CEREC Ortho SW
Contents
1. Export scan You can save an optical impression in a compressed format in any location Y You have opened a scan 1 Select Export Scan A standard Windows file dialog box opens Select the target folder to which you want to export the scan Assign any name to the scan Click on the Save button The scan is exported m d mI If you want to transfer the scan from your acquisition unit to another PC e g in order to upload it to the Web portal you can use a USB stick for this purpose 65 26 060 D3534 D3534 208 06 01 02 01 2015 21 5 System menu Sirona Dental Systems GmbH 5 5 License manager Operator s Manual CEREC Ortho SW License manager on the USB license stick To do this start the license manager via the system menu and follow the instructions on the screen Keep the license certificate with 25 digit license key ready which you either obtained with the unit or ordered separately from your dealer F The license manager is used for the installation of new software licenses i4 Tip You can also start the license manager via Start All Programs Sirona Dental Systems CADCAM CEREC Ortho Tools License Manager To activate the license you must have an Internet connection and the USB license stick must be connected Licenses and code libraries For information on licenses and code libraries from third parties see licenses pdf The file is in the installation directory under C Pr
2. The software prompts you to proceed to the next latching 7 Execute steps 5 and 6 a total of 11 times The software provides status updates on the calibration and informs you once the procedure is complete You will be prompted to measure the position of the exit window 65 26 060 D3534 D3534 208 06 01 02 01 2015 17 4 Configuration Sirona Dental Systems GmbH 4 2 Settings Operator s Manual CEREC Ortho SW Measuring the position of the exit window 1 Mount the bottom side of the calibration set to the tip of the camera 2 Click the OK button on your gt The calibration process is continued Once the calibration is complete a message is displayed indicating this 3 Confirm the message by clicking the OA button on your The CEREC Omnicam is calibrated Error message during calibration The software indicates if an error occurs during calibration If the calibration process resulted in errors restart the process End calibration Y The software indicates that the calibration was completed successfully gt Click on the OK button The CEREC Omnicam is calibrated Settings The menu item Settings has the following subitems e Reset notes e Selecting a language e Selecting a seating position 65 26 060 D3534 18 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 4 Configuration 4 2 Settings Operator s Manual CEREC Ortho SW Reset notes Warnings may appear
3. step Guide the camera in the direction of the arrow over the marked area Ifthe connection is scanned a signal tone will be heard The software changes to the Complete Jaw step Rescan the areas marked in yellow in the model This step is optional The upper jaw scan is finished 35 8 ACQUISITION phase Sirona Dental Systems GmbH 8 3 Perform a scan Operator s Manual CEREC Ortho SW Buccal registration Y The lower jaw and upper jaw have been scanned 1 Click the Buccal scan object The sequence bar for the buccal registration is displayed 2 Perform the buccal registration on the right side by having the patient bite as required and guiding the camera from the start marker A to the target marker B As soon as the buccal registration is finished a signal tone will be heard The software changes to the next Scan buccal left step 3 Perform the buccal registration on the left side by guiding the camera from the start marker A to the target marker B As soon as the buccal registration is finished a signal tone will be heard Finishing the phase Y All required scans have been performed lower jaw upper jaw buccal registration Y The MODEL phase can be selected 2 Click on the MODEL phase NN or A gt Click on the double arrow 36 The program switches over to the VODEL phase 65 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 9 MODEL ph
4. Click into the main window with the left mouse button The system menu is closed Save scan In this dialog you can save the current scan gt Select Save Scan in the system menu The scan is saved 65 26 060 D3534 20 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 5 System menu Operator s Manual CEREC Ortho SW 5 2 Save the scan under a different name Save the scan under a different name This dialog allows you to save the current scan under a new name or assign it to a different patient 1 Select Save Scan as in the system menu gt The Save as dialog box opens 2 Enter a name for the scan 3 Select the appropriate patient 4 Click on the OA button Import scan Y A CEREC Ortho SW scan is located on your acquisition unit or removable media 1 Click the mport Scan button in the system menu A standard Windows dialog box opens 2 Select the folder where the scan is located 3 Select the relevant file NOTICE File types If the selected file ortho or dxd is an optical impression with the CEREC Ortho SW software it will be opened If not it will not open and an error message will be displayed Scans which were not created with the CEREC Ortho SW Guided Scanningprocess can not be imported for orthodontic use 4 Click the Open button 5 Assign a name to the scan 6 Click on the OK button The scan is now imported and opened
5. aate dida dol hasos 41 10 22 Enter order dale ons Dus eiiam ea e ER E RA Ra DDR Ced ai 41 10 2 3 Add additional InfoMine 41 10 24 SING DIN GICAL ses ats Seats O SH UE ORO 41 10 25 Order lS Eos casi tin 41 1055 Send Modelo Dolphin 3D nara 42 INdEX ene E 43 65 26 060 D3534 D3534 208 06 01 02 01 2015 Copyright Trademarks 65 26 060 D3534 D3534 208 06 01 02 01 2015 Introduction Dear Customer Thank you for purchasing CEREC Ortho SW software from Sirona With CEREC Ortho SW and in combination with the CEREC AC Omnicam acquisition unit you can create scans and send them to your partner laboratory via the Sirona Connect portal Any type of orthodontic apparatus can be manufactured from your impressions there Improper use and handling can create hazards and cause damage Therefore please read and follow this document carefully You should always keep it within reach To prevent personal injury or material damage it is important to observe all safety information Your CEREC Ortho SW team Copyright and trademark Sirona Dental Systems GmbH All rights reserved The information contained in this manual may be changed without notice The software and all related documentation are protected by copyright You must therefore handle it in the same way as any other protected material Anyone who copies this software to any medium for any purpose other than his own personal use without the written permission of Siro
