Spectrafuge 7M Instruction Manual Book Form
Contents
1. Technical Services Department at 732 417 0700 Our Service Department is staffed from 8 30am to 5 00pm EST Monday through Friday Our 24 hour fax number is 732 417 1750 Electronic mail may be sent to labnet labnetlink com Please be sure to have the unit s serial number located on the back panel of the instrument available when calling Should an item require return to Labnet for service a return authorization RA number must first be received from Labnet Items sent without an RA number will not be accepted 8 Determination of g values The centrifuging radius of the 1 5ml rotor is 6 3cm See the follow ing page for the formula for determining g values and a quick reference chart 1 3 personnel The rotor has not been installed securely on the motor shaft Rotors and accessories not belonging to the standard range are being used without permission being obtained from the manu facturer to use such rotors and or accessories in the centrifuge Exception Microcentrifuge tubes made of plastic normally available in the laboratory The centrifuge is located in an explosive atmosphere Materials to be centrifuged are combustible and or explosive Materials to be centrifuged are chemically reactive The rotor load is not properly balanced Damage to the rotor or centrifuge is evident The rotor has been dropped or abused even though damage cannot be observed Technical data Manufacturer Labnet International PO Box 841 Woodbri
2. fuses Troubleshooting Where to Call Determination of g values 12 12 13 1 General Information This manual provides important safety information for the Spectrafuge 7M laboratory microcentrifuge It should be kept near the centrifuge for quick and easy reference 1 1 Description The Spectrafuge 7M is a small benchtop centrifuge designed for separation of various research samples A 12 x 1 5ml rotor for micro samples is supplied with the unit Adapters are available for tubes smaller than 1 5ml The Spectrafuge 7M reaches speeds of up to 10 000rpm 7176 x g 1 2 Safety precautions Note All users of the centrifuge must read the Safety Precautions section of this manual before attempting to operate the unit manufacturer the protection provided by the equipment If this equipment is used in a manner not specified by the may be impaired Thoroughly inspect the centrifuge rotor and accessories for mechanical and chemical damage prior to each use Mechanical damage can often be visually observed and includes cracks scratches or gouges on the centrifuge lid or rotor Chemical damage can often be visually observed and includes crazing peeling or similar deterioration of the inner cavity lid rotor or surfaces See section 5 3 for detailed instructions on inspecting the rotor The rotor cover provided is not intended to be a bioseal as defined by IEC 21010 2 020 for containment of pathogenic microbi
3. lock Open manually and have unit serviced No power from PC board Call for service Lid lock is jammed Call for service Centrifuge is not receiving power See Centrifuge will not start Centrifuge cannot be started although power is on Possible reason Solution Possible reason Solution 7 Where to call Lid not closed correctly Close lid correctly No speed or time has been selected Set speed and or time Should you have any questions about the Spectrafuge 7M or its accessories please call Labnet s Customer Service Department at 732 417 0700 Customer Service is staffed from 8 30am to 5 30pm EST Monday through Friday Our 24 hour fax number is 732 417 1750 Inquiries may also be sent via our electronic mailbox at labnet labnetlink com 12 5 4 5 5 The rotor may be cleaned using methyl or isopropyl alcohol soap solutions heptane hexane naptha or Freon TF cleaner Do not use corrosive cleaning agents with partially halogenated hydro carbons such as 1 1 1 trichloroethane phenol methylene chlo ride or ketones such as MEK When necessary the rotor and rotor cover may be autoclaved at 115 kPa absolute pressure 121 to 124 C Note autoclaving decreases the usable life of the rotor If the rotor is auto claved frequently once a day the rotor should be replaced every 3 months If the rotor is autoclaved more than once a day it should be replaced after 60 autoclaving cycles All other rotors sh
