1. chattanooga Alliance Stand Assist Lifts Models Stand Assist 1929 shown Legacy Series Performance Series 1906 400 Ib 1933 1932 400 Ib 1931 500 Ib 1930 1929 500 Ib 1907 600 Ib 1928 1927 600 Ib Operator Instructions Alliance Stand Assist Thank you for choosing Chattanooga Alliance Patient Lift To better serve you please record the following information Dealer Name Telephone Serial Date of Purchase INSPECT YOUR MERCHANDISE Upon receipt of your Patient Lift verify that all merchandise is complete and free from any shipping damage Refuse delivery if the packaging appears to be badly damaged If the merchandise is re ceived damaged or is missing components contact the shipping company immediately and file a claim For further assistance contact your local dealer or Product Support at the following 1 800 494 3395 USA chattproductsupport djoglobal com DJO Inc 1430 Decision Street Vista CA 92081 USA T 1 800592 7329 USA DJOGlobal com 2013 DJO LLC All rights reserved Alliance Stand Assist SAFETY The Stand Assist sit to stand patient lift allows a caregiver to lift and transfer a patient safely with minimum physical effort Before attempting to lift anyone you should practice operating the lift and be able to explain to the patient the lifting and transfer procedures Always keep the Operator s Instructions Manual available with the lift SAFETY INSTRUCTIONS
2. Special care must be taken with patients who are comatose spastic agitated or otherwise severely handicapped A competent caregiver must always be present during any patient lifting The sole function of the patient lift is for transferring a patient from one utility beds bathtubs toilets etc to another The lift should not be used to transport or move patients from room to room or over longer distances Transfers should normally take place within the same patient room or bathroom Do not put anything e g cushion pad etc between the user and the sling This may cause user to slide out of the sling and cause injury The user patient being transferred should be positioned in the center of the area being transferred from i e bed wheel chair chair commode etc before being lifted as well as when he she is returned Keep lifter base widened use foot pedals to accomplish this and brakes unlocked during lifting If the brakes are locked during the lift process the lift may become unstable While the patient is suspended always keep the patients centered between the base legs Do not attempt to roll the lift or casters over floor obstacles while the user is in the sling and on board Use only Chattanooga slings manufactured by BestCare specially designed and tested for Chattanooga patient lifts Do not use slings manufactured by other companies on Chattanooga Stand Assist sit to stand patient lift Chattanooga is no
3. LONG LOOP reclined position CENTER LOOP CENTER LOOP semi reclined SHORT LOOP SHORT LOOP most upright FOR SLING APPLICATION Place sling behind seated patient with the 2 red pads under the arms Cross the adjustable chest belt over and fasten it securely using the buckle adjust for comfort e Cross the short black strap over and pull the left loop strap through the black strap e Hook the two loop straps to the hooks on the lift boom using the same hook and loop sequence on each side e The black strap will be straight across in front of patient to stabilize the sling during lift e Have patient grasp handles on boom and tilt head back slightly e Make sure the patients feet are firmly on the platform and knees are touching the kneepads with kneecaps just above the padding e Use the lift to complete the lift and pivot transfer 18 Alliance Stand Assist FITTING SUPPORT STRAP DIAGRAM 1 DIAGRAM 3 LIFTING HINTS Check sling weight capacity for the patient being lifted Do not exceed maximum safe working load Review following procedures carefully prior to attempting lift Call with any questions Position the sling around the buttocks of the patient This sling is designed to support the middle and lower part of the body Attach the sling strap to the lower pig tail A on the mast of the Stand Assist lift as in Diagram 1 Bringing the other side around the back of the patients
