble free operation Years of engineering and practical experience went into manufacturing this appliance which will dispense water exceeding every requirement for Type water purity To assure that you operate and maintain this unit properly and safely please read this manual carefully In order that we may keep you informed regarding your unit please take a moment to complete and mail the owner registration card provided with this appliance If you have any questions concerning the operation of the unit or need service refer to section 11 0 REQUESTING SERVICE When requesting service you will need the model and serial number as well as the purchase date of your unit For your convenience record this information below Model Number Date Serial Number Dealer Contact Phone Cartridge 1 Part Number Cartridge 2 Part Number Cartridge 3 Part Number Options Installed ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com Section 1 0 2 0 3 0 3 1 4 0 5 0 5 1 5 2 5 3 5 4 5 5 5 6 5 7 5 8 5 9 6 0 7 0 8
3. ASTM Specific Conductance micromho cm max Specific Resistance megohms cm min pH 25_ C Silicate Soluble Ig L max Culture Colony Count cfu mL Total Organic Carbon TOC max Ig L Sodium max Ig L Chlorides max Ig L Note CAP College of American Pathologists ASTM American Society for Testing and Materials NCCLS National Committee for Clinical Laboratory Standards EU Endotoxin Units N L No Limit The measurements of pH in Type I Type Il and Type III reagent water is meaningless and has been eliminated from the procedure since the electrodes used in this test contaminate the water in When bacteria levels need to be controlled reagent grade types should be further classified as follows Microbiological contamination Max heterotrophic bacteria count Type A 10 1000 mL 0 03 EU Type B 10 100 mL 0 25 EU Type C 100 10 mL References CAP Reagent Water Specifications 1985 pg 5 11 NCCLS Document C3 A2 Vol 11 No 13 Pg 8 Annual Book of ASTM Standards 1992 Vol 11 01 D1193 91 Pg 45 13 0 WARRANTY Systems and options have a one year warranty from the date of purchase any component fails to function normally within the first year return the unit to the factory prepaid Ifin the opinion of the factory failure was due to defects i materials or workmanship repair or replacement will be made without charge reasonable service charge will be made for diagnosis or rep
4. may be operated independently by pressing the corresponding DISP or RECIRC KEYS 3 DE PRESSURIZATION amp CARTRIDGE REMOVAL Sanitizing solutions are not compatible with the DI and final filter cartridges and therefore these filters must be removed during sanitization Close the supply water isolation valve and press the DISP KEY to relieve system pressure until no water dispenses Remove the basin grille and filter access door held in place with magnetic catches then remove the cartridge bowls and cartridges note the order of cartridges left to right Though we recommend cartridge replacement after sanitization if the cartridges are not exhausted and will be re used be sure to handle them with sterile gloves and store them in sterile plastic bags to prevent contamination and label each cartridge as it is removed 1 2 3 for correct relocation 4 RE FILL amp BLEED AIR After removing the cartridges replace the bowls Open the feed water valve and open the dispensing valve to purge air from the bowls ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 10 12 Sanitization Procedure Continued Replace the front panel and grill and place the sanitizing tank on the grill directly below the dispensing nozzle The tank should be half full from step one Close the feedwater supply valve and disconnect the female tubing qui
5. screws that attach it to the top platform Gently lift the top cover up and left and set it aside Carefully remove the UV bulb it from its protective package Holding the lamp by the plastic ends slide the 4 prong end of the lamp into the 3 4 hole on the right side of the UV housing Gently guide the lamp fully into the housing up onto the lamp guide far side and into the socket Leave the top cover off until start up is complete e NOTE The UV Bulb may have to be rotated so the 4 prongs align correctly with the bulb socket ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 5 3 Installing the Water Purifying Cartridges NOTE For easy installation the left and right hand cartridges must be installed first Install the center filter last Remove the stainless steel basin grill and set it aside Remove the black front panel by pulling the bottom forward and out it is held in place with four 4 magnetic catches and set it aside This accesses the water purifying cartridge compartment There are three 3 cartridge housings with removable bowls having top O ring seals The bowls have right hand threads Thread the bowl to the left to loosen and remove cartridges and thread to the right to close and seal Remove the three bowls at this time and set aside being careful not to lose the O rings store inside each bowl e Locate
