Home

HYDROSENSE 2410 User Manual - Can

image

Contents

1. 4 0 CALIBRATION CALIBRATION MUST BE PERFORMED AFTER INSTALLATION AND ANY LAMP REPLACEMENT Allow a minimum of 24 hours of power up prior to calibration to allow the lamp to stabilize Using actual process stream flows for the calibration is desirable to provide the best accuracy This will zero out any background influences that may be present in the process water At least two points of entry at different contaminant levels are required These levels may initially be unknown for calibration purposes and confirmed to the instrument after laboratory results are returned One of these points may be clean process water 0 PPM hydrocarbons Prepared samples may be used for calibration and testing although variations in personal blending techniques the source of the hydrocarbon and background water the container used and the retention time prior to use will all play a part in the concentration reliability and repeatability 4 1 BEFORE CALIBRATION Open the sample chamber door Lift the Lamp Blue Box off the tray and place it on the door brackets Check that the flow tile is resting securely in place against the tray The flow tile has a special reflective core and MUST be placed in the tray with the etched UP side facing out toward you Slowly open the process stream valve to allow a steady and even flow over the flow tile Check that the outfall is draining well and water is not building up in the drain tray Verify that the flow tile is evenly
2. X ZL Jois Bugunow 1dN4 SI jeiu ejduies Lal z Z X LdNd v jemno Jeubis Y JOU JOMOd seoeld Z dA 1 191 29 ajoH BununoN MALA AdIS 1HOIY UOHEIQIED pue 15981 JOJ LANA 8 1914 J9JEM Uea g NIA 3GIS 1331 LdN3 8 1e u ajdwes sseooJg y soo peiddns eA eA eM Z uuuu pue seuour u ae suoisu9euuip IV SON Lpz s 6 soe zL MALA LNOHJ N IAJ AO UJejs S uoneiqileO pue sa JO lul J9jeM Ue9J2 oBing pue esuesoJpAH neque pue 1591 JOJ dNA 8 19141 J9 E Ma SR ldN4 8 1 lul ejduies sse9old V esoo poljddns aA eA eM Z GALINII wu pue sayoul ui ee SUOISUSLUID Ily L 3SN3SOHGAH ONIHHANIDNA 9 ON 1O3POYd AVIAV TT o T TO pavo NOSIS V avo aa se eld Z dAL AE X ZIL 10 s Bununoj Jojeoipu eBung 1dN4 8 Jeu ajduwes 466 Sc 6 Z SON 99S Lg z 1dN4 81 p 1 3e u Ajddns Z X LdNd v jepno eubis 3 1 u 1SAOd seoeld z dAL 9L e m loH Bununoyy HOO LNOHLIM ve s6 NIA INOX MAA 3GIS 1H I3 MIA sais 1331 soe zu NIA INOUS 02c c8S00V Hed em OMV 81 BuulM Dat Joj Buiuioou OZZ Z8v00V Hed SEN punoig uren ounen OVA0ZZ su soup H Fu Asoc ss IST Ms JSNJSDATAH GALINII seuosseooe ou sa 28 64 ge JuBI AA JON ONIHAANIDNA TT IVDIV tens XVIN WA 002 uonduinsuo 1emog 240 INN Sms p puno ule eApoetoid JOJ uoneinsu d Ravaxo
3. This filter narrows the HydroSense response to only those that fluoresce from 254 nm 3 This limited number of compounds that do fluoresce from 254 nm light may emit light at any number of wavelengths such as 290nm 310 nm 350 nm 480nm etc Aromatic hydrocarbons happen to fluoresce at approximately 350 nm By filtering the light sensor from all light except 350 nm only compounds that emit light at 350 nm are indicated at the receiver 4 Oiland Grease in water may be made up of hundreds or thousands of different hydrocarbon compound structures The aromatic compounds tend to be the fluorescing compounds The proportion of aromatics within the total hydrocarbons is generally consistent in a product or process The aromatics are therefore used as an indicator to correlate the monitor to total hydrocarbons in water Changing Oil Types and Sources Different oils have a different make up of compounds and the fluorescing strength may vary between oil types For instance diesel fuel may fluoresce much stronger than transformer oil If the HydroSense is calibrated using 100 PPM of diesel 100 PPM of transformer oil may only give a display reading of 95 PPM Crude oil may vary from one well to another lubricating oils from different manufacturers may vary in their make up oils may be dissolved or free and so on The calibration is therefore site selective and should be done using actual process water or with samples of oil that are
4. t ee etr er Ret i eres 11 3 3 MINIMUM SETUP lt CONTROL gt Key emm 12 3 3 1 Operation Range 4 20 mA Gettng 12 3 3 2 Confirm 4 20mA output direction 12 3 3 3 Relay Setpolnts x tette ki Ne pte detenta ts 12 3 3 4 Relay Time Delay it Re e Pen Re EGRE A SE 12 3 3 5 Relay Enable m u y tue geen del td ee jeta 13 4 0 CALIBRATION ae iid tec pite beUi 14 4 1 BEFORE CALIBRA TON un ette HERD elt eas 14 4 2 CALIBRATION Secs rotate erit eae ain etal ett rate 14 4 2 1 Automatic Calibration n snssssssssssssssssssa 14 4 2 2 Manual Calibration With Laboratory Results 15 4 2 3 Gain Setting visi ip uu nee eut ede valove ja ede eda ljala rejce 15 4 2 4 Light Reference and Temperature Setting 15 4 2 5 Change calibration location ooooooonnccccnonnciccnonacccccnnannncnnnnarnnr nano ncnrnnnn nn 15 5 0 OPERATION 8 DIAGNOSTICS ana 16 5 1 AMPLIFIED SIGNAL uan rl Ee 16 5 2 UNAMPLIFIED SIGNAL Su ans ii nine 16 5 3 DIAGNOSTIGS i at dei m ici nita ne 16 54 MIN T MA escisiones fete 16 5 5 SETTINGS omen menm en ment 16 6 0 CONTROLLER SETTINGS SHEET see een eene 18 7 0 SAMPLE PREPARATION For Calibration or Testing 19 8 0 PERIODIC TESTING AND MAINTENANCE A 20 8 1 PERIODIG TESTINQ z ite ence teet tee eene eal Avge das Ate 20 8 2 FLOW TILE CLEANING ana 20 8 3 EAMP REPEAG
5. a position that will receive a continuous representative sample from the process stream The farther from the stream the greater the lag time of readings vs actual process variance Mount the enclosure on a vertical wall or support of minimal vibration The sample will be flowing over a flow tile A bubble level will assist in this installation Be sure the unit is mounted level This is necessary for a proper flow across the flow Tile Open the Lamp Blue Box and secure the new lamp into place with the metal band and knurled screw Plug the lamp terminal into the socket Close the Lamp Blue Box The flow tile has a special reflective core and must be placed in the Tray with the etched UP side facing out toward you Place the Lamp Blue Box against the Tray and secure it with retaining latch Be sure the 3 way valve supplied loose is installed as per drawing 20020019 or 20020019A Make sure the inlet sources process sample and clean water are valved for future maintenance and testing operations Connect the inlet sources to the 3 8 inlet of the 3 way valve DO NOT TURN ON THE INLET SOURCES Connect the outlet source to drain or a sump For non purged pressurized models the outlet of the tube should be open to air not submerged in water or a process which would cause a backpressure This would result in an overflow inside the chamber For purged pressurized models a slight and constant backpressure is required to maintain a ca
