Home
User Manual
Contents
1. in manufacturer s package at ambient temperatures from 5 to 40 C and relative humidity up to 80 at 25 C unpacked under conditions at ambient temperatures from 10 to 35 C and relative humidity up to 80 at 25 C 6 2 The storage location should be free of dust chemical vapours aggressive gases and other substances that may cause corrosion Beta spectrometer should be kept away from direct sunlight 7 TRANSPORTATION 7 1 Beta spectrometer in the original package of the manufacturer can be transported by all means of transport at any distance transportation by rail should be carried out in clean boxcars during transportation in the open motor transport boxes should be covered by the water proof material during transportation by air boxes should be placed in airtight heated compartment during maritime and river transportation boxes should be placed in the hold 7 2 Arrangement and fastening of boxes on all means of transport have to ensure stable condition during transportation absence of relocation and shocks resulted from collisions with each other 7 3 It is necessary to follow the requirements indicated on the package during loading and unloading During loading and unloading the beta spectrometer should not be exposed to atmospheric precipitations 7 4 Transportation conditions aPC EE from minus 50 to 50 C provided a smooth thermal stabilization is ensured at unloading to temperatures from m
2. conditions of operation are ensured There are no special requirements for the room where the beta spectrometer is installed 2 1 2 Scintillation detector unit with shielding shall be mounted in a location within a room where the possibility of vibrations is minimal closer to a corner or a wall ofthe room The laboratory bench has to be placed so as to exclude problems with connecting units to the wall outlets and to the protective earth terminals It is recommended to place the monitor so as not to expose its screen to the direct sun light and to ensure protection in the place where the detector is located 2 1 3 Beta spectrometer should be used in rooms where increasing of the natural background is not possible 2 1 4 Beta spectrometer has to be used in rooms where the presence of constant and or alternating magnetic fields with strength more than 40 A m is excluded 2 1 5 Servicing of the beta spectrometer does not require participation of specially trained personnel 2 2 Preparation of the product for use 2 2 1 Assemble the beta spectrometer consulting with the assembling diagram in the Appendix A 2 2 2 Connect cable ofthe detector unit to the USB port of the personal computer in accordance with the wiring diagram in the Appendix B When connecting several detector units of the complex Progress to the same PC units are connected independent of each other to vacant USB ports Use USB splitter if necessary 2 2 3 Turn on the PC Warm
3. in the energy calibration mode During measurement process the software shows on the screen positions channel numbers n and nm determined from the calibration source spectrum these positions correspond to boundary values of apparatus beta spectra of radionuclides Sr and PY 540 and 2280 keV accordingly and the test count rate within energy range from 250 to 500 keV 3 After 150 seconds make a record in the row 1 ofthe Table 4 2 In the columns Reference mark 540 keV and Reference mark 2280 keV channel numbers are to be recorded corresponding to the boundary values of the apparatus spectra and in the column Test count rate the value of measured test count rate Table 4 2 Results of the energy calibration by source Sr Y Reference mark 540 keV Reference mark 2280 keV Test count rate measurement 4 Perform at least five successive measurements of calibration source and record the results into the rows of the table 4 2 4 5 3 2 Measurements of background 1 Remove calibration source out of the detector unit and start spectrum acquisition in the background measurements mode During measurement the software shows on the screen values of the background count rate in the current and previous measurements 2 In case that the difference between current and previously measured count rates in at least one of the checked intervals exceeds the criterion preset in the software it will warn the user about change ofthe
4. the product shall be determined FVKM 412131 002 02RE 11 Appendix A mandatory ASSEMBLING OF BETA SPECTROMETER Detector FVKM 412131 002 02RE 12 Appendix B mandatory CONNECTION LAYOUT Function of USB port contacts PIN SIGNAL 1 5 2 Data 3 Data 4 GND USB m 1 A2 220 V 50 Hz lt 5 Position Item Q ty Note Al Detector unit BDEB3 2U 1 A2 System unit of the PC 1 FVKM 412131 002 02RE 13
