1. Leica Geosystems Viva TPS October 2010 TS11 15 Total Stations Check and Adjust Procedure Summary Leica builds total stations to the highest quality and calibrates each instrument before it leaves the Factory After the instrument is shipped or used in the field under changing temperature conditions the instrument will physically change In order to provide the customer with the ability to field calibrate every total station Leica offers the Check and Adjust routine on the TS11 15 instruments This guide will explain why the Check and Adjust is needed and how to perform the procedure This procedure will allow the user to assure that the instrument is providing measurements to standards eica Geosystems 2010 Leica Geosystems Inc All rights reserved when it has to be right TS11 15 Check and Adjust Why is procedure needed Why is the procedure needed When should you perform procedure What adjustments are determined by procedure TS11 15 Total Stations FW Ver 3 00 Check and Adjust Procedure Leica instruments are manufactured assembled and adjusted to the best possible quality Quick temperature changes shock or stress can cause deviations and influence the instrument accuracy The physical relationship of the axis of the instrument are physically different if the operating temperature of the instrument is 32 F 0 C versus 95 F 35 C Materials contract and expand Survey instr
2. Adjust You will then be reminded to make sure the instrument is acclimatized to the surroundings and shaded from direct sun Information o Before adjusting the instrument ensure the instrument is acdimatised to the surrounding temperature amp shaded from direct sunlight TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 5 22 Continued on Next Page TS11 15 Check and Adjust Combined Test Adjustment Performing The combined adjustment procedure determines the following instrument errors in one process pie Lompinee Lt C tor longitudinal and t ind 7 Adjustment ompensator longitudinal and transversal index errors P e Vertical index error related to the standing axis rocedure e c Horizontal collimation error also called line of sight error e ATR Hz ATR zero point error for horizontal angle option e ATRV ATR zero point error for vertical angle option Select the top option for the combined procedure Check amp adjust the compensator index error line of sight error amp automatic target aiming O Check amp adjust the compensator View the current values Configure Check amp Adjust The Wizard will now remind the user what should be done at the first measurement The user needs to manually aim at the center of the Leica Round Prism being sighted The target prism should be placed approximately 330 feet away If you instrument has ATR please check the box for th
3. for calibration amp store the results Combined Check amp Adjust Wizard Adj ustment It is recommended to repeat the last calibration Procedure routine at least three times O Add another calibration loop Finish the calibration amp store the results Hz 18070003 W 270 00 11 You will now view the results of the Check amp Adjust procedure you just performed It may be helpful to view the old correction values and the new correction values you have just determined in the same view To do this select the F5 More key until the middle column is displaying the values from the previous procedure under the Old column You should notice some changes between the values The amount change between the Old and New values depend on the change in the condition For example if you only performed this procedure every 6 months and the difference in temperature between these times is say 70 F there might be a large difference If you follow the recommended procedures carefully for the instrument setup manually aim at the target prism carefully each time and observe that the standard deviation in your points are within the angular resolutions of the model of instrument you are using you can accept the values from your procedure by selecting the F1 Finish button 0 00 02 0 00 02 0 00 00 i V index 0 00 01 0 00 00 c Hz col 0 00 01 0 00 01 ATR Hz 0 00 06 0 00 00 ATR V 0 00 07 0 00 00 1000 Ww 270 O0 11 Finish
4. measurements Procedure Check amp Adjust Wizard It is recommended to repeat the last calibration w routine at least three times Add another calibration loop O Finish the calibration amp store the results Hz 180 0014 wW 270 00 S4 The process of measuring continues Please note that the instrument starts the second measurement set in the FACE 2 position This is how the application was designed in order to provide efficiency in operation Aim accurately at a target positioned horizontally at a distance greater than 330ft 180 00 14 270 00 54 Hz 180 00 14 V 270 00 54 Continue with pointing to the target prism in FACE 2 and FACE 1 as prompted by the application Please remember to manually point at the center of the target prism Continue with the measurements until you complete a FACE 2 and FACE 1 measurement Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 8 22 Performing the Combined Adjustment Procedure TS11 15 Total Stations FW Ver 3 00 Check and Adjust Procedure Once you complete the second calibration loop the system can now start to report the standard deviations These values tell the user how closely they are pointing at the same spot on the target as the calibration loops progress Do not expect these values to be reported as 0 00 00 Your measurements will have error and that is why you repeat the steps Check amp Adjust Wizard No o