4. This will shorten the life of the batteries e Do not let the batteries run down to the last red light This might shorten the life of the batteries or damage the batteries e Unplug the charger first before using the lift Alliance Stand Assist WORKING WITH THE BATTERY CONT KNOW WHAT YOUR BATTERY IS TELLING YOU RED AMBER GREEN GREEN os fon fou fon Jon eurem NEED CHARGE C O eee a orr orr orr orr orr BAD BATTERIES OR BAD CONNECTION WARNING BUZZERS The control box has a buzzer that provides you with information on the status of the batteries and the lift To stop the buzzer push down the RESET button or press the Battery Test button BUZZER TYPE INDICATION NOTE red amber lights on red light on no light on other second no light on does not lift IMPORTANT NOTES ON CHARGING e Push lift to charging location and charge the batteries with the charger provided Avoid unplug hand control and motor from control box Frequent plug unplug of the hand control and motor into from the control box may damage the control box e Do not charge batteries over 12 hours A Patient Lift will not operate while charging Alliance Stand Assist OPERATING INSTRUCTIONS A d Double check all assemblies for tightness and read operating instructions carefully prior to use Do not attempt to use patient lift unless the patient exhibits control over the upper body strength to grasp the handles
5. ability to bear some weight and flexibility in knees hips and ankles PREPARATION BEFORE LIFTING e Turn on the power by twisting the RED RESET BUTTON clockwise e Press the UP or DOWN button on the hand control once and check if the battery indicator lights are on e Position the base of the lift around or under the object e Widen the base and engage the caster brakes Do not attempt to transfer a patient without prior approval of the person s nurse Also do not transfer without having studied the instructions and performed several practices in operating the product Together with the patient s doctor nurse or medical attendant select a sling that is both practical and comfortable The sling selected should be one that serves the needs of the patient while providing the patient with optimal safety Never interfere with the lift unless instructed by the attendant Have a doctor nurse or medical attendant experienced in the use of the Stand Assist Lift present during the first few times the lift is used to transfer a new user Alliance Stand Assist FITTING STAND ASSIST SLING e Position the patient in a sitting position e Slide the sling down patient s back to lumbar position e Draw the shoulder straps to the front of the patient close to the chest e Draw the waist belt around patient s waist and press together to fasten Attach the shoulder straps to the hooks Velcro waist belt Sling Loop Options LONG LOOP
6. defects or damage caused by labor furnished by anyone other than Company Company s dealer or Company s certified service center or 3 any malfunction or failure in a Product caused by abuse accident or misuse including but not limited to the failure to provide required maintenance or any use that is inconsistent with the Product User Manual Company is not responsible for injury death or damage resulting from Product modifications or repairs performed by service personnel which have not been authorized by Company This warranty extends to the original customer and is not transferable This paragraph states the entire warranty relating to the Product and Company does not authorize any person or representative to modify this warranty COMPANY HEREBY DISCLAIMS ANY OTHER EXPRESS OR IMPLIED WARRANTIES NOT SET FORTH IN THE FOREGOING LIMITED PRODUCT WARRANTY INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE COMPANY WILL NOT BE LIABLE FOR ANY INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOST PROFITS CAUSED BY ANY PRODUCT DEFECT WHETHER CLAIMS ARE BASED UPON TORT INCLUDING NEGLIGENCE WARRANTY CONTRACT OR OTHERWISE EVEN IF COMPANY HAS BEEN ADVISED OF SUCH POTENTIAL LOSS OR DAMAGE TO THE EXTENT THE FOREGOING DISCLAIMERS ARE NOT ALLOWED BY APPLICABLE LAW ANY IMPLIED WARRANTIES WILL BE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY PERIODS APPLICABLE TO THE PRODUCT S
7. s buttocks to fit like a park swing and attach lower pig tail B Position sling so it gives the greatest support to middle and lower part of the patient being lifted Notice in Diagram 2 and 3 that the Buttock strap MUST be used with the Stand Assist sling These slings are designed to compliment one another in providing the greatest comfort and security for the patient being transferred Do not lock caster brakes to allow lift to properly balance weight e Prior to beginning a lift have the patient lean slightly back and look at the a 45 degree angle to the ceiling this help position the sling this is recommended only at the start of a lift e Try the sling straps in various color positions to establish best fit e Always insure the same color straps on each side are attached to lift e Should patients have a larger girth in mid section or back side area try the optional buttock strap to help with initial lift leverage e Remember this lift is designed for pivot transfers and it is not necessary to lift to full upright position as long as the patients body clears bed or chair and vice versa you can Itis not uncommon for the strap to loosen and drop down as lift proceeds 19 Alliance Stand Assist LIFT AND TRANSFER FROM BED e Fit sling as described in Fitting Stand Assist Sling e Push lift towards patient Open the base of the lift Apply the brakes in both rear casters e Position patient
