CT-1000 Long Web User Manual - HKA Technologies leak test

image

Contents

1. Technologies Inc Mfrs of leak test instruments Full turnkey solutions CT 1000 Leak Test System Installation and Operation Manual Table of Contents Page GENERAL OVERVIEW OF LEAK TESTING 2 CT 1000 LEAK TEST SYSTEM 6 OPERATION OVERVIEW 11 SETUP MODE 13 CALIBRATION MODE 18 TEST MODE 22 CHAINED SEQUENCE TESTING 26 RS232 DATA OUTPUT 27 PNEUMATIC CONNECTIONS 30 ELECTRICAL CONNECTIONS 32 SPECIFICATIONS 32 TROUBLESHOOTING 36 WARRANTY 37 CONVERSION CHARTS 38 SYSTEM SETUP LOG SHEET 39 HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page2 GENERAL OVERVIEW OF LEAK TESTING Leak testing systems can be divided into three widely used basic methods The simplest method is water bath testing or bubble testing where the part to be tested is filled with air at the required pressure and then immersed in water An operator visually inspects for bubbles appearing from any leak This method is relatively inexpensive and can detect small leaks less than 1 0 sccm The advantage is that one can often pinpoint where the leak is on a given part The disadvantage of this system is its high human error factor due reliance on the operator to detect leaks the inability of the system to quantify the leak rate and the need to dry most parts after testing There are some parts that cannot be allowed to get wet internally when leak tested The most expensive method helium leak testing is used to detect very small leaks le
2. seconds to 999 seconds while the 0 1 SEC time base allows time parameters to be varied from 0 5 seconds to 99 9 seconds The bottom line of the LCD will now show FillTime followed by a numeric value Use the Up or Down keys to select the desired time for this state Note the range of values for FillTime is 0 1 to 25 5 seconds changeable in 0 1 second increments When the correct value is displayed push the ENTER key again The bottom line of the LCD will now show StabTime followed by a numeric value This is the time for the stabilization state and it also can be varied from 0 1 to 25 5 seconds in 0 1 second increments When correct value is displayed push the ENTER key again The bottom line of the LCD will now show ReadTime followed by a numeric value This is the time for the leak reading to reach its final value or to reach an acceptable value approaching the final leak value As for the previous parameters the range is 0 1 to 25 5 seconds with 0 1 second increments With the correct value displayed push the ENTER key again The bottom line of the LCD will now show Leak Lim followed by a numeric value This value corresponds to the lowest leak reading that will be termed a failure by the leak test system With the leak readout rounded off to 0 1 sccm a value of 0 05 sccm lower than the leak limit would be shown as a failure For example if the leak limit was set to 2 0 sccm a leak reading of 1 95 sccm would be termed as a fail
3. CT 1000 test system The horizontal axis represents seconds of test time The vertical axis represents the leak rate in calibrated units of sccm The scaling for both axes is taken from the part set up information sent to the PC through the RS232 port on the leak test unit The lower curve is taken with a no leak part on the test port The upper curve has a 1 sccm calibrated leak added to the no leak test part In the set up for the part being tested the fill time is 5 0 seconds the stabilization time is 5 5 seconds and the read time is 10 seconds HKA PART TEST 06 24 1999 08 43 41 PRESS 55 PSI LEAK 1 0 SCCM HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page9 Figure 2 shows the valve states and air flow during the fill state STAB 1 FILTER PRESSURE AIR FILTER REGULATOR FILL PPLY REFERENCE PART CALIBRATED LEAK STAB 2 EST PART an ne FILTER REG SUPPLY REFERENCE PART CALIBRATED LEAK STAB 2 TEST PART OPEN DRAIN Figure 4 shows the conditions for the read state Note that the curve during the read state exponentially approaches the final value of either for the no leak case or 1 0 for the 1 sccm leak case As can be seen from the graph the difference between no leak and a 1sccm leak is clearly evident after 3 seconds of the read state but to get a true sccm value for the final leak a time of 10 seconds is required HKA Technologies Inc Jun
4. CT 1000 leak test system is guaranteed to be free from defects in materials and workmanship for a period of one year from date of purchase The warranty does not cover units that have been modified abused or improperly maintained or operated During the warranty period HKA will repair or replace at its option any unit that has been found to be defective Any product to be returned under this warranty must have an authorization number from HKA and must be shipped prepaid to HKA in Burlington Ontario or the nearest authorized representative HKA s liability for the leak test system is limited solely to the replacement or repair of the product and HKA will not be liable for any consequential damages HKA reserves the right to make on going changes and upgrades the CT 1000 leak test unit and has no obligation to make modifications to products previously sold Any units returned to HKA after the warranty has expired will be subject to charges for labour and materials and any shipping costs incurred For warranty contact HKA Technologies Inc 3536 Mainway Burlington Ontario L7M 1A8 905 336 8668 or FAX 905 336 8805 HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 38 CONVERSION CHARTS FOR COMMON LEAK RATE UNITS The CT 1000 leak test unit uses sccm standard cubic centimeters per minute units for display and limits The following chart gives conversion factors for other commonly used units for leak r