6. and press lt ENTER gt The unit will prompt you to enter a FLR value in mV This was recorded for you when you first took the grab sample Press lt ENTER gt to accept this Press ENTER to enter the 277 point value Key in the lab result in PPM of Sample 2 And press lt ENTER gt The unit will again prompt you to enter a FLR value that corresponds to your sample Press lt ENTER gt to accept this value Press lt ENTER gt to continue calibration and repeat the above procedure for each sample Press lt 0 gt to finish calibration then press lt DISPLAY gt to main display menu Calibration is completed 4 2 3 Gain Setting The FLR signal gain is typically factory set and should only be modified if directed by an Arjay Engineering representative The procedure to change the gain is as follows Press the lt CALIB gt key then press lt 4 gt to Set Gain Enter the new gain value then press lt ENTER gt Note the actual gain set may differ than the desired value due to the resolution of the gain setting circuitry The actual gain will be displayed if the gain setting menu is re entered Press DISPLAY to main display menu 4 2 4 Light Reference and Temperature Setting With automatic calibration the light reference and temperature are automatically recorded REF CAL TEMP CAL REF CAL is used to compensate the FLR as the lamp ages The procedure to change these values is as follows Press the lt CALIB gt key t
7. condition cleared if PPM is below this value Alarm Enable Master Alarm Relay Enable Off prevents relays from reflecting the Alarm condition The relays will remain in the non alarm condition based on the Fail Safe switch setting on the relay modules Filter Moving Window Filter tracks the average of the last N samples Higher values provide more stable readings Max 1000 Temperature This value is set at the factory The value may be Coefficient altered at any time via the keypad when authorized by Arjay Representative Alarm Delay Amount of time in seconds the PPM value must be above the Hi Setpoint for the Alarm condition to activate Maximum 99 seconds Adjustable gain on the unamplified FLR SIGNAL same as Should only be modified if directed by an Arjay fixed gain Representative 18 Model HS2410 HYDRO39 DOC Rev 3 9 7 0 SAMPLE PREPARATION FoR CALIBRATION OR TESTING Samples may be prepared in a number of ways 1 An unknown concentration may be used to provide a response test This does not verify the accuracy or calibration of the instrument but does confirm that it will respond and alarm to a high concentration condition An unknown sample with a concentration of contaminant can be sent to a lab for analysis A concentration blend may be made using the stream fluid and a known concentration of contaminant If a sample concentration is to be made the contaminant must be made to emulsify in the stre
8. details 3 3 MINIMUM SETUP lt CONTROL gt Key 3 3 4 Operation Range 4 20 mA Setting Determine the desired operating range of the instrument This will reflect the 0 to 100 OUTPUT display on the main menu and the 4 to 20 mA output range of the instrument For example if the operating range of the process is 0 to 30 PPM a range of 0 to 50 PPM may be desirable The display will show 0 50 PPM equal to 0 100965 The control signal of 4 20 mA will represent 0 50 PPM Press CONTROL on the keypad then press 2 for 4 20 mA Settings The display will prompt you to enter the zero point in PPM This will typically be 0 0 Enter the value and press lt ENTER gt The display will prompt you to enter the span value i e 50 0 Key in the desired value and press lt ENTER gt 3 3 2 Confirm 4 20mA output direction The display will require a confirmation if the 4 20 mA signal is to be direct acting 4 20 mA 0 100 or if inverse acting is required 20 4 mA 0 100 The lower left display will read mA DIR If this is required press lt ENTER gt If inverse is required press 1 3 3 3 Relay Setpoints Four relay alarm points are available for remote alarm Of these two are general purpose alarm relays with user settable alarm points dead band differential alarm points and time delay The remaining two relays are to indicate lamp failure R4 and flow tile maintenance R3 If the general purpose alarm relays are to be used
9. ensures a representative and stable snapshot of the water conditions The unit can tolerate suspended solids up to 400 mg l response curves overflow baffle de aerates bubbles Available accessories include air pressurization The backlit 4 line display provides easy menu purging systems sample coolers and pumps driven commands for set up calibration and diagnostics ppm Oil in Water Monitor E Performance E Technical Specifications Control Unit The performance is based on the site Range user selectable 0 10 to 0 300 ppm linear calibration to a known hydrocarbon trending from 300 ppm to 1 000 ppm concentration in stable background water Display Resolution 0 1 ppm Changes in hydrocarbon make up and Instrument Accuracy 0 1 ppm background stability may affect the Process Accuracy 1 0 ppm under stable conditions output Through a simple calibration this Ambient Operating Temp 10 C to 50 C unit correlates well with laboratory ISO Ambient Process Temp 0 C to 40 C and EPA methods optional cooler for temperatures gt 40 C Power Input 24 vdc or 110 vac or 220 vac Alarm Relays 4 x 10 amp SPDT dry Output 4 20 mA Isolated Interface RS 485 Modbus optional HART Standards UL CSA CE ABS CSA Div 2 T3C Groups A B C D Zone 2 Pressurization Purge available for use in Zone 2 NFPA ATEX Enclosure 316 SS Type 4X IP65 HydroSense 2410 gt o c 2 S o v K 5 Laboratory Spectrophoto
10. EMENT cidcid e eer dE 21 Model HS2410 HYDRO39 DOC Rev 3 9 SPECIFICATIONS Power Display Alarm Setpoints Relay Time Delay Range Accuracy Resolution Calibration Signal Output Communication Output Signal Filtering Process Temperature Environment Temperature Temperature Comp Maintenance Alarms Sample Inlet Sample Outlet Inlet flow rate Inlet pressure Enclosure Rating Enclosure Dimensions Pressurization Purge Shipping Weight 115 VAC 50 60 Hz 150 VA including lamp power supply 10 VA controller only optional 220 VAC or 24 VDC four line LCD with simultaneous display in PPM Cal Location current temperature and bar graph two independent SPDT 7 amp 120VAC dry relay contacts with LED panel indication 0 99 second delay on 0 300 ppm Hydrocarbon in Water linear up to 1000 ppm trending 1 PPM The performance is based on a calibration to a known hydrocarbon concentration and a stable background water The hydrocarbon make up and variations in the contaminating compounds may affect the actual output 0 1 PPM 10 Locations Up to 5 concentration entry points per location 4 20 mA DC 900 ohms isolated RS 485 modbus or optional HART protocol 20 1000 samples average 0 100 C with cooler 0 40 C without cooler 10 40 C without Air conditioning 10 55 C with Air conditioning compensation 10 40 C ambient air 2 x SPDT 7 amp 120 VAC dry One for lamp fail