4. Instruction Manual Labnet Spectrafuge 7M Microcentrifuge Labnet International PO Box 841 Woodbridge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 email labnet labnetlink com IMPORTANT NOTICE The use of IsoQuick brand nucleic acid extraction kits with this product is not advised If the IsoQuick extraction matrix or dye components come in contact with the Ultem rotor used in this product they may cause cracks or striations Cracked rotors may disrupt during use If any IsoQuick brand components come in contact with the rotor in this product immediately replace the rotor Please fully read the Rotor Care or Rotor Usage Guide section of this manual to learn more about rotor safety Copyright August 2001 Labnet International All rights reserved Reproduction without prior written permission is prohibited Information contained herein is subject to change without notice Notes Table of Contents General Information Description Safety precautions Technical data Accessories supplied with unit Warranty Installation Unpacking the centrifuge Required space Installing the centrifuge Installing the Rotor Rotor and accessories Mounting and securing the angle rotors Removing the rotor Overloading rotors Operation Closing the lid Lid release Lid lock Speed selection Selection of operating time Service and Maintenance Service Cleaning Rotor care Disinfection Replacing
5. bling it to achieve speed more guickly Spinning without the rotor cover in place may result in motor failure Although the rotor cover is not intended to be a bioseal it does provide an extra layer of protec tion against contamination in the event of tube breakage Removing the rotor The rotor is removed simply by grasping the sides and pulling up firmly Be sure to lift upward in a straight vertical motion Overloading rotors The maximum load of the rotor and the maximum speed have been established by the manufacturer Do not attempt to exceed these values The maximum speed of the rotor has been mea sured for liquids having a homogeneous density of 1 2g ml or less In order to centrifuge liguids with a higher density it is necessary to reduce the speed Failure to reduce the speed may result in damage to the rotor and centrifuge The 6 revised maximum speed can be calculated with the following formula _ 1 2 Reduced speed n TETE TNI ENET x max speed na Example Where the density of the liquid is 1 7 the new maximum speed would be calculated as follows n gt x 10 000 7058 rpm If in doubt concerning maximum speeds please contact the manufacturer for assistance 4 Operation ATTENTION Never attempt to operate the centrifuge with rotors or adapters that show signs of corrosion or mechanical damage Never centrifuge strongly corrosive materials that may damage the rotors or accessories Fi
6. dge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 Type Spectrafuge 7M Dimensions Width 8 25 inches Depth 8 9 inches Height 7 6 inches Maximum speed 10 000rpm Maximum RCF 7176 x g Maximum volume 12 x 1 5 2 0ml Admiss density 1 2kg dm Electrical fuse rating 120VAC 50 60Hz 0 5 amps 250V 0 5AT 230V 50 60Hz 0 25 amps 250V 0 25AT N 1 4 Accessories supplied with centrifuge 1 5 Each unit is supplied with one 12 x 1 5ml rotor instruction manual rotor usage guide warranty card and power cord Warranty This centrifuge has been subject to thorough testing and quality control In the unlikely event of a manufacturing fault our one year warranty from the date of delivery covers the centrifuge and rotor This warranty becomes invalid in the case of incorrect operation use of nonstandard spare parts or accessories and unauthorized modification of the rotor or centrifuge The rotor warranty does not cover normal wear and tear or damage or weakening caused by radioactive contamination or frequent autoclaving Labnet reserves the right to make technical modifications Statements contained herein are not to be considered binding 2 Installation 2 1 2 2 Unpacking the centrifuge Before unpacking the centrifuge inspect the outside of the carton for any shipping damage The centrifuge is delivered in a carton with protective cushions Remove the centrifuge from the carton Retain the carton and cushion
7. ervice should be performed by authorized qualified personnel only Repairs performed by unauthorized personnel may void the warranty Cleaning the centrifuge Always keep the centrifuge housing rotor chamber rotor and rotor accessories clean All parts should be wiped down periodi cally with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied with a soft cloth Excessive amounts of liquid should be avoided Liquid should not come into contact with the motor After cleaning ensure that all parts are dried thoroughly by hand or in a warm air cabinet maximum temperature 50 C Rotor care The rotor should be inspected before each use for mechanical or chemical damage The inspection should focus on the following areas The underside of the rotor including the area around the hub See figure 4 e The web area between the sample tube slots See figure 5 e The outer rim of the rotor The rotor is comprised of two pieces which have been ultrasonically welded together any movement felt between the bottom and top pieces indicates that a weld has been broken rendering the rotor unsafe See figure 6 Mechanical damage can often be visually observed and includes cracks scratches or gouges Hidden mechanical damage may be detected by an in crease in noise or vibration during a spin Again check the rim for looseness See figure 6 Chemical damage appears as discolorat
