Standard Procedures to Use the Tissue Dryer
Contents
1. Electron Microscopy Centre Title STANDARD PROCEDURES TO USE THE TISSUE DRYER Equipment Edwards Tissue Dryer Model ETD4 Revision 1 0 Effective Date 12 21 2005 Author X Yang Warning Before you attempt to operate this equipment for the first time please make sure you are aware of the precautions that you must take to ensure your own safety Working Environment Do not use electrical equipment in e Rain or excessive moisture environment e The presence of flammable or explosive gases The equipment is not designed to be water or splash proof or to be used in area where there are flammable or explosive gases or fumes Standard Procedures to Use the Tissue Dryer The purpose of freeze dryer is to remove water from a frozen material by sublimation so to preserve biological materials Sensitive media such as tissues can be dried and preserved with very low moisture contents Reconstruction of freeze dried products can be achieved by adding distilled or sterile water Checkouts before a Drying run 1 Thermo electric module cooling check 1 1 Having connected the unit to the electrical and water supplies switch on mains supply Mains electricity indicators on the front panel should light 1 2 Ensure all valves are closed and the variable element controller is set at the zero position 1 3 Turn on water supply 1 4 Fill the desiccant tray with fresh molecular sieve and place in position on the base plate 1 5 Place th
2. e chamber in position on the base plate 1 6 Depress the buttons marked RP element and cold on the temperature selector switch and turn the variable element controller in a clockwise direction to position 7 The temperature should start falling to 50 C 2 Vacuum check 2 1 Switch on the rotary pump by depressing the switch button on the front panel marked Rp 2 2 Switch the chamber isolation valve from isolation to pump position on the front panel 2 3 The pressure should fall to 10 2 mbar within 6 minutes Note The pump down time can only be expected with a clean empty chamber containing molecular sieve The first pump down after a long exposure to atmosphere will be prolonged due to the degassing of vapour from the surface of the system Subsequently pump down times should be shorter 2 4 If the above pumping performance cannot be obtained switch off the element and turn the variable element controller knob to zero position Depress the OFF button on the front panel and contact EMC staff immediately The following steps only apply to advanced users only 2 5 Switch the chamber isolation valve from pump to isolation position and open the air admittance valve to vent the chamber 2 6 Open the gas ballast valve on the front panel and allow the pump to operate on full gas ballast for about 1 hour This operation will clean the pump oil if contaminated and improve the subsequent performance 2 7 Repea
3. n fresh desiccant is used it is advisable to allow the surface to absorb a small quantity of atmospheric moisture this will prevent its subsequent migration in the vacuum system and possible contamination of the specimen 10 11 12 13 14 Turn on cooling water valve mounted on the wall turn the valve anticlockwise and check that sufficient water flows The minimum flow required is 1 0 L min at a maximum pressure of 4 2 kgf cm Check both cylinder valve fully clockwise and regulator valve anticlockwise are closed Turn on argon slowly turn the cylinder valve clockwise until there is a reading at the first meter Slowly turn the regulator clockwise until the second meter is reading about 5 psi Note Please refer to a separate document for detailed operation procedures for handling high pressure gas cylinder Switch on the power by depressing both green buttons marked mains and Switch the chamber isolation valve to pump Switch on the rotary pump by depressing the button marked RP The rotary pump will start immediately and the needle of the vacuum gauge will register falls If the needle is falling below 0 1 mbar it indicates that there are no major leaks in the system If the needle could not fall under 0 1 mbar for a quite while contact the EMC staff for support or refer to the vacuum check section Depress the buttons marked element and cold on the temperature selector swi
4. on the log book Drying times for mammalian tissues as a function of thickness and module temperature Thickness of tissue Temperature maintained Spleen Kidney Liver 1 0 mm 10 C 55 min lh 45 min 2 0 mm 10 C 1h 2h 1h 3 0 mm 10 C 3h 30 min 3h30min_ 3h 30 min 4 0 mm 10 C 1 0mm 20 C lh 1h10min 45 min 2 0 mm 20 C 2h 30min 3h 1 h 30 min 3 0 mm 20 C 4h 4h 4h 4 0 mm 20 C 5h 5h30 min 5h 1 0mm 30 C 1h 20min 1h 20min 1h 20min 2 0 mm 30 C 3h 3h 3h 3 0 mm 30 C 3h45 min 4h 3h 45 min 4 0 mm 30 C 1 0 mm 40 C 2h 2h 2h 2 0 mm 40 C 3h45 min 3h45 min 3 h45 min 3 0 mm 40 C 4h 30 min 4h30 min 4h 30 min 4 0 mm 40 C 6h 30min 6h45 min 6h 30 min 1 0mm 50 C 3h 3h 3h 2 0 mm 50 C 4h45 min 4h45 min 4h 45 min 3 0 mm 50 C 6h15 min 6h15 min 6h 15 min 4 0 mm 50 C 1 0mm 60 C 5h30 min 5h30 min 5h 30 min 2 0 mm 60 C 8h 8h 8h 3 0 mm 60 C overnight overnight overnight 4 0 mm 60 C overnight overnight overnight Reference 1 Edwards Tissue Dryer User Manual Edwards 1983
