1. 0 00o0ooooooooc cece reece eee eee eee essen neneeees 8 1 Thermo Scientific Touchscreen User Interface for Sorvall LYNX i ee i Screen Views The following manual describes the touchscreen or graphical user interface GUI for Thermo Scientific Sorvall LYNX superspeed centrifuges Contents e Main Screen on page 1 1 e Lighthouse Mode on page 1 5 Main Screen The main screen is subdivided into the following sections Ready o Status O som 20 ACCEL max 00 02 00 TEMPERATURE 20 Parameters I EIA gt Control and Configuration USER NAME Thermo Scientific Touchscreen User Interface for Sorvall LYNX 1 Screen Views Main Screen Status The status of the centrifugation process is displayed at the top of the main screen Ready Status Progress IDLE 0 20 Temperature rpm L Speed Status In the Time mode the remaining time for the centrifugation process is shown here In the Hold mode the elapsed time is displayed Progress The curve diagram is divided into the sections Accelerate Centrifugation and Deceleration Temperature The current temperature in the rotor chamber is shown here Speed The current speed of the rotor is shown here The remaining time is shown for a centrifugation process in progress The progress bar is used to determine in which phase the centrifugation process is currently in The acceleration or deceleration curve correspon
2. Configuration 4 18 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Tooltip Mode Thermo Scientific As with the configuration menu the Tooltip mode cannot be accessed while a centrifugation run is in progress You can activate the Tooltip mode while the rotor is at standstill to view information about the individual buttons Use the button at the bottom right hand corner of the main screen to activate the Tooltip mode Touch any element on the screen to view detailed information about it The actual function of the element is not activated or initiated while in the Tooltip mode Touching the screen again de activates the Tooltip mode The Tooltip mode is available for every button in the main screen and most of the elements on the configuration screens Ready IDLE O mn 20 Acceleration Profile Opens a dialog to select the acceleration profile TEMPERATURE 20 USER PROGRAM F12 6X500 LEX i Touchscreen User Interface for Sorvall LYNX 5 1 Rotor The rotor is detected automatically by the centrifuge If the centrifuge is familiar with the rotor type the key parameters for the rotor being used will be displayed in the rotor data field Ready 2 0 oc ROTOR PARAMETERS IDLE 0 Inserted Rotor ACCEL TIME DECEL F12 6x500 LEX max 00 0 2 00 max Material Carbon Fiber Capacity 6 x 500 mi k Factor SPEED TEMPERATURE l Tube Angle y Speed at 4 C 20 i 1 000 2 Tg rpm
3. Thermo Scientific for Sorvall LYNX Superspeed Centrifuges Instruction Manual 50136520 c 07 2015 ee i Table of Contents Chapter 1 SCICOM VICWS wise A o Ou ds 1 1 MCA AA 1 1 Lighthouse Mode s oie ies acceso nec ndos tes EESE ERa EEE doo ae 1 5 Chapter2 APA E yg seer es ceasncespeesanncdaesess eee ase aes 2 1 Chapter3 Entering Parameters 00 cece eee eee eee eee eee eees 3 1 Wemperiire e eponesa lares essa eee beds adorado aa se 3 1 idiota To lb Y AREA PU II E 3 2 MAA UNES II 3 3 Spestand ACE Vales met get posesiones aporto sent 3 4 Acceleration Deceleration Profile o o oooooooooooooooooooorrrr eee eens 3 6 Chapter4 Configuration a lt eoocorrtasdsrccanaizo rrollo sabed 4 1 PICS MES Op rn a eeee ye cae ph Sect h elise ngewe oes 4 2 Rus Consul cuele prototipos apeadero eat 4 3 Senos ee ee osos odos dale ee o eee ee aso a 4 8 ChapterS Tooltip MOGG coescicios toners ira 5 1 Ghaplero Ol ass cesraci arrasaron serio EE pe bane cdeses 6 1 ROtOrIMDOlES cueorccee cots Pops sep Peed eee pesen EER EEE E E e a 6 1 Chapter7 Thermo Scientific Centri Vue Application 0 0 c cece eee eee eee eens 7 1 ReQUIRCMVENES yoo ott ews dete dci Eee eh wee de de wee eee ahs gee aman dk eee 7 1 OUI E o IIA 7 1 Connectivity Plug In Touchscreen User Interface ooooooooooroorcrrrrr eee 7 6 E o o5 a da oes oe be cbse Oakes iad take eh eet sea ee ce beaehessas 7 7 Chapter8 REST Webserver
4. ace 0 00E00 powerDown false rer putt roms 0 Stace STOPPED temperature 0 time null i error null name My Centrifuge progran TH rotorName F10 4x1000 LEX setValues accelerationProfile 9 ace T2 22E02 decelerationProfile 9 rer MULA FpD 900 temperature 0 time null y user 8 4 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific GET lt device ip gt lt port gt getstate Query status and name of the centrifuge Request No data Response t name lt centrifuge name gt powerDown lt true or false gt state lt state identifier gt I Example t name My Centrifuge powerDown false state STOPPED i 8 REST Webserver Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 8 5 Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg 37520 Osterode am Harz Germany thermoscientific com centrifuge 2015 Thermo Fisher Scientific Inc All rights reserved Delrin TEFLON and Viton are registered trademarks of DuPont Noryl is a registered trademark of SABIC POLYCLEAR is a registered trademark of Hongye CO Ltd Hypaque is a registered trademark of Amersham Health As RULON A and Tygon are registered trademarks of Saint Gobain Performance Plastics Alconox is a registered trademark of Alconox Ficoll is a registered tradem
5. centrifuge Speed specifies the target speed of the centrifuge Temperature specifies the target temperature of Overview the centrifuge Program display shows if a preset program is used Rotor display shows the name of the installed rotor Request remote select button to switch to the Request Remote Access menu Overview select button to switch to the Overview menu Discovery select button to switch to the Discovery menu Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 9 7 Thermo Scientific Centri Vue Application Centri Vue App Detail View during Error If an error occurs to the centrifuge a message will pop up in the detail view The message opens every time you switch to the detail view until the error is fixed Selecting the big red cross in the detail view opens up the message manually Serious Centrifuge Error 2015 04 16 04 17 00 PM x A serious centrifuge error occurred Please restart the centrifuge Close Adding a Centrifuge Search on Network If the IP address of the centrifuge is not known the centrifuge can be discovered with discovery functionality Follow the provided steps 1 Switch to Discovery menu 2 Open the list New Discovered devices 3 The found centrifuges are displayed in a list of names and IP addresses NOTE If no name is entered in the centrifuge the name field remains empty 4 Select the desired centrifuge 7 10 Touchs
6. Discovery to switch to the discovery menu q Q 4 7 8 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific Detail View 7 Thermo Scientific Centri Vue Application Centri Vue App To enter the detail view of a locally stored centrifuge select a centrifuge entry The detail view is very similar to the main screen of the touchscreen user interface software and shows the current values of the centrifuge at a refresh rate of one second All parameters are read only they can only be changed by accessing the LYNX centrifuge directly l 10 11 12 13 14 15 16 Thermo Scientific Acceleration specifies the acceleration of the Time specifies the duration of the centrifuge Deceleration specifies the deceleration of the User Optional If a user login is necessary to start a Msn N E the centrifuge the username is shown here Name of centrifuge shows the name of the 3 Door Open centrifuge If no name is set yet for the centrifuge the IP address of the centrifuge is shown 4 IDLE Time display shows processing time or remaining time depending on the program selection Status display shows the progress of the centrifuge graphically TIME Current speed shows the current speed of the 00 20 00 centrifuge 8 Current temperature shows the current JO SPEED 7 TEMPERATURE temperature of the centrifuge 29 000 rpm 20 centrifuge
