Home

MFK1 Kappabridge

image

Contents

1. HILIAS u3 d ISNI y3 mod POR Xcec su E LINN dN ADId IZEZ SH E HOSNIS 1vV1S0AHJ d iviodH Od yS3 LLAJA HOSN3IS 3univuddHWil sani 0S HdIDAH3S3H H31V A 33vNurid EN YOSNAS 3univu3dHW3l ved anjq SNILV3H Y3M 0d NI H31V 100 831v uw u 00L NOSHY 18 Connection Kappabridge to computer m Copy the software SAFYR to your computer exactly in the same structure as it is on original diskette and run program SAFYR EXE After the program is started the communication of the instrument with your computer via serial channel RS 232C COMI or COM2 is tested automatically each time you run the program SAFYR If there is something wrong in the communication the following message appears on the screen RS 232 COMMUNICATION ERROR In this case it is recommended to switch the instrument off and to check the connection of the instrument with the computer or to check whether the serial port COMI or COM2 is enabled in your computer 7 If the communication is O K the initial procedure is started and the messages received from Kappagridge subsequently appear on the screen of the computer These information inform the user of the current activities of the instrument They differ with MFK version and with curr
2. After the above information are input the question appears for the way of inputting the geological orientation data Select Using geological file 1 Manual input from memo book 2 Non oriented specimen 3 If you selects 1 the data are read from the geological data file created earlier the geological data file can be created using the ANISOFT program package which must be located in the same directory as the standard AMS data file The geological data are used in the calculations and also copied into the standard AMS file see Appendix 2 If you select 2 two sampling angles have to be entered from keyboard Input angles of the orientation of the specimen the first is azimuth of the fiducial mark of the specimen the second is the dip or plunge of the fiducial mark separated by comma for details see the AGICO Print No 6 27 Number of tectonic systems 0 to 2 If 0 is input for example if non foliated and non lineated volcanic or plutonic rock is measured no other geological data are input If 1 or 2 is input the following data must also be input 1 Code 4 tectonic angles The two character code characterizes the measured mesoscopic foliation and lineation the angles are azimuth of the dip or strike if the orientation parameter P4 is 90 dip of the first mesoscopic foliation trend plunge of the first mesoscopic lineation respectively If only foliation exists the second character in the code must be
3. Measuring Anisotropy in Pos 2 rotation about axis 2 easuring Anisotropy in Pos 3 rotation about axis 3 Measuring Total directional Susceptibility in Pos 3 n Measuring Total Susceptibility in a Current Position Set Field When measuring anisotropy is completed in all positions lt Eval gt is on The specimen data can be entered from the keyboard or read rom the Geological file he anisotropy tensor and anisotropy factors are computed Fh HA 4 d Enter Actual Volume of a specimen in mode Volume Enter Actual Mass of a specimen in mode Mass This key invokes this HELP Set Frequency instrument options FA and FB The current measuring action is stopped The previous data is preserved Sets the Holder and Rotator to initial position Breaks the current activities the data is cleared Sets the Holder and Rotator to initial position Switches to AUX menu 22 Measuring of AMS static specimen method A The measurement of AMS using static specimen method with manual holder is available with the MFK1 FB or MFK1 B Kappabridges and also with MFK1 FA or MFK1 A with user option adjustment U D DISABLED and ROTATOR DISABLED Do not forget to insert the plastic cylinder into the coil before the measurement The plastic cylinder ensures the placing of the specimen fixed in manual holder in the middle of measuring coil During measurement process
4. This procedure serves for inputting the actual volume or mass of the measured specimen If all the specimens measured in a particular collection have the same volume it is sufficient to input this volume only once If the volume varies from specimen to specimen it is necessary before or after the measurement but before the evaluation of each specimen to input the correct volume or mass of the measured specimen The Mass is possible to input for field variation measurement only not for AMS After starting this procedure the Volume or Mass depending on Volume or Mass mode saved in the configuration file appears on the screen 7 If the volume of the measured specimen is the same one only hits ENTER while if the volume is different one should input new actual volume of the measured specimen Function Key 7 Help This key invokes the help procedure To quit help page press ESC key Function Key 9 Kill This key breaks the current activities and clears the measured and input specimen data Function Key 10 Aux This key switches the program to the AUXILIARY MENU 40 Brief description of the Auxiliary Menu Fl lt BSus gt F2 lt ACmd gt F2 lt CStd gt F2 lt OrPar gt F2 lt AnFac gt F2 Vol F2 Mass F2 Sigma F2 List F3 Cal F4 Holder F5 Field F6 lt ActVol gt F6 Mass F7 Help F8 lt Freq gt F9 Kill F10 Main Convention Measuring the Bulk
5. To help you quickly find the information the name of the Key of the Auxiliary Menu is denoted as Function AuxKey instead of Function Key in Measuring Main Menu to underline that the key of Auxiliary menu is mentioned Examples of measurement values are expressed in talic 41 Function AuxKey 1 Bsus Bulk Susceptibility This procedure serves for measurement of the bulk or mass directional susceptibility in current position of a group of specimens for example in monitoring the susceptibility changes due to the demagnetization steps in palaeomagnetism or for measurement of field variation of the specimen in current position 7 After starting the procedure the following information appear on the screen Measurement of susceptibility in current field or field variation The current holder susceptibility Re 2 57 E 6 Im 1 25 E 9 New measurement of holder Y N If Y is selected the procedure Key 4 Hol is made Than the following selection prompt appears Field variation measurement Standard Reduced None S R CR If one inputs N or CR the procedure continues by bulk measurement in current Field In this case of individual bulk measurement any measurement is started by pressing CR after entering the specimen name If no name is entered the current time is used as the name The bridge is zeroed first you should wait for a beep insert ver FB and B the specimen into the pick up coil wait for the seco
6. and remove the cover of the motor carefully by pulling it up 4 4 check the belt strain to set it later in the same level and remember the belt mark position loosen a little bit do not remove them two screws after that the strain of the tooth belt eases and the specimen shell can be removed from its bearing and remove the belt Clean the black wheel with 64 notches optocouples and photosensor from dust using soft dry brush Check if all 64 notches are transparent 4 2 release gently the ring pay attention to the position and trapezoid shape of rubber washer ERE 4 3 and remove the perspex ring 4 5 Clean the shell the belt and the bearing use pure water with a small amount of non powdered detergent and soft brush Check if the belt mark is clear white Before assembling dry up everything properly Do not use alcohol for cleaning 61 Rotator assembling MFK versions FA and A During assembling pay attention to the following three points the strips on the upper wheel are in the same line the white belt mark is approximately in the centre of the photosensor the strips on specimen shell are in the same line If all three conditions are met adjust the proper belt strain by picking up slightly the part with motor and fixing the two screws Connect the rotator switch MFK on run SAFYR program select AUX menu and check the number following the message ROTATOR is Enabled This val
7. Down and or Rotator is disabled versions FA and A the enabled function is performed 44 AuxKey 2 CStd Calibration Standard This key is used to check or input the susceptibility ies of the calibration standard Please note that the instrument is calibrated by two values versions FA and A This is because the calibration standard gives not only the directional bulk susceptibility value but also the anisotropy which is derived from the susceptibility along the x3 vertical axis max bulk of the calibration standard and from that along the direction perpendicular to the x3 axis min bulk The value Bulk min is not used for Kappabridge versions FB and B Calibration Standard Bulk max Bulk min 136 7 E 03 27 00 E 03 The Bulk max and Bulk min values are the same as those written on the calibration standard to be used for the instrument calibration If you change the calibration standard nominal value s the Holder correction value s is reset to zero AuxKey 2 Orpar Orientation Parameters The scientists use different ways of sampling oriented specimens In order to respect these differences we have developed such a software solution of the data transformation from the specimen coordinate system to the geographic palaeogeographic and tectonic coordinate systems that it is controlled through the so called orientation parameters In this way any oriented sampling is possible For definition of these orientation para meters and more
8. directional or Mass Susceptibility on a group of specimens in current field Measuring field variation of Bulk or Mass Susceptibility of the specimen in current position Auxiliary Command Up Down set the Holder position out of coil into the coil Enable Disable Up and Down options FA and A Enable Disable Rotator options FA and A Zeroing set the bridge nucleus to zero condition Init set the Holder Up and set the Rotator to index pulse Entering the Nominal Directional Susceptibilities of the Calibration Standard Defining the specimen sampling Orientation Parameters i e the meaning of two sampling angles Defining the set of Anisotropy Factors Set the Volume mode for bulk susceptibility measurement Set the Mass mode for mass susceptibility measurement Execute the test measurement with standard error evaluation List the configuration files Calibration i e setting the proper instrument gain Measuring of the Holder real and imaginary susceptibility These values are subtracted from measured susceptibilities Set Field nter Actual Volume of a specimen depends on mode Vol Mass nter Actual Mass of a specimen E EH Invokes this HELP Set Frequency options FA and FB Breaks the current activities The data is lost Sets the Holder and Rotator to init position Return to MAIN menu