5. up the beta spectrometer for 30 minutes 2 2 4 Start the Progress software by clicking on the corresponding icon on the desktop FVKM 412131 002 02RE 5 2 2 5 Detailed description of steps during operation of the beta spectrometer is presented in the following documentation that comes with spectrometer Procedure for performance testing of the spectrometric complex Progress for activity measurement of alpha beta and gamma emitting radionuclides beta channel Procedure manual for measurement of activity of radionuclides using the scintillation beta spectrometer Progress with software Progress Software for complexes Progress for spectrometric and radiometric measurements User manual 3 MAINTENANCE 3 1 General notes 3 1 1 Maintenance of the beta spectrometer shall be performed with the purpose of ensuring operability during all life time Types and terms of maintenance are shown on the section 3 3 3 1 2 Performance testing is carried out during annual calibration of the beta spectrometer in accordance with the procedure described in the section 4 3 1 3 Safety measures indicated in the section 3 2 should be followed during maintenance of the beta spectrometer 3 2 Safety precautions 3 2 1 All work associated with the operation must be performed in accordance with instructions for safety and radiation safety applicable in the enterprise 3 2 2 Maintenance of the beta spectrometer is carried out by perso
6. 25 to 0 31 mg m Chlorides not more than 0 3 mg m day FVKM 412131 002 02RE 3 1 2 14 Casings of the beta spectrometer ensure protection against penetration of solid objects and water in accordance with the IP23 1 2 15 Beta spectrometer withstands electromagnetic interference in compliance with the IEC 1000 4 8 93 IEC 1000 4 9 93 IEC 61000 3 2 95 IEC 610003 3 94 IEC 61000 4 2 95 IEC 61000 4 3 2006 IEC 61000 4 4 95 IEC 61000 4 5 95 IEC 61000 4 6 96 IEC 61000 4 11 94 IEC 61000 4 12 96 IEC 61000 4 13 2002 IEC 61000 4 14 99 IEC 61000 4 28 99 design group III performance criterion A Electromagnetic interference does not cause false triggering and restart of the measurement channel 1 2 16 With regard to seismic stability the beta spectrometer stability against earthquake intensity 7 according to the MSK 64 scale at 30 m relative to the grade level 1 2 17 With regard to protection against electric shock beta spectrometer is classified as class I according to the IEC61010 1 2001 1 2 18 With regard to fire prevention characteristics beta spectrometer complies with fire probability of no more than 10 year 1 2 19 Beta spectrometer withstands the exposure to decontaminating solutions first solution caustic soda 50 60 g l potassium permanganate 5 10 g l second solution oxalic acid 20 40 g l third solution detergents 5 solution of the citric acid in the rectified ethyl alco
7. C Oo3a Scientific Production Company Doza SPECTROMETRIC COMPLEX FOR MEASUREMENT OF ACTIVITY OF ALPHA BETA AND GAMMA EMITTING RADIONUCLIDES PROGRESS SCINTILLATION BETA SPECTROMETER PROGRESS BETA User manual FVKM 412131 002 02RE Content 1 Description and operation of the product srserrerererrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrra 3 1 1 Product ee Oe 3 1 2 Technical characteristics gea ENEE 3 1 35 Configuration EE 4 1 4 Design and operation WEE 4 1 5 Marking and s EE 5 EO Pate ea ts aaea att rs EMNE al Seas EAN EA GEDE Bee een La E 5 2 Intended ET 5 21 Operating TEE een 5 2 2 Preparation of the product for use 3 3 Mant nan EE 6 EE 6 202 Saletv PIECAUHONS I snes corro edet er SA KRA cob dott vere opa SRS 6 3 3 Reeder ee 6 4 Calibration TOVE sses d de ene EN REA bre SRA meos a LER UE Redi m dads 7 Seel requitetments iive gruber eat cepa Eur ege ege EE I 4 2 Preliminary arrangements EE Eege 7 43 Safety precautions EE 7 AA CONGIIONS suse e en vea E a E alata Brr RENS EEN ER 7 AD Ptocedute sot ee e 7 5 Routine Tepas eege teen 10 Dr OOA Earn a n Send a E ee 10 PRA B EDOT OLONA ENA OD AEE S E E 10 e EE 10 Appendix A Assembling of beta spectrometer ssssssssssserererrrrrrsrrrrrrrrrrrrrrrrrrrrrr ent 12 Appendix B Connection layout cis eet t nente x oer aa dpa OQ vL a Retos EE 13 FVKM 412131 002 02RE 2 This user manual FVKM 412131 002 02RE contains information on design pri