8. f heat may occur within the chamber Installation Make certain that the timer is set to the off position Before operating the centrifuge check that the power supply corre sponds to that on the manufacturer s rating label then connect the power cord to the centrifuge and the power supply ATTENTION The timer must be in the OFF position before connecting the power cord Failing to place the timer in the off position may result in damage to the centrifuge and injury to personnel 3 Installing the rotor Rotor and accessories The following rotor and accessories are available for the Spec trafuge 7M Angle rotor for 12 x 1 5ml tubes Order no C 9312 supplied with unit Tube measurement 1 5ml 10 x 40mm Max speed 10 000rpm Centrifuging radius 6 3cm RCF g value 7176 x g Replacement rotor cover pack of 2 Order no C 9312 RC Adapter for 0 5ml tubes Order no C 1205 Tube measurement 8 x 30mm Max speed 10 000rpm Adapter for 0 4ml tubes Order no C 1206 Tube measurement 6 x 47mm Max speed 10 000rpm Adapter for 0 2ml tubes Order no C 1222 Tube measurement 6 x 21mm Max speed 10 000rpm Always use the serosal seal for Melk yin Ppottorrmanca Fi 6 Checking the outer ri the rotor Figure 3 Balanced rotor load TREE PERNE panoji ke TOLON 3 10 5 Service and Maintenance 5 1 5 2 5 3 Centrifuge service The motor in the Spectrafuge 7M requires no routine mainte nance Any required s
9. gure 4 Spectrafuge 7M control panel layout 4 1 Closing the lid After the rotor has been properly secured and loaded close the centrifuge lid making sure that the interlock has been engaged 4 2 Lid release The lid button located on the front panel above the speed control must be pressed to open the lid after a centrifuge run has been completed Note that the lid button will not operate until the rotor has come to a full stop 7 WARNING Do not attempt to open the lid of any centrifuge until the rotor has come to a complete stop In the event of a power failure or malfunction it may be neces sary to open the lid manually 1 Disconnect the power cord from the wall socket 2 Remove the plastic plug located on the left side panel of the unit near the front 3 Pull the cord attached to the plug to open the lid lock manually 4 3 Lid lock The centrifuge can be started only with the lid securely closed When the rotor begins to accelerate the lid button becomes inoperable Do not attempt to open the lid until the lid lock indicator turns off 4 4 Speed selection The speed rpm can be selected to 10 000rpm with the knob labeled speed The scale is directly proportional to the speed a setting of 5 corresponds to 5000rpm a setting of 9 corresponds to a speed of 9000rpm etc 4 5 Selection of operating time Operation of the centrifuge begins when the timer knob is turned clockwise to set a run time For ru
10. ion crazing granulation peeling or similar deterioration of the finish If any of the above are noted discontinue use immediately 3 2 3 3 3 4 Mounting and securing the angle rotor The rotor supplied with the Spectrafuge 7M has been precision balanced for optimum performance and guiet operation at maximum speed To minimize vibration and maintain guiet operation the rotor should be installed with the arrow mark on the bottom of the rotor aligned with the mark on the rotor snap With the markings on the rotor and snap aligned center the rotor over the hub and push down slightly keeping the rotor flat and evenly spaced within the centrifuge chamber Push the rotor down until it locks into position When loading the rotor refer to figure 3 located on the opposite page Loading in the pattern indicated will ensure a balanced load Tubes to be loaded should be filled egually by eye The difference in the weight between the tubes should not exceed 0 1 gram A partially loaded rotor may be centrifuged if the loading scheme for balancing a rotor given in figure 3 is followed Never spin samples which are flammable explosive or which chemically interact vigorously when mixed Discard any rotor or rotor cover that becomes contaminated by radioactive samples as the exposure may reduce the strength of the rotor material Always use the rotor cover The rotor cover is designed to minimize wind resistance on the spinning rotor ena