9. last specimen measured File Bulk TMP contains the screen contents in measuring bulk susceptibility using routine Bulk invoked by function AuxKey 1 This file can be saved as xxx MFK for future post processing File Safyr HLP contains the Help page File MFK VAR contains predefined fields for field variation measurement File Safyr MON contains information useful for service diagnostics this file is appended each program run File Safyr TMP contains current day information useful for service diagnostics this file is overwritten in each program run 55 Structure of Standard AMS File The STANDARD ANISOTROPY FILE is a random access file with the length of the record being 64 bytes Structure of the First Line Record N 2 LOCALITY LONGI LATI ROCK STRATI LITHO REGIO ORIENT P EOL The first record contains the locality data N 2 LOCALITY LONGI LATI ROCK STRATI LITHO REGIO ORIENT EOL number of specimens in the file 2 name of locality geographical longitude of the locality geographical latitude of the locality rock type stratigraphical position lithostratigraphy regional position orientation parameters P1 P2 P3 P4 end of line sequence Structure of the Other Lines Records SPEC CHAR NORM KIH to K33 CI FOLII LINEI C2 FOLI2 LINE2 The second and the following records contain the specimen data Each record contains SPEC name of the specimen CHAR mean susceptibility in the order
10. missing or damaged contact the manufacturer or your dealer immediately You may want to retain the box and other packing material in case you need to ship the instrument Storage and Transportation The properly wrapped instrument can be stored and transported at a temperature 20 C to 55 C and relative humidity up to 80 In both cases the instrument should be stored in suitable place free of dust and chemical evaporation 16 Installation Procedures The first installation and training is performed exclusively by the AGICO technician or by the authorised representative If you need later to reinstall the apparatus due to the moving the instrument to another place or any other reasons be sure the following conditions are met to achieve guaranteed parameters Choosing the place Place the apparatus to a room with relatively magnetically clean environment N The instrument must not be placed near sources of alternating magnetic field e g big transformers electric motors electricity power source wires thermal sources etc Do not place the instrument near thermal and electrical sources and prevent the pick up coils from direct sunshine The pick up unit must not be exposed to heat from the sun or from other sources which would affect the precision of measurement Do not place the pick up coils near the other instruments or computer monitors Do not place the instrument to a draughty room Air condition may somet
11. of 10 SI NORM Freq 1000 Field K11 K22 K33 K12 K23 K13 components of normed AMS tensor in the geographic system Cl code for the pair of mesoscopic foliation and lineation FOLII azimuth of dip and dip of the 1 foliation LINE1 trend and plunge of the 1 lineation C2 code for the 2 pair of mesoscopic foliation and lineation FOLI2 azimuth of dip and dip of the 2 foliation LINE2 trend and plunge of the 2 lineation 56 Structure of Geological Data File The GEOLOGICAL DATA FILE is a random access file with the length of the record being 64 bytes The numerical data are for practical reasons recorded as strings so that they can be directly checked Structure of the First Line Record N 2 LOCALITY LONGI LATI ROCK STRATI LITHO REGIO ORIENT P EOL The first record contains the locality data N42 number of specimens in the file 2 LOCALITY name of locality LONGI geographical longitude of the locality LATI geographical latitude of the locality ROCK rock type STRATI stratigraphical position LITHO lithostratigraphy REGIO regional position ORIENT orientation parameters P1 P2 P3 P4 EOL end of line sequence Structure of the Other Lines Records SPEC ORIENTATION CODE FOLII LINE1 CODE2 FOLD LINE2 The second record and the following records contain the specimen data Each record contains SPEC name of the specimen ORIENTATION azimuth and dip of the fiducial mark orienting the specimen CODEI co
12. the measurement of the empty holder starts and the following information subsequently appear on the computer screen AUX 200 A m BULK ANISO HOLDER Bulk Cos Sin Old values 4 138E 06 10 E 09 2 7E 09 4 179E 06 29 E 09 19 E 09 4 140E 06 36 E 09 13 E 09 4 113E 06 11 E 09 14 E 09 New values 4 144E 06 25 E 09 15 E 09 Std error 33 E 09 13 E 09 3 5 E 09 Press lt CR gt to save the data In the beginning of the procedure the Old values data appear on the screen These are the data stored in the configuration file obtained in the last measurement of the empty holder 7 The bulk susceptibility of the empty holder is automatically measured three times its mean value and standard error of the average are calculated Than the anisotropy of the empty holder versions FA and A is measured three times and the mean value and standard error of the average are also calculated If the measurements are inconsistent for example if holder bulk susceptibility does not lie within the expected interval or the standard error is greater than 0 1x10 the New values are blinking It is upon the operator whether to save the holder or not depending also on how strong specimens are measured m If you switch to frequency F2 F3 or change the calibration standard nominal value s the Holder Correction is reset to zero If you switch back to frequency Fl the last saved holder values for Fl are recalled from configuration file SAFYR S
13. the measuring coil from where pull it out after the second beep Then change the specimen position and continue analogously until the first 5 directional susceptibilities are measured Then fix the specimen to initial position for next 5 directional susceptibilities P6 P10 be sure the slot is up and black strips are visible front of you and follow the next five directions with the same design but now using the double arrow Continue with last 5 directional susceptibilities set the initial position for P11 P15 and follow the same design with triple arrow Static specimen measuring positions Initial position P6 P10 Initial position P11 P15 The users of the Kappabridges MFK1 FA or MFK1 A can use the arrangement with manual holder adapter This is useful for automatic field susceptibility variation measurement but can be used also for AMS static specimen measurement if one prefers such a way In this case do not insert the plastic cylinder into the coil Set the ROTATOR DISABLED and U D ENABLED AuxKey 2 in Auxiliary Menu 24 The results look like in the following example DATA MEASURED RESIDUALS in 30 41E 03 32 25E 03 31 54E 03 0 12 0 19 0 03 31 27E 03 31 42E 03 31 79E 03 0 11 0 13 0 05 30 60E 03 31 20E 03 32 63E 03 0 13 0 28 0 12 30 44E 03 32 33E 03 31 60E 03 0 02 0 05 0 24 30 29E 03 31 45E 03 31 85E 03 0 03 0 02 0 22 Std error 0 18 Anisotropy test 356 1 3229 135 6 Confidence angles 3 3 5 1 2 0
14. the susceptibility of the specimen is measured subsequently in 15 directions following the rotatable Jelinek s design in exactly the same way as in the KLY 2 KLY 3 and KLY 4 Kappabridges Using the least squares method the susceptibility tensor is fit to these measurements of the 15 directional susceptibilities and the errors of the fit are calculated The results of the measurement in the form of various parameters derived from the susceptibility tensor and orientations of the directions of the principal susceptibilities in various coordinate systems are presented on the screen or written on the disk The tensor elements together with orientations of mesoscopic foliations and lineations can be also written on the disk into standard AMS file which is binary random access file with extension ran from where they can be read in advanced processing Function Key 1 Dir 15 The procedure serves for the measurement of 15 directional susceptibilities The Jelinek s design of the 15 directions is shown in the following scheme The position design is the same for the cubic and cylindrical specimens After pressing Fl the following appears on the screen DATA MEASURED RESIDUALS Next direction 1 Press lt SpaceBar gt to continue Put the specimen into the holder in the Pos 1 see following scheme be sure the slot is up and five black strips are visible front of you press the SpaceBar key and wait the computer s beep Then insert the specimen into
15. to 0 3 SI 0 to 0 7 SI 2x 10 6x 10 12x 10 15 to 35 C 0 1 0 3 Holders for specimens of slightly different size can be supplied on request 2 MFK1 versions FA and A 3 MFKI all versions Accuracy of the absolute calibration HF Electromagnetic Field Intensity Resistance Power requirements Power consumption Relative humidity Dimensions Mass MFK1 Control Unit MFK1 Pick up Unit Rotator 12 3 1 Vm 240 230 120 100 V 10 96 50 60 Hz 40 VA max 80 230 mm x 210 mm x 130 mm 4 kg 280 mm x 355 mm x 320 mm 11 kg 320mmx 70mmx 65 mm 1 kg 13 CS4 CSL Description The CS4 and CSL High Low Temperature Apparatuses have been designed for measurement in connection with MFK1 FA or MFK1 A Kappabridges The CS4 High Temperature Furnace Apparatus is used for measurement of the temperature variation of low field magnetic susceptibility of minerals rocks and synthetic materials in the temperature range from ambient temperature to 700 C The apparatus consists of non magnetic furnace with a special platinum thermometer CS4 temperature control unit laboratory power supply EA PS cooling water reservoir with pump and argon flow meter The specimen is placed in a measuring vessel which is heated by a platinum wire in three selectable heating rates The temperature is measured by special platinum thermosensor The protective argon atmosphere can be applied during heating to prevent oxidation of mea
16. 1 350 system I 59 17 25 0 0315 0 0650 0 2305 Paleo 1 D 337 75 172 0 9205 1 1277 0 9518 System I 45 8 44 0 0106 0 0595 0 3107 Tecto 1 D 37 135 232 1 0668 0 9815 0 9518 System T 45 8 44 0 0950 0 2393 0 2069 12 07 2007 The data page can be left by pressing ESC key 31 Measuring of AMS spinning specimen method The measurement of AMS using spinning specimen method with rotator is available with the MFKI FA or MFKI A Kappabridges options U D ENABLED and ROTATOR ENABLED During measurement the specimen slowly rotates subsequently about three perpendicular axes in the same way as KLY 3S and KLY 4S Kappabridges The bridge is zeroed after inserting the specimen into the measuring coil so that susceptibility differences are measured during specimen spinning 64 measurements are made during one spin which results in very sensitive determination of the anisotropic component of the susceptibility tensor profiting from the measurement on the lowest possible and therefore most sensitive range Then one bulk susceptibility value is measured along one axis and the complete susceptibility tensor is combined from these measurements The measured data in the form of various parameters derived from the susceptibility tensor and orientations of directions of the principal susceptibilities in various coordinate systems are presented on the screen The tensor elements together with orientations of mesoscopic foliations and lineations can be wr
