Contents

1. TABLE OF CONTENTS 1 INSTALLATION Ge 6 1 1 Scope of SUPDIN aa 6 12 ole ej ae 6 13 Mainsoomeiona ANA 6 1 4 COMPresSsSOr ee ee AA 6 1 5 Connecting GANG panel engsten 8 1 6 Connection to a Peine 8 2 OPERATION en een 8 21 SWIENINB ON ernennen 8 2 2 Program EO nsa AA 10 2 3 Preset Programs Kasa ae een 11 24 Positi ning the Wat h nee 11 SB 1 iN AA AA AA 13 2 6 Interpretation of the FSU IES ins cisssnnssioncstneenanannnnmassarnsescnenurnddseanamshsanusascnexwexcunn 13 3 LEAKFINDER PROGRAM amma ANGKAN 14 3 1 How does the Leakfinder Program works uuennensensannannennannannunnunnunnunnunnunnunnunnen 14 3 2 Test SONGS aaa ANAN AA GA 15 4 CUSTOMER SPECIFIC PROGRAMS 00a nee eine 16 ca BI Cola LAAN S 16 4 2 Test with one OF TWO pressures ne KAG 17 4 3 LOWEL PLESSUNG maa NAA AA N ARGENEEREEERENENEROREUERERREERHEUEEFORERNEN MINERORAEE 17 4 4 Higher pressure Ga NAAAGNAS 17 4 3 Stabilisation Imb n iaa 18 4 6 Meas ring UNG aaa aNG ANAN an AN AGANG 18 4 7 Tiantness Miami E Ruhr uuhe uhren ehe 19 4 8 Case DANS ee 19 4 9 Waka AA AA 19 4 10 Test oe E T GG LALAKAD LAKAN NALNG 20 4A TI Save paramMetetS een 20 5 SYSTEM PARAMETER 20 51 Lang age Selection 21 5 2 Display of the numerical results unnnennsennnennanannannnnunnannnnunnannnnunnannnnunnannnnunnennnnn 21 5 3 AUtO PINUNA ANAN 21 ee een 21 5 5 Identification MUmber ain nennen nennen nee 21 55 Calibration da 21 5 7 Calibration Valle nein 21 58 Programmi
2. If the test pressure is not 2 bar the value is converted accordingly see next chapter The default value is 1 min for medium sized watches and 2 min for small watches based on 2 bar The value for customer specific programs can be selected between 0 1 and 3 0 The lower the selected water resistance limit the longer the measuring time intervals needed for an accurate measurement 6 6 REFERENCE TO ISO STANDARD NIHS92 10 The ISO 2281 Standard states that a watch is to be considered as water resistant if less than 50 ug of air per minute penetrates into the watch at a test pressure of 2 bar A leak test based on the deformation principle does not allow the weight of the inflowing air to be directly measured Under specific assumptions it is however possible to establish a relation between the weight of the inflowing air and the deformation reversal With a medium size wristwatch free volume in the watch 2140 mm and a test pressure of 2 bar a flow of 50 ug of air per minute correspond to a deformation reversal of 1 per minute This value is set as the water resistance limit in the pre set programs The ISO 2281 standard does not take into account the volume of the watch Therefore the leak rate value of 50 ug per min is valid for all watches irrespective of their size In the test according to the deformation principle PROOFMASTER S the deformation reversal depends on the free volume in the watch A watch with a smaller
3. In the case of a small to medium leak it takes several minutes until the excess pressure has been released Thus a short test under water does not present any risk that water penetrates into the watch If the leak is very big however the excess pressure is released faster and there is some risk that water penetrates into the watch PROOFMASTER S measures the size of the leak with the Leakfinder Program and indicates to you if the watch can be immersed in water without risk Based on the size of the leak and the selected pressure the instrument computes the time during which the pressure has to be applied to the watch to create a high enough excess pressure in it Note Certain leaks due to a deformation of the case only manifest themselves at higher pressure and close again when the watch is under a normal pressure environment Such leaks cannot be located with the water test 3 2 TEST SEQUENCE Selecting the Leakfinder Program The Proofmaster S has two possibilities for selecting and using the Leakfinder Program The information for using the Leakfinder Program automatically appears on the display once a watch has completed the previous test with a poor result leaking The customer now has the choice to prepare the watch for locating the leakage in water by means of the Leakfinder Program or not The Leakfinder Program may be used as an independent test unrelated to other programs i e without previous test in Type Stnd Case another te
4. This allows deformations to reverse themselves If a watch is not leak tight an excess pressure builds up in it and according to the leak size it can take a very long time until the excess pressure has fully dropped In the case of test programs with 2 to 3 pressure values the previous pressure must always be lower than next pressure TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 7 FAULTY OPERATION 7 1 ERROR MESSAGES Implemented error messages are listed below Their meaning and fault removal are described The chamber opens again after it has been closed The test Measurement stopped vacuum or pressure cannot be built up Check the output No air pressure of the compressor It must exceed the selected test pressure by at least 1 bar If the input pressure is less than 8 bar or if the instrument is used at high altitude the highest achievable vacuum is reduced Switch to a program with lower vacuum Measurement stopped The chamber opens immediately again after it has been Sensor out of range closed The sensor exceeds the measuring range Check that the watch is correctly positioned The yellow LED sensor lights up when the sensor is correctly lowered Measurement stopped A big loss of pressure or of vacuum has been detected The chamber opens automatically Check the cleanliness of the sealing ring and of the supporting surface of th
5. H x D e Weight 6 Kg TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO Compressed air connection e Quick coupling for connection to a compressor or an existing compressed air line e Minimum input pressure 1 bar higher than test pressure e Maximum input pressure 12 bar e Air consumption when testing with pressure 10 per test when testing with vacuum 40 per test e The air must be dry and free of oil Mains connection e Mains adapter 9 V 1 2A optionally for 230 V or 120 V 12VA e Function range 200 V 250 V and 100 V 130 V respectively Operating conditions e Operating temperature 10 C 40 C e Rel humidity 10 80 non condensing e Storage temperature 0 C 50 C Standard accessories e Mains adapter e Compressed air hose 6 4 mm screw connection G 1 8 e Dust cover e Operating Manual Optional accessories e Compressor JUN AIR Type 6M Maximum pressure 12 bar e Printer type MCP 7810 e Filter oil separator for compressed air connection e CD with PC Software Auto Print for printing out the results via the PC e CD with PC Software Proofmaster S M for data recording and parameter setting EC DECLARATION OF CONFORMITY The instrument conforms to the following regulations of the EC Directives 89 336 EEC EMC EN 55011 1991 Interference emission EN 50082 1 1992 Interference immuni