6. and sequence bars 30 The Guided Scanning process contains 3 scan objects A e Lower jaw scan e Upper jaw scan e Buccal registration The selected scan object is highlighted in yellow Every scan object contains several steps which are displayed in a sequence bar B The selected step is highlighted in yellow and marked with a red line When a step is finished it is marked with a green line in the sequence bar and the software switches to the next step You can repeat the last step by clicking on the double arrow on the left side of the sequence bar In this case the current scan step is discarded and must be repeated 65 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 8 ACQUISITION phase Operator s Manual CEREC Ortho SW 8 3 Perform a scan User guidance You are guided through the acquisition steps with tone signals and graphic displays e The start point of a partial scan is displayed with a start marker green e The end point of a partial scan is displayed with a target marker black white e fthe start marker or target marker are marked with an hourglass symbol the camera must be held in this position until the start marker or target marker disappears e fthe camera has identified the start point of a partial scan the tone signal changes and the start marker is hidden The exposure begins e The guide direction for the camera is displayed with arrows e Ifthe autom
7. in The USB license stick is included in the scope of supply of the acquisition unit If you require additional licenses please contact your dealer Always keep the USB license stick near the acquisition unit All authorizations interface and software licenses can be installed as electronic licenses on the USB license stick You must enter a 25 digit license key for this purpose You will receive the license key along with the acquisition unit Alternatively you can order it separately from your dealer Following an update you may require a new license that is not available on your USB license stick For more information refer to the section on License manager Downloading software ServicePacks To keep your software updated regularly check whether new ServicePacks are available You can find information on current product versions and a description of the improvements and enhancements made in the ServicePack under htfp cerec ortho com Starting the software Y The CEREC Ortho SW software is installed You will find the start icon on the desktop X The USB license stick is connected with a valid current license gt Double click the CEREC Ortho SW start icon Or gt Click on Start All Programs Sirona Dental Systems CEREC Ortho CEREC Ortho SW The software is started 13 License update For more information on the license manager refer to the section on License manager Installation of the Li
8. more for 3 seconds to mark the end point of the partial scan While doing so the target marker slowly disappears The software automatically changes to the next Scan occlusal right step 4 Position the camera above the last molar on the right The starting point is marked with a green start marker A and is automatically recognized Guide the camera occlusally in the direction of the arrow as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan vestibular right step 5 Position the camera above the last molar on the right The starting point is marked with a green start marker A and is automatically recognized 6 Turn the camera towards the buccal and guide it in the direction of the arrow over the arch as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan transversal right step T Guide the camera in the direction of the arrow over the marked area from the start marker A to the target marker B Ifthe connection is scanned a signal tone will be heard The software changes to the next left hand quadrant step 65 26 060 D3534 32 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 8 ACQUISITION phase Operator s Manual CEREC Ortho SW 8 3 Perform a scan Scan the left hand quadrant and complete th
9. saved it to the device Navigate to the location of the installation file Unzip the zip file and save the installation files Double click the setup exe file PDPA Select the language of the installation and click the button marked OR The installation wizard opens 5 Click on Next The license agreement is shown 6 Confirm the license agreement with the Yes button The program continues the installation routine 7 Select Full Installation NOTICE Installing DirectX If DirectX is not yet installed on your computer it will be installed now Accept the license agreement and decide whether the computer is to be restarted now or later 8 After installation you can display the ReadMe file by ticking or unticking the appropriate check box 9 Click on the Finish button Uninstalling the software X The program is closed 1 Click on Start All Programs Sirona Dental Systems CEREC Ortho Tools Deinstallation to uninstall the software During the uninstall procedure you will be asked whether you want to delete the patient data or the entries in the registration database e g the calibration data 2 Depending on your decision click either the Yes or No button The software is uninstalled 65 26 060 D3534 12 D3534 208 06 01 02 01 2015 65 26 060 D3534 D3534 208 06 01 02 01 2015 Copy protection The software can be started only when the USB license stick is plugged
10. will be heard The software changes to the next Scan transversal left step 6 Guide the camera in the direction of the arrow over the marked area from the start marker A to the target marker B Ifthe connection is scanned a signal tone will be heard The software changes to the Complete Jaw step 7 Rescan the areas marked in yellow C in the model This step is optional The lower jaw scan is finished 65 26 060 D3534 D3534 208 06 01 02 01 2015 33 8 ACQUISITION phase Sirona Dental Systems GmbH 8 3 Perform a scan Operator s Manual CEREC Ortho SW Upper jaw scan Scan the right hand quadrants Y The lower jaw has been scanned 1 Click on the Upper Jaw scan object The step list for scanning the upper jaw is displayed V The first step Scan lingual right is selected in the step list 2 Position the camera above the last molar on the right of the upper jaw and hold the camera in this position to mark the start of the scan The green start marker A which marks the starting point for the partial scan slowly disappears and the scan begins The guide direction for the camera is shown with an arrow 3 Turn the camera lingually and guide it in the direction of the arrow over the lingual surface of the teeth as far as the target marker B on the centerline When the center line has been reached hold the camera in position once more for 3 seconds to mark the end point of the partia