5. are given at the end of this manual HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page6 CT 1000 LEAK TEST SYSTEM The standard configuration CT 1000 is a differential mass flow leak test system It uses a mass flow transducer to accurately measure flow from a reference part to the test part Other configurations of the CT 1000 use simple mass flow for those low pressure high flow test situations such as mufflers catalytic converters air cleaners manifolds etc The CT 1000 also incorporates a pressure transducer to enable accurate setting of the test pressure and to monitor the pressure during the test cycle The system uses advanced microprocessor circuitry to deliver superior performance at lowest cost INCREMENT PUSH BUTTON DECREMENT PUSH BUTTON PROG CAL RUN PUSH BUTTON ENTER PUSH BUTTON TESTON PRESSURE ADJUST ACCEPT _ REJECT CT 1000 LEAK TEST FRONT PANEL FUNCTIONS A large backlit LCD readout displays all readings and set up parameters Setup parameters for up to ten different parts are stored in a non volatile EEPROM which does not require battery backup All operations including set up are made through a four button keypad located on the front panel HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page7 All process valves are air operated eliminating any heating effect on leak measurement caused by power dissipation in the solenoid coils Most internal pneumati
6. setup parameters for any part are saved only under the following conditions 1 Part number changed in SETUP mode 2 After completion of test cycle in TEST mode 3 After completion of a successful CAL cycle The display will now show Pressure followed by a numeric value When this parameter is accessed the valves will change state to allow the pressure transducer to read the test pressure The system must be connected to an air supply to enable the pressure parameter to be set To set pressure adjust the regulator knob on the front of the test until the correct reading is displayed Note the regulator has a one to two second response time after adjustment and that time should be allowed for a final pressure to be reached Once the pressure has stabilized at the correct value push the ENTER key again Note The pressure readout is in psig units and has a resolution of one psig HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 17 The last option software version 4 20 or later is Fast Cycle followed by either Yes or No With the fast cycle enabled Yes the system will give a pass output during the read cycle as soon as the leak value is lower than the leak limit and the slope of leak vs time is zero or negative This allows a much shorter test cycle in many applications It is important to note that the short cycle ensures the part being tested is within the leak limit parameter but the fi
7. M WILL NOT START IN TEST MODE e Check connections to user I O connector 3 SYSTEM STOPS PART WAY THRU IN TEST MODE e P displayed top right of LCD e Verify air pressure to test system e Gross leak in part under test e Pressure dropped below limits during test 4 SYSTEM STOPS PART WAY THRU IN TEST MODE A displayed at top right of LCD e Abort input turned on during test 5 SYSTEM DISPLAYS ERROR CODE U AT END OF READ CYCLE e Upward positive slope above limit at end of read cycle 6 SYSTEM DISPLAYS ERROR CODE D AT END OF READ CYCLE e Downward negative slope above limit at end of read cycle 7 SYSTEM DISPLAYS ERROR CODE AT END OF READ CYCLE e Leak below zero at end of read cycle 8 SYSTEM WILL NOT ACCEPT CHANGES IN SETUP MODE e Verify correct Password entered 9 SYSTEM STOPS PART WAY THRU IN CALIBRATION MODE e Z displayed at top right of LCD e Two zero readings not within correct range or differ from each other by more than allowable amount e Leak in reference port or test port plumbing e Part temperature changing during test cycle e SETUPmed times too short for stable readings 10 SYSTEM STOPS PART WAY THRU IN CALIBRATION MODE e S displayed at top right of LCD e Two span readings not within correct range or differ from each other by more than the allowable amount e SETUPmed times too short for stable readings HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 37 WARRANTY The
8. an be changed from 0 1 CCM to to 5 0 CCM The value entered is the allowable leak below 0 Leak rates below 0 can occur with cold parts being tested relative to the reference part This parameter puts a limit on how much negative flow will be accepted and a pass output given at the end of the read cycle If the value is outside limits an will appear at the top right LCD display As was the case for slope errors a pass output is not given at the end of read if this limit was exceeded and a fail output may or may not be energized depending on specifications made at time of purchase The next parameter to be displayed is Pres Tol followed by a numeric value This is the tolerance for the test pressure at which the system will indicate a fault and abort the test sequence The value is in percent of the set test pressure and can be varied from 0 to 20 in increments of 1 With mass flow leak sensing a given leak reading will change directly as the test pressure a 196 psi change in pressure will result in a 196 change in a given leak rate The precision regulator used in the test system should be able to maintain test pressures within 2 and therefore any value of 2 or more for pressure tolerance would be acceptable Again use the Up and Down arrow keys to select the desired value and then push the ENTER key HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 16 PART NUMBER MODE Pressure 1 15 PSI The
9. and the part under test After a short delay to ensure all of the pressurized air is allowed to escape from both parts the test sequence is automatically repeated There is no need to push the start test button again HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 20 When the test sequence starts this second time the bottom LCD will show the parameter Zero 2 followed by a numeric value This value will change to 0 as soon as the test sequence begins At the end of the test cycle the will be replaced by a new value The two values Zero 1 and Zero 2 are compared and if close enough in value to each other are averaged together to establish a zero leak offset called F Zero Again after a short delay for venting the test cycle will repeat automatically This time the calibrated leak valve will be open and the bottom line of the LCD will read Cal 1 followed by a numeric value This number will change to 0 as soon as the test starts and will be replaced by a new value at the end of the test The process will be repeated for a second test where a new value for Cal 2 will be established The values of Cal 1 and Cal 2 are compared and if their difference is in limits the values are averaged and divided by the known calibrated leak to obtain a constant that is used to calculate the leak rate in sccm called F Fact The following summarizes the parameters that can be displayed when the system is i