6. the DI resin cartridges have exhausted their capacity to perform Disconnect or turn off the water supply and push the DISP key until water no longer dispenses to relieve water pressure within the system Remove the stainless steel basin grill and set it aside Remove the black front panel by pulling the bottom out and up it is held in place with four 4 magnetic catches and set it aside This accesses the water purifying cartridge compartment Locate and replace the DI resin cartridges for your system refer to sections 5 3 Installing the Water Purifying Cartridges and 5 4 Applying Water Pressures and Bleeding Air for procedure and sequence to open load and purge air from the filter cartridge housings For Ultrafilter option top cover must be removed to gain access to the filter Remove in line filter and replace with new filter Follow steps as stated in sections 5 4 and 5 6 The used cartridges may be disposed of in the same manner as any water filter cartridge ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com Section 9 SANITIZATION CAUTION Prior to sanitizing the GEMINI it is important to become familiar with the owners manual sections 5 0 amp 8 0 for the correct method of cartridge removal and replacement and bleeding air from cartridge housings Always use protective gloves and eye wear when handling chemicals and care
7. to stop dispensing water We recommend a sample be collected in an appropriate sterile sample container and tested for water purity Periodic laboratory water testing to confirm the effectiveness of the treatment process is the owners responsibility ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 5 7 Manual Dispensing of Water e Press the DISP to start and stop dispensing water e Pressing the 1 2 3 or 4 batch buttons will automatically dispense the programmed amount of water as set by the user SEE SECTION 5 8 for BATCH PROGRAMMING INSTRUCTIONS 5 8 Programming and Dispensing a Batch of Water PROGRAMMING A BATCH e Press the MODE key once and the C LED will flash This indicates that you are now in the batch programming mode e Push any batch button to start the flow of water e Dispense the desired amount of the batch e Push the batch button a second time to stop the flow of water and lock in the batch e All 4 batch buttons may be programmed in the PROGRAMMING mode e Press the MODE key once to return to the NORMAL operating mode The A LED will be light continuously in the normal operating mode USING A BATCH Press the desired batch to dispense 1 2 3 or 4 The GEMINI will automatically dispense the preprogrammed quantity of water 5 9 Entering the SERVICE Mode e Push and hold MODE but
8. 0 8 1 8 2 9 0 10 0 11 0 12 0 13 0 Table of Contents Description Important Safeguards Glossary of Terms General Description Principles of Purifying Water Product Specifications Setting up and Using Your GEMINI System Setup Installing the Water Purifying UV lamp Installing the Water Purifying Cartridges Applying Water Pressure and Bleeding Air Applying Electrical Supply Initial Polishing of new cartridges Manual Dispensing of Water Programming and Dispensing a Batch of Water Entering the SERVICE Mode Display amp Keypad Functions Component Location amp Identification Maintenance Replacing the UV lamp Replacing the Resin amp Filter Cartridges Sanitization Procedure Options and Accessories Requesting Service Charts and References Warranty and Sales Contacts ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 1 0 Important Safeguards When using this electrical appliance basic safety precautions should always be followed including the following READ ALL INSTRUCTIONS before use Follow all WARNINGS and INSTRUCTIONS marked on the product DO NOT immerse appliance in water or other liquid Use a soft cloth and mild detergent when cleaning the cabinet or inside of the basin DO NOT remove the top cover with unit power applied DO NOT remove the ultraviolet lamp cover with unit power applied DO NOT look d
9. SERVICE mode ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com ELWOWABLE Pu DOVER k CONMPARTWEMT FOR PAYHTED UA Pagi UY STERLIZATIDN ATOR ATOM Greer be aed CDNTWOLS OPERATOR MMRR ACE Fa PulTt MONITOR gna f WORE SCE SR DOTI Ae 12 CLEARAMCE FP TEN TED MV PROTECHED OPENS HOTIE MONDED Pet TARH amp REMOVABLE FLL RENOVARE PAEL COM MEPMENMF Pt T PNL j THREE FILTER H usinecs FOR WED BED D ond Saa C BON FILTER CAR TRDCOE 5 20 GDDDin FRONT BeOS CCE SiR qas CONTROLLER dc PURSTY wire TOR PC B AeOS F 10 7500in rel Cs 1B 6250in 4 s I 7 0 Component Location and Identification PATERE CHUM PARS Wi 20 3750n ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 8 0 MAINTENANCE This unit was designed and manufactured to have the ability to produce the purest water that can be obtained for laboratory usage It purifies water that meets or exceeds the specifications recommended by CAP ASTM for Type Reagent Grade Water when it is operated and maintained according to these instructions e The ultraviolet lamp must be changed once each year e The resin cartridges must be replaced when the purity meter indicates the water purity is below your requirem
