Home

1730R User`s Manual

image

Contents

1. Speed setting unit 1 rcf or 10 rcf 1 Touch the RPM RCF button once gt RPM MODE is generated with touching a RPM RCF button Once gt RPM LED is flickering at the display window 2 Touch the A V buttons to change input value gt After keeping holding finger on the AW buttons for 5 seconds the unit of setting value is changed to 10 rcf from 1 rcf gt If you do not touch the A VW button for 5 second the setting mode is cleared 3 Touch the Enter button to complete the setting 13 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 4 3 Setting Time Speed setting unit Imin or 10 min 1 sec or 10 sec Action v Time is down counted after starting centrifugation 1 Touch the TIME button once gt MIN on LED is flickering 2 Touch the A VW buttons to change the minute value gt After keeping holding finger on the AW buttons for 5 seconds the unit of setting value is changed to 10 min from 1min gt If you do not touch the A V button for 5 second the setting mode is cleared 3 Touch the Enter button to pass the SEC value setting 4 Touch the A VW buttons to change the second value gt After keeping holding finger on the AW buttons for 5 seconds the unit of setting value is changed to 10 sec from 1sec gt Ifyou do not touch the AW button for 5 second the setting mode is cleared 5 T
2. The Aerosol tight buckets and rotors prevent contamination and ensures safety Autoclavable and corrosion resistant rotors ensure safety and long life High quality cabinets with scratch resistant powder coated finished 2 LC yaY RELIZCZEN Refrigerated high speed micro centrifuge 1730R CONVENIENCE IN OPERATION Intuitive touch button with easy to follow controls Easy to read LCD display with bright white lettering on a blue LCD background Time control of pulse timed and continuous Automatic RPM RCF conversion for prompt detection of g force C 2 eo o ee lt amp Easy to check actual rpm through the top window of door 2 25 Program memory for up to 100 programs HOGG 956 Automatic rotor identification function m Automatic door opening function when rotor is static for easy retrieval of samples A large assortment of rotors fixed angle or swing out with buckets and adaptors ECO FRIENDLY MANUFACTURING M Dust free AC induction motor Mi Eco safe refrigerant R404a E Very quiet operation at lesser than 56 EASY CUSTOMIZATION Any rotors sample containers and adaptors can be manufactured according to customer s Specification a Flexibility of including any additional functions or programs in need 3 Liy mRLUIZE EN Refrigerated high speed micro centrifuge 1730R 1 1 1 2 2 1 2 2 3 1 3 2 3 3 4 1 4 2 4 3 4 4 4 5 4 6 4 7 4 8 4 9
3. 3 Shaft 1 Always make special attention to clean the motor shaft to avoid any imbalance problem due to the contaminants 2 After using the instrument take out the rotor from the shaft and clean the shaft with dry soft cloth Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 18 to keep dry 5 4 Rotor 1 If any parts are contaminated with samples clean the rotor with soft wet cloth and keep the rotor dry 2 Be careful not to make scratches inside or on the surface of rotors Any small scratches can cause corrosion of the rotor and big damage to the instrument 3 If you do not use the instrument keep the rotor separately from the motor shaft and stand it upside down 5 5 Transportation of Instrument 1 If you need to move or ship the instrument be cautious to protect the motor shaft from any physical impact or turbulence 2 Do not mount a rotor in any cases of movement Fill inside the chamber with proper materials to keep the motor shaft on place and not to be influenced by physical pressure 6 Trouble Shooting 6 1 Check list Symptom Power failure Can t be started Can t open the door Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R Check list Connect the AC Power cord and make sure that the line is completely connected between the instrument and power outlet Check the power switch is turned on Please refer to 3 1 Power On Off and Door Release If th
