Operating manual
Contents
1. If the device is set to the resistance measurement function Q check that the display indicates A P O for auto power off SCAN for automatic search for the quantity to be measured dashes to indicate search in progress Withdraw the two leads from their compartment and establish a good contact between the metallic parts of the two test probes the value measured and displayed must change from to 0 00 accompanied by the message on the value displayed may not be exactly zero but must remain very close to zero The instrument is operational when the various states described above have been checked and are correct The rotary switch can then be set to the desired function or to OFF to switch the instrument off 2 3 Deactivating Auto Power Off In order to extend the life of the batteries the device switches itself off automatically approximately 10 minutes after it is switched on if no key or change of function has been activated 6 The device warns that it is about to switch off by emitting 3 series of 2 audible beeps and if no action is taken then emits a long beep and switches itself off When the multimeter has switched itself off automatically it can be reactivated by any action on a key or on the rotary switch Automatic switching off can be deactivated by holding down the Hz MAX MIN or R H key while switching on To eliminate the slight residual drain of the batteries in
2. The 600mV range can be used only in the manual range change mode by repeated presses on the RANGE key The other ranges can be used in either the manual or the automatic range change mode Withdraw the test probes and leads from their compartment Switch the instrument on in the V setting Apply the metallic part of the black test probe on the part of the circuit to be tested assumed closest to the earth potential Apply the metallic part of the red test probe on the partof the circuit to be tested assumed farthest from the earth potential In the SCAN mode if the voltage to be measured is greater than 0 5V the multimeter automatically determines if an AC or DC measurement must be made If necessary press the SEL key briefly to force an AC voltage measurement display m of the Ne symbol instead of mm m Read the measurement result after stabilization 2 6 Frequency measurement Hz optimum safety measurements of voltages between 300V and 600V must be made only on category Ill installations or on circuits of which the possible overvoltage levels are known to be less than those of category Ill The instrument has 3 measurement ranges 600 0Hz 6 000kHz and 60 00kHz Measurement of the frequency of the voltage measured start the AC voltage measurement method then press the Hz key display of the Hz symbol instead of V Read the measurement result after stabilization 2 7 Resis
3. ITALIA Amra SpA Via Sant Ambrogio 23 25 20050 Bareggia di Macherio MI Tel 039 245 75 45 Fax 039 481 561 OSTERREICH Chauvin Arnoux GmbH Slamastrasse 29 2 4 1230 Wien Tel 01 61 61 961 0 Fax 01 61 61 961 61 SCANdINAVIA CA Matsystem AB Box 4501 SE 18304 TABY Tel 46 8 50 52 68 00 Fax 46 8 50 52 68 10 SCHWEIZ Chauvin Arnoux AG Einsiedlerstrake 535 8810 Horgen Tel 044 727 75 55 Fax 044 727 75 56 UNITEd KINGdOM Chauvin Arnoux Ltd Waldeck House Waldeck Road Maidenhead SL6 8BR Tel 01628 788 888 Fax 01628 628 099 MIddLE EAST Chauvin Arnoux Middle East P O BOX 60 154 1241 2020 JAL EL dIB Beirut LEBANON Tel 01 89 04 25 Fax 01 89 04 24 CHINA Shanghai Pu Jiang Enerdis Instruments Co Ltd 3 F 3 rd Building N 381 Xiang de Road 200081 SHANGHAI Tel 86 21 65 21 51 96 Fax 86 21 65 21 61 07 USA Chauvin Arnoux Inc d b a AEMC Instruments 200 Foxborough Blvd Foxborough MA 02035 Tel 508 698 2115 Fax 508 698 2118 RANGE key Your multimeter has a range change function that is normally automatic but can be made manual After the quantity to be measured is chosen manually SEL key the default mode is automatic range change the AUTO message is then displayed Briefly press the RANGE key to change to manual mode the AUTO message is then replaced by MANU Successive brief presses are used to reach the desired range Ho