11. 08 06 01 02 01 2015 Sirona Dental Systems GmbH Operator s Manual CEREC Ortho SW Index C V Code libraries 22 Views Cut Change 25 Invert model area 26 Enlarge 25 Remove model area 26 Global 25 D Reduce 25 Display objects W Lower jaw 27 warm up time Transparency 27 Camera 29 Upper jaw 27 G Guided scanning 30 L Licenses 22 M Manual html format 8 PDF format 8 P Phase EXPORT 39 MODEL 37 S Safety information 7 Scan objects 30 Software Uninstallation 12 T Tool wheel 10 Tools 25 65 26 060 D3534 D3534 208 06 01 02 01 2015 43 We reserve the right to make any alterations which may be required due to technical improvements Sirona Dental Systems GmbH 2015 Sprache englisch Printed in Germany D3534 208 06 01 02 01 2015 A Nr 000 000 a NUN Sirona Dental Systems GmbH in the USA FabrikstraBe 31 Sirona Dental Systems LLC Order No 65 26 060 D3534 D 64625 Bensheim 4835 Sirona Drive Suite 100 Germany Charlotte NC 28273 WWW sirona com USA
12. 5 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 4 Configuration Operator s Manual CEREC Ortho SW 4 1 Devices Configuration The Configuration button is located in the system menu The Configuration menu contains the following submenus e Devices e Settings Close the Configuration menu by clicking on the Exit Configuration button Devices All connected devices can be displayed and configured under the menu item Devices A green check mark on a device indicates its availability CEREC Omnicam Audio feedback Using the Sound selection box you can switch the audio feedback for acquisitions on or off You can control the volume using the slide bar 3 different tone varieties are available Accepting settings gt Click on the Ok button Discarding settings gt Click on the Cance button Resetting settings gt Click on the Reset Camera Settings button The settings are reset to factory settings 65 26 060 D3534 D3534 208 06 01 02 01 2015 15 16 Calibration The measurement procedure used by the system requires the use of a calibrated CEREC Omnicam The CEREC Omnicam is factory calibrated Then calibrate the CEREC Omnicam after every reinstallation and after every transport The calibration set supplied with the CEREC Omnicam is available for the calibration process The supplied calibration set and a USB stick form a single unit Recalibrate the C
13. 6 060 D3534 24 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH Operator s Manual CEREC Ortho SW 7 Menu field 7 1 Views 65 26 060 D3534 D3534 208 06 01 02 01 2015 Menu field The menu field offers you various different functions depending on the current step Views Global You can use the View Options button to display six predefined views in the Global area e Top e Bottom e Right e Left e Anterior e Posterior Changing the view 1 Click on the View Options button 2 Click on one of the proposed views The virtual model rotates to the corresponding view Enlarge or reduce the view 1 Click on the View Options button 2 Position the mouse pointer over the center tooth icon and press and hold the left mouse button The icon then changes to a magnifying glass 3 Pull the mouse button up or down The virtual model is then enlarged or reduced Tip You can also use the center mouse button and the trackball directly on the virtual model to enlarge or reduce a view Tools The most important tools are also offered to you in the tool wheel For more information on the tool wheel refer to the section Tool wheel You will find all tools as a submenu under 7oo s Tip You can cancel the current tool with the stop icon top right Undo and reset With the Undo button in the tools you can undo the last change made With the Reset button in the
14. EREC Omnicam in the following cases e following transport shaking stress or during first commissioning e after storage in unheated or un air conditioned rooms temperature differences exceeding 30 C e with temperature differences of over 15 C between the last calibration and operation e n general carrying out a calibration is the correct process in the event of errors in the acquisition process Such as poor image quality or the lack of a 3D preview In many cases the errors can be corrected in doing so Prior to the first calibration Ensure to take note of the serial number of the calibration set when starting the calibration which must be used at this point The number displayed in the software and the serial number on the calibration set must be identical The serial number of the calibration set to be used has been set on the CEREC AC with Omnicam at the factory Prior to the first calibration after changing the PC and when switching the calibration set you must enter the serial number of the calibration set to be used in the software If you do not wish to do this you can continue with Start calibration gt 16 When entering the serial number the following steps should be followed 1 Start the CalibRegistry exe application on the USB stick included with delivery of the calibration set 2 Enter the 8 digit Sirona ID You can find the ID on the sticker on the calibration set v With all future calibration
15. New as of 01 2015 SIron a The Dental Company CEREC Ortho SW Software Version 1 0 Operator s Manual Table of contents INTFOdUCUO Ne ss oi M ienida 5 1 1 Dear GUSTO MMC a NETTE 5 1 2 Copyright and trademark ooccconcconccccncocnconnconoconnnonnnncnoncnonnnnnnnnononornoncnonones 5 ho Opensource SO Wa Eurasia 5 Ee gol ee Ro fs kc eae ne ne ee te te doo os 6 2 1 General safety information ooccocccccccncconncnoncconncnncnnnonoconnnnannonnnonannnnaninanos 6 2 2 jouucture orthie miallualk tanda ds 7 2 2 1 Identification of danger levels ooccconccccnncoccnconnccnnnccnnnconnnoncnnnnos T 2 2 2 Formats and symbols used occcccncocncccncocncccncnonconncnnnonnnonanonaninaninors 7 2 2 9 CONVENIO Sra Indo eS Eu 8 2 24 PORMALSOT THe TAMU scisco ute vites ii 8 2 3 User Interld6eB uti ee tias 9 ZA MESS Dal MR 9 Za VSS CHOU QE TP EUR 9 2 951 2 JAGOUISITIONU ssec itas las 9 A e unte OR ae MA eae 9 2 e EN P ol Parto m ER Oo 9 2 92 WOOL Wii 10 2 939 ep Mei adit eame Soda ctu Eolo ed eae Eve 10 2 94 AV SECC sssescennced peni eoa centu ab teen crine asc Eon e ex cua cuarta 11 2 9 9 estar WIN O West ticas Hte coe tet db 11 So Ee EU NP 12 3 1 Installing the SOM ANC scii rea Port nce ated die toss 12 3 2 Uninstalling the software ooocccocccooccconcooncoonnconocononornonnnonnnnonocnnoconncannnannnns 12 So UOODVDIOIGCHOLD sas nrbt teta bee a erede tt
16. Position the camera above the last molar on the left of the upper jaw and hold the camera in this position to mark the start of the scan The green start marker A which marks the start of the scan slowly disappears and the scan begins The guide direction for the camera is shown with an arrow Turn the camera lingually and guide it in the direction of the arrow over the lingual surface of the teeth as far as the target marker B on the centerline When the center line has been reached hold the camera in position once more for 3 seconds to mark the end point of the partial scan While doing so the target marker B slowly disappears The software automatically changes to the next Scan occlusal left step Position the camera above the last molar on the left The starting point is marked with a green start marker A and is automatically recognized Guide the camera occlusally as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan vestibular left step Position the camera above the last molar on the left and hold the camera in this position to mark the start of the partial scan Turn the camera towards the buccal and guide it in the direction of the arrow over the arch as far as the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan transversal left
