OPERATORS GUIDE RocketCRAFT™ Oocyte
Contents
1. 1991 2014 Unwrap the footswitch and attach the air tubing to the male female ports on the front of the pump casing The footswitch is normally left attached to unit in daily use unless removal for storage is required Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 A N WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements WARNING Ensure the set vacuum level is appropriate to the patient s needs taking into consideration the needle and tube set configuration Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 10 11 12 Copyright 1991 2014 Connect the correct mains lead to an electrical supply 110 240VAC 50 60Hz Model dependant Turn the power switch 0 I front panel to on The green light illuminate Connect the short limb of the water trap set to the port indicated or Using aseptic technique un pack the R5 685 Rocket CRAFT Water Trap Set and attach the short tubing length to the spigot indicated Pass the longer patient connection tube into the operative field The R57685 water trap set is for single patient use and must be replaced for each patient Refer to Table 1 for recommend vacuum values for various needle and tube set combinations Occlude the patient tube set distal to the water trap Activate the foot switch Rotate the control knob clockwise
2. Units manufactured before 2014 The year in which the device was manufactured is indicated by the first 2 numbers of the serial number For example a serial number starting 11180776 indicates the device was manufactured in 2011 For units manufactured after 01 04 14 the year of manufacture is shown on the rear rating plate label opposite the model number 10 RETURNING THE PUMP FOR SERVICE All devices to be returned must be prepared as described below for the protection of the servicing team and for safety during transport 1 eg dS Surface clean the pump as described in the Section above Seal in a plastic bag and seal within a second plastic bag Place in the original packaging Enclose the following information e Contact name e Centre address e Decontamination Certificate e Description of the fault or service required e Accompanying Order to authorise servicing contact your local Customer Services Team for details E Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide I 1 STORAGE The device must be stored and operated in temperatures 5 C to 35 C The device must be stored in a clean dry condition ideally in its original packaging which should be retained to return the unit for servicing Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediate
3. Rocketmedical OPERATORS GUIDE Rocket CRAFT Oocyte Aspiration Pumps SG CL Ce E Rocket Medical plc Sedling Road Washington Tyne amp Wear NE38 9BZ England T 44 0 191 4196988 F 44 0 191 419 6989 E customerservices rocketmedical com www rocketmecdhcal com Registered Office Rocket Medical plc Imperial Way Watford WD24 4XX England Company No 03276608 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide I CONTENTS TOONTEND sce at ss cep E te etn eae an ie ade ec ead ences E ee nes eee 2 2 GENERALAS OG 3 2 1 ROcket CRAFT M DUO VAC SUCTION PUMP sarracenia a a e aoa aao TAa a anaa 3 2 2 ROCKket CRAFT Suction PUMD sca teccctsatieisinat aen a a a E ea a A aaa na A AE E 4 GENERA CIR eege 5 E er El A d En BER 5 32 MODEL NUMBER S ene en en ne ee ne ane ee 5 EU E REV ION EE 5 IA MANUFACTURER EE 5 6 Peo ered ERVICE E RE 5 e OAFE TY NIN Eege 6 74 Wi Ute GA ee Ed WY ee GR a erent ne ene ene eee ee een eee 7 4 2 ELECTROMAGNETIC COMPATIBILITY scaccseticcnccastegnrcastasncteelie brcactetgeraedcnncanit ios beneniadjccceiseereaniocaciveebanntesscartcere 7 4 3 PACKAGING EE T 4 4 POSITIONING and PLACEMENT of the DEVICE nannnnnnnnnnnnnnennnnnnsrnnnnnsennnnssnnrnrnsrnrrresennrrrnsnnrrrrserrrnsesrnrrrnnennenno f 4 5 SYMBOLS USED ON CRAFT SUCTION amp DUO VAC PUMPS n sasannnnnnennnnunsnnnnnnennnnnsennrrensnnrrrnssnrrnrssnnrrnsrnrrrenne 8 4 6 SYMBOLS USED ON R57685 PATI
4. ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent N WARNING Ensure the set vacuum level is appropriate N WARNING No user serviceable parts inside SVE WARNING Device can cause explosion in the presence of flammable gases 4 SAFETY INSTRUCTIONS This manual describes the operation and intended use of the device and the associated consumables and it is essential that you use this document to familiarise yourself with the correct function and operation of the device before use Failure to follow these instructions may result in serious injury to the patient or operator and can lead to damage or breakdown of the device In case the device fails during an operation a replacement device and replacement disposables should be kept within reach so that the operation maybe completed This manual does not provide a detailed description of the oocyte harvesting procedure and is not intended as a training guide for users inexperienced in the technique The device must only be used with R5 685 Rocket CRAFT Water Trap Patient Connection Sets or the water trap set supplied with the pump The usage of non approved tubing or filter sets may impair pump performance lead to increased risk to patients and harvested oocytes and will invalidate the Warranty The water trap sets are designed to prevent fluid contamination of the vacuum pump If the devi
5. LRQA CE mark Revision 24 Gauge images water trap cleaning ins 3 4 MANUFACTURER Rocket Medical plc Sedling Road WASHINGTON Tyne amp Wear NE38 9BZUK 3 5 SERVICE AGENTS Rocket Craft Pumps typically require little routine maintenance however they must be serviced and calibrated annually at a Rocket Medical plc service facility 25 05 10 07 06 10 28 10 11 21 11 11 01 12 11 05 12 11 07 12 11 07 12 11 30 05 12 29 06 12 30 09 13 06 11 13 14 11 13 15 04 14 16 07 14 27 08 14 17 12 14 Failure to service the pump at the indicated intervals may invalidate the Warranty Customer Services Rocket Medical plc Sedling Road WASHINGTON NE38 9BZ ENGLAND Tel 44 0 191 419 6988 Fax 44 0 191 419 6989 Email homesales rocketmedical com US Office Rocket Medical Unit 1 150 Recreation Park Drive HINGHAM MA 02043 USA Tel 1 781 749 6223 Fax 1 781 749 6235 Email usa rocketmedical com Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide WARNING This manual contains important information Please familiarise yourself with these Safety instructions before using the device WARNING This device should only be operated by appropriately qualified personnel AD WARNING This device must only be operated with approved consumables J WARNING Protect the device from
6. and the Human Fertilisation amp Embryology Authority HFEA April 2010 as amended 5 3 CONTRAINDICATIONS Not intended for use where ovarian aspiration or the aspiration of ovarian fluid is contraindicated For short term operation only NOT for continuous drainage 5 4 REFERENCES Craft McLeod F Edmonds K Human embryo transfer technique Lancet 1961 ii 1104 5 Craft I Diahanbakch O McLeod F et al Human pregnancy following oocyte and sperm transfer to the uterus Lancet 1992 i 1031 3 Craft I 1984 Clinical Methodology British Journal of Hospital Medicine 90 102 Reeves G Scott R T et al 1989 Journal of Assisted Reproduction and Genetics Volume 6 Number 6 December 1989 Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 ko E WARNING This device must only be operated with approved consumables W d p J WARNING Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent O Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 6 OPERATING THE PUMP Rocket CRAFT Suction Pump 1 Unwrap the footswitch and attach the air tubing to the ports on the front of the pump casing The footswitch is normally left attached to unit in daily use unless removal for storage is required Rocket CRAFT Suction Pump 2 Copyright
7. in cartons 10 units Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide A N WARNING ELECTRIC SHOCK HAZARD Do not immerse the device N WARNING Do NOT attempt to sterilise the device e d P J WARNING Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent N WARNING No user serviceable parts inside ko E HAZARD Do NOT include used consumables as these pose a significant contamination risk e IMPORTANT A decontamination certificate MUST be included with every returned pump Repair or servicing cannot be commenced unless the service agent is in possession of this certificate 8 CLEANING THE PUMP CASING At the end of each clinical session turn off the device at the front panel and disconnect the device from mains power supply Using an aqueous solution of 70 alcohol IMS or isopropyl BP moisten a cloth and wipe all external surfaces of the device If the surface has become contaminated with proteinaceous material remove with a light detergent solution before surfacing cleaning with alcohol Do NOT use a 100 alcohol or any other solvent to clean the device as this may cause damage to the casing surface and display Prevent any fluid from entering the device 9 YEAR OF MANUFACTURE
