1. a stable interferogram is another indication that the HWG accessory is ready for measurements An insufficiently cooled detector will produce a low and unsteady signal The PN 350 048100 01 Page 3 red LED light indicates the accessory is not ready for use If the red LED is illuminated during use check that the detector Dewar is filled with liquid nitrogen Place the Diffuse Gold Reference on the sample window to maximize the light level in the sphere The gain of the IntegratIR is set to around 4 5 V with open aperture PN 350 048100 01 Page 4 Adjustment The accessory has been aligned and tested to ensure that it performs to specifications and no additional alignment is necessary Sample Preparation Sampling with the IntegratIR is as easy as placing the sample on the optical window on the top of the accessory For plastic sheets painted test panels and other flat samples the reproducible sample positioning has to be assured If the sample does not lay flat under its own weight a small weight could be placed on the sample to press it against the window or open port if the window plate is removed to aid in the accurate and repeatable sampling Other materials such as powders and pastes can be directly placed on the ZnSe window The choice whether to use the window or to place the sample directly on the integrating sohere depends on the type of sample The window is sealed using epoxy compound and can be cleaned preferably with
2. alcohol or with acetone in the case of more difficult samples In all cases the accessory is measuring the reflected light thus transparent materials films produce very small signal In certain cases a reflective mirror can aid in returning the near infrared light to the sphere detector Please note that the reflectance of the ZnSe affects the measurement the reflected light is added to the reflectance of the sample For absolute reflectance measurement place the sample directly on the sample port of the sphere Use Gold Reference placed on the window for background spectra and for interferometer alignment IMPORTANT NOTE After use please replace protective cap to avoid damage or contamination to gold surface Samples need to be presented to the spectrometer in an optically uniform and reproducible fashion to obtain the same spectrum every time Parameters that have significant effect on the samples are particle size and distribution crystallinity phase melted or solidified oils for example have very different spectra compression temperature especially water containing samples and other physical or chemical differences PN 350 048100 01 Page 5 Sample Reference Selection Mirror On the front of the IntegratIR module there is a knob that selects the sample or reference positions IMPORTANT NOTE Please move the selector in between the SAMPLE and REFERENCE positions only When the selected positions are reached it will click in pla
3. 53719 Printed in the United States of America All world rights reserved No part of this publication may be stored in a retrieval system transmitted or reproduced in any way including but not limited to photocopy photograph magnetic or other record without the prior written permission of PIKE Technologies Inc Address Comments to PIKE Technologies Inc 6125 Cottonwood Drive Madison WI 53719 Phone 608 274 2721 Fax 608 274 0103 E mail info piketech com Web Site www piketech com Jan 1 2012 Contents Introduction Unpacking Your Accessory Packing List Installation MCT Liquid Nitrogen Filling and Stabilization Adjustment Sample Preparation Sample Reference Selection Mirror Measurement without Substitution Typical Spectra Precautions Replacement Parts and Options APPENDIX CO CO CO YJ M DD OW UT WW WI NI N PP Introduction IntegratIR series Mid Infrared Integrating Sohere module is a compact sampling accessory for measuring the diffuse reflectance of a wide variety of solid materials The IntegratIR module features a 3 high reflectivity gold coated integrating sphere and a high speed low noise liquid nitrogen cooled Mercury Cadmium Telluride MCT detector with transfer optics and interface electronics The IntegratIR fits in the sample compartment of the FTIR spectrometer and connects to the internal detector port of the spectrometer The 18 mm diameter opening on top of the integrating spher
4. Installation and User Guide TECHNOLOGIES Specticecopie Creativity j PE IntegratIR Mid IR Integrating Sphere The information in this publication is provided for reference only All information contained in this publication is believed to be correct and complete PIKE Technologies Inc shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing performance or use of this material All product specifications as well as the information contained in this publication are subject to change without notice This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under the patent rights of PIKE Technologies Inc nor the rights of others PIKE Technologies Inc does not assume any liability arising out of any infringements of patents or other rights of third parties This document contains confidential or proprietary information of PIKE Technologies Inc Neither this document nor the information herein is to be reproduced distributed used or disclosed either in whole or in part except as specifically authorized by PIKE Technologies Inc PIKE Technologies Inc makes no warranty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Copyright 1991 2012 by PIKE Technologies Inc Madison WI
5. ce to produce reproducible repositioning of the gold mirror inside the sphere There are two different ways of using the Mid IR IntegratIR The mirror can be left in the SAMPLE position and the background can be measured first using the Gold Reference supplied with the IntegratIR After the background is collected the reference is substituted with your sample and the spectrometer should collect the sample spectrum In this mode the Sample Reference Mirror should be left in SAMPLE position This is called the substitution method because the reference is substituted with the sample Measurement without Substitution The IntegratIR has another measurement mode The Taylor method of measurement is made by LEAVING THE SAMPLE on top of the sphere at the sample port and moving the Selector Mirror to REFERENCE position to measure the background and back to SAMPLE measuring the sample spectrum PN 350 048100 01 Page 6 Typical Spectra The spectral results using the IntegratIR depend to some degree on the spectrometer it is used with Source intensities shape of the infrared beam the beam splitter characteristics vary from bench to bench thus the single beam spectra could be slightly different Using the Gold Reference or other well characterized highly reflective standard and ratioing against this background corrects for the individual throughput characteristics of the spectrometers A typical single beam spectrum and a zero reflectance line are shown b