17. TION phase 8 1 Camera view 8 2 3D Preview 8 3 Perform a scan 8 3 1 8 3 2 8 3 3 8 3 4 65 26 060 D3534 D3534 208 06 01 02 01 2015 Camera warm up time occoncccccccocnconoconnonnncononanoconoconocanonaronarnnnnonone Taking acquisitions with the CEREC Omnicam GUIDES SCAMIMING isso dao A ES 8 6 1 gt eL T eee 8 3 3 2 Start the scan proCesS coocccocccocncocccocnconncococonnnonnnnononononans 9 9 9 9 LOWER JAW SCdllusisecn tieniti eio eed ecs dou dae pode d aee dO 0 304 UUDDER JAW SC dai 9 9 3 5 B ccal regISIFaltlOFi canti rta Finishing the phase 19 19 19 20 20 21 21 21 22 22 22 22 22 23 23 24 24 24 24 25 25 25 26 2f 28 28 28 29 29 29 30 30 31 32 34 36 36 Table of contents Sirona Dental Systems GmbH 10 Operator s Manual CEREC Ortho SW MODE DISC did 37 9 1 SeEmodel AXIS a cis of 92 Oheck DIG oen uod a lindos 38 Lo 8 0910 02 M e eee ee 38 gA A A 38 9 0 EIISniNg ANG Pasionaria 38 EXPORTA S Ganaa A 39 10 1 Send model to Sirona Connect portal occocccocccccnconncocnnocnncononanononanonos 39 TOMA LORO PU es 39 1012 Enter Orgel Dalarnas 39 10 1 3 Add additional Information ero EUR bb id 40 TOMA SAO DIN EGA dara a rea 40 TOMES Orla 40 10 2 Send model to ClearCorrect ccoccconnicnncccnconncoonoconccnnnonnnonnnnononononononcncnannns 40 10 21 Log IM TO he portal seio oos tete ate
18. The upper jaw is displayed or hidden Lower jaw With the Lower Jaw button you can display and hide the lower jaw a 1 Click on the View Options button LOT 2 Click on the Lower Jaw button The lower jaw is displayed or hidden Display upper lower jaw transparently You can adjust the transparency of the upper lower jaw continuously 1 Click on the View Options button 2 Place the mouse pointer on the appropriate button press and hold the left mouse button and move the mouse up or down The transparency of the jaw concerned is changed 65 26 060 D3534 D3534 208 06 01 02 01 2015 21 8 ACQUISITION phase Sirona Dental Systems GmbH 8 1 Camera view Operator s Manual CEREC Ortho SW ACQUISITION phase Camera view You can adjust the size of the camera view proportionally 1 Click the arrow at the right edge with the mouse and hold down the mouse button 9 2 Drag the camera view to enlarge or reduce it 9 3D Preview You can freely select the viewing direction of the virtual model in the 3D preview window by using the mouse Rotating a 3D preview 1 Click on the 3D preview with the left mouse button and hold it down 2 Move the mouse The 3D preview is rotated Moving the 3D preview 1 Click on the 3D preview with the right mouse button and hold it down 2 Move the mouse The 3D preview is moved Zooming into out of the 3D preview 1 Click on the 3D preview with the middle mou
19. To do so under Verification you must enter your password and confirm it with Ok The order list is then inserted automatically Order list All sent orders are displayed in the order list The most recently sent order is always at the top You can filter the orders by the various statuses via the filter below the list In order to view the order details you must click the relevant order in the list so that it is highlighted in orange Then you can view the details by clicking on View Order in the step menu Via the step menu you can call up individual items of information on the highlighted order Send model to ClearCorrect Exporting a model to ClearCorrect occurs via the Sirona Connect Portal Model data in the Sirona Connect Portal The Sirona Connect Portal saves the data only briefly The data is promptly deleted for data protection reasons Y You are registered on www sirona connect com as a dentist Y in your account settings C earCorrect is selected as a favored laboratory gt Click on the C earCorrect button 65 26 060 D3534 40 D3534 208 06 01 02 01 2015 65 26 060 D3534 D3534 208 06 01 02 01 2015 Log in to the portal 1 Enter user name and password 2 Activate the relevant option if you want to save the user name and password 3 In the step menu click on Ok v The data are uploaded in parallel to the information being entered in the portal Enter order data You can enter the d
20. a Cure iav t vao Scb Ra d ct 13 3 4 Downloading SoftWaLe exegit Ee ata de obe a rece tb Rod b ol C da Rr 13 20 tard the SOMWal msan stop odes pet URL eo Oo 13 20 e A a 14 3 6 1 Installation of the License Manager Individual o 14 3 6 2 License update without Internet access eeeeeeeeee 14 COMUNA TE 15 4 1 BI Le PA TT EU T UU m 15 411 XBERBEGXOmmIO ITI usara idas 15 4 1 1 1 Resetting Sellos pl dd 15 4 112 CAMA OM aii 16 4 2 SS CIN eee mE 18 65 26 060 D3534 D3534 208 06 01 02 01 2015 4 2 1 4 2 2 4 2 3 5 System menu 5 1 Save scan Reset notes Selecting a language ccoccccccnccocnccocnnocnccnnnonanonnnnnonnnonannonanonanonnnnns Seating position 5 2 Save the scan under a different name cssee menn 5 3 Import scan 5 4 Export scan 5 9 License manager 5 6 Configuration 5 7 Window mode 5 8 Current program version 5 9 Close the software 6 SR cones aideigs a 6 1 Creating a new patient 6 2 Patient search 6 3 Editing patient data o9 dEd ndgapalernbedf ss aee pet aet ai biie 0 9 2 Deleting patients oor roni net a bn aput ed us 7 Menu NIG cc n 7 1 uc A etal DP EE c X X ZW CUR OULMIOUE arcas ens na itd tet sette eoe boe nd USO Oo cen 7 3 Displaying objects 8 ACQUISI
21. ace Phase bar Description For CEREC Ortho SW the digital scan process is divided into three phases ACQUISITION In this phase you can perform the following e Create optical impressions with CEREC camera Lower jaw Upper jaw Buccal bite registration e View a 3D preview of the acquisitions MODEL In this phase you can perform the following e Set the model axis e Check the bite situation e Adjust the virtual models as required EXPORT In this phase you can perform the following e log on to the Sirona Connect portal and transfer models to the laboratory of your choice e send model to Dolphin 3D e send model to ClearCorrect 2 General data Sirona Dental Systems GmbH 2 3 User interface Operator s Manual CEREC Ortho SW Tool wheel The tool wheel makes the standard tools available in the MODEL phase in order to simplify access The tools currently available vary depending on the current step 1 Right click in the workspace The tool wheel opens 2 Click with the right mouse button anywhere in the workspace The tool wheel moves to the position of the mouse pointer 3 Select a tool The selected tool is available The tool wheel closes automatically You also can close the tool by clicking in the workspace with the left mouse button Step menu The ACQU SITION and MODEL phases are divided into steps They are shown in the step menu at the bottom edge of the screen The
