Home
850MT User Manual here - VIGILANT Generation 6 Range
Contents
1. JdHz Fast DIN iHe BANS LE EE X E ms At diz Fast filter Yodalarm Con Bell 7 T hea PM ly Fig 26 Tone Selection Screen 5 6 850EMT Settings SET TONES Press the icon on the Main Screen to navigate to the AUDIO FOR 850EMT Settings screen as shown in Figure 27 TOUCH SCREEN BUTTON AUDIO FOR IR COMMS SUCCESSFUL Please select an oj SET TIME POWER SETTINGS UEM Fig 29 Set Tones LCD BRIGHTNESS 1 Negative option is highlighted in red 2 Positive options is highlighted in green aud icta EE When you select the options by pressing it on the Main menu screen a tone emanates from the E e i 850EMT To enable disable this tone select Yes No respectively When the 850EMT successfully communicates with the detector via the IR Mode a tone emanates from it To enable disable this tone select Yes No respectively Fig 27 850EMT Settings e Set Date This option is disabled e Set Time Press this option to set the time or change the currently displayed time to an acceptable format The following screen appears as shown In Figure 28 e Power settings Use this option to find out the status of the Real Time Clock RTC battery and the Main battery e Access Management This function is disabled Press this tab the following screen appears as shown in Figure 30 POWER SETTINGS RTC Battery Status 3 03 V Main Battery Status 7 52V
2. Initially PLEASE WAIT will show on screen then the results will appear as shown in Figure 24 The overall result of the self test across all supported sensors will be shown at the bottom of the screen in this case PASSED N A 254 N A Photo Value Heat Value CO Value PASSED Fig 24 Self Test Results Screen For sensors that aren t present in the MX device or don t support self test N A will be shown as the value for that sensor Many detectors support self test on only one of their sensors and some don t support self test on any sensors at all 0r 5 5 6 Change Tones This function is for changing the tones and beacon settings for the MX Symphoni sounders and AVBase devices The function will be greyed out on the Test Functions menu unless a supported MX device is connected On the main change tones screen that appears as shown in Figure 25 1Hz and Hz options are for changing the flash rate of the built in beacon VOL LOW and VOL HIGH change the volume of the tone RESET will reset the settings back to the factory default The current settings are highlighted in green TONES RESET Fig 25 Main Change Tones Screen Pressing TONES will bring up another screen shown in Figure 26 that allows the particular tone type to be selected Temp 3 Temporal 3 is the standards compliant tone for Australia ULM THz Fast oo tn March T rh z we m mum l Tenp 3
3. Fig 17 Reset Table pattern displayed on screen matches what is required press the WRITE button to send the digital output pattern to the MX device e Press Y to clear the used addresses table and make all addresses available The message Session Table has been Reset appears e Press N to continue with the currently displayed values It leads you to the Session Program screen as shown in Figure 14 5 5 Test Functions A number of test functions are available for MX devices from the Test Functions menu as shown in Figure 18 which can entered from the main screen E n j i 7c TTE PES kni Fig 18 lest Functions Menu 17 5 5 2 Test Outputs This function allows the Remote LED and L2 functional base outputs of detectors to be tested Note this function will fail on MX ancillaries If the device is on the MX loop and connected via the IR mode the Test Outputs screen appears as shown in Figure 20 TEST OUTPUTS CHECK REMOTE LED IS ON FUNCTION BASE 1 ON NOTE B LST Fig 20 Test Outputs IR 1 Selected option is momentarily highlighted 2 Not Applicable option turns the function to Off Press the Test R LED or Test L2 buttons to activate the output and then YES or NO to indicate if the output is turned on To stop the test function press YES NO or N A as shown in Figure 20 The highlighting of the result selected is only visual and is not recorded NOTICE In the above scre