8. base of the lift to go around the chair Apply the brakes in both rear casters e Position patient s feet on the foot platform and knees against the knee pad e Attached the sling straps to the hooks e Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest e Press the UP button on the hand control e Before the patient s body is lifted from the chair stop and make sure the sling is secured and patient s knees are against the knee pad e The patient should be comfortable with the pressure under the arms If not adjust the sling and try another loop option to release pressure e Press the UP button until the patient s body has completely left the chair e Transfer to another object can be done at this position without the patient attending a full standing position e Release the brakes close the base and pull the lift away from the chair e Lower the patient to the object intended e To lift patient in a standing position continue to lift until the patient s knees are locked ina standing position e At the standing position patient s knees may move away from the knee pad e Release the brakes close the base and pull the lift away from the chair e Lower the patient to the object intended e Reverse the above procedures when lowering the patient to the chair 21 Alliance Stand Assist LIFT AND TRANSFER USING SANI S
9. s feet on the foot platform and knees against the knee pad e Attached the sling straps to the hooks e Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest e Press the UP button on the hand control e Before the patient s body has lifted from the bed stop and make sure the sling is secured and patient s knees are against the knee pad e The patient should be comfortable with the pressure under the arms If not adjust the sling and try another loop option to release pressure e Press the UP button until the patient s body has completely left the bed e Transfer to another object can be done at this position without the patient attending a full standing position e Release the brakes close the base and pull the lift away from the bed e Lower the patient to the object intended e To lift patient in a standing position continue to lift until the patient s knees are locked in a standing position e At the standing position patient s knees may move away from the knee pad e Release the brakes close the base and pull the lift away from the bed e Lower the patient to the object intended e Reverse the above procedures when lowering the patient to the bed 20 Alliance Stand Assist LIFT AND TRANSFER FROM SEATED POSITION e Fit sling as described in Fitting Stand Assist Sling e Push lift towards patient Open the
10. slotted bolt head inside the L arm then tighten to secure Repeat these steps for the other side U Bracket Threaded Knob With both L arms securely attached to the mast slide the kneepad onto the L arms and fasten into place with U bracket and knob bolt Knob Bolt Bracket Alliance Stand Assist BATTERY INSTALLATION FOR PERFORMANCE Series Install the control box Slide the control box over and onto the flat metal hook mounted at the bottom of the Mast Line up the control box with the hole at the top of the mast and insert the screw to hold the control box in place Secure the control box by turning the lock mechanism into the lock position CONTROL BOX Plug Ins for Performance Series LCD display LCD Lifting actuator UP Lifting actuator DOWN AC Power Plug Battery pack release Battery indicator Actuator Pluc Emergency button Pendant Plug Alliance Stand Assist CONTROL BOX FEATURES e Smart charging function for longer battery life e Battery capacity indicator on battery pack e Soft start and stop for lifting actuator e Soft and Hardware over current protection e Actuator s Over duty protection with sign on LCD e Emergency stop button interrupts the power supply to the actuator and makes the actuator stop immediately in case of sudden danger e Internal charger e LCD display showing battery capacity and indica