4. 5 10 3 2 4 Continuity measurement 600 0 The buzzer is actuated Q 0 10 at values lt 300 approx 0 8V 3 2 5 Diode test Measurement uncertainties No load voltage 2 000V 0 001V 1 0 5 approx 2 0V 3 2 6 Capacitance measurement Range 6 000nF 60 00nF 600 0nF E 2 5 5 60 00pF 600 0pF 6 000mF 60 00mF 3 2 7 Frequency measurement 6 000kHz 0 001kHz 0 1 3 60 00kHz 0 01kHz Sensitivity Minimum input signal sine wave Range 11 3 3 Environmental conditions Environmental conditions Temperature 0 C to 50 C 10 C to 60 C 32 F 122 F 14 F 140 F Relative humidity RH lt 70 without condensation lt 80 without condensation Altitude lt 2 000m 3 4 Constructive characteristics up to 10 000m Housing Rigid polycarbonate shell with moulded elastomer covering Screen LCD display Dimension L 41 x H 18mm Dimension H 130 x L 81 x P 24mm Weight 150g without battery 3 5 Power supply Batteries 2x1 5V AAA Mean battery life gt 250 hours Auto power off delay After approximately 10 minutes without action on the switch and or on the keys 3 6 Compliance with international standards Compliant with standards IEC61010 1 and IEC61010 2 033 300V CAT Ill Class 2 double insulation Electric safety Protection class Electromagnetic Compliant with standard EN61326 1 compatibility Level of protection of fie the housing Housing IP40 according to the stand
5. ohms Withdraw the test probes and leads from their compartment Switch the instrument on and set to Q on IHH F Apply the metallic tips of the test probes to the terminals of the resistance or of the circuit to be tested In the SCAN mode the multimeter can automatically determine and select the continuity mode when the resistance measured remains low otherwise press the SEL key briefly until the continuity symbol n appears Read if necessary the measurement result after stabilization 8 2 9 Diode measurement Q resistance continuity diode test and capacitance measurements must be made only on circuits completely disconnected from any power supply and after any capacitors have been discharged The instrument indicates the voltage across the terminals of the semiconducting junction Withdraw the test probes and leads from their compartment Switch the instrument on and set to pA F Apply the metallic tips of the test probes to the terminals of the diode or circuit to be tested n the SCAN mode the multimeter can automatically determine and select the diode test mode when the voltage measured is less than approximately 0 8V otherwise press the SEL key briefly until the diode test symbol pf appears The test probes must be applied to the terminals of the diode or junction to be tested according to the following polarities 1 forward direction conducting apply the blac
6. 2 leads 1 red and 1 black each terminated by a test probe of the same colour The leads and test probes are not detachable they are permanently connected and when not in use can be stowed in the compartments provided for them on the right side of the device The tips of the test probes once taken out of their compartments by pressing on them with a finger allow voltage resistance continuity diode test and capacitance measurements The test probes have physical guards that show the operator where the hand grip part ends beyond which the fingers must not be placed 5 2 USE 2 1 Commissioning Place the batteries supplied with the device as follows 1 Using a cross headed screwdriver unscrew the screw of the compartment cover on the back of the housing and open the cover 2 Place the 2 batteries in their compartment with the correct polarities 3 Close the cover and screw it back to the housing 2 2 Starting up and check of operation We recommend performing this procedure at the time of commissioning and if the device is used only occasionally each time it is used Hold the HOLD key down and switch the instrument on by turning the switch from OFF to Q Keep the HOLD key pressed and check that the various symbols and segments are correctly displayed the 1st digit on the left may be a 6 or an 8 Releasing the HOLD key restores the normal display for the function chosen
7. Multimetrix DMM 13 Multimeter SELECT RANGE HOLD CHAUVIN ARNOUX 190 rue Championnet 75876 PARIS Cedex 18 FRANCE Tel 33 1 44 85 44 85 Fax 33 1 46 27 73 89 info chauvin arnoux fr Export Tel 33 1 44 85 44 38 Fax 33 1 46 27 95 59 export chauvin amouxfr You have just acquired a DMM13 multimeter and we thank you for your confidence PRECAUTIONS FOR USE This device is compliant with safety standards IEC61010 1 and 61010 2 033 for voltages up to 300V in category III or 600V in category Il at an altitude below 2 000m indoors with a degree of pollution of not more than 2 These safety instructions are intended to ensure the safety of persons and proper operation of the device If the device is used other than as specified in this data sheet the protection provided by the device may be impaired The operator and or the responsible authority must carefully read and clearly understand the various precautions to be taken in use f you use this instrument other than as specified the protection it provides may be compromised thereby endangering you Do not use the instrument in an explosive atmosphere or in the presence of flammable gases or fumes Do not use the instrument on networks of which the voltage or category exceeds those mentioned Do not exceed the rated maximum voltages between terminals or with respect to earth Do not use the instrument if it seems to be damag
