1. DEVON MEDICAL PRODUCTS CircuFlow 5200 Quick Setup Guide www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 LB04 0001 Rev B 20140314 DEVON MEDICAL PRODUCTS Setup CCS TABLE OF CONTENTS Therapy Setup 4 Gradient Mode 5 Pressure Mode 08 Lock Device __________ 1 Unlock Device 1 1 Treatment 1 13 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS SETUP 01 Open the shipping box and lift device up and out of box 02 Check to make sure the box also contains 1 power cord 1 air blocker and a user manual 03 Remove the protective foam covers on each side and the plastic bag Not Pictured 4 Place the device on a flat and stable surface near the patient 05 Unwrap the power cord from the box and plug it into the back of the machine Plug the device into a safe and properly grounded outlet www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS TH ERAPY SETU P Settings can only be modified or restored before or in between treatments There are three operating modes with this device e Factory Default Mode Device comes with a factory default setting providing 45mmHg pressure on distal chamber and a
2. DEVON MEDICAL PRODUCTS GRADIENT MODE 1 Upon this display the user can access GRADIENT MODE by pushing the SHUTTLE KNOB circular button This will take you to THERAPY MODE SETTINGS The indicator light will be yellow setting 1 1 This display shows THERAPY MODE SETTINGS Note The asterisk shows which mode is currently selected 1 Turn the SHUTTLE KNOB to the right to select GRADIENT MODE Then push the SHUTTLE KNOB to access GRADIENT MODE It will take you through various parameters PRESSURE GRADIENT TREATMENT TIME CYCLE TIME AND SKIP CHAMBER 1 This display shows the first parameter PRESSURE turn the SHUTTLE KNOB to the left Push the SHUTTLE KNOB to select PRESSURE Note User can access other parameters by turning the SHUTTLE KNOB to the right left at any time 1 You can increase the pressure by turning the SHUTTLE KNOB to the right and decrease the pressure by turning to the left pressure can range from 25 120 mmHg increments of 1 mmHg 1 b Push the SHUTTLE KNOB to confirm the new pressure www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS GRADIENT MODE continued 1 b Access the second parameter GRADIENT by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select GRADIENT 1 You can increase the gradient by turning the SHUT TLE KNOB to the right and decrease by tu
3. mmHg increments of 1 mmHg 3h Push the SHUTTLE KNOB to confirm the new pressure This will take you to the next chamber Follow step 33 to set the pressure in that cham ber Repeat until you have reached Chamber 4 08 www devonmedicalproducts com e 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE continues 3 After chamber 4 the screen will read EXIT Push the SHUTTLE KNOB to confirm 3 Access the second parameter TREATMENT TIME by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select TREATMENT TIME 30 You can increase the treatment time by turning the SHUTTLE KNOB to the right and decrease time by turning to the left treatment time can range from 15 180 minutes increments of 15 minutes 39 Push the SHUTTLE KNOB to confirm the new treatment time Af Access the third parameter CYCLE TIME by turning SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select CYCLE TIME 41 You can increase the cycle time by turning the SHUTTLE KNOB to the right and decrease time by turning to the left cycle time can range from 20 80 seconds increments of 10 seconds 4 Push the SHUTTLE KNOB to confirm the new cycle time www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 09 DEVON MEDICAL PRODUCTS PRESSURE MODE ineo 43 Access the fourth param
4. not fill with air Push the SHUTTLE KNOB to confirm the skipped chamber To exit and complete GRADIENT MODE setup turn the SHUTTLE KNOB to the right Push to select EXIT Turn the SHUTTLE KNOB to the right in THERAPY MODE SETTINGS until you see EXIT push SHUTTLE KNOB to select and you will be taken to the Ready Press lt Start gt display Push the green START button to start your custom treatment proceed to step 48 2 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE 3 Upon this display User can access PRESSURE MODE by pushing the SHUTTLE KNOB That will take you to THERAPY MODE SETTINGS 3 1 This display shows THERAPY MODE SETTINGS Note The asterisk shows which mode is currently selected 3 Turn the SHUTTLE KNOB to the right to select PRESSURE MODE Then push the SHUTTLE KNOB to access PRESSURE MODE it will take you through various parameters PRESSURE TREATMENT TIME CYCLE TIME AND SKIP CHAMBER 33 This display shows the first parameter PRESSURE SET Push the SHUTTLE KNOB to select PRESSURE SET then push the SHUTTLE KNOB on the individual chamber you want to set Note Pressure can be set in all 4 chambers 34 You can increase the pressure by turning the SHUTTLE KNOB to the right and decrease the pressure by turning to the left pressure can range from 25 120