8. alibration certificate after correction for decay Table 4 4 Results of activity measurements specific activity of the working standard Measured activity Radionuclide specific activity Uncertainty Reference activity 4 5 4 Processing of results 1 Calculate the mean value of five measured results obtained with calibration source see 4 5 3 1 2 Calculate the mean value A of measured activity of the working standard see 4 5 3 3 3 Calculate relative measurement error of the working standard activity by the following formula AA ER 100 4 1 0 where Ao activity of the working standard 1SO indicated in the calibration standard after correction for decay Bq A mean value of measured activity of the working standard Bq Calculate limits for the basic relative error of activity measurement in percents by the following formula 11 02 657 4 2 where 0 uncertainty of the working standard indicated in the calibration certificate for the standard Calibration results are considered positive if the value of the basic relative of measurement error does not exceed limits stated in the section 1 2 4 FVKM 412131 002 02RE 9 5 ROUTING REPAIRS 5 1 Defective detector units of the beta spectrometer are to be replaced or repaired by the manufacturer 6 STORAGE 6 1 Before putting into operation beta spectrometer shall be stored in the heated and ventilated warehouse
9. atmospheric pressure TE EE from 30 to 80 e ee 86 0 to 106 7 kPa 4 4 2 Operations conducted with calibration facilities and beta spectrometer under test shall comply with guidelines detailed in relevant operation manuals 4 4 3 Before switching the beta spectrometer on and performing the calibration keep it in conditions specified in 4 4 1 for 4 hours 4 5 Procedure 4 5 1 External examination The following should be confirmed during the external examination completeness of the delivery kit presence of operational documentation absence of defects affecting the operation of beta spectrometer 4 5 2 Testing Turn the beta spectrometer on computer and power supply unit and wait until warming up for 30 minutes Start Progress software and initiate viewing mode of the apparatus spectrum of beta spectrometer Insert the calibration source Sr Y in standard position and start measurement in the energy calibration mode in accordance with the user manual for the software Progress A live spectrum will appear on the screen typical shape of the spectrum is shown in the Figure 4 1 FVKM 412131 002 02RE 7 Figure 4 1 Apparatus spectrum of the Sr Y source measured by beta spectrometer 4 5 3 Determination of metrological characteristics 4 5 3 1 Measurements of the calibration source 1 Place the calibration source Sr Y in the detector unit for performing energy calibration 2 Start measurement
10. background spectrum If this is the case eliminate the reason of increased background and perform two successive measurements of background in accordance with step 1 above In case that during the last background measurements a high background warning will appear again results of the calibration should be considered negative with the wording instability of the background readings It is recommended to dispatch beta spectrometer for repair 3 Provided that background change warning does not appear upon completion of acquisition of the background spectrum 3600 seconds record the results of background measurements in the Table 4 3 FVKM 412131 002 02RE 8 Table 4 3 Results of background measurements Count rate in the energy intervals cps 1 2 3 4 Measured background spectrum Previously measured background spectrum 4 5 3 3 Measurement of the working standard activity 1 Perform energy calibration in accordance with section 4 5 3 1 and record results in the table 4 2 2 Remove calibration source out of the detector unit and place working standard 1SO instead 3 Start acquisition of the spectrum in the activity measurement mode by selecting a task MEASUREMENT OF Sr Y 90 ACTIVITY Record measured activity values specific activity and uncertainty of measurement results for the Sr Y working standard used in the Table 4 4 together with reference activity of this standard indicated in the c
11. hol for handling electronic devices 1 220 Shielding Weight siot esse xeu vente tes vice etri Fade Eege Fue E REA rg 30 kg 1 2 21 Overall dimensions of the shielding sssssmssessrssrrrrrrrrsrrerrrrrrrrrr es 220x190x160 mm 1 3 Configuration 1 3 1 Beta spectrometer is a stationary unit with scintillation detector unit personal computer and software Progress intended for control in all modes of operation and all steps of the measurement process for processing of measurement data and recording results 1 3 2 Beta spectrometer includes combined into single case scintillation detector block with plastic scintillator power supply unit and pulse amplifier analog to digital converter ADC counting sample holder and lead shielding block for shielding of natural background radiation Power supply units and pulse amplifiers as well as ADC are placed directly in the detector block Standard planchets are used for measurement of counting samples 1 3 3 Calibration source Sr Y on a special disk for placing the source in measurement position included in the delivery kit is intended for performing energy calibration and regular checks of stability of beta spectrometer parameters 1 4 Design and operation 1 4 1 Mode of operation of the beta spectrometer is based on acquisition of apparatus spectrum of pulses from detector which detects radiation emitted from the counting sample exposed under fixed measurement conditions Activity of a radionucl