5. the results Hz 0 00 04 W 60 15 57 Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 13 22 TS11 15 Check and Adjust Performing Tilting Axis Adjustment Tilting Axis Once you have performed the second set you will then be able to see the standard deviation for Adjustment the sightings you have made Again this value is only representing a value that tells the user Procedure how closely they are pointing to the same spot on the target during the observations 2 fa Ra P Check amp Adjust Wizard No of measurements o a T axis 0 00 06 180004 Wi 299 44 04 Continue the process until you have performed a minimum of 3 observation sets Please be sure to manually aim the instrument at the target each time you observe the target Check amp Adjust Wizard No of measurements d a T axis 0 00 03 y 60 15 56 Fn abe 15 54 Once you have observed the last observation set choose the F3 Next key if the standard deviation is acceptable Acceptable values would be within the resolution of the model of the instrument For example if you have a 3 second model of the instrument you should be within 3 to 6 in standard deviation Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 14 22 TS11 15 Check and Adjust Performing Tilting Axis Adjustment Calibration Results for Titling Axi
6. to be setup in a place that is protected from direct sunlight in order to avoid one side of the instrument being heated more than the opposite side of the instrument The best conditions can be found usually early in the morning and with overcast sky Setting the instrument up under a tree for shaded area or on a shaded side of a building are good common locations that can be used If you are also performing the test for ATR adjustments you will need to set up a Leica circular prism approximately 330 ft 100 m away from the instrument You can setup the prism being used as the target for the testing in direct sunlight Setup up the prism within 9 of the horizontal plane of the instrument you need to use a Leica circular prism Do not use a 360 prism for this procedure Note f you are not performing the procedure for ATR you do not need to use a prism for the procedure The combined test procedure the titling axis procedure and compensator only test will be explained TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 4 22 T 11 15 Check and Adjust Performing Combined Test Procedure Accessing After you have prepared the instrument as explained in the Preparation and Field Requirements Check and section you can follow the Check and Adjust Wizard Adjust Wizard From the main menu select 4 User t Go to Work a sebup total station Survey amp stake pbs Next select 5 Check amp
7. Redo More Back Continue out of the Check amp Adjust Wizard and you now ready to perform measurements with you TS instrument which is now automatically applying these new corrections to measurements You now only need to perform the Tilting Axis Adjustment a to have a fully calibrated instrument TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 10 22 TS11 15 Check and Adjust Performing Tilting Axis Adjustment Tilting Axis You will now perform the Tilting Axis a adjustment the tilting axis error a is caused by the Adjustment change between the mechanical tilting axis and the line perpendicular to the vertical axis Procedure Please refer to Leica Viva Series Technical Reference Manual version 3 Pages 622 647 for detailed explanation and diagrams of axis relationships The requirements for acclimating the instrument and keeping the instrument from direct sunlight are still the same as required by the Combined Adjustment Most users perform the Combined Adjustment and then follow with the Tilting Axis procedure To avoid influences from the Hz collimation error c you should determined prior to the tilting axis error which is done in the Combined Adjustment You do not need a prism target to perform the Tilting Axis adjustment To access this procedure select the Check amp adjust the tilting axis option in the Check and Adjust Wizard SR 8 Loy p Check amp Adjust Wizard What do y
8. a building if sighting a steep zenith angle The user just needs to find a target that can be identified and sighted only requires an optical sight repetitively The user needs a target that falls with the angular ranges that is clear enough to identify the same spot when sighting in FACE 1 and FACE 2 during the procedure It is also helpful to remove the instrument carrying handle before trying to sight with the peep sights on the telescope housing At this point in the procedure the user needs to manually point to the center of the target After pointing to the center of the target prism select the F1 Meas key to record the reading The instrument will pause for a moment while internal readings are being recorded You have now completed the first measurement in FACE 1 If it is a motorized model the instrument will automatically turn to FACE 2 to the target If the instrument is not motorized the user will be required to manually point at the target in FACE 2 Once the instrument has turn into FACE 2 you will see the following screen Check amp Adjust Wizard Face II measurement Aim accurately at the same target in this face 180 00 04 299 44 04 Fo abe 15 31 Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 12 22 TS11 15 Check and Adjust Performing Tilting Axis Adjustment Tilting Axis Adjustment Once you are in FACE 2 Procedure Do not accept