8. ard IEC60529 12 inuse instorage 4 MAINTENANCE The instrument has no parts that can be replaced by personnel who are not trained and approved Any non approved repair or other work or replacement of a part by an equivalent may severely compromise safety 4 1 Cleaning Disconnect the unit completely and turn the rotary switch to OFF Use a soft cloth dampened with soapy water Rinse with a damp cloth and dry rapidly with a dry cloth or forced air Dry perfectly before putting back into use 4 2 Replacement of the batteries The symbol indicates that the batteries are spent When this symbol appears on the display unit the batteries must be replaced The measurements and specifications are no longer guaranteed To replace the batteries proceed as follows 1 Disconnect everything connected to the device 2 Set the switch to OFF A using a cross headed screwdriver unscrew the screw of the battery compartment cover on the back of the housing and open the cover 4 Always replace both batteries 5 Close the cover and screw it back down 4 3 Repair Return the instrument to your distributor for any work to be done whether under the warranty or not If you have to ship the instrument it is best to use its original packaging and to state as clearly as possible in a note attached to the equipment the reasons for the transfer 5 WARRANTY The equipment is warranted against defects of materials or wo
9. confirmed by an audible signal The functions are described in the table below o e AC or DC voltage measurement Frequency measurement VIHz Automatic detection of the AC or DC voltage to be measured o1 Resistance measurement continuity measurement with buzzer or diode test and capacitance measurement p Automatic detection of the quantity to be measured When switched on the multimeter is in SCAN mode automatic detection and analyzes the input signals to determine what quantity is to be measured In this mode the RANGE HOLD and MAX MIN keys are deactivated 1 2 The keypad Here are the functions of the keys of the keypad SEL SELECT key The SEL key is used to access the functions of the RANGE HOLD and MAX MIN keys and choose the quantity to be measured manually stops the SCAN mode Press the SEL key repeatedly to obtain the following functions according to the setting of the rotary switch DC voltage AC voltage Functions obtained SCAN gt Vv Z V C Resistance measurement continuity measurement diode test capacitance measurement Functions obtained SCAN gt Q gt e1 gt gt gt 4k 06 2014 694242A02 Ed 5 DEUTSCHLAND Chauvin Arnoux GmbH StraRburger Str 34 77694 Kehl Rhein Tel 07851 99 26 0 Fax 07851 99 26 60 ESPANA Chauvin Arnoux Ib rica S A C Roger de Flor N 293 Planta 1 08025 Barcelona Tel 902 20 22 26 Fax 934 591 443
10. ed incomplete or poorly closed Before each use check the condition of the insulation on the leads housing and accessories Any item of which the insulation is deteriorated even partially must be set aside for repair or scrapping Use leads and accessories rated for voltages and categories at least equal to those of the instrument If not an accessory of a lower category reduces the category of the combined multimeter accessory to that of the accessory Observe the environmental conditions of use Do not modify the instrument and do not replace components with equivalents Repairs and adjustments must be done by approved qualified personnel Replace the battery as soon as the symbol appears on the display unit Disconnect all leads before opening the battery compartment cover Use personal protective equipment when conditions require Keep your hands away from the unused test probes of the device When handling the test probes crocodile clips and accessories keep your fingers behind the physical guard Disconnect the test probes from the measurement circuit before changing functions MEASUREMENT CATEGORIES CAT Il Circuits directly connected to the low voltage installation Example power supply to electro domestic devices and portable tools CAT Ill Power supply circuits in the installation of the building Example distribution panel circuit breakers machines or fixed industrial devices CAT IV Circ
