MODEL AWQ-105 PRO

image

Contents

1. 14 Precaution before the Treatment Pee eee eee eee eee ee eee eee 16 Instructions For Use eee eee eee eee eee ee eee eee ee eee eee eee eee ee eee eee eee ee eee eee 16 ELECTRIC STIMULATION SETTINGS eee eee ee eee eee eee eee eee eee eee eee eee 18 Acupuncture therapy Needles or Electrodes Pee eee eee eee eee eee ee eee eee ree ey 20 Therapy with Electric Stimulation probe amp 21 Precaution After the Treatment eee eee eee eee eee eee eee eee ee eee eee eee eee 22 Disposal of the Device ee eee eee eee eee ee eee rer eee ee eee eee eee eee eee eee eee eee eee ree 22 23 Trouble 24 Storage and Transportation eee eee eee eee eee ere eee eee ree eee ee eee eee 25 Environmental condition eee eee ee eee eee eee eee eee ee eee eee eee eee eee eee eee eee 25 25 Limited Warranty 26 AWQ 105 PRO User Manual Contents of the AWQ 105 PRO Package AWQ 105 PRO ACUPUNCTURE STIMULATOR 5 1 pe Search stimulation probe with handgrip electrode 1 pc Connecting wire 4 assorted colors 4 pcs Pin lead to duck beak clip adaptor 4 pairs DC 1 5V C size
2. LED lights will illuminate 15 AWQ 105 PRO User Manual Precaution before the Treatment INSTALLATION Use the device in an area free from any exposure to water Avoid storing the device in an environment of high humidity and temperature and free from dust or chemical contamination Place the device on a flat and stable surface free from any inclination and vibration Do not operate the device in an environment where flammable substances are present Check to be sure the device is installed with new batteries and check that all LCD displays on the device are working properly Connect all cables correctly Only use accessories that are supplied with the device Check that all output intensity knobs are turned OFF Never operate the device with wet hands to avoid an electric shock Avoid operating the device consecutively for more than 60 minutes in each treatment Instructions For Use POINT DETECTION BASIC PRINCIPLE and DIRECT STIMULATION On the body surface the electric resistance of an acupuncture point is lower than its surrounding area It can be readily located by means of an instrument which is sensitive to resistance changes When the metal probe touches a low resistance point the loudspeaker gives a sound signal and the digital display shows a numerical value Its pitch rises and digital display number increases with the lowering of the electric resistance of the skin The pitch will get h
3. battery not included in package 6 pcs Instruction manual 1 pc The device and its accessories are packed in a sturdy paper carton box Optional accessories Different types of connecting clips are available for selection Crocodile Clip Duck Beak Clip Grasp Clip Micro Hook Clip AWQ 105 PRO User Manual Technical Specifications Output channel 5 Channels including one channel for APD STIM Normal Output channels APD STIM output channel Output Voltage Hi 0 22 5V 0 40V Low 0 10V Output Current Hi 0 45mA 0 80 Low 0 20mA Pulse width X1 50 500 uS adj 260 US fixed X5 10 100 US adj Pulse Rate X1 4 100 Hz Adj 10 Hz fixed X5 5 500 Hz Adj APD amp Stim mode Waveform Pulse width Pulse Rate Mode LED lights up in Stim mode and is flashing when Stim button is pressed Asymmetric Square Wave with a negative spike 50 500 5 adjustable when in X1 position 10 100 5 adjustable when in X5 position 1 100 Hz adjustable when in X1 position 5 500 Hz adjustable when in X5 position Constant pulse rate and width are adjustable from 1 100 Hz amp 50 500 US respectively at X1 position and from 5 500 Hz amp 10 100 5 respectively at X5 position Modulation Preset Pulse Rate and width ramp up and down for 6 seconds per one cycle both pulse rate amp width are adjustable Burst Preset Pulse Ra
4. or decrease the stimulation intensity direct stimulation with preset waveform frequency and pulse width will be provided by the probe head BUZZER ON OFF switch the buzzer will emit sounds during point detection The buzzer can be turned ON or OFF by switching it to OFF position ELECTRIC STIMULATION SETTINGS This section of the instrument produces stimulating impulses which act on acupuncture points or through electrodes for therapeutic purpose Operating instructions Ensure that all batteries are inserted correctly see the battery information section Turn on the device by pressing the POWER ON OFF switch on the left side of the unit Insert connecting wires to corresponding output jacks Connect the wire to acupuncture needles with the clips 18 AWQ 105 PRO User Manual Select the appropriate mode and adjust frequency pulse rate and pulse width as follows iii iv MODE selection switch the desired waveform mode CONT continuous MODULATE dense disperse and BURST Intermittent may be selected by switching this switch Each waveform mode has a corresponding LED that will illuminate when the waveform mode is selected The Frequency and Pulse Width of Ch1 2 and Ch3 4 can be adjusted separately To adjust the frequency and pulse width of Ch1 amp Ch2 Turn clockwise increase or anticlockwise decrease of the lt Frequency Knob CH1 CH2 gt to adjust the frequency and lt Pulse Width knob Ch1 C