11. Alliance Stand Assist EMERGENCY LOWERING MECHANIM Gi The following instructions apply to PERFORMANCE series ONLY In case there is a failure with the actuator or electronics and the user is left suspended in mid air please follow the procedures below to safely lower the user to a safer position The Quick Release mechanism is activated by the RED lever mounted at the top end of the actuator Activate the Quick Release by pulling the RED lever with the arrow symbol in the direction indicated by the arrow This will de clutch the actuator and safely lower the user down NOTE When the Quick Release is activated the boom will not lower itself unless there is sufficient weight or force applied in the downward direction Contact your dealer immediately if standard troubleshooting techniques do not correct the failure Do not attempt to lift until all failure and safety issues have been resolved NOTE The emergency lowering device is intended for use during lift failure This device will allow lowering of patients only Please contact your dealer immediately in case of failure 26 Alliance Stand Assist WARRANTY LIMITED PRODUCT WARRANTY Subject to the terms below DJO LLC Company provides the following limited product warranty relating to Chattanooga Alliance Patient Lifts Products e Lift frames and spreader bar 3 years e Actuator 2 years e Parts including casters control box PCB pendant charger 2 year
12. LING SANI TOILETING SLING The Toileting Sling can be used with the Stand Assist to lift a patient in a sitting position When in use the Stand Assist Toileting Sling supports the client under the thighs and across the mid back area The Sani Toileting Sling is not suitable for patients who have involuntary movements or patients with limited postural control Fitting the Sani Toileting Sling e Slide the sling down patient s back to lumbar position e Draw shoulder straps under the arms and around the chest into the front e Draw Velcro waist belt around waist and press together to fasten e Draw each leg support under the thigh e Crisscross leg straps by feeding one of the leg straps through the other strap e Attach shoulder straps to hooks in upper ends of the lifting arms e Attach leg straps to hooks in the lower ends of the lifting arm 22 Alliance Stand Assist MAINTENANCE AND INSPECTION The operator of the lift shall inspect the Stand Assist Lift before each use e Check all bolts for tightness e Make sure the base can be easily widened and that all lift parts are in place e Make sure that casters can be turned freely and that caster brakes can be engaged e Make certain all necessary items i e slings wheelchairs etc are ready for use At least once a month the lift should be thoroughly inspected by a person qualified to recognize any signs of wear and tear and looseness of bolts or parts Replace any worn par
13. TATED ABOVE SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS OR THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES SO THE ABOVE LIMITATIONS MAY NOT APPLY TO CUSTOMER 27 Alliance Stand Assist DJO Inc 1430 Decision Street Vista CA 92081 USA T 1 800 592 7329 USA DJOGlobal com Alliance Stand Assist Manual 2013 DJO LLC All rights reserved 28803 RevA 9 2013 28
14. by pulling the battery pack without pressing the release button B Fig 2 JO Alliance Stand Assist BATTERY ASSEMBLY FOR LEGACY SERIES Attach the Control Box Hanging Bracket to the Mast and tighten the nuts Once the bracket is secure then place the Legacy Electronic Box inside CONTROL BOX INSTRUCTIONS FOR LEGACY SERIES Battery Indicator Battery Test Button Emergency Button Alliance Stand Assist WORKING WITH THE BATTERY Charger Plug Pendant Plug Actuator Plug CHECKING THE BATTERY e Check batteries by pressing the Battery Testing Button blue circle with a battery sign on the control box e Batteries are fully charged when all lights on the Battery Indicator are ON e Charge batteries when Indicator shown only one GREEN light Do not use the lift when no GREEN light is shown Charge the batteries immediately e Replace batteries when frequent charging is needed CHARGING THE BATTERY e Ensure the power is switched ON the red RESET button should be up e Insert charging plug into charging port on the control box e Plug charger to power supply e All lights of battery indicator should be ON while charging e It takes approximately 2 3 hours to fully charge the batteries from one green light e It takes approximately 7 9 hours to fully charge the batteries from the red light e Do not leave the batteries on charging for an extensive period of time