11. k probe to the cathode and the red probe to the anode Read the measurement result after stabilization see figure A the threshold voltage of a silicon diode is between 0 5V and 0 7V it is between 0 2 and 0 3V for a germanium diode High voltage diodes some Zener diodes and LEDs cannot be tested by this method 2 reverse direction non conducting apply the black probe to the anode and the red probe to the cathode Read the measurement result after stabilization see figure B a diode or junction in good condition causes the display of OL With some types of diode a measured value outside the stated voltage ranges does not necessarily mean that the diode or junction is defective 3 CHARACTERISTICS 3 1 General conditions Peak factor of the applied AC signal 3 2 Characteristics under the conditions of use Sampling rate approximately 3 measurements per second The uncertainties are expressed in x L y digit 3 2 1 DC voltage measurement Range Resolution Measurement uncertainties Input resistance 600 0mV 6 000V 0 001V 1 0 3 approx 10 MQ 60 00V 0 01V 600 0V 3 2 2 AC voltage measurement Measurement uncertainties Range Resolution Input resistance 45 60Hz 60 400Hz 6 000V 0 001V approx 10 MQ 60 00V 0 01V 1 5 5 600 0V 3 2 3 Voltage Measurement Range Resolution maximum uncertainties applicable 1 0 5 600V Resistance measurement 6 000MOQ 0 001MQ 2 0 5 60 00MQ 0 01MQ 3 0
12. ld the RANGE key down for 2 seconds to return to the automatic range change mode the MANU message reverts to AUTO HOLD key In the HOLD mode the device freezes the display of the last value measured Pressing the HOLD key briefly during a measurement freezes the display the message is then displayed A second brief press on the HOLD key is used to return to the normal measured value display refresh mode the message disappears from the display unit MAX MIN key The device records the maximum and minimum values of the measurements made The automatic range change does not change the range when this mode is entered if the maximum values exceed the display range indicated by the OL message the appropriate higher range must be chosen manually before reactivating the MAX MIN mode Successive brief presses on the MAX MIN key produce in order the following actions qst press the device records and displays the maximum value measured The AUTO message is replaced by MANU and the MAX message is displayed 2nd press the device records and displays the minimum value measured the MAX message is replaced by MIN 3rd press the device records the maximum and minimum values measured simultaneously it is the value currently measured that is displayed the MAX and MIN messages blink simultaneously on the display unit Additional presses serve to display the values recorded one by one by re
13. producing the actions of the previous successive presses Press the MAX MIN key for two seconds to exit from the mode the MANU MAX and MIN messages are replaced by AUTO indicating that the device once again changes range automatically Hz Hertz key When the switch is set to VHz a brief press on the Hz key displays the frequency of the voltage being measured A 2nd press restores the display of the voltage being measured 1 3 The display unit SCAN AUTO MANUCHD A P0 gt C2330 SE Us caBIG I Lins SCAN Automatic search for the quantity to be measured AUTO Automatic range change MANU Manual range change Data Hold Freezes the display en Continuity test function APO Auto Power Off automatic switching off activated Diode test function Display of the maximum and minimum values DE DC measurement and display AC measurement and display Presence of hazardous voltage above 30V E Display of a negative value Low battery indicator battery must be replaced OoOo E O Hertz unit of frequency ohm kilo ohm Megohm unit of resistance Volt unit of voltage MIQyF milliFarad nanoFarad microFarad units of capacitance The OL message OverLoad indicates an overshoot of the measurement or display capacity A display of dashes indicates waiting for the determination and selection of the quantity to be measured 1 4 The leads and test probes The multimeter is equipped with
14. rkmanship in accordance with the general terms of sale During the warranty period 1 year the instrument must be repaired only by the manufacturer who reserves the right to choose between repairing it or replacing it entirely or partially 13 If the equipment is sent back to the manufacturer carriage is paid by the customer The warranty does not apply in the following cases 1 Inappropriate use of the equipment or use with incompatible equipment Modifications made to the equipment without the explicit permission of the manufacturer s technical staff 3 Work done on the device by a person not approved by the manufacturer 4 Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user s manual 5 Damage caused by shocks falls or floods 6 DELIVERY CONDITION The DMM13 multimeter is delivered in a blister pack with 2 1 5V batteries a user manual in several languages 14