4. Daejeon 305 301 Korea declare under our sole responsibility that the product Description of Equipment Centrifuge Model Name 1730R to which this declaration relates is in conformity with the following standards or other normative boratory use Part 1 General Safety requirements for electrical and laboratory use Part 2 Particular requiremen Electrical equipment for measurem EMC requirements Industrial scientific and medical ISM disturbance characteristics Limits an measurement 2004 108 EC Directive the European Parliament and of t Council of 15 December 2004 on the approximation of the laws of the Member States relating to electromagnetic compatibility and repealing Directive 89 336 EEC 2006 95 EC Directive 2006 95 EC of the European Parliament and of the Council of 12 December 2006 on the harmonization of the laws of Member States relating to electrical equipment designed for use within certain voltage limits issued date May 20 2008 Confirmed by PA 24 Li i EL EN Refrigerated high speed micro centrifuge 1730R
5. Product Description Door Chamber gt Pa i od Emergency Door Open Hole D Power le Switch Control Panel Liy mRmRLUIZE EN Refrigerated high speed micro centrifuge 1730R Emergency Open Tool 6mm Wrench User manual AC Power Cord Rotor Locking Tool 2 2 Technical Specifications i This instrument has following functions for safety 1 Automatic rotor identification function 2 Automatic detection and alarms for imbalance excess speed and heating Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 3 Installation 3 1 Power On Off and Door Release 3 1 1 Power On Off A 1 Connect the AC Power cord at the power socket on the right back of the instrument 2 Turn on the instrument by pressing a switch on the right side of the ee instrument 3 1 2 Door Release 1 For opening the door touch the DOOR button gt Should touch the DOOR button When the door is closed Door LED shows off gt Close the door until hearing clank shut gt When the door is opened the door LED turns on v The door is not opened while the instrument is running v If the door is opened the instrument could not be operated even with pressing the Start button v For operational safety this instrument has the automatic rotor recognition function v When you supply the power Searching Rotor Change to Rotor ID will be appeared If the rotor is
6. 4 10 5 1 5 2 5 3 5 4 Liy mRmRLUIZE EN Refrigerated high speed micro centrifuge 1730R CONTENTS Page Meanings of Symbols Ge Safety PrecautioriS EEE EEE ERR 6 Mean oe PP 6 EE Me GJEN 6 Product Description and Technical Specifications mmmmmmersmvsssssssssssvsssssssssssseessseessesserssersersserseessenns 8 Product Description msmrssesssnsssssssssssssessrseseseeseessrssesseeseeseessrssesserseessessrsserserssesserserserseessene 8 Technical SPECIFICATIONS m mmmnemmssvsvsvsssssssssssesersserseessersersserserssrssersserseessersersserssesserssrsserssessenssesssnne 9 Installation esesenssenseseeneensenseensenneensennennsennsnnsennsnnsnnnsnnsnnnsnnsennsnnsensennsnnssnnsenssnnsenssnnsnnssnnsenssnnsenssnnsenssnnsenssnnsnnssnnsnnssnssenssnnenn 10 Power On OTF and DAG ee an 10 Rotor Coupling and Disassembling mmmrssrsssssessssrsrsssessrssrsrssrsssesessrsssseesessesseseeseessessrsserseesse 10 PT EET hhv 11 1 UI ESAEREN 12 EVU TON O CONTO el ao cesccescesccgssesscussesscussesscussesscossesseusaueseuzsueseussueseussusseussaescussesscuseazscueaeesueaaeccecseeceseesesc 12 MR e To E E T O ee 13 Setting EIA AlE 14 Setting Temperature and Fast COOL ierra eet 14 Acceleration Deceleration s srsursrssrsessrsrsssvssrsesssessisesssessessessessrssesssesssessisessesessessessesseessersse 15 PROC MNT a LA e RE oE EE ET cemansessuasaaeaesseaeeneetis 16 PUISE nnnnrnnernsrrssrssrnssrssrrssrsserssrsssssrsssssrssres
