Easy Air 15 PM15P user manual

image

Contents

1. 15 Models 15 1 5 shown Outlet OF Air Only SAVE THESE INSTRUCTIONS ACAUTION Federal USA law restricts this device to sale by or on the order of a physician PRECISION MEDICAL 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 www precisionmedical com 5 RECEIVING 2 INTENDED USE che 2 READ ALL INSTRUCTIONS BEFORE USING 2 EXPLANATION OF 2 SAFETY INFORMATION WARNINGS AND CAUTIONS 2 SPEGIBIGATIONS a 4 COMPONENT IDENTIFICATION 5 GROUNDING 6 OPERATING 6 CLEANING MAINTENANCE _ 7 FYE DURING es staan A E 8 REPLACEMENT PARTS eeni 8 9 LIMITED WARRANTY 10 EASY AIR 15 RECEIVING INSPECTION Remove the Precision Medical Inc Easy Air Compressor PM15F or PM15P from the packaging and inspect for damage If there is any damage DO NOT USE and contact your Provider INTENDED USE The basic function is to apply a portable compressed air s
2. Outlet Regulator Knob Gauge Air Inlet Filters Power Cord lt Power Cord Fuse Drawer AC Inlet Hour Meter GROUNDING INSTRUCTIONS The Easy Air Compressor should be grounded In the event of an electrical short circuit grounding reduces the risk of electrical shock by providing an escape wire for the electrical current This Compressor is equipped with having a grounding wire with a grounding plug The plug must be plugged into an outlet that is properly installed and grounded ADANGER Improper use of the grounding plug can result in a risk of electrical shock AWARNING Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked Hospital Grade OPERATING INSTRUCTIONS CAUTION Inspect the Easy Air Compressor for visible damage DO NOT USE if damaged 1 Become familiar with the Compressor see Component Identification 2 Position the Compressor near a properly grounded electrical receptacle Allow a minimum of 4 in 10 cm clearance on all sides to ensure adequate air flow for cooling of the Compressor 3 Be sure the On Off I O Switch is in the OFF O position Be sure the Power Cord is properly connected 5 Attach appropriate device to the Outlet Fitting securely without overtightening 6 Pull Regulator Knob OUT to unlock regulator then turn Knob fully counterclockwise Failure to do so may cause t
3. the Compressor immediately after using DO NOT use while bathing DO NOT place or store the Compressor where it can fall or be pulled into a tub or sink DO NOT place or drop into water or other liquid DO NOT reach for the Compressor that has fallen into water Unplug immediately DO NOT attempt to repair the Easy Air Compressor Any attempt could result in electrical shock AWARNING TO REDUCE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS e Close supervision is necessary when this Compressor is used by on or near children or invalids e Use this Easy Air Compressor only for its Intended Use as described in this manual e DO NOT use attachments not recommended by Precision Medical Inc DO NOT use the Compressor with any equipment that requires dry air including ventilators gas blenders or flow generators without a device to properly dehumidify the discharged air of the PM15 DO NOT use if the Compressor has a damaged cord or plug DO NOT use if the Compressor has been dropped or damaged Keep the cord away from heated surfaces DO NOT use when sleepy or drowsy DO NOT drop or insert any object into any opening or hole DO cover air vents Restriction of air flow will cause unit to overheat causing the thermal reset to shut down the Compressor EASY AIR 15 SPECIFICATIONS Dimensions Length 16 in 40 6 cm Width 7 25in
4. 18 4 Height 10 25 26 0 Weight 20 4 Ibs 9 3 kg Shipping Weight 24 Ibs 10 9 kg Electrical PM15F PM15P 108 132 VAC 60 Hz 5 Amps Outlet Grounded 9 16 18 Male Threaded Fitting Operating Range PM15P Output PM15F Output 1 Ipm Models Equipped with a Pressure Gage Models Equipped with a Flow Gage minimum 15 lpm lpm as set and read on the gauge Maximum Rated Sustained Pressure 60 psi Operating Temperature Range 50 F to 104 F 10 C to 40 C Storage Temperature Range 4 F to 104 F 20 C to 40 C Storage Operating Humidity Maximum 95 Noncondensing Safety Systems High Temperature Motor Shutdown Thermal Protector Class B Thermal Limit Automatic Reset Equipment Classification Class with respect to protection from electrical shock Type B degree of protection against electrical shock IPXO degree of protection against ingress of liquids Mode of Operation continuous The Compressor meets the EMC requirements of EN 550011 1998 Group 1 Class B The Compressor meets the requirements of CGA V 5 2005 section 5 4 This Compressor is not suitable for use in the presence of a flammable anesthetic mixture with air with oxygen or with nitrous oxide 15 COMPONENT IDENTIFICATION ACAUTION Missing or illegible labels must be replaced contact Precision Medical Inc Carry Handle On Off I O Switch Air
5. TATEMENTS DO NOT CONSTITUTE WARRANTIES The representatives of Precision Medical Inc or any retailers are not authorized to make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and exclusive statement of the terms of that contract THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY WHETHER EXPRESS OR IMPLIED Precision Medical Inc shall not under any circumstances be liable for special incidental or consequential damages including but not limited to lost profits lost sales or injury to person or property Correction of non conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical Inc whether based on contract negligence strict tort or otherwise Precision Medical Inc reserves the rightto discontinue manufacture of any product or change product materials designs or specifications without notice Precision Medical Inc reserves the right to correct clerical or typographical errors without penalty EASY AIR 15 10 5 MEDICAL Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Toll Free Tel 800 272 7285 Toll Free Fax 800 353 1240 Tel 001 610 262 6090 Fax 001 610 262 6080