10. ate measurement To convert to sccm from the following units multiply by the given factors c c sec c c hour litres min cu in min cu ft hour 02 Technologies Inc 60 0 0167 1000 16 39 472 0 011 air at standard atmospheric conditions June 2004 HKA CT 1000 Leak Tester Manual Page 39 CT 1000 SETUP LOG SHEET USER PART IDENTIFICATION 1 PASSWORD 2 PART NUMBER 3 TIME BASE 4 FILL TIME 5 STABILIZATION TIME 6 READ TIME 7 LEAK LIMIT 8 CALIBRATED LEAK 9 FLOWZERO 10 MAX SLOPE 11 MIN FLOW 12 PRESSURE TOLERANCE 13 TEST PRESSURE 14 FAST CYCLE HKA Technologies Inc June 2004
11. c connections are made through a single internal manifold thus eliminating a large number of pneumatic lines and connections that could be subject to potential leaks in the system LJ El 1 LI Lj E J E 00000 CT 1000 LEAK TEST REAR PANEL FUNCTIONS All external pneumatic connections are made to the manifold located at the rear of the leak test unit These include ports for the reference part and the part under test a port for the calibrated leak air supply in and exhaust ports All electrical connections are made through a removable twelve pin connector located on the rear of the unit The unit can be powered either by applying 24 volts through the twelve connector or by the optional AC adapter There is also an RS 232 port on the rear panel allowing for connection to a PC for initial set up and for data logging Two software SETUPs are currently available HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page8 The first SETUP plots the graphs of the operation of the tester during the leak test cycle This is very useful for set up and for ongoing monitoring of system performance The graphical representation of the leak test process is ideal in determining the optimum test time for the part under the given test conditions The second SETUP produces production records on a daily or other periodic basis Figure 1 illustrates two typical leak test curves as taken from the RS232 port of the
12. d even though no parameters are changed in the SETUP menu To enter the calibration function push the MODE key until CAL RDY appears on the top right of the LCD display Make sure proper air pressure is applied known no leak parts are attached to the reference and test ports and that the correct calibrated leak is in place Also ensure a jumper is connected between terminals 2 and 3 of the user I O connector or a switch between these terminals is in the closed position There are a number of parameters that may be displayed on the bottom line of the LCD when in CAL mode The parameter displayed is the same as the last time the CAL mode was accessed It does not matter which parameter is displayed to start the CAL procedure The calibration procedure automatically TESTs through four complete test cycles two with no leak added and two with the calibrated leak applied When completed the system will have established offset and gain constants named F Zero and F Fact Push the start test button Close contacts between terminals 2 and 11 of the user I O connector The bottom line of the LCD display will now read Zero 1 followed by a numeric value This value will change to 0 as soon as the test sequence begins The system will now go through a complete test cycle and when complete will display a new value in place of the 0 reading The system will also go through a drain cycle discharging the air from both the reference part
13. e 2004 HKA CT 1000 Leak Tester Manual Page 10 PRESSURE AIR FILTER REGULATOR SUPPLY FILTER REFERENCE PART CALIBRATED LEAK EST PART HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 11 OPERATION OVERVIEW The CT 1000 system can be SETUP med to test up to ten different parts as long as the test pressures are all within the range of the leak test unit The calibrated leak used with the system must be in the correct order of magnitude for the parts family 2 LINE POWER SWITCH LCD DISPLAY INCREMENT PUSH BUTTON CT 1000 LEAK TEST SYSTEM DECREMENT PUSH BUTTON PROG CAL RUN PUSH BUTTON ENTER PUSH BUTTON test on PRESSURE ADJUST ACCEPT REJECT CT 1000 LEAK TEST FRONT PANEL FUNCTIONS The selected part number is displayed at the left of the top line of the LCD display The test unit can be in one of three conditions as set by the Mode button the button to the right of the down arrow SETUP Mode TEST Mode or Calibrate Mode When power is removed and reapplied to the unit it will always start in the same mode as it was in when power was turned off Repeatedly pushing the mode button will scroll from TEST to SETUP to Calibrate and back to TEST The selected Mode is indicated on the top line of the LCD display In TEST or Calibrate Mode the left top of the LCD display is used to indicate any fault conditions or if there are no faults present RDY will be di
14. eak test unit the calibrated leak will be factory installed Once set up the calibrated leak is not required to TEST the test system but it is required to perform periodic tests on the performance of the leak test system It is recommended that a system performance check be done at the start of each shift of operation The calibrated leak can be enabled by removing the jumper between terminals 2 and 3 on the user connector More information on this function will be detailed in the section entitled Electrical Connections This port is used to exhaust the air from the reference part the test part and the internal valves in the test system Normally this port is left open but in very clean or quiet environments the user may wish to connect tubing to this port and lead the exhaust air to a different area HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 32 ELECTRICAL CONNECTIONS The user connector has screw terminals and is removable for easy wiring The terminal designations are from 1 to 12 starting at the top of the connector This connector is polarized and cannot be inserted in any but the correct position All interfacing to the test unit is done through this connector The CT 1000 leak test system may be powered through the 12 pin user connector with 24 volts DC or using the AC adapter plug which delivers 24 volts The current draw for both is 500 mA If powered with 24 DC the positive power input is termi