22. anyone s skin or mucosal membrane After removing the CEREC Omnicam from its camera holder the temperature of the mirror sleeve drops within a number of minutes 5 minutes to less than 43 C The CEREC Omnicam is therefore suitable for use in the patient s mouth for an unlimited period of time At an ambient temperature from 30 C only select the three lower heater settings NOTICE Image brightness The image brightness during the acquisition is controlled automatically so that there is always optimum image brightness largely independent of the distance between the CEREC Omnicam and the tooth The surroundings of the tooth to be scanned should be as weakly illuminated as possible Avoid any type of external light Switch off the operating light 29 8 ACQUISITION phase 8 3 Perform a scan Sirona Dental Systems GmbH Operator s Manual CEREC Ortho SW IMPORTANT Do not use cotton rolls in the scan area Do not use any cotton rolls in the vicinity of the scan area Should any pieces of cotton roll contaminate this area the acquisitions will be inaccurate Guided scanning Sequence Description For the CEREC Ortho SW Guided Scanning process the complete lower jaw complete upper jaw and the bite situation are scanned The Guided Scanning process guides the user through several partial scans whereby the single quadrants are scanned A complete model is then rendered from the partial scans Scan objects
23. arallel to the information being entered in the portal Enter order data In this step you can select the lab to which the case is to be sent and enter the desired delivery date IMPORTANT Maximum 5 labs You can save a maximum of 5 labs in the list of your favorite labs Via Edit your favourite laboratories you can add labs to or delete labs from the list Under Return Date you can select the delivery date by clicking on the desired day in the calendar Under 7 me you can also state a delivery time Then click on Add Additional Information to go on to the next step 39 10 EXPORT phase Sirona Dental Systems GmbH 10 2 Send model to ClearCorrect Operator s Manual CEREC Ortho SW Add additional information In this step you must add the patient s gender and insurance type Germany only Under Additional Instructions there is a free text field available for additional written information Using the Additional Files function you can send additional files photos Then click Update ist to move to the shopping cart Shopping cart In the shopping cart you can check the order details modify them or delete the order from the shopping cart Your current delivery and invoice address is displayed In the top left corner you can see the progress for loading the model file As soon as the file has been fully uploaded and all the information is correct you can send the order to your lab via Submit Cart
24. ase Operator s Manual CEREC Ortho SW 9 1 Set model axis MODEL phase In the MODEL phase the virtual models are reconstructed based on the acquired image catalogs The Set Model Axis step is required The Check Bite Edit Lower and Edit Upper steps are optional Set model axis Set the axes for model alignment Please ensure a consistent representation here This alignment is required to create optimal view options and initial suggestions 1 Align the model to the orientation of the jaw A Each tooth must be in its respective quadrant Align the incisors based on the displayed center line 2 Align the incisal edges along the displayed line B If necessary rotate the model by clicking on the display holding the mouse button down and dragging the model 3 Align the model for the occlusion plane C If necessary tilt the model by clicking on the display holding the mouse button down and dragging the model 65 26 060 D3534 D3534 208 06 01 02 01 2015 37 9 MODEL phase Sirona Dental Systems GmbH 9 2 Check bite Operator s Manual CEREC Ortho SW Check bite In this step the contact points of both jaws are displayed in order to allow a check against the natural situation The colored displays of the contact points have the following meanings Penetration pressure EN gt 100 um 100 50 um E 50 Oum Distance ES 0 50 um EE 50 100 um E 100 um Edit upper In the Edit Upper ste
25. atic data flow is interrupted the tone signal changes and a start marker is displayed In this case guide the camera to the start marker As soon as the correct position has been found the tone signal changes and the start marker disappears The acquisition procedure continues Switching Omnicam on off You can switch the camera on off by clicking on the Omnicam symbol in vm the bottom left Start the scan process Y The correct working position is taken up Y The teeth are blow dried 2 Change to phase SCAN 65 26 060 D3534 D3534 208 06 01 02 01 2015 31 8 ACQUISITION phase Sirona Dental Systems GmbH 8 3 Perform a scan Operator s Manual CEREC Ortho SW Lower jaw scan Scan the right hand quadrants 1 Remove the CEREC Omnicam from its holder The step list for scanning the lower jaw is displayed V The first step Scan lingual right is selected in the step list 2 Position the camera above the last molar on the right of the lower jaw and hold the camera in this position to mark the start of the scan The green start marker A which marks the starting point for the partial scan slowly disappears and the scan begins The guide direction for the camera is shown with an arrow 3 Turn the camera lingually and guide it in the direction of the arrow over the lingual surface of the teeth as far as the target marker B on the centerline When the center line has been reached hold the camera in position once
26. cense Manager Individual Y The PC is powered up and all programs are terminated 1 Insert the DVD in the DVD drive v The setup program starts automatically 2 If this is not the case run the Sefup exe file in the root directory of the DVD The installation wizard opens 3 Click on the OK button 4 In the next dialog click the Nex button gt The license agreement is shown 5 Read through the license agreement carefully 6 If you accept the license agreement then activate the accept the terms in the license agreement option button and click the Next button 7 In the next dialog click the Custom button 8 Uncheck all options apart from the license manager 9 Inthe next dialog click the Nex button 10 In the next dialog click the nsta button v The program continues the installation routine This may take several minutes 11 Click the Finish button once installation is complete gt The license manager is installed License update without Internet access If the Sirona acquisition unit does not have Internet access itself you can run the license manager on another PC with Internet access You need to remove the USB dongle from the Sirona acquisition unit and plug it into the PC with Internet access The USB dongle is behind the lower cover at the rear side of the Sirona acquisition unit Install the license manager on the PC with Internet access and run the license update 6