7. Refrigerated High Speed Microcentrifuge 1 30R User s Manual j Gyrozen Co Ltd B Station 30 12 Gyeryong ro 141 gil Yuseoung gu Daejeon 305 301 Korea Tel 82 42 719 8200 Fax 82 42 826 9848 WWwwW gyrozen com DOC No CO2DC00204 11 GYROZEN Desiged to suit your exact needs Wide range of modern centrifuges for a wide variety of laboratory applications THE BEST FUNCTIONAL QUALITY E Manufactured and tested to IEC standards stable spining operation within 2 variation E Steady and soft deceleration with dynamic brake technology M High capacity strong compressor ensures fast cooling of chamber and samples E Fast cool function 4 C in 5 minutes for fast start up and fast cool down of sample M Automatic recognition and alarms for imbalance over speed and over heat E Automatic rotor identification functions Triple or double laminated ABS Steel door construction minimized noise and heat trnasmissions SAFETY and ROBUSTNESS Safety door lock mechanism ensures the door is locked whilst in operation mode Unique door drop protection protects the operator and samples when loading and unloading Automatic rotor identification secures operational safety Automatic detection and alarms for imbalance excess speed and heating Emergency door lock release helps to open the instrument when power blackout or suddun stoppage occurs The compressor off function during door open minimizes frosting and rusting
8. YS position the instrument with enough space on each side of instrument to ensure proper air circulation ALWAYS install the instrument within a temperature and humidity controlled environment Permissible ambient temperature 5 C 35 C Relative humidity lt 85 Before connecting the power check the rated voltage Should not use unapproved rotors and accessories v Only use rotors from Gyrozen Co Ltd with appropriate centrifugal tubes and suitable adaptors to embrace sample containers tightly enough inside rotors Before operating the instrument check if the rotor and the rotor lid are securely fastened v Should operate the instrument with a rotor properly installed and secured to the motor shaft Mount the rotor on the motor shaft properly check it with spinning manually Do not stop the rotor by touching with hand during the instrument is running Emergency door open should be performed only when spinning is completely stopped Should not exceed the rated speed or specific gravity Samples whose density is greater than 1 2g ml must have reduced maximum rotational speed to avoid rotor failure The sample content should not exceed 80 of total capacity of a tube Otherwise it would cause spillage of sample fluid and even the tube breakage ALWAYS load the tubes symmetrically with evenly weighted samples to avoid rotor imbalance If necessary use the water blank to counterbalance the unpaired sample The opera
9. absent the Error 9 will be appeared This message will be cleared after rotor coupling and running v The door is not automatically opened after finishing operation to keep the sample at proper temperature v Power Failure If there is any power failure during operation door is not opened with touching Door Door can be opened only when the operation is completely stopped and the power is on again If you want to open the door at the power failure please refer to 4 9 Emergency Door Open 3 2 Rotor Coupling and Disassembling Action 1 Before coupling a rotor clean the motor shaft and chamber with soft dry towel 3 2 1 Fixed Angle Rotor PCR Rotor 2 Mount a proper rotor into the motor shaft 10 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 3 Grasp the rotor with one hand and place Rotor Locking Tool at the center hole of the rotor gt To assemble the rotor Rotate the Rotor Locking Tool clockwise until tightly assembled gt To disassemble the rotor Rotate the Rotor Locking Tool counterclockwise 4 To close of the rotor lid rotate the rotor lid nut clockwise gt For opening lid rotate the rotor lid nut counterclockwise A When you run a fixed angle rotor make sure that the rotor lid is tightly closed If you don t close the rotor lid completely it will be crushed v For operational safety this instrument has the automatic rotor recognition function v Wh