6. and maintenance in its manual If you have any question kindly contact our customer servic TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO 9 DISPOSAL OF THE INSTRUMENT This electronic instrument should not be disposed of as household refuse If it is not accepted by public collecting points kindly bring it back to the sales outlet It will provide suitable disposal according to legal directives Your equipment suppliers in the EU will take back free of charge all instruments manufactured after the 13 8 2005 as well as an older instrument provided an equivalent new instrument is purchased 10 TECHNICAL DATA Test principle e Analysis of the deformation of the watch case under pressure and or vacuum Predefined test programs e 7 non modifiable test programs optimised for various watch types Customer specific test programs e 10 test programs with free choice of test parameters Test sequence with 1 or 2 different test pressures Vacuum and pressure range e Vacuum selectable from 0 2 to 0 8 bar in steps of 0 1 bar e Pressure selectable from 0 2 to 1 0 bar in steps of 0 1 bar e Pressure selectable from 1 0 to 2 0 bar in steps of 0 2 bar e Pressure selectable from 2 0 to 5 0 bar in steps of 0 5 bar e Pressure selectable from 5 0 to 10 0 bar in steps of 1 0 bar e Accuracy of test pressure 3 Measuring time e Automatic determ
7. bar The number and the program name are displayed in the first line The two test pressures are displayed in the second and third lines Test 2 2 0 bar Select Too high test pressure or vacuum can damage the watch being tested TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 2 3 PRESET PROGRAMS P1 Safe Test This program can be used with all types of water resistant watches Test 1 0 2 bar without risk of damages Test 2 2 0 bar This programme gives unreliable results with hard watches P2 Men Standard Test 1 0 4 bar Standard program for gent s watches plastic or metal Test 2 3 0 bar P3 Ladies Test 1 0 4 bar Program for small watches lady s watches Test 2 2 0 bar P4 Hard Case Test 1 0 7 bar For hard stable watches with ceramic case and mineral glass Test 2 3 0 bar P5 Jewelery Test 1 0 2 bar For jewellery watches and other watches with less stable cases Test 2 1 0 bar P6 Diver 100m Test 1 0 7 bar Threefold test for diver watches that are specified for at least 100 m Test2 10 0 bar water depth P7 Pressure only Test 1 2 0 bar Pressure only for quick tests P8 Two pressures Test 1 0 5 bar 1 low pressure and a higher pressure double pressure test Test 2 3 0 bar For watches that should not be vacuum tested C01 C10 Test p
8. different languages see Chapter 5 1 Language selection page 21 e Before the start of a test the selected test program with the relevant parameters is displayed e You will receive information concerning the test progress during the test e At the end of the test the results as well as the date and time are displayed TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO In addition the 3 light emitting diodes inform you about the status of the test The functions of the 3 LED s are e signal Glows yellow as soon as the sensor has been correctly lowered onto the watch e Test 1 good green poor red or uncertain yellow If Test 2 good green poor red or uncertain yellow 2 2 PROGRAM SELECTION With the PROOFMASTER S you can select the optimum test program for testing a certain watch from 17 different programs The programs P1 to P8 are predefined programs whose parameters are optimised for different watch types These programs can not be changed In the case of the customer specific programs C01 to C10 you may determine the parameters in accordance with your requirements Select the desired program by pressing the cursor button up or down several times Press escape if the instrument is not on the menu level for the program selection e g on the menu level result display or parameter programming P1 Safe Test Test 1 0 2
9. ge f co converts the limit based on the test pressure Quit esc For a test in accordance with the ISO standard the water resistance limit must be adapted to the free volume in the watch We recommend using the following values e 0 5 for large watches e 1 for medium watches Standard value e 2 for small watches 4 8 CASE ANALYSIS To determine the minimum necessary deformation the C01 Case analysis Proofmaster needs the information about the ductility of the Soft Stnd Hard Stnd watch case Change t Param gt This parameter affects Big leakage Quit esc You can select one of the following case types e Soft for plastic cases and thin walled metal cases e Sind for normal and robust watches with metal case e Hard for watches with hard case hard metal ceramics and for diver watches 4 9 WATCH SIZE C01 Watch size In the case of the watch size the dimensions refer to the glass mm 20 40 mm or bottom diameter To determine the minimum necessary deformation the Proofmaster S needs the information the watch size Change Param lt Quit esc You can select from the following watch sizes e Smaller than 20 mm e 20 40 mm e Larger than 40 mm TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 4 10 TEST MODE co1 Test mode Proofmaster S offers 3 different modes for the measurement Lab St
10. instrument based on the test parameters The watch either has a large leak which causes an instantaneous pressure equalisation or the watch is so rigid that it can be sufficiently Leakfinder 14 23 compressed with the selected pressure With the Leakfinder Program you can locate a possible leak see Chapter 3 ZEAKFINDER PROGRAM page 13 TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Important leakage detected The watch is leaking The size of the leak has exceeded the numerical measuring range of the PROOFMASTER S With the Leakfinder Program you may locate the leak Leakfinder 5 10 58 P1 0 2 142 0 bar Test 1 poor 5 2 The deformation reversal with both pressures is greater than 12 5 the set leak proof limit the watch is leaky Test 2 el ade With the Leakfinder Program you may locate the leak Leakfinder 11 10 P1 0 2 2 0 bar Test 1 okay 0 4 Test 2 okay 0 2 26 07 2008 11 18 The deformation reversal is less than the programmed leak proof limits or the watch shows a low increase of the deformation the watch does not leak With the C01 C10 test programs a fixed measuring time may also be programmed instead of the automatic measuring time If the programmed fixed measuring time is too short the result becomes uncertain This is indicated by a question mark next to the result and by a go