7. in this menu Ready Centrifuge XYZ THF 012 345 6789 1234V2 1 0 8 589V125 4719V226 My Laboratory IPV6 DHCP 10 48 137 244 255 299 292 0 10 48 136 2 10 48 137 32 Settings Device Contact Configuration Touchscreen User Interface for Sorvall LYNX 4 17 4 Configuration Settings Contact Contact data for Thermo Fisher Scientific is given in the contact section in the event that you have any queries about the device or experience technical difficulties If a caretaker is given for your centrifuge you can enter the data for this person in this section so that your employees know who to contact for any questions they may have Also contact the person specified here if changes are to be made which required the master PIN Any changes to the name and contact data of the person responsible for the centrifuge require the master PIN Ready Instrument Owner NAME PHONE NUMBER resremesanone voa2sesere9 gt E_MAIL ADDRESS mail caretaker com Thermo Fisher Technical Service First Name Surname Phone Number 00 0 1234 567 8 Fax Number 00 0 1234 567 9 E Mail Address technical service thermofisher com Postal Address 123 Road City State 45678 Country Thermo Fisher Sales Service First Name Surname Phone Number 00 0 1234 567 8 Fax Number 00 0 1234 567 9 E Mail Address sales service thermofisher com Postal Address 123 Road City State 45678 Country User IDs Settings Device Contact
8. until the DS we rotor has come to a complete stop See Acceleration Deceleration Profile on page 3 6 6 566 10 Centrifugation has been successfully completed The rotor has Complete come to a complete stop and the centrifuge door can be opened pe amp O on 4 Parameters You can set the setpoints for centrifugation in the Parameters section for more information see Entering Parameters on page 3 1 Touch one of the buttons to open a new window to enter the corresponding value Thermo Scientific Touchscreen User Interface for Sorvall LYNX 1 3 1 Screen Views Main Screen Acceleration mode Time Braking profile Here you can select an Select the duration for centrifugation and select the acceleration profile Level the Time mode here deceleration profile 1 9 0 9 here TIME Speed 00 02 00 Set the speed as RPM or RCF here SPEED TEMPERATURE 1 000 rpm 20 Temperature Set the temperature in C for the rotor chamber here Program Select a program for centrifugation here The reviously selected program will always be P y prog y Rotor displayed initially You can change the parameters o once you have selected a program Changing a The rotor currently in use is displayed parameter will cancel the program here in this case F12 6X500 LEX Control amp Configuration In this section you can start and halt centrifugation You can also defin
9. z Pb Start centrifugation Time expired Stop Error Standard mode Standard mode centrifuge halted centrifuge running 1 6 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Centrifugation You can define the parameters for centrifugation via the main screen see section Entering Parameters on page 3 1 Touching a button opens a corresponding window in which you can enter the new value A rotor does not have to be already in use to enter this data The speed is adjusted automatically to the maximum when a rotor is installed As an alternative to manual input of parameters you can also use a program in which values have already been defined Refer to section Programs on page 4 3 for information on how to define and activate a program For example you can select the predefined program PRE COOLING for details see Precooling on page 3 2 Centrifiguation can be started when all of the input parameters are acceptable when a rotor is in use and when the door is closed If an error occurs a large red X will be displayed in the main screen in the Lighthouse mode Touch the screen to obtain information about the exact information about the error It is only possible to open the centrifuge door using the Open Door button in the main screen if the centrifuge is at standstill Device Settings on page 4 17 contains information about activation and de activation of the acoustic signal on completio
10. C 24471 g ll os USER NAME Rotor import When using a rotor with which the centrifuge is not familiar the window for importing rotor data via a USB stick is displayed This dialog guides the user through the import process Import of the rotor data is begun when the USB stick is connected to the centrifuge Progress of the import is indicated by the progress bar The remaining progress is indicated as a percentage under the progress bar along with the remaining time in minutes and seconds MM SS On successful completion of rotor data import the rotor information field is displayed containing all of the data for the newly added rotor You must then restart the centrifuge by touching the Restart button Thermo Scientific Touchscreen User Interface for Sorvall LYNX 6 1 oe Thermo Scientific Centri Vue Application Sh Requirements e Thermo Scientific Sorvall LYNX centrifuge with Thermo Scientific Touchscreen User Interface software version 1 2 2 or higher e Thermo Scientific Centri Vue application version 0 1 2 or higher e Local Area Network LAN Quick Guide This quick guide describes the steps how to connect the centrifuge with the Centri Vue application Detailed descriptions are given in Chapter 4 1 Install the touchscreen user interface software on the centrifuge 2 For remote control of the centrifuge you have to make some option changes in the touchscreen user Activate Remote Access i
11. adable by the centrifuge Users can be exported via USB stick for further processing To do this connect an USB flash drive to the centrifuge and select the option Export List Existing users will then be exported to the USB stick as CSV files Users in this CSV format can be edited and imported using any spreadsheet program such as Excel The USB stick that you use must have FAT32 formatting in order to be readable by the centrifuge Access Control In the Settings submenu of the configuration menu you can define whether log in is required to start a centrifugation run Device policy settings can only be made after entering a master PIN 5 digit see User Roles on page 4 15 The factory default for the device master PIN is 12345 As the first step change the device master PIN The following access control options are possible e Open environment 4 10 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Thermo Scientific 4 Configuration Settings e Log in is not required in this mode Centrifugation can be started by any user e Closed environment PIN protected e In this mode log in using a PIN must be entered to make any settings and to start a centrifugation run Access control Options Open environment e No user specified e Optional user selection e Mandatory user selection Closed environment e Login with name and PIN I 11 ACCESS CONTROL Protected by User PIN a Closed environment E Lock af
12. arameters Speed and RCF Value RCF Relative centrifugal force Units Xg G FORCE Ez 14 000 Maximum set value 24 471 09 The defined speed may not exceed the maximum speed of the rotor being used A notice is given below the button for the defined speed indicating the maximum permissible speed for the rotor If the entered value for the speed exceeds the value entered for the rotor a window is displayed with an error message indicating that the entered speed must be matched to the maximum permissible speed for the rotor Maximum set value 10 500 rpm 1112 3 4 ILs ILe L7 ILs Lo 000 o cE If the speed is changed while centrifugation is in progress the current speed is adjusted immediately to the new value by means of acceleration deceleration The newly set speed then applies for the remaining duration of centrifugation Touchscreen User Interface for Sorvall LYNX 3 5 3 Entering Parameters Acceleration Deceleration Profile Acceleration Deceleration Profile There are 9 profiles available for acceleration 1 9 and ten for deceleration 0 9 You can select the acceleration deceleration profiles via the main screen TIME DECEL 5 5 00 02 00 1 SPEED TEMPERATURE 1 000 m 20 USER PROGRAM F10 4X1000 LEX After clicking on the appropriate button a window opens in which the required profile can be selected by moving a slide selector over it You can also define a pr