17. 2 Corr 3 4 5 Eval 6 7 8 9 Kill 10 Aux The three columns DATA MEASURED show the values of 15 directional susceptibilities measured The data RESIDUALS represent the deviations of the measured and fitted data After fitting the susceptibility ellipsoid to the measured data using the least squares method the susceptibility in each measuring direction is calculated from the fitted tensor and subtracted from the measured value this is the residual The residuals are the lower the higher is the measuring accuracy and better fit Ideally the residuals are as low as the measuring errors of individual directional susceptibilities Std error is the mean value of the absolute values of the residuals The quality of the measurement can be evaluated also from the values of Anisotropy test and Confidence angles The Anisotropy test values are the values of the F test for anisotropy isotropy and for triaxial rotational prolate and for triaxial rotational oblate ellipsoids If the left value is higher than 3 48 then the differences between the principal susceptibilities determined by measurement compared to measuring errors are great enough that the specimen can be considered anisotropic from the statistical point of view on the 95 level of significance If the central and right values are higher than 4 25 then the ellipsoid is triaxial The Confidence angles values are the angles defining the statistical accuracy of the determination of the individual princip
18. 2 k3 k1 k2 2k3 53 Usual Abbreviation P In P In P In L In F k1 gt k2 gt k3 are principal normed susceptibilities and n1 n2 n3 are their respective natural logarithms the symbol r gt means default set of AGICO stored in PAFA SAV configuration file of a new instrument 54 Structures of Data Files The program SAFYR works with the following data and configuration files The following notation is used for variables denotation n nbytes string variable 2I two bytes integer variable 4R four bytes real variable The classical sequential ASCII file with extension ASC contains the printable output results of measured specimen s in the same form as on the screen The Standard Anisotropy File AMS contains the results of the anisotropy measurement in binary form and has extension RAN The Geological Data File contains only the geological data orientations of specimens and of mesoscopic fabric elements its extension is GED File Pafa SAV contains current set of orientation parameters and anisotropy factors Files Safyr SAV contains serial port number calibration standard values instrument configuration parameters instrument gain and phase coefficients calculated during calibration procedure for field 200 A m peak holder components measured during holder correction routine and actual current volume or mass of specimen File Aniso TMP contains the screen contents results of AMS measuring of the
19. 5 different orientations following rotatable design From these values six independent components of the susceptibility tensor and statistical errors of its determination are calculated The specimen positions are changed manually during measurement The MFKI FA and MFKI A versions measure the AMS of a spinning specimen fixed in the rotator In the spinning method the same as in KLY 3S and KLY 4S bridges the specimen rotates with small speed of 0 4 r p s inside the coil subsequently about three axes From these data the deviatoric susceptibility tensor can be computed This tensor carries information only on anisotropic component of the specimens For obtaining complete susceptibility tensor one complementary measurement of bulk susceptibility must be done The zeroing of the bridge which works over the entire measuring range enables to zero the bridge prior to the anisotropy measurement after inserting the specimen into the measuring coil The background bulk susceptibility is eliminated and the bridge measures only the susceptibility changes during specimen rotation and thus the most sensitive range can be used The result is high precision of measurement and determination of principal directions of susceptibility tensor The specimen is adjusted only in three perpendicular positions Thus the specimen measurement time was dramatically shortened The measurement is rapid about two or three minutes per specimen depending on range and pre
20. AV 50 Function AuxKey 5 Field Set Field Set the field from 2 A m to maximum field depending on current frequency and properties of pick up coils in step of 1 A m The nominal maximum fields are as follows F1 976 Hz 700 A m peak value F2 3904 Hz 350 A m peak value F3 15616 Hz 200 A m peak value The real maximum field may be slightly higher than nominal maximum field The real maximum value is displayed at enter field prompt Set Field is not available during AMS measurement in individual positions Function AuxKey 6 ActVol Actual Volume Set the actual specimen volume which is used for bulk susceptibility calculation This key is active if the Volume Mode is selected see AuxKey2 ACmd lt Vol gt Function AuxKey 6 Mass Actual Mass Set the actual specimen mass which is used for mass susceptibility calculation This key is active if the Mass Mode is selected see AuxKey2 ACmd lt Mass gt Function AuxKey 7 Help Help Page Pressing this key invokes the list of help page To quit help page press ESC key 51 Function AuxKey 8 Freq Set Frequency Fl 976 Hz 200 A m peak F2 3904 Hz 200 A m peak F3 15616 Hz 200 A m peak This routine allows to select one of three available frequencies FA and FB options If the frequency is changed the field is set to default 200 A m In case of F2 and F3 the holder correction values are zeroed If you return to F1 the holder values are restored from configuration file where they were save
21. EYS ANS cei BAe hid m m Su a aG Raat ed umola Guanes eS eh 34 Function KEPA TS aus Sun ate la a edes 34 FUNCION Keko EUCLA p usa nd let e uu 34 Function Rey 5 Eval ocio ettet ient ee ee Hee Pete Ie RE de ae nestles eo Credere Eee denotes 35 Function Key 6 lt ActVol or Mass oe epe eere ne ec pend etes 39 Function Key Help saa nn n aa p uuu eg Re EE bad bug a Pel ERE Qe iaa 39 Function Key 9 Kill a sees eed tees seas r deus cec tedio be deep 39 FUNCION KEV LO AUN soe n e teer ete e m ee eren 39 BRIEF DESCRIPTION OF THE AUXILIARY MENU 2 S an a teeth dla Sl tr ete ero eee 40 Function AuxKey 1 Bsus Bulk Susceptibility sss 41 Function AuxKey 2 Aemd Auxiliary Commands seen eterne 43 AuxKey 2 CStd Calibration Standard seen 44 AuxKey 2 Orpar Orientation Parameters sess eerte rennen 44 AuxKey 2 Anfac Anisotropy Factors essent ener nrennet rennen nretre eret eene 45 AuxKey 2 Vol Set Volume Mode sse eene nnns 45 AZuxKey 2 Mass Set Mass Mode e ae Hehe ur hi usuka ote A a ete 45 AuxKey 2 Sigma Standard Error sse ener nette netten enne 46 AuxKey 2 List Parameters List aia m teo e te en tim get 46 Function AuxKey 3 Cal Calibration sse rene 47 Function AuxKey 4 Hol Holder Correction sse nnne 49 Function AuxKey 5 Field Set Field so it iae eee ig d abe bb eite 50 Function AuxKey 6 ActVol Actual Volume essent
22. For this test 1t is recommended to use manual holder fixed in the adapter in case the test is performed with a particular specimen For magnetic environment test 46 the Up Down mechanism is usually disabled or empty clean manual holder is used Do not disturb test measurement by moving anything in the vicinity of instrument The temperature in the room should be also stable AuxKey2 List Parameters List This function prints the contents of the configuration files SAFYR SAV and PAFA SAV to see the current configuration and calibration parameters 47 Function AuxKey 3 Cal Calibration This procedure serves for the calibration of the instrument This calibration is made as for the bulk susceptibility value along the vertical x3 axis of the calibration standard as well as for the anisotropy represented by the susceptibility difference between the standard susceptibility along the x3 axis and the perpendicular direction The standard is fixed in the holder vertically in the first measuring position for spinning specimen Perform the calibration after at least 10 minutes of warm up time Generally it is recommended to calibrate the bridge every day before beginning the work However since the gain changes of the instrument are usually very small and in the case the absolute value of susceptibility is not precisely important e g in AMS measurement of principal directions it is not quite necessary to calibrate the bridge every day The instrum
23. GICO Kappabridge MFK1 series Kappabridge and its optional accessories represent modular system designed for measurement of magnetic susceptibility and its anisotropy in variable fields and optionally at three different frequencies In conjunction with furnace or cryostat apparatus also for measurement of temperature variation of magnetic susceptibility Preface The User s Guide is divided into two parts a The Part 1 Kappabridges MFKI FA FB A B contains general common information description and specifications of individual modules and describes the capabilities of the system The attention is focused on measurement of anisotropy of magnetic susceptibility AMS and automatic field variation measurement using the Kappabridges MFK1 FA or MFKI A equipped with an up down mechanism and a rotator for spinning specimen measurement method The options MFKI FB or MFKI B are intended for AMS measurement using static specimen method with manual holder 15 directions method and manual holder susceptibility measurement of individual specimens and or field variation The Part 2 Apparatus CS4 CSL describes the measurement of temperature variation of magnetic susceptibility using the high temperature furnace CS4 and low temperature cryostat CSL KAPPABRIDGE MFK1 FA MFK1 A MFK1 FB MFK1 B User s Manual Instrument for measuring magnetic susceptibility and its anisotropy in variable fields at three frequencies AGICO Advanced Ge
24. ION vested etie eere mer ee iie ero e ede d ae ina AL ENDE GR 15 INSTALLATION PROCEDURES wisscssesiccecdeciseesissusassenecassseelecceensersessensessonceseucesssursuscapsuasadeotacetarsnenssossssuacecsesceseuceunsusse 16 Choosing th place ii E E pe ee a E a w aus ette 16 Tnterconnection of Unis te deno Ge n PP RESO E Co RU e e Eb de NET ete DEDE 16 Connection Kappabridge to computer sees tette retener nnne nnns 18 Testing the magnetic environment a ales eratac dies bed hts re 19 OPERATING BASIC 20 BRIEF DESCRIPTION OF THE MAIN MENU at Jose dre eee tenta ah erp pecu eq reset cene e renean 21 MEASURING OF AMS STATIC SPECIMEN METHOD ccssccsscssecsscesecssecsecesscneesecscesseccessecssesseessecseceasoneeeasensesecaeeatonseeseenaes 22 Function Key Ll DWTS ees sees ale aaa E tete esee daca hu aan aan s 22 Static specimen measuring positions esses nan cananea anna nana nana nana 23 Function Key 2 COM rer team en eb nae tdt datae taedas aie 26 Function Key F SE VAL 2 a veio Reed e ca n oie dieu eeiam 26 MEASURING OF AMS SPINNING SPECIMEN METHOD nennen enhn tette ter enne nn nono corner ee tnen nee Eer nne tne nnne tee tenete enne 31 Spinning specimen measuring positions and laboratory marking sese 32 Function Key L AxL 0 Rede ii idera na idas 33 TFunction Key 2 4x2 ont A ette odit en dco tute 34 FUNCION K