10. airs not covered b this warranty ARIES ResinTech Inc assumes no other responsibility or liability This warranty does not cover any product which has been subject to misuse neglect accident or used in violation of operating instructions nor does it extend to any units altered or repaired for warranty defects by anyone other than ARIES ResinTech Inc IN NO EVENT SHALL ARIES RESINTECH INC BE LIABLE FOR ANY INCIDENTAL CONSEQUENTIAL SPECIAL INDIRECT PUNITIVE OR EXEMPLARY DAMAGES OR LOST PROFITS FROM ANY BREACH OF THIS WARRANTY OR OTHERWISE ARIES ResinTech Inc reserves the right to make changes in design specifications and prices without notice
11. ck connector from the Gemini water supply Connect the female Gemini supply fitting to the male connector on the sanitizing tank Press the RECIRC KEY to start the pump and recirculate the water in the system Then press the DISP KEY to continuously dispense the recirculated water into the sanitizing tank and draw water back through the unit While the system is recirculating the water measure 40 ml of sanitizing solution and pour it into the wide mouth sanitizing tank Allow the solution to circulate for two 2 hours see step 10 After 1 hour use a sanitizer strip to check for sanitant If sanitant is not present return to step 6 If santant is present continue recirculating for an additional hour When sanitization is complete press the DISP KEY to stop dispensing to the tank and the RECIRC KEY to stop the pump Disconnect the tubing from the tank Remove the tank and dispose of the sanitizing solution Remove the three housing bowls and dispose of the sanitizing solution then replace the bowls Connect the unit to the feed water supply Open the feed water isolation valve slowly and allow the system to fill Press the DISP KEY to purge any air form the unit and dispense approximately 5 liters of water to rinse the system of sanitizing solution Press the RECIRC KEY to start the pump and recirculate the rinse water through the system Allow the pump to run for 5 minutes Press the DISP KEY to dispense another 5 li
12. ents e The dispensing system must be sanitized if HPC tests indicate unacceptable bacteria counts or at preventative maintenance intervals such as annually or semi annually It is recommended to replace sub micron filters and ultrafilters at scheduled intervals not to exceed one year e The basin area below the stainless steel grill should be cleaned regularly to prevent odors even if the BASIN DRAIN KIT or BASE CABINET and WASTE TANK options are installed e If the unit is equipped with the optional WASTE TANK it must be emptied when full e The exterior surfaces may be wiped clean with a mild detergent and a soft cloth to keep the unit clean and new looking 8 1 Changing the UV Lamp WARNING ELECTRICAL SHOCK HAZARD Before removing the top PVC cover disconnect the power cord FAILURE to do so could result in electrical shock or personal injury Hold the UV lamp by the plastic ends ONLY DO NOT touch the glass Finger prints and oils from your hands will etch the glass and reduce UV life amp performance If the lamp should become oily or dirty clean it with a clean soft tissue or cloth wetted with isopropyl alcohol e It is not necessary to turn off the water supply to replace the lamp e For convenience the UV lamp may be turned off by advancing the control to the MAINTENANCE mode However for safety to prevent electrical shock it is advisable to unplug the unit when changing the UV lamp Remove the top cove
13. eps above If the leak persists or other leaks develop in the unit s plumbing please contact customer service see section 11 0 Requesting Service e For Ultrafilter option itis necessaryto bleed of the airin the in line filter Filter is located in the top portion of the system Remove the top cover and bleed air of ultrafilter as stated in section 5 6 e If there are no leaks re install the black front panel bent edge down and facing you the stainless steel basin grill and the PVC top cover with the six 6 phillips head screws ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 5 5 Applying Electrical Supply The electrical outlet supplying the Gemini should be properly grounded and fused and not be able to be switched off during facility vacancies Plug in the Gemini The digital display will come on indicating purity and the A and B LED will alternate flashing 5 6 INITIAL POLISHING of New Cartridges NOTE An initial rush of air will escape the system until the unit is completely filled with water Some splashing may occur Place a beaker or collection vessel on the grill directly below the dispensing nozzle Press the DISP function key to purge the trapped air and to rinse the cartridges and housings Press the DISP function key a second time to stop the flow of water once all the air has been purged from the s