11. negative percentage indicates that the deformation has been reversed during the measuring time This means in general that the watch is not leak tight The higher the percentage the bigger the leakage In the case of a vacuum test a negative percentage is also obtained if the watch is cooling down during the test In other terms it may occur if the temperature of the watch exceeds the ambient temperature at the beginning of the test The watch is classed as not leak tight if the negative value exceeds the programmed water resistance limit Positive percentage A positive percentage means that the deformation has still shown a small increase after the stabilisation time The deformation of a water resistant watch with plastic behaviour still increases slightly during the measuring time and the result will be a positive value Small positive values for the deformation do not affect the reliability of the measurement process A significant increase of the deformation during the measurement process could however mask a small leak and compromise the measurements PROOFMASTER S monitors the variations of the deformation during the stabilisation time and only starts the measurement process when it has become reliable TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO 6 5 TOLERANCE LIMIT The programmed tolerance limit refers to the standard 2 bar test pressure
12. the following important points e Only use oil free dry and dust free air use a filter e The lowest input pressure must exceed the working test pressure used by at least 1 0 bar e For the vacuum test the compressor must include a pressure vessel with a volume of at least 4 litre e Before connecting to the mains ensure that the voltage of the compressor corresponds to the voltage of your mains e A PROOFMASTER S can be damaged if condensation water penetrates the instrument together with the compressed air Therefore carry out a regular maintenance of the compressor and air filte see Chapter 8 2 Compressor JunAir e Maximum input pressure 12 bar set pressure reducing valve on compressor to 11 bar e The sudden discharge of compressed air can cause injuries Exercise care when manipulating the compressor and the compressed air hoses e Before a possible disassembly of parts of the instrument disconnect it from the compressor Never open pressurised parts of the instrument using force e Normal pneumatic instruments and fittings are appropriate for a maximum pressure of 8 bar They may not be connected to the compressed air supply of 11 bar Compressed air connection Introduce the open end of the flexible tube up to the catch into the plug in coupling labelled max 12bar To unplug the compressed air tube press the black ring of the plug in coupling toward the rear panel of the instrument Simultaneously unplug the t
13. volume shows a greater deformation reversal for the same quantity of inflowing air For a test in accordance with the standard it is necessary to program the water resistance limit according to the free volume in the watch The approximate values are 0 5 for large watches 1 for medium watches and 2 for small watches The standard refers to a pressure of 2 bar Limits for other pressure values are not specified Let us assume that the limit of 50 ug per minute is valid for all test pressure values The water resistance limit set for a pressure of 2 bar is hence adjusted as a function of the test pressure that corresponds to the 50 yg per minute limit The following limits are obtained as a function of the pressure for the standard limit of 1 Pressure Limit 0 5bar 4 min 1bar 2 min 2bar 1 min 5bar 0 4 min 10bar 0 2 min The corresponding limit adapted to the pressure is shown on the display during the test TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Note As explained above the leak test can be performed with PROOFMASTER S according to the ISO standard with results being in good agreement However since the test principle of PROOFMASTER S does not correspond to the ISO 2281 standard it is not possible to ensure a 10096 conformity 6 7 MEASUREMENT RESOLUTION PRECISION The measurement reso
14. ALL LEAK CONTINUE WATER TEST You can immerse the watch for the water test Water test Extract immediately the watch from the instrument and place it in a transparent receptacle filled with water If the watch has a leak bubbles appear at the position of the leak If the leak is small the bubbles form very slowly It may take several minutes to identify the position of the leak If possible use distilled water for the test It avoids the formation of limestone deposits on the watch Remove the watch from the water if the formation of bubbles slows down in the case of a bigger leak Water might penetrate into the watch if the excess pressure in the watch has entirely vanished 4 CUSTOMER SPECIFIC PROGRAMS The customer specific programs C01 to C10 can be adapted according to your own needs to take into account the specific characteristics of your watches The next chapters provide you with hints for programming the test sequence and selecting the parameters 4 1 PROCEDURE e Select one of the programs C01 C10 by pressing the cursor button up or down e Press enter to edit the parameters e Press the cursor button to the left or right to select the desired parameter TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO e Press the cursor button up or down to change the parameter value e Press the cursor button to the right to get to the next p
15. Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Zertified Management System Ses ISO 9001 2000 Reg Nr 12228 PROOFMASTER S OPERATING MANUAL Document No 31 21D35e Rel 2 0 Witschi Electronic Ltd CH 3294 Biren a A N Tel 41 0 32 352 05 00 a a Fax 41 0 32 351 32 92 WITSC h www witschi com welcome witschi com NER TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO Congratulations You made the right choice You bought PROOFMASTER S a test instrument answering to the highest technical standards while being easy to handle and operate PROOFMASTER S has been especially designed for repair service centres specialised trade outlets and for testing small series Pre set universal programs custom programmable test cycles and a friendly menu oriented approach provide the highest possible ease of use and ensure that all watches will be reliably tested Watches with or without wristlet can be tested The special Leak Finder Program opens a new dimension in the hunt for watches leaks under water The instrument equipped with a USB interface can be networked by using the optional WiCoTrace software This manual helps operating PROOFMASTER S efficiently and adequately It also provides important advice regarding safety and maintenance of the instrument Always keep this manual ready to hand and pass it on to
