Home

PERFORMER III

image

Contents

1. CHAIR LIFT DOWN FOOTSWITCH Performer Programming Positions 0 1 or 2 NOTE To stop the chair at any point push any button on the footswitch or touch pad PROGRAM BUTTON PRE PROGRAMMED PROGRAMMABLE FOR ENTRY EXIT x POSITION BUTTONS CUSPIDOR RETURN MANUAL FACTORY SET POSITION ORA BUTTONS PROGRAMMABLE PROGRAM BUTTON POSITION BUTTONS MANUAL POSITION BUTTONS Chair Touch Pad e Move the chair to the desired operating position e Press the program button A tone will sound e Press the programmable position button for the desired setting 0 1 or 2 within 4 seconds e A tone will sound 3 times confirming the position has been reprogrammed Position 3 is factory set in the cuspidor return mode Pressing the position 3 button moves the chair back up allowing patient access to the cuspidor Pressing the position 3 button a second time returns the patient to the previous operating position Contact an authorized A dec Dealer to have Position 3 reconfigured to a third Pre Position or as a last position recall Performer Chair LED The chair LED indicates the status of the chair ON Normal operation SLOW BLINK The cuspidor or stop plate limit switches have been activated Remove any obstructing object Performer Chair Stop Plate 0 9 CHAIR STOP O PLATE The chair stop plate stops the chair immediately when any part of it is pressed Should a
2. PERFORMER a dec 85 2638 00 Warranty Information Serial Number Model Number Date Purchased Date of Technician s Service Model Description of Service Initials ALPHABETICAL EQUIVALENT TO THE NUMERAL OF THE MONTH MANUFACTURED A January 2601 CRESTVIEW DRIVE B February NEWBERG OREGON 97132 esignated epresentative A dec Dental U March Uber Way Attleborough Fi SERIAL D Apri neon irt mosso NUMBER E May SN J828287 REI F June a G n MODEL H August NUMBER September J October s November YEAR MANUFACTURED L December Serial Number Identification Performer Equipment e Delivery System On the bottom of the control head e Cuspidor Underneath the cuspidor on the post e Assistant s Instrumentation Underneath the assistant s instru mentation module on the post e Light On the light post near the mounting end e Chair On the upper structure under the seat upholstery For service information contact your local authorized A dec dealer Check with local codes and A D A Americans with Disabilities Act requirements for installation of this product Warranty A dec warrants all products in this owner s guide against defects in material or workmanship for one year from time of delivery A dec s sole obligation under the warranty is to provide parts for the repair or at its option to provide the replacement product excluding labor The buyer sh
3. Holder Tension Adjustment 18 Swivel Tension Adjustment 18 Performer Accessory Tray Holder Optional for Performer ChairONLY Tray Holder Height Adjustment 19 Swivel Tension Adjustment 19 Performer Dental Light Dental Light On Off Toggle 20 Dental Light Intensity Switch 20 Light Head Positioning 20 Dental Light Bulb Replacement 21 Light Head Pivot Adjustment 21 Floor Box Air Filter Element Replacement 22 Water Screen Replacement 22 Accessories Bitewing Viewer 23 Bitewing Viewer Bulb Replacement 23 80 Watt Power Supply 23 Fiber Optic Bulb Replacement 24 Dual FIVE Su 24 General Information Adjustments and Specifications 25 Mallare 26 Safety Considerations for Accessory Equipment 26 Identification of Symbols 27 Classification of Equipment EN 60601 1 27 Performer PERFORMER CHAIR Back Positioning CHAIR BACK a DOWN a OAA CHAIR e BACK UP TOUCH PAD FOOTSWITCH CHAIR LIFT UP