4. Fig 28 Set Time 1 Field to enter time Use the numeric keypad to enter the time in the i format hours minutes e Set Language This option is disabled Fig 30 Power Settings e Set Tones To set audible indications for touch screen operation and successful IR communications Press this button the following screen appears as shown in Figure 29 Ae e LCD Brightness Use this option to increase or decrease the clarity of the LCD display screen Press this tab the following screen appears as shown in Figure 31 LCD BRIGHTNESS oetthe Brightness level ofthe LCD display Fig 31 LCD Brightness 1 Scroll Up arrow to increase the brightness 2 Scroll Down arrow to decrease the brightness Slide the red indicator or press the up or down arrows to adjust the visibility of the LCD display screen as required e About 850EMT Use this option to find out the model number and the software version of the service tool Model Number 850EMT Software Version V1 3 Fig 32 About 850EMT ZZ 5 7 Ordering Information Pat Order Number 850EMT Service Tool Kit 850EMTK including carry case 5 7 1 Spares Accessories Order Number Spare USB Key 516 850 902 spare Marine Download 516 850 903 Lead Stylus 516 850 904 spare Ancillary Lead 516 800 922 Carry Case and Accessories 516 800 923 Ancillary Lead Spare Pins 516 800 924 bag of 10 Table 5b Order Numbers 5
5. Session Program button to display the session program screen as shown in Figure 14 Session Program SESSION PROGRAM MANUAL PROGRAM RESET SESSION TABLE AUTO PROGRAM B B Fig 14 Session Program Manual Program 5 4 1 Use this option to manually navigate to the required address and select it for programming into the device The screen appears as shown in Figure 15 Manual Current Device Address Next Available Address 3 4 Check Address Fig 15 Manual Session Program To change the device s address to the Next Available Address press the Address button Ihe message PLEASE WAIT PROGRAMMING DEVICE is displayed followed by the message Address changed successfully Io manually change the device s address to a different address press the Check Address button The Enter Address screen appears See section 5 3 Type a different address and press Enter for the 850EMT to accept and program the device 16 If you attempt to program an address that has already been used by an existing device the message appears as ADDRESS ALREADY IN USE In such a scenario e Press Y to change the device s address to the required address e Press N to retain the existing address 5 4 2 Auto Program Use this option to automatically program the device to the next available address The screen appears as shown in Figure 16 Use this option to automatically pro
6. 7 2 Specifications 6x AA NIMH rechargeable 240VAC to 12Vdc Adaptor 0 C to 50 C 1096 to 9596 non cond Battery Mains Environmental Ambient Temperature Relative Humidity Table 6 Specifications 6 Appendix 6 1 Additional Information 6 1 1 Compatibility The 850EMT supports USB 2 0 The 850EMT is not compatible with all USB memory sticks It is known to be compatible with the following memory sticks e Freecom Traveller supplied with this tool e Kingston Data Traveller e Lexar FireFly The 850EMT may not work properly with USB card readers The 850EMT s IR link is compatible with the 850 4098 625X series detectors and the Quad 850 ancillaries The MX1 panel needs V1 5 firmware or later for IR functionality to be available 6 1 2 Known Issues e There is no auto power off or standby function e For USB compatibility see 6 1 1 Compatibility section e For the 850 detectors the SCI Short Circuit Isolator and the Test outputs are triggered by the same command 6 2 Updating the Firmware New versions of firmware can be downloaded by placing the file on the USB memory stick placing the stick into the USB socket of the tool and turning off on the tool Errors in this process will not damage the tool The file is a GZ type It must not be uncompressed before use Consequently the existing firmware version is replaced with the new version The status message Is displayed as Updating firmw