5. points The device utilizes a low intensity low frequency pulse multimode generator and either electrodes or alligator clips to deliver electronic stimuli to the patient When using the device with electrodes the electrodes may be placed on the skin at the treatment location When using the device with alligator clips the alligator clips should be connected to acupuncture needles that penetrate the skin at the acupuncture points AWQ 105 PRO User Manual Safety Precautions Warnings Heart Patient Adequate precautionary measures should be considered prior to stimulating patients suspected of having heart disease Current clinical data cannot sufficiently rule out the possibility of adverse results for such patients Carotid sinus Do not stimulate over the carotid sinus nerves especially in patients with known sinus reflex sensitivity Neck Stimulation Severe spasm of the laryngeal and pharyngeal muscles may occur when the electrodes are placed across the neck or the mouth This may be strong enough to close off the airway Cardiac Pacemakers Stimulation will inhibit the output of some demand cardiac pacemakers and therefore it is not recommended for patients with this type of pacemaker Pregnancy The safety of electrical nerve stimulation for use during pregnancy or delivery has not been established Other Electrical nerve stimulation as presently understood is a symptomatic treatment and as such may suppress the progress
6. 2 10 Frequency Adjustable Turn clockwise increase or anticlockwise decrease to adjust the knob for Ch3 amp Ch4 frequency pulse rate of Ch3 amp Ch4 11 Pulse Width Adjustable Turn clockwise increase or anticlockwise decrease to adjust the Pulse knob for Ch1 amp Ch2 Width of Ch1 amp Ch2 12 Pulse Width Adjustable Turn clockwise increase or anticlockwise decrease to adjust the Pulse knob for Ch3 amp Ch4 Width of Ch3 amp Ch4 13 Ch1 Ch2 Frequency LCD Displays the frequency selected for Ch1 amp Ch2 14 Ch3 Ch4 Frequency LCD Displays the frequency selected for Ch3 amp Ch4 15 APD LED It illuminates during point detection and flash continuously which indicates the probe reaches or closer to the acupuncture point 16 ISTIM LED It illuminates when Search to treat push button is depressed for immediate stimulation 17 SEN STIM LCD This LCD displays the detection accuracy when using the point location feature of the device the higher the number the closer the probe is to the acupuncture point It also shows the frequency of stimulation when the Search to treat push button is depressed for immediate stimulation on a point right after its location is found 14 AWQ 105 PRO User Manual 18 Sensitivity Knob Fine tune the sensitivity of the point location feature of the device Adjust the sensitivity knob of the Loc to the appropriate position it t
7. MODEL AWQ 105 PRO ACUPUNCTURE STIMULATOR T E N S 5 channels with digital displays INSTRUCTION MANUAL AWQ 105 PRO User Manual AWQ 105 PRO ACUPUNCTURE STIMULATOR T E N S It is a professional brand new unit with 5 channels It comes up with a modern design and inherits the features of AWQ series acupunctoscope This professional unit provides multiple functions and features which are essential in daily acupuncture practice Before using this device please read the instruction manual carefully as well as read all Safety Precautions and Warnings carefully to ensure the device is operated properly AWQ 105 PRO User Manual Table of Contents Contents of the AWQ 105 PRO Package 3 Technical Specifications eee eee eee eee eee eee eee eee eee eee eee eee 4 Wave Form Pictures eee eee eee eee eee eee eee rere eee eee eee eee eee eee eee eee eee eee eee 6 LRP LUCIO Fates deen ttn 7 Indication and Its Intended Purpose 7 Safety Precautions Warnings 8 Features amp Advantages eee eee eee reer eer ee eee rere eee eee ree eee ee eee eee 9 Battery 11 Indications and Controls eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee 12 Operation of 8
8. and convenient connection to needles search to treat probe for point detection amp electric stimulation and optional accessories which expand its function to a powerful T E N S device as well Search Stim Probe Search to treat push button 10 AWQ 105 PRO User Manual Battery Information Your stimulator requires 6 pcs 1 5V battery not included use C type LR14 or its equivalent type of battery The batteries can be replaced by opening the battery door on the bottom side of the device Slide the battery door off remove the old batteries and insert the new batteries Insert the batteries correctly into the battery compartment by making sure the direction of the positive and negative poles of each battery match the illustrations inside the battery compartment Be sure that each battery used is a fresh new battery Once the batteries have been properly seated inside the battery compartment the battery door should be closed Battery life expectancy is between 4 to 20 treatment hours depending on current settings and treatment length Many short treatments yield longer battery life than fewer longer treatments To avoid battery leakage remove the battery if the device will not be used for extended period of time A specially designed battery compartment is used in this device to ensure easy and proper insertion as well as securely hold the batteries in place u ge xT N AN N Special design of batte