15. for changes in ambient temperature and long term drift of electronic components The Cal option is used when short cycle times do not allow sufficient time to obtain a stable flow reading and this option is also used for all applications of simple mass flow as compared to the standard differential mass flow The next parameter to be displayed is MaxSlope With mass flow technology the leak reading does not vary with time after transient conditions are allowed to settle out The value displayed can vary from 0 1 to 10 0 CCM The value corresponds to the allowable change in leak during the last one third of the READ cycle The value is calibrated in SCCM per minute A low value such as 0 1 CCM will give a test error as indicated by an U or D at the bp right of the LCD display if there is a very small slope in the leak reading during the last one third of read time Conversely a high value such as 10 0 will allow a very large change in the Leak value during the last one third of the read time without indicating any error condition The U indicates an upward slope to the leak readings while an D indicates a downward slope to the leak readings If either an U or D appears the pass output is not turned on at the end of the Read time Depending on customer requirements at the time of order the fail output may or may not be energized in the event of a slope error as indicated by U or D The next displayed parameter is Min Flow This value c
16. he two parts being tested were separate Each will only get updated when its part number is selected in TEST mode HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 26 RS232 DATA OUTPUT Ver 3 90 or lower When in TEST mode the test system continuously outputs a data string to the RS232 port The information sent via the RS232 port consists of the following ASCII data HXMTCPPPPFFFFSSSSQQQE where H SOH X STX M Mode P SETUP R TEST C Calibrate T Test Status F Fill S Stabilization R Read Ready Completion Status N Normal Aborted P Pressure fault Z Zero cal error S Span cal error P Pressure value 4 digits F Flow rate 4 digits S Flow scale factor Q Sum of Fill Stab Read times EOT Technologies June 2004 HKA CT 1000 Leak Tester Manual Page 27 RS232 DATA OUTPUT Ver 4 20 or higher ABCCCDDEFGHHHIIIJJJJKKKKLLLLMNNNOOOPPPQQQRRSSSTTTUUUUVVVVW XXXXYYYZZA where A SOH HKA Technologies Inc B STX Version Major Minor Maintenance D Part Number 1 10 E Mode P SETUP R TEST C Calibrate F Test Status F Fill S Stabilization R Read Y Ready G Completion Status N Normal Aborted P Pressure fault Z Zero cal error S Span cal error D Neg slope error during Read U Pos slope error during Read Negative flow error during Read X Reserved for future use Y Reser
17. ible seal under some form of clamping pressure In setting up times for the leak test cycle care has to be taken to ensure flexible seals are still not creeping under pressure when the leak test measurement is made The creep in the seals can effectively reduce the volume of the part under test while maintaining the same mass of gas thereby increasing the pressure in the test part Larger volumes in the test part reflect in longer test times given other test conditions are identical It is therefore prudent to minimize the volume of the plumbing between the test part and leak test system as this volume is in effect added to the test part volume All leak test systems require calibration for initial setup A means for verifying calibration on a routine basis is also required The most common method of calibration employs a known no leak part and a known calibrated leak orifice For the calibration process a known no leak part is tested with also a no leak reference part and the system should read 0 leak Then the known calibrated leak is added to the known no leak test part and the system should read the value of the calibrated leak Leak rates are usually given in sccm units which corresponds to a flow of standard normal atmospheric pressure cubic centimeters of air or other test gas per minute HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page5 Other units of volume and time are also used and conversion rates
18. ications Air Requirements Air Connections HKA Technologies Inc Mass Flow Transducer 10 Non Volatile EEPROM 2 Line x 16 Characters High Contrast Backlit LCD Standard 0 125 psig 9 0 bar g Other ranges available 0 25 5 sccm Using 0 1 sccm resolution Fill Time 0 5 25 5 sec 0 5 999 sec Stabilize Time 0 5 25 5 sec 0 5 999 sec Read Time 0 5 25 5 sec 0 5 999 sec 1 version 4 20 or higher Total cycles 0 65 536 Passed Test 0 65 536 Failed Test 0 65 536 Start Test Abort Test TEST Next Part Chained Test in Progress Passed Test Failed Test 12 pin removable connector supplied with unit RS232 Serial Port 9 pin subminiature D conn 60 150 psig 4 10 bar g Clean Dry Air 1 8 NPT Female Ports June 2004 HKA CT 1000 Leak Tester Manual Power Requirements Dimensions Weight PassWord Calibrated Leak HKA Technologies Inc Page 35 24 VDC at 300 mA 24 VAC at 300 mA optional adapter 12 wide 6 high 8 deep 533 mm wide 152 mm high 203 mm deep Approx 6 4 lbs 2 9 kg 22 factory Default or Required for set up Specify leak rate and pressure option with test unit June 2004 HKA CT 1000 Leak Tester Manual Page 36 TROUBLESHOOTING 1 LCD DISPLAY BLANK e Make sure power rocker switch in on position e Verify 24 volt supply to unit e Make sure no shorts in field wiring to user I O connector 2 SYSTE
19. ity none Data 8 bits Stop Bits 1 HS request to send suppressed CS CS 0 DS DS 0 CD CD 0 The RS232 uses a standard 9 pin sub miniature female D connector with the following pin assignments Pin 1 nc Pin 2 data out Pin 3 data in Pin 4 joined to 6 Pin 5 common Pin 6 joined to 4 Pin 7 joined to 8 Pin 8 joined to 7 HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 29 Pin 9 nc HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 30 PNEUMATIC CONNECTIONS All pneumatic connections are made through a manifold located at the back of the test System H L H L H TN A NN 20000 DT D D D UN D D CT 1000 LEAK TEST REAR PANEL FUNCTIONS All connections have 1 8 female threads From left to right facing the rear of the test unit the ports have the following functions TEST PORT This port is connected to the part being leak tested The fittings used should provide a reliable leak free seal from the port to the part under test The tubing should be kept as short as possible to reduce effective added volume of the test part AIR SUPPLY The user must provide clean dry air to the leak test system Air should be filtered with a 50 micron maximum element and depending on the level of moisture present in the air may require the use of a desiccant type dryer While any significant leaks on this port would waste air there wou