25. Ver 4 0 Mar 2009 MFK1 FA CS4 CSL MFK1 A CS4 CSL MFK1 FB MFK1 B User s Guide Modular system for measuring magnetic susceptibility anisotropy of magnetic susceptibility and temperature variation of magnetic susceptibility AGICO Advanced Geoscience Instruments Co Brno Czech Republic Contents INTRODUCTION TO THE USER S GUIDE 4 PREPACE RE 4 WARRANTY 2 40 eere e reget e PEN aM E deett a T Re ee SR UU RR DAR EROR 6 GENERAL SAFETY SUMMARY wigiccccccesssccccsccctscscsessessesedeacnssstecentsasecdocstsoessecsesdesevetucdeedudesdaaietssedoecisdecesedsececdsatzatenssies 7 SAVAIT EE Z Injury PrecautiQns 5 u S uei ISA eR S D S S hu E PERRO baka ERE u UE 7 Product Damage PrecautiOns 5 iiio rcm S D RP Herr er ne treni hed dac ee IC OL MULIER EVI OL J eissa stss 9 MFKI FA FB A B DESCRIPTION eese eene eene a nenne ernnr sese teens esse ete ensn sss esee nns s eset einn asse ete e enn n s eterna nnn 9 MEK 1 SPECIFICATIONS i rrr ERE ERE UR q Aaa GS A EIER A rers 11 CS4 CSE DESCRIPTION 222 Heri nies GREGG EE E AA a nanus 13 St CSTs SPECIFICATIONS d e oet eu doa RO en eu y esto Mao to o tes 13 EC DECLARATION OF CONFORMITY 5 555255 pter Eten trt S egre apy isa RS AE OEE run ki deg e oae UNE RENE RE teo Reese reed 14 UNPACKING INSTRUCTIONS STORAGE AND TRANSPORTAT
26. al susceptibilities directions on the 95 level of significance for more details see AGICO Print No 1 26 Function Key 2 Corr This key may be activated during or after the 15 directional susceptibilities are measured It enables any imprecisely measured directional susceptibility to be re measured during the measurement pressing Corr sets the position number to the current position minus one After complete measurement and pressing F2 input the Direction to be repeated and re measure the corresponding directional susceptibility The proper specimen position should be prepared before pressing F2 key The re measurements in various directions can be repeated until the expected accuracy is reached Function Key 5 Eval This procedure evaluates the measured data through the determination of the susceptibility tensor and its related parameters Before this procedure is activated it is possible to repeat measurement of any of the 15 directional susceptibilities in order to get the best data for the evaluation After the evaluation is once started none of the directional susceptibilities can be re measured only the whole specimen can be re measured n If the Eval procedure is started for the first time the following questions subsequently appear on the screen Path drive dir1 dir2 lt CR gt current Name of file without extension 8 chars max Each of associated files contains x record s Specimen name means new file
27. cise profiting from many susceptibility determinations in each plane perpendicular to the axis of specimen rotation The static method of the measurement can also be used Software SAFYR combines the measurements in three perpendicular planes plus one bulk value to calculate a complete susceptibility tensor The errors in determination of this tensor are estimated using a special method based on multivariate statistics principle MFK1 Specifications Specimen Size Cylinder Diameter Length Cube Cube ODP box Fragments bulk susc Pick up coil inner diameter Nominal specimen volume Up Down mechanism ver FA and A ver FB and B Operating frequency ver FA and FB ver A and B Field intensity Field homogeneity Fl and F2 F3 Measuring range Fl 976 Hz F2 3904 Hz F3 15616 Hz Sensitivity F1 400 Am peak SI units F2 200 Am peak SI units F3 200 Am peak SI units Operating temperature range Accuracy within one range Accuracy of the range divider Spinning Specimen 25 4 mm 02 1 5 22 0 mm 40 5 1 5 20 mm 0 5 1 5 11 Static Specimen 25 4 mm 41 0 1 0 22 0 mm 2 0 2 0 20 mm 40 5 2 0 23 mm 40 5 2 0 26x25x19 5 mm 40 cm spinning spec static spec 43 mm 10 cm yes no Fl 976 Hz F2 3904 Hz F3 15616 Hz Fl 976 Hz 2 Am to 700 Am peak at Fl 2 Am to 350 Am peak at F2 2 Am to 200 Am peak at F3 0 5 1 0 to 0 9 SI 0
28. ction omission of which may cause lost of properties damage or injury Injury Precautions Use Proper Power Cord To avoid fire hazard use only the power cord specified for this product Do Not Operate Without Covers To avoid electric shock or fire hazard do not operate this product with covers or panels removed Fasten Connectors Do not operate the instrument if all connectors are not properly plugged and fixed by screws Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or damp conditions Do Not Operate in an Explosive Atmosphere To avoid injury or fire hazard do not operate this product in an explosive atmosphere Disconnect Power Source To avoid risk of electric shock unplug the instrument from mains before reinstalling or removing unit Product Damage Precautions Use Proper Power Source Do not operate this product from a power source that applies more than the voltage specified Use Proper Fuses only Do not use fuses which are not specified by the manufacturer If a fuse with a different characteristics or value is used the protection is not effective Operator s Training Operator should be familiar with operation of the instrument and Safety Regulations Use Manufacturer s Cables Only Other devices can be connected to the instrument via the appropriate cables only Do Not Disconnect Connectors To avoid damage of the instrument never disconnect any con
29. cturer AGICO s r o Jecn 29a CZ 621 00 Brno ICO 607 313 54 Place of producing AGICO s r o Je n 29a CZ 621 00 Brno ICO 607 313 54 fulfils the applicable requirements of following regulations normative documents and technical specifications CSN EN 55022 class B CSN EN 61326 1 for basic requirements CSN EN 61000 4 2 criterion B CSN EN 61000 4 3 criterion A CSN EN 61000 4 4 criterion B CSN EN 61000 4 5 criterion B CSN EN 61000 4 6 criterion A CSN EN 61000 4 11 criterion A for short time dip criterion B for short time interrupt N 55022 1998 CSN EN 55022 1999 A1 2001 A2 2001 Z2 2001 N 61326 1 2006 CSN EN 61326 1 2006 N 61000 4 2 1995 A1 1998 CSN EN 61000 4 2 1997 A1 1999 N 61000 4 3 2006 CSN EN 61000 4 3 ed 3 2006 N 61000 4 4 2004 CSN EN 61000 4 4 ed 2 2005 N 61000 4 5 2006 CSN EN 61000 4 5 ed 2 2007 N 61000 4 6 1996 A1 2001 CSN EN 61000 4 6 1997 Z1 2001 N 61000 4 11 2004 CSN EN 61000 4 11 ed 2 2005 E E E E E E E E The judgement of conformity was performed in co operation with the ITI TUV s r o Prague Place and date of issue Brno March 2009 Responsible person Ing Petr SUZA development engineer 15 Unpacking Instructions Remove carefully the instrument and its accessories from the box and packing material referring to the packing list included to confirm that everything has been delivered Briefly inspect each item for shipping damage If anything is
30. d After pressing the F5key the required integer value of peak Field can be entered Setting is not available during AMS measurement procedure 35 Function Key 5 Eval This key is available if all positions of AMS measurement procedure have been completed This procedure evaluates the measured data through the determination of the susceptibility tensor and its related parameters Before this procedure is activated it is possible to repeat any of the procedures Ax1 Ax2 Ax3 TSus3 in order to get the best data for the evaluation When any of the above procedures is completed the denotation of the respective key is supplemented by an asterisk If the evaluation is started none of the above procedures can be repeated only the whole specimen can be re measured If the Eval procedure is started for the first time the following questions subsequently appear on the screen Path drive dir1 dir2 lt CR gt current Name of file without extension 8 chars max Each of associated files contains x record s Specimen name means new file After the above information are input the question appears for the way of inputting the geological orientation data Select Using geological file 1 Manual input from memo book 2 Non oriented specimen 3 One selects 1 if the data should be read from the geological data file created earlier the geological data file can be created using the ANISOFT program package which is located i
31. d during last holder correction procedure Program than waits 10 min to eliminate coil drift caused by frequency change The time can be reduced by the user by pressing lt Esc gt key This is not recommended in case of measuring weak specimens and or using low fields Function AuxKey 9 Kill The program breaks the current activities and clears current specimen data Function AuxKey 10 Main Return from AUXILIARY menu to the MAIN menu Appendices This chapter covers the following topics Structures of Data File CJ Selection of Coordinate Systems CJ Geological Locality Data 1 List of Magnetic Anisotropy Factors 52 List of Magnetic Anisotropy Factors Factor No DIARRA WN A DoD Mathematical expression 15 2 k1 k 2 k2 k 2 H K3 k 2 3 k 2 exp sqr 2 n1 n 2 n2 n 2 n3 n 12 sqr 2 n1 n 2 n2 n 2 n3 n 2 k1 k3 In k1 k3 100 k1 k3 k1 k1 k3 k2 k1 k3 k k1 k2 In k1 k2 k1 k2 k 2k1 k2 k3 k2 k3 In k2 k3 k1 k2 2k3 k1 k3 2k2 2k2 k1 k3 1 k3 k2 2k1 k2 k3 k1 k3 k1 k2 2 k3 k k2 k3 k k1 sqr k2 k3 k1 k3 k2 2 k1 k2 k1 k2 2 k3 k1 k2 k2 k3 k2 k3 k1 k2 arcsin sqr k2 k3 k1 k3 k2 2 k1 k3 k2 k1 k2 k1 k2 k3 k2 k3 1 k1 k2 1 2n2 n1 n3 n1 n3 2k2 k1 k3 k1 k3 k1 k2 2k3 k1 k2 sqr k1 k 2 k2 k 2 k3 k 2 3Y k k1 k2 k3 1 3 k3 k1 k2 k1 k2 k3 k3 k1 k2 k2 2 k1 k3 k1 k2 2k1 k2 k3 k
32. d mesoscopic foliation trend plunge of the second mesoscopic lineation respectively If only foliation exists the second character of the code must be zero and the last two angles are also zeros If one selects 3 no angle data are necessary After the geological data are input the program displays the results and after pressing ESC key the program asks Output to file Y N Enter YES Data are written as an ASCII file in the same format as they appear on the screen later they can be re printed on the paper if necessary The extension of this file is ASC and the file is located in the same directory as the standard AMS file After measuring the second or later specimen only the question for the specimen name appears on the screen The data are handled in the same way as those of the first specimen If one wishes to change the file one inputs instead of the specimen name and the inputting is made as in the first specimen Then the calculated data are shown on the screen in the form whose example is shown on the page 38 The meaning of the presented results is as follows Azi first orientation angle mostly azimuth of the dip or strike of the fiducial mark on the specimen Dip second orientation angle dip of the fiducial mark or plunge of the cylinder axis O P orientation parameters see the section OrPar Nom vol nominal volume of the used pick up unit mostly 10cm Act vol the volume of the specimen measured i