9. charged to the customers on the basis of rates which are available on request Except for personal injury no liability is held in either tort or contract for any loss or damage direct consequential or incidental arising out of the use misuse or inability to use this product Serial No located on the rear side of the device next to the battery compartment door Date Purchased Customer Please record this information 26
10. h2 gt to adjust the pulse width To adjust the frequency and pulse width of Ch3 amp Ch4 Turn clockwise increase or anticlockwise decrease of the lt Frequency Knob CH3 CH4 gt to adjust the frequency and lt Pulse Width knob Ch3 Ch4 gt to adjust the pulse width Ch1 Ch2 Frequency LCD will display the frequency for Ch1 amp Ch2 when selected Ch3 Ch4 Frequency LCD will display the frequency for Ch3 amp Ch4 when selected A X1 X5 multiplier switch is provided for selecting the frequency and pulse width of the output in normal range of frequency at 1 100 Hz pulse width at 50 500us or multiplies 5 i e 5 500 Hz frequency and 10 100 us pulse width The channel output intensity for all channels is adjusted by rotating the knob of Ch1 4 INT respectively in clockwise direction it will first turn on the output channel and then increase the current output or vise versa Two ranges High Low of output intensity are provided If maximum intensity in Low range is not sufficient switch it to High range for stronger intensity output Select the timer duration of treatment for 15 30 minutes or continuously by switching the Timer switch to the selected position 19 AWQ 105 PRO User Manual Acupuncture therapy Needles or Electrodes Turn all intensity knobs to OFF position and slide the HI LOW switch to the Hi or Lo position as required HI for electrode stimulation or LOW for needle st
11. igher and the numerical value will increase as the probe moves closer to the acupuncture point With this instrument and method of detection acupuncture points located usually produce a strong sensation and reaction to electric stimulation and will achieve the best therapeutic results 16 AWQ 105 PRO User Manual Ensure that all batteries are inserted correctly see the battery information section Turn on the device by pressing the POWER ON OFF switch on the left side of the unit Insert the Search stimulation probe with handgrip electrode into the APD STIM socket Adjust the sensitivity by turning the sensitivity knob When the air humidity is wet turn the knob anti clockwise to decrease sensitivity If it is too dry turn the knob clockwise to increase sensitivity Sensitivity knob Touch the hand grounding pole with the probe tip the buzzer will give a sharp sound signal the digital display will show increasing numbers and the ADP LED will flash continuously indicating that the point location capabilities of the device are working properly Let the patient grasp the hand pole in one hand Search for the acupuncture point with the probe tip on the patient s skin in the appropriate anatomical location When doing this the probe should be held in a vertical position perpendicular to the surface of skin The pressure and duration in touching the point with the probe should be kept constant If too much pressure is ap
12. imulation Apply acupuncture needles or electrodes to patient Insert alligator type connecting wires into 1 2 3 or 4 sockets and fasten their clips to the acupuncture needles or electrodes The number of channels used depends on the number of needles or electrodes used in each treatment CH4 CH3 SEN STIM CH2 CH1 Set the desired mode frequency pulse rate pulse width and operation time with the switches as mentioned above Adjust the intensity of each channel to appropriate setting The intensity current of each output channel can be adjusted from 0 mA to 20 mA when the HI LOW switch is switch to the LOW range or 0 to 45 mA when the HI LOW switch is switch to the position Rotating the INTENSITY knob will increase or decrease the intensity 20 AWQ 105 PRO User Manual Therapy with Electric Stimulation probe amp Settings Use the Search Stimulation probe point detector with hand pole by placing the probe tip onto the acupuncture point or another selected part of the skin STI in Search to treat mode il APD in Search mode Search Stim Probe Search to treat ME a Insert the Search stimulation probe with handgrip electrode into the APD STIM socket To stimulate the acupuncture point with the probe tip make sure the probe tip is touching the treated point or area e and press the search to treat button on the probe Socket for Search Stim The i
