BIOMEDOS M Radio Test Report (FCC143961)

image

Contents

1. 30 000 40 50 60 70 80 MHz 300 400 500 600 700 1000 000 Reading Correct Measure No Mk Freq Level Factor ment Limit Over MHz dBu dBim dBuVim dBuV m dB Detector 1 81 2117 91 84 23 21 28 63 40 00 11 37 peak 2 106 0126 50 73 21 85 28 88 43 50 14 62 peak 3 199 9856 56 19 20 39 35 80 43 50 7 70 peak 4 290 3012 54 10 18 11 39 99 46 00 10 01 peak 5 314 3765 51 51 16 54 34 97 46 00 11 03 peak 6 425 0280 50 60 12 92 37 68 46 00 8 32 peak Maximum data x Over limit over margin Emission Level Read Level Correct Factor TB RF 074 1 0 Report No TB FCC143961 TOBY Page 17 of 18 EUT Device for generating modulated Model Name BM1 signals BIOMEDIS M Temperature 29 C Relative Humidity 55 Test Voltage DC 12V Ant Pol Horizontal Test Mode Mode 2 Normal Playing Mode Remark Only worse case is reported 80 0 dBu m FCC 15B 3M Radiation Margin 6 dB 30 y a 20 30 000 40 50 60 70 80 MHz 300 400 500 600 700 1000 000 Reading Correct Measure No Mk Freq Level Factor ment Limit Over MHz dBu dB m dBuV m dBuV m dB Detector 1 47 9940 54 14 23 54 30 60 40 00 9 40 peak 2 96 0986 46 48 22 16 24 32 43 50 19 18 peak 3 119 8556 48 23 22 50 25 73 43 50 17 77 peak 4 143 8295 50 15 21 67 28 48 43 50 15 02 peak 5 N 167 8243 60 89 21 04 39 85 43 50 3 65 peak 6 l 191 7450 60 02 20 81 39 21 43 50 4 29 peak Maximum data x Over limit over margin Emission Level Read Level Corr
2. of confidence of approximately 95 Test tem Expanded Uncertainty Ua Level Accuracy Conducted Emission 9kHz 150kKHz 3 42 dB 150kHz to 830MHz 3 42 dB Level Accuracy Radi Emissi 4 B 9kHz to 30 MHz Level Accuracy Radiated Emissi 4 40 dB Level Accuracy Radiated Emissi 4 20 dB Above 1000MHz aae OO 1 7 Test Facility The testing report were performed by the Shenzhen Toby Technology Co Ltd in their facilities located at 1A F Bldg 6 Yusheng Industrial Zone The National Road No 107 Xixiang Section 467 Xixiang Bao an Shenzhen Guangdong China At the time of testing the following bodies accredited the Laboratory CNAS L5813 The Laboratory has been accredited by CNAS to ISO IEC 17025 2005 General Requirements for the Competence of Testing and Calibration Laboratories for the competence in the field of testing And the Registration No CNAS L5813 FCC List No 811562 The Laboratory is listed in the United States of American Federal Communications Commission FCC and the registration number is 811562 IC Registration No 11950A 1 The Laboratory has been registered by Certification and Engineering Bureau of Industry Canada for radio equipment testing The site registration Site 11950A 1 May 22 2014 certificated by TUV Rheinland China Co Ltd with TUV certificate No UA 50282953 0001 and report No 17026822 002 The certificate is valid until the next scheduled audit or up to 18 months at t
3. 143961 Page 9 of 18 Device for generating modulated EUT l Model Name BM1 signals BIOMEDIS M Temperature 25 C Relative Humidity 55 Test Voltage AC 120V 60Hz Terminal Line Test Mode Mode 1 Remark Only worse case is reported 0 0 dBu haa RIRE x L 10 0 150 05 MHz 5 30 000 Reading Correct Measure No Mk Freq Level Factor ment Limit Over MHz dBu dB dBuV dBuV dB Detector 1 0 7140 35 30 10 12 45 42 96 00 10 58 QP 2 0 7140 21 61 10 12 31 73 46 00 14 27 AVG 3 1 8900 28 49 10 06 38 55 56 00 17 45 QP 4 1 8900 19 59 10 06 29 65 46 00 16 35 AVG 5 3 3500 25 95 10 02 35 97 56 00 20 03 QP 6 3 3500 19 02 10 02 29 04 46 