16. any new user If properly handled and maintained your new instrument will last for many years Using it will be both enjoyable and cost effective TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N PROOFMASTER S is a powerful and accurate instrument for seal integrity testing of watches The instrument must be exclusively used for its intended aims and purposes PROOFMASTER S is manufactured according to the current state of the art and standard rules of safety technique There remain however a number of residual risks mostly due to inappropriate use or breach of the instructions provided in the user manual The Manufacturer Witschi Electronic Ltd in CH 3294 BUREN a A SWITZERLAND takes no responsibility whatsoever for damages due to inappropriate handling such as damages to the test instrument to watches and or people Tests with Proofmaster S are in general extremely reliable The Witschi Electronic Ltd Company does not however offer any warranty that a watch having successfully passed the test with PROOFMASTER S will be truly water resistant when immersed in water No responsibility will be taken for damages that can have been caused by the lack of water resistance of a watch having been tested TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N
17. arameter e Press escape to leave the Editing menu Note The changed parameters are only stored if you have selected Save parameters yes last point in Editing menu and confirmed with enter 4 2 TEST WITH ONE OR TWO PRESSURES In the case of a vacuum test the bottom of the case and the co1 Test with 1 2 glass show a tendency to be lifted This provides a way to Pressures 2 detect certain faults that are not coming to light under a Change Parami pressure strain Let us mention e g insufficient pressure of the case bottom or of the battery cover on the O ring Quit esc Therefore to obtain an exhaustive test we recommend performing one vacuum test and one pressure test 4 3 LOWER PRESSURE The first test is normally performed with vacuum You can also select a lower pressure instead of vacuum if it better fits Vacuum bar 0 2 the characteristics of the watch Change Param lt Proceed with care when testing jewellery watches and very flat watches with vacuum Vacuum could crack the bottom or the glass We advise you to test such watches with a C01 Lower pressure Quit esc vacuum not exceeding 0 3 bar Robust watches in particular watches with a hard case diver watches must be tested with a vacuum between 0 5 to 0 7 bar The resulting deformation is large enough for a fast and reliable test The vacuum generator can provide a vacuum up to 85 of the normal pressure environment Therefore
18. constant pressure is applied if the watch is leak tight If the watch is not leak tight the air flowing into the watchcase or out of it causes a pressure equalisation that reverses the deformation The speed of the deformation reversal depends on the leak size 6 2 TEST SEQUENCE Locking After the start the sensor is lowered automatically on the watch The chamber is closed and locked Initial value of the deformation The initial value of the deformation is measured after the chamber has been closed This value constitutes the zero of the deformation measurement Pressure build up The first test pressure usually vacuum builds up in the test chamber after the zero of the deformation has been determined Big leakage Important leakage Big leakage If no significant deformation is observed following a pressure build up this means in general that the watch has a big leak that causes an immediate pressure equalisation In this case the test is aborted The limit depends on the measurement parameters and is displayed during the stabilisation process Important leakage The test is also aborted if the leakage exceeds the range detectable by the instrument TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Vv Stabilisation Most watches show a more or less distinctive plastic behaviour In other terms the deformation still increases so
19. cy class Test mode may be programmed e The practical achievable accuracy is determined by the stability and reproducibility of the deformation of the watch to be tested e Ifthe measured result is within the scattering range of the tolerance value the measurement is automatically repeated to obtain a secure result Handling e Test programs and parameters are selected and programmed with a cursor button e The operator is assisted in with messages and instructions e 7 selectable languages for the operator interface e The test is started by closing the cover e The cover is closed manually It is automatically locked and automatically opened at the end of the test e The sensor is manually lowered on to and lifted off the watch Result print out e A test protocol with configurable title line date and time can be printed out via the optional printer Connection to PC e For storing and statistical evaluation of the results a PC can be connected via the USB interface The appropriate demo software is provided The full version is available as accessory Maximum watch size e Diameter 70 mm Thickness 28 mm Width across the watch strap 85 mm Height of support 75 mm Display e Alphanumerical LC display with 4 lines of 20 characters each with LED background illumination Casing e Test chamber and cover of aluminium painted silver grey Stable plastic casing dark grey Front panel aluminium e Dimensions 250 x 250 x 240 mm W x
20. e chamber Clean them if necessary Chamber not tight Never use any cleaning agent or grease that might attack rubber It is enough to wipe off with a soft micro fibre cloth Power interruption A brief power interruption has taken place Switch the Switch off instrument off and then on again The test chamber cannot be locked or unlocked due to a Motor mechanical or electrical fault Switch off Switch the instrument off and on again the motor will attempt to open the chamber once more Switch off the instrument if the chamber remains closed Proceed to Chapter Fehler Verweisquelle konnte nicht gefunden werden Fehler Verweisquelle konnte nicht gefunden werden on page Fehler Textmarke nicht definiert TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 7 2 EMERGENCY RELEASE Never try to force the chamber open Proceed as follows if a watch is held in the chamber e Switch off the instrument with the mains switch O This releases the remaining pressure in the chamber e Disconnect the instrument from the compressed air supply e When the chamber is no longer under pressure the hook can be pressed downward by applying some amount of force with the finger The test chamber can then be lifted manually In the unlikely case where the outlet valve doesn t open when the instrument is switched off and the press