7. 850 Engineering Management Tool Left Side View 1 DC IN 12V External Connector 2 Power On Off Toggle Switch The 850EMT Engineering Management Tool is a powerful to be interrogated tested and programmed Its easy and flexible tool used in the installation commissioning to navigate screens capture user requirements in an diagnostics and servicing of Tyco MX detection systems intuitive manner The 850EMIT allows all the MX addressable devices Fig 2 850 Engineering Management Tool Right Side View 1 COM Port 2 AUX Cable Port 3 Stylus The 850EMTK includes the following e 850EMT MX Service Tool e Pack of ten spare pins for Ancillary lead e MX Service Tool to Ancillary connector lead e Carry case e Mains charger adaptor e Shoulder Strap for 850EMT e 6xrechargeable AA size NIMH batteries e 12V car charger adaptor USB Memory Stick 2 1 Key Functions and Features The 850EMT can be used as a desktop unit clipped to a trouser belt or be carried using a shoulder strap The key functions and features of the 850EMT are 2 2 Two way communication via infrared IR to 850 Series detectors and modules Touch screen Backlit colour LCD display Portable with built in charger Accepts detectors onto the tool or ancillary programming lead for modules Read write detector ancillary address Displays model number and the SKU Displays temperature CO levels smoke obscuration Tests the detector remote LED an
8. MIX 850 Engineering Management Tool Australia New Zealand User Manual LTO586 Doc version 1 0 6 May 2015 LS exer Copyright 2015 Tyco Australia Pty Limited All rights reserved Tyco reserves the right to make changes to any aspect of this publication at any time without notice VIGILANT is a trademark of Tyco New Zealand Limited or its affiliates MX TECHNOLOGY is a trademark of Thorn Security Limited or its affiliates TYCO is a trademark of Tyco International Services GmbH 2 Contents 1 Guide through this manual ssseescs ecepevieurarterixbnet rettet ebd eut at ener ebackeswpentierdaclasise semester edet nteeeies 4 1 1 gzyswleteictcok eed m 4 1 2 Who is this Manual fOr RETE 4 13 What Products are Covered by this Manual ccccccccccceceeeeee emen 4 2 PECA e NN REN 5 PM IG TCU SQ lg TERRE diac seers te getuseaha E esee eee east eschensacy 6 P EE i AN ste Oru a oo D m em Rm eer TR 6 3 Operating SFC TONG a arseeaacetente ccunceieaaeantesetetaeeideadeeedactaceacadearatniennseceqnee boniendatigeatectanseenes 7 3 1 Indicators and Controls cece cc ceccc cece eee e cece e ne eme hme ne nennen nn eren nne nne nnn nnns 7 owAii ui i c E TEE ER 8 4 DUGG FASO OOEIVIT E 9 AT OGG IN OMOM t 9 42 Main IVES Ce RE 10 4 3 Connecting to an VX C
9. TO586 Version 1 0 6 May 2015 Subject to change without notice 26
10. V COs MNT 11 AA Co nn cingtoaDeteCior sctansanonesceacccarenaneannsanaccemasnanennneonsancacanaetnanensaveuceaesensepenenianausanececas 12 45 Connecting to an Ancillary DEVICE inncdinconssanccbsnduaedsneoantwawannndsteais tonansactsinlnniGamsetntbaceenegmeaamea 13 AG USNR MOGE speserei eaei E Je D MisqgR E stowteneascn 14 cx MEN I REDE TITETT 14 S MER Ie Felici RO 14 5 3 Change Tise c RETRO 15 5 4 Session Program m 16 Biel Avril reget csmeiersistninei na r a aa i e aE 16 DAL Ato FOGTA geese stosstrcus isana aetna ira e n T 16 924 5 Reset session laD Eiras na aut pasa r e OENE ues ies E EErEE Suances 17 o I SPITUDEDIOFS a E TUM e eet aU Emi cM EM D MI DE sacsae anode ianes 17 5 5 1 Test Digital COS M gl ctieeeeeehetecerehsaa a sleet nt et atte erkent artakoak seater chante taco vesaiwtnant tata hcdeantuansdaeteneana 17 o c MMI RU kc ee nee eee eee eer 18 5 5 3 Short Circuit Isolator PETERET 19 o0 siia oO ue TET 19 D eI NITITUR A 20 ING Eon POE E 20 o cS s EN SG RENNES 21 5 7 Ordering INfOrmMation peirtscertee occa oawersnnteadate ences cee concen aes aec ouseeesentpaievedewaussnemiwnaenesniruanitaasactes 22 SUME ERUNT ee ee ne 22 5 72 Specifications INEO T 22 6 POC IOI PN u 23 Bk Adaon IPPON ea MERE 23 SMNEMEC Cs ee er eee m
11. al User Interface This is used to display user interaction information such as alarms status messages and responses to user input 10 Buttons or Icons Buttons Icons are used for entering access codes text strings or general information Used to exit the current screen and display the home screen or main menu Used to accept the inputs as entered by the user or return to the previous screen Used to scroll up the list of devices Used to scroll down the list of devices Table 2 Buttons e Battery Status The battery icon changes its display as shown in Table 3 to indicate the different states of charging Battery Status Empty battery Charging in progress flashes between different charge states 5096 charged battery 75 charged battery Completely charged battery not charging Fully charged battery charging Error in charging battery Nearly empty battery need to recharge immediately Table 3 Battery States 4 3 Connecting to an MX Device The 850EMT can be connected to MX detectors or ancillary devices using the tool base the Ancillary lead or infrared communications The 5 different modes used by the 850EMT to communicate with the device are e IR High The 850EMT communicates with the detectors placed at a distance between 1m to 15m via the infrared link It is indicated by the icon on the bottom left hand corner of the main screen e IR Low The 850EMT communicates w