00 16 96 AVG T 5 2500 23 87 9 97 33 84 60 00 26 16 QP 8 5 2500 17 08 9 97 27 05 50 00 22 95 AVG 9 18 5419 16 49 10 19 26 68 60 00 33 32 QP 10 18 5419 11 27 10 19 21 46 50 00 28 54 AVG 11 24 0020 20 53 10 16 30 69 60 00 29 31 QP 12 24 0020 19 70 10 16 29 86 50 00 20 14 AVG Emission Level Read Level Correct Factor TB RF 074 1 0 Report No TB FCC143961 Page 10 of 18 gun for generating modulated atl Were Sue signals BIOMEDIS M Temperature 25 CU Relative Humidity 55 Test Voltage AC 120V 60Hz Terminal Neutral Test Mode Mode 1 Remark Only worse case is reported 90 0 dBuV eo peak AVG 10 0 150 05 MHz 5 30 000 Reading Correct Measure No Mk Freq Level Factor ment Limit Over MHz dBuV dB dBuV dBuV dB Detector 1 0 2260 3
4. 8 98 10 11 49 09 62 59 13 50 QP 2 0 2260 29 79 10 11 39 90 52 59 12 69 AVG 3 0 4980 37 73 10 02 47 75 96 03 8 28 QP 4 0 4980 21 50 10 02 31 52 46 03 14 51 AVG 5 0 5860 37 37 10 02 47 39 56 00 8 61 QP 6 0 5860 22 41 10 02 32 43 46 00 13 57 AVG fj 0 7140 34 85 10 03 44 88 56 00 11 12 QP 8 0 7140 21 27 10 03 31 30 46 00 14 70 AVG 9 0 8740 33 88 10 10 43 98 56 00 12 02 QP 10 0 8740 22 62 10 10 32 72 46 00 13 28 AVG 11 0 9700 30 65 10 15 40 80 96 00 15 20 QP 12 0 9700 20 83 10 15 30 98 46 00 15 02 AVG Emission Level Read Level Correct Factor TB RF 074 1 0 Report No TB FCC143961 TOBY Page ll of 18 5 Radiated Emission Test 5 1 Test Standard and Limit 5 1 1 Test Standard FCC Part 15 109 5 1 2 Test Limit Radiated Emission Limit Frequency Field Strength Measurement Distance MHz dBuV m meters Above 960 Note Emission Level dBuV m 20log Emission Level uV m For unintentional radiators FCC Part 15 section 15 33 1 Except as otherwise indicated in paragraphs b 2 or b 3 for an unintentional radiator including a digital device the spectrum shall be investigated from the lowest radio frequency signal generated or used in the device without going below the lowest frequency for which a radiated emission limit is specified up to the frequency shown in the following table Highest frequency generated or used Upper frequency of measurement in the device or on which the device range MHz operate
5. Hz 5 0 30 0 Notes 1 Decreasing linearly with logarithm of the frequency 2 The lower limit shall apply at the transition frequencies 4 2 Test Setup RECEIVER 4 3 Test Procedure The EUT was placed 0 8 meters from the horizontal ground plane with EUT being connected to the power mains through a line impedance stabilization network LISN All other support equipments powered from additional LISN s The LISN provide 50 Ohm 50uH of coupling impedance for the measuring instrument Interconnecting cables that hang closer than 40 cm to the ground plane shall be folded back and forth in the center forming a bundle 30 to 40 cm long I O cables that are not connected to a peripheral shall be bundled in the center The end of the cable may be terminated if required using the correct terminating impedance TB RF 074 1 0 Report No TB FCC143961 Page 8 of 18 TOBY The overall length shall not exceed 1 m LISN at least 80 cm from nearest part of EUT chassis The bandwidth of EMI test receiver is set at 9kHz and the test frequency band is from 0 15MHz to 30MHz For the actual test configuration please refer to the EUT test Photos 4 4 EUT Operating Mode 1 Setup the EUT and peripherals refer to the description of test mode 4 5 Deviation The test is no deviation from the standard 4 5 Test Data Please see the next page TB RF 074 1 0 TOBY Report No TB FCC