11. HYDROSENSE 2410 On line Oil in Water Monitor User Manual Arjay Engineering Ltd Oakville Toronto Canada L6H 6C9 Tel 1 905 829 2418 Fax 1 905 829 4701 North America 1 800 387 9487 ARJAY www arjayeng com arjay arjayeng com The HydroSense 2410 is the engineered combination of three unique designs by Arjay The sensing chamber contributes a continuous controlled water sample while the optical sensor package hovers above the passing stream The Arjay controller then monitors the multiple signals to provide a reliable ppm concentration output non contacting sensor design minimizes system maintenance fluorescence technology is selective to petroleum hydrocarbons by targeting their aromatic faction continuous on line monitoring without chemicals or lag time To ensure a long term and reliable operation a routine maintenance schedule should be implemented Arjay has made this operation quick and easy The following are a few of the standard maintenance and design features built into every unit The sensing chamber is hermetically sealed from the upper control unit This not only keeps the control components dry but also allows keypad and wiring access without opening the sensing chamber Operator clean time is less than 2 minutes and no tools are required for any procedure Power or sample flow do not need to be turned off Lamp replacement is equally easy A sensor independently m
12. L SNOILISNNOD i TIOALNOI JSNJSOATAH 30 MIA JdIS IHDId WILIVA WwO9LL AOS2 CAVI3A OZ1 vL670OV Hed Je O 0u02 SSUSSOIP H SO 3SN3SOHQAH HOS AIdANS YAMOd xe vlog indino Age eBung ON Y ZHE9O LV OVAOZL OLL indu Jamog abund TVNOILdO NI ZH09 06 INdNI vse NeTJI NO A A G90GAH Hed JexeoJg NIJO JaMod Vp0 TJ 3t Hed A ddns Jewog due
13. am fluid This can be difficult as the concentrate often will not break down enough to provide an even distribution in the low PPM range For example a sample of 50 PPM parts per million is egual to 1 oz in 20 000 oz 125 gallons To use an amount of stream for the base that is manageable such as one gallon a syringe is reguired to inject a small enough amount of the concentrate to make 50 PPM This droplet of oil will tend to separate float to the surface A typical approach to making a concentration is as follows Acquire 10 liters of 0 PPM process water in a glass jar plastic containers will draw the hydrocarbons out of the prepared sample Separately thoroughly mix 1 ml of the sample oil type of oil to be found in stream with 1 ml of acetone The oil will readily mix with the acetone and the acetone will act as an agent to distribute the oil throughout the water Thoroughly mix the oil acetone sample with the 10 liters of O PPM process water This will make a 100 ppm sample This is not a precise method Standing time the volatility of the oil operator measurements and equipment will all contribute to errors in the blend This should be used quickly and only when actual process conditions cannot be used for calibration 19 Model HS2410 HYDRO39 DOC Rev 3 9 8 0 PERIODIC TESTING AND MAINTENANCE CALIBRATION MUST BE PERFORMED AFTER INSTALLATION AND ANY LAMP REPLACEMENT 8 1 PERIODIC TESTING The HydroSense
14. and completely covered with the flowing water The sample flow may not initially cover the whole tile by itself Stir up the flow tile by rubbing the flow tile so the water sheens evenly and completely across the flow tile as it flows Place the Lamp Blue Box back against the tray and secure it with the retaining latch Close the chamber door Make sure the Sample Chamber Door is closed tightly The door MUST be closed to eliminate any background light interference during and after calibrations 4 2 CALIBRATION For Calibration the Hydrosense2410 can accept up to 5 sample points to draw a calibration curve A calibration curve is used because some samples may not be linear as concentrations increase 4 2 1 Automatic Calibration Press the lt CALIB gt key then press 1 for automatic calibration Press ENTER to enter ppm value of the 1 point Normally the e point is clean process water 0 0 ppm Manually put the handle of the 3 way valve to up position and let the clean process water going through the Hydrosense 2410 Enter the ppm value 0 0 Once the reading FLR rdng has stabilized press ENTER key A different concentration of contaminated water is required If the process stream has a different concentration than 0 ppm put the handle of the 3 way valve to down position to allow the process stream to go through the Hydrosense 2410 If the concentration of process stream is close to 0 ppm a prepared sample refer to Se
15. binet pressure To accomplish this submerge the effluent tubing approximately 10 cm below the water surface or run the outlet tube with a slight upward grade prior to the downward outfall Be sure the tube never rises higher than the bottom of the chamber outlet Note that this is gravity feed line only The outlet tube must NOT rise above the chamber height or be connected to a pressure process Confirm that the tray and flow tile are securely in place Slowly open the inlet water source to allow a steady and even flow over the tray and onto the flow tile Once the flow connections are verified and operational turn the water source off to continue with the electrical installation Model HS2410 HYDRO39 DOC Rev 3 9 2 4 1 4 Oo C 7 2 5 PERMANENT POWER CONNECTION AC POWERED MODELS ONLY Connection to the building wiring system shall be in accordance with the Canadian Electrical Code CEC Part 1 in Canada the National Electrical Code ANSI NFPA 70 in the USA or the local electrical codes of the country where the equipment is being installed An external mains switch or external over current protection circuit breaker device is required as a disconnect device This mains disconnect device shall be specified as complying with the requirements of IEC 947 1 and IEC 947 3 The external main switch shall be in close proximity to the equipment and within easy reach of the operator The switch shall be marked as th
16. cted into the sample chamber using a pumped or process pressure source A stable flow rate is required The sample is released from the sample chamber by a gravity flow to a drain or sump The sample flow is dispersed evenly down an 88 mm x 200 mm flow tile The tile is angled forward to allow the sample to flow forward beneath the source receiver port This dispersion accomplishes two favorable results The sample is spread over a wide area providing a larger surface area for the ultraviolet light to penetrate resulting in an increased excitation of the oil molecules for the receiver As well the gravity flow over the tile width minimizes the sample depth against the tile This reduces the effect of suspended solids interference Fewer oil molecules can hide from the light source and oil molecules adsorbed onto the suspended solids have a reduced effect on the overall reading After the sample leaves the tile it is allowed to gravity flow to the outlet port The sample outfall must not be obstructed from gravity flow The ultraviolet light source is positioned directly in front of the sample flow The receiver is positioned at an angle to the flow direction Both the emitter and the receiver are equipped with precision light filters to control the wavelengths of the ultraviolet light being emitted and the fluoresced light being received A relationship between the measured fluoresced light and the amount of oil in the sample is mathematical