13. ntensity is regulated by the STIM Channel probe direct stimulation with fixed waveform frequency and pulse width will be provided by the probe head NOTE The output intensity of Stim is in High range intensity 0 80 mA and frequency is fixed at 10 Hz SEN STIM LCD will display the Hz during direct stimulation 21 AWQ 105 PRO User Manual Precaution After the Treatment After use turn all output knobs to OFF position and press the POWER SWITCH to turn off the device Keep the device clean and Store it in a clean and dry place Explanation of symbols The following symbols are used in this device 6 A A a Read Instruction Year of Part Manufacture 5 N 0197 Serial Number Authorised CE 0197 mark of conformity Representative in The device is complied with the Essential Requirements for the European Medical Devices according to EC Directive 93 42 EEC Community Disposal of the Device Please follow the local environmental requirements when disposing of the device or its accessories or packing materials 22 AWQ 105 PRO User Manual Maintenance Maintenance of the stimulator is limited to cleaning the stimulator battery contacts and replacing lead wires when they become old The stimulator operates on DC 1 5 volt x 6 pcs Batteries not included When adequate stimulation can no longer be maintained change the batteries Remove the old batteries and replace them
14. ocal authorized dealer or to the manufacturer as listed in this manual for repair or service Problem Solution Indicator lights up but device Check control settings are they set to values prescribed as does not function properly desired IL Check if frequency control is turned on not too low or at zero setting II Are needles or electrodes firmly fixed and in proper position IV Check connecting wire Make sure all alligator clips are firmly connected to the needles or electrodes V Replace connecting wire with another to insure the wire is not broken Low battery lamp illuminates Replace batteries with new ones No indicators light up Replace batteries with new ones 1 Battery replacement Batteries should be replaced whenever sufficient stimulation cannot be maintained 2 Care of Device The stimulator and connecting wires should be kept clean The stimulator should not be immersed in any liquid Avoiding rough use will help prevent premature failure 24 AWQ 105 PRO User Manual Storage and Transportation Store the device in a dry location free from dust and contamination where the temperature remains fairly constant and within the range of 10 to 50 C 50 F to 122 F Do not drop mishandle or expose to temperature or humidity extremes lt outside the range of 10 C to 50 C 50 F to 122 F 15 95 RH non condensing gt Do not use if the device malfunctions o
15. of pain which would otherwise serve as a protective influence on the outcome of a disease process The potential for physical and or psychological dependence upon nerve stimulation as a means of relieving pain has not yet been determined It has been noted that some patients find the sensation of electrical stimulation extremely unpleasant and should probably be excluded from further use of the stimulator Do not apply electrical nerve stimulation when pain syndromes are undiagnosed until etiology is established Do not apply electrical nerve stimulation current transcerebrally Electrical nerve stimulation devices should be used only under the continued supervision of a physician Electronic monitoring equipment such as EKG monitors and EKG alarms may not operate properly when electrical nerve stimulation is in use Avoid use in post operative recovery rooms when a heart monitor is on Keep out of reach of children AWQ 105 PRO User Manual Features amp Advantages i 5 outputs including an output exclusively for the use of search stimulation probe for point detection and immediate stimulation Frequency and Pulse Width are adjustable amp separated into two groups Group 1 for Ch 1 amp 2 and Group 2 for Ch 3 amp 4 Instead of one Frequency one Pulse Width for all channels 2 large LCD each measures 15X30 mm for displaying precise frequency of each group and one LCD 15x30 mm fo
16. plied to the skin the electrical resistance measured by the device effectively becomes lower and the device will behave as though an acupuncture point was found When the probe touches an acupuncture point the buzzer will buzz the LED of the ADP will flash continuously at the same time the digital display will show a higher number and a very light electric stimulus may be felt by the patient Since an acupuncture point is very small it should be located with patience and care If the electrical resistance of the general surface of the skin is low or when the skin is wet or sweating which would lower the skin resistance the buzzer would give a sound signal on 17 AWQ 105 PRO User Manual the skin surface wherever the probe touches When this happens dry the skin with a towel maneuver the probe with lighter pressure and readjust the sensitivity setting to a lower level so that the point to be located can be discriminated from the surrounding areas of the skin There is no definite standard for the adjusting the sensitivity of the device and how much pressure to apply to the skin surface the appropriate sensitivity will be found through experience WARNING Never use the probe on open cuts or wounds If the located point is to be stimulated by the probe tip make sure the probe tip is touching the located point and press the search to treat button on the probe adjusting the intensity setting of STIM INT Channel will increase