12. ide in the sample is determined by means of spectrogram processing in the personal computer using the Progress software 1 4 2 The Progress software allows controlling the beta spectrometer operation analyzing spectrogram and identifying radionuclides determining activity of corresponding nuclides in the sample computing measurement error and registering measurement results FVKM 412131 002 02RE 4 1 5 Marking and sealing 1 5 1 A label should be attached to the products which constitute beta spectrometer with the following designations trade mark and name of the manufacturer designation ofthe product product type unit works number of product or unit according to the manufacturer s system of numeration year of production degree of ingress protection provided by casings made in Russia 1 5 2 Place and method of marking of the products which constitute beta spectrometer should comply with the design documentation 1 5 3 All products which constitute beta spectrometer should be sealed in accordance with design documentation 1 6 Packing 1 6 1 Packing meets the design documentation and provides protection against penetration of atmospheric precipitation and aerosols splashing water dust sand solar ultraviolet radiation and limits the penetration of water vapor and gases 2 INTENDED USE 2 1 Operating limitations 2 1 1 Beta spectrometer shall be stationary installed in the laboratory room where normal
13. inus 10 to 40 C and subsequent storage under normal conditions during 24 hours e nonu up to 98 at 35 C sinusoidal vibrations ososeeeerci oreet edita da in the frequency range from 10 to 55 Hz with displacement amplitude 0 35 mm 8 DISPOSAL 8 1 After a full life of beta spectrometer products included in its composition before sending it for repairs or for calibration should conduct a survey products for the presence of radioactive contamination of surfaces Criteria for deciding on the cleanup and future use are taken in accordance with the mandatory requirements of national standards 8 2 Decontamination should be carried out with solutions in accordance with 1 2 20 in cases where the level of radioactive contamination of surfaces products including accessible for repair can be reduced to acceptable values FVKM 412131 002 02RE 10 8 3 The product for which the permission was granted for further use after decontamination is subject for repair or replacement in case of malfunction Beta spectrometer unsuitable for operation with radioactive contamination levels below permissible values should be dismantled to prevent its further use and transferred to a special site for disposal of industrial wastes 8 4 Product with expired intended life time which is considered suitable for further use after decontamination shall be inspected for proper technical state In case of satisfactory technical state extended operation term of
14. n of the beta spectrometer is considered acceptable in case the deviation of the measured count rate from the reference count rate indicated in the certificate of calibration does not exceed 10 96 FVKM 412131 002 02RE 6 4 CALIBRATION ROUTINE 4 1 General requirements 4 1 1 Calibration of the beta spectrometer is performed in accordance with IEC 61453 2007 4 2 Preliminary arrangements 4 2 Operations that should be performed during calibration and required equipment are listed in the Table 4 1 Table 4 1 List of calibration operations Operation Section Calibration equipment and its characteristics External examination 4 5 1 Visual 2 Testing 4 5 2 3 Determination of 4 5 3 Calibration source OSr Y included in the delivery kit of metrological characteristics the beta spectrometer Working standard 1 SO with activity 100 500 Bq PC Software Progress 4 Processing of results 4 5 4 Note It is acceptable to use other calibration equipment with characteristics that ensures determination of metrological parameters with required precision 4 3 Safety requirements 4 3 1 It is necessary to follow safety requirements described in section 3 2 and in documentation accompanying calibration tools and equipment 4 4 Conditions 4 4 The following normal operating conditions shall be met during calibration EE doc CECIDIT TTE TD IUUELELTTLETDTCMT 20 5 C relative air humidity