17. ction 7 0 and a sample pump will be used Disconnect the inlet of clean water and connect the prepared sample into the inlet Pump the prepared sample going throught the Hydrosense 2410 Press lt ENTER gt to input ppm value of the 2 point Enter the known or unknown ppm value of stream concentration i e 30 0 PPM Once the reading FLR rdng has stabilized press the lt ENTER gt key If the PPM value is unknown enter a random value that would be indicative of what is flowing through the unit and take a grab sample at the same time Send the sample for lab analysis promptly Record the FLR signal value on the bottom line of the display The laboratory value will be entered later Press lt ENTER gt to continue calibration and repeat the above procedure for each concentration level A minimum of two inputs is required 14 Model HS2410 HYDRO39 DOC Rev 3 9 Press lt 0 gt to finish calibration then press lt DISPLAY gt to main display menu THIS COMPLETES THE CALIBRATION IF KNOWN VALUES WERE ENTERED If random unknown values were entered the laboratory results will need to be entered to correct the values of the two samples 4 2 2 Manual Calibration With Laboratory Results When the lab results have been returned press the lt CALIB gt key and then press lt 2 gt for Manual Calibration Press lt ENTER gt to enter the 1 point value Your initial unknown will be displayed Key in the lab value of Sample in PPM
18. e disconnecting device for the equipment and include the symbols to its ON and OFF positions using the following symbols C Power Off D Power On The wiring for AC power should be 16AWG 300V or as required by local country codes The equipment is suitable for connection to a 15A protected branch circuit Wiring diagram for permanent connection See drawings at the back of this manual Use copper conductors only ELECTRICAL INSTALLATION See drawings at the back of this manual 10 Model HS2410 HYDRO39 DOC Rev 3 9 3 0 STARTUP AND SETTINGS CALIBRATION MUST BE PERFORMED AFTER INSTALLATION AND ANY LAMP REPLACEMENT RELAY LEDS ON ALARM R1 GEN PURP R2 GEN PURP R3 MAINT ALRM R4 INSTRUMENT FAILURE 4 m char STATUS LED O STATUS GRN OK RED ERROR S005 S008 S008 ZE s B Par v www arjayeng com Membrane keypad DISPLAY KEY Displays oil concentraion Information Also used as backspace in value entry CALIBRATE KEY For calibration menus CONTROL KEY For 4 20mA output and Alarm Relay settings SETUP KEY For configuration and diagnostics USRINT2k dsf USER INTERFACE Figure 1 0 3 1 NOTES ON VALUE ENTRY When entering in numeric values the cursor can be backspaced to correct mistakes by pressing the DISPLAY key This is only true if the cursor is not at the beginning of the displayed value in which case the DISPLAY menu is entered The decimal point is the d
19. ed Jeoue 3 OvAOZ L esuesoJp H GALINIT ONIHAANIDNA 193 Dad AV HV TT T To pevowo wua avd a 9I Gupuda JSNJSOACAH G31V1OSI SINHO 006 LNdLNO VU 02 7 pasn ON pasn ON pasn ON OLy 00V Hed xog due esuesoJp H a w L sl a Pic Dal AIOOE Sa201y 2 dA uadunc aui MATA ACIS dal ILI ENLIO eJeld 492g duu ajgnog JSUSEM JES INN punolg UHEJ eAnoejoJg 104 UOBJEJNSUJ 9444 OMY 81 BuulM pial Joj Buruoou punoio m seuosseooe ou sal zg DM ge JuBlem JON XVIN VA DOS uonduunsuo Jemod 810 1 spyepuejs 931 0 Ajajeg eouioe 3 1n pue INI 0cL 8cS00V Hed SPIEPUEJS 931 0 Ajayes jeoujoe a IN pue IN z euoz Z AIC sse O VSO o pe ouddy UuOngJe eg 32 0c L C8v0OV Hed jddns Jamod ajejedes aen pjnous esuesoJp H pue waj s s eBung SNOILO3NNOO YAWOLSND DONDE 18 dlAV S 9 OVAOSZ 18 dINV Z QVO13O9NVISIS3S SLOV LNOO AYA SLAdLNO AVT3 zs AES salas ou S EUJULI zuaza p Jo sindino e 12euuoo1eju jou op sous uno l JO 91 Jo ASI 94 sonpa1 o NOLENILLLV NOILNVO BAS Je O NOILYSTILN A TANNVW 37 UJAHVIJU IVNNYW Y3SN 33 O XVIN VA 002 ZHO9 0S OVAOZL OLI aqeoddy ejeuM JeujeBo paJlM lo 9e y suue y uonenueouo WO Lad 28 SAVIJA suue v bil Lg 138 Lead Lead SOUBUdIUIEY iy Y SAVT3H FA
20. eguipment An optional HART interface board is available 15 Shielded wiring is required for the output alarms and signals to avoid EMI and RFI interference from other equipment near the sample unit 16 The Monitor is housed in a Type 4X Stainless Steel enclosure Extremes in temperature and humidity should be avoided Indoor installation or a heated or cooled instrument housing for outdoor use may be required 2 2 GLOSSARY OF SYMBOLS Attention consult accompanying documents Attention veuillez consulter les documents ci joints Protective Earth Terre de protection Fuse DO P gt Coupe circuit fusible Direct Current DC Courant continu Normally open relay contacts Contacts travail Normally closed relay contacts Contacts Repos Power off Arr6O mise hors tension Power on Marche mise sous tension 8 Model HS2410 HYDRO39 DOC Rev 3 9 2 3 Live Sous tension Neutral Neutre Ground Terre UNIT INSTALLATION Locate an area that is environmentally protected from wide variances in temperature and humidity The unit will not operate in below freezing temperatures A heated enclosure may be required for these applications As well the sample line may require heat tracing When selecting the location consider that regular maintenance and testing is desirable for the proper and accurate operation of the instrument Ifthe sample input is not from a pumped source locate the instrument in
21. ergy and fluoresce light energy back out of the water at a signature wavelength The intensity of light energy at this wavelength is measured to provide an indication of the ppm concentration E Features and Benefits Bi Technical Specifications The special UV absorbing flow plate non contacting optics minimizes maintenance sheens the water over a large surface area compensation for temperature and lamp The resulting high surface area to depth deterioration minimizes recalibration ratio provides many benefits requirements alarm warns of impending lamp replacement long life lamp expectancy of 18 months continuous on line monitoring reads the water 50 times second with an averaged display update every one second no consumables or chemical used sample flow gravity outfalls to drain available with CSA Zone 2 approval or with NFPA ATEX Pressurization Systems for Zone 2 designed for harsh environments with a 316 SS housing no tools necessary for routine maintenance or lamp replacement The flow or power does not require to be turned off during routine maintenance multi point calibration available to customize The sensitivity to oil molecules is increased by maximizing the optical viewing area The minimal depth discourages oil molecules from hiding behind particulates in the water The large lamp source targets the water from multiple angles to get a representative sampling of all oil The large sample target area