13. ark of GE Healthcare Haemo Sol is a registered trademark of Haemo Sol Triton X 100 is a registered trademark of Sigma Aldrich Co LLC Valox is a registered trademark of General Electric Co All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Shown pictures within the manual are examples and may differ considering the set parameters and language Australia 61 39757 4300 Italy 39 02 95059 552 UK Ireland 44 870 609 9203 Austria 43 1 801 400 Japan 81 3 5826 1616 USA Canada 1 866 984 3766 Belgium 32 53 73 42 41 Netherlands 31 76 579 55 55 China 800 810 5118 or New Zealand 64 9 980 6700 Other Asian Countries 852 2885 4613 400 650 5118 Nordic Baltic CIS countries Countries not listed 49 6184 90 6000 France 33 2 2803 2180 358 10 329 2200 ermo Germany national toll free 0800 1 536 376 Russia 812 703 42 15 Germany international 49 6184906000 Spain Portugal 34 93 223 09 18 SCIENTIFIC India 91 22 6716 2200 Switzerland 41 44 454 12 12 l al Part of Thermo Fisher Scientific 50136520 c original instruction manual published in 07 2015
14. as hh mm The set time is counted down when centrifugation is in progress Initial value Defined duration hh mm 00 HOURS MINUTES 1 00 02 HOLD Run Unlimited duration of centrifugation The time elapsed so far is displayed while centrifugation is in progress Initial value 00 00 00 TIME Time HOLD ACE w dt Accumulated Centrifugal Effect Input as x y 10 X Whole digits Y Decimal numbers Z Power TIME 2 50 08 If you change the time setting while centrifugation is in progress the newly defined countdown is restarted Any time that has already elapsed is disregarded and the acceleration phase is skipped 00 00 23 ING 1 000 17 Centrifugation in prog ress The centrifugation period including acceleration Is 1 minute Speed and RCF Value Hold swat 00 02 Time change while centrifugation in progress The centrifugation time is changed to 2 minutes _ 00 02 00 1 000 17 E Countdown restarting The new centrifugation time is now 2 minutes You can set the Speed and the RCF value using the Speed button in the main screen When you click this button a window opens in which you can enter the Speed as rpm or the RCF Value 3 4 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Thermo Scientific rpm Revolutions per minute SPEED MES 2 500 m Maximum sel value 12 000 rpm 3 Entering P
15. asons this PIN should not be disclosed to persons besides the owner An administrator is authorized for user account management see User Management on page 4 9 The device master PIN is required to make changes in the Settings Device and Contact menus All of the settings in the open menu are enabled for editing after the master PIN has been entered If a menu is closed completely and then re opened the master PIN must be entered again to make changes to any of the settings in that menu General Settings Thermo Scientific The Settings submenu of the Configuration menu consists of two pages On the first page you define settings for device policy and general settings such as the energy saving option Signal settings and languages can be defined on the second page The master PIN must be entered before any changes can be made in this menu see User Roles on page 4 15 Signal settings do not require input of the device master PIN Overview of possible settings in this menu e Device Policy e Define whether operation of the centrifuge requires a PIN to be entered Closed environment or not Open environment In an open environment you can also define whether a user is to be assigned to each centrifugation run e Language e Language e Energy Saving Options e Define various energy saving modes Possible options Green Mode Balanced Mode max Availability e Centrifuge Vacuum e Define whether a vacuum is to be emplo
16. cientific 4 Configuration Runs Configuration For programs and rotors the following applies If the rotor is installed after the program has been chosen than the speed of the centrifuge will be limited to the maximum speed of the rotor and the program name will not be shown any longer If the rotor is set in a program and another rotor is installed than a dialogue is displayed showing the inserted rotor and the expected rotor The program name will not be shown any longer If a program is loaded after a rotor has been installed and the set speed in the program is too high than the program will not be loaded and an error message is displayed If a program is loaded after a rotor has been installed than a dialogue is displayed showing the inserted rotor and the expected rotor The program name will not be shown any longer ROTOR PARAMETERS Inserted Rotor F12 6x500 LEX Material Carbon Fiber Capacity amp 500 ml k Factor 2302 Tube Angle 20 Speed at 4 C 12000 rpm Speed max 12000 rpm 24471 xg Expeched Rotor BIOFlex HC Material Aluminum i Capacity 421000 ml k Factor 5542 Tube Angle 90 Speed at 440 5500 rpm Speed max 5500 rpm 2068 q Thermo Scientific Touchscreen User Interface for Sorvall LYNX 4 5 4 Configuration Runs Configuration The table below lists possible cases for loading a program A program is either assigned to a specific type of rotor when the program has been defined with a rotor in
17. cientific Centri Vue Application Quick Guide 11 The application switches to the overview menu Overview and shows the formerly added centrifuge entry in an information block name information status Research 1 DOOR OPEN Basement Lab 12 Select the added centrifuge entry to switch to detail view 13 For remote control select Request Remote in detail view Research 1 Ready 0 rpm 18 C ACCEL TIME DECEL max 00 20 00 max SPEED TEMPERATURE 29 000 rpm 20 c NO PROGRAM T29 8X50 AAA A Overview Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 3 7 Thermo Scientific Centri Vue Application Quick Guide 14 Generate a new Remote ID Select Generate new ID in the settings screen N O Roto r REMOTE CONTROL Activate Remote Access AUTHENTIFICATION Generate new ID QR CODE NUMERICAL CODE 6F65GN3BUC User IDs Device Contact 15 In the request remote access screen of the Centri Vue application you can enter the ID by hand or by scanning the QR Code 16 Select Verify Request remote access Cancel ID 169U58l4NR Scan OR Code Verify 7 4 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo Scientific Centri Vue Application Quick Guide 17 If you have remote access to the centrifuge the start 0 rpm 18 C button is green ACCEL TIME DECEL max 00 20 00 max SPEED TEMPERATURE 29 F 000 r
18. creen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo Scientific Centri Vue Application Centri Vue App This opens the Add New Device menu Add New Device 1 Cancel closes the Add New Device menu without saving 2 Name shows the name of the centrifuge The name is detected automatically B cannot be IP Address edited If no name is retrievable Device name 1s 192 168 178 272 Example is shown 3 IP Address the IP address is entered automatically Information It can be edited subsequently 4 Information additional information can be entered later in order to identify the centrifuge da Add Device better 5 Add device selecting this button adds the centrifuge in the overview menu If saved succesful the application switches to the overview menu and shows the added centrifuge Add Manually If the IP address of the centrifuge is known after selecting the found device the centrifuge can be manually saved in the overview menu Select New in the overview menu This opens the Add New Device menu 1 Cancel closes the Add New Device menu Add New Device Cancel without saving 2 IP Address the IP address will be entered 2 IP Address Da 3 Information additional information can be Information entered in order to better identify the centrifuge later 4 Add device selecting this button adds the 4 f i Add Device centrifuge in t
19. ds to the selected profile see Acceleration Deceleration Profile on page 3 6 Possible statuses Ready Centrifugation can be started Door open The centrifuge door is open Error An error has occurred Stopped Centrifugation has been halted manually Completed Centrifugation or precooling has been successfully completed No rotor No rotor has been placed in the centrifuge 1 2 Touchscreen User Interface for Sorvall LYNX Thermo Scientific 1 Screen Views Main Screen Progress Status The rotor has been installed and the centrifuge door is closed 00 01 14 precooling in progress The remaining time for cooling is Pa displayed See Precooling on page 3 2 4 050 6 The rotor is in place and the centrifuge door is closed press the Ready button E to start centrifugation rl O on 20 Centrifugation begins with the acceleration phase The displayed 00 17 30 curve corresponds to the selected profile The time remaining for the complete centrifugation without deceleration phase is Bj ff displayed see Acceleration Deceleration Profile on page 3 6 4 852 rpm 9 c Centrifugation is taking place atthe set speed The time remaining 00 09 for centrifugation is displayed 10 000 10 Centrifugation has been successfully completed The centrifuge is 00 00 00 in the deceleration phase The displayed curve corresponds to the selected profile The centrifuge door cannot be opened