6. T Equipment Under Test sss sss eee ee 3 1 3 Block Diagram Showing the Configuration of System Tested sese eee eee 4 1 4 Description Of Support UNIS 000 ee eee ee 4 RUR Onen Ol OSE MOO Soes E E eens eeaneine eeine 4 1 6 Measurement Uncertainty ee eee ee 5 Tc OG FCN T 5 2 11 25 III Voo cageck cope cnacaemecnaeteasutesisdeecscestsavuncs iunesscvsnssavencegnsaesssascandevcivuctsnstuscesnseoatsese 6 3 TEST EQUIPMENT oasis ETE E EE EE 6 4 CONDUCTED EMISSION TES ssc ivsncssctcccosecececesectesesussetscoscstscsscscvavcssectsscesucevacssisisccpesssssesvecsse 7 4 1 Test Standard and Limit 7 A2 E nO ae E E Cn ee 7 4 3 Test Procedure 7 Aa EUTODE ANO MOOG T 8 Ao VD CUA Oi T 8 de 8 5 RADIATED EMISSION PES C 11 5 1 Test Standard and Limite eee 11 SE dE E E ce een ne A eee ene ee ee eee ee ee 11 So Rl see lT 13 PE UI ODS Fat CONO O asser A NA 14 Mi DOVON eera E E E AEE EE TEE AEE E EA TR 14 oho Ros AB e E E NE E E eee E eee 14 TB RF 074 1 0 Report No TB FCC143961 TOBY Page 3 of 18 1 General Information about EUT 1 1 Client Information Applicant g BIOMEDIS TECHNOLOGIES CO LIMITED Address UNIT E223 3 F WING TAT COMM BLDG 97 BONHAM STRAND EAST SHEUNG WAN HONG KONG BIOMEDIS TECHNOLOGIES CO LIMITED TECHNOLOGIES CO LIMITED ee 810 E223 3 F Celli E TAT COMM BLDG 97 BONHAM STRAND EAST SHEUNG WAN HONG KONG 1 2 General Description of EUT Equipment Under Test EUT Name A Device for generatin
7. TOBY Shenzhen Toby Technology Co Ltd Report No TB FCC143961 Page L of 18 FCC 15B Test Report FCC ID 2AEMC BMT Original Grant Report No TB FCC143961 Applicant BIOMEDIS TECHNOLOGIES CO LIMITED Equipment Under Test EUT EUT Name Device for generating modulated signals BIOMEDIS M Model No BM1 Serial No N A Brand Name N A Receipt Date 2015 04 22 Test Date 2015 04 22 to 2015 04 24 Issue Date 2015 04 28 Standards FCC Part 15 2014 Subpart B Class B Test Method ANSI C63 4 2014 Conclusions PASS In the configuration tested the EUT complied with the standards specified above The EUT technically complies with the FCC eeii Test Witness Engineer Approved amp Authorized This report details the results of the testing carried out on one sample The results contained in this test report do not relate to other samples of the same product The manufacturer should ensure that all products in series production are in conformity with the product sample detailed in the report TB RF 074 1 0 1A F Bldg 6 Yusheng Industrial Zone The National Road No 107 Xixiang Section 467 Xixiang Bao an Shenzhen China Tel 86 75526509301 Fax 86 75526509195 Report No TB FCC143961 Page 2 of 18 Contents CONTENTS on E E E T 2 1 GENERAL INFORMATION ABOUT EUT 0 csssccssssccssscscsssccsssscssssesssscssssssssssssoes 3 TN MST TEOMA A OV T 3 1 2 General Description of EU
8. ect Factor TB RF 074 1 0 Report No TB FCC143961 TOBY Page 18 of 18 Device for generating modulated B signals apial N aali aul Temperature 29 C Relative Humidity 55 Test Voltage DC 12V Ant Pol Vertical Test Mode Mode 2 Normal Playing Mode Remark Only worse case is reported 80 0 dBuV m EEREN FCC 15B 3M Radiation Margin 6 dB A e 30 5 4 i 4 20 30 000 40 50 60 70 80 MHz 300 400 500 600 700 1000 000 Reading Correct Measure No Mk Freq Level Factor ment Limit Over MHz dBu dB m dBuV m dBuV m dB Detector q1 47 9940 61 63 23 54 38 09 40 00 1 91 peak 72 0843 39 08 23 54 15 54 40 00 24 46 peak 96 0986 46 07 22 16 23 91 43 50 19 59 peak 119 8556 47 60 22 50 25 10 43 50 18 40 peak 167 8243 55 75 21 04 34 71 43 50 8 79 peak 191 7450 53 24 20 81 32 43 43 50 11 07 peak oO mm j Mj Maximum data x Over limit over margin Emission Level Read Level Correct Factor TB RF 074 1 0