14. erter Combination UV for Bacteria and TOC Basin Drain Kit Wall Mounting Kit Base Cabinet and Waste Tank Pre treatment Packages in Base Cabinet Teflon Spray Gun with Recirculation 0 05 through 0 45 micron Final Filtration Ultrafiltration Sanitizing Kit Teflon and Kynar are registered trademarks Typical Applications Media Prep Tissue Culture Glass Wash DNA NRA testing Liquid amp lon Chromatography Atomic Absorption Physical Properties Dimensions WxHxD Shipping weight Power Requirements 50 8x55 3x47 6 cm 20x21 75x18 75 in 25 kg 55 Ibs 115vac 60hz 105 watts 230 vac 50 hz available Feed Water Requirements Reverse osmosis distillation deionization or tap water that meets the following Conductivity TOC Silt Density Index Pressure Connection Temperature Treated Water Dispense flow Purity TOC reduction Temperature rise Microorganisms lt 150 uMhos lt 100 ppb lt 2 20 50 psi 1 4 3 5 bar 6 3 8 OD pp tubing amp 1 2 MPT ftg lt 80 F 27 C up to 3 7 Ipm 1 gpm gt 18 M cm gt 94 96 lt 5 c 9 F above feed 10 cfu 1000ml ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 5 0 SETTING UP AND USING YOUR GEMINI SYSTEM This section includes instructions for initial installation and start up of the unit Desc
15. fully read labels and warnings on chemical packaging The GEMINI with its multi pass UV protection is virtually bacteria free however it is recommended that periodic sample testing be performed to assure that the GEMINI components continue to control bacterial levels Filters and DI cartridges can be breeding grounds for bacteria whether required or not these cartridges and the UV lamp should be replaced at least once per year to prevent any bacterial growth from exceeding the kill rate of the UV system Should the Gemini dispensing system require sanitizing as determined by unacceptable bacteria levels in the dispensed water or to satisfy regular FDA requirements a sanitization kit is available that allows the user to introduce a sanitizing solution into the system while it recirculates to sanitize all wetted components KIT CONTENTS sanitizing instructions sanitant recirculation tank sanitizing solution with MSDS sheet sanitant test strips and disposable gloves SANITIZATION PROCEDURE 1 FILL SANITIZING TANK Prior to sanitization use the GEMINI to fill the sanitization tank approximately half way with DI water and set the tank aside 2 ENTER MAINTENANCE MODE Press and hold the MODE KEY for five seconds to enter the MAINTENANCE mode In this mode the UV lamp is disabled so that the UV lamp will not degrade the sanitizing solution and both the recirc pump and dispensing valve are set into a toggling ON OFF mode so each
16. irectly at the ultraviolet lamp with unit power applied DO NOT operate any appliance with a damaged or frayed power cord To disconnect the power cord grip the plug and pull it from the wall outlet NEVER PULL ON CORD DO NOT use this appliance outdoors This appliance is intended for commercial industrial institutional and professional use Use of this product in a manner other than recommended herein may void the warranty DO NOT operate this appliance without it being connected to an adequate water supply and purged of air or damage to the pump may occur DO NOT allow the cord to hang over the edge of a counter top or to touch hot surfaces NOTE If an extension cord is used verify that the rating of the extension cord is equal to or greater than the rating of the appliance Refer servicing of a defective or damaged unit to the factory or an authorized dealer The use of attachments not recommended by the manufacturer may adversely affect the water purity or cause appliance damage and void its warranty Avoid contacting the dispensing outlets with any foreign materials or hands to prevent contaminating them ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 2 0 GLOSSARY of TERMS The following terms are used throughout this text Please take a few moments to familiarize yourself with them so that the information and instruction
17. r bacterial removal A 05 micron Pyrogen removal filter is also available HPA 002 185 nm UV combination for Bacterial and TOC reduction Mounted external to the GEMINI unit a stainless steel electropolished UV chamber is plumbed into the recirculation loop The 185 nm wavelength is especially good at TOC destruction HPA 003 185 nm Replacement bulb for the combination Bacterial and TOC reduction HPA 004 254 nm Replacement bulb for the GEMINI Multi Pass system HPA 005 Sanitization kit Kit includes the recirculating sanitant tank sanitant test strips and rubber gloves HPA 009 Sanitization Refill Kit sanitization refill and test strips HPA 006 Wall Mounting Bracket Wall kit intended to be mounted on a wall capable of supporting a minimum of 55 pounds AFK 006 Cartridge kit with 2 Low TOC high purity DI filters and 1 2 micron filter AFK 008 Cartridge kit with 2 Low TOC high purity DI filters and 1 05 hollow fiber filter AFK 009 Cartridge kit with 3 Low TOC high purity DI filters for GEMINI 110U PF 00 6505 In line Pyrogen Ultrafilter 0 05 micron for GEMINI 110U 1200103 Replacement Pump 110 VAC HPL RO Reverse Osmosis Pretreatment system for the Gemini 100 GPD with 14 Gallon Bladder Tank AFK 005 Cartridge kit for HPL RO 5 0 Micron Prefilter GAC 1 0 Micron Post Filter AM 12 7010 Membrane for HPL RO 100 GPD TFC Many other options and customizations are available for the Gemini Please con