11. n times less than 5 minutes turn the knob clockwise past the halfway point and then counter clockwise to the desired time For run times longer than 5 minutes turn the knob clockwise to the desired time When the preselected time expires the centrifuge will stop automatically To stop the centrifuge prior to the expiration of set time turn the timer knob to the zero position Some models are equipped with a timer that includes a hold position Continuous operation of these models may be achieved by turning the timer knob firmly to the left The centrifuge will continue to operate until the knob is turned to the zero position Note The timer knob may be turned in either direction during operation of the centrifuge without damage to the timer mecha nism
12. ological specimens during use While a small amount of vibration during rotor acceleration or deceleration is normal immediately stop the centrifuge is excessive vibration is evident at any time Do not operate the centrifuge if any of the following conditions exist The centrifuge has not been installed properly The centrifuge is partially dismantled Service has been attempted by unauthorized or unqualified 1 RELATIVE CENTRIFUGAL FORCE Radis Speed qin cm tin HERS pas poe i p Relative Centrifugal Force Force Relative 15 000 Centrifugal The relative centrifugal force Force z G Force can be estimated using ix qh the chart on this page or by i applying the following ne TAG MO yo lt IK r pi Z i BAR E whore r radius in centimeters E a speed in RPM The radius from the canlar of the rotation axis ta the bottom or outermost partion of the Lest tubo i should be used AGE is expressed relative to the force of thg earth s gravity ie d To use this chart find the radius value on the radius scale Place the edge of a ruler on the value Place the right side edge of the ruler on the speed scale at the desired speed The estimated RCF can then be read from the RCF scale where the ruler edge passes through it This chart can also be used to determine the proper speed for the desired RCF value z 14 Should your Spectrafuge 7M require service please call Labnet s
13. ould be replaced on an annual basis The rotor should mot be immersed in any solution below 10 C such as LN2 or LCO2 Handle the rotor carefully It should not be tossed or dropped Dropping can create small cracks that may or may not be visible to the eye A crack in the rotor can result in a rotor disruption during a spin Any rotor which has been subject to abuse regardless of any evidence of physical damage should be dis carded Any rotor and or rotor cover that becomes contaminated by radioactive samples should be discarded as the exposure may reduce the strength of the rotor material Disinfection Should a spill of infectious materials occur within the rotor or rotor chamber the unit should be disinfected This should be performed by gualified personnel with proper protective eguip ment Replacing fuses Check the fuse when it is recommended in the Troubleshooting Guide located in this manual The fuse holder is located in the power inlet on the rear of the unit Disconnect the power cord from the power inlet Open the fuse holder drawer by inserting a small screwdriver under the tab and prying it open Remove the innermost operative fuse from its retaining tabs and replace the fuse if necessary A spare fuse is located in the outermost chamber of the fuse drawer Replace only with a fuse of exactly the same value as the original Fuse type may be found in the 11 2 3 3 1 exhaust of a compressor as a buildup o
14. s until it has been established that the centrifuge is working properly Inspect the centrifuge for any visible signs of shipping damage Shipping damage is the responsibility of the transportation carrier Any claims for damage must be filed within 10 days The accessories and rotor usage guide supplied with the centri fuge should be kept with the instruction manual near the centrifuge s place of installation Required space The centrifuge should be installed on a rigid even surface such as a stable laboratory bench cabinet etc To guarantee suffi cient ventilation ensure that the centrifuge has at least 15cm 6 inches of free space on all sides including the rear The centrifuge should not be located in areas subject to exces sive heat such as in direct sunlight or near radiators or the 3 Technical data section of this manual 6 Troubleshooting Guide Please refer to this guide before calling for service Centrifuge will not start Possible reason Solution Possible reason Solution Lid lock will not release Possible reason Solution Possible reason Solution Possible reason Solution Possible reason Solution No power supply Check that power is being supplied to the outlet Check that the power cord is plugged into both the wall outlet and the back of the centrifuge Check that the power cord in intact and not damaged Blown fuse Check fuse and replace if necessary Defective lid