6. ched to the 15 pin D connector in the sample compartment of the spectrometer Select the appropriate source optical path and detector Please consult the software manual provided by the spectrometer manufacturer concerning how to perform the detector and source selection For best results please use parameters in Appendix A To connect the IntegratIR please turn off the spectrometer After the accessory is connected power up the instrument The accessory can be directly purged via the purge port on the front panel MCT Liquid Nitrogen Filling and Stabilization Fill the MCT detector with liquid nitrogen using the funnel included with the accessory WARNING Wear safety goggles and protective clothing when handling liquid nitrogen Exposure to liquid nitrogen will cause severe skin or eye injury Pour a small amount of liquid nitrogen into the funnel and wait for it to drain into the Dewar of the detector Add additional small amounts of liquid nitrogen until the detector is filled Allow the detector to stabilize for about 10 minutes and then top it off with a final amount of liquid nitrogen to completely fill the liquid nitrogen Dewar A lit green LED light indicates the Dewar contains liquid nitrogen and is beginning the detector cooling process or is sufficiently cooled It is not an indication that the accessory is necessarily ready for use Be sure to wait a 10 minute equilibrium time after the Dewar has been completely filled Observing
7. e is the sample window through which the sample is illuminated The samples are placed horizontally on the sphere for direct illumination An additional sample plate is available with a built in ZnSe window Using the ZnSe window as a sampling port powders can also be measured on the top of the integrating sphere The window is bonded and sealed to protect the sphere from corrosive materials and contamination In addition to the main function of diffuse reflectance measurements the IntegratIR can also be used to measure diffuse transmittance On the illumination port side of the device there is a standard 2 x3 slide position which allows thin samples such as polymer films or other thin sheets of materials to be brought into the infrared beam The integrating sphere positioned very close to the partially transmitting partially scattering sample collects all infrared light emerging from the sample The gold plated sample slide included with your IntegratIR serves as a sample mounting surface and can also serve as an aperture controlling the beam for minimum background scatter rr Detector Cable Connection Temperature Indicators gt Detector Dewar Sample Reference Position Knob E Diffuse Gold Reference Purge Tube Connector with Protective Cap Figure 1 Components of the IntegratIR Accessory PN 350 048100 01 Page 1 Unpacking Your Accessory In order for you to quickly verify receipt of your accessory we have incl
8. elow The enclosed spectrum was recorded at 4 cm resolution showing ambient water vapor and CO Purging the sample compartment will also reduce this effect for increased spectral reproducibility in the water vapor and CO regions Art ary ums Absorbance 0 8 Y __ ____e __ 9 _ o__ 0 0 0 1 Y __ 8 ___ _ 0 _ g __ RU 0 0 8 __ a _ 6 0 0 Figure 2 Single beam energy of the IntegratIR Typical noise level around 1800 2200 cm should be less than 900 uAbs RMS PN 350 048100 01 Page 7 Precautions In order to provide the maximum signal in the infrared with the minimum spectral interference the mirrors used in this device are gold coated Since the coatings are soft care must be taken to avoid damage Normally these mirrors will not need cleaning since they are contained within the housing of the accessory If they do need cleaning they may be gently wiped with a lint free abrasive free cloth such as lens tissue or with a camel hair brush Under no circumstances must the mirrors be rubbed with paper products such as Kleenex since this may produce scratching of the mirror coating Replacement Parts and Options Part Number Description 048 0100 Sample Plate with 18 mm x 2 mm ZnSe Window 048 3000 Diffuse Gold Reference APPENDIX A Recommended Measurement Parameters Range for FT IR Spectrometers Wavelength Range 5000 to 600 cm Velocity select appropriate for MCT detector for your FTIR
9. spectrometer Aperture Open Resolution 4 or 8 cm Number of scans 64 NOTE Before installation please turn the FTIR Bench off In order to install the IntegratIR the detector cable has to be connected to the FTIR spectrometer electronics Please consult your FTIR User Manual for instructions about connecting a sample compartment accessory INSTALLATION OF IntegratIR DETECTOR CABLE PN 350 048100 01 Page 8 PICE TECHNOLOGIES Spectroscopic Creativity 6125 Cottonwood Drive Madison WI 53719 5120 608 274 2721 TEL 608 274 0103 FAX sales piketech com www piketech com
10. uded a packing list Please inspect the package carefully Packing List IntegratIR User Manual IntegratIR Accessory Detector Cable PN 350 0481000 01 PN 048 10XX 048 3000 Quantity 1 Quantity 1 Quantity 1 Gold Reference Mirror with Liquid Nitrogen Funnel Camel Hair Brush Protective Cap Quantity 1 Quantity 1 PN 048 3000 Quantity 1 Spatula Quantity 1 Y PN 350 048100 01 Page 2 Installation Before installation make sure the spectrometer is working in the mid infrared region without the IntegratIR Please consult the manufacturer of the FT IR spectrometer for proper operation This may involve replacing the beam splitter and source to optimize for the 4000 to 600 cm wavelength region of the IntegratIR It is advisable to perform an alignment of the interferometer before inserting the IntegratIR into the sample compartment Each IntegratIR unit is custom manufactured to interface with a specific model FTIR Spectrometer Operation with a different model may require a factory conversion to be done by PIKE Technologies Place the accessory in the sample compartment of the spectrometer The pin mounting of the accessory plate assures the alignment of the accessory relative to the beam coming from the spectrometer Connect the detector cable to the spectrometer CAUTION Make sure power to the spectrometer bench and the IntegratIR are off before y connecting the detector cable For ThermoFisher instruments use the cable atta