8. C CEC and NEC requirements N CAUTION Any adjustment modification or repairs to the equipment should be carried out by authorised service agents Disposal of this device must be undertaken with regard to the WEEE directive 2002 96 EC 3 GENERAL INSTRUCTIONS 3 1 COPYRIGHT This manual contains information that is subject to copyright All rights reserved This manual should not be photocopied duplicated or distributed completely or in part without the approval of Rocket Medical plc 3 2 MODEL NUMBERS Rocket Craft Suction Pump Complete 110v Rocket Craft Suction Pump Complete 240v Rocket Craft Duo Vac Suction Pump 240v Rocket Craft Duo Vac Suction Pump 110v R29654 R29655 R29660 R29661 3 3 MANUAL REVISION Update to consumables info voltage Amend voltage statement in Section 4 Update to latex statement and symbol Update to YOM consumption footer Revision 8 New specification based on Rev 7 Revision 9 Updated CSD address graphics Revision 10 Symbol tables amp storage transport Revision 11 CE Mark manufacturer data Revision 12 Revision 13 Terminology change Low Medium Revision 14 Revision 15 CE mark amendment Revision 16 Symbol correction Revision 17 Correction to annotated images Revision 18 Update images with new footswitch Revision 19 Revision 20 Image to clarify footswitch controls Revision 21 Amend EMC revision Revision 22 Revision 23 Update to
9. ENT CONNECTION SET 8 Oe rev INFORMATION esc css Sass eset ace ieine ea see enc E EEEEE 9 Ske OENER AC DE SCRIP TION eee ee ee ee er eee eee 9 Oe IN CP THON E 9 TC ONTRAINDICATION S iraro enn EE EEE eee ere ee ee eee 9 E 4 ma a e 9 6OPERATING THE PUMP EE 10 6 1 TABLE 1 RECOMMENDED VACUUM SETTINGS 2 0 0 iccccccccceeeeeeeeeseeeceeeesaeeseeeeeaeeeeeeeeeaeeeeeessaeaeeeesaeaaaees 12 6 2 CLEANING amp STERILISING THE REUSABLE WATER TRA 13 FON UNABLE EE 14 6 GCCEANING THE PUMP E EE 15 GE al A CO ar ER 15 10 RE TURNING THE PUMP POR SERVICE i rrczevicciincecsacemencncaisanessccdieteitwetnccwencsennaase ctu AEAEE AEN EAEE TETA a 15 Teo TIRAGE EE 16 T TRANSPORTA E EE 16 13 WARRANT Y ircen ria EENEG 17 14 TECHNICAL CIFIC A le 18 DE Wa COIN see teees ete tase cette E E E E E E E E E 18 EE 18 Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 2 GENERAL ASSEMBLY 2 1 Rocket CRAFT DUO VAC Suction Pump oo PF E e SE Illuminated O I Mains Power On Off Vacuum Control Dial clockwise to increase anticlockwise to decrease the set value Vacuum Display mmHg Footswitch connection ports Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets Water Trap Set comprising glass bottle bung angled connectors silicone tube set Recommended for use with R57685 Rocket Craft Pump Water Tra
10. LE EO Device is sterilised by Ethylene Oxide The device is Latex Free CE 0088 Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 5 GENERAL INFORMATION 5 1 GENERAL DESCRIPTION The Rocket CRAFT Oocyte Aspiration Pumps have been developed to provide smooth low volume vacuum at a pre determined negative pressure Vacuum is activated by a foot operated toggle air switch controlled by the surgeon performing the oocyte collection The range of vacuum is infinitely variable from 10 250mmHg in medium vacuum mode and at a pre set 440mmHg 1 in high vacuum mode Rocket CRAFT Oocyte Aspiration Pumps require either a water trap set supplied or R57685 Disposable Rocket CRAFT Water Trap Sets supplied separately sterile and for single patient use A suitable collection test tube for use with oocyte needle sets is B D Falcon test tube No 2001F 17 x 100mm 9 2 INDICATIONS For the generation of medium vacuum between 10 250 mmHg to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures This device may only be used by or under the supervision of appropriately trained personnel in conjunction with clinical practice guidelines such as those published by the National Institute of Clinical Excellence NICE 2004