21. g Glass Bottom 28 mm 1 2kg 2 5kg 4 3kg 6 2kg 12 3kg 30 8kg 61 6kg Glass Bottom 36 mm 2 0kg 4 1kg 7 1kg 10 2kg 20 4kg 50 9kg 101 8kg 4 5 STABILISATION TIME The deformation of the watch still shows some increase after the pressure has been applied plastic deformation It is therefore necessary to wait for a certain time before the actual measure can start C01 Stabilis time 15 600s Auts Auts Change Param We recommend using the Aut s mode In this case the shortest measuring time needed is automatically determined according to the behaviour of the watch Quit esc 4 6 MEASURING TIME The time needed for a reliable measurement depends on the measured deformation and the programmed water resistance limit C01 Measuring time 15 600s Auts Auts We recommend using the Aut s mode In this case the shortest measuring time needed is automatically determined according to the measurement parameters and the Change f Param Quit esc deformation of the watch TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO 4 7 TIGHTNESS LIMITS CO1 Tightness limits Allows entering the water resistance limit in per minute Standard min 1 0 based on a 2 bar test pressure If the programmed test pressure is different from 2 bar the instrument automatically Change Param HA
22. ination of the of the minimum required test time based on the programmed tolerance values the measurement resolution and the measured deformation Alternatively manual input from 10 s to 600 s e Automatic repetition of any measurement with an uncertain result Stabilisation time e Automatic transition from stabilisation to measurement when the necessary stability is reached Alternatively manual input of 10s to 600s Leakage limit e Programmable from 0 1 to 3 min based on a test pressure of 2 bar Automatic conversion to other test pressures Measurement of the deformation e Resolution 0 01pm internal resolution 0 02um e Minimum watch deformation required Depending on the watch type and the test parameters typically 0 5um TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO Result display e Percentage regression of the deformation based on one minute measuring time Range 0 to 9 9 resolution 0 1 e Automatic evaluation leak proof leaking e During the measurement continuous display of the deformation status display for the measurement and count down counter for the remaining measuring time e Display of date and time Accuracy e The resolution of the measurement is dependent on the programmed tolerance value With the predefined programs the resolution is 2596 of the tolerance value With the customer specific programs the accura
23. installed firmware version Calibration date Last calibration date 30 07 2008 The locking mechanism opens at the same time and the test chamber is lifted After this the program which was selected prior to the last time the instrument was switched off will be activated To change the display language to the one you want select another language according to chapter 5 1 Language selection page TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Operating Elements and Display The operating elements on the front panel offer the following functions print Press this button to print current test results measurement and system parameters if the optionally available printer is connected e escape Press this button to stop a running test From a special test program this takes you back to the program 7 Safe Test Back to main menu if you edit a program e select By pressing the cursor button down or up you select a test program or change a selected parameter By pressing the cursor button to the left or right you select a parameter in the programming mode and the Leakfinder program in test and result mode enter By pressing the cursor button vertically you confirm the selection and continue to the next input step The PROOFMASTER S is equipped with an illuminated LCD display with 4x20 characters The instrument communicates in 9
24. it is impossible to generate the maximum vacuum of 0 8 bar at high altitude The table below indicates the highest possible vacuum values Altitude 0m 500 m 1000 m 1500 m 2000m Maximum vacuum 0 85 bar 0 81 bar 0 76 bar 0 72 bar 0 68 bar 4 4 HIGHER PRESSURE C01 Higher pressure 0 2 10 bar 2 0 f For the last test we recommend testing at the highest pressure Change Param admissible for this watch The second test pressure must be higher than the first one Quit esc TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N So called water resistant wristwatches should according to the corresponding standard be tested at a pressure not exceeding 2 bar Watches wearing a label for the maximal pressure or the maximal water depth can be tested up to this pressure 10 m water depth 1 bar pressure WATER RESISTANT 30M Do not test the watch at too high a pressure The plastic case of a watch in particular can be so heavily compressed that the movement will be damaged The table below indicates the internal Vacuum and external Pressure forces acting on the glass and bottom of the watch case Vacuum Pressure Watch size 0 2bar 0 4bar 0 7bar 1bar 2bar 5bar 10bar Glass Bottom 20 mm 0 6kg 1 3kg 2 2kg 3 2kg 6 3kg 15 7kg 31 4k
25. lution determines the smallest measurable unit for deformation reversal With PROOFMASTER S the measurement resolution is determined by the programmed water resistance limit and the selected test mode Lab Stnd Prod see Chapter 4 7 Tightness limits page 19 With the pre set programs the instrument measures with a 25 resolution of the water resistance limit i e 4 times more accurately than the programmed tolerance limit The higher the measurement resolution the more accurate the measured leak rate with correspondingly longer measuring times The value for customer specific programs can be selected between 5 Labo 25 Stnd and 40 Prod measurement resolution values TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO 6 8 DETRIMENTAL FACTORS IN TEST Badly positioned watch The watch when it is tested must be in a stable position on all three supporting points of the support table It is for instance impossible to carry out a reliable test if one side of the watch rests on the wristlet Indeed the band undergoes an unpredictable deformation during the measurement process The wristlet must be removed if necessary If the size or the unsuitable shape of the watch prevents it to be placed correctly use one of the three supplied supports It is often difficult to position watches with a protruding battery case It may be better to