20. key or the Down arrow key to change the display to Yes and then push the ENTER key When in TEST Mode the bottom line of the LCD can be left displaying any of the parameters listed for TEST mode but the most common parameter to be displayed would be Leak indicating the measured leak rate in sccm of the last part tested PARTNUMBER _ VALUE PARAMETER The test is started in TEST mode by momentarily connecting terminals 2 and 11 on the I O connector Once the test is in progress it will complete the entire sequence unless the Abort input is made by momentarily connecting terminals 2 and 9 on the connector The top line of the LCD display will show the part number on the left side of the display PN xx followed by TEST followed by RDY on the left side HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 24 When the test has been started the RDY will change to FILL indicating the sequence is at the Fill state When the Fill state is finished the FILL will change to STAB indicating the Stabilization state At the end of the stabilization state the display will read READ until the test cycle is finished When the Read cycle is complete the top right display will again show RDY indicating the system is ready for the next test cycle The test system will also automatically exhaust pressure from both the reference part and test part at the end of the Read state Duri
21. ld be no effect on the accuracy of the leak test measurement HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 31 REF PORT INT REF CAL EXH PORT This port should be connected to a known no leak part identical to the part under test Other reference volumes may be used which are similar in size to the parts being tested but they may add to the test times to obtain the same accuracy as an identical part All connections must be leak free and again care should be taken to minimize any added volume from tubing length As an option the CT 1000 leak test system may have a fixed volume reference built into the unit While not as optimum as an identical reference part the fixed volume may be used where the same level of accuracy in the same test time is not required or where there are several different parts of similar volumes to be tested and changing reference parts for each different part may not be warranted If an internal reference is used the test system will be shipped with connectors and tubing joining the INT REF port to the REF PORT In many CT 1000 leak test units this port is marked AUX PORT Please consult the documentation shipped with the particular CT 1000 unit to determine the function if any of this pneumatic connection In order to set up the system to test any part an appropriate calibrated leak must be connected to this port again using quality no leak fittings If ordered with the l
22. leak parts are required one for the reference port and one for the test port The calibrated leak must be attached to the leak test unit and the correct leak rate in sccm must have been entered into the SETUP for the part to be tested In the calibration process the test system first goes through two complete test sequences with the valve closed to the calibrated leak This gives two numbers that correspond to a zero leak reading If these two numbers are the same or within a given tolerance of each other the two numbers are averaged to obtain a zero leak reference After the zero leak reference is obtained the system again goes through two complete test sequences but this time with the valve open to the calibrated leak HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 19 The system now has two new readings which correspond to the known leak rate Again the two numbers are compared to each other and if within a given tolerance they are averaged together and used as the factor to calculate the leak rate It is important to note that the calibration test sequence is identical to the TEST test sequence in terms of test pressure and times for the various states in the test sequence If any of the parameters except Leak Limit or Pressure Tolerance are changed in SETUP mode the system has to be recalibrated If the parts to be tested are different from those used during calibration the system has to be recalibrate
23. n Calibration mode To view each parameter press the ENTER right key P PART NUMBER MODE VALUE PARAMETER HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 21 Note that none of the parameters can be changed by the Up and Down arrow keys all parameter values are automatically generated by the system in the calibration procedure All of these values are dimensionless 1 Zero 1 first no leak value read 2 Zero2 second no leak value read 3 Cali first reading with calibrated leak 4 Cal2 second reading with calibrated leak 5 FZero offset average of Zero 1 Zero 2 6 F Fact gain value calculated from average of Cal 1 and Cal 2 and the value of the calibrated leak If the times for the various states Fill Time Stabilization Time Read Time are not long enough for the test conditions the values of Zero 1 and Zero 2 or Cal 1 and Cal 2 may not be close enough for the system to get valid numbers for F Zero and F Fact If this is the case an asterisk followed by a letter A P S Z will appear at the top right of the LCD display If this occurs the calibration procedure will be aborted and no new values of F Zero and F Fact will be calculated For more information on calibration fault conditions please refer to the section entitled Troubleshooting HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 22 TEST MODE TEST Mode is the normal test state of the leak te