11. SELECTION The device operates at a voltage 220 240 VAC 50Hz 0 12A or 110VAC 60Hz as appropriate to the model Ensure that the correct power cord is connected 4 2 ELECTROMAGNETIC COMPATIBILITY Rocket CRAFT Oocyte Aspiration Pumps comply with the electromagnetic compatibility EMC limits for medical devices as specified by IEC 60601 1 1990 The device must be operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility 4 3 PACKAGING The packaging has been designed to allow secure transportation of the pump and its accessories After unpacking re assemble and retain the packaging for transport for servicing when required 4 4 POSITIONING and PLACEMENT of the DEVICE Rocket CRAFT Oocyte Aspiration Pumps must be placed on a secure level surface away from sources of heat water splashes mists or cooling vents Do not expose to direct sunlight Do not expose to flammable gases Operating temperature Range 5 C and 35 C Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 4 5 SYMBOLS USED ON CRAFT SUCTION amp DUO VAC PUMPS These tables describe their respective meanings Mains Power ON OFF CE Mark 4 6 SYMBOLS USED ON R57685 PATIENT CONNECTION SET Read the Instruction for Use before connection and use Device is for Single Use Only STERI
12. acuum of 440mmHg 1 Rocket CRAFT Suction Pump To activate the HIGH vacuum occlude the RED button and depress the footswitch the pump will immediately deliver a vacuum of 440mmHg Removing the occlusion will cause the vacuum to revert to its previously set level 17 12 14 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 Rocket CHA TTVN Oocyte Aspiration Pumps Operators Guide ko E WARNING This device must only be operated with approved consumables which are specifically designed to provide the correct flow rates and vacuum characteristics Should any liquid enter the device discontinue use immediately and refer to an authorised service agent lf water media or any other material is allowed to enter the water trap chamber the pump must be stopped immediately Remove the water trap from the device empty any fluid Decontaminate the water trap if required following local hospital guidelines and refit before use The pump must NOT be operated if fluid is present in the water trap Should the water trap become full and there is a risk that fluid has entered the pump it must be withdrawn from service and returned to an authorised service agent for inspection and repair H a LES l ig E a x e TA A AN gt 6 2 CLEANING amp STERILISING THE REUSABLE WATER TRAP This device should be cleaned in accordance with local hospital policy and w
13. arranties herein above set forth Rocket Medical plc makes no warranty of merchantability or fitness for a particular purpose with respect to the Products or parts thereof 7 Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 8 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 14 TECHNICAL SPECIFICATIONS 14 1 CLASSIFICATION IEC 60601 1 Degree of protection against electric shock Type B 14 2 SPECIFICATIONS Power Input to Pump 220 240 VAC 50Hz 0 12A Maximum current 2 5A Consumption 0 6KW h Environmental conditions e Temperature 5 C to 35 C e Atmospheric Pressure Range 7 00 1060hPA Dimensions e W 264mm e H 124mm e D 164mm Weight e Unit 1 9Kg e Foot Switch 0 51Kg Vacuum Ranges e Medium vacuum 20mmHg to 250mmHg in 20mmHg increments e High vacuum 440mmHg Vacuum Range Accuracy 10 Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14