10. amp Safety Precautions 1 1 Meanings of Symbols 1 1 1 Symbols on the device Symbol Meaning Symbol Meaning Attention and warning Attention and warning for electric shock Attention and warning nl AN r Attention and warning mon Door ma l A Please fix the rotor ag Please be careful for door opening and firmly on place SAE for rotor coupling ioe closing Attention and warning Indicate a hole for for correct way of manual door opening in sample balancing in the case of emergency rotor 1 1 2 Symbols in this document This symbol refers to safety Note relevant warnings and indicates This symbol refers to the important possible dangerous outcomes reminder 1 2 Safety Precautions Before using the instrument please read this operation manual to ensure correct usage through understanding Incorrect handling of the instrument could possibly result in personal injury or physical damage on the instrument or its accessories Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 10 11 12 13 14 I Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R ALWAYS locate the instrument on a flat rigid and stable table capable of withstanding the weight of the instrument and its spinning operation ALWAYS make a safety zone of 30 cm around the centrifuge to indicate that neither hazardous materials nor persons should be permitted within the area during operation v ALWA
11. e door is not closed completely the instrument can t run Check the Door LED on the display window and close the door completely If the power is out check the main fuse for the laboratory to supply the power If it is not solved in shortly open the door with Emergency Door Open Tool manually for safety of sample Please refer to 4 9 Emergency Door Open 19 Remove the dirt at the door latch and then close the door completely Can t close the door again If the door seems not being closed by mechanical reason please contact our service team Please check the balanced status of both the table and the instrument Please re check the coupling status of the following three matches to minimize the noise 1 the balanced way of coupling of the rotor into the motor shaft 2 the completeness of fixing of the Rotor Locking Nut on the Noise and vibration during ar runnin I 3 the matching status of Rotor Lid with the rotor Please refer to 3 2 Rotor Coupling and Disassembling Check balances of samples in the rotor Please refer to 3 3 Positioning of Sample of Tubes and load the same weight of samples symmetrically 6 2 Error Code If the instrument shows the error code with beeping sound press STOP button to stop the beeping sound and press Enter button to release of the error status and make the instrument go to the default setting again Error Possible Causes Actions Shut off the power supply and the
12. en you supply the power Searching Rotor Change to Rotor ID will be appeared If the rotor is absent the Error 9 will be appeared This message will be disappeared after rotor coupling 3 3 Positioning of Sample Tubes 1 Before loading sample tubes check the water drop or dirt in the rotor hole or inner adaptor gt If there is a water drop or dirt in the rotor hole or inner adaptor remove it with soft dry cloth 2 Tubes should be placed in the rotor with same amount of samples at symmetrical positions gt Only use appropriate centrifugal tubes and do not exceed the speed beyond the tube s max g force gt For safety fill the sample for 70 80 in the tubes Correct Way of Sample Balancing amp Tube Usage 9 8 000xg 10 000 xg 7 1 000 xg 70 80 filling for safety Check the g force of sample containers NG 4 j Ke If the number of samples is not in pair please load the control tubes at each symmetrical position Otherwise it results noise and vibration which eventually damages the instrument 11 Liy mRmRLUIZE EN Refrigerated high speed micro centrifuge 1730R Ke For safety the Imbalance Cut Off function will be occurred if there is imbalance of loading tubes Error 8 Imbalance error Please refer to 6 Trouble Shooting 4 Operation 4 1 Key Functions of Control Panel RPM RCF TIME Temp ACC DEC PROG Fast Cool Pulse Ente
13. er supply and then check the voltage of the Power supply V H2 Use AVR to provide proper power If the instrument is soun with over speed there will be some problems in the overload of motor and the output of motor Shut off the power supply and then turn on the power switch again to check the instrument If the installed software has bugs this message is appeared Tuning the firmware Download Check weight balances of samples Please refer to 3 3 Positioning of Sample Tubes and then turn off and on the instrument for checking If the function of rotor recognition is failed this message is appeared This message will be cleared by coupling an appropriate rotor Please refer to 3 2 Rotor Coupling and Disassembling