24. nal 2 and the power common is terminal 1 There are three inputs to the test system start test abort test output mode select TYPICAL USER USER CONNECTOR CONNECTIONS POWER COMMON 1 24 2a 24 VDC USER INPUT 3 3 CAL LEAK CLOSE NORMAL OPERATION CAL LEAK ENABLE CAL LEAK SOL 4 RELAY COMMON 4 RELAY COMNION 5 5 NG 6 6 TEST ON p a TEST ON 7 7 PASSED TEST sil TYPICAL k FAILED TEST 8 EALES TEST ie ABORT TEST ABORT TEST CHAINED TEST SWITCH CLOSE FOR HIGHER PART SELECT START TEST These inputs can be powered from an external 24 DC source or used with isolated contacts along with the 24 volts available on the user connector To use an external 24 volts for these inputs connect terminal 12 to the output common of the user system and connect the positive signal outputs to terminals 9 10 11 The inputs draw 10 mA current each When driven from isolated contacts join terminals 1 and 12 of the user connector HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 33 The relay contacts are connected to terminals 2 and 11 for start test 2 and 9 for abort test and terminals 2 and 10 to set the output mode With no input on terminal 10 the passed test or failed test output will be on relay contacts closed from the end of the test to the start of the next test With terminal 10 energized or joined to 2 the relay contacts for the passed test
25. nal value of the leak for the part being tested remains unknown The SETUP ming is now complete for the selected part number and the bottom line of the display will again read PassWord followed by the correct number The following lists all the parameters accessible in SETUP mode and also indicates the range of values and units for each parameter 1 PassWord 0 99 22 factory default 2 Part Number 1 10 3 Tim Base 0 1 or 1 0 SEC 4 FillTime 0 5 25 5 seconds or 5 255 seconds 0 5 99 9 seconds or 5 999 seconds 5 StabTime 0 5 25 5 seconds or 5 255 seconds 0 5 99 9 seconds or 5 999 seconds 6 ReadTime 0 5 25 5 seconds or 5 255 seconds 0 5 99 9 seconds or 5 999 seconds 7 Leak Lim 0 1 25 5 sccm 8 Cal Leak 0 5 25 5 sccm 9 FlowZero Stab or Cal 10 MaxSlope 0 1 10 0 sccm 11 Min Flow 0 1 5 0 sccm 12 Pressure Tol 0 20 13 Pressure Standard Model 0 125 psig 14 Fast Cyc Yes or No Version 4 20 firmware or later only subject to available air supply pressure HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 18 CALIBRATION MODE After all the parameters have been SETUPmed for a particular part the system has to be calibrated to read the correct leak It is accessed by pushing the Mode key until CAL appears PART NUMBER MODE To perform the calibration process in a differential mass flow system like the CT 1000 two known no
26. ng the test sequence the test pressure is monitored during the Fill state and the Stabilization state If the test pressure is outside limits as specified by the Pressure and Press Tolerance parameters the test sequence will be aborted and the top right of the LCD will display RDY P indicating a pressure fault has occurred The CT 1000 has three isolated output contacts for user interfacing These are test in progress passed test and failed test The test in progress contacts are closed during the Fill Stabilization and Read states of the test sequence Either the test passed or test failed contacts will be closed at the end of the read cycle depending on the results of the leak test These contacts will be maintained until the start of the next test cycle or until power is turned off in the test unit HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 25 CHAINED SEQUENCE TESTING For certain applications the leak test system is required to perform sequential testing on two different parts for example a valve in open positions could be Part 1 and the valve in a closed position could be Part 2 To be able to test a different part without going thru the SETUP mode there is an auxiliary input terminal 10 on the I O connector which when energized will select the next test SETUP to be TEST The following outlines considerations when using this chained test sequence 1 The two part numbers must be con
27. om the pressure source The leak rate can be calculated indirectly based on the difference in pressure between the two parts over the test period Mass flow testing is a further improvement over the pressure decay systems above Differential mass flow also uses a no leak reference part but the differential pressure transducer is replaced with a mass flow transducer In this case the leak rate can be measured directly without having to take time into consideration For many applications the use of mass flow can allow for shorter test cycles than differential pressure testing or pressure decay testing HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 3 Systems that use either differential pressure or mass flow to determine leak rates have four distinct states in the test process First the reference part and test part must both be pressurized to the required level This first state is often referred to as the Fill time In the second state referred to as Stabilization time both parts are isolated from the pressure source but air is allowed to freely flow from one part to the other If insufficient time is allowed for stabilization accurate leak readings may not be possible In the third state the free flow path between the reference part and the test part is closed and the difference in pressure between the two parts or the flow of air between the two parts is measured This state is referred to as the Measuremen
28. or failed test outputs will be on for 100 milliseconds after the completion of the test cycle There are three user outputs available on the 12 pin connector Terminal 4 is the common for all outputs that have isolated relay contacts rated at 5 amperes 250 volts Terminal 6 relay contacts are closed for the duration of the test sequence Test On output Terminal 7 is connected to the passed test relay contacts while terminal 8 is connected to the failed test relay contacts Terminal 3 is connected to the calibrated leak enable solenoid valve For normal operation in TEST mode and Calibration mode this terminal is joined to terminal 2 To enable the calibrated leak disconnect the terminal 3 from terminal 2 If desired a switch may be used between terminals 2 and 3 With the switch closed normal operation takes place With the switch open the calibrated leak is applied to the test part This provides an easy method of verifying system performance by reading the value of the calibrated leak in TEST mode Note that a known no leak test part must be used to read the correct leak rate of the calibrated leak HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 34 SPECIFICATIONS STANDARD UNIT Measurement system Number of SETUPs SETUP Storage Display Pressure Range Leak Range Test Parameters 1 1 1 Internal Counters User Inputs User Outputs User Connections Commun