6. ar A 4 Blown Fuse Replace Fuse 5 Power Interruption 5 Turn off Compressor When Power restored repeat Start Up Procedure 6 Compressor does 6 BEFORE STARTING not start UNIT Pull OUT regulator Knob to unlock then turn Knob fully counterclockwise Unable to adjust 1 Locked Regulator 1 Pull Regulator Knob pressure flow Knob out Poor Ventilation 1 Allow a minimum of 4 in 10 cm clearance on all sides to ensure adequate air flow for cooling of the Overheating Compressor 2 Dirty Air Filters 2 Clean Air Filters Low pressure 1 Loose connection at 1 Tighten connection flow outlet fitting 2 Improper regulator 2 Re adjust regulator adjustment EASY AIR 15 LIMITATION OF LIABILITY Precision Medical Inc warrants that EASY AIR 15 Compressor the Product will be free of defects in workmanship and or material for the following period One 1 year from shipment Should any failure to conform to this warranty appear within the applicable period Precision Medical Inc shall upon written notification thereof and substantiation that the goods have been stored installed maintained and operated in accordance with Precision Medical Inc s instructions and standard industry practice and that no modifications substitutions or alterations have been made to the goods correct such defect by suitable repair or replacement at its own expense ORAL S
7. he device not to start 7 Press the On Off Switch to the ON I position 8 Adjust pressure flow To INCREASE pressure flow Turn Knob CLOCKWISE To DECREASE pressure flow Turn Knob COUNTERCLOCKWISE 9 Push Regulator Knob IN to lock pressure flow setting 10 Compressor is now ready to use 11 When the compressor is not in use press the On Off I O Switch to the OFF position EASY AIR 15 CLEANING MAINTENANCE AWARNING e Turn the Easy Air Compressor off and disconnect from any external power source before cleaning DO NOT attempt to repair There no serviceable parts inside To clean the filter weekly 1 Unplug Compressor 2 Check air inlet filters for dust buildup 3 If dust buildup occurs remove air inlet filters and wash in warm soapy water and rinse 4 Absorb excess water from filter with a dry towel 5 Return the air inlet filters Cleaning for Exterior Components Use a clean cloth or sponge dampened with soap and water Wipe dry with a clean cloth or paper towel CAUTION e Excessive dust buildup on filter will reduce performance of Compressor If this occurs clean or replace with a new filter e The air filter does not filter out saline Treatments using nebulized saline will draw the saline mist back into the Compressor Saline causes rapid deterioration of the compressor seals and will void the warranty A
8. ource adequate to drive the high volume nebulizers READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the Easy Air Compressor PM15F or PM15P This is provided for your safety and to prevent damage to the Easy Air Compressor If you do not understand this manual DO NOT USE the Easy Air Compressor and contact your Provider EXPLANATION OF ABBREVIATIONS psi Pounds Per Square Inch Ipm Liters Per Minute SAFETY INFORMATION WARNINGS AND CAUTIONS Indicates an imminently hazardous situation which if not avoided will result in death or serious injury AWARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injury ACAUTION Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury CAUTION Used without the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage CONSULT ACCOMPANYING DOCUMENTS Type B Equipment Protective Earth Ground Power On Power Off EASY AIR 15 ADANGER e Caution must be exercised when using this Compressor where oxygen is being administered DO NOT use near any type of flame or flammable explosive substances DO NOT allow the Compressor to exceed Operating or Storage Temperature specifications TO REDUCE THE RISK OF ELECTROCUTION e Always unplug
9. t 8500 Hours Return the Easy Air Compressor to your Provider for Compressor Maintenance including rebuilding pump Contact Precision Medical Inc for Preventative Maintenance Services EASY AIR 15 RETURNS Returned products require a Return Goods Authorization RGA number contact Precision Medical Inc All returns must be packaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in transit Refer to Precision Medical Inc Return Policy available on the Internet www precisionmedical com NOTE When returning this Easy Air Compressor disconnect the power cord from the back of the unit DISPOSAL INSTRUCTIONS This device and its packaging contain no hazardous materials No special precautions need to be taken when disposing the device and or its packaging Please Recycle LY REPLACEMENT PARTS PART DESCRIPTION 503237 Large Filter 502391 Small Filter 15 TROUBLESHOOTING If the Easy Air Compressor fails to function consult the Troubleshooting Guide below If problem cannot be solved consult your Equipment Provider Problem Probable Cause Remedy Compressor 1 On Off Switch in 1 Push On Off Switch to does not operate OFF position ON position 2 Power Cord not 2 Check Power Cord connected properly connections 3 Compressor 3 Shut Compressor Off overheated Wait 15 minutes Ensure vents are cle
10. www precisionmedical com 502597 Rev 9 07 16 09 PRINTED IN USA