33. de for the 1 pair of mesoscopic foliation and lineation FOLII azimuth of dip and dip of the 1 foliation LINEI trend and plunge of the 1 lineation CODE2 code for the 2 pair of mesoscopic foliation and lineation FOLD azimuth of dip and dip of the 2 foliation LINE2 trend and plunge of the 2 lineation FREE four characters free string Note The orientation of mesoscopic foliation should be measured in terms of azimuth of dip and dip or strike and dip and this is indicated by the orientation parameter P4 see the section Orientation Parameters The azimuth of dip or strike should be measured as angles ranging from 0 to 360 degrees not from zero to 180 degrees and they are recorded in the geological data file as measured values from 0 degrees to 360 degrees values from 0 degrees to 180 degrees for the reason of compatibility with the AMS file only 2 characters are used values from 0 degrees to 90 degrees 57 Selection of Coordinate Systems The orientations of magnetic foliation and magnetic lineation can be presented not only in the standard geographical coordinate system but also in the so called palaeo geographical system after rotation of the mesoscopic foliation under consideration into the horizontal position about the corresponding lineation or in the so called tectonic coordinate system mesoscopic lineation and foliation are the coordinate axes The program can work with up to 2 pairs of mesoscopic fol
34. details see AGICO Print No 6 m The program shows the set of current orientation parameters Orientation parameters Any changes Y N Ifyou enter N the shown parameters are used in the subsequent calculations 7 If you enter Y new parameters are set up Computer asks for inputting the P1 P2 P3 and P4 parameters and check them for their validity These new parameters are written into a configuration file 45 AuxKey 2 Anfac Anisotropy Factors Magnetic fabric can be visualized by the shape and orientation of the anisotropy ellipsoid The eccentricity and shape of the ellipsoid can be characterized by conveniently chosen parameters derived from the principal values parallel to the axes of the anisotropy ellipsoid Unfortunately more than 30 parameters have been suggested for this purpose even though 2 parameters are sufficient to characterize the eccentricity and shape Some of them are listed in the enclosed Table As it is not reasonable to present them all our program selects up to 8 parameters according to the demands of the user a The selection is made as follows First the set of the previously used parameters appear on the screen Current anisotropy factors together with the question Any changes Y N If you wish to change this set the program shows the table of factors from which you can select new set and asks for Count of factors Input the number of selected factors in our case 8 Then enter the sequen
35. e susceptibility is measured 64 times during one revolution As the bridge is zeroed with the specimen inserted into the measuring coil before the specimen starts spinning the susceptibility differences are measured between the susceptibilities along the respective directions and that of the direction in which the bridge was zeroed This way of measurement is very advantageous because one measures only the anisotropic component of the susceptibility which is much lower than the bulk component and one can profit from the higher accuracy of the measurement made on the more sensitive range Before pressing Key F1 the specimen should be fixed into the specimen holder in the measuring position No 1 After pressing F1 the specimen is automatically inserted into the specimen coil the bridge is zeroed and the specimen starts spinning during spinning the specimen susceptibility is measured The results are presented in the form as in the following example Ax Range Cosine Sine Error Error 1 1 5 709E 06 2 102E 06 8 2E 09 0 14 Ax means that the specimen spins about the x axis the measurement in the x2 x3 plane Position No 1 Range informs us of the range on which the anisotropy was measured this is only formal information because the instrument has a fully auto ranging feature Cosine and Sine give the values of the cosine and sine components respectively of the average anisotropy curve Error gives the standard devia
36. eal component of susceptibility and its standard error calculated as the average of averages of individual measurement sets 10 x 10 measurements Check also all individual averages of each measurement set and its distribution Operating Basics This chapter covers the following topics 1 Main Menu of the SAFYR Brief description of the Main Menu Measuring of AMS using static specimen method in 15 Directions Measuring of AMS using spinning specimen method with Rotator Field and Frequency Setting O Auxiliary Menu of the SAFYR a Brief description of the Auxiliary Menu Measuring Bulk Susceptibility and Field Variation Setting Parameters Configuration Calibration Standard Volume Mass Calibration Correction for the Holder Appendices List of Magnetic Anisotropy Factors Structures of Data Files Selection of Coordinate Systems Geological Locality Data 20 21 Brief description of the Main Menu F1 F1 F2 F3 F4 F4 F5 F5 F6 F6 F7 F8 F8 F9 15dir lt Ax1 gt lt Ax2 gt lt Ax3 gt lt TSus3 gt lt TSus gt lt Field gt lt Eval gt lt ActVol gt lt Mass gt lt Help gt lt Freq gt lt Stop gt lt Kill gt F10 lt Aux gt Measuring Anisotropy in 15 Directions using manual holder instrument option FB and B Measuring Anisotropy in Pos 1l rotation about axis 1 instrument option FA and A
37. efining the specimen sampling Orientation Parameters i e the meaning of two sampling angles lt AnFac gt Defining the set of Anisotropy Factors lt Vol gt Set the Volume mode for bulk susceptibility measurement lt Mass gt Set the Mass mode for mass susceptibility measurement lt Sigma gt Execute the test measurement with standard error evaluation lt List gt List the configuration files AuxKey2 Ctrl Cmd Control Commands UP DOWN is used for checking the Up Down Mechanism performance and speed The movement up takes a little bit longer time than down but it should not exceed 3 6 s The timeout for error message is 4 s This option is not available for versions FB and B ENABLE DISABLE Up Down Mechanism is used in case you wish to use manual holder measurement routine AMS in 15 directions for instruments versions FA and A ENABLE DISABLE Rotator is used in case you wish to use manual holder measurement routine in 15 directions for instruments versions FA and A It is useful also for field variation measurement using manual holder fixed in the adapter instead of rotator particularly at frequencies F2 and F3 where the influence of rotator may be more significant for weak specimens ZEROING is used for checking the zeroing capability and speed The timeout is about 12 s INIT sets the rotator to initial position checks the belt adjustment and speed of rotation This option is not available for versions FB and B In case the Up
38. ent configuration In case there are no problems the offer of the MAIN MENU appears For detailed information and explanation of the main menu see chapter Operating Basics 19 Testing the magnetic environment 7 See chapter Choosing the place before running test If you do not use notebook or computer with LCD display we recommend it take attention to your PC monitor The monitor distance from pick up coils and its azimuth position can have sometimes great influence Be sure there are no other instrument in the vicinity of pick up coils Run program SAFYR EXE 7 In the MAIN menu select function Keyl0 Aux then press Key2 Acmd If you have Kappabridge versions FA or A disable Up Down Mechanism disable Rotator and Select S to start the test measurement routine called Sigma Test see AuxKey2 Standard Error Test p 45 The measured data empty coil measurement at F1 and 400 A m peak are stored in the files which names are derived from current time in format HHMMSS with extensions K00 and R00 The file K00 contains the all measured data file ROO contains only the results average and standard error of 10 repeated measurements in one set Number of sets is also 10 The test takes approx 40 minutes Do not disturb test during measurement by moving anything in the vicinity of instrument Check the results saved at the end of files K00 and R00 denoted as Total Average the total average of measurement of the r
39. ent should be always calibrated when the frequency was changed and small susceptibility frequency variation of the specimen measured are expected The program displays the day of the last calibration and recommends the calibration in case the last one was performed more than 30 days ago After activating this procedure through pressing the AuxKey F3 the calibration procedure starts and the following information subsequently appear on the computer screen AUX 200 A m F1 ANISO Auto Range Calibration 200 A m PEAK Bulk Cos Sin Delta GainA GainB OLD 136 7 E 03 54 85 E 03 0 00 E 00 22 77 3 5023 3 5005 MEAS 136 6 E 03 54 84 E 03 38 51E 06 22 77 3 5023 3 5005 NEW 136 7 E 03 54 85 E 03 0 00 E 00 22 81 3 5027 3 5008 gt Press Enter to save calibration data Bulk the values of the bulk susceptibility of the calibration standard along the vertical x3 axis Cos the values of the cosine component of the anisotropy of the calibration standard Sin the values of the sine component of the anisotropy of the calibration standard Delta value represents the phase lag of the measured signal relatively to the position of the spinning specimen This lag is mainly due to the phase characteristics of the output low pass filter 7 GainB is the correction for getting the total gain for the bulk susceptibility to be measured precisely 48 GainA is the correction for getting the total gain for the anisotropy to be measured preci
40. eterne nennen entes 50 Function AuxKey 6 Mass Actual Mass essere enne e nenne nne nnne 50 Function AuxKey 7 Help Help Page c ccccecceccscsecsessssessessesseeeseeseesensenscescnseneceecnesnecesaessessesseeseeseeseesenseneeeeeateats 50 Function AuxKey 8 Freq Set Frequency sse eene nemen eret renhe trente erre tet rentre 51 Function AuxKey BUM ui ten t EHE Rd oretenus 51 Function AuxKey I0 Matuqa Sama mo dee e HE tet e e u e p C RDU 51 NAAA A X 52 List OF MAGNETIC ANISOTROPY FACTORS 2 7 e tee kiqa au a au RE ka debts 53 STRUCTURES OF DATA FILES oer Fert e a HEN ORO yaaa ha hanh ere eerte 54 Structure of Standard AMS File eese eene eere nne tee treten inrer eren nnt rennen 55 Siructure of Geological Data File er bte ree duae m ette es 56 SELECTION OF COORDINATE SYSTEMS vcn ON CERE CREE HUE GENDER UE IET ROGER UNI TU ERIT IEEE 57 GEOLOGICAL LOCALITY DATA IIS u e de tem RE er e LR e E e A Eee REL rete iti eris 58 APIDAE N DT LO E A 59 Cleaning the Holders and Rotator sees eerie tette tritt trennen ener 59 Cleaning the Rotator and Belt MFK versions FA and A seen 60 Rotator assembling MFK versions FA and A sese enne nennen enne 61 Cleaning the Up Down Mechanism eee ente ee oe gere ice siete ette the bete ioca lose pev did 62 Introduction to the User s Guide Thank you for purchasing magnetic susceptibility meter A