22. hen press 3 to ManRef Enter the new REF CAL value then press lt ENTER gt Enter the new TEMP CAL value then Press ENTER Press DISPLAY to main display menu 4 2 5 Change calibration location There are 10 calibration locations in Hydrosense2410 Cal 0 Cal 9 Each location can have up to 5 points to draw a calibration curve The procedure to change calibration location is as follows Press the lt CALIB gt key then press 5 to Sel Cal Enter the location number 0 9 the press lt ENTER gt Press DISPLAY to main display menu 15 Model HS2410 HYDRO39 DOC Rev 3 9 5 0 OPERATION 8 DIAGNOSTICS There are a number of diagnostic screens to monitor the performance of the unit 5 1 AMPLIFIED SIGNAL Press the lt SETUP gt key then press lt 1 gt for AmpSig This screen displays the FLR Sample Fluorescence and the REF Lamp Reference signal values AFTER amplification The values when the lamp is ON as well as OFF for both the FLR and the REF are displayed as well as the respective difference between the ON and OFF values The OFF values are typically a measure of the background fluorescence and are therefore subtracted from the ON values The bottom line shows the REF value at calibration The REF value slowly decreases as the lamp ages The software compensates the FLR as the REF decreases using the REF CAL value 5 2 UNAMPLIFIED SIGNAL Press the lt SETUP gt key then press lt 2 gt for UnA
23. igher values provide more stable readings Max value is 1000 The default value is 960 Press lt 3 gt for Units 2 LCD display units are available PPM and FLR Press lt 1 gt for PPM and 2 for FLR Press lt 4 gt for Tcomp 16 Model HS2410 HYDRO39 DOC Rev 3 9 Tcomp temperature coefficient is used for temperature compensation It is factory set This value should only be modified if directed by an Arjay Engineering Representative 17 Model HS2410 HYDRO39 DOC Rev 3 9 6 0 CONTROLLER SETTINGS SHEET PARAMETER DESCRIPTION FACTORY USER SETTING SETTING REF at To compensate for fluorescence light source Calibration degradation the lamp light intensity value is captured during calibration and is subsequently used to compensate the fluorescence signal This captured reference value may be viewed in the Calibrate menu NOTE altering the REF at Cal value will alter the unit calibration and should only be done under Arjay authorization Zero Oil PPM value for full scale mA output 20 mA 100 0ppm mA Action Direct 20mA when PPM is at Span or Inverse 4mA when PPM is at Span Relay1 Hi Set Alarm Relay 1 High Setpoint Alarm condition if PPM is above this value Relay1 Lo Set Alarm Relay 1 Low Setpoint Alarm condition cleared if PPM is below this value Relay2 Hi Set Alarm Relay 2 High Setpoint Alarm condition if PPM is above this value Relay2 Lo Set Alarm Relay 2 Low Setpoint Alarm
24. is an electronic device used for environmental and personal protection as well as general process monitoring As with any calibrated sensing device wetted parts may become contaminated and the light source can deteriorate over time The Arjay system has a Failure Lamp alarm relay R4 feature included however a scheduled periodic test is recommended to ensure that the unit and remote alarms and devices are operating to specifications A Clear Flow Tile is used for most applications and is shipped as the standard with the unit Film build up of algae and silt will not affect the operation of the unit unless it accumulates oil from contaminated water to a point at which the unit acknowledges and reads this oil It is suggested that routine maintenance be scheduled to verify that this build up is not extreme and the flow tile is cleaned as necessary There are three standard tests to assure the operation and accuracy of this unit The first is a BUMP TEST To confirm that the unit is responding to the contaminant a higher or lower concentration than normal may be manually inserted into the stream at the actual process or at the sample chamber Divert the 3 way valve and provide clean water and different concentration sample into the flow Observe that the reading increases or decreases and alarms accordingly A second test is a FLUORESCENCE ACCURACY TEST A colored TEST TILE is provided with the unit to confirm the accuracy and stability of the uni
25. ly predictable over the measurement range of the instrument The precision light filters maximize the predictability over alternative non filtered methods of measurement The sample tray is easily accessed for any necessary cleaning of the tile The ultraviolet light source is easily accessible for replacement as required For periodic testing and calibration a 3 way valve in the inlet line is provided to manually input a fresh water source to confirm the instrument response zero and clean the instrument Model HS2410 HYDRO39 DOC Rev 3 9 13 INTERFERENCES AND AFFECTS TO ACCURACY The UV fluorescence technique monitors the intensity of light emitted from the passing stream at a selected wavelength band This technique can be quite selective by eliminating the light affect of compounds in the water that do not share the same fluorescence characteristics of hydrocarbons 1 When chemical compounds in the water are excited with light energy only certain compounds will emit the light back out of the water at a higher wavelength than excited with These are referred to as fluorescing compounds The HydroSense does not respond to most chemicals because it only responds to fluorescing compounds of which aromatic hydrocarbons are included 2 The light used to excite the compounds is filtered to 254 nm Of all the fluorescing compounds only certain ones will respond to this wavelength Some respond to higher and some to lower wavelengths
26. meter 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 documented field results from offshore oil platform Jan 24 Mar 04 02 All calibration relays signal outputs and The HydroSense 2410 correlates well against laboratory power wiring are available at the main methods and is ideal for process trending and continuous control unit on line monitoring Arjay SS 06 AR A i tel 1 905 829 2418 m Arjay Engineering Ltd fax 1 905 829 4701 2851 Brighton Road N America 1 800 387 9487 Oakville Ontario email arjay arjayeng com w SBS Canada L6H 6C9 web www arjayeng com Model HS2410 HYDRO39 DOC Rev 3 9 TABLE OF CONTENTS Please read the HydroSense Installation Notes 2 1 prior to locating and mounting the enclosures SPECIFICATIONS otr n e ett e ali 3 1 0 INSTRUMENT OVERVIEW aaa 4 1 1 FEATURES u uu asun ayu ai 4 1 2 DESCRIPTION 6 Sains Shae uu RE a assets 4 1 3 INTERFERENCES AND AFFECTS TO ACCURACY seem 5 1 4 ROUTINE CLEANING PROGRAM 6 2 0 aus 7 2 1 HYDROSENSE INSTALLATION NOTES 1 mem een 7 2 2 GLOSSARY OF SYMBOLS 1 eene nc nr ran nnne 8 2 3 UNIT INSTALLATION ests spi la sadi ini la ra wan a niku ME 9 2 4 PERMANENT POWER CONNECTION AC POWERED MODELS ONLY 10 2 5 EEEGTRIGAE INSTALLATION u ul uu triti Hebe tet etit iet bance geseis 10 3 0 STARTUP AND SETTINGS taa fe cane aetna eae E dee nod ne 11 3 1 NOTES ON VALUE ENTRY u eri Rete Ier qb e Re e aati 11 3 2 POWERUP DISPLAY 5