27. e entered enough information the bar in the Edit Patient step turns from red to green 3 Click on the Add New Case button eis The program switches over to the ADMINISTRATION phase 65 26 060 D3534 D3534 208 06 01 02 01 2015 23 6 Start window Sirona Dental Systems GmbH 6 2 Patient search Operator s Manual CEREC Ortho SW Patient search Displaying all patients The Show All Patients function can be used to display all patients Searching for individual patients You can view individual patients by searching for them n 1 Click into the search text box 2 Enter the surname or the patient ID P 3 Click the magnifying glass to start The program now shows all the search results Editing patient data Editing a patient card Y You have found the patient with the search function P a 1 Click on the patient card ed 2 Click on the Edit Patient step in the step menu The patient card case view is opened for editing 3 Carry out the changes 4 Confirm your changes by clicking the Ok button The changes are saved in the memory 5 Click the double arrow on the left side of the step menu The patient card case view is displayed Deleting patients Y You have found the patient with the search function 1 Click on the patient card 2 Click on the Delete Patient step in the step menu C 3 Confirm the deletion by clicking the Ok button sn e The patient is deleted 65 2
28. e lower jaw scan 1 Position the camera above the last molar on the left of the lower jaw and hold the camera in this position to mark the start of the scan The green position marker A which marks the start of the scan slowly disappears and the scan begins The guide direction for the camera is shown with an arrow 2 Turn the camera lingually and guide it in the direction of the arrow over the lingual surface of the teeth as far as the target marker B on the centerline When the center line has been reached hold the camera in position once more for 3 seconds to mark the end point of the partial scan While doing so the target marker B slowly disappears The software automatically changes to the next Scan occlusal left step 3 Position the camera above the last molar on the left The starting point is marked with a start marker A and is automatically recognized Guide the camera occlusally as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan vestibular left step 4 Position the camera above the last molar on the left The starting point is marked with a green start marker A and is automatically recognized 5 Turnthe camera towards the buccal and guide it in the direction of the arrow over the arch as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone
29. esired delivery date in this step Under Return Date you can select the delivery date by clicking on the desired day in the calendar Under 7 me you can also state a delivery time Then click on Add Additional Information to go on to the next step Add additional information In this step you must add the patient s gender and insurance type Germany only Under Additional Instructions there is a free text field available for additional written information Using the Additional Files function you can send additional files photos Then click Update ist to move to the shopping cart Shopping cart In the shopping cart you can check the order details modify them or delete the order from the shopping cart Your current delivery and invoice address is displayed In the top left corner you can see the progress for loading the model file As soon as the file has been fully uploaded and all the information is correct you can send the order to your lab via Submit Cart To do so under Ver fication you must enter your password and confirm it with Ok The order list is then inserted automatically Order list All sent orders are displayed in the order list The most recently sent order is always at the top You can filter the orders by the various statuses via the filter below the list In order to view the order details you must click the relevant order in the list so that it is highlighted in orange Then you ca
30. in pop up windows when using the software Many of these messages can be deactivated by clicking on the Don t show this message again check box If this check box is already selected or if a new user uses the software all warnings can be reset here All warnings are then displayed if required Description YES Displays all the deactivated warnings in the workflow again NO Warnings which were previously hidden remain hidden Selecting a language Here you can set the language of the software Once the software is rebooted the language is set to your choice Seating position Here you can adjust the scan display to the seating position of the patient You have the following options e Always next to or facing e Facing for lower behind for upper e Always behind 65 26 060 D3534 D3534 208 06 01 02 01 2015 19 5 System menu Sirona Dental Systems GmbH 5 1 Save scan Operator s Manual CEREC Ortho SW System menu In the system menu you can e Close the opened scan and switch to the start window e Save scan e Save scan as e Import scan e Export scan e Open license manager e Configure hardware and software e Change window mode e Retrieve software information e Close the software Opening system menu 2 Move the mouse cursor to the top of the window or gt Click the start window button The system menu is displayed Closing system menu gt Click the start window button or gt
31. l scan While doing so the target marker slowly disappears The software automatically changes to the next Scan occlusal right step 4 Position the camera above the last molar on the right The starting point is marked with a green start marker and is automatically recognized Guide the camera occlusally as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan vestibular right step 5 Position the camera above the last molar on the right The starting point is marked with a green start marker A and is automatically recognized 6 Turn the camera towards the buccal and guide it in the direction of the arrow over the arch as far as the target marker B on the centerline Ifthe centerline has been scanned a signal tone will be heard The software changes to the next Scan transversal right step T Guide the camera in the direction of the arrow over the marked area from the start marker A as far as the target marker B Ifthe connection is scanned a signal tone will be heard The software changes to the next left hand quadrant step 65 26 060 D3534 34 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH Operator s Manual CEREC Ortho SW 8 ACQUISITION phase 8 3 Perform a scan Scan the left hand quadrant and complete the upper jaw scan 65 26 060 D3534 D3534 208 06 01 02 01 2015 1