18. r by removing the six 6 phillips head screws that attach it to the top platform Gently lift the top cover up and left and set it aside e Grasp the UV lamp by the plastic end on the right side of the UV chamber and pull the lamp to the right until it is removed from the unit It can then be discarded like any light bulb ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 8 1 Changing the UV lamp continued Using the plastic end as a handle carefully remove the new UV lamp from its protective package Holding the lamp so that the 4 metal prongs are on the left slide the lamp into the 3 4 hole on the right side of the UV housing Gently guide the lamp fully into the housing up onto the lamp guide far side and into the socket The bulb may have to be rotated that the 4 prongs match up with the bulb connector Replace the top cover Plug the unit in or turn the UV on by advancing to the NORMAL mode and confirm that the lamp is lit by observing the dispensing spout for a blue purple glow If it is not lit the lamp probably is not seated into the socket Repeat the steps above to assure correct seating 8 2 Replacing the Resin and Filter Cartridges The cartridges are expendable items which have a finite capacity to deionize and polish the water When the digital display indicates the purity is less than your minimum requirements
19. reductions to below 0 05 units Dispenser outlet is continuously exposed to UV light to prevent retrograde contamination within the unit Microprocessor controller keypad and display provide the user with a variety of operating modes system monitoring and flexibility for future enhancements A high resolution digital purity monitor displays the ionic purity of the treated water in Megohms to two decimal places The user never has to guess how pure the water is The built in pump will recirculate the water for the highest available purity ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 4 0 PRODUCT SPECIFICATION Design Features e Up to 3 7 lpm Continuous Flow On Demand Exceeds 18 Megohm Type Water Microprocessor Based Control Touch Key Dispensing 4 User Programmable Batch Dispense Modes Digital Purity Monitor standard e Automatic UV and DI filtration polishing e Natural FEP Kynar or Polypro Fittings amp Tubing to Prevent Contaminants amp Leaching e Silent Magnetic Drive Polishing Pump e UV Protected Dispensing Nozzle e Multi Pass UV for Prevention of Bacterial Growth e Built in Basin standard e Easy Access Replaceable Cartridge Compartment e Provisions for up to 3 Cartridges in Combinations of Deionizing Resins amp Filtration Options amp Accessories 230 volt Operation power conv
20. ribed herein are selection of unit location unit unpackaging and connection of unit to the water and electrical supplies Also described are UV lamp installation water treatment cartridge installation purging of trapped air and initial flush 5 1 Setup 5 2 Select a location for the unit which is in close proximity to the water supply and electrical supply connections If the unit has been purchased with the BASIN DRAIN option locate the unit near a sink or waste connection If the unit has been purchased with the WALL MOUNTING option select a wall which will support a 55 pound load on 16 inch stud centers or provide wall backing material If the water supply does not already have an isolation valve for servicing the Gemini install one ahead of the unit connection Carefully remove all of the packaging material and tape used for shipping purposes and position the unit in its operating location Using the water supply connection kit provided with the unit connect the unit to the supply DO NOT open the supply isolation valve at this time Installing the Water Purifying UV Lamp WARNING Hold the UV lamp ONLY by the plastic ends DO NOT touch the glass bulb Finger prints and oils from your hands will etch the glass and reduce UV life amp performance If the lamp should become oily or dirty clean it with a clean soft tissue or cloth wetted with isopropyl alcohol Remove the top cover by removing the phillips head
21. s within are clear TERM LED LCD TOC TDS SDI HPC DESCRIPTION Light Emitting Diode is a semiconductor indicating device used for the status indicators on the operator control panel of this unit Liquid Crystal Display is a semiconductive display device used in the operator control panel of this unit Total Organic Carbon The identification of organic carbon in water is helpful to assure successful water system operation Trace organic chemicals can interfere with highly sensitive reactions and alter biological processes Total Dissolved Solids conduct electricity This conductivity allows you to measure the concentration of salts directly in parts per million ppm Higher TDS increases the load on the deionizing cartridges Silt Density Index is a scale for determining the amount of suspended solids in water that exceed 0 45 microns in size A regulated flow is passed through a 0 45 membrane and the flow time and pressures are plotted to arrive at the index number Higher SDI numbers result in premature fouling of resin and filters Heterotrophic Plate Count formerly known as the standard plate count is a procedure for estimating the number of live heterotrophic bacteria in water and measuring changes during water treatment and distribution ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 3 0 GENERAL DESCRIPTION