9. ccuracy from procedure and Internal Instrument Temperature Int T recorded at time procedure was completed Check amp Adjust Wizard i Component Current Date 13 10 2010 1 0 00 03 13 10 2010 i V index 0 00 01 11 10 2010 c Hz col 0 00 01 11 10 2010 a T axis 0 00 15 12 10 2010 ATR Hz 0 00 06 11 10 2010 ATR V 0 00 09 11 10 2010 Hz 180 0004 W 29904904 Fn abc 16 41 TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 20 22 TS11 15 Check and Adjust Check and Adjust User Options Configure To set the user options for the Check amp Adjust Wizard choose the Configure Check amp Adjust Check and option from the main Wizard menu Adjust Options amp _ iG oe What do you want to do Boi O Check amp adjust the compensator index error line of sight error amp automatic target aiming Check amp adjust the tilting axis Check amp adjust the compensator View the current values Configure Check amp Adjust Hz 0 00 04 W 60 15 56 Fn abc 16 51 Next Set User The user needs to choose the interval of time that he would like to be reminded to perform the Reminder Check and Adjust procedure If one or more adjustment values date values are older than Interval the time specified with this parameter then a reminder message is displayed each time the instrument is turned on Leica recommends that this is done at least once a month and also for th
10. e Calibrate the automatic target aiming option After you have carefully sighted the center of the Leica Round Prism both vertically and horizontally select the F1 Meas key The instrument will pause for a moment as it reads its systems for the horizontal circle vertical circle compensator and the image seen in the ATR components if calibrating the ATR amp T a Ke Check amp Adjust Wizard Face I measurement Aim accurately at a target positioned horizontally at a distance greater than 330ft Hz 0 00 04 89 59 08 Fn abe 16 12 TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 6 22 Continued on Next Page TS11 15 Check and Adjust Combined Test Adjustment Performing You have now completed the first measurement in FACE 1 If it is a motorized model the the Combined _ instrument will automatically turn to FACE 2 to the target If the instrument is not motorized the Adiust t user will be required to manually point at the target in FACE 2 Once the instrument has turn pee into FACE 2 you will see the following screen Procedure 180 00 15 270 00 54 Hz 180 00 15 W 270 00 54 At this point in the procedure the user needs to manually point to the center of the target prism Note Do not accept the pointing to the target prism that the instrument performs automatically if the instrument is motorized Always make a completely manual pointing to the target prism This procedure de
11. e list of situations such as immediately after the instrument has been shipped via a courier service as listed at the beginning of this guide After making your selection choose F1 Finish to continue e a ie Check amp Adjust Wizard 5 Choose when you want to be reminded to check amp adjust the instrument O Never 2 weeks 1 month 3 months 6 months 12 months 0 00 04 W 60 15 56 Fn abc 16 53 Finish TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 21 22 TS11 15 Check and Adjust Performing Compensator Procedure Create a The next option is one that can be used to create a ASCII file containing a report for the Report Sheet information from the Check and Adjust routine If selected the user must give the name of the file to be created in the Report Sheet field and will also need to specify the format file that will be used to generate the report in the Format file to use field The ASCII file created will appear in the Data directory on the internal memory of the TS The format file can be created by using the Format Manager application in the Leica Geo Office software suite The format file must be transferred to the internal memory of the TS in order to be utilized by this routine _ i 8 bite es Create report sheet Report sheet Adjust ttet a Format file to use M Hz 0 00 04 W 60 15 57 Fn abc 17 33 Finish Back If you have questions
12. f measurements 2 Sigma Comp 0 00 01 Sigma t Comp 0 00 00 Sigma i V index 0 00 01 Sigma c Hz col o 00 03 Sigma ATR Hz o 00 01 Sigma ATR V 0 00 01 Hz 0 0008 W 9 59 56 Fn abc 14 49 Select F1 Next and then choose to Add another calibration loop to perform the last of the recommended of 3 loops of the process Check amp Adjust Wizard It is recommended to repeat the last calibration routine at least three times Add another calibration loop Finish the calibration amp store the results Fn abc 14 51 Once you have completed the F1 and FACE 2 of the measurements you will again view the standard deviations of the now all three of the calibration loops If the values are acceptable select the F1 Next key Acceptable values would be within the resolution of the model of the instrument For example if you have a 3 second model of the instrument you should be within 3 to 6 in standard deviation a 3 oe e Check amp Adjust Wizard No of measurements Sigma Comp o 00 01 Sigma t Comp 0 00 00 Sigma i V index o o0 01 Sigma c Hz col 0 00 02 Sigma ATR Hz o 00 01 Sigma ATR V o 00 02 18050008 W 270 00 11 Continued on Next Page OCT 2010 Page 9 22 TS11 15 Check and Adjust Combined Test Adjustment Procedure Calibration This time when prompted to add more calibration loops or finish the process choose Finish the Results