29. secutive The first part number can be in the range of 1 to 9 Part 10 is not valid for the first part number of a chained test and errors will occur if this part is selected The second part will be the next higher part number from the first part 2 The second part is selected in TEST mode by turning on the auxiliary input on the I O connector 3 The auxiliary input only works in TEST mode Do not turn on the Auxiliary input until both parts have been SETUP med and calibrated Ensure the first part is selected when the system is put into TEST mode and that the auxiliary input is off Failure to follow this procedure may result in setup parameters being mixed between the two parts selected for chained operation 4 After SETUP ming the setup parameters for the second part change the part number to any other part to save the setup information This is good practice for any part being SETUPmed 5 Calibrate both parts as would be done if the parts were not to be tested in a chained sequence 6 When the auxiliary input is on in TEST mode the display will show the new part number 7 The parameter being displayed in TEST mode will be the same for both parts in the chained sequence For example if the parameter LEAK is displayed for the first part it will also be displayed for the second part The parameter to be displayed can be changed when in TEST mode when either part is being tested 8 The GOOD BAD TOTAL part counters work as if t
30. splayed indicating system is ready The bottom line of the LCD is used to display variable parameters for the test system The left half of the display identifies the parameter while the right half of the display identifies the numeric value of that parameter HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 12 Pushing the enter button right hand button will scroll thru the various parameters available in the Mode selected In SETUP Mode only the Up and Down arrow buttons can be used to change the numeric value of the parameter being displayed In order to change any value the correct PassWord must have been entered Without the correct PassWord the various parameters can only be examined not changed In TEST Mode the Up and Down arrow keys are only used to toggle between YES and NO for the parameter Clear Counters The Up and Down arrow keys have no function in Calibrate Mode The illustration shows the location of the four push buttons and LCD display used to SETUP and control the leak test unit ut PART NUMBER MODE 1 15 PSI HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 13 SETUP MODE SETUP Mode is used to select the desired part number 1 thru 10 and to input or change any of the parameters associated with that part number The new or revised parameters are saved in EEPROM any time the Mode is changed from SETUP to TEST or TEST to Calibrate To acce
31. ss than 0 1 sccm In these systems the part is filled with helium gas and placed in a container often called a bell jar or clam shell that traps any helium gas escaping through the leak A mass spectrometer is used to measure the relative amount of helium in the container surrounding the part and a leak rate can be determined These types of systems are fast and accurate but are also the most expensive with typical systems costing over 100 000 Sniffer type helium leak test systems no container hand held helium sniffer are also used but they may also suffer from the same kind of human error factors as water testing The third method of leak testing uses air as a means of leak measurement Either air pressure or air flow are measured to detect leaks The simplest method of air leak testing is pressure decay testing where the test part is pressurized to a given pressure and then isolated from the pressure source A pressure transducer is used to measure the pressure at the beginning of the test period and at the end of the test period The pressure drop over the time period is considered to be due to the leak in the part A leak rate can be calculated indirectly using this method An improvement over this method is called differential pressure testing This method uses a differential pressure transducer to measure the difference in pressure between a known no leak part and the part under test after both have been pressurized and then isolated fr
32. ss the SETUP function push the MODE key until SETUP is displayed on the top line of the LCD display In order to select a different part number or change any parameter value push the ENTER far right button until the parameter PassWord appears Use the Up and Down arrow keys to enter the two digit factory set default 22 If your test system has a different PassWord it will be shown in the specifications section of this manual When the display reads the correct PassWord push the ENTER PART NUMBER MODE Pressure 115 PSI The bottom line of the LCD display should now read Part No followed by a number from 1 to 10 Note the same number is shown at the top left of the LCD display PN followed by number 1 to 10 If the number displayed is the desired part number to HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 14 input new or revised parameters push the ENTER key again If the part number is to be changed use the Up and Down arrow keys to select the correct part number Note that the value displayed for part number will not exceed 10 When the correct part number is displayed push the ENTER button The top left of the LCD will show the new part number The bottom line of the LCD will now show Tim Base followed by either 1 0 SEC or 0 1 SEC Use the Up or Down arrow keys to select the desired time base for system parameters Note with a 1 0 SEC time base each time parameter can be varied from 5