If the error code shows continuously please call Gyrozen Field Service Engineer If the instrument is not reached to setting temperature within an hour this message is appeared No user action Please call Gyrozen Field Service Engineer If there is a faulty in the temperature sensing of chamber or over heated this message is appeared No user action Please call Gyrozen Field Service Engineer If the motor temperature sensor can t recognize this message Is appeared No user action Please call Gyrozen Field Service Engineer If the temperature of compressor is over heated up this message is appeared No user action Please call Gyrozen Field Service Engineer An
14. h the PROG button shortly less than 1 sec gt Check the message of PROGRAM CALL at the display window 2 Touch the Aw buttons to select program number you want to recall and then touch the Enter button 95 5 gt If you do not touch the AW button for 5 second the PROGRAM CALL 00 setting mode is cleared gt When you touch the Enter button display window shows the saved setting parameters RPM RCF TIME TEMP 16 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 4 7 Pulse It is for quick and short spin down Action 1 If you touch Pulse button and release at the point you want to stop the centrifuge decelerates immediately 4 8 Start Stop Action 4 8 1 Start 1 After setting RPM RCF Time and Temp touch Start Stop button During running a Start LED is turned on The instrument is running only when the door is closed 4 8 1 Stop 1 In case of touching the Start Stop button the operation is stopped 4 9 Emergency Door Open For emergency door open you can use the Emergency Door Open Tool only when the instrument is completely stopped The door can be unlocked manually with Emergency Door Open Tool through the emergency opening hole 1 Find the emergency hole on the left side of the instrument 2 Insert the Emergency Door Open Tool into the hole and push it until the door is released Manual opening
15. n turn on the power switch again to check the instrument Error 1 RPM Sensor If the error code shows continuously although you try to operate again please contact us If the door is not closed completely this message is appeared Remove the dirt at the door latch and then close the door completely Error 2 Door again Check the Door LED on the display window If it is not solved in shortly open the door with emergency door tool manually for safety of sample Please refer to 4 9 Emergency Door Open If the motor is overheated this message Is appeared Shut off the power supply for an hour and then turn on the power Error 3 Motor Overheating switch for checking the instrument If the error code shows continuously please contact us 20 LOY ROUEN Refrigerated high speed micro centrifuge 1730R Error 4 Error5 Error 6 Error Error 8 Error 9 Error 11 Error 12 Error 15 Error 16 Low Voltage High Voltage Over Speed Software Imbalance Rotor ID or RPM Sensor Chamber Temp Error Temp Sensor Error Motor Temp Sensor Comp Temp Sensor If the power input of Power supply V Hz is 10 less than required power this message is appeared Shut off the power supply and then check the voltage of the Power supply V H2 Use AVR to provide proper power If the power input of Power supply V Hz is 10 more than required power this message is appeared Shut off the pow
16. on levels to protect sensitive samples Action 1 Touch ACC DEC button 2 Touch the AW buttons to change input ACC value gt ACC blinks on the ACC DEC display gt Input the desired level of ACC from 1 to 9 Level 9 The fastest acceleration gt If you do not touch the Aw button for 5 second the setting mode is cleared 3 Fix the ACC level by touching Enter button 15 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 4 Touch the AW buttons to change input DEC value gt DEC blinks on the ACC DEC display gt Input the desired level of DEC from 0 to 9 Level 0 Natural deceleration Level 9 The fastest deceleration gt Ifyou do not touch the 4 W button for 5 second the setting mode is cleared 5 Fix the DEC level by touching Enter button 4 6 Program Saving amp Recalling Action 4 6 1 Saving 1 Set parameters Refer to 4 2 4 4 2 Touch the PROG button longer than 3 seconds gt Check the message of PROGRAM SAVE on the display window 3 Touch the AW buttons to change input Program number l gt If you do not touch the Aw button for 5 second the PROGRAM SAVE 00 setting mode is cleared gt Save up to 100 programs Program numbers from 00 to 99 4 Touch the Enter button to complete the saving 4 6 2 Recalling 1 To recall the saved program just touc