12. are for a few seconds Firmware Upload i The PIN will not be reset when new firmware is uploaded such that access is continued and re registration Is not required Any saved changes made to the settings of the 850EMT automatically apply to the updated firmware 233 Index Numerics 850EMT SETTINGS A About 850EMT Access Management Activate the Isolator Address Auto Auto Program AUX Cable Battery C CAUTION Change Address Charging and Mains Use Charging Batteries Check Address Communication Mode Connecting to a Detector Connecting to a Device Control Keys Controls D DANGER Device Status and Change Settings Dirtiness 24 21 14 14 Enter Address G Guide H Hold Indicators Indicators and Controls Information IR High IR Link IR Low K Key Functions and Features Keywords L LCD Brightness LCD display Main Menu Manual Program Menu Details More N NOTICE O Obscuration 15 16 15 22 10 14 16 14 14 15 P Pass Point 850EMT at Detector Point text Power Settings Reset Session Table HIC S session Program oet Language Set Tones Short Circuit Isolator opares opecifications otatus oymbols T Test Options Test Outputs Tool Base U Using the Menus 18 13 21 17 21 18 18 11 W WARNING Z Zone text 14 254 tyc CJ Safer Smarter Tyco Fire Protection Products NZX TECHNOLOGY L
13. d control outputs Monitors ancillary outputs Benefits of Using 850EMT Advantages of using the 850EMIT are Simplifies installation and commissioning Bidirectional IR link allows remote detector programming interrogation and testing 6 Reads the device status Changes the device settings session Programming enables programming of multiple devices in succession Test Functions Allows the device s inputs and outputs to be tested and the blink on poll setting to be changed Entry buttons have black border or red when selected which respond to touch DC IN 12 V External connection from the car 12V socket or mains adaptor In built stylus to select menu options by touching the LCD screen Improves health and safety by reducing the need to work at height e Reduces the possibility of Engineering error 3 Operating Instructions 3 1 Indicators and Controls The 850EMT is shown in Figure 3 The numbered items in the figure are explained in the following paragraphs Fig 3 850 Engineering Management Tool 1 Tool Base 2 Escape Button 3 OK Button 4 Up Arrow Button 5 Down Arrow Button 6 LCD Display Screen Tool Base The circular area is used to fit an MX detector by rotating the detector in the clockwise direction until it reaches the lock position For additional details see 4 4 Connecting to a Detector on page 12 4 3 2 Batteries The 850EMT is powered by 6 AA NiMH batteri
14. e activating an isolator when there Is another open circuit fault on the loop all the other loop devices between those two nodes will lose power Some loop devices may lose power If an isolator is opened and the IR mode is enabled however the isolator that is opened remains operational When the Short Circuit Isolator function is selected the screen as shown in Figure 22 appears SHORT CIRCUIT ISOLATOR CHECK YELLOW LED I5 ON CHECK YELLOW LED Bee E E E s NL i a r a f i l OE p ai a i EA Y Lr 4 i Fig 22 Short Circuit Isolator Pressing OPEN ISOLATOR will open the MX device s isolator and break the loop the yellow LED will become lit Pressing CLOSE ISOLATOR will close the MX device s isolator and close the loop the yellow LED will not be lit Any result buttons pressed will be visually highlighted on screen but no results will be recorded at all 5 5 4 Blink On Poll The blink on poll setting for a MX device can be changed using the Blink On Poll screen shown in Figure 23 To change the setting press YES or NO on the touch screen Fig 23 Blink On Poll Screen When the setting is NO the MX device will not blink its LED when polled even when the panel is configured to enable blink on poll for the MX loop to which the MX device is connected 19 5 5 5 Self Test Entering the SELF TEST function performs a self test of the supported sensors of a detector