14. ce has been used with a non approved filter set or there is any evidence or suspicion that the pump may have been contaminated with fluid during use it must be removed from service and returned for examination immediately Please contact your nearest Service Centre for advice The use of high vacuum levels may lead excessive fluid flow rates which may result in damage to the oocyte and reduced fertilisation rates Damage to oocytes in harvesting systems is principally caused by turbulent flow which can lead to physical shearing stresses on the cumulus sufficient to denude or damage the fragile zona Reeves et al 1989 Flow rate is a function needle set configuration and vacuum applied HIGH vacuum setting must only be used to clear blockages from a needle set Refer to Table 1 for recommended vacuum setting for given needle set configurations Refer all servicing to the manufacturer s authorised service agent Do not use in an area where flammable gases are present Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements L WARNING ELECTRIC SHOCK HAZARD Do not immerse the device WARNING Device can cause explosion in the presence of flammable gases 4 1 SUPPLY VOLTAGE
15. d guidelines laid down in ISO 17665 1 2006 Sterilization of health care products Moist heat Part 1 Requirements for the development validation and routine control of a sterilization process for medical devices and ISO TS 17665 2 2009 Sterilization of health care products Moist heat Part 2 Guidance on the application of ISO 17665 1 E Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide AD WARNING This device must only 7 CONSUMABLES 1 The following items should be replaced annually oo epee WIN earlier if subjected to repeated autoclaving and 4 approved consumables i we ds which are specifically showing signs of wear or poor sealing qualities designed t ide th milena Ge Ge a R28406 Water trap bottle amp bung set ee vacuum characteristics b R28135 Silicone tubing including Should any liquid metal luer fitting Oe enter the device 3 discontinue use immediately and refer ee to an authorised K service agent i 2 For users who prefer not to reprocess the water trap a sterile single patient use option is available and LI replaces the non sterile bottle set which is supplied ilies with the pump 7 R57685 Rocket CRAFT Water Trap Patient Connection Set glass water trap bottle 2 5m patient connection tube Va with male luer connector Supplied sterile for single a gt use
16. ith regard to all applicable regulations including but without limitation to those pertaining to human health amp safety and care of the environment such as those published by the Institute of Decontamination Sciences Standards and Practice 3rd ed as amended 1 2 Remove the water trap bottle and tubing from the pump Separate the tubing and bung set from the glass bottle The silicone tubing does not need to be detached from the bung set unless it has been contaminated in use Wash the tubing and bung set and bottle under warm running water Very hot water will cause coagulation of proteinous substances and should not be used Mechanical washing machines and ultrasonic cleaners using instrument detergents of strength recommended by the manufacturer may be used followed by a clean rinse Neutral ph7 detergents should be used during manual mechanical or ultrasonic cleaning The concentration and volume of detergents used should be in line with the detergent manufacturer s instructions Immediately after washing the bottle and tubing and bung set should be dried completely either by hand or in a suitable warm air cabinet The bung tubing set and bottle should be packed separately in autoclavable pouches Do not sterilise assembled The recommended method for sterilisation is steam sterilisation 134 C 3 C 0 C for minimum of 3 minutes This recommendation should be used in conjunction with the standards an
17. ly and refer to an authorised service agent 12 TRANSPORTATION The device must be transported in temperatures 5 C to 35 C The device is FRAGILE and must be transported in its original packaging to ensure protection If the original packaging is not available please contact your local Customer Services Agent who will provide replacement packaging Weight i i e Unit 1 9Kg Dimensions Foot Switeh e W 264mm 0 57Kg DUOVAC plastic e H 124mm 0 34kg Craft Suction plastic D 164mm 0 51Kg DUOVAC metal 0 38kg Craft Suction metal O O O O Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 13 WARRANTY Rocket CRAFT Oocyte Aspiration Pumps are sold by Rocket Medical plc under the warranties set forth in the following paragraphs Such warranties are extended only with respect to the purchase of the Products directly from Rocket Medical plc as new merchandise and are extended to the first Buyer thereof other than for resale For a period of twelve 12 months from the date of shipment the Products are warranted to be free from functional defects in materials and workmanship and to conform to the description of the Products c