27. mpSig This screen displays the FLR value the difference between the ON and OFF values BEFORE amplification This is a calculated value based on the measured amplified signal and the gain value The Total Gain Fixed Gain and Pot Adjustable Gain values are also displayed on the same screen 5 3 DIAGNOSTICS Press the lt SETUP gt key then press lt 3 gt for Diags The screen displays the currently applied temperature compensation correction and the temperature at calibration In addition to the above information the mA output can be forced to either 4 or 20mA regardless of the ppm value Forcing the mA value is useful to check the response or performance of external eguipment such as chart recorders or control devices 5 4 MIN MAX Press the SETUP key then press 4 for Min Max This screen displays the FLR and REF minimum and maximum values for the last 10 seconds as well as their difference These figures give an indication of the reading stability 5 5 SETTINGS Press the lt SETUP gt key then press lt 5 gt for Settings Press lt 1 gt for mA Trim This procedure trims the mA output for maximum accuracy by compensating for the mA output circuitry tolerances THIS PROCEDURE IS PERFORMED AT THE FACTORY AND IS TO BE PERFORMED BY AUTHORIZED PERSONNEL ONLY IF IMPROPERLY DONE THE ACCURACY OF THE MA OUTPUT CAN BE IMPAIRED Press lt 2 gt for Filter The moving average filter tracks the average of the last N samples H
28. oking directly at the ultraviolet light source Model HS2410 HYDRO39 DOC Rev 3 9 2 0 INSTALLATION NOTE If any damage to the instrument is found please notify an Arjay Engineering representative as soon as possible prior to installation QUALIFIED PERSONNEL MUST UNDERTAKE ALL INSTALLATIONS WARNING If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired 2 1 HYDROSENSE INSTALLATION NOTES IMPORTANT Read these notes before installation The system is comprised of two main components the Sample Chamber and the controller 1 The Sample Chamber receives the flow sample from the process and outputs it to drain The chamber should be wall mounted on a vertical and horizontal plane to allow a proper flow through the sensing unit 2 The Sample Chamber should be located close to the process to reduce the lag time of the sample to the unit This will offer more instantaneous readings and real time recording 3 The outlet gravity flows to drain and consideration of a close proximity to the drain is important Also mount the unit where it is readily accessible for maintenance and periodic testing by manual insertion of known samples 4 Ifthe process flow is not under pressure the chamber should be mounted below the process level so the sample can free flow down and through the unit A pump may be used The maximum input pressure is 100 psi The input flow
29. onitors the lamp life The LCD display and a maintenance relay will warn of an impending need for replacement so an operator can schedule this replacement without any downtime Full diagnostics are accessed on the LCD display The system automatically compensates for temperature and zero calibration offset shifting due to lamp aging The system automatically compensates for background or stray light energy The controller reads the sample over 50 times per second and averages these readings to provide an updated output The unique glass flow plate design provides a stable representative sample of the passing stream All modular components are plug in for easy servicing The unique sensing chamber design allows easy access to the controls and wetted components The lamp receiver unit is simply lifted and placed onto the convenient door rack For any routing cleaning the flow plate can be wiped in place or removed Arjay SS 06 Hvdrosense 2410 The HydroSense 2410 uses a UV fluorescence technique to target the aromatic component of the oil contamination Through a site calibration this aromatic tag provides an indication relative to total oil A continuous sample flow is tapped or pumped off the process line and directed through the HydroSense chamber It passes behind the non contacting UV light source and is targeted with filtered light energy The soluble and emulsified oils in the water will excite from this light en
30. ot e key Values may be entered with any number of decimal places If the entered value is out of the allowed limits the system displays the limiting value for 2 seconds For example if the alarm time delay value is entered as 5000 seconds then MAX 99 is displayed for 2 seconds then entry is allowed again The current value is not changed unless the entered value is within limits During value entry the oil concentration is constantly updated in the background 3 2 POWERUP DISPLAY After mechanical and electrical installation of the sample chamber and the controller have been successfully completed power up the unit The LCD should show a similar screen Display Screen to 11 Model HS2410 HYDRO39 DOC Rev 3 9 OIL MONITOR Calo 50 0 PPM 20 C nu NOTE The shown values are for example only The right corner of the 1st line shows the calibration location Cal0 Cal9 The unit can save 10 different calibration curves in each curve up to 5 points can be entered Press 0 9 to view Cal Cal9 location respectively The 2nd line shows the oil concentration value in PPM and current temperature of sample chamber The 3rd line is blank The 4th line displays a bar graph of the oil concentration as a percent of the Output Full Scale The bar graph resolution is 5 The Status Indicator see figure 1 0 should be green If this is red the LCD displays the System Error See the troubleshooting guide for