22. sult with the Dealer or Factory for additional options 11 0 REQUESTING SERVICE Before calling the factory to request service search the troubleshooting guide below for symptoms which your unit exhibits and verify the cause and possible remedies listed If your symptom is not listed or the remedy does not provide a cure Contact 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com Email aries resintech com and ask for a customer service representative Be prepared to provide the representative with the model number serial number date of purchase and any options that may have been purchased with or added to the unit after purchase as recorded on page 1 of this manual Next be prepared to describe to our service representative the symptoms that the unit exhibits TROUBLESHOOTING GUIDE SYMPTOM CAUSE REMEDY Display is off No power to the unit Check the unit is plugged into a live outlet Check the unit s fuse replace it if itis blown Unit blows fuses An electrical short exists in the Return the unit to the factory unit for repairs Unit will not dispense water Supply water is not on Confirm that the supply water is turned on The purity is below the set Allow the purity to rinse up to point above the set point or replace DI cartridges A filter cartridge is loaded with Replace the filter cartridge particles and plugged System is bound
23. ters of rinse water Use the sanitizer residual test strips supplied with the kit to verify that no residual sanitizing solution is being dispensed and rinse if necessary After all traces of sanitizing solution are gone stop the pump close the supply valve and press the DISP Key to depressurize the system Remove the housings and re install the cartridges Refer to sections 5 3 and 5 4 for the proper cartridge loading and purging air information Press the MODE KEY to advance to the NORMAL operating mode the A LED is on amp the B LED is off ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 10 0 OPTIONS amp ACCESSORIES The following options amp accessories are supplied as kits and include detailed instructions for assembly onto the basic design The use of attachments or accessories not approved by this manufacturer may adversely affect the water purity or cause appliance damage and void the warranty For questions regarding these and other options amp accessories that may be available for system specific needs or for components to aid installation please contact an authorized dealer in your area or contact the factory HPA 001 Dispensing gun and tubing kit Fully recirculating gun for remote dispensing with concentric coiled tubing Gun is equipped with a standard 2 micron final filter fo
24. the water purifying cartridges selected for your system They must be installed in the correct sequence to perform properly water flows left to right The left and center cartridge positions always contain either AF 10 4010 or AF 10 4011 deionizer filters The right position will contain either the submicron filter or another DI filter depending upon the system options and setup e Drop the filter into the housing bowl Check that the O ring is on the upper seat inside the bowl the O ring must be correctly in place or it may become pinched or cause the housing to leak Align the bowl and cartridge under the left 1 position so that the cartridge top opening slips over the housing center post Gently thread the bowl to the right onto the housing Use the filter wrench supplied with the unit to gently snug the housing closed Install the right 3 cartridge using the same care as above e Install the middle 2 cartridge using the same care as above 5 4 Applying Water Pressure and Bleeding Air e Slowly turn on the water supply to the unit and correct any leaks at the supply connections Water will fill the left 1 bowl then the middle 2 bowl and the right 3 bowl in sequence While the pressure on the unit slowly increases to supply pressure inspect bowl seals for leaks If a leak develops turn off the water supply release the pressure from the system and then loosen the bowl to inspect the o ring and reseat then repeat the st