4. a week Handpiece Control Adjustments Control Head Height Adjustment Performer Pivot Point Tension Adjustment Us s Se M Ut a EA E 13 Performer A DEC AUTOCLAVABLE SYRINGE Using the Autoclavable Syringe Syringe Flow Adjustments v X BI 4 gt SY ais IN PN 7 NL j m e d fi i 4 d Performer PERFORMER CUSPIDOR Cup Fill Bowl Rinse 15 Performer Cuspidor Bowl Rinse Flow Adjustment Lae MA LOCATED IN THE FLOOR BOX 16 Performer ASSISTANT S INSTRUMENTATION Autoclavable HVE HVE Screen 17 Performer PERFORMER ASSISTANT S ARM Option for Performer Chair ONLY Holder Tension Adjustment HOLDER TENSION ADJUSTMENT SETSCREWS Swivel Tension Adjustment REMOVE SEAT UPHOLSIERY 5 16 HEX KEY SWIVEL TENSION ADJUSTMENT BOLT SG BACKREST IN NEAR FULL DOWN POSITION 18 Performer PERFORMER ACCESSORY TRAY HOLDER Option for Performer Chair ONLY Tray Holder Height Adjustment TRAY HOLDER HEIGHT ADJUSTMENT RING 1 Lift the tray holder to access the height adjustment ring 2 Slide the height adjustment ring to desired position 3 Lower the tray holder onto the arm Swivel Tension Adjustment REMOVE SEAT 5 16 UPHOLSTERY HEX KEY ACCESSORY TRAY HOLDER ARM 19 Performer PERFORMER DENTAL LIGHT 20 Performer Dental Light Bulb Replac
5. all have no other remedy All special incidental and coincidental dam ages are excluded Written notice of breach of warranty must be given to A dec within the warranty period The warranty does not cover damage resulting from improper installation or maintenance accident or misuse The warranty does not cover damage resulting from the use of cleaning disinfecting or sterilization chemicals and processes The warranty also does not cover light bulbs Failure to follow instructions provided in the A dec owners guide operation and maintenance instructions may void the warranty A dec warrants A dec dental chair cylinders both lift and tilt for ten years from the date of purchase of the chair or the cylinder This war ranty is retroactive to A dec chair cylinders already in the field The warranty covers chair cylinders A dec finds to have manufacturing related irregularities Stool cylinders are covered under A dec s one year warranty NO OTHER WARRANTIES AS TO MERCHANTABILITY OR OTHERWISE ARE MADE Performer Return Policy U S and Canadian dealers wishing to return overstock unopened merchandise to A dec for credit consideration must include a copy of the original invoice number A return authorization form from an A dec Territory Manager must be included with serial numbered equipment A dec W amp H handpieces 15 restocking fee will be assessed Merchandise that cannot be returned for credit includes parts assem
6. ar Tension Adjustment 5 Upholstery Replacement 6 Double Articulating Headrest Upholstery 6 Backrest Upholstery Armrest Caps 6 Narrow Backrest Upholstery with Slings 7 Seat Toeboard Upholstery 7 Performer Handpiece Control Master On Off Toggle 8 Foot Control Wet Dry Toggle 8 Drive Air Adjustment 9 Coolant Air Adjustment 10 Coolant Water Adjustment 10 Handpiece Tubing Flush 11 Handpiece Oil Collector Cleaning 12 Handpiece Control Adjustments 12 Control Head Height Adjustment 12 Control Head Positioning 12 Pivot Point Tension Adjustment 13 A dec Autoclavable Syringe Using the Autoclavable Syringe 14 Syringe Flow Adjustments 14 Performer Cuspidor de _ 15 BOW 15 Cuspidor Bowl Rinse Flow Adjustment 16 Assistant s Instrumentation Autoclavable HVE 17 CONTENTS Autoclavable Saliva Ejector 17 FIVE NR T 17 Performer Assistant s Arm Optional for Performer ChairONLY
7. bled to the dental unit chair light or dental furniture obsolete parts and specials Dental furniture cannot be returned for credit Standard color upholstery ordered for obsolete chairs or stools cannot be returned for credit In the case of a defective warranty item a copy of the replacement in voice serial number of the unit under which it was replaced and a description of the symptoms of the defect must be returned with the part to A dec Inc 2601 Crestview Drive Newberg Oregon 97132 USA Performer Equipment Alterations Policy Certain modifications or alterations of A dec equipment which expand the use of A dec equipment beyond its design and intent or which override any safety features of A dec equipment may jeopardize doctor patient or staff safety Field modifications that alter the electrical and or mechanical safety of A dec dental devices are in conflict with Underwriters Laboratory UL construction file requirements and are not sanctioned by A dec Examples of field modifications that diminish safety design include but are not limited to rendering access to the line voltage without the use of tools modification of supporting elements that increase or shift loading characteristics and the addition of any powered device that exceeds the design limits of the dental system The use of accessory equipment not complying with the equivalent safety requirements of A dec equipment may lead to a reduced level of safet
