Home
User`s Guide
Contents
1. O Equipment Alert Always use nylon or plastic bristle scrubbers The use of steel wool or steel brushes may cause steel particles to become embedded in the surfaces and will cause rust Avoid prolonged surface exposure to chlorides bromides iodides and thiocyanates Salty solutions allowed to evaporate on stainless steel surfaces may cause pitting corrosion or metal discoloration Maintenance Troubleshooting If the table will not go up or down check the following e Unit is plugged into a wall outlet Power to the outlet No visible damage to wiring s The table is at its maximum or minimum height Hand control is fully inserted and making proper connections If the table won t drain check the following The filter is clogged If the pump won t start check the following The filter is clogged Unit is plugged into a wall outlet d Power to the outlet e The water sensor is out of adjustment contact factory If the pump won t stop running e The water sensor is out of adjustment contact factory If the scale does not operate check the following e Refer to scale section elsewhere in this manual Preventive Maintenance Periodically inspect the following areas Power cord should be free of cuts or other visible damage e Allfasteners should be present and tightened securely e All mechanical functions should operate properly Have an authorized service technician inspect your tabl
2. 3 products considered to be of a consumable nature 4 accessories or parts not manufactured by Midmark 5 damage to mat caused by chewing 6 charges by anyone for adjustments repairs replacement parts installation or other work performed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark EXCLUSIVE REMEDY Midmark s only obligation under this warranty is the repair or replacement of defective parts Midmark shall not be liable for any direct special indirect incidental exemplary or consequential damages or delay including but not limited to damages for loss of profits or loss of use NO AUTHORIZATION No person or firm is authorized to create for Midmark any other obligation or liability in connection with the products THIS WARRANTY IS MIDMARK S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS Registration To register your product warranty go to www Midmark com Midmark Corporation 60 Vista Drive Versailles OH 45380 0286 1 800 643 6275 1 937 526 3662 MINMARI lt www midmark com Because we care 12
3. The scale is under loaded Ensure that tabletop cover pins are properly seated on load cells If the problem cannot be resolved using the above corrective actions please contact Midmark at 1 800 Midmark 1 800 643 6275 or 937 526 3662 8 Maintenance Cleaning DebrisTrap A debris trap exists beneath the drain of the stainless steel table tub It should be visually inspected daily and removed and cleaned weekly if table is used only for examinations The debris trap must be used and should be removed and cleaned after each procedure Inspect debris trap If any debris is evident remove cap A N Clean debris trap m Remove yellow trap ey Xx clean and replace ANS y VA101500i Painted Surfaces Clean the table weekly wiping the painted metal and plastic surfaces with a clean soft cloth and mild cleaners Periodic applications of common furniture wax will ease cleaning and maintain the finished luster of the table s painted surfaces Stainless Steel Surfaces For best results stainless steel should be cleaned weekly or whenever films or deposits become apparent Use a mild cleaning agent and whenever possible rinse thoroughly and towel dry More stubborn deposits or tightly adhering debris may require vigorous rubbing or brushing Always scrub with the grain of the stainless steel If a commercial cleaner needs to be used make sure it is approved for use on stainless steel surfaces
4. lowest height 9 5 in 24 1 cm the clearence between the bottom of the table and the floor is approximately 4 5 in 11 4 cm Make sure patient and all objects are clear of this area before lowering table If table malfunctions immediately release control unplug the table power cord and assist the patient from the table If table continues to malfuction call for service Failure to comply with these instructions may result in injury to patient and or damage to equipment J Clearance Approx 4 5 in 11 4 cm VA1966i Operation Table Up Down MESS E DOWN To move the table top UP or DOWN Press and hold the appropriate button Note Movement automatically stops at the max min geed o v S VA101400i Water System To acquire water pressure Turn water valve knob to side and utilize spray nozzle D 7 Equipment Alert Always close shutoff valve when table is not in use t VA101200i To adjust water temperature Pull table away from wall loosen screw on cap turn mixing valve knob until desired temperature is achieved then tighten cap screw factory setting 70 F 21 C adjustable range 70 120 F 21 49 C VA101300i S Operation e Scale optional The 300 pound capacity platform scale utilizes four 4 built in load sensors