41. iation and lineation which are described by a two character code The first character of the code describes the foliation while the second character describes the lineation for the proposal of the codes see the enclosed table For example the code characterizing the existence of metamorphic schistosity and mineral alignment lineation is SA If only the foliation and no lineation exist the second character in the code is zero For example the system characterized by the bedding only has the code BO zero Codes characterizing mesoscopic foliations and lineations Code Foliation Code Lineation B bedding A mineral alignment C cleavage D bedding cleavage intersection K cataclastic schistosity F fold axis S metamorphic schistosity R striation J joint W wave hinge lineation G igneous banding P current direction E fluidal foliation M beta axis H schlieren foliation L lava flow lineation N lava flow foliation O schlieren lineation 58 Geological Locality Data The inputting of the locality data is not compulsory The ANISOFT package of programs for advanced AMS data processing does not work with the locality data These data serve only for storing the locality geological characteristics on the disk In inputting the locality geological data the following data are asked for NAME OF LOCALITY max 16 characters ENTER means no data This is the literary name of the locality it serves only as a note characterizing the l
42. imes cause higher thermal drift of coils prevent the direct air flowing in the room The temperature in the room should be stable as much as possible The temperature variation in the room should not exceed 2 C hour Place the instrument and pick up unit on a wooden table with good stability which has no iron part under working desk It is recommended to place the pick up unit on a separate stand or a small table which should be of such a height so that the middle of the pick up unit coincides with the level of the working table This arrangement makes the operation easier During measurement prevent motion of magnetically objects metal parts of chairs doors furniture watches rings tools components of your clothes etc Interconnection of Units Fig 1 shows the Interconnection Scheme If you are installing only Kappabridge do not consider CS4 laboratory power supply EA PS and its accessories Be sure the instrument is unplugged from mains during connecting the cables Do not manipulate with any connector while the instrument is ON Fix the connectors by screws plug the mains socket and switch the Kappabridge on 17 Fig 1 Interconnection Scheme MFK1 FA CS4 CSL O RJ VOL vsQuog asna H3 0d vu Fer asepayuj 5939 tt ES 198q Sq v3 Sd 4 dANd veg asn4 Sd Y3 91 TS vSO LYAW aui u3S uonaauuo212 u DIA 01 FSO Quo Sq v3
43. itten on the disk into standard AMS file which is binary random access file from where they can be read in advanced processing Do not forget to remove the plastic cylinder from the coil in case the previous measurement was done using the manual holder and with the option U D DISABLED Spinning specimen measuring positions and laboratory marking POS 1 spinning about axis x1 POS 2 spinning about axis x2 POS 3 spinning about axis x3 and Tsus3 SPECIMEN MARKING POSITION 1 POSITION 2 POSITION 3 D The specimen laboratory marking defines the specimen right handed oriented coordinate system and must be performed in the way displayed on the above picture and in the agreement with orientation parameters used in the particular laboratory Please note the orientation of the fiducial mark on the frontal side of cylinder which defines the X1 of specimen coordinate system may or may not be measured in the field and the orientation of the arrow on the case of cylinder drawn in the laboratory and drawn in the field may be in general different For this reason different orientation parameters are used for the information how the fiducial mark is drawn and for the information how the azimuth is measured 32 33 Function Key 1 Ax1 This procedure serves for the measurement of the AMS in the x2 x3 plane the specimen spins about the x axis The spinning is very slow one revolution per 2 5 seconds and th
44. ll screw remove the arm of the holder Check the position of the white teflon ring to reinstall it later into the same position and angle orientation Then remove the screw and the washer and at last the white teflon ring You can remove the rear panel of the Pick Up Unit only for the reason to see inside Using cleaning paper clean the space inside the cylinder Insert the paper into the gap between the inner cylinder and outer tube and clean the inner cylinder space by moving papers along inner surface of the cylinder and simultaneously by moving them a little bit up and down Clean the ring also and reinstall it checking its proper position and angle orientation inside the cylinder Tighten the screw fixing the washer gently For easier manipulation it is recommended to use the following size and shape of cleaning paper Paper strip about 30 cm in length of trapezoid shape with bases 5 and 2 cm thickness of about 0 1 mm standard office paper Do not lubricate the inner cylinder by any lubricant Do not use alcohol for cleaning may cause damaging of cylinder due to evaporated heat In case the cleaning does not fix the problem do not try to solve it yourself contact the manufacturer
45. n Dip second orientation angle dip of the fiducial mark or plunge of the cylinder axis O P orientation parameters see the section OrPar Nom vol nominal volume of the calibration standard 10 cm Act vol the volume of the specimen measured in cm Demag fac information whether the demagnetizing factor of the specimen was considered in the calculation of the mean susceptibility 28 Holder susceptibility of the holder measured in the section Hol Tl code for the first pair of mesoscopic foliation and lineation Fl orientation angles for the first foliation Ll orientation angles for the first lineation T2 code for the second pair of mesoscopic foliation and lineation F2 orientation angles for the second foliation L2 orientation angles for the second lineation Field and Frequency peak field in A m and F1 F2 or F3 Mean mean susceptibility SI units Standard err error in fitting the susceptibility tensor of the measured data F F12 E25 statistics for anisotropy triaxiality and uniaxiality testing Normed principal susceptibilities principal susceptibilities normed by the norming factor and errors in their determination 95 confidence angles E12 E23 E13 confidence angles on the 95 probability level in the determination of the orientations of the principal susceptibilities 29 Anisotropy factors Principal directions Normed tensor values of the selected anisotropy parameters orien
46. n cm Demag fac information whether the demagnetizing factor of the specimen was considered in the calculation of the mean susceptibility 37 Holder susceptibility of the holder measured in the section Hol Tl code for the first pair of mesoscopic foliation and lineation Fl orientation angles for the first foliation L1 orientation angles for the first lineation T2 code for the second pair of mesoscopic foliation and lineation F2 orientation angles for the second foliation L2 orientation angles for the second lineation Field and Frequency peak field in A m and F1 F2 or F3 Mean Standard err F F12 F23 Normed principal susceptibilities 95 confidence angles for rotation axis Ax1 Ax2 Ax3 Anisotropy factors Principal directions Normed tensor mean susceptibility SI units error in fitting the susceptibility tensor of the measured data statistics for anisotropy triaxiality and uniaxiality testing principal susceptibilities normed by the norming factor and errors in their determination semi axis of the confidence ellipse angles of the orientations of the principal susceptibilities values of the selected anisotropy parameters orientations of principal susceptibilities in decreasing succession as declination D and inclination I in various coordinate systems values of the normed susceptibility tensor in the appropriate coordinate system the upper line gives the diagonal tensor elemen
47. n the same directory as the standard AMS file being measured The reading is made automatically by program The geological data are used in the calculations and also copied into the standard AMS file see Appendix 2 If one selects 2 the following questions appear on the screen MANUAL INPUT FROM MEMO BOOK 2 sampling angles One inputs the angles of the orientation of the specimen the first is azimuth of the fiducial mark of the specimen the second is the dip or plunge of the fiducial mark for details see the AGICO Print No 6 Number of tectonic systems 0 to 2 If 0 is input for example if non foliated and non lineated volcanic or plutonic rock is measured no other geological data are input If 1 or 2 is input the following data must also be input 36 1 Code 4 tectonic angles The two character code characterizes the measured mesoscopic foliation and lineation the angles are azimuth of the dip or strike if the orientation parameter P4 is 90 dip of the first mesoscopic foliation trend plunge of the first mesoscopic lineation respectively If only foliation exists the second character in the code must be zero and the last two angles are also zeros If 21s input the following data must also be input 2 Code 4 tectonic angles The two character code characterizes the measured mesoscopic foliation and lineation the angles are azimuth of the dip or strike if the orientation parameter P4 1s 90 dip of the secon
48. nd beep and pull ver FB and B the specimen out The measurement is terminated by pressing Esc key The bulk or mass susceptibility is calculated using the specimen volume which is entered in procedure Key 6 ActVol or Mass To finish measurements press Esc key N Specimen Bulk 1 XY 4 58E 06 2 STANDARD 82 75E 03 3 12 45 07 82 75E 03 If you select lt S gt or lt R gt the standard up to 20 fields susceptibility measurements or reduced short curve field variation measurement is performed in the fields predefined in the file MFK VAR This file contains for each frequency the number of field points and fields The standard curve for frequency F1 is marked e g F1S the reduced FIR etc The maximum length of the reduced curve may be also 20 fields points So you can use the reduced curve for reduced measurement in number of field points or use this possibility as field zoom for certain field range reduced in field range or both The number of points has the priority it is not necessary to edit the rest list of individual fields if you wish to decrease the number of points only The Standard and Reduced are the names of two pre defined curves only they have no significant meaning 42 Examples Standard 20 points 25102030 4050607080 100 150 200 250300 350 400 500 600 700 Reduced 10 points 5 20 40 60 100 200 300 400 500 700 Reduced 20 points 1520253035 4045506065 75808590100 110120 130 140 150 Standard 15 point