33. st system It is accessed by pushing the Mode key until TEST appears PARTNUMBER MODE mi _3 6sccm VALUE Ec FA When in TEST Mode the top left of the LCD displays the part number while the top right displays TEST followed by RDY The bottom line of the LCD can display a number of different parameters which are scrolled through by pushing the ENTER key The following summarizes the parameters that can be displayed along with the corresponding range of values and units when in TEST Mode Leak 0 0 25 5 sccm Pres 0 0 125 psig Flow 0 0 25 5 sccm Pass Cnt 0 65 535 Fail Cnt 0 65 535 TotalCnt 0 65 535 Clr Cnts No Yes toggle with Up Down keys PACEM BUE d DC COD dw Leak is the measured leak rate in sccm units of the last part tested HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 23 Pres is the test pressure during both the Fill and Stabilize states Pressure is not read during the Read state Flow corresponds to the current reading of the mass flow transducer at any point in time during the test At the end of each test Flow and Leak have the same value The last four parameters refer to the counters in the system used to track the total number of tests the number of passed tests and the number of failed tests The Clr is an abbreviation for Clear Counters and the default when this parameter is viewed is No To clear the counters use either the Up arrow
34. t or Read time Finally there must be the capability to exhaust the pressure from the part under test The reference part may also be exhausted at this time There are two transient conditions which effect leak rate measurement in systems using air pressure The first condition is caused by adiabatic temperature changes as the air under pressure expands cools to fill the test and reference parts and then is compressed inside the parts to the test pressure heats up Depending on part volume and test pressure it could take a significant time for the temperature effects to settle out The higher the pressure and or the higher the volume the greater the impact of these temperature changes One advantage of using a no leak reference part is that it should behave the same thermally as the part under test and the adiabatic temperature effects should cancel out While the temperature effects rarely fully cancel test times are greatly reduced The second transient effect is caused by the disturbance introduced when valves are opened or closed Apart from the sudden change in flow caused by operating a valve the opening or closing action tends to pump a small volume of air in one direction or the other Slowing down valve response helps but some effects always remain Depending on the type of valve used the valves may contribute to the temperature effects also One of the largest sources of error in air pressure leak test systems is due to ex
35. ternal factors that directly influence the temperature of the part under test If the reference part could be subjected to exactly the same external conditions then this error would be cancelled but this is seldom practical in the shop environment Often parts to be tested come directly from a manufacturing process and even simple operations such as installing a screw can create enough heat from friction to heat up the part enough to create errors in leak testing measurements Since air follows the basic gas laws any change in the temperature of one of the parts will cause a change in pressure directly proportional to the temperature change of that part A small leak in a HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page4 part can be masked by the increase in pressure from external temperature increases to that part The first two transient effects can be overcome by having sufficient time in each test state to allow the system to become stable The external effects of temperature on the test part may take many minutes or more to become nonsignificant depending on part mass The total test time for any part depends on the following factors 1 Required leak test resolution and accuracy 2 Volume of test part 3 Test Pressure 4 Thermal responsiveness of the part 5 External temperature variations Pneumatic connections from the part under test to the leak test system are usually made with some form of flex
36. ure The leak limit can be varied from 0 1 to 25 5 sccm in 0 1 sccm increments using the Up and Down arrow keys When the correct leak limit is displayed push the ENTER key again The display will now show Cal Leak followed by a numeric value on the bottom line This number must correspond to the leak value in sccm of the calibrated leak attached to the test unit Note that since the leak rate of the calibrated leak will vary with pressure the value displayed for the parameter Cal Leak must be the leak value at the set test pressure The range of allowable calibrated leak values varies from 0 5 to 25 5 sccm More information on the calibrated leaks is detailed in the section Calibration Mode When the correct value for the calibrated leak is displayed push the ENTER key again HKA Technologies Inc June 2004 HKA CT 1000 Leak Tester Manual Page 15 The next selection to be made is the choice of zero to be used during the test cycle The displayed parameter is FlowZero and the two options are Cal or Stab With the Cal option selected the zero reference used for each test will be the 0 value obtained during the first two passes of the calibration cycle With the Stab option selected the zero reference used for each test will be the flow transducer reading at the end of the stabilization time It is generally preferable to use the Stab option as a new zero is obtained during each test cycle which compensates
37. ved for future use Reserved for future use H Current Pressure Value 0 500 Pressure Reference Value 0 500 J Scaled Flow rate 0 4095 in 0 1 LPM units K Flow scale factor 0 4095 M Time base 1 1 0 Second 2 0 1 Second N Reserved for future use FILL time 0 255 0 999 for ver 4 20 or later STAB time 0 255 0 999 for ver 4 20 or later Q READ time 0 255 0 999 for ver 4 20 or later R Reserved for future use S Reserved for future use T Reserved for future use U Leak limit 0 0 25 5 V Calibrated Leak 0 25 5 W Zero Selection 1 Cal 2 Stab X Leak slope limit 0 1 10 0 Y Leak negative limit 0 1 5 0 Z Pressure Tolerance 0 20 A EOT June 2004 HKA CT 1000 Leak Tester Manual Page 28 The data from the RS232 port can be input to any IBM compatible computer system PC There are currently two versions of software offered by HKA the first will graph the reading of the mass flow transducer The second SETUP will summarize leak test results using SPC type format The user can also input the data from this port into their own custom SETUP using the format as outlined data transfer is unidirectional from the CT 1000 test system to the PC monitoring system There is no effect on the leak test system if the PC is connected to the RS232 port or not The serial data has the following format Speed 9600 baud Par