5. 7 of gradient pressure and a 30 minute treatment time with a 30 second cycle time with no chambers being skipped e Gradient Mode This mode allows you to set your starting pressure and the percentage drop of pressure gradient between chambers e Pressure Mode This mode allows you to set the pressure in each of the four chambers in the sleeve Follow steps 6 8 before pressing the green START button 0 Attach left and or right quick connectors of the garment to the quick connect ports which are located on the front panel of the device During a single gar ment session i e left only or right only insert the air blocker into the unused quick connect port 07 Putting the garment on a For leg garments unzip the garment all the way to the end Place the foot at the bottom of the garment and pull up the zipper while supporting the garment to wrap around the leg b For arm garments slide the arm through the internal cavity of the garment 08 Press main POWER SWITCH up to the ON position which is located on the rear panel The green power indicator on the front panel will then illuminate 00 Upon this display User can directly push START to run the device on FACTORY DEFAULT mode e f GRADIENT MODE is desired continue to steps 10 29 e f PRESSURE MODE is desired proceed to steps 30 46 m www devonmedicalproducts com e 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092
6. STOP the treatment Will not pick up from where it left off at b After the treatment the device will vacuum air from the garment for 1 minute so that it will facili tate removal of the garment Not Pictured Note If using one of the larger size sleeves all the air may not vacuum out within 1 minute The vacuum setting can go through another 1 minute cycle by pressing STOP b Once the sleeve is deflated press the power switch on the backboard to the OFF position and then unplug the power cord b Once the power indicator light is off it is safe to remove the garment Not Pictured b The garment should be loose enough now that you can unzip the garment and remove Not Pictured www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS NOTES www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS NOTES www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 E DEVON Innovation in Medical MEDICAL PRODUCTS Products Engineering
7. eter SKIP CHAMBER by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select SKIP CHAMBER AA Select to the chamber you wish to skip by turning the SHUTTLE KNOB to the right or left Select one only Chamber 1 Chamber 2 Chamber 3 Chamber 4 or No skip Note Skipping a chamber does not mean the chamber selected to skip will not fill with air but less air Ah Push the SHUTTLE KNOB to confirm the skipped chamber 46 To exit and complete PRESSURE MODE SETUP turn the SHUTTLE KNOB to the right Push to select EXIT 4 Turn the SHUTTLE KNOB to the right in THERAPY MODE SETTINGS until you see EXIT Push SHUT TLE KNOB to select and you will be taken to the Ready Press Start display Push the green START button to start your custom treatment 10 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS LOCK DEVICE 4 Upon this display User can access Lock Device by pushing the SHUTTLE KNOB That will take you to TherapyModeSettings The indicator light will be yellow setting 40 This display shows TherapyModeSettings Note The asterisk shows which mode is currently selected P After setting device to desired treatment turn the SHUTTLE KNOB to the right to select Lock De vice Then push the SHUTTLE KNOB to access Lock Device b 1 A second screen will appear to verify that you would like to
8. lock the device Move the cursor over to Yes and select using the SHUTTLE KNOB b Another screen will appear to notify you that the device is now locked This screen will also show the password to unlock the device and then the revert back to the Start Menu Password is DM for Devon Medical www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 Hid DEVON MEDICAL PRODUCTS UNLOCK DEVICE b In order to unlock the device press down the SHUTTLE KNOB Instead of taking you into the menu a screen will appear to Input Password HA Turn the SHUTTLE KNOB to select the first char acter of the password press down and select the second character Press down the SHUTTLE KNOB a second time to input the password If the password is correct the device will unlock and return to the menu screen Hh If the password is incorrect the device will display a screen to notify you and return back to the start menu www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS TREATMENT b The pump will go through its initial prefill stage of reaching 10 mmHg in all 4 chambers Each treatment will end after its set treatment 5 time has elapsed The user can also end the treatment at any time during a treatment session by pressing the STOP button on the front panel Note If the user presses
9. rning to the left gradient can range from 5 15 incre ments of 1 1 1 Push the SHUTTLE KNOB to confirm the new gradient 1 0 Access the third parameter TREATMENT TIME by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select TREATMENT TIME 20 You can increase the treatment time by turning the SHUTTLE KNOB to the right and decrease time by turning to the left treatment time can range from 15 180 minutes increments of 15 minutes 21 Push the SHUTTLE KNOB to confirm the new treatment time 22 Access the fourth parameter CYCLE TIME by turning SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select CYCLE TIME 06 www devonmedicalproducts com e 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS GRADIENT MODE continues You can increase the cycle time by turning the SHUT TLE KNOB to the right and decrease time by turning to the left cycle time can range from 20 80 seconds increments of 10 seconds Push the SHUTTLE KNOB to confirm the new cycle time Access the fifth parameter SKIP CHAMBER by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select SKIP CHAMBER Select to the chamber you wish to skip by turning the SHUTTLE KNOB to the right or left Select one only Chamber 1 Chamber 2 Chamber 3 Chamber 4 or No skip Note Skipping a chamber does not mean the chamber selected will