31. press CONTROL then press 1 for Relay Settings Menu Press 1 for setpoints Enter the value in PPM for the desired alarm relay to activate Press ENTER Now enter the value for the relay to de activate and press ENTER A small differential between Relay Hi and Relay Lo may be desired to eliminate relay chatter if the PPM concentration is fluctuating at the alarm level Similarly enter the high and low alarm points for the second relay if it is to be used Note for a single point alarm with no differential enter the low alarm value identical to the high alarm value 3 3 4 Relay Time Delay To delay the relay alarm for a preset time in seconds press 2 for Enter On delay in the Relay Setting Menu This will suppress the alarm to eliminate a spurious momentary alarm that may be caused by an oil globule or process variable not indicative of an alarm condition 12 Model HS2410 HYDRO39 DOC Rev 3 9 3 3 5 Relay Enable The relays may be disabled from operating for maintenance purposes Confirm the relays are activated in the lower right corner of the display in the RELAY SETTINGS menu If the display reads ON in the lower right corner the relays are activated Press lt 3 gt to enable the relays or press lt 4 gt to disable the relays The control functions are now entered Press lt CALIB gt to calibrate or lt DISPLAY gt to return to the main display 13 Model HS2410 HYDRO39 DOC Rev 3 9
32. rate is a continuous and stable 1 litre minute 5 The inlet connection to the unit is a 3 8 female thread A barb connection may be threaded to this when flexible inlet tubing is used Clear flexible 3 8 or 1 2 inlet tubing is suggested for the inlet sample This will provide a visible indication of the sample as well as an indication of contaminant build up To minimize the contamination in the tubing Teflon lined tubing may be desirable Hazardous Locations or local regulations will dictate materials to be used Note Clear tubing should not be used outdoors where algae build up from sunlight is increased 6 An on off valve at the process is recommended to shut down the system for maintenance and or sample tube replacement An on off valve is included within the unit for throttling flow This can be used for internal maintenance A 3 way valve is included at the inlet for a fresh water input for zeroing cleaning and testing 7 The sample gravity flows out of the Sample Chamber The outlet tube must only be installed in a downward vertical or downward graded horizontal direction Any excessive upward direction will cause the sample to back up and could flood the sample chamber 8 The outlet fitting is a 2 male thread Do not reduce this size This will cause a restriction and flood the system 9 For non purged pressurized models the outlet of the tube should be open to air not submerged in water or a process which would cau
33. rescing chemicals are often included in detergents to enhance the visual affect of a cleaned product such as clothes Not all of these commercial dyes will affect the wavelengths of the HydroSense however green dyes have proven to be a common interferent Suspended Solids and Turbidity The unit is calibrated to a passing stream of water The amount of light fluoresced by the aromatic hydrocarbons determines the calibration parameters The light received by a hydrocarbon and then sent to the receiver is based on a stable light path through the water If suspended solids or turbidity block the light getting to the hydrocarbon light cannot be fluoresced back to the sensor Readings can be dampened by an increase in solids or turbidity When process water is used during the calibration the offset affect of solids is taken into account and zeroed out The design of the large surface sensing area verses the small sensing depth minimizes the affect of turbidity in the HydroSense In effect the hydrocarbons have little place to hide behind solids In circumstances of dramatic changes in turbidity sample conditioning techniques prior to the HydroSense should be considered Temperature Compensation Temperature can affect the light sensor The value is negative i e the measured oil readings fall as the temperature increases Temperature compensation is built into the unit The temperature compensation coefficient will be preset at the factory but ma
34. se a backpressure This could result in a restriction of the effluent flow and spillage from the overflow ports at the chamber outlet For purged pressurized models a slight and constant backpressure is required to maintain a cabinet pressure To accomplish this submerge the effluent tubing approximately 10 cm below the water surface or run the outlet tube with a slight upward grade or drain trap prior to the downward outfall Be sure the tube never rises higher than the bottom of the chamber outlet 10 The Sample Chamber must be mounted indoors or in a heated housing to avoid sample freezing The inlet and outlet tubing must not be exposed to freezing environments 7 Model HS2410 HYDRO39 DOC Rev 3 9 11 The controller operates using 115 VAC 50 60 Hz 220 VAC 50Hz or 24 VDC as ordered 12 The monitor provides LED indication of the relay status The relays are dry contacts and will accept AC or DC inputs Setpoint alarm levels are on Relay 1 8 2 Relay 3 is an offset drift alarm which indicates cleaning or re calibration is necessary Relay 4 indicates a Lamp Replacement reguirement 13 A 4 20 mA DC output signal proportional to the PPM levels is provided This is an isolated signal capable of driving 900 ohms Remote indicators receiving devices and their distances should be considered when choosing a location for the Arjay Controller 14 A RS 485 Modbus output is provided This can be used to link to computers and other
35. t After a calibration has been completed lift the Lamp tray to access the glass flow tile Turn off the inlet water and replace the glass tile with the ceramic tile shiny side facing out Leave the water off Secure the lamp box back in position and close the chamber door Read and record the display in PPM Label or note the reading and date of the test tile Periodically or in conjunction with the BUMP TEST insert this test tile as above and confirm that the reading is within 5 of the initial reading If the unit is not within specification re calibration of the unit should be initiated IMPORTANT Each time the unit is re calibrated the test tile should be inserted and the new reading recorded It is recommended that the above tests be done initially with a high frequency to record a history of the unit stability The frequency can be reduced to a level comfortable to the application and customer A one month minimum check is recommended in conjunction with a tile cleaning A third test is an actual CALIBRATED SAMPLE TEST Since the two above tests confirm both a response to a calibrated sample tile and the contaminant using an actual sample may only be necessary in applications where precise data records and monitoring is required To test the accuracy and calibration of the unit have a sample analyzed by a lab and compare it to the observed reading The lab procedures must be the same as the ones used for calibration NOTE Any sample sen