15. efer to Figure 1 2 How to log on to the 850EMT 1 Inthe PIN Entry screen enter the 4 digit PIN 7240 by using the numeric keypad as shown in Figure 6 and press Enter 2 Ifthe PIN is correct the message PIN ACCEPTED is displayed on the screen The user will remain logged in until the 850EMT is turned off mzx TECHNOLOGY The PIN Entry screen is as shown in Figure 6 850EMT Engineering Management Too Enter PIN Fig 5 850EMT Start up Screen Fig 6 PIN Entry Screen 1 Enter PIN Field 2 Enter key Invalid PIN If the PIN entered is incorrect the message INVALID PIN is displayed on the screen 9 4 2 Main Menu Screen After the 4 digit PIN number has been successfully entered the main menu screen appears as shown In Figure 7 Detail on using the main menu is covered in oection 5 REPORT GENERATOR LA RETURN Fig 7 Main Screen 1 Battery Status 2 USB Icon Indicates Disconnected 3 Four Disabled Options 4 8b50EMT Settings Button Each option leads to a further screen which may present further options When using the menus you use combinations of the same basic methods to navigate between the various settings in a screen for example To avoid repetition this guide does not provide full step by step details for every screen LCD Display A 3 5 inch OVGA TFT backlit LCD alphanumeric display with a high resolution of 320 x 240 pixels and 262k colours is provided as the Graphic
16. en the Function base remains in the ON state until all the available tests for the TEST L2 have been completed It is advisable to turn OFF the Function base test before continuing with other tests If the device is connected via the tool base the Test Outputs screen is displayed as shown in Figure 21 18 TEST OUTPUTS REMOTE LED PASS TESTL2 FAIL NOTE Bau us Fig 21 Test Outputs Toolbase Press the required Test button and check the result on the screen After all the tests have been performed on the device when connected to the Toolbase the output of the test is either of the below options e Pass Result of the test is positive Yes and is shown in Green e Fail Result of the test is negative No and is shown in Red 5 5 3 Short Circuit isolator This option is used to activate and test the isolator function of a 850 Series detector or MX Quad module Short Circuit Isolator This option works for only the 850 Series Generation 6 detectors and range of Quad modules It can be used to isolate a particular section of the loop because some planned activity requires that section to be without power Use the Short Circuit Isolator SCI function at two devices when you would like to disable the loop between those two devices The device s LED is illuminated yellow to indicate that the isolator is tripped This status remains until the isolator is closed Activate the Isolator Whil
17. es located underneath a cover in the base of the unit How to fit replace the batteries 1 Remove the battery compartment cover located on the base of the 850EMT use a flat bladed screwdriver if desired 2 Insert the NIMH AA sized batteries ensuring correct polarity as marked on the PCB or as shown in Figure 4 Fig 4 Batteries placed in the battery compartment 3 Refit the battery compartment cover by gently pushing it down hard enough such that it audibly clicks into place 8 CAUTION Battery Status e Fully charge the batteries for 4 hours before using for the first time e Recharge the batteries as soon as the Low Battery indicator appears e Do not open the battery compartment cover while the unit is switched on or charging e Ensure that only Tyco specified Nickel Metal Hydride rechargeable batteries are used and fully charged before use Charging and Mains Use The 850EMT has its own built in charging circuit powered by an external adaptor Adaptors for mains and 12V car charging operation are included Charging while the 850EMT is OFF i If the 850EMT is switched off and you connect a charging adaptor the 850EMT will automatically turn on while it is in the state of charging If you disconnect the adaptor to stop charging it the 850EMT will switch off automatically 4 Starting the 850EMT The screen shown in Figure 5 appears for about 3 4 1 Logging On seconds after the power is turned on r