32. n of physicians dentists or licensed professionals 65 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 2 General data Operator s Manual CEREC Ortho SW 2 2 Structure of the manual Structure of the manual Identification of danger levels To prevent personal injury and material damage please observe the warning and safety information provided in this document Such information is highlighted as follows DANGER An imminent danger that could result in serious bodily injury or death A possibly dangerous situation that could result in serious bodily injury or death A possibly dangerous situation that could result in slight bodily injury NOTICE A possibly harmful situation which could lead to damage of the product or an object in its environment IMPORTANT Application instructions and other important information Tip Information on making work easier Formats and symbols used The formats and symbols used in this document have the following meaning Y Prerequisite Prompts you to do something 1 First action step 2 Second action step or gt Alternative action Result gt Individual action step see Formats and symbols Identifies a reference to another text used gt 7 passage and specifies its page number Command menu item Indicates commands menu items or quotations 65 26 060 D3534 D3534 208 06 01 02 01 2015 T 2 General data Sirona Dental System
33. n view the details by clicking on View Order in the step menu Via the step menu you can call up individual items of information on the highlighted order 41 10 EXPORT phase Sirona Dental Systems GmbH 10 3 Send model to Dolphin 3D Operator s Manual CEREC Ortho SW Send model to Dolphin 3D IMPORTANT In order to send a virtual model to Dolphin 3D the Sirona Server software must be installed and configured with the Sirona Registry Server Sirona Launcher and RequestScan components You will find more information on this in the Sirona Server user handbook IMPORTANT Exporting to Dolphin 3D is available if CEREC Ortho SW has been started via the Dolphin 3D software 1 Click on the Dolphin 3D button IMPORTANT If the scan is requested via Dolphin 3D exporting to Dolphin 3D starts automatically when switching to the EXPORT phase In this case clicking on the Do ohin 3D button is not required 2 Click the Add button in the window that appears 3 Inthe menu that appears select the menu option mport Upper and Lower Arch Models from CEREC Ortho Scanner While the data is being sent to Dolphin 3D the Loading Arch Models Please wait message is displayed When the data is sent a confirmation message appears 4 Click the OK button in the confirmation message The data has been sent to Dolphin 3D and can be processed there 65 26 060 D3534 42 D3534 2
34. na Dental Systems will be liable to prosecution Microsoft and Windows 79 are registered trademarks Windows is a trademark of Microsoft Corporation All other trademarks are the property of their respective holders Open source software This product contains open source software For more information please refer to the Licence pdf in the installation directory General data Please read this document completely and follow the instructions exactly You should always keep it within reach Original language of the present document German General safety information Only use original software Only use original software or software which has been released by Sirona To produce restorations and equipment manipulated or non released software components must not be used Software and software components must not be installed using incorrect data Please check that each installed component has been granted approval in its country Contact your dealer for more information Models to be checked by trained personnel Every model created with this software must be checked for suitability by a trained person e g orthodontist or dentist before it is digitally saved or exported for further use All orthodontic apparatus manufactured using models from this software must be given to the patient by a trained individual For the USA only CAUTION According to US Federal Law this product may be sold only to or by instructio
35. ograms Sirona Dental Systems CADCAM Configuration The configuration is described in the chapter Configuration JU EE LE oh E E Ll Window mode The Window Mode function can be used to exit full screen mode or enter it again Current program version jE The About function contains information about the current program version Close the software The Exif function can be used to close the software 65 26 060 D3534 22 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 6 Start window Operator s Manual CEREC Ortho SW 6 1 Creating a new patient Start window In the start window you can perform the following e Search in the database e Display patients e Add patients Switch to the start window If you have a scan open you can switch to the start window by closing the scan 1 Open the system menu 2 Click on the C ose button Ifthe current scan has not been saved a prompt appears asking if the scan should be saved 3 Confirm the prompt to save the scan The start window is displayed Creating a new patient In the data structure a patient is uniquely identified by one of the following two entries e Surname first name and date of birth Or e Patient ID Adding patients 1 Click on the Add New Patient button An empty patient card is opened 2 Enter a surname first name and date of birth Or gt Enter the patient ID Once you hav
36. p you can work with the following tool e Cut Use of the individual tools is described in the section Menu field gt 25 Edit lower In the Edit Lower step you can work with the following tool e Cut Use of the individual tools is described in the section Menu field gt 25 Finishing the phase Y The EXPORT phase can be selected gt Click on the EXPORT phase or gt Click on the double arrow The program switches over to the EXPORT phase 65 26 060 D3534 38 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 10 EXPORT phase Operator s Manual CEREC Ortho SW 65 26 060 D3534 D3534 208 06 01 02 01 2015 10 1 Send model to Sirona Connect portal EXPORT phase In this phase you can perform the following e Send model to SIRONA Connect portal e Send model to ClearCorrect e Send model to Dolphin 3D Send model to Sirona Connect portal Model data in the Sirona Connect Portal The Sirona Connect Portal saves the data only briefly The data is promptly deleted for data protection reasons Y You are registered on www sirona connect com as a dentist Y n your account settings at least one favored laboratory is selected gt Click on the Sirona Connect Portal button Log in to the portal 1 Enter user name and password 2 Activate the relevant option if you want to save the user name and password 3 In the step menu click on Ok The data are uploaded in p