20. e other settings such as for the centrifugation programs in this section for detailed information refer to the Section Programs on page 4 3 If you are not sure of the general use of a button you can use the Tooltip mode which provides information about all of the operator control elements see Tooltip Mode on page 5 1 Start centrifugation Centrifugation can only be started when a rotor starts this button switches has been installed and to Stop IO identified when all a When the rotor comes to a parameters are gt Open Door complete stop this button acceptable and when the switches back to Open centrifuge door is closed Open Door Stop When centrifugation Door USER NAME Configuration Operator Tooltip Mode The configuration menu is In certain modes This button is used to activate the displayed when this button is touching the Tooltip mode All functional clicked This button is not active operator name elements are de activated with while centrifugation is in progress opens a window for this mode When an element is selecting other selected in the Tooltip mode a field operators is displayed with information about that particular element If you press the Tooltip button again this mode Is canceled 1 4 Touchscreen User Interface for Sorvall LYNX Thermo Scientific 1 Screen Views Lighthouse Mode Lighthouse Mode Thermo Scientific The Lighthouse mode i
21. e rotor data and the number Speed max 12 000 rpm of centrifugation runs performed with this rotor here for each rotor type and rotor name Number of Cycles 21 F10 4x1000 LEX ever used in the centrifuge The type of rotor i Carbon Fib i ee esas a being used is indicated by the open round Speed max 10 500 rpm arrow Number of Cycles 0 Programs Rotor log Configuration Settings You edit the settings for user management general settings device and contact in the setting section of the configuration menu 4 8 Touchscreen User Interface for Sorvall LYNX Thermo Scientific 4 Configuration Settings User Management Thermo Scientific A centrifugation run can be assigned to a user That user is then saved in the centrifuge log enabling subsequent evaluation of which user s conducted which centrifugation run s You can define in the menu Settings whether a log in is required to start centrifugation see Settings on page 4 8 and Access Control on page 4 10 Go to the submenu User IDs in the configuration menu to determine which users are listed Here you can create new user accounts or edit existing ones Any user that is logged in can edit his her own user information The user may also change his her user name and PIN 4 digit for the corresponding configuration Only users with Admin rights may create new users or manage other users A user needs not be logged in the open environmen
22. ers Acceleration Deceleration Profile If a profile is changed while centrifugation is in progress the new profile is applied immediately acceleration or deceleration of centrifugation is modified accordingly Touchscreen User Interface for Sorvall LYNX 3 7 eo Configuration Contents A Help on page 4 2 e Runs Configuration on page 4 3 e Settings on page 4 8 You can open the configuration menu using the button at the bottom left hand corner of the main screen This menu is subdivided into the sections Help Runs and Configuration Touching one of these buttons will open a second level with the corresponding submenus All of the sections of the configuration menu can be viewed by any user However to make or change any settings a PIN must be entered for authentication Details about distribution and assignment of user rights are given in User Management on page 4 9 The Configuration submenu is closed automatically if there is no interaction with the user for 30 seconds The Manual and Training videos submenus are closed automatically if there is no interaction with the user for 60 seconds The submenu will not be closed automatically when a training video is being played Any input or changes which have not been saved up to then will be lost Therefore always be sure to save any changes at once b Open Door Opening the configuration menu The buttons for USER NAME centrifugation are covered while the con
23. figuration menu is open The contents of the selected menu are displayed instead User IDs Settings Device Contact Configuration Thermo Scientific Touchscreen User Interface for Sorvall LYNX 4 1 4 Configuration Help Help 4 2 Help and assistance for a number of areas is provided in this menu The first two submenus Training Videos and Manual contain helpful videos and informative texts Values for two different rotors can be converted in the Calculators submenu First select a rotor on the left a Thermo Scientific rotor or from other manufacturers and then select a Thermo Scientific rotor for Thermo Scientific Sorvall LYNX centrifuges All of the parameters input for the rotor on the left will then be converted automatically for the rotor on the right Besides selecting the rotors you can also enter speeds and times for conversion for the particular centrifugation Ready Select Rotor 1 Sorvall F8 6X1000Y Material Carbon Fiber Capacity 6 x 1000 ml Tube Angle zas Speed max 8 500 rpm SPEED rpm 8 500 rpm 8 500 rpm TIME SPEED rcf 15 900 xg 15 824 xg k FACTOR 4 496 Select Rotor 2 Thermo Scientific A27 8X50 Material Aluminum Capacity 8 x 50ml Tube Angle 34 Speed max 27 000 rpm SPEED rpm 27 000 rpm SPEED rcf 87 207 xg 87 207 xg k FACTOR 408 Training Videos Manual Calculators Runs Configuration Touchscreen User Interface for Sorvall LYNX T
24. for acceleration and deceleration profile time and speed are set from the centrifuge If the values are changed the precooling program is exited Ready mame LYNX TEST No rotor 00 02 00 1 000 rpm 20 C ACCEL max DECEL max Precooling PRE COOLING marked line is stored as a fixed program Use the Load button to apply the program You can set the target temperature after this in the main screen Programs Run log Rotor log 00 01 27 1 800 12 DECEL ACCEL TIME max TEMPERATURE A SPEED 1 800 mn 3 2 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Time Thermo Scientific 3 Entering Parameters Time You can set the time for the centrifugation process using the Time button in the main screen When you click the Time button a window opens in which you can enter the required time The Time ACE w dt or HOLD modes can be selected The selected mode is clearly displayed in the main screen The selected mode is displayed on the Time button As shown in the diagram below the Time mode has been activated ACCEL TIME DECEL 2 00 02 00 1 SPEED TEMPERATURE 1 000 20 LU y USER PROGRAM F10 4X1000 LEX TIME rime vea sree HOURS MINUTES a 2 L3 ENESENN 7 L 8 9 00 o cE Touchscreen User Interface for Sorvall LYNX 3 3 3 Entering Parameters Speed and RCF Value Time Duration of centrifugation input
25. ge supplies target and actual values Request No data Response actualValues ace lt ace value in x xxExx gt powerDown lt true orfalse gt rcf lt rcf value in xg gt rpm lt rotation speed in rpm state lt state identifier gt temperature lt temperature in C time lt time format hh mm ss gt i error lt error object gt name lt centrifuge name gt program lt program name gt rotorName lt rotor name gt Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 8 1 8 REST Webserver 8 2 setValues 4 accelerationProfile lt profile number gt ace lt ace value in x xxExXx gt decelerationProfile lt profile number gt rcf lt rcf value in xg gt rpm lt rotation speed in rpm gt temperature lt temperature in C gt time lt time in hh mm ss gt user lt user name gt lt error object gt code lt error code gt description lt error description in gui language gt title lt error title type of error gt time lt error occurrence time in year month day hh mm ss gt If a value is not available the value is set to zero This can be used to distinguish between the following modes of operation e LYNX in RPM Mode rpm set rcf contains the value zero e LYNX in RCF Mode rpm contains the value zero rcf set The distinction of operation mode time hold and ACE is