49. nector while the device is on Do Not Operate With Suspected Failures If you suspect there is damage to this product have it inspected by qualified service personnel Getting Started In addition to a brief product description this chapter covers the following topics Specifications of Individual Modules Declaration of Conformity Unpacking Instructions Storage and Transportation MFK1 FA FB A B Description The MFK1 Kappabridges are probably the most sensitive commercially available laboratory instruments for measuring magnetic susceptibility and anisotropy of magnetic susceptibility AMS The Kappabridges have the following features High sensitivity Automatic zeroing over the entire measuring range Automatic compensation of both real and imaginary susceptibility components Auto ranging Measuring at three different frequencies version FA and FB Measuring of in phase and relative change of out of phase component Slowly spinning specimen version FA and A Quick AMS measurement FA and A Easy manipulation Automated field variation measurement FA and A Only three manual manipulations for measuring AMS FA and A Built in circuitry for controlling the furnace CS4 and cryostat CSL Full control by computer Sophisticated hardware and software diagnostics 10 The Kappabridge apparatus consists of the Pick Up Unit Control Unit and Computer In principle the instrument represents a
50. nstall the instrument without explicit written permission of AGICO company AGICO shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than AGICO representatives to install repair or service the product b to repair damage resulting from improper use or connection to incompatible equipment or c to service a product that has been modified or integrated with other products when the effect of such modification increases the time or difficulty of servicing the product This warranty is given by AGICO with respect to this product in lieu of any other warranties expressed or implied AGICO and its vendors disclaim any implied warranties of merchantability or fitness for a particular purpose AGICO s responsibility to repair or replace defective products is the sole and exclusive remedy provided to the Customer for breach of this warranty AGICO and its vendors will not be liable for any indirect special incidental or consequential damages irrespective of whether AGICO or vendor has advance notice of the possibility of such damages General Safety Summary Review the following safety precautions to avoid and prevent damage to this product or any products connected to it Only qualified personnel should perform service procedures Convention aN Symbol Attention is used to draw attention to a particular information Symbol Prohibition is used to accent important instru
51. ocality location etc z LOCALITY GEOGRAPHICAL LONGITUDE DECADICAL EXPRESSION LOCALITY GEOGRAPHICAL LATITUDE DECADICAL EXPRESSION Both in the format xxxx xx These data are input as decadical expressions not using minutes and seconds ROCK TYPE max 4 characters STRATIGRAPHY max 4 characters LITHOSTRATIGRAPHY max 4 characters REGIONAL max 4 characters These data are recommended to be input as 3 character codes their purpose 1s to characterize geologically the locality investigated they are not used in further calculations ORIENTATION PARAMETER Pl 5 ORIENTATION PARAMETER P2 g ORIENTATION PARAMETER P3 z ORIENTATION PARAMETER P4 See also Appendix Selection of Coordinate System 59 Maintenance Cleaning the Holders and Rotator It is recommended to clean the specimen holders regularly especially if you continue with measurement of relatively weak specimens after measuring strong ones For cleaning use pure water with a small amount of detergent and soft brush Do not use alcohol for cleaning may cause damaging due to heat of vaporization 60 Cleaning the Rotator and Belt MFK versions FA and A It is necessary to clean the shell in which the specimen is fixed and the other part of the rotator if rotator makes suspicious noise during spinning and or if the belt strain value see below is higher than 1750 4 1 Unscrew three plastic screws and washers
52. oscience Instruments Co Brno Czech Republic Warranty AGICO warrants that this product will be free from defects in materials and workmanship for a period of usually 1 one year from date of installation However if the installation is performed later than 3 three months after the date of shipment due to causes on side of Customer the warranty period begins three months after the date of shipment If any such product proves defective during this warranty period AGICO at its option either will repair the defective product without charge for parts and labour or will provide a replacement in exchange for the defective product In order to obtain service under this warranty Customer must notify AGICO of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service AGICO will decide if the repair is to be performed by AGICO technician or AGICO delegated serviceman in customers laboratory or product shall be sent for repair to the manufacturer In latter case customer shall be responsible for packaging and shipping the defective product to the AGICO service centre In both cases all the costs related to a warranty repair shall be at expenses of AGICO The warranty becomes invalid if the Customer modifies the instrument or fails to follow the operating instructions in case of failure caused by improper use or improper or inadequate maintenance and care or if the Customer attempts to i
53. precision fully automatic inductivity bridge It is equipped with automatic zeroing system and automatic compensation of the thermal drift of the bridge unbalance as well as automatic switching appropriate measuring range The measuring coils at frequency Fl are designed as 6th order compensated solenoids with a remarkably high field homogeneity Special diagnostics was embedded in MFK1 Kappabridges which monitors important processes during measurement with MFK1 and also with CS4 or CSL unit The digital part of the instrument is based on micro electronic components with two microprocessors controlling all functions of the Kappabridge The instrument has no control knobs it is fully controlled by external computer via serial channel RS 232C The main advantage of the new models MFK1 FA and MFKI FB is the possibility to measure bulk susceptibility and AMS at three different frequencies The auto ranging and auto zeroing work over the entire measuring range Automatic zeroing compensates real and imaginary components the zeroing circuits are digitally controlled by firmware The output signal from pick up coils is amplified filtered and digitalized raw data are transferred directly to the computer which controls all the instrument functions The MFK1 FB and MFKI B versions measure the AMS of a static specimen fixed in the manual holder In the static method the same as in KLY 2 KLY 3 and KLY 4 bridges the specimen susceptibility is measured in 1
54. s the curve finished at 300 A m 25102030 4050607080 100 150 200 250 300 350 400 500 600 700 In case the file MFK VAR contains any wrong data the number of points is greater then 20 or field is not available for current frequency the default field variation set is used and user file MFK VAR is ignored The default field variation set factory set MFK VAR 20 No of points Fl standard 2 5 10 20 30 40 50 60 70 80 100 150 200 250 300 350 400 500 600 700 FlS 10 No of points Fl reduced 5 20 40 60 100 200 300 400 500 700 FIR 16 No of points F2 standard 2 5 10 20 30 40 50 60 70 80 100 150 200 250 300 350 F2S 10 No of points F2 reduced 5 20 40 60 80 100 150 200 250 350 F2R 13 No of points F3 standard 2 5 10 20 30 40 50 60 70 80 100 150 200 F3S 7 No of points F3 reduced 5 20 40 60 100 150 200 F3R 43 Function AuxKey 2 Acmd Auxiliary Commands Auxiliary Commands procedures serve for setting and checking the SAFYR configuration parameters for control commands and test some instrument functions Ctrl Cmd gt Up Down set the Holder position out of coil in the coil Enable Disable Up and Down options FA and A Enable Disable Rotator options FA and A zeroing set the bridge nucleus to zero condition Init set the Holder Up and set the Rotator to index pulse lt CStd gt Entering the Nominal Directional Susceptibilities of the Calibration Standard lt OrPar gt D
55. sely The line headed OLD gives the above data of the last calibration corresponding to those written in the configuration file The line MEAS gives the data actually measured standard assuming that Delta GainB and GainA values equal OLD values The line headed NEW shows the result of the above measurement but with proper new corrections Delta GainB GainA The constants Delta GainB GainA are also written into the configuration file The calibration results are verified to prevent writing erroneous values into the configuration file Thus the GainB and GainA values should be within the interval of 65 to 135 of the nominal values otherwise the error is indicated If you change the calibration standard nominal value the GainB and GainA are undefined until proper calibration is performed successfully After successfully performed and saved calibration the all holder values are zeroed For instrument versions FB and B the part of calibration using rotator is skipped 49 Function AuxKey 4 Hol Holder Correction This procedure consists of the measurement of the bulk susceptibility and anisotropy of the empty holder in the field 200 A m PEAK value In case of frequency Fl the susceptibility and anisotropy are written into the configuration file The holder correction values are subtracted from the measured values after measuring the specimen After activating this procedure through pressing the AuxKey F4
56. sured specimen In order to perform susceptibility measurement at a chosen temperature range the equipment moves the furnace automatically into and out of the pick up coil of the Kappabridge The quasi continuous measurement process is fully automated being controlled by the software The CSL Low Temperature Cryostat Apparatus is used for measurement of the temperature variation of low field magnetic susceptibility of minerals rocks and synthetic materials in the temperature range from minus 192 C to ambient temperature The apparatus consists of non magnetic cryostat with a special platinum thermosensor CS4 temperature control unit and laboratory power supply EA PS The specimen is placed in a measuring vessel which is cooled inside the cryostat by liquid nitrogen and then heated spontaneously to a given temperature The argon gas is needed for deplenishing the liquid nitrogen out of cryostat Temperature is measured by platinum thermosensor The quasi continuous measurement process after cooling the specimen is fully automated being controlled by the software CS4 CSL Specifications Maximum specimen volume fragments or powder 0 25 cm Inner diameter of measuring vessel 6 5 mm Sensitivity to susceptibility changes 976 Hz 400 Am 1x IO SI Temperature range CS4 ambient temperature to 700 C Temperature range CSL 192 C to ambient temperature Accuracy of temperature sensor 2 C see also IEC 751 Pt100 Argon gas flow requiremen