5. t the original vacuum test after gas ballasting has been completed Close the chamber air admittance valve Depress the element and cold temperature selector switches and turn the variable controller to maximum 2 8 When the temperature reaches 40 C open the chamber isolation valve and repeat test as ZT 2 9 If the correct pump down times still cannot be obtained check the system for leaks Thermoelectric Module heating check 3 1 With the chamber under vacuum and the thermoelectric element cooled to about 58 C turn the variable element controller to zero position 3 2 Depress the element and the heat button on the temperature selector switch 3 3 Turn the variable element controller in a clockwise direction Heating will now take place and with the controllers in a fully clockwise position the temperature will rise to approximately 60 C in about 12 minutes Note If the variable element controller has been preset to certain value for cooling down purpose the heat function will not start You will have to turn it counterclockwise to zero position then reset it to the desired value Operation of Freeze Dryer 1 Check the waterline is proper connected Check both air admit and gas ballast valves are closed If not close the valves Check the variable element controller is set at zero position If not set to zero Ensure the tray within the chamber is filled with fresh molecular sieve Note Whe
6. tch and turn the variable element controller to the desired value If not sure a value of 7 is always a good start point The temperature on the base plate should start falling down to a certain value depending on the setting Prepare and freeze the specimen by quenching it in R12 refrigerant or iso pentane cooled by liquid nitrogen Switch the chamber isolation valve to isolation and open the air admittance valve by turning it counter clockwise slowly to vent the chamber with dry Argon gas When the chamber is vented open the metal cover and transfer the frozen specimen onto the base plate Note During the transferring please remember to wear power free gloves provided to avoid potential contamination to the specimen and vacuum chamber Caution A cold burn could result is the cold plate or the liquid nitrogen is in contact with bare skin when transferring the frozen specimen 15 16 17 18 19 20 21 22 23 Replace the chamber cover in position Close the air admittance valve and switch the chamber isolation valve to Pump Caution the air admittance valve and gas ballast valve and needle valves which is very sensitive to the force When operating with such valve please be gentle The treads on air admittance valve have been found to be worn out The tissue should start to dry Note The drying times for individual specimen are different depending on the tissue type frozen temperat
7. ure and thickness of the tissue Please refer to available references Practical tests might also have to be contacted in order to obtain more relevant information Examples of drying time are given at the end of the manual When the specimen is dry depress the cold button and set the variable element controller back to zero position The specimen should start to warm up to room temperature Close the cylinder valve while the regulator valve remains on Switch the isolation valve to isolation and slowly open the air admittance valve to vent the chamber with residue dry argon gas By doing this user not only vents the work chamber but also release the residue high pressure argon gas remained in the regulator and tubing Caution if the valve is opened too quickly the specimen may be blown from the plate by the inrush of air to the chamber Remove the work chamber and top plate with powder free gloves Transfer the specimen to vapour fixative to embedding medium as desired Replace work chamber and top plate Close the air admittance valve fully clockwise and switch the chamber isolation valve to pump so the chamber could maintain at vacuum Switch the isolation valve to isolation after evacuation Switch off rotary pump and power by depressing RP and two green buttons Close the regulator valve by turning the knob counter clockwise when regulator meters read Zero 24 Record the usage of freeze dryer
Download Pdf Manuals
Related Search
Related Contents
USER'S MANUAL discriminateur de sens de rotation 2551.090 Rev N Spanish Blind Magmeter.indd HSR-240L HK7-505 DVD-2683W13 Betriebsanleitung End User Manual - Micron Communications Slant/Fin Sentry S-120 User's Manual 取扱説明書 - トーヨーキッチン Manual de Instalación N Classic Controla Copyright © All rights reserved.