15. tance measurement Q resistance continuity diode test and capacitance measurements must be made only on circuits completely disconnected from any power supply and after any capacitors have been discharged The instrument has 6 measurement ranges 600 0Q 6 000kQ 60 00kQ 600 0kQ 6 000 MQ and 60 00 MQ All of the ranges can be used with either automatic or manual range change Withdraw the test probes and leads from their compartment Switch the instrument on and set to Qm gt H F Apply the metallic tips of the test probes to the terminals of the resistance or of the circuit to be tested In the SCAN mode the multimeter automatically determines the nature of the quantity to be measured f necessary press the SEL key briefly to change to the desired measurement mode Read the measurement result after stabilization NB Touching the elements or circuit during the measurement may lead to erroneous results For resistance measurements on nonlinear circuits it may be necessary to select the measurement mode and the range to be used manually 2 8 Continuity measurement Q resistance continuity diode test and capacitance measurements must be made only on circuits completely disconnected from any power supply and after any capacitors have been discharged The instrument indicates continuity by emitting a steady sound when the resistance measured is less than 30Q The value displayed is the value measured in
16. the sleep mode it is always best to switch the device OFF 2 4 DC voltage measurement V EE E 6 optimum safety measurements of voltages between 300V and 600V must be made only on category III installations or on circuits of which the possible overvoltage levels are known to be less than those of category Ill The instrument has 4 measurement ranges 600mV 6 0V 60V 600V The 600mV range can be used only in the manual range change mode by repeated presses on the RANGE key The other ranges can be used in either the manual or the automatic range change mode Withdraw the test probes and leads from their compartment Switch the instrument on in the V setting Apply the metallic part of the black test probe on the assumed negative part of the circuit to be tested Apply the metallic part of the red test probe on the assumed positive part of the circuit to be tested Read the measurement result after stabilization Display of the sign in front of the numerical value indicates that the value measured is negative the test probes are reversed with respect to the polarity of the voltage 2 5 AC voltage measurement V Ny 6 optimum safety measurements of voltages between 300V and 600V must be made only on category III installations or on circuits of which the possible overvoltage levels are known to be less than those of category Ill The instrument has 4 measurement ranges 600mV 6 0V 60V 600V
17. uits supplying the low voltage installation of the building Example power feeders counters and protection devices For best results from your instrument read these operating instructions carefully comply with the precautions for use Meaning of the symbols used Risk of danger The operator agrees to refer to these instructions whenever this danger symbol appears Battery The CE marking indicates conformity with European directives Double insulation or reinforced insulation Selective sorting of wastes for the recycling of electrical and electronic equipment within the European Union In conformity with directive WEEE 2002 96 EC this equipment must not be treated as household waste xare DC Direct current EEE 4 AC Alternating current PaF AC and DC Alternating and direct current la Earth Q Instructions that must be read and understood 1 PRESENTATION This is an instrument for the measurement of electrical quantities that groups the following functions AC or DC voltage measurement Frequency measurement Resistance measurement continuity measurement with buzzer diode test Capacitance measurement Automatic determination of the quantity to be measured from the setting of the switch Test probe 1 red and 1 black 1 1 The switch The switch has three positions To access the various functions set the switch to the corresponding positions Each active position is