15. missing Parts List for PERFORMANCE Series Base Mast Control Box Pendant Actuator Lifting Boom Kneepad Assembly Footrest Castor with Brake 1 2i 3 4 5 6 7 8 9 1 Front Castor Parts List for LEGACY Series Base Mast Control Box Pendant Actuator Lifting Boom Kneepad Assembly Footrest Castor with Brake Front Castor O ONAN AWN gt Alliance Stand Assist STEP BY STEP ASSEMBLY Step 1 Engage the brakes Remove the bolts from the bottom of the mast and base Pull the Mast to an upright position Re insert the bolts into the holes at bottom of the mast and tighten the nuts Step 2 Attach boom to top of mast Re insert the bolts into the holes at bottom of the mast and tighten the nuts Alliance Stand Assist STEP BY STEP ASSEMBLY CONT Step 3 A Attach the bottom of Actuator to the bracket on the Mast and insert pin Insert keeper ring through hole in pin to secure base of actuator B Attach the top of the Actuator to the bracket on the Boom and insert pin Insert keeper ring through hole in pin to secure actuator to upper bracket J A A CARA Cy Alliance Stand Assist STEP BY STEP ASSEMBLY CONT Step 4 KNEEPAD ASSEMBLY FOR PERFORMANCE Series Insert slotted bolt through bracket with washer on the outside and loosely attach threaded knob to the mast by turning the knob two times L Shape Arms Slide L arm inside bracket with
16. nents or parts on your lift as this may void your warranty or cause further problems that may result in patient injury Stop using your lift immediately if it is not functioning correctly or any warning beeps are heard need to re charge my batteries often or they fail to hold a charge when I charge them e Replace the batteries or battery pack as they are at the end of their life cycle e Batteries should be changed every 18 24 months depending on usage The actuator will either go up or down but not both e Replace your hand control as it is at the end of its life cycle e Your control box may be malfunctioning There is a grinding sound inside the actuator when lifting e Replace your actuator as the internal gears are stripped The actuator stops and starts while lifting or lowering e Replace the batteries or battery pack as they are at the end of their life cycle My lift will not operate even though it shows the batteries are charged e Unplug the AC power cord from the control box as the lift will not work when the control box charger is connected to AC Power e Check to make sure the hand control is properly inserted all the way into the control box port e Check to make sure the actuator plug is properly connected and inserted all the way into the control box e Your hand control needs to be replaced as it is at the end of its life cycle If there is a problem you cannot correct please contact the dealer for assistance 25
17. olution for patient transfers between bed chair or commode The lift reduces injury risk associated with antiquated manual patient lifting practices Competitively priced and combined with a unique design the lift provides a new level of comfort and safety The Stand Assist lift has set the standard for value and is an option for residential use e 400 500 600 Lb safe working loads e Low base e Fully adjustable knee pad e Removable foot platform e Four sizes of deluxe padded slings e Optional buttock support strap e Knee pad strap for extra leg support Features alternative Sani sling hook up that eliminates underarm discomfort during transfer Alliance Stand Assist SPECIFICATIONS AND OPTIONS Control Box Legacy Series Performance Series 220 240VAC 50Hz 110 220 240VAC 50Hz 120VAC 60Hz 110 120VAC 60Hz Output Voltage 24 VDC 24 VDC Battery Pack 24 VDC 5AH 24 VDC 5AH Panel and Handset Panel and Handset e S by LED color indicator by LCD Input Voltage Maximum Safe Load Capacity 400 Lb 500 Lb 600 Lb Max heightofBoom e e Chattanooga is committed to continuous improvements of our products therefore the specs dimensions and features listed above are for guidance only and subject to change without prior notice Alliance Stand Assist ASSEMBLY Prior to assembly unpack all parts from the shipping carton and check for any missing parts Contact your dealer immediately if a part is
18. s e Batteries 1 year e Weigh Scale 2 years e Stand Aid all components 2 years e Hydraulic pumps 1 year e Slings 6 months The above warranty periods commence on the date of the original customer purchase of the Product from Chattanooga or an authorized Chattanooga dealer In the event of a defect in material or workmanship Company will repair or replace the defective component or the Product at Company s option without charge to customer and Company shall be responsible for shipping expenses for defective Products that are returned by the customer at Company s request and for shipping expenses for repaired or replaced Products that are shipped to the customer To make a warranty claim Company s Customer Care Department or Company s authorized dealer which sold the Product must be notified of the defect during the applicable warranty period above Products may not be returned by customer without a return authorization number Products that are returned for a warranty repair are typically repaired within 30 days from the date the Product is received by Company Company s dealer or Company s certified service center This warranty will be void if repairs or Product modifications are made by anyone other than Company Company s dealer or Company s certified service center This warranty does not cover 1 replacement parts or labor furnished by anyone other than Company Company s dealer or Company s certified service center 2