13. he instrument acclimate to the surrounding temperature have a good solid setup that will not be disturbed during the procedure and you need to have the instrument setup in a place where it will not be exposed to direct sunlight during the procedure One of the most important things that the operator needs to do is to remember to manually point at the target when prompted by the routine on the TS The needed preparation and field requirements in detail are as follows 1 First get the instrument acclimated to the current working environment temperature This needs to be started first and then can continue while you are performing other steps in the preparation When you arrive at the place where the instrument will be setup to perform the procedure open up the instrument case to expose the instrument to direct contact with the working environment If you are taking the instrument out of a building or out of a vehicle that has a different temperature other than the one of the working environment you will need to wait until the instrument has become acclimated to the ambient air temperature of the work site The rule of thumb or this is to wait 2 minutes per every 2 F or per 1 C of difference in the temperature of the place where the instrument originated and the ambient air temperature of the work site Setup the instrument in a very stable place that will not be effect by vibrations or other disturbances during the procedure The instrument needs
14. hree times Add another calibration loop Finish the calibration amp store the results p a a Hz 180 00 04 W 299 44 04 Fn abc 13 59 Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 17 22 TS11 15 Check and Adjust Performing Compensator Procedure Compensator You will again enter the measurement mode of the procedure Repeat the measurements by land t selecting F1 Meas Procedure amp Wi 4G ae Check amp Adjust Wizard i ist tilt measurement in any face Tilt L 070010 Tilt T 0 00 02 After you complete the second measurement loop you will get the screen with the number of measurement 2 and will now need to repeat one more calibration loop to gain the minimum of three sets of measurements Check amp Adjust Wizard No of measurements 2 Sigma Comp 0 00 00 Sigma t Comp 0 00 00 Hz 0 00 04 W 60 15 56 You will finally get to the third calibration loop and will have 3 sets of measurements Select F1 Next to continue Check amp Adjust Wizard No of measurements Sigma Comp o 00 00 Sigma t Comp 0 00 00 Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 18 22 TS11 15 Check and Adjust Performing Compensator Procedure Compensator Select the option Finish the calibration amp store the results to co
15. l angle option if instrument is equipment with ATR e ATR V ATR zero point error for vertical angle option if instrument is equipment with ATR e a Tilting axis error This check must be done separate from all other procedures The results from check and adjust are displayed as errors but used with the opposite sign as corrections when applied to measurements There is a combined procedure that will determine the values for t l c Hz and ATR zero point You can also perform a test for the compensator errors alone if needed The tilting axis a check must be done as a completely separate test Please refer to Leica Viva Series Technical Reference Manual version 3 Pages 622 647 for detailed explanation and diagrams of axis relationships OCT 2010 Page 2 22 Continued on Next Page TS11 15 Check and Adjust Why is procedure needed Summary of Instrument error Effects Effects Elimination Automatically Errors to be V with two face corrected with adjusted measure proper adjust electronically ment ment Hz y vw y v F amnas p o N Gompensatoringeror e T Ww Vertical index error vw ATR Collimation error a ae TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 3 22 TS11 15 Check and Adjust Preparation and Field Requirements Prepare the instrument and target The procedure for performing the Check and Adjust procedure is very simple You need to let t
16. ntinue to the results of your landt procedure Procedure It is recommended to repeat the last calibration lt routine at least three times O Add another calibration loop amp Finish the calibration amp store the results Hz 180 00 04 The results of your procedure are now displayed in the New column If you wish to discard all of the measurements from the procedure and start over you can do so by selecting F2 Redo If you are satisfied with the results then select F1 Finish Hz 180000 W 299 44 04 Fn abc 16 20 Finish Redo Back TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 19 22 TS11 15 Check and Adjust Viewing Current Values from Calibrations Current If you wish to review the values being used in the TS from the Check and Adjust procedures Calibrations these are always available for reference In the Check amp Adjust Wizard select the View the current values option being used by Instrument Check amp Adjust Wizard What do you want to do Check amp adjust the compensator index error line of sight error amp automatic target aiming Check amp adjust the tilting axis Hz 18070004 V 299 44 04 Fn abe 16 36 The current values being applied to measurements by the TS are displayed in the Current column The user can toggle the right column to display the Date of Calibration Date Standard Deviation of Value A