26. he detail view for a centrifuge 3 Select Overview to switch to the overview menu 4 Select Discovery to switch to the discovery menu To edit or delete a locally saved centrifuge select a list element and keep it touched Detailed information to this screen is listed below at 4 5 Overview Centrifuge States The centrifuge can show following states e DOOR OPEN if the door is open e READY if the centrifuge can be started e ACCELERATING if the centrifuge accelerates e RUNNING if the centrifuge is running e STOPPING if the centrifuge decelerates e COMPLETE when a run has been completed successful e STOPPED If a run get canceled e POWER DOWN centrifuge is switched off Note This message is only available when the centrifuge is connected to a network with PoE Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 7 7 Thermo Scientific Centri Vue Application Centri Vue App e DISCONNECTED Network timeout If a centrifuge error occurred a red cross is displayed Discovery Menu 1 New discovered devices To start a new discovery gt E Device Discovery open the item list 2 List view all centrifuges found in the network will New discovered devices 4 be displayed For each centrifuge following information is shown Research 3 a Name of the centrifuge b IP address of the centrifuge 3 Select Overview to switch to the overview menu 4 Select
27. he overview menu If saved succesful the application switches to the overview menu and shows the added centrifuge Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 11 7 Thermo Scientific Centri Vue Application Centri Vue App Error Adding Centrifuges If a centrifuge cannot be added to the overview menu there are two possible reasons 1 A centrifuge with the same IP address is already stored in the local list A pop up message indicates the error See figure 19 IP Address already exists Please choose another IP address OK 2 The IP address has no valid format The entered IP address must have the format x x x x In addition only numbers may be included If an IP address is entered in an invalid format the outline IP Address turns red and the add device button will be 192 178 168 25 disabled See figure 20 Add New Device Information Add Device 7 12 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo Scientific Centri Vue Application Centri Vue App Edit Centrifuge Entry The centrifuge entries in the overview menu can be edited or deleted To edit a centrifuge entry select the entry and hold it about 3 seconds The edit menu will open with the following options Overview Research 3 Basement Lab STOPPING 1 Select Edit to open the edit menu of the centrifuge 2 Select Delete to delete a centrifuge entr
28. hermo Scientific 4 Configuration Runs Configuration Runs Configuration Programs Thermo Scientific You can define settings for programs and for log processes in the Runs section of the configuration menu You can save set values for a centrifugation process in a centrifugation program Avoid multiple identical input for similar centrifugation runs by defining the values in advance in individual programs Programs may only be administered when the rotor is at standstill PRE COOLING is stored as a default program and cannot be deleted You can only change the target temperature in this program all other values are calculated automatically when PRE COOLING is loaded For more details refer to Precooling on page 3 2 The list of available programs is given in the first submenu Programs in the Runs menu If the program list should exceed the length of the screen you can use the pager to easily navigate through the individual pages of the list A maximum of 120 programs can be contained in the list The list can also be called up in the main screen by touching the button with the name of the program Ifa program has already been selected a blue arrow is displayed in front of the program name Before that program can be activated you must first select a program by touching the display showing the program name and the parameters Touch the Load button to activate the particular program Now the blue arrow is displayed in front of the
29. ifuge and or smartphone are not on the same network OR Code is not Recognized by the Scanner e To scan the QR code quickly and correctly the phone must be held vertically over the QR code e The entire QR code should be in the bright area of the scanner e Ifyou have problems to focus the QR code check if the camera of the phone works and if the camera lens needs to be cleaned Closing the Remote Session The remote session cannot be terminated manually However the remote session terminates automatically at the following events e A new remote ID is generated at the centrifuge e Centrifuge without User login If the centrifuge door has been opened 5 minutes without interaction in the application on the smartphone or the centrifuge e Centrifuge with User login If the current user is logged off manually or automatically depending on the selection of the combo box in the configuration menu The session management for the smartphone is directly attached to the user management of the LYNX centrifuge This means that the smartphone session becomes invalid in case any user logs in or out at the LYNX centrifuge Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 15 7 Thermo Scientific Centri Vue Application Centri Vue App Connection to a Network with active PoE A 1 kaos The user interface of the LYNX centrifuge is a PoE enabled device If it is connected to a PoE network port it ma
30. ins no information about the user who started the centrifugation run e Optional user selection e In this mode a user may be selected when starting the centrifuge The centrifugation run is assigned to the selected user and the user name is listed in the centrifugation log Specifying a user is optional If no user is selected this log is equivalent to the first option above for the open environment No user e Mandatory user selection e In this mode a user must be selected in order to start the centrifuge As a result each centrifugation run is allocated uniquely to a specified user 4 12 Touchscreen User Interface for Sorvall LYNX Thermo Scientific 4 Configuration Settings Mandatory user selection RST Gu er Allocation of a user to a centrifugation run A ERARIO is obligatory for mandatory user selection User 1 6 24 2010 Laboratory Group B 6 24 2010 User 2 6 24 2010 User 3 6 24 2010 User 4 6 24 2010 Usergroup 6 24 2010 Usergroup 1 6 24 2010 Usergroup 2 6 24 2010 Usergroup 3 6 24 2010 v Usergroup 4 6 24 2010 Closed Environment Thermo Scientific In a closed environment users must log in using their PIN before they can operate the centrifuge The login window is opened by touching the user name at the bottom of the screen LOG IN USER ID USER PIN 1 J 2 3 eee When you press the arrow in the USER ID 9 field a list of all available users is opened In the ce
31. mapped as follows e Time Mode time set ace Value is zero e ACE Mode time is zero ace is set e Hold Mode As in Time Mode but the value for time at setValues is also zero Examples Centrifuge in Time RPM Mode and an error occurred actualValues ace null powerDown false rer null rpm 0 state EREADY temperature 0 time 00 02 00 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific Thermo Scientific i error code 36575 description Error Text title Centrigue Error time 2015 03 23 03132137 PH 3 name My Centrifuge program o TH rotorName F10 4x1000 LEX setValues 4 accelerationProfile 9 ace null decelerationProfile 9 Fer mul Ls rom 500 temperature 0 tame 00 02 00 y user Centrifuge User Centrifuge in Hold RCF Mode actualValues ace null powerDown false oor 0 rom i null state STOPPED temperature 0 time 00 00 38 y error null hame My Centrifuge program i T rotorName F10 4x1000 LEX setValues accelerationProfile 9 ace null decelerationProfile 9 8 REST Webserver Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 8 3 8 REST Webserver ref 1000 rpm null temperature 0 time null y user Centrifuge in ACE RPM Mode actualValues