8. e Types of shock JCLASS EQUIPMENT Dental chairs dental lights and power supplies protection CLASS II EQUIPMENT Chair wall and cart mounted delivery systems Degree of shock TYPE B APPLIED PART Delivery systems only protection Degree of ORDINARY EQUIPMENT All products tection against water ingress CONTINUOUS OPERATION All models except dental chairs CONTINUOUS OPERATION WITH INTERMITTENT LOADING Dental chairs 5 duty cycle Flammable Not suitable for use in the presence of a flammable anesthetic mixture with air oxygen or nitrous Gasses oxide where such gasses may accumulate in concentration closed space Performer Electrical Rating Type Specification Volts 100 110 120 220 240 VAC Frequency 50 60 Hz As configured and specified in equipment manual products labeled 15 or greater require dedicated circuit identified in distribution panel Environmental Specifications Temperature Humidity Specification Storage Transportation 40 to 70 C 40 F to 158 F Relative humidity 80 for up to 31 C decreasing Temperature linearly to 50 at 40 C Operating Temperature 10 C to 40 C 40 F to 104 F Relative humidity 80 for up to 31 C decreasing linearly to 50 at 40 C Indoor Use Altitude up to 2 000M 6 563 ft installation category pollution degree 2 UL 61010A 1 and CAN CSA C22 2 No 1010 1 92 only 28 Performer 29 NU 0 a dec USA and Ca
9. ement 21 Performer FLOOR BOX Air Filter Element Replacement MANUAL AIR SHUTOFF VALVE turn completely OFF when replacing the air filter element MANUAL WATER SHUT OFF VALVE turn completely OFF when replacing the water filter screen COMPRESSION REGULATOR WATER SCREEN black body water shutoff valve only 22 Performer ACCESSORIES Bitewing Viewer No cleaning or maintenance required If the floor box has an indicator light it will illuminate when the power supply is plugged in 23 Performer Fiber Optic Bulb Replacement Dual HVE 24 Performer GENERAL INFORMATION Adjustments and Specifications Performer III Chair page 1 Headrest Positioning page 5 Headrest Glide Bar Tension Adjustment page 5 Drive Air Adjustment page 9 Coolant Air Adjustment page 10 Coolant Water Adjustment page 10 Control Head Height Adjustment page 12 Control Head Positioning page 12 Pivot Point Tension Adjustment page 13 Using the Autoclavable Syringe page 14 Syringe Flow Adjustment page 14 Cuspidor Bowl Rinse Flow Adjustment page 16 Dental Light Intensity Switch page 20 Light Head Positioning page 20 L
10. ight Head Pivot Adjustment page 21 Minimum air water and vacuum service require ments for proper unit operation Air 2 50 cfm 70 80 L min at 80 psi 551 Water 1 50 5 68 L min at 40 psi 276 Vacuum 12 cfm 339 84 L min at 8 inches 27 kPa of mercury Maximum weight specifications Maximum control head weight is 55 lbs 30 kg Maximum cuspidor and dental light weight is 27 lbs 12 25 kg Maximum Tray Holder optional weight is 12 lbs 5 44 kg Weight of optional scaler 2 lbs 91 kg 25 Performer Maintenance Chair Capacity Patient Load 300 lbs 135 kg maximum Accessory Load 130 lbs 59 kg maximum Handpiece Tubing Flush page 11 Handpiece Oil Collector Cleaning page 12 HVE SCS 17 Dental Light Bulb Replacement page 21 Air Filter Element Replacement page 22 Bitewing Viewer Bulb Replacement page 23 80 Watt Power Supply page 23 Fiber Optic Bulb Replacement page 24 Also refer to the following Owners Guides for more maintenance information Safety Considerations for Accessory Equipment 26 Autoclavable Syringe 85 0680 00 Equipment Asepsis 85 0696 00 Maintenance Parts 85 2634 00 Self contained Water System 85 0675 00 The use of acce
11. nada 2601 Crestview Drive Newberg Oregon 97132 USA Phone 1 800 547 1883 1 503 538 7478 Fax 1 503 538 0276 www a dec com International Phone 1 503 538 9471 Fax 1 503 538 5911 Distribution Centers A dec Australia 41 43 Bowden Street Alexandria NSW 2015 Australia Phone 61 0 2 9699 4600 Fax 61 0 2 9699 4700 www adec com au EU Authorized Representative A dec United Kingdom Austin House 11 Liberty Way Nuneaton Warwickshire England CV11 6RZ Phone 0800 ADEC UK 2332 85 Within UK 44 24 7635 0901 Outside UK Fax 44 24 7634 5106 C 0843 Copyright 2006 A dec Inc 85 2638 00 Rev F All Rights Reserved