17. r has been damaged in any manner There is generally no restricted shelf life of the device Environmental condition Temperature Humidity Pressure 10 C to 40 C 30 75 RH During Use 700 1060 hPa 50 F to 104 F non condensing 10 C to 50 C 15 95 RH Storage 700 1060 hPa 50 F to 122 F non condensing 10 C to 50 C 15 95 RH Transportation 700 1060 50 F to 122 non condensing Cleaning DO NOT IMMERSE THE STIMULATOR IN ANY CLEANING SOLUTION The stimulator should be periodically wiped clean using a damp cloth and a solution of mild soap and water Use of other cleaning solutions may damage the housing The wires should be wiped clean with a cloth dampened with a mild soap solution and then wiped dry AWQ 105 PRO User Manual Limited Warranty This warranty is in lieu of any other warranty expressed or implied This AWQ 105 PRO ACUPUNCTURE STIMULATOR T E N S is warranted to the initial purchase purchases and to no other person against any defects in material and workmanship for a period of one year from the date of purchase If the stimulator is found to be defective within the warranty period it will be repaired or replaced if returned prepaid to an authorized service center This warranty does not cover damage caused by rental misuse negligence accident abuse alteration or modification of the device Repairs after the warranty period will be made and
18. r showing sensitivity during acup point detection This LCD also displays the frequency fixed at 10 Hz during direct stimulation by the search stimulation probe 3 different Mode Cont continuous Modulate and Burst High amp Low output levels for selection X1 and X5 multiplier of pulse rate and width for adjustments Timer selection for different period of treatment session Continuous 15 minutes and 30 minutes Buzzer ON OFF for sound control during point detection Music for timer alert and error warning The Search Stim Probe includes a Search to treat push button selector which when depressed allows immediate stimulation on a point right after its location is found Search to treat push butt n in Search to treat mode il APD in Search mode AWQ 105 PRO User Manual Improved design of accessories CE marked plug connecting wires 4 assorted colors with pin leads so that each can be fitted to alligator clip adaptors or different types of needle connectors such as Grasping clips duck beak clips micro hook clips or crocodile clips A large range of optional accessories is available ii Optional accessories Fa Se N Pa N 2 N 4 N NG 7 f 0087 A Crocodile Clip Duck Beak Clip Grasp Clip Micro Hook Clip iii Alligator clip wires which provide very sturdy
19. ry compartment for easy and proper insertion of batteries 11 AWQ 105 PRO User Manual Indications and Controls AWQ 105 PRO 30 MIN 12 AWQ 105 PRO User Manual AWQ 105 PRO Peat Tha Pet ef 190 whe 99 em ine 21 AWQ 105 PRO User Manual Operation of Controls No Descriptions Functions 1 Model Name AWQ 105 PRO 2 Power ON OFF switch Turns the power ON or OFF 3 BUZZER ON OFF switch Turns the buzzer sound during point detection ON or OFF 4 HI LOW switch For selection of HI range or LOW range output intensity 5 ERROR LED When turn ON the unit if all channels setting do not reset to OFF output Red LED lights up amp music sounds The unit resumes work after reset all channels to OFF position 6 LOW BATTERY LED When the battery voltage drops to under 7 volts this LED will illuminate indicating that new batteries need to be installed 7 Mode selection switch To select different modes of waveform CONT continuous MODULATE dense disperse and BURST Intermittent 8 Mode indication LED The selected waveform mode LED will illuminate indicating which waveform is currently active 9 Frequency Adjustable Turn clockwise increase or anticlockwise decrease to adjust the knob for Ch1 amp Ch2 frequency pulse rate of Ch1 amp Ch