32. n for this however The log settings can be loaded using the Load button The parameters used are then applied in the main screen The list of the last 120 entries is available for viewing under Run log If required this list can be exported to a USB stick for systematic evaluation of the centrifugation runs using the button Export list The following values are logged for each centrifugation run e Starting time e Program used if one was used e Run time e Speed e Temperature e Profiles for acceleration and deceleration e Rotor e Termination of centrifugation completed successfully halted manually error occurred e User name if a user was logged in Touchscreen User Interface for Sorvall LYNX 4 7 4 Configuration Settings Rotor log All of the types of rotor known by the centrifuge are listed in the Rotor log submenu The maximum values and the number of centrifugation runs that have already been performed with this type of rotor are displayed for each rotor type The following values are saved in a Rotor log e Rotor type and rotor name e Rotor material Capacity maximum number and volume of specimen containers e Maximum speed Number of centrifugation runs performed up to this time in this centrifuge with this type of rotor The centrifugation runs performed with each rotor type are saved Ready F12 6x500 LEX l Rotor log Material Carbon Fiber Capacity 6 x 500 ml You can view th
33. n of centrifugation Thermo Scientific Touchscreen User Interface for Sorvall LYNX 2 1 AR Entering Parameters Contents e Temperature on page 3 1 e Precooling on page 3 2 e Time on page 3 3 e Speed and RCF Value on page 3 4 e Acceleration Deceleration Profile on page 3 6 You can change any parameter prior to and during a centrifugation process Temperature You can set the temperature using the Temperature button in the main screen When you press this button a window opens in which you can enter the temperature in C Either you p p y p positive or negative values can be entered here The temperature limits upper lower for the centrifuge may not be exceeded however An error message is issued to indicate any unacceptable values TEMPERATURE 20 When centrifugation is in progress the temperature will be adjusted immediately to the newly defined temperature by means of cooling This does not affect the duration of centrifugation Thermo Scientific Touchscreen User Interface for Sorvall LYNX 3 1 3 Entering Parameters Precooling Precooling Precooling is used to bring a rotor to a desired temperature prior to centrifugation of a sample Precooling is defined as a set program and is called up via the configuration menu Programs on page 4 3 This program is activated using the button Load In the main screen you can then set the required target temperature The settings
34. nlocked Instead of using the user PIN it is possible to enter the device master PIN instead Ready Parameters cannot be changed until a IDLE user logs in O 20 Read only mode ACCEL TIME DECEL max 00 02 00 max Lock after start The user interface is locked after a SPEED TEMPERATURE centrifugation has been started 1 000 son 20 e oe Ho In the closed environment you can also specify the setting for whether and when the registered user is logged out automatically These settings can be defined under Settings in the configuration menu using the device master PIN see User Roles on page 4 15 The factory default for the device master PIN is 12345 The following settings are required for automatic logout e Directly after a completed centrifugation run as soon as the cover has been opened e 5 minutes without any interaction after a completed centrifugation run e The user is not logged out automatically and remains logged in until he she logs off or until a different user logs in 4 14 Touchscreen User Interface for Sorvall LYNX Thermo Scientific User Roles 4 Configuration Settings A distinction is made between the following user roles Administrator and User The centrifuge includes the device master PIN which can be used to define all settings for the centrifuge The device master PIN is handed over to the owner of delivery of the unit and may be changed Therefore for safety and security re
35. nterface To activate the remote control access select Settings Step 1 3 and select the checkbox aurmenmiricarion Remote Control Step 4 QR CODE NUMERICAL CODE User IDs Settings Device Contact Configuration Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 1 7 Thermo Scientific Centri Vue Application Quick Guide 3 Download and install the Centri Vue application on your smartphone 4 Smartphone and centrifuge have to be in the same LAN with the same IP range 5 Start the Centri Vue application sq 6 Select Discovery Screen Q Discovery 7 Open the New Discovered Devices list in the discovery menu to search for new centrifuges in the network Step 1 and 2 Device Discovery Device Discovery Device Discovery New discovered devices New discovered devices Research 3 Pull to refresh New discovered devices Q 11 Lo 11 ip 8 Centrifuges with installed Connectivity Plug In will be identified automatically in the same LAN Step 3 9 Select the identified centrifuge 10 Select Add Device to add the centrifuge to the device list in the overview menu Add New Device Cancel Optional You can add some additional IP Address information about the centrifuge Select the 192 178 168 25 Information field Information Add Device 1 2 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo S
36. ntrifugation log the user logged in for the specific centrifugation run is documented If no user is logged in the centrifuge cannot be started in a closed environment Any user can halt an ongoing centrifugation run or open the Tooltip mode and the configuration menu without having to be logged in All displays and screens are however available as read only Only users that have logged in may define change settings Touchscreen User Interface for Sorvall LYNX 4 13 4 Configuration Settings A user can log in out at any time when the rotor is not spinning simply by touching the user name field at the bottom of the main screen As an alternative the user can also log out under User ID in the configuration menu using the Log out button In a closed environment it is possible to activate an automatically locking of the user interface so that no other user is able to change the settings while a run is active This option is called Lock after start and has to be enabled in the configuration menu under Acces control It can only be enabled if the option Protected by user PIN is enabled When the user interface is in the Lock after start mode no changes to the centrifuge are possible except stopping it by pressing the Stop button To unlock the user interface touching the button for the set points or the one for the user name will bring up the log in dialog When the correct user PIN has been entered the user interface will be u
37. ofile by direct point and select Touch the Apply button to define the profile for the next centrifugation process or cancel your selection using Cancel The profile with the smallest number has the most gradual slope and is indicated by min the profile with the number 9 has the steepest slope and is denoted by max In the main screen the current profile is indicated by the number on the Profile button and the slope curve in the Status section see Status on page 1 2 Presentation of the profiles must be taken symbolically The exact slope is complex and is a function of the rotor being used and the set speed The curve that is displayed represents an abstract simplification for distinguishing between the different profiles gradual slope steep slope ACCELERATION PROFILE DECELERATION PROFILE Acceleration Brake Select the profile by clicking the slide on the screen You can also screen on both ends using your finger I EY B move the slide on the 3 6 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Thermo Scientific Ready O on 20 min 00 02 00 min Ready O 2 00 02 00 20 DECEL max Ready dm kh Oren 20 3 00 02 00 3 Profile display in status sec tion The profiles selected for acceleration and deceleration are indicated by profile numbers on the buttons and by the curve in the status section 3 Entering Paramet