12. nything inadvertently become lodged under the chair press Base UP on the footswitch or touch pad to raise the chair so the object can be removed As long as pressure is applied to the stop plate the chair base will not go down any further Performer Headrest Positioning DECREASE gt INCREASE Performer Upholstery Replacement Double Articulating Headrest Upholstery Backrest Upholstery Armrest Caps Performer Narrow Backrest Upholstery with Slings MOUNTING Seat Toeboard Upholstery Performer PERFORMER HANDPIECE CONTROL Master On Off Toggle Performer Drive Air Adjustment NOTE The drive air control must be adjusted to meet the handpiece manufacturer s drive air pressure specification Refer to the documentation that came with your handpiece for drive air pressure The handpiece tubing flush system flushes more water through the tubings in less time than is Performer Coolant Air Adjustment 10 Performer Handpiece Tubing Flush normally possible when operating the foot con trol only The handpieces should not be connected when flushing the tubings After each patient How Often Should the Handpiece Libings be Flushed Flush the tubings for about 20 30 seconds At the beginning of each day Flush the tubings for 2 3 minutes 11 Performer Handpiece Oil Collector Cleaning Change the 2 50mm square gauze pad once
13. ssory equipment not complying with the equivalent safety requirements of this equip ment may lead to a reduced level of safety of the resulting system Consideration relating to the use of accessory equipment shall include Evidence that Safety Certification of the accessory equipment has been performed in accordance to the appropriate IEC 60601 and IEC 60601 1 Harmonized National Standards Performer Identification of Symbols Al Recognized by Underwriters Laboratories Inc with respect to electric shock fire and mechanical hazards BB only in accordance with UL 60601 1 2601 1 and under mutual recognition agreement with CAN CSA C22 2 No 601 1 A C22 2 No 601 1 SSIF oz lt Classified Underwriters Laboratories Inc with respect to electric shock fire and mechanical hazards C us only in accordance with UL 60601 1 2601 1 and under mutual recognition agreement with CAN CSA C22 2 UL 2601 1 No 601 1 C22 2 No 601 1 UL listed to UL 61010A 1 BS EN 61010 2 010 and Canadian CAN CSA C22 2 No 1010 1 92 safety standards Conforms to applicable European Directives refer to Declaration of Conformity Protective earth ground Functional earth ground Attention consult accompanying documents No user servicable parts Attention line voltage Only licensed electrician should remove cover Type B applied part Class Il equipment Caution Metal surfaces can be hot during and following the dry cycl
14. y of the resulting system It is the responsibility of the equipment distributor and the installer to assure that the installation complies with all building code requirements The responsibility to determine whether a modification or alteration of A dec equipment falls within these constraints is with the person s who initiates approves and or performs such modification or alteration A dec will not respond to inquiries on an individual basis This person s will be deemed to have assumed all associated risks with such alteration or modification and will hold A dec harmless from resulting claims including product liability claims Additionally such modification or alteration voids A dec s warranty and may invalidate UL or other regulatory agency approval ardec All product names used in this document are trademarks or registered trademarks of their respective holders Printed in USA e Copyright 2006 All Rights Reserved Performer Serial number location service information and warranty information are located on the inside front cover and front page Performer Chair CONTENTS Performer Ill Chair Back Positioning 1 1 Programming Positions 0 1 or 2 2 Cham 3 Chair Stop Plate 4 Headrest Positioning 5 Headrest Glide B