19. t responsible for any consequence resulting from using non Chattanooga products Use only Chattanooga or approved Chattanooga parts and components Chattanooga is not responsible for any adverse consequence resulting from using parts and components from other manufacturers Service and repair of the patient lift should be performed by Chattanooga or its authorized dealers ONLY Chattanooga is not responsible for any consequence resulting from any unauthorized service or repair Never attempt to lift over the maximum capacity of the liftand any accessories used in conjunction with the lift The maximum safe working load for the lift will be clearly labeled on the lift Alliance Stand Assist DEFINITIONS AND SYMBOLS In this manual the user refers to the patient or resident and may be used interchangeably at different times Caregiver refers to the operator or person who is assisting with the transfer Symbols used in this manual and their meanings Warning Failure to heed this warning may result in damage to the product or serious injury to the operator and or user Important instructions follow Read and understand the instructions in the manual before using the product z gt NOTE Important information concerning the product and or its correct and proper usage follows Alliance Stand Assist PRODUCT FEATURES DESCRIPTION amp APPLICATIONS Chattanooga presents the Stand Assist Series of sit to stand patient lifts a s
20. tion when the battery needs recharging e Audible alarm for low battery capacity e Standard protection class IP 66 e Pulling proof design to lock the AC plug LCD INDICATORS Capacity Function Battery is charged to O Overload i 70 Battery is charged to around 70 Battery Volume 50 Battery Overused Capacity Remaining Pi 50 Charge batteries Battery Low Blink immediately Battery is charged to around 10 Buzzer beeps Only lowering possible Low Batt 10 25 Battery Capacity Remaining O ea Charge batteries immediately Alliance Stand Assist CONTROL BOX FEATURES CONT EMERGENCY STOP FOR PERFORMANCE ELECTRONICS ONLY Emergency stop button interrupts the power supply to the actuators and makes the actuators stop immediately in case of sudden danger Press RED button Turn RED button clockwise i for over 3 days to maintain optimal battery life and performance FEATURES OF POWER PLUG C CLAMP e when the battery is being charged insert this C clamp to secure the power plug WATERPROOF COVER e when control is not being used the cover should be closed to prevent water ingress Waterproof cover REMOVE AND INSERT BATTERY PACK REMOVE press button A and pull out in the up direction Fig 1 INSERT put the battery pack all the way down to position as shown in Fig 2 e slide into direction C while having the release button pressed Fig 3 e make sure the battery pack won t slide back off
21. ts immediately To lubricate put a drop of oil on the following points when the Stand Assist Lift is placed into service and every month thereafter Top of Mast Spreader Hinge and Caster Axles FIRST EVERY 3 coe for improper connection looseness or wear Check the Boom for bending and deflection XT o MAST ma MA a a aaa a Checkthe Mast for bending or deflection x XT BASE amp FOOT PEDAL Check bolts and nuts for looseness x Jo o o x o Check casters and axle bolts for tightness XX Check rubber parts on the casters for defletion x o x 23 Alliance Stand Assist MAINTENANCE AND INSPECTION CONT FIRST EVERY 3 HERI RECEIVED i MONTHS BASE amp FOOT PEDAL Apply grease to caster ball bearings if needed e a ae Check welding joints for rust and crack re ae CLEANING ACTUATOR amp CONTROL BOX ae a a mast Souaeaee ff and mast with pins and key rings Make sure actuator plug into control box is not ee loose SLINGS amp SLING HARDWARE Check sling for wear XxX Check sling hardware every time before use XK 24 Alliance Stand Assist TROUBLESHOOTING GUIDE The following list of encountered problems and solutions will assist you in determining what may be causing your patient lift not to function as designed If you have a problem occurring which is not listed below please contact your dealer or technical support for help Do not attempt to repair or replace compo

Download Pdf Manuals

image

Related Search

Related Contents

Graco 312540E- User's Manual  Westinghouse 42-Inch Instruction Manual  ChondroCelect, Characterised viable autologous cartilage  BENDIX TCH-001-044 User's Manual  Ditec Traffic C  Recherches en Éducation - n°20  SEH PS105  Manual: Jumbo Matic - Implementos Agrícolas Jan S/A  HV-AMP400FN-4+400-D_1  Bosch PT-400 User's Manual  