9. g modulated signals BIOMEDIS M BrandNome NA Model No Aem o sanle G Power Supply DC Voltage supplied from PC System by USB Cable DC power by Li ion battery Power Rating USB DC 5V DC 3 7V by 300 mAh Li ion Battery Note 1 The EUT with a USB port can communicate with PC by USB Cable The EUT considered as an ITE Computing Device 2 For more detailed features description please refer to the manufacturer s specifications or the User s Manual TB RF 074 1 0 Report No TB FCC143961 Page 4 of 18 TOBY 1 3 Block Diagram Showing the Configuration of System Tested Cable 1 1 4 Description of Support Units Equipment Information OPTIPLEX380 DOC Name LCD Monitor Keyboard Mouse M UARDEL7 DOC DELL Printer 0 1 5 Description of Test Mode Mode Description USB Charging and Loading Data Normal Playing To investigate the maximum EMI emission characteristics generates from EUT the test system was pre scanning tested base on the consideration of the EUT operation mode and the maximum emission levels of the conducted and radiated emissions are compared to the FCC Part 15 Subpart B Class B limits TB RF 074 1 0 Report No TB FCC143961 Page 5 of 18 TOBY 1 6 Measurement Uncertainty The reported uncertainty of measurement y U where expended uncertainty U is based on a standard uncertainty multiplied by a coverage factor of k 2 providing a level
10. he discretion of TUV Rhineland TB RF 074 1 0 Report No TB FCC143961 TOBY Page 6 of 18 2 Test Summary FCC Part15 Subpart B Test Method Test Item Judgment Radiated Emission 15 109 ANSI C63 4 2014 Class B PASS gria 30M 1GHz Gass pass Conducted Emission 15 107 ANSI C63 4 2014 Class B PASS eira 150KHz to 30MHz wan PASS Note N A is an abbreviation for Not Applicable 3 Test Equipment Conducted Emission Test Cal Due Model No Last Cal Date EMI Test Rohde amp Schwarz ESC 100321 Aug 08 2014 Aug 07 2015 Receiver RF Switching Compliance Direction RSU A4 34403 Aug 08 2014 Aug 07 2015 Unit Systems Inc SCHWARZBECK NNBL 8226 2 8226 2 164 Aug 08 2014 Aug 07 2015 LISN Rohde amp Schwarz ENV216 101131 Aug 08 2014 Aug 07 2015 Radiation Emission Test Cal Due Model No Last Cal Date Spectrum Agilent E4407B MY45106456 Aug 08 2014 Aug 07 2015 Analyzer Spectrum Rohde amp Schwarz FSP30 DE25181 Aug 08 2014 Aug 07 2015 Analyzer EMI T Rohde amp Schwarz ESCI 101165 Aug 08 2014 Aug 07 2015 Receiver Signal Rohde amp Schwarz SML03 IKW682 054 Feb 10 2015 Feb 09 2016 Generator Positioning ETS LINDGREN 2090 N A N A N A Controller TB RF 074 1 0 Report No TB FCC143961 TOBY page Tf 18 4 Conducted Emission Test 4 1 Test Standard and Limit 4 1 1 Test Standard FCC Part 15 107 4 1 2 Test Limit Conducted Emission Test Limit Frequency M