36. t to a lab may incur separation and evaporation during transit Advise the lab to thoroughly mix the sample prior to testing 8 2 FLOW TILE CLEANING The unit is designed for guick and easy cleaning Remove the Lamp Blue Box and place on the chamber door to access the flow tile Wipe the flow tile with a clean damp cloth The tile may be removed for cleaning if desired Do not use soap as this may cause an inaccurate reading if not rinsed completely Replace the flow tile Verify an even water flow across the tile Place the Lamp Blue Box back in the Tray and secure it with the retaining latch Close the door snugly 20 Model HS2410 HYDRO39 DOC Rev 3 9 8 3 LAMP REPLACEMENT Power down the unit Allow 15 minutes for the lamp to cool Open the chamber door Open the Blue Box lid to view the lamp The lamp is connected to a socket and the unit is held down with the single knurled screw Undo the screw and carefully lift the lamp up Remove the lamp from the socket unscrew the connectors mounting block and insert the new lamp Note the socket has a polarity to direct you Secure the new lamp into place with the metal band and knurled screw Close the lid A calibration must be performed after this procedure 21 N IAJ AO UJejs S JU9A pue esuesoJp H 3SN3SOMHGAH 193POYdA AALINTI ONIHAANIDNA AVIAV E e YT pevowo wan avid a c BON 99S HOOG LNOHLIM MALA LNOHJ seoeld Z dAL vI
37. to be targeted by the monitor The calibrated accuracy relies on the oil type and conditions being consistent The HydroSense will respond positively to aromatic hydrocarbons but the display accuracy may be affected by variations in the types and sources of these hydrocarbons Other Chemicals in the Water The light sensor is selective to compounds in the water that emit light at 350 nm when excited from 254 nm light If there is a background chemical in the water that fluoresces at these wavelengths the HydroSense will respond to them If this background chemical concentration is consistent this interference will be zeroed out during calibration Calibration is recommended using process water so that any background interferents are zeroed out If an interfering background chemical changes in concentration the HydroSense will sense this change Consideration to this affect is important for alarms and recording Filtering of the water changes to chemical use or special light filtering may be required to provide more stable readings Model HS2410 HYDRO39 DOC Rev 3 9 The periodic introduction of fluorescing chemicals into the water may also affect the reading During these conditions operators and alarms should be acknowledged that nuisance alarms may occur Soap manufacturers will often include fluorescing dyes in the product for appearance and identification Green dyes are typical in industrial degreasers and commercial soaps Fluo
38. ure alarm one for zero offset drift 3 8 NPT female 2 NPT male outfall must be unrestricted gravity to drain minimum 5 0 litre minute continuous and stable minimum 2 psi maximum 100 psi minimum 20 psi when equipped with optional coooler Type 4X Stainless Steel with viewing window CSA Approved Ex Zone 2 12 0 X 37 5 X 9 25 D Eexp available for Zone 2 where CSA Zone 2 is not applicable 34Kg 75 Ibs Model HS2410 HYDRO39 DOC Rev 3 9 1 0 INSTRUMENT OVERVIEW 1 1 FEATURES e Fluorescence technology e Non contact sensor e Isolated inputs e Continuous On Line monitoring e Instantaneous readings e Multi point automatic or manual calibration e Temperature and light strength compensation e 4 relays SPDT 7A contacts R1 and R2 for Alarms R3 and R4 for maintenance e Isolated 4 20 mA output RS 485 Modbus output optional HART e LCD display in PPM e No moving parts complete maintenance without turning off the sample stream 1 2 DESCRIPTION The HydroSense Hydrocarbon in Water monitor from Arjay Engineering Ltd has been designed for municipal and industrial applications to measure PPM levels of hydrocarbons in aqueous solutions Typical applications include PPM trace amounts of oil in effluent water from storm water runoff oil in cooling water produced water and oil water separators Other measurements and mediums can be monitored on request i e colorants in fluids etc A continuous sample is dire
39. vola aa Ws spaepueys Oa 0 4948S 1899913 IN pue 109 O 0ZZ 9ZS00V Hed Y 30N pue punojo UHEJ peppy pozad L spJepuejs 931 o Ajajes e2ujoe 4 IN pue 109 Z euoz Z AIC SSBID WSO 0 ponoJddy uoneJlej9ag JO OZZ Z8700Y Hed Ajddns Jamod ajejedas esn pinoys esuesoJup H pue wW9 sAs oBINJ 4 8 0N SNOLLIINNO9 HNAMOLSNI T xew VWOSL qaa1v10Sl DONDE 1 dNWV S 9 OvAOSZ JE dINV Z XVN VA 002 Indo een eBind ON O SINHO 006 QVO SONVLSISSY SLOVLNOO AJO ZH09 0S8 OVAOrC 0cC ZHE9 y OvVAOvZ 0zZ LAdLNO vui 02 7 SLAdLNO AV TJA JSNJSOHJAH HOJA AIdANS Y3IMOd indu Jamo dg eBung ZI L1 1vNOILdO 9 ge91 ddy eueuM JoujoBo palm fuojoe y O anj 1i DN suue y UOHBLJUHIUOL Wo Gad 8 SAVTIJA TAVIJA 5 salas ou I ETS ASSIS sindino ew Eer jou op 204s ous JO 91 Jo ASIA 943 snp ot NOLLNALLW NOILNWD ZHO9 0S suue v INdNI eoueuejure y 9 SAVTJM a gran stet A062 vez S90CAH Hed arga 122919 WOU Je Mog 0ZZ VL6r00V Hed Ing paJlM e Joj 041u02 SSUSSOIP H A At SeD01g 2 GAL uadunr M 00000 ASIA STIS EP TO NE S E Vp0ETI 4 ued 3QIS 1HDIY WILavd Addns Jewog due pasn ON pasn ON pasn ON OLVEOOV 8 Hed BIAS De xog due esuesoJp H NOILYSTILN A TANNVW 371 38303V933 IVANV A Y3SN 33 O 22504002 ov oud sa uf OZ 1 Z8900V Hed OZL Z8v00V H
40. y be adjusted at any time via the keypad 1 4 ROUTINE CLEANING PROGRAM The HydroSense relies on a constant flow of water across the sensing plate Excessive particulates and algae in the water can build up on the flow tile and in the overflow tray This will eventually affect the performance of the unit Setting up a Routine Cleaning Program is vital to the successful performance of the unit Each application will vary in the frequency of cleaning Some may require daily wipes and some may require monthly cleaning A basic wipe down of the flow tile can be done without having to shutdown the stream or power The wipe procedure will take 2 to 3 minutes To set up a schedule it is recommended to program a daily wipe of the glass using a clean paper towel After two weeks of daily cleaning determine if every other day may be adequate If so set this program in place for two weeks Slowly extend the frequency between cleanings until an adequate program frequency is determined for your individual site conditions Cleaning the flow tile and plastic surface should be done with clear non fluorescing chemicals An appropriate glass cleaner is Windex brand anti fog Low concentrate HCL or grout cleaners are also effective Use only chemicals approved for your site personal safety and disposal Be sure to thoroughly rinse the flow tile with clean water before installing CAUTION Repeated visual contact with the light source can be harmful Avoid lo

Download Pdf Manuals

image

Related Search

Related Contents

  C 150 C 155  BENUTZERHANDBUCH Thermo-Hygrometer Modell RH30  User-Manual - Newegg.com  SilkCentral® Test Manager™ 2010 R2  Phocos SI 350 Series  Scientific Atlanta DPC2434 User's Manual  Oster TSSTTRWF4S User's Manual  Carrier 50PG03-07 Air Conditioner User Manual  

Copyright © All rights reserved.
Failed to retrieve file