18. ontained in the operating manual and accompanying labels provided the same is properly operated under conditions of normal use that annual maintenance and service is performed at an authorised Rocket Medical plc service facility Removal of any QC seal voids the warranty The foregoing warranties shall not apply if the Products have been repaired other than by Rocket Medical plc or other than in accordance with written instructions provided by Rocket Medical plc or altered by anyone other than Rocket Medical plc or if the Products have been subject to misuse negligence or accident Rocket Medical plc s sole and exclusive obligation and Buyer s sole and exclusive remedy under the above warranties is limited to repairing or replacing free of charge at Rocket Medical plc s option Products which are reported to Rocket Medical plc by mail telephone or email and which if so advised by Rocket Medical plc is thereafter returned with a statement of the observed deficiency not later than seven 7 days after the expiration date of the warranty to Rocket Medical plc during normal business address transport charges prepaid and which upon Rocket Medical plc s examination is not found to conform with the above warranties Rocket Medical plc shall not be otherwise liable for any damages including but not limited to incidental damages consequential damages or special damages There are no express or implied warranties which extend beyond the w
19. p Sets Silicone tube set with metal luer connector Medium Vacuum Standard 50 250mmHg_ amp High Vacuum 440mmHg foot switch Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide Rocket Craft Suction Pump with earlier metal footswitch Illuminated O I Mains Power On Off Vacuum Display mmHg High Vacuum 440mmHg 1 control button SC oe SS Vacuum Control clockwise to increase anticlockwise to decrease the set value 5 Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets 6 Water Trap Set comprising glass bottle bung angled connectors silicone tube set Footswitch connection port Medium Vacuum Standard 50 250mmHg footswitch Silicone tube set with metal luer connector Copyright 1991 2014 Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide WARNING READ THIS MANUAL CAREFULLY Please familiarise yourself with the contents of this manual before attempting to use the device Failure to observe these instructions may result in damage to the pump or cause injury to the patient or user This device should only be used by suitably qualified personnel A N WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IE
20. to increase to the desired value on the gauge To decrease the set value rotate anticlockwise When the desired value is reached release the footswitch and tubing set occlusion Repeat from Step 8 to ensure stable value Rocket Medical plc All rights reserved ZDOCK105 Rev 24 17 12 14 Rocket CRAFT Oocyte Aspiration Pumps Operators Guide 6 1 TABLE 1 RECOMMENDED VACUUM SETTINGS Single Lumen Double Lumen Tube Set Length Recommended Vacuum mmHg 85cm 8 110 130 150 70m 130 150 170 N WARNING Ensure the set vacuum level is appropriate to the patient s needs taking into consideration the needle and tube set configuration 13 Connect the patient filter set to a suitable luer fitting indicated such as that found on the Rocket Oxford Tube Set found on Rocket Single and Double Lumen Oocyte Aspiration Needles and attach a collection tube to the bung If vacuum lines become blocked due to debris or viscous fluids a temporary high vacuum 440mmHg 1 can be obtained to unblock the needle and tubing set by following the procedures below WARNING HIGH VACUUM MODE Rocket CRAFT DUO VAC must NOT be used to ee ees 14 To activate the HIGH vacuum depress may result in damage to the oocyte and lead to reduced fertilisation rates 2 Copyright 1991 2014 15 16 the WHITE HIGH VACUUM Pedal the pump will immediately deliver a v