26. mewhat after the application of the pressure and needs a certain time to reach a stable final state The increase of the deformation can mask a small leakage Therefore accurate measurement cannot be performed during this time The stabilisation time ends automatically when the deformation is stable enough for allowing accurate measurements to take place The limit depends on the programmed watch characteristics and on the test pressure Measurement PROOFMASTER S automatically computes the shortest measuring time needed for a reliable measurement When the measurement process is started the last deformation value is stored as the zero The variation relative to this initial value is continuously evaluated during the measurement process Should a watch not be leak tight the measurement process will be terminated as soon as a deformation reversal corresponding to the water resistance criterion is detected The measurement is automatically repeated if the instrument does not reach an unambiguous result In such a case Repeat appears on the display for a few seconds The percent modification of the deformation converted to a measuring time of 60s and the pass fail evaluation are displayed at the end of the measurement process Test at the next pressure The test with the first pressure value is followed by a test with the second pressure value according to the sequence described above Pressure release chamber opening The test relea
27. n the display By pressing the cursor button up or down you can select the Change language desired value from the available parameter values In some English settings you can choose between Yesand No By pressing the button to the right or by pressing enter you Change f Param lt get to the next parameter in the menu System parameters At Quit esc the end you return to the main menu You may also return to the main menu at any time by pressing the escape button TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 5 1 LANGUAGE SELECTION You can select one of the following languages DANSK DEUTSCH ENGLISH ESPANOL FRANCAIS ITALIANO NORSK SUOMI SVENSKA 5 2 DisPLAY OF THE NUMERICAL RESULTS You can select if the numerical results should be displayed and printed out in addition to the good poor evaluation 5 3 AUTO PRINT You can select if after each test the results are printed out automatically or only by pressing the print button The standard setting is AutoPrint no 5 4 COUNTER Counter for the measurement cycles that have been carried out Can not be reset to to 0 5 5 IDENTIFICATION NUMBER Serial number of the instrument Set by the Witschi Company 5 6 CALIBRATION DATE Last calibration date Set by our calibration service 5 7 CALIBRATION VALUES With the next 3 menu points the values of the last calib
28. nd Prod MStnd accuracy Change ft Param lt The greater the measurement accuracy is selected the longer are the measuring times Quit esc LAB Very accurate measurement with long measurement time intervals for lab experiments and tests Corresponds to a 5 resolution of the set water resistance limit STND Normal test mode with adequate accuracy for a reliable measurement Corresponds to a 25 resolution of the set water resistance limit PROD Accelerated test mode for the production with a somewhat reduced accuracy Corresponds to a 40 resolution of the set water resistance limit Take into account the fact that the measurement accuracy and reproducibility reached in practice primarily depend on the behaviour of the watch under the test conditions and not on the accuracy limits of PROOFMASTER S CO1 Save parameters 4 11 SAVE PARAMETERS Yes No Yes Select Save param and acknowledge with Yes if you want to Change Param gt store the modified parameter Save If the parameters are not stored these are only active for as long as the relevant program is selected 5 SYSTEM PARAMETERS The instrument can be configured to meet your requirements with the System parameters menu This menu also provides important information concerning your instrument To access the menu hold the cursor button enter pressed down approx 4 s when switching on the instrument until the setting Change language appears o
29. ng MG NGANGA AA 21 5 9 Default parang AANGAL ALANGAN 21 Tlf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk 6 10 Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N INFORMATION RELATED TO TESTING 22242222220200020200000n20000nnn0n00nnnnnannnnnnnnnnnnnnnnnnnnnnnnn 22 6 1 Meas UnInG FING ale een 22 6 2 Jest sequence ee ee 22 6 3 DTN ON 2AM NANAMAN ANA NGA 24 64 Lesk Ga NAA GA ANU 24 65 Tolerance lit ea 25 6 6 Reference to ISO Standard NIHS92 10 222 2u22202u0n20n200n0nnu0n00nnnnnannunnnnnnnnn ann 25 6 7 Measurement resolution precision uruasre0nnannennannnannannnnnnunnnnnnunnnnnnannnnnnnnnn 26 6 8 Detrimental Factors in Test anne 27 Faulty Operating 28 7 4 Error MES58g8 nu nn 28 212 Emergency IGISA AABANG AGAIN 29 MAINTENANCE AND CALIBRATION iman ana kA AA 29 8 1 POU EES ESI Sue gene 29 8 2 Compressor JUMAIT waaa neuen anne euere 29 DISPOSAL OF THE INSTRUMENT 2 aan 30 TECHNO NTN AGA LAAL 30 TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 1 INSTALLATION 1 1 SCOPE OF SUPPLY e Testing instrument PROOFMASTER S e Compressed air hose 6 4 mm with plug nipple e Set of 3 carriers for watches e Mains adapter 9 V 1 2 A optionally for 230 V or 120V e Dust protection cover e UBS Cable for connection to a PC e Operating Manual e Demo software Proofmas
30. od poor display illuminated yellow In this case we recommend increasing the measuring time or setting to Aut Leakage rate If the measurement is within the numerical measuring range a positive or negative percentage value appears next to the good poor display as a result in the display at the end of the test This value represents the percentage change of the deformation during the measurement The value is converted to a measuring time of 60s The display of the numerical results may be switched on or switched off see Chapter 5 2 Display of numerical results Additional information relating to the leakage rate may be found in Chapter 6 4 Leakage rate 3 LEAKFINDER PROGRAM PROOFMASTER S is the first instrument of its kind worldwide with the so called Leakfinder program developed by Witschi Electronic This program conditions a watch that isn t water resistant for locating the leak under water 3 1 HOW DOES THE LEAKFINDER PROGRAM WORKS The watch is set under pressure for a significant amount of time Let us suppose that the watch is not water resistant In this case the air penetrating into it causes an excess pressure When the TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N watch isn t under pressure anymore the incoming air escapes again through the leak Bubbles arising when the watch is under water indicate the position of the leak