18. gram the address of any newly fitted device with the next unused address AUTO PROGRAM Current Device Address Next Available Address 4 5 Check Address To change the value of the Current Device address to a new value click on the Check Address button Refer to section 5 3 for the Change Address function aan Fig 16 Manual Session Program 5 4 3 Reset Session Table Some functions may be greyed out and unavailable depending on the MX device if any that is connected Use this option to reset the values displayed for the Current Device Address and the Next Available Address of the device The screen appears as shown in Figure 17 The operation of each test function is detailed in the following sections of this manual RESET SESSION TABLE 5 5 1 Test Digital Outputs The Test Digital Outputs function allows the digital output signals of the MX integrated circuit to be controlled As this function controls the actual ASIC outputs knowledge of each output s function on the specific MX device Is required Currently the only field application for this Want to Reset the Table is configuring AZM800 modules in New Zealand This procedure is documented in LT0459 AZMS80O Installation N and Setup Instructions 4 The screen which is displayed on the 850EMT is shown iss n P in Figure 19 To toggle a bit between 0 and 1 or vice versa press that bit on the touch screen When the bit
19. ion dirtiness temperature etc so that they do not fluctuate according to the environmental conditions While in the Hold state the 850EMT stops beeping To return from this state press hold again e More Select this option to show additional information about the device Device Status Change Settings CH 2 255 CH 3 1 CH 4 0 Node Voltage N A MFG Date 24 03 11 Serial 027017CCB Hs Fig 12 Device Status More LPSY800 R 5 3 Change Address The Change Address function allows the address of the MX device to be changed Press the Change Address button to show the Change Address Screen see Figure 13 The Enter Address screen appears lype the required address 1 250 using the numeric keypad To program the address you typed press the Enter button The message Address is changed Successfully is displayed In order to delete the digits typed press the back arrow On exit the Device Status Change Address screen is updated to reflect the new address ENTER ADDRESS 129 MEE EVEREX Ea ae Fig 13 Enter Address Screen Maximum Address Value The Enter Address field accepts a maximum value of 255 197 5 4 Use this option to program multiple devices with consecutive addresses The 850EMT checks a table of addresses used so the next unused address can be assigned to each device his session table can be cleared From the Main menu press the
20. ional commissioning tests on the device and alter the default settings for Visual Alarm Devices See 5 5 Test Functions on page 17 e 850EMT SETTINGS cog wheel used to change the 850EMIT settings See 5 6 850EMT Settings on page Fi The following sections detail the menu options 14 5 2 Device Status Use Device Status option to display the status configured settings of the device connected to the 850EMT An example Device Status screen is shown in Figure 11 Device Status Change Settings Address 128 Ch Type SYS800 angs Address SKU 516 800 960 Point Text Sounder Beacon Device Zone label Bldg 2 Block A Dirtiness N A a O foh HOLD Fig 11 Device Status amp Change Settings Screen 1 Channel analogue values 2 Home Icon Returns to the main screen 3 Return Icon Goes back to the previous screen 4 Displaying more information The properties shown on this screen and the More screen are as follows e Address Displays the address of the device To change the address of the device press the Change Address button See Section 5 3 e Type Displays the type of the device e SKU Shows the stock code of the device for a replacement e Point Text Represents the point area where the device Is located Usually shows N A e Zone Label Reflects the Zone where the device is located Usually shows N A e Dirtiness This represents