17. ouch the Enter button to complete the setting 4 4 Setting Temperature and Fast Cool Action 4 4 1 Setting Temperature gt Temperature can be set from 20 C to 40 C Temp setting unit 1 C 1 Touch the TEMP button gt Temperature value blinks on the display window 2 Touch the AW buttons to change input value gt If you do not touch the Aw button for 5 second Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 14 the setting mode is cleared 3 Touch the Enter button to complete setting 4 4 2 Fast Cool 1 Setting the temperature Please refer to 4 4 1 Setting Temperature 2 Touch the Fast Cool buttons for fast cooling gt Fast Cool on LED is turned on gt The display shows a message as follows Searching Rotor gt gt recognition OK gt By touching the Fast Cool button the instrument is refrigerated down to the set temperature in a short time During the fast cooling the rotor runs at low speed 1 000 rpm gt The passed time is showed on the display window v If you d like to load your sample tubes before pressing the Fast Cool button please check if the sample is safe during spinning at 1 000 rpm v Before starting Fast Cooling please check the rotor coupling and symmetry of sample tubes 4 5 Acceleration Deceleration Use the adjustment function of acceleration amp decelerati
18. r Start Stop Door RPM EER HUL dn I TER HIN LAL RPM AEC TIME fremp For automatic conversion of RPM RCF and to set the speed Use to set time available range up to 99 min 59 sec 00 00 continuous Use to set temperature 20 C 40 C Use to set the acceleration amp deceleration level from 1 to 9 steps 0 in deceleration step means natural deceleration Larger number means faster acceleration or deceleration Use to save a set of setting values or recall the saved program number Use to reach rapid refrigeration up to the setting temperature Use for quick runs Use for completion of data setting Use to start and stop operation Use to open instrument lid LiYRUZEN 12 Refrigerated high speed micro centrifuge 1730R 4 2 Setting RPM RCF Action 4 2 1 Setting RPM gt Speed setting unit 1Orpm or 100rpm 1 Touch the RPM RCF button once gt RPM MODE is generated with touching a RPM RCF button once gt RPM LED is flickering at the display window 2 Touch the A VW buttons to change input value gt After 5 seconds from touching the A VW buttons the unit of setting value is changed to 100 rpm from 10rpm gt If you do not touch the A V button for 5 second the setting mode is cleared 3 Touch the Enter button to complete the setting gt Touch Enter to save the setting value 4 2 1 Setting RCF
19. should be performed only when spinning is completely stopped Otherwise harmful A damage will be accompanied to not only operators but samples After opening the door manually it is recommended to wait until normal electricity comes back 17 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 4 10 Fuse replacement When the power is not turned on please check the connection of power consent power switch Replace the fuse as following instruction if the power is still not turned on Action 1 Separate the AC Power Cord at the back of the instrument and push the flat head screwdriver for bring out the fuse case Connected Fuse in the power 2 Replace the damaged fuse with new one from the fuse case and then connect in the power 5 Maintenance 5 1 Outer Part of Instrument 1 Clean the outside of the instrument with dry soft cloth If necessary dip the cloth in neutral detergent and clean contaminated area Keep completely dry after cleaning 2 Do not use any volatile chemicals such as alcohol and benzene etc 3 Be careful not to make scratches on the surface of the instrument The scratches can cause corrosion on the surface of the instrument v Tf any rust appears clean it with neutral detergents and keep dry 5 2 Chamber 1 Keep dry inside the chamber after every use 2 If the chamber is contaminated dip the cloth in neutral detergent and clean contaminated area 5