17. or option in the Check and Adjust Wizard What do you want to do Check amp adjust the compensator index error line of sight error amp automatic target aiming Fn abc 13 47 Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 16 22 TS11 15 Check and Adjust Performing Compensator Procedure Compensator The application will display the current and t values being read from the system Select F1 landt Meas to continue to the Face 2 measurement Again motorized instrument will automatically Procedure turn 180 but non motorized instruments must be manually turn 180 to perform the second reading amp _ m8 Check amp Adjust Wizard ist tilt measurement in any face N Tilt L 070010 Tilt T 070002 Hz 00m W 6041556 The procedure will now report the standard deviation of the measurement s sets performed Since you now only have one set no standard deviation can be reported Please select F1 Next to continue to the next step in the procedure Check amp Adjust Wizard No of measurements 1 Sigma Comp Soe Sigma t Comp Puen 150 00 04 VW 299 44 04 Fn abc 13 57 Choose Add another calibration loop and then F1 Next Leica recommends again that the user makes a minimum of 3 measurement sets for this test a W a Check amp Adjust Wizard It is recommended to repeat the last calibration routine at least t
18. or comments about the information in this guide please contact Leica Technical Support at Toll Free 1 800 327 4773 Email survey support leicaus com You can also find more information on Leica MyWorld at https myworld leica geosystems com TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 22 22
19. ou want to do Check amp adjust the compensator index error line of sight error amp automatic target aiming The Wizard will remind you that the Horizontal Collimation test should be done before performing this Titling Axis test If you have completed and accepted the combined test select the F4 OK function key and continue into the procedure Che ern It is recommended to determine the Hz collimation error run the combined adjustment routine before determining the tilting axis error Continued on Next Page TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 11 22 TS11 15 Check and Adjust Performing Tilting Axis Adjustment Tilting Axis To perform the measurements for the titling axis you do not need a prism as the target The Adjustment instrument will not be measuring any distances or ATR images You only need a target that is P d at a very steep sight You can sight a target that is high above the horizontal plane of the POCELE instrument or very low below the horizontal plane The ranges are High Target must be in the range of a zenith angle 63 27 Low Target must be in the range of a zenith angle 117 153 Check amp Adjust Wizard Face I measurement Aim accurately at a distant high 27 to 63 or low 117 to 153 target 0 00 04 90 00 07 This target could be something like a corner of a tall building or corner of a window frame on
20. pends on the user manually point to the target so that the instrument can calculate the error corrections When the motorized instruments automatically change face the aim of the instrument will be very close to the center of the target Please always manually perform the pointing to the target After you point the instrument to the center of the target prism select the F1 Meas key to record the readings You have now performed the first of a set of FACE 1 and FACE 2 observations The application will now report the standard deviations of the measurements you have taken At this time you only have one measured set and the application cannot report these values until multiple sets are measured The user now needs to continue measuring sets by selecting F1 Next Check amp Adjust Wizard i Sigma Comp ial Sigma t Comp Sigma i V index Sigma c Hz col Sigma ATR Hz Sigma ATR V Hz 18070015 W 270 00 S3 TS11 15 Total Stations FW Ver 3 00 OCT 2010 Check and Adjust Procedure Page 7 22 Continued on Next Page TS11 15 Check and Adjust Combined Test Adjustment Performing The application will then prompt the user to either Add another calibration loop or Finish the the Combined calibration amp store the results The user needs to choose to add another calibration loop Leica Adjust t suggests that the user repeats the process until they have a minimum of three 3 sets Select HSIEH the F1 Next key to continue
21. s Procedure TS11 15 Total Stations FW Ver 3 00 Check and Adjust Procedure values shown Choose the option Finish the calibration amp store the results to accept the results of the procedure and have a chance to review It is recommended to repeat the last calibration routine at least three times Add another calibration loop Finish the calibration amp store the results Hz 0 00 04 W 60 15 57 Fn abc 16 03 Next Back The next page shows the old correction value and the new correction value for comparison The user can choose to accept the results by choosing F1 Finish or observe the entire procedure again and throw out all the current observations by selecting F2 Redo Check amp Adjust Wizard Store tilting axis adjustment results Press Redo if you want to repeat all calibration loops New oJ 0 00 4 g Old 0 00 14 Hz 070004 W 60 1556 Fn abc 16 05 Finish Redo If you follow the recommended procedures carefully for the instrument setup manually aim at the target prism carefully each time and observe that the standard deviation in your sightings are within the angular resolutions of the model of instrument you are using please accept the You should see some change in the old versus the new and the amount of change will depend on the difference in the conditions the instrument was working in when the previous check and adjust was performed versus the curren