57. t protect atmosphere approx 100 ml min Amount of liquid nitrogen cooling cryostat approx 0 25 for one cooling Power requirements 240 230 120 100 V 10 50 60 Hz Power consumption 350 VA Dimensions Mass Temperature control unit 230mmx190mmx130mm 1 7kg Laboratory power supply EA PS 310 mmx 240 mmx 130mm 8kg Water container with Pump 380 mm x 380 mmx 700 mm 2 kg without water Argon flow meter 32 mmx 32 mmx 140 mm lkg Furnace Cryostat diameter 60 mm length 220 mm 0 5 kg 14 EC Declaration of Conformity We AGICO s r o Je n 29a CZ 621 00 Brno ICO 607 313 54 declare that the product Name Modular system for measuring magnetic susceptibility anisotropy of magnetic susceptibility and temperature variation of magnetic susceptibility Modulus Type MFKI indicator of susceptibility and its anisotropy at variable fields Modifications MFKI FA indicator of susceptibility and anisotropy of susceptibility at 3 frequencies with rotating sample MFKI A indicator of susceptibility and anisotropy of susceptibility with rotating sample MFKI FB indicator of susceptibility and anisotropy of susceptibility at 3 frequencies MFKI B indicator of susceptibility and anisotropy of susceptibility Modulus Type CS4 indicator of temperature variation of susceptibility from room temperature to 700 C Modulus Type CSL indicator of temperature variation of susceptibility from 192 C to room temperature Manufa
58. tations of principal susceptibilities in decreasing succession as declination D and inclination D in various coordinate systems values of the normed susceptibility tensor in the appropriate coordinate system the upper line gives the diagonal tensor elements consecutively K11 K22 K33 while the lower line gives the non diagonal elements K12 K23 K13 30 PYR B ANISOTROPY OF SUSCEPTIBILITY Program SAFYR ver 1 7 Azi 9 QS PL uo oua 90 6 0 Nom vol 10 00 Dip 20 Demag fac YES Holder 1 69E 06 Act vol 10 00 TI F1 L1 T2 F2 L2 CD 10 20 30 40 0 0 00 00 00 00 Directional Total Susc Residuals 2322 03 72 45E 04 81 87E 04 0 05 0 18 0 12 10 275 03 88 80E 04 90 29E 04 002 O Sl 0 46 11 22E 03 10 36F 03 59 48E 04 0 69 0 60 0 71 35E 03 71 86E 04 81 93E 04 0 35 0 82 0 17 10 31E 03 88 81E 04 89 72E 04 0 42 0 49 0 17 Field Mean Standard Tests for anisotropy A m susc err F F12 F23 200 F1 9 186E 03 0 586 2593 2 262 2 3625 0 Normed principal 95 confidence angles susceptibilities F12 F23 F13 142515 1 1222 0 6203 3 6 128 0 8 470 0037 0 0037 0 0037 Anisotropy factors principal values positive L F P uP T U Q E 1 121 1 809 2 027 2 136 0 678 05515 0237 1 614 Principal directions Normed tensor Specimen D 334 67 245 1 2327 1 1256 0 6417 system I 0 12 78 0 0571 0 0911 0 0444 Geograph D 314 1 173 0 7308 1 1342 12
59. tial number of the factor and the name abbreviation of the factor delimited by comma This is repeated till the whole set is entered After entering the last factor the program displays again the whole set and asks Any changes Y N In the case you need to do any change you have to repeat the whole procedure AuxKey 2 Vol Set Volume Mode Set the Volume mode for bulk susceptibility measurement In case the volume mode is active the pre set actual volume of specimen is displayed and Bulk susceptibility is calculated during susceptibility measurement AuxKey 2 Mass Set Mass Mode Set the Mass mode for mass susceptibility measurement In case the mass mode is active the pre set actual mass of specimen is displayed and Mass susceptibility is calculated during susceptibility measurement This option is not available for AMS measurement AuxKey 2 Sigma Standard Error Executes the test measurements with standard error evaluation The data are stored in the files which names are derived from current time in format HHMMSS with extensions K00 and ROO The file K00 contains the all measured data file ROO contains only the average and standard error of 10 repeated measurements in one set Number of sets is also 10 The measurements takes approx 40 min It is used for testing the instrument sensitivity and or the magnetic environment in the room where the Kappabridge is installed The measurement is performed at frequency Fl and the field 400 A m peak
60. tion of the individual curves from the average curve 7 Error gives this deviation divided by the amplitude value The Error you obtain in each of three AMS axes measurement is standard deviation of the individual curves there are two sine wave curves for one physical revolution from the average curve and the Error gives this deviation divided by the amplitude value This error has only informative meaning and reflects the ratio between the noise and anisotropy signal for measurement in one plane only Thus it depends not only on absolute susceptibility of the specimen measured but mainly on the degree of anisotropy in an individual plane perpendicular to the axis of rotation In case there is no anisotropy in one of the three planes this error may be over 100 and has no physical meaning In case the anisotropy in one plane has reasonable value the usual value is lower 5 but it does not reflect the quality of the measurement but the level of anisotropy in one 34 plane On the other hand it is clear that the sensitivity of the instrument influences this error For judgement of the quality of AMS measurement use F test numbers and 95 confidence angles The general rule is as follows If the F numbers are high let say at least above 5 the confidence angles are low and principal direction directions is are very well defined The sensitivity of AMS measurement for field 400 Am on MFK1 is 2x10 the anisotropy of the specimens
61. ts consecutively K11 K22 K33 while the lower line gives the non diagonal elements K12 K23 K13 9 4 1 KkKKKK Azi 30 Dip 60 T CD Field A m 420 F1 1 0304 0 0003 1 032 Specimen system Geograph system Paleo 1 system to 1 tem Tec sys Paleo 2 system to 2 tem Tec sys 11 06 2007 ANISOTROPY OF SUSCEPTIBILITY 38 Program SAFYR ver 1 0 O P 12 0 3 90 Nom Vol 10 00 Demag fac Yes Holder 1 67E 06 Act vol Al O 0 F1 L1 T2 F2 L2 100 20 30 40 SO 140 60 70 80 Mean Standard Tests for anisotropy susc Err F F12 F23 127 9E 06 0 042 29592 2055 3 1564 5 Normed principal 95 confidence angles susceptibilities Axl Ax2 Ax3 0 9985 0 9711 1 6 1 9 0 9 0 0003 0 0003 0 9 1 6 1 9 Anisotropy factors principal values positive F P HP T U Q E 1 028 1 0611 061 0 063 0 078 0 738 0 996 Principal directions Normed tensor D 283 1 93 68 1 0000 1 028 DELS I 4 3 85 0 0069 0 0046 0 0004 D 40 146 305 1 0095 0 9973 0 9932 I 9 60 28 0 0254 0 0124 0 0028 D 34 152 284 0153 0 9890 0 9957 dE 26 44 35 0 0162 0 0194 0 0074 D 94 212 344 0 9815 1 0228 0 9957 I 26 44 35 0 0033 0 0161 0 0131 D 229 42 133 0 9878 0 9868 1 0254 I 67 23 2 0 0160 0 0095 0 0068 D 249 62 153 0 9774 0 9972 1 0254 I 67 23 2 0 0126 0 0112 0 0031 The data page can be left by pressing ESC key 39 Function Key 6 ActVol or Mass
62. ue is proportional to the mechanical resistance belt strain and should be in the range 1550 to 1700 If necessary adjust the belt strain than disable and enable the rotator by commands lt T gt and lt E gt to update the belt strain value Switch MFK off and finally set the motor cover and fix it by screws install the perspex ring and check the proper position of trapezoid rubber washer AUX 200 fi n xx END OF ZEROING Enter Command 4 up Exec time 3 35 s FC down Exec time 3 24 s lt i init rotator lt B gt U D disaBle U D is Enabled D RBolator Disable ROTATOR is Enabled 1675 2 zeroing Exec tine 2 26 s lt C gt Calibration Standard lt 0 gt Orientation Paraneters lt A gt Anisotropy Factors 4 Mass mode VOLUME is ON Vol 10 00 ccm 4 Signa Test lt L List of Parameters lt Esc gt Escape 62 Cleaning the Up Down Mechanism The Up Down Mechanism should be cleaned in case the mechanism makes suspicious noise during movement and or if the time of the movement of the holder from lower to upper position is longer than 3 6s and or if the massive linear white contamination trace from teflon is present on the inner surface of the cylinder The Up Down execution time can be checked using AuxKey 2 in Auxiliary Menu of SAFYR program Set the Up Down Mechanism to down position AuxKey 2 lt down gt and switch off the instrument Last Revison 02 Mar 09 After loosening the sma
63. with mean susceptibility about 5x10 SI units can be measured but the confidence angles may be in some cases higher it depends on type of anisotropy The sensitivity is approximately linearly decreasing with decreasing field Due to the influence of rotator motor the AMS measurement may be problematic at frequencies F2 and F3 in case of specimens weaker than 100x10 SI units and with degree of anisotropy lower than 596 For this case at F3 it is recommended to use manual measurement method in 15 directions to eliminate the influence of motor of the rotator Function Key 2 Ax2 This procedure serves for the measurement of the AMS in the x x3 plane the specimen spins about the x2 axis Position No 2 in the same way as in the previous case Function Key 3 Ax3 This procedure serves for the measurement of the AMS in the x x2 plane the specimen spins about the x3 axis Position No 3 in the same way as in the previous case Function Key 4 Tsus3 This procedure measures the bulk susceptibility along the x axis corresponding to the specimen in the third measurement position After pressing F4 the bridge is zeroed the specimen is inserted into the measuring coil and the bulk susceptibility is measured The knowledge of the bulk susceptibility along the x axis is necessary in the construction of the complete susceptibility tensor from the deviatoric tensor based on susceptibility differences and one bulk value Function Key 5 Fiel
64. zero and the last two angles are also zeros If 2 is input the following data must also be input 2 Code 4 tectonic angles The two character code characterizes the measured mesoscopic foliation and lineation the angles are azimuth of the dip or strike if the orientation parameter P4 is 90 dip of the second mesoscopic foliation trend plunge of the second mesoscopic lineation respectively If only foliation exists the second character of the code must be zero and the last two angles are also zeros After the geological data are input the program displays the results and after pressing ESC key the data can be saved in the data file They are written on the disk as an ASCII file in the same format as they appear on the screen later they can be re printed on the paper if necessary The extension of this file is ASC and the file is located in the same directory as the standard AMS file After measuring the second or later specimen only the question for the specimen name appears on the screen The data are handled in the same way as those of the first specimen If one wishes to change the file one inputs instead of the specimen name and the inputting is made as in the first specimen Then the calculated data are shown on the screen in the form whose example is shown on the page 30 The meaning of the presented results is as follows Azi first orientation angle mostly azimuth of the dip or strike of the fiducial mark on the specime

Download Pdf Manuals

image

Related Search

Related Contents

  テレビ受信用機器 〈屋外用混合器〉  Kenwood KRC-PS979R Car Stereo System User Manual  Nivona CafeRomatica 855  Life Fitness FZHAD User's Manual  DECEMBER 1999 Issue - James Madison University  Manuel d`utilisation et d`entretien  Transcend 512MB DDR266 ECC Registered Memory  Land Pride 322-215M Spreader User Manual  Kenwood DV-402 DVD Player User Manual  

Copyright © All rights reserved.
Failed to retrieve file