25. to prevent contamination via airborne bacteria A continuously cycling recirculation pump polishes the water assuring the highest quality water is always available and a high resolution digital monitor is built in for continuous display of water purity A microprocessor based controller and operator keypad provides the user with several dispensing modes while controlling the recirculation pump monitoring overall system performance and providing system diagnostics An easily accessible module compartment makes cartridge replacements a simple task The built in basin keeps any drips or spills under control Optional pretreatment systems are available to operate from tap water that does not meet feed water specifications They can be packaged into the optional base cabinet Reverse osmosis distillation or deionizing pretreatment may be necessary depending on the quality of your tap water for optimum performance of 3 1 PRINCIPLES OF PURIFYING WATER e Multi Pass Ultraviolet Irradiation is used for organic reduction amp bacterial reduction Low TOC Mixed Bed Resins are used for polishing purity levels from pre treated filtered RO distilled or DI water to 18 2 megohms resistivity Microporous Resins are used if necessary for removing colloids and silica from supply water Absolute rated cartridges for critical applications requiring particle retention of 0 05 through 0 45 micrometers Ultrafiltration for pyrogen
26. ton for 5 seconds to enter maintenance mode e LED B flashes indicating that the unit is in the Maintenance Mode e All unit operations are off upon entering the maintenance mode e The UV light LED FL is off at all times in the maintenance mode e DISP key toggles solenoid valve on and off e RECIRC key toggles pump on and off ARIES a division of ResinTech Inc 160 Cooper Road West Berlin NJ 08091 Phone 856 768 9600 Fax 856 768 9601 www arieswater com 6 0 DISPLAY AND KEYPAD FUNCTIONS Be sure you understand the function of each key before using them e Function Keys DISP Single key DISPENSING In NORMAL operating mode pressing this key will start the polishing pump and DISPENSE ultrapure water RECIRC Press to start stop the POLISHING pump when the unit is in MAINTENANCE SERVICE mode 1 2 3 amp 4 User programmable batch dispensing modes FLUSH For future enhancements MODE Press to step through operating modes Push and hold for 5 seconds to enter the SERVICE mode e LED Indicators DI on when the DISPENSING solenoid valve is energized R1 on when the POLISHING pump is running A B C indicates which MODE the unit is in STATUS MODE ALTERNATING FLASHING INITIAL POWER UP POWER LOSS NORMAL OPERATING MODE IB FLASHING SERVICE MAINTENANCE MODE FLASHING BATCH PROGRAMMING MODE FL on when the UV lamp in energized R2 on when in the
27. with air Bleed off canister amp utlrafilters continued next page 11 0 TROUBLESHOOTING GUIDE continued SYMPTOM CAUSE REMEDY Display indicates water purity is below set point Resin cartridges exhaust quickly Sub micron filter plugs rapidly Bacteria exceeds operating specifications Bacteria exceeds operating specifications Pyrogen levels are excessive after the optional ultrafilter The supply water is improperly treated The resin cartridges are not seated Resin is exhausted The supply water is improperly treated The supply water has colloids and is improperly treated UV lamp is not working The system is fouled with contaminants The feed water supply is grossly fouled Test the supply water TDS add additional treatment Open cartridge housing and reseat cartridge Install fresh cartridges Test the supply water TDS Add additional pretreatment to obtain higher quality Perform a Silt Density Test If it is higher than 1 add additional pretreatment Confirm that the lamp is seated in the socket Confirm that the lamp is not burned out Confirm that the lamp has not been in service for more than one 1 year Replace the lamp with a new one Autoclave the sub micron cartridge Sanitize the feed water supply Replace the resin cartridges with fresh new ones The ultrafilter is contaminated Sanitize the UF 12 0 REAGENT GRADE WATER
28. ystem For Ultrafilter option top cover should be removed In line ultrafilter should be visable with two side port plugs To bleed air in the ultrafilter unit should be in the dispensing mode With a towel present slight loosen the coarse screw plug to bleed off air Air should come out of the port with a small amount of water Bleed until all air is removed and tighten plug Note avoid over tightening plug as it is a coarse thread Remove power and clean up any water that is present Re install top cover and power up unit and proceed Press the MODE function key The A LED will light continuously the polishing pump R1 LED and the UV light FI LED will come on The unitis in the INITIAL POLISHING mode and the pump will run for 2 hours This will allow the system to polish and rinse the system components with high purity water After recirculating for two 2 hours the system will automatically change to its NORMAL operating mode During the NORMAL operating mode the polishing pump and the R1 LED will come on every 10 minutes and run for 2 minutes minimum pump running time to polish the water and keep it up to the highest purity Then the pump will shut off and relax until the next cycle In the NORMAL operating mode pressing the DISP key will initiate the polishing cycle the pump comes out of the relaxed state and the Gemini will dispense ultrapure water as indicated by the DI LED Press the DISP key a second time