31. place the watch with the glass looking downward Adhesive label or protecting foil on the bottom of the case An adhesive label on the bottom of the case causes a strong plastic deformation during the test since the adhesive layer is slowly pushed in by the pressure of the support points This effect can simulate a leak with vacuum testing It can hide a small leak with pressure testing Therefore any label and protecting foil must be removed prior to testing A thin protecting lacquer applied to the bottom of the case usually doesn t have any significant influence on the test The watch temperature and the ambient temperature differ A watch cooling down or warming undergoes a deformation in a completely unpredictable way This may completely alter measurement results Therefore a watch that has been worn on the wrist cannot be tested immediately To carry out reliable tests we recommend Iying the watches to be tested for at least an hour near the test instrument so that they can reach ambient temperature The PROOFMASTER S too must be at ambient temperature when testing Thus the instrument should not be placed close to heaters or other heath sources Shocks and vibrations during the test Shocks and vibrations can cause some shift on the transducer whereas the test result will be invalidated Previous test with higher pressure Following a high pressure test it is necessary to let the watch recover first for a few minutes
32. ration are displayed which are of interest to our calibration service 5 8 PROGRAMMING THE HEADER Witschi Electronic The header for the printout of the results may be programmed according to your desires Upon delivery of the instrument the CH 3294 Bueren a A following header is programmed Change Param lt Save The selected character flashes Press the cursor button down and up to change the character Press enter to move the cursor to the next character which you can change as the previous one etc Once you have entered your header max 2 x 20 characters press the cursor button to the right to get to the next parameter 5 9 DEFAULT PARAMETERS Allows resetting all modified parameters to their default values Caution All customer programs are reset to their initial values TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO 6 INFORMATION RELATED TO TESTING 6 1 MEASURING PRINCIPLE PROOFMASTER S tests the seal integrity of watches according to the deformation principle The watch to be tested is fixed in a very sensitive path measuring system that continuously measures and monitors the changes in the deformation of the watch The watch is submitted to a pressure or a vacuum after the measuring chamber has been closed This causes the watch to undergo some contraction or expansion This deformation stays constant as long as a
33. rograms adaptable to your requirements See Chapter 4 CUSTOMER SPECIFIC PROGRAMS 2 4 POSITIONING THE WATCH The watch must be at ambient temperature for the test to be reliable Results can be invalidated by variations in the watch temperature during the test Test results can be invalidated by adhesive protection on the bottom or on the glass of the watch Lift the sensor with knob A When lifting the knob the sensor is unlocked and can be freely moved up and down TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Place the watch on the three point support of the support table Check that the watch lies horizontally If the bottom of the watch is not flat battery case it may be better to place the watch with the glass looking downward Wristlet should be placed on the three point support in such a way that the watch is in a stable position Move the sensor at the knob A slowly downward until it rests upon the watch The sensor must rest upon the watch through its own weight When releasing the knob the sensor is blocked The signa LED lights up when the sensor is correctly lowered TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 2 5 TEST Once you have positioned the watch press the test chamber down until it locks The test
34. sequence starts automatically The test sequence consists of the following steps e Filling up the first test pressure e Stabilising e Measurement e Filling up the second test pressure e Stabilising e Measurement e Pressure release The test sequence can be followed on the PROOFMASTER S display The program number the current test and the current test pressure are displayed in the first line P1 Test 1 0 205 bar Measure 77s The second line shows the status of the test stabilising Deformation 2 0 0 04u measuring and the remaining maximum time to the end of the current test step During stabilisation the current deformation is displayed in the third line During the measurement the current change of the deformation in and in ym is displayed At the end of the measurement the results and the good poor 0 2 2 0 bar evaluation are displayed Date and time are displayed in the Test 1 okay 0 4 last line Test 2 okay 0 2 At the end of the test the test chamber opens automatically 26 07 2008 09 16 Move the sensor up with knob A and remove the watch Comprehensive information concerning the test procedure is described in Chapter 6 2 7est sequence page 22 2 6 INTERPRETATION OF THE RESULTS The possible results of a test are listed in the following Once the pressure has been applied the deformation is less Big leakage or not than the minimum value which is determined by the Compressible watch