37. s GmbH 2 2 Structure of the manual Operator s Manual CEREC Ortho SW Conventions Pressing once and releasing the left mouse button or the left trackball button on the acquisition unit or foot switch Double clicking Pressing twice quickly in succession and releasing the left mouse button or the left trackball button on the acquisition unit or foot switch Moving the mouse On the acquisition unit Moving the trackball in the in one direction corresponding direction Seizing a point Pressing the left mouse button left trackball button on the acquisition unit and ves it pressed Cir N On the keyboard Press the Ctrl and N keys simultaneously Drag amp drop Drag amp drop Press and hold an element e g pictograph and drop onto new potential destination Formats of the manual You can access the manual via the Help button or by pressing F1 You will find the Operator s Manual on the Internet in PDF format e hitp cerec ortho com You can download the current manual in the DOWNLOADS I Downloads area e http www sirona com manuals This format is page oriented and is well suited for printing out the desired pages 65 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 2 General data Operator s Manual CEREC Ortho SW 2 3 2 3 1 2 3 1 1 2 3 1 2 2 3 1 3 2 3 1 4 65 26 060 D3534 D3534 208 06 01 02 01 2015 2 3 User interface User interf
38. se button and hold it down 2 Move the mouse up or down as desired The 3D preview is enlarged or reduced 65 26 060 D3534 28 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 8 ACQUISITION phase Operator s Manual CEREC Ortho SW 65 26 060 D3534 D3534 208 06 01 02 01 2015 8 3 Perform a scan Perform a scan Camera warm up time When switching on the system the camera needs to warm up for 15 20 minutes If the sapphire glass of the Omnicam is not sufficiently warm it may steam up during the acquisition As such it is not possible to carry out the exposure Following use always position the Omnicam on the heater plate From CEREC Ortho SW software version 4 0 4 and unit serial number 121 001 you can now set the end temperature to which the camera heater warms the Omnicam mirror sleeve 1 In the software navigate to the system menu and click on the Configuration button Click on the Devices button Click on the Omnicam button Click on the Camera Heater Settings button Use the slider to adjust the temperature E E i Taking acquisitions with the CEREC Omnicam Hot surface The output window of the CEREC Omnicam is preheated in the camera holder When removing the CEREC Omnicam from its holder the surface temperature of the mirror sleeve can be up to 51 C This may cause an unpleasant heat sensation on contacting a person s skin or mucous membrane These temperatures will not damage
39. step menu changes depending on which phase the current scan is in This menu guides you through the process step by step The double arrow keys can be used to switch between steps and phases Mandatory steps Mandatory steps are marked with a red or green bar Red bar The step has not yet been completed successfully Green bar The step has been completed successfully Optional steps Optional steps do not have colored bars If you switch between the steps using the double arrow keys the optional steps are skipped 65 26 060 D3534 10 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 2 General data Operator s Manual CEREC Ortho SW 2 3 User interface System menu In the system menu you can e Switch to the start window to start a new scan e Save scan e Save the scan under a different name e Import scan e Export scan e Open license manager e Configure hardware and software e Change window mode e Retrieve software information e Close the software The start window In the start window you can perform the following e Search in the database e Display patients e Add patients 65 26 060 D3534 D3534 208 06 01 02 01 2015 11 3 Getting started Sirona Dental Systems GmbH 3 1 Installing the software Operator s Manual CEREC Ortho SW Getting started Installing the software Y The PC is powered up and all programs are terminated Y You have downloaded the installation file from the Internet and
40. tools you can reset changes that were made with the tool 25 7 Menu field Sirona Dental Systems GmbH 7 2 Tools Operator s Manual CEREC Ortho SW Cut out model areas Regions of the model can be cut off using the Cuf function The cut plane lies orthogonal to the screen plane Remove model area The Discard Part function enables model areas to be removed When performing this activity be careful not to accidentally cut out any areas that e g are located behind the model or are otherwise cut away from the line 1 Click on the Cut button 2 Begin the cut line with a double click 3 Click to set additional points 4 Finish the cut by double clicking The model area is cut off Invert model area With the nvert Selected function the model area that is cut out can be inverted X The Cut tool is selected X You have created a cut gt Click on the nvert Selected button The model area which was cut out is displayed The rest of the model area is hidden Tip You can invert the model area that is cut out by double clicking on the semitransparent cut out area 65 26 060 D3534 26 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 7 Menu field Operator s Manual CEREC Ortho SW 7 3 Displaying objects Displaying objects Upper jaw With the Upper Jaw button you can display and hide the upper jaw 1 Click on the View Options button 2 Click on the Upper Jaw button
41. work involving the CEREC Ortho SW software you will be prompted to use this calibration set Start calibration 1 In the software navigate to the system menu and click on the Configuration button 2 Click on the Devices button 0 Click on the Omnicam button 4 Click on the Ca ibrate button v The camera view is displayed in one window 5 Follow the prompt to set up the calibration set with the corresponding ID number At this point check that this ID matches the ID on your calibration set If this is not the case go back to Prior to the first calibration 16 65 26 060 D3534 D3534 208 06 01 02 01 2015 Sirona Dental Systems GmbH 4 Configuration Operator s Manual CEREC Ortho SW 4 1 Devices Calibrate the camera 1 Remove the protective cap from the calibration set 2 Mount the calibration set on the tip of the camera until it locks into place 3 Secure the CEREC Omnicam in the calibration set using one hand Ensure that the external calibration set screw is fully screwed in in a clockwise motion until it gently locks into place 4 Click the OK button on your The measuring process starts The software prompts you to proceed to the next latching 5 Turn the screw counter clockwise until you reach the next latching point 6 Click the OK button on your In doing so ensure that the CEREC Omnicam does not move The software confirms the calibration process