20. srssrsssrssenssessenssrssrsssessrsssessrsssrssrsssrssrsssrssrsssessssssssssnssrssnssessrsssessesssrssesssessrsssessesssessesssessesssessensens 17 SLEI D S100 enantio 17 ar 19 EE 0 EE 17 Fuse Replacement mmsrsmnsrssssssssssseserseesenssrnserseeseessrsserserseeseessrssrsserseeseessrsserserserssessersrsserssesee 18 Maintenance mssrssesserserssrssrssessessersssssssssssessersesssssssssessernesssssssssessesnerssssssssssseenernessssssssseeneenesssssssssesnesnernssssssssssenernessssssssseeneenessssssssssenesne 18 Smr D O E essen 18 Chamber mmemmsrsesssesessesssssesessssssesesessesessssessesssesseeseesseeseesseeseessessersseeserssesserssesserssersersserseersessersserssrssersersseessesssene 18 SS EM KE er OK KE EK AE OE eee EE OR Ke eee ee ree 18 Fe 19 4 5 5 Transportation of the instrument ssn ervnvevsrvesrvessvssvssvsssssssssssesssessesesserssesserssesseessrrseesserseessensee 19 6 Trouble Shooting mmmsvrsssssssssssssssssesessesseseseserserssrsserseeseessessrssessesseesserserserserssessessersersersenn 19 6 1 Bae AE 19 6 2 Error Code nsmnssrnnrnnsmmsmmmsmmsmmsmmmmmnmmmnmmmmmveseererrsesrsseresessesenesessesenessssenesessssenen 20 7 Rotors amp Accessories 22 8 Product Range mmemmsmrsssssssssssssesesssrsesesesessseserseessessessrsserseeseessessrsserserseessessrrsersesssrssessrrserserssene 23 9 E EEEE EEEE EEEE 24 5 Liy ROUUIZEN Refrigerated high speed micro centrifuge 1730R 1 Meanings of Symbols
21. tion speed should not exceed the highest value of the individual guaranteed g forces of each centrifuge rotor bucket or adaptors and sample container especially the guaranteed g force of sample container should not be neglected The rotors should be cleaned and kept dry after every use for longer life and safety ALWAYS disconnect the power supply prior to maintenance care and service to avoid electrical shock 16 ALWAYS use proven disinfection procedures after centrifuging biohazardous materials 17 Should not centrifuge flammable toxic radioactive explosive or corrosive materials 18 When it is necessary to use toxic or radioactive materials or pathogenic micro organisms which belong to the Risk Group II of WHO Laboratory Bio safety Manual should follow national regulations v Do not place dangerous materials within 30 cm distance around the instrument and that is also recommended by IEC 61010 2 020 Use the emergency door open function only when the door button on the control panel is dumb under the condition of complete stop of rotor running Never try to open or move the instrument if it is not completely stopped If the power input is more than 10 of the recommended voltage or fluctuates frequently it may cause malfunction of the instrument and often result serious damage Install the instrument at the place without any kinds of corrosive gases 2 Product Description amp Technical Specifications 2 1
22. y wire disconnection or tuning of the instrument must be performed only by a service engineer who is authorized by GYROZEN Co Ltd LY ROZEN 21 Refrigerated high speed micro centrifuge 1730R 7 Rotors amp Accessories Fixed Angle Rotor GRE M m2 0 24 24x 1 5 2 0 m 45 Hole diameter mm 11 1 Max height for tube fit mm 56 Supplied with aerosol tight O ring Fixed Angle Rotor GRF M m2 0 30 30 x 1 5 2 0 ml R45 Hole diameter mm 11 1 Max height for tube fit mm 56 Supplied with aerosol tight O ring N N amp Si Tube i u L Tube capacity m 0 2 0 5 15 20 f 7 un Adaptor A None s B Cat No GAS m0 2 2 GAS m0 5 2 __ Adaptor bore Ox L mm 6 5x 23 8x31 Radius mm 24 30 67 79 73 86 843 94 Max RPM 24 30 17 000 14 000 PCR Rotor GRF M s0 2 6b4 8 x B tube PCR strips 64x 0 2 me PX 45 Hole diameter mm 6 5 Max height for tube fit mm 16 i KENE 4 Ew Tube HU 8 u W Tube capacity m B tube PCR strip 0 2 Radius mm Ist 58 6 2nd 67 1 Max RPM 12 500 Max RCF g force Ist 10 237 2nd 11 722 22 1YROUOISZEN Refrigerated high speed micro centrifuge 1730R 8 Product Range 23 YR A N Refrigerated high speed micro centrifuge 1730R 9 CE Declaration Doc No STC A08 025 KELO8 C02041 EC Declaration of Conformity We Gyrozen Co Ltd B station 544 1 Bongmyeong dong Yuseong gu