11. s or tunes MHz Below 1 705 1 05 108 1000 108 500 2000 900 1000 9000 Above 1000 5 harmonic of the highest frequency or 40 GHz whichever is lower 5 2 Test Setup TB RF 074 1 0 Report No TB FCC143961 Page 12 of 18 RX Antenna Metal Full Soldered Ground Plane Spectrum Analyzer Receiver Bellow 30MHz Test Setup RX Antenna Ant feed point 1 4m an gt U GE Ei Metal Full Soldered Ground Plane Spectrum Analyzer Receiver wane oe 30MHz to 1000MHz Test Setup TB RF 074 1 0 Report No TB FCC143961 Page 13 of 18 3 Antenna yr tower gt 3m lt 4 Horn EUT Y antenna 4m p fo Spectrum A pei E analyzer B r L Y Turntable f im A E 30cm pi Se Above 1GHz Test Setup 5 3 Test Procedure 1 The measuring distance of 3m shall be used for measurements at frequency up to 1GHz and above 1 GHz The EUT was placed on a rotating 0 8m high above ground the table was rotated 360 degrees to determine the position of the highest radiation 2 Measurements at frequency above 1GHz The EUT was placed on a rotating 0 8m high above the ground RF absorbers covered the ground plane with a minimum area of 3 0m by 3 0m between the EUT and measurement receiver antenna The RF absorber shall not exceed 30cm in high above the conducting floor The table was rotated 360 degrees to determine the position of the highest radiation 3 The Test antenna shall var
12. y between 1m and 4m Both Horizontal and Vertical antenna are set to make measurement 4 The initial step in collecting conducted emission data is a spectrum analyzer peak detector mode pre scanning the measurement frequency range Significant peaks are then marked and then Quasi Peak detector mode re measured 5 If the Peak Mode measured value compliance with and lower than Quasi Peak Mode Limit Bellow 1 GHz the EUT shall be deemed to meet QP Limits and then no additional QP Mode measurement performed But the Peak Value and average value both need to comply with applicable limit above 1 GHz 6 Testing frequency range below 1GHz the measuring instrument use VBW 120 kHz with Quasi peak detection 7 Testing frequency range above 1GHz the measuring instrument use RBW 1 MHz and VBW 3 MHz with Peak Detector for Peak Values and use RBW 1 MHz and VBW 10 Hz with Peak Detector for Average Values 8 For the actual test configuration please see the test setup photo TB RF 074 1 0 Report No TB FCC143961 TOBY Page 14 of 18 5 4 EUT Operating Condition 1 Setup the EUT and peripherals refer to the description of test mode 5 5 Deviation The test is no deviation from the standard 5 6 Test Data Remark The Highest frequency of the device operates or tunes below 108MHz so no requirement for test the emission frequency of above 1GHz Test data please refer the following pages TB RF 074 1 0 TOBY 1 Bellow 1GH
13. z EUT Device for generating modulated signals BIOMEDIS M Model Name Report No TB FCC143961 Page 15 of 18 Temperature 25 C Relative Humidity Test Voltage Ant Pol AC 120V 60Hz Horizontal Test Mode Mode 1 USB Charging and Loading Data Remark Only worse case is reported 80 0 dBu m 20 30 000 40 No Mk 50 60 70 80 MHz Correct Factor Reading Level dBuV Freq MHz dB m 300 Measure ment dBuV m 199 9856 239 9874 323 3204 379 9141 475 4991 776 8778 56 64 92 01 47 76 42 87 42 06 39 75 20 39 18 59 16 20 14 14 11 60 6 72 36 25 33 42 31 56 28 13 30 46 33 03 Maximum data x Over limit over margin Emission Level Read Level Correct Factor EEREN FCC 15B 3M Radiation 400 Limit dBu m 43 50 46 00 46 00 46 00 46 00 46 00 Margin 6 dB 500 600 700 Over dB 7 25 12 58 14 44 17 27 15 54 12 97 1000 000 Detector peak peak peak peak peak peak TB RF 074 1 0 Report No TB FCC143961 TOBY Page 16 of 18 Device for generating modulated signals wali N meen Nance aui Temperature 29 C Relative Humidity 55 Test Voltage AC 120V 60Hz Ant Pol Vertical Test Mode Mode 1 USB Charging and Loading Data Remark Only worse case is reported 80 0 dBu m se FCC 15B 3M Radiation Margin 6 dB ji 6 S 7 FA 20