15. nciple of operation characteristics of the scintillation beta spectrometer Progress and instructions essential for correct and safe use of this product intended use maintenance servicing storage and transportation as well as information regarding the utilization of the product 1 DESCRIPTION AND OPERATION OF THE PRODUCT 1 1 Product functionality Scintillation beta spectrometer Progress beta FVKM 412131 002 02 hereinafter beta spectrometer represents a part of a spectrometric complex Progress for measurement of activity of alpha beta and gamma emitting radionuclides and can be used independently Beta spectrometer is intended for measurement of activity of beta emitting radionuclides in samples of the foodstuffs samples of biological origin water and soil samples and other environmental samples in the laboratory Beta spectrometer can be used for radiation monitoring at nuclear plants at the enterprises and facilities which produce or use sources of ionizing radiation 1 2 Technical characteristics 1 2 EE 200 to 3000 keV 1 2 2 Measurement range of specific activity of Sr in a sample weighted 10 g placed in a standard planchet with presence of K in a sample with activity up to 5 Bq g from 100 to 10 Bq kg without presence of K in a sample ssssssssrrsssrrrrssrrrrssrrrrrsre ra from 50 to 10 Ba kg 1 2 3 Highest possible count rate in the measurement channel u
16. ns trained in operation of radiometric and spectrometric instrumentation having permission for work with high voltage supplies having permission for work with sources of ionizing radiation 3 2 3 When performing work one should pay special attention to condition of the power cord and power switch dangerous voltages can be present in that locations 3 3 Maintenance routine Maintenance is divided into routine and periodic 3 3 1 Routine maintenance Routine maintenance is carried out during regular operation of the beta spectrometer it includes daily dust removal from outer surfaces and measuring beta radiation background Upon increasing of the beta radiation background perform decontamination of outer surfaces of the detector unit and inner surfaces of the lead shielding It is also necessary to wipe measuring planchets using a pad moisten with ethyl alcohol after each measurement Ethyl alcohol consumption is 50 g per 10 samples It is necessary to inspect connecting cables switches and connectors wipe leads of plugs and sockets with ethyl alcohol total ethyl alcohol consumption 150 g 3 3 2 Periodic maintenance Periodic maintenance includes check of the technical condition of the beta spectrometer it shall be performed on a regular basis at least once a month The check is performed by means of measurement of a count rate from calibration sources using specified mode of operation of the spectrometer Technical conditio
17. sse 5 10 cps 1 2 4 Limits of the permissible basic relative error of measured activity 10 95 1 2 5 Relative energy resolution uide cb nad eoe e Oe at te RR NET EO IE ard Ra at peak of conversion electrons with energy 624 keV 1 2 6 Integral non linearity of the conversion characteristic sse 5 99 ELAN Ar MUD DEBE 2e obe ceret tipa ex et rebate ER TERRE Ese s Eb e dm PEE 1 2 8 Continuous operation EE 8 hours 1 2 9 Instability during 8 hours of continuous operation cesse 2 96 1 2 10 Power supply of the PC from mains power supply 220 25 V 50 Hz 1 2 11 Power consumption including PC 400 VA 1 2 12 Power consumption of the measurement channel 0 5 VA 1 2 13 Microclimatic factors of the environment during operation of the beta spectrometer operating temperature range eon cx er NEE ge from 10 to 40 C Telativ humidity dT PP up to 98 at 35 C atmospheric pressure range oue ce eei ee EEN dee from 84 0 to 106 7 kPa content of the corrosive agents in the ambient air corresponds to the values in table 1 1 Table 1 1 Type of atmosphere Designation Description Content of the corrosive agents Sulfur dioxide gas no more than 20 mg m day I Relatively clean not more than 0 025 mg m Chlorides not more than 0 3 mg m day Sulfur dioxide gas no more than 20 to 250 mg m day II Industrial not more than 0 0
Download Pdf Manuals
Related Search
Related Contents
Deutsch - Kristall Form Spiegel GmbH Sitecom USB 2.0 to VGA adapter PDF Screen - About This Particular Macintosh Manual - The BBQ Store User's Manual Automatic Ice Maker User`s Manual OEM6® Family Firmware Reference Manual 取扱説明書[NR-F468T/F478TM] (2.12 MB/PDF) LITEMAX SLD0832 LCD Monitor Copyright © All rights reserved.