38. pm 2 0 C NO PROGRAM T29 8X50 z Overview iscove Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 5 7 Thermo Scientific Centri Vue Application Connectivity Plug In Touchscreen User Interface Connectivity Plug In Touchscreen User Interface Remote Settings For remote control ability start stop by smartphone application you need to change the settings as y p Dy p pp y g g described below If you just want to check the status of the centrifuge on your smartphone read onl you J g y p y access you do not need to select the checkbox Activate Remote Access in the touchscreen user interface The settings for the remote control feature are located on the last page of the touchscreen user interface settings menu Following options are available 1 Remote Control Checkbox to general allow deny a remote control Master passcode required ae Su li Mess 2 Authentication Selecting Generate new ID generates a new code ID for a remote session Depending on the setup a user authentication may be required 3 QR Code The generated ID is shown in a QR Code 4 Numerical Code The generated ID is displayed in text form I NUMERICAL CODE n 6F65GN3BUC Access Control Settings The centrifuge has 2 different user modes which influence the behavior of the remote control feature 1 Closed centrifuge If the option Access Control is selected it is not
39. possible to start the centrifuge without login A Remote ID can just be generated if a user is logged in 2 Open centrifuge The option Access Control is not checked It is possible to start the centrifuge without login There are 3 different options for user ID logging a No user logging b Optional user logging c Mandatory user logging Depending on the selection a user login is mandatory to use the remote control feature as the user is able to start the centrifuge remotely and the centrifuge needs to know the user for logging purposes 7 6 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo Scientific Centri Vue Application Centri Vue App Centri Vue App The Centri Vue application can discover available centrifuges in the LAN and create a list of local centrifuges with optional additional user information This information can be used to add location information to the local centrifuge entry Overview Menu 1 Select New to manually add new centrifuges New Overview 2 List view shows the added centrifuges in a list showing following information Research 1 Basement Lab a Name of the centrifuge if no name is entered for the centrifuge the IP Research 2 address of the centrifuge is shown instead Basement Lab b Optional information e g for location Research 3 information Basement Lab c Centrifuge status Select a centrifuge entry to open t
40. s activated while centrifugation is in progress if you do not touch the screen for at least 30 seconds The buttons for parameters are concealed when the Lighthouse mode is active The status for centrifugation is displayed larger instead so that you can read the remaining time for centrifugation from a good distance Thermo ECIENTIFIC a Lighthouse Mode 0 0 n 0 0 5 5 Centrifugation can be halted at any time in the Lighthouse mode by touching the Stop button You can exit this mode a _ by simply touching the 1 000 y 20 screen USERNAME The Stop button is the only button active in the Lighthouse mode in order that you can manually halt centrifugation at any time All other buttons are de activated The diagram below illustrates graphically the statuses for the main screen such as the switching from the Standard to the Lighthouse mode The Standard mode is shown at the bottom both with the centrifuge in motion and at standstill the Lighthouse mode is shown at the top of the diagram If the centrifuge is stopped manually it switches immediately from the Lighthouse mode back to the Standard mode Normally you must touch the screen to switch back to the Standard mode Touchscreen User Interface for Sorvall LYNX 1 5 1 Screen Views Lighthouse Mode Lighthouse Mode Lighthouse Mode centrifuge halted centrifuge running Centrifugation completed Error occurred N 0n S 5 7 S 3 O S A gl
41. s successfully scanned the application returns to The remote ID is decoded from the QR code and shown in the ID Select Verify If successfully verified the application returns to the Select Cancel to return to the overview menu Request remote access Cancel entered manually ID 169U58l4NR A new session for this smartphone will be created in case of a successful verification Request remote acces ID e Press in the ID field of the ae PA E cc Request Remote Access AUTHENTIFICATION Scan OR Code Verify menu The keyboard appears the input field on the smartphone Select Verify MUMERICAL CODE FAS GNABLIC sl Generate a new remote ID in the centrifuge s 3 2 2 Select Scan QR Code in the Request Remote Access menu figure 25 the Request Remote Access menu field The ID can be checked again by comparing it to the numerical code on the centrifuge Detail view menu 7 14 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific 7 Thermo Scientific Centri Vue Application Centri Vue App Error Establishing the Remote Connection If a message Request Error pops up when selecting Verify no remote connection is established to the centrifuge Possible error sources e The remote ID was not transferred correctly from the centrifuge e Another user has already built up a remote session with this remote ID e The centr
42. selected program indicating that it is activated Ready PRE COOLING Fh LYNX TEST No rotor L 00 03 00 1 000 rpm 20 C ACCEL max DECEL max Import List Export List A AA PP AE Programas Run log Rotor log Configuration Yo li Use the button New to create a new program When you touch this button the standard program editor is displayed that you can use to set all relative parameters for centrifugation Touchscreen User Interface for Sorvall LYNX 4 3 4 Configuration Runs Configuration PROGRAM EDITOR IDLE TIME 00 02 00 SPEED TEMPERATURE 1 000 rpm 20 E Ea Ee a Y Apply Prog rams Aun lo g Rotor l g The rotor that is currently in use in the centrifuge will be assigned automatically to the program Note here that the parameters that are entered may not exceed the limits for the rotor A program can still be defined even if no rotor is currently in use The parameters will be checked as soon as the program is loaded To save the changes made in the program editor use the button Apply When you click the Load button the parameters for the program are set as setpoints in the main screen The parameters are then subjected to a plausibility check for the type of rotor being used A warning is issued if any values are not acceptable These values can then be adjusted automatically to match the type of rotor being used 4 4 Touchscreen User Interface for Sorvall LYNX Thermo S
43. t see Open Environment on page 4 12 New 88 p p pag user accounts can be created by any users in this environment These new users are not assigned a PIN however These user accounts can as a result not be used in the closed environment for which a password must be entered PINs must be assigned to user accounts at a later time for working in a closed environment Ready Admin 05 31 2013 User Management The logged in user is indicated by an open round arrow All users holding a valid PIN are indicated by 4 dots to the right of his her name User IDs Settings Device Contact Configuration Touchscreen User Interface for Sorvall LYNX 4 9 4 Configuration Settings Import Export Editing a user Selection of the corresponding USER 1D user role determines whether a user possesses Admin rights EDIT USER ACCOUNT USER PIN REPEAT USER PIN USER ROLE Administrator The factory default Admin PIN is 0000 First create a new administrator profile with a new PIN Then delete the factory default profile Users can be imported via USB stick To do this connect an USB flash drive to the centrifuge and select the option Import List All users with new names will be added to the internal list If users with the same name do exist on the internal list of the centrifuge they will be replaced during the import process The USB stick that you use must have FAT32 formatting in order to be re