21. ith the detectors placed at a distance between 0 and bm via the infrared link It is indicated by the icon oee Section 4 6 for using IR mode IR Link The IR mode needs to be enabled at the fire panel before the IR link can be used CAUTION Turn off the Sounders before enabling the IR Mode to avoid the devices from overloading the loop and from being accidentally activated by the 850EMT Address Programming Tool e AUX Cable Communication between the 850EMT and the ancillary is possible by attaching the auxiliary cable to the AUX socket of the 850EMT and plugging it on to the auxiliary device 3 pins Refer to 4 5 Connecting to an Ancillary Device on page 13 It is indicated by the icon on the main screen a NOTICE If the 850EMT is not in the Auto mode it cannot automatically detect the Communication mode to be AUX for legacy ancillary leads You have to manually select AUX e ool Base Communication between the 850EMT and the device is possible by placing the detector in the lock position on the tool base of the 8bOEMT Refer to 4 4 Connecting to a Detector on page 12 It is indicated by the icon f e Auto The 850EMT automatically detects the kind of mode with which it has to communicate with the device by considering each of the following scenarios f the Ancillary lead is plugged in AUX Mode is selected f a detector is fixed onto the tool base the Tool Base mode is chose
22. munications If successful IR comms are enabled see section 5 6 the 850EMT will beep with a specific chirp when it can successfully communicate with the device using IR Keep the 850EMT pointing at the device and use its screens and buttons to select the required functions Note that if the 850EMT no longer points directly at the device IR communication will stop working 13 5 Vienu Details The Main screen is displayed as shown in Figure 10 DEVICE STATUS SESSION PROGRAK 2 AUTO Fig 10 Main Menu Listed Options 1 Device Status and Change Settings 2 Session Program Manual and Auto Modes 3 Communication Mode For this example shown as AUTO 4 Test Functions 5 Detector connected to 850EMT 6 Device connected via an Ancillary to the 850EMT Navigate to 850EMT Settings 5 1 Main Menu Options There are five options as follows e DEVICE STATUS used to display the previously configured device settings and allows you to change these settings as required See 5 2 Device Status on page 14 e SESSION PROGRAM allows the user to manually automatically programme addresses into the device and reset the session table See 5 4 Session Program on page 16 e COMMUNICATION MODE pressing this changes the device communication mode between IR High IR Low Aux Cable Tool Base and Auto See 4 3 Connection to a MX Device on page 11 e TEST FUNCTIONS used to perform addit
23. n f no auxiliary cable is present and the detector is not fixed onto the tool base the 850EMT chooses IR Low as the default mode of communication f the ancillary is connected to the 850EMT via the Auxiliary cable and a detector is fixed on the tool base the 850EMT does not choose any communication mode In this scenario an error message is displayed as Setup not allowed Remove detector or disconnect AUX cable lt is indicated by the icon Q This is due to a conflict between the AUX mode and the Tool base modes of communication Auto mode Not Available This functionality is not applicable for e The 813P and the 801F 801 FEx Flame detectors e ntrinsically Safe detectors such as the 801 PHEx 801 HEx 801 CHEx etc hese detectors can communicate with the 850EMT via only the Tool base mode of communication 11 4 4 Connecting to a Detector Detectors are connected to the 850EMT via the Toolbase as shown in Figure 8 Use the marking on the service tool above ESC button to align the detector Place the detector in position 1 to engage and then twist clockwise to position 2 to lock Fig 6 Connecting to a detector 1 Position 1 2 Position 2 3 Marking 12 4 5 Connecting to an Ancillary Device Ancillary devices are connected to the AUX socket using the ancillary lead as shown in Figure 9 Fig 9 Connecting to an Ancillary Device Connecting to an Ancillary De