35. ses the pressure after normal or abnormal termination If a bigger leak has been detected during a test the pressure will be lowered gradually and slowly to reduce an excess pressure inside the watch that could damage it Once the pressure has been released t the chamber is automatically opened TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 6 3 DEFORMATION With every PROOFMASTER S test one must keep in mind the fact that the measured deformations are extremely small and that even minimal influences can alter the measurements For instance A stable watch diver watch undergoes a deformation of about 3 um 3 thousandth parts of a millimetre under a pressure of 0 5 bar This watch will be rejected as bad if the deformation has regressed by more than 0 06 um 0 06 thousandth parts of a millimetre after one minute The more stable the watch being tested and the smaller the selected test pressure therefore the smaller the deformation and the trickier the measurement Therefore a fast and reliable measurement always requires the highest possible pressure or vacuum VAKUUM 6 4 LEAK RATE PROOFMASTER S computes the deformation reversal as a percentage of the initial deformation based on a measuring time of 1 minute The percentage deformation reversal for a given leak size depends on the test pressure Negative percentage A
36. st program Time Auto Pr 2 0 bar The Leakfinder Program is selected by pushing the cursor to Start gt gt Locking the chamber the left or to the right Leak Finder Progam Selection of parameters When selecting the Leakfinder Program the following parameters are automatically set e Type The case type is automatically taken over from the test program previously used e Time This parameter is set on Auto by default The time needed is automatically determined based on the leak size We recommend using Auto for a riskless water test You can set here the time during the watch will be under pressure from 60 to 600 s and Auto e Pr Normally the pressure is taken over from the program previously used It is also possible to enter another pressure Test sequence If the watch has not already been placed on the sensor put it in position and lower the sensor Select the Leakfinder Program and close the chamber The watch check and preparation for the water test runs down Depending on the size of the leak this can take up to 10 minutes TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers NO One of the messages below appears at the end of the test BIG LEAKAGE WATER TEST IMPOSSIBLE The watch leak is too big The watch must not be immersed in water In the case of a very big leak this message is displayed before the end of the check NO OR SM
37. ter M S for PC 1 2 LOCATION A reliable test requires the instrument to be at the ambient temperature Thus the instrument should not be placed close to heaters open windows or exposed to the sun 1 3 MAINS CONNECTION A mains adapter supplies PROOFMASTER S with a 9 V output alternating voltage and a maximum current of 1 2 A The mains adapter is available for a mains voltage of 230 V range 210 V 240 V or for a mains voltage of 120 V range 110 V 130 V Before connecting for the first time ensure that the voltage of the mains adapter corresponds to the voltage of your mains Only use the original mains adapter Connect the mains adapter to the 9 7 24 socket on the rear panel of the instrument We recommend disconnecting the mains adapter from the mains during longer idle periods e g vacations 1 4 COMPRESSOR PROOFMASTER S only needs compressed air for operating since a vacuum generation device is already included The instrument can be connected to various compressed air sources To benefit from the entire pressure range of PROOFMASTER S it is necessary to dispose of a compressor with an output pressure of 11 bar Since standard compressors only provide a pressure of 6 bar we recommend using our special JunAir 6 M compressor instead TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N Kindly never fail to pay attention to
38. ty residential area EN 50082 2 Interference immunity industry 73 23 EEC Low voltage directive EN 61010 1993 Electrical safety 89 392 EEC Machinery EN 292 1991 Page32 TIE 86 42 52 99 Fax 89 41 37 45 lindholts email dk www liA High Fipctronic Ltd
39. ube Only unplug the tube when it is not under pressure Remove the fast coupling from the compressor before unplugging the tube The supplied compressor is equipped with a safety coupling compatible with the connecting tube of the PROOFMASTER S Plug in the extremity of the tube with the plug in nipple into the safety coupling of the compressor To unplug the plug in nipple first shift back the black sleeve of the coupling pressure relief and then forward release TIf 86 42 52 99 Fax 89 41 37 45 lindholts email dk www lindholts dk Lindholts Fourniturehandel ApS Sportsvej 6 8930 Randers N 1 5 CONNECTING THE PRINTER The printer interface for connecting the optional printer with the supplied connecting cable is located on the rear panel of the instrument The results can also be printed with the printer of a PC by connecting this PC to the printerinterface This requires the Auto Print software available as an option with the suitable RS cable 1 6 CONNECTION TO A PC The instrument can be connected to a PC with the USB interface on the rear panel The transfer of results to the PC and their evaluation requires the Proofmaster M S software available as an option 2 OPERATION 2 1 SWITCHING ON Switch on the instrument with the I O switch on the rear side of the instrument The following display briefly appears for approximately 3 seconds Witschi Electronic Proofmaster S V1 00 Instrument name and number of the
40. ure is not released contact our customer service Warning An instrument that doesn t open is the symptom of a serious technical problem Kindly contact our customer service 8 MAINTENANCE AND CALIBRATION 8 1 PROOFMASTER 5 PROOFMASTER S does not require any special maintenance Clean the O ring and the sealing surfaces from time to time with a soft cloth After 2 years or after 100 000 test cycles at the latest a message appears on the screen when the instrument is switched on It reminds the user to have the instrument undergo a calibration and a functional check The date of the last calibration appears on the display when the instrument is switched on Kindly contact our customer service at our headquarters or one of our agencies to have a calibration performed 8 2 COMPRESSOR JUNAIR Warning PROOFMASTER S can be damaged if condensation water penetrates into the instrument together with the compressed air Carry out a regular maintenance of the compressor and air filter Draining of condensation water It is necessary to drain the condensation from the air filter and pressure vessel after about every 1000 test cycles or at least once a month Kindly refer to the manual included with the compressor Oil level It is necessary to check the oil level after about every 1000 test cycles or at least once a month The oil level must appear in the glass window If you use another compressor type refer to the information for operation