22. t conditions Select F1 Finish to accept the final results of the procedure Continued on Next Page OCT 2010 Page 15 22 TS11 15 Check and Adjust Performing Compensator Procedure Compensator land t Procedure The compensator adjustment procedure determines the following instrument errors Compensator longitudinal index error t Compensator transversal index error These are also determined in the Combined Procedure explained earlier If the user desires to determine these values for each setup he can perform the compensator procedure The compensator index errors occur if the vertical axis of the instrument and the plumb line are parallel but the zero points of the compensator and the circular level do not coincide The calibration procedure electronically adjusts the zero point of the compensator The plane of the dual axis compensator of the TPS1200 is defined by a longitudinal component in the direction of the telescope and a transversal component perpendicular to the telescope The requirements for acclimating the instrument and keeping the instrument from direct sunlight are still the same as required by the Combined Adjustment This procedure requires no target Motorized version of the TS instruments will automatically turn the instrument after measuring but non motorized version will need to be manually orientated between measurements To access this procedure please choose the Check amp adjust the compensat
23. the pointing to the target prism that the instrument performs automatically if the instrument is motorized Always make a completely manual pointing to the target This procedure depends on the user manually point to the target so that the instrument can calculate the error corrections When the motorized instruments automatically change face the aim of the instrument will be very close to the center of the target Please always manually perform the pointing to the target Carefully point to the center of the target now that you are in FACE 2 and then select the F1 Meas key to record the readings You have now performed the first of a set of FACE 1 and FACE 2 observations You will now be able to review your standard deviations of your sightings However at this point in the procedure you only have a single set of observations and this cannot be calculated Leica recommends that you perform a minimum of three sets of observations for this test which is the same as the combined test To start the procedure of observing additional observations sets please select the F1 Next key No of measurements oa T axis Hz 0 00 04 W 60 1556 You will be navigated to the next screen and here please choose the option Add another calibration loop to continue with additional observation sets amp _ ff 8 Le Check amp Adjust Wizard It is recommended to repeat the last calibration routine at least three times O Finish the calibration amp store
24. uments are also exposed to shocks and stress during transportation to the project site and on the project site These stresses will also change the relationship of the instrument components Please refer to Leica TS11 TS15 User Manual Pages 74 88 and the Leica Viva Series Technical Reference Manual version 3 Pages 622 647 Leica recommends that the Check and Adjust Routine is performed for the following situations Before the first use after the instrument has been shipped by any courier service Before every high precision survey for example high precision traverse After any rough or long transportations to a job site After a long period of storage of the instrument After a long period of use of the instrument a reminder can be set on the instrument for different periods of time e When the change in temperature of the working environment is more than 36 F 20 C between the working temperature of the last calibration and the current working temperature In other words if the working environment temperature changes by 36 F 20 C since the last calibration you need to perform the Check and Adjust routine The following instrument errors can be determined electronically by the Check and Adjust procedure t longitudinal and transversal index errors of compensator Vertical index error related to the standing axis c Hz Collimation error also called line of sight error ATR Hz ATR zero point error for horizonta
Download Pdf Manuals
Related Search
Related Contents
Photo pleine page Manual da Rede do Ar 2010_1 - Agência Portuguesa do Ambiente TIPO “A” - Normateca JVC RX-730R User's Manual Accessory Parts Manual (All Vehicles)incl Beverage HP Generating Set User's Manual RC17xxHP-RC232 User Manual PRESENTATION DE L` INSTITUTION Project Proposal for an Automatic Tracking System for Effort Saving GHC 10 Copyright © All rights reserved.