24. n nnn re ne eee eee 23 NN ME Ie I NE E ee ne ne ee ere ee 23 S MEE 918 aie Did mal AB LA s C nen E A ree eee eee 23 ho WC 24 ey 1 Guide to this manual 1 1 Keywords and symbols This documentation uses special notations that you can use for better orientation Symbols in the margins indicate warnings information or instructions You can find an explanation of these symbols in Table 1 Symbol n A k A Table 1 Keywords and symbols in this manual 4 Warning Imminent danger Death or severe Injury when disregarded Warning Potentially dangerous Situation Death or severe injury possible when disregarded Warning Potentially dangerous Situation Minor injury possible when disregarded Warning Potentially dangerous situation Material damage possible when disregarded Helpful Information 1 2 Who is this Manual for This user manual is aimed for the Installation Commissioning and Service Engineers in Australia and New Zealand who are using the 8b5OEMT What Products are Covered by this Manual This manual covers the 850EMT Engineering Management Tool with Australia and New Zealand specific features enabled This manual describes the product range and features as used in Australia and New Zealand 1 3 For use with marine detectors or UK sourced MZX systems please see manual 120 515 058 2 Introduction Fig 1
25. the percentage of dirt or dust that is present inside the sensor accumulated over a period of time This field cannot be altered The percentage value obtained may be different from the value displayed for the Dirtiness field on the panel screen CAUTION If the percentage of dirtiness is shown as 80 or above the device may have degraded pertormance and should be replaced as soon as practicable e Obscuration he response of an optical detector is normally measured with reference to the obscuration produced by smoke Obscuration is measured in percent per metre This field cannot be altered e Temperature The temperature for those detectors with a heat element e CO Level Carbon Monoxide CO level ppm for those detectors with a CO sensor e CH X Channel analogue value for input X The MX ASIC has up to 4 analogue inputs and these are assigned different functions depending on the device type The fields CH 1 to CH 4 show the raw analogue value 0 to 255 of the respective input Refer to the MX device details for an explanation of each input s function e Node Voltage For 850 series detectors in IR mode it IS possible to read the MX loop voltage present at the detector This value can help in fault finding or checking for voltage drops around the loop e Mfg Date The date of manufacture of the device e Serial Unique serial number of the device e Hold use this option to retain the values for fields such as obscurat
26. vice Plug the 3 pin header into the matching three holes or connector on the MX device The device can be powered off or connected to an operational MX loop The ancillary lead will power the device sufficiently to communicate with it and override the MX loop communications Available Connections It is good practice to connect only a detector or one ancillary device at a time The 850EMT is equipped with a port interlock feature When the ancillary lead is connected to the AUX socket communication via the IR link and base connection are disabled When the ancillary lead is removed the 850EMT will be able to communicate via the IR link and the tool base connection will be re enabled The 850EMT may be connected to an ancillary device that is also connected to and powered from the addressable loop However a No Response fault for that device may be generated at the Control Panel under these conditions 4 6 Using IR Mode Using Infrared Mode Infrared communications can be used with certain MX detectors and modules that contain the IR transceiver currently 850 Series detectors and Quad I O modules The devices need to be connected to and powered by the compatible MX panel and IR mode needs to be enabled on the panel before IR mode can be used Refer to the specific Fire Panel Operator s Manual With IR mode enabled carefully point the 850EMT at the MX device until the yellow LED on the device flashes to indicate successful com
Download Pdf Manuals
Related Search
Related Contents
La recherche clinique, mode d`emploi Lite-On 52x 32x 52x + 16x Half-Height COMBO 取扱いについて Télécharger la plaquette de Gutermann Multimètre analogique - GMC par 取扱説明書 User`s Manual 1vv0300714_GE863 Welcome to your AHSCA USER Manual Copyright © All rights reserved.
Failed to retrieve file