44. ter start The closed environment is activated by checking this box This box must remain LOG OUT unchecked blank for an open environment USER ID LOGGING MASTER PASSCODE Change Master Passcode ENERGY OPTIONS Max Availability gt VACUUM OPTIONS High Speed Only gt User IDs Settings Device Contact Configuration Touchscreen User Interface for Sorvall LYNX 4 11 4 Configuration Settings Open Environment A PIN is not required to operate the centrifuge in an open environment Any user can enter parameters and perform centrifugation runs Various options are available for the open environment scenario to establish which centrifugation runs were started by which users A distinction is drawn in these options as to whether the user is also to be documented in a centrifugation log l 11 ACCESS CONTROL E Protected by User PIN pi N Open environment The open environment scenario is Lock after start activated User may change You can now define whether the user is to be included in every centrifugation log here logging of user is activated LOG OUT USER ID LOGGING Mandatory User Logging MASTER PASSCODE No User Logging Optional User Logging _ a ENERGY OPTIONS Mandatory User Logging gt VACUUM OPTIONS High Speed Only User IDs Settings Device Contact Configuration e No user specified e No user is selected prior to starting the centrifugation run The centrifugation log conta
45. the same name do exist on the centrifuge they will be replaced during the import process The USB stick that you use must have FAT32 formatting in order to be readable by the centrifuge Export Programs can be exported via USB stick for further processing To do this connect an USB flash drive to the centrifuge and select the option Export List Existing programs will then be exported to the USB stick as CSV files Programs in this CSV format can be edited and imported using any spreadsheet program such as Excel The USB stick that you use must have FAT32 formatting in order to be readable by the centrifuge 4 6 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Run log Thermo Scientific 4 Configuration Runs Configuration Every centrifugation run is documented as a log A new entry in the log list is registered as soon as a centrifugation run is started Ready Admin 06 14 2013 08 32 14 AM Stopped manually 00 02 00 ACCEL max DECEL max F12 6x500 LEX 1 000rpm 20 C Admin 06 14 2013 08 31 39 AM Stopped manually 00 02 00 ACCEL max DECEL max F12 6x500 LEX 1 000 rpm 20 C Programs Run log Rotor log Configuration Run log Every centrifugation process is logged In addition to the defined parameters such as Time Speed and Temperature an operator can also be registered for each configuration process as an option An operator must be logged on prior to the start of centrifugatio
46. use or not program defined without a rotor in use Normally centrifugation can only be started when its assigned parameters do not exceed the limits defined for the centrifuge or the type of rotor being used A warning will be issued if this is not the case No program has been loaded Centrifugation can be started MO ME E The program has been loaded No rotor is currently being used In the rotor information field the notice No rotor is displayed MT No program has been loaded or a program has been loaded and the parameters were subsequently changed in the main screen so F10 4X1000 LEX i norrocram KEN that at least one parameter no longer corresponded to the value in the loaded program A rotor Is in use and all parameters are acceptable Centrifugation can be started In the program information field the notice No program is displayed The program has been loaded Centrifugation can be started Once the values for the program have been applied in the main screen you can adjust the parameters for the next centrifugation run The values stored for the program will not be affected by subsequent changes in the main screen Program parameters can only be edited using the program editor in the configuration menu Import Programs can be imported via USB stick To do this connect an USB flash drive to the centrifuge and select the option Import List All programs with new names will be added to the internal list If programs with
47. y Deletion is done if the subsequent question is confirmed with Yes 3 Select Cancel to switch back to the overview menu Name Research 3 IP 192 168 178 24 Edit Delete Cancel If Edit is selected an Edit Device menu opens Edit Device Cancel 1 Select Cancel to return to the overview menu 2 Name the name of the centrifuge can only be changed directly at the LYNX centrifuge and is therefore not editable IP Address 3 IP Address the stored IP address can be changed 192 168 178 24 here Information 4 Information the information field can be edited Basement Lab here e ann Soe na Remote Control Prerequisites The centrifuge was enabled for remote access For detailed information see chapter 3 Thermo Scientific Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges 7 13 7 Thermo Scientific Centri Vue Application Centri Vue App Request Remote Access Menu The Request Remote Access menu opens when Request Remote Access is selected in the detail view l Input by Hand Input by OR Code l ID the Remote ID from the centrifuge can be Scan QR Code opens the integrated QR Scanner Verify the entered or scanned code will be verified Generate a new ID on the 0 centrifuge No Rotor Enter the numerical ID into The integrated QR scanner opens Scan the QR code on the centrifuge with the QR scanner See If the QR code i
48. y remain active after the main power switch of the centrifuge has been turned off While it is in this state the following screens will be displayed at the centrifuge and in the Centri Vue application My Centrifuge S C Ve ING el a SOF P ey Me ed a Centrifuge main power is turned off A y Centrifuge main power turned off To start centrifuge turn on main power switch After turning on the main power switch the centrifuge will return to the normal state of operation To avoid keeping the user interface in the active state disable PoE on the used network port Power over Ethernet or PoE describes a system which passes electrical power along with the Ethernet cabling 2 F e Disabling PoE is recommended to prevent premature wear of the user interface display components 7 16 Touchscreen User Interface for Sorvall LYNX Superspeed Centrifuges Thermo Scientific lg REST Webserver Port of RES T Webservers 800 TCP The data is exchanged via defined JSON objects Resource Overview In the table below the REST interface provided methods are listed URL Method allowed Description lt device ip gt lt port gt getall GET Information and data about the current state of the centrifuge lt device ip gt lt port gt getstate GET Brief information only state and name of the centrifuge Detailed description of the resources GET lt device ip gt lt port gt getall Query the current state of the centrifu
49. yed Possible options High Speed only Smart Energy Optimized e Brightness e Brightness Control e Sound Signal e You can define different acoustic signals in order to distinguish between various centrifuges e Volume Control e Use this regulator to adjust the volume of the signal Touchscreen User Interface for Sorvall LYNX 4 15 4 Configuration Settings Ready LANGUAGE English DATE 05 13 2013 Settings Menu In the Settings menu you can modify TIME 03 14 00 PM the volume of the signal for example BRIGHTNESS CONTROL J END OF RUN ALERT Off gt VOLUME CONTROL User IDs Settings Device Contact Configuration 4 16 Touchscreen User Interface for Sorvall LYNX Thermo Scientific Device Settings Thermo Scientific page 4 9 network DEVICE TYPE SERIAL NUMBER SOFTWARE VERSION HMI MAIN CONTROLLER PARAMETERS DEVICE NAME DEVICE LOCATION IP ADDRESS SUBNET MASK STANDARD GATEWAY DNS SERVER User IDs 4 Configuration Settings In the Device submenu of the configuration menu you can specify special data about the location of the device The master PIN is required to make any changes in this menu see User Management on If the centrifuge is incorporated into a network you can specify the address of the centrifuge here These settings are important when rotor types are to be imported or exported from a server via the A device master PIN must be entered to make any changes