20. set pulse rate and width ON 3 seconds OFF 3 seconds Pulse Width range at different Pulse Rate settings x1 x5 Pulse Rate 1 100 Hz 5 500 Hz Pulse Width 50 500 US 10 100 uS AWQ 105 PRO User Manual Introduction AWQ 105 PRO ACUPUNCTURE STIMULATOR 5 is a professional brand new unit with 5 channels It comes up with a modern design and inherits the features of AWQ series acupunctoscope This professional unit provides multiple functions and features built in with 5 outputs channels and with separated groups for adjusting frequency and pulse width The acupuncture point detection is with sensitivity control and the Search Stim Probe includes a Search to treat push button selector which when depressed allows immediate stimulation on a point right after its location is found 2 large LCD each measures 15X30 mm for displaying precise frequency of each group and one LCD 15x30 mm for showing sensitivity during acup point detection and the frequency during direct stimulation It provides 3 different Mode Cont continuous Modulate and Burst and a High Low selection of output Lastly a timer is equipped for selection of different period of treatment session and a Buzzer ON OFF control for shutting down the sound during point detection Indication and Its Intended Purpose The AWQ 105 PRO ACUPUNCTURE STIMULATOR T E N S is intended to use in applying electronic stimulation to acupuncture
21. te and width lt ON gt 3 seconds lt OFF gt 3 seconds both pulse rate amp width are adjustable AWQ 105 PRO User Manual Timer Display Low battery lamp Power supply DC Power Jack Output socket Error LED Weight Size Accessories Cont 15 min 30 min 3 LCD each measure 15x30 mm The LED lights when voltage drops to 7V C type LR 14 batteries x 6 pcs Or DC 9V 500mA adaptor optional Use correct adaptor comply with standard and spec 6 3mm x 2mm centre pin 1 5mm diameter socket Red LED lights up amp music sounds when turn ON the unit without reset all previous channels setting to zero output The unit resumes work after reset all channels to zero position 726g main unit only 255 x 175 x 75 mm main unit only Search stimulation probe with handgrip electrode Connecting wire 4 assorted colors 4 pcs Pin lead to duck beak clip adaptor 4 pairs Instruction manual 1 pc AWQ 105 PRO User Manual Wave Form Pictures FC A CONTINUOUS Pulse Rate FC Continuous Frequency Pulse Width range at different Pulse Rate settings x1 x5 Pulse Rate 1 100 Hz 5 500 Hz Pulse Width 50 500 uS 10 100 5 Preset pulse rate and width ramp up and down for 6 seconds per one cycle Pulse Width range at different Pulse Rate settings X1 x5 Pulse Rate 1 100 Hz 5 500 Hz Pulse Width 50 500 uS 10 100 uS C BURST J Hb Pre
22. urns with increasing sensitivity during location of acupuncture point 19 Timer switch Select the timer duration for 15 30 minutes or continuously 20 Search Stimulation probe This jack is used for attaching the search stim probe to the device insertion jack socket 21 Output jacks socket For insertion of output connecting wires 22 Output Intensity control Rotating the knob clockwise will first turn the output channel and then knob increase the current output or vise versa 23 1 X5 switch X1 X5 multiplier switch to select the frequency and pulse width of the output in normal range of frequency at 1 100 Hz pulse width at 50 500 S or multiplies 5 5 500 Hz frequency and 10 100 yw S pulse width 24 Battery compartment door The battery compartment door can be removed to expose the battery compartment The device uses 6 pcs of 1 5 V C type LR14 batteries 25 DC 9V Input This input can be used to plug in an external 9V 500mA input source 26 Output Intensity control Output Intensity control knob is applied only at APD amp STIM mode when knob for search stim operating stimulation at APD amp STIM mode rotating the knob clockwise probe turns on the output channel and then increase the current output intensity or vise versa of the search stim probe 27 Output Intensity LED When turning clockwise output intensity knob at ON position output Lights intensity
23. with new batteries The device will provide sufficient levels of stimulation only if the batteries are properly installed To avoid battery leakage remove the batteries if the device will not be used for an extended period of time To ensure the device works properly please check the device on a periodic basis with the following PARTICULARS DESCRIPTIONS PROCEDURES Appearance check for broken or damaged parts By visual inspection Il check all LCD displays make sure all digits are legible Operation I check all LEDs light when device is turned on By operating the check that the buzzer sounds device Ill check that the device functions properly Accessories check for broken damaged parts of all accessories By visual inspection including the search stim probe Il check all connecting wires for any broken or damaged parts Safely Turn on one of the intensity knobs press POWER By operating the precaution SWITCH ON so an error occurs Verify that the device device resets to normal operation when turn OFF all intensity knobs no error light no music to be function 23 AWQ 105 PRO User Manual Trouble Shooting If your device appears to be functioning improperly check the procedures shown below to determine what may be wrong If none of these measures correct the problem the device should be serviced Do not attempt to repair the device by yourself Return the device to your l