5. MIIDIMIAIRIK Canis Major Wet Dental Treatment Lift Table For models CMDT 45 CMDT WS 45 CMDT 60 CMDT WS 60 U S e S G U d e 003 1725 00 Rev G nd Product Information The information below is required when calling for service Dealer name phone Date of Purchase m Y Model Serial Number B Midmark Authorized Service Company name phone Model Serial Number Location PL KS VA101100i Table of Contents Important Information Safety de 4 Transportation Storage Conditions nnn 4 Intended USG c 4 Electrical PrecaltiONS T 4 e Elle 5 Operation Table pis EE 6 Water System irine M 6 Scale Opina m 7 Maintenance nep EE 9 TroubleshiooIO me PEE 10 Preventative Maintenance mico IE BRA RES ERRURRRRRRUREE 10 Calling ee 10 Specifications Weights Dimensions Electrical Specifications ccoonoocccnnnnnncocccnncccnnnanncnnccnnnnnn noo 11 Rangeof MOTOR EC 11 Warranty Information Ringel 12 STE e cR TL ERE 12 CO Important Information Safety Symbols a N DANGER Indicates an imminently hazardous situation which will result in serious or fatal injury if not avoided This symbol is used only in the most extreme conditions J N WARNING Indicates a hazardous situation which could result in serious injury if not avoided fe N CA
6. Range of Motion amp Dimensions 61 Se 154 9 cm 46 5 118 1 cm 60 t 152 4 cm ss WI 64 3 cm Hi 43 5 Max 110 5 cm 62 5 PN 158 8 cm end 77 195 6 cm 9 5 Min a j 24 1 cm cm Water Inlet Height jn 27 9 Drain Outlet Height SE VA1699i 11 Warranty Information Limited Warranty SCOPE OF WARRANTY Midmark Corporation Midmark warrants to the original purchaser its new Animal Care products and components except for components not warranted under Exclusions manufactured by Midmark to be free from defects in material and workmanship under normal use and service Midmark s obligation under this warranty is limited to the repair or replacement at Midmark s option of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which upon examination by Midmark prove to be defective APPLICABLE WARRANTY PERIOD The applicable warranty period measured from the date of delivery to the original user shall be one 1 year for all warranted products and components EXCLUSIONS This warranty does not cover and Midmark shall not be liable for the following 1 repairs and replacements because of misuse abuse negligence alteration accident freight damage or tampering 2 products which are not installed used and properly cleaned as required in the Midmark Installation and or User s Manual for this applicable product
7. These electronic sensors are accurate within 5 grams and are protected against impacts moisture and liquid spills The scale has been calibrated and all parameters are preset so it should never need calibration CENTER OF ZERO O Equipment Alerts Prevent flammables and liquids from entering the scale head e Allow adequate clearance on all sides of scale platform to ensure accuracy Do not drop large loads onto the scale platform Never exceed the rated capacity of the scale Do not operate in close proximity to RF transmitters There are two 2 controls located on the display UNITS Toggles the displayed weight between pounds Ibs and kilograms kg ZERO Resets the display to indicate zero 0 0 Note The scale will not respond to the ZERO control under these conditions Motion on Scale An under load condition less than 0 1lbs or 50g An over load condition more than 300lbs or 1kg e Operation Scale continued Weighing an animal without a cage Zero scale 0 0 place animal on center of platform and read the weight displayed Note A blank scale while weighing means excess motion is detected and weight averaging is in process Once completed the scale will display and HOLD the recorded weight Adding additional weight while in the HOLD condition will NOT change the display Weighing an animal with a cage Place empty cage on center of platform then zero sc
8. UTION Indicates a potentially hazardous situation which could result in minor or moderate injury if not avoided It may also be used to alert against unsafe practices J b Equipment Alert Note Indicates a potentially hazardous Amplifies a procedure practice or condition situation which could result in equipment damage if not avoided Dangerous voltage shock hazard A Consult User Guide for important information Transportation Storage Conditions Ambient Temperature Range AN 5 C to 38 C 41 F to 100 F Relative Humidity seeesesesessss 10 to 90 non condensing Atmospheric Pressure cocccoccccnoncccnonnnonancnnnnancnnnnos 500hPa to 1060hPa 0 49atm to 1 05atm Intended Use This product is intended to be used as a table to provide positioning support and cleaning of animals during examination or minor surgery procedures conducted by qualified professionals Electrical Precautions Use nonconductive material to insulate patient from metal portions of table Consult operating instructions for the device before use WARNING When using high frequency surgical devices or endocardial catheters Failure to comply may result in electric shock or burns to patient 4 Safety Instructions CAUTION Be sure that all personnel and equipment are clear of the table before activating any function Failure to do so could result in personal injury At the table s
9. ale 0 0 Remove cage from platform scale will read negative weight Place cage with animal on center of platform and read record weight To continue to weigh animals in same sized cage place next cage with animal on center of platform and read record weight do not zero scale Troubleshooting Symptom or Error No Display Incorrect weight Display segment s missing illegible ZERO key does not reset to 0 0 Display locked amp will not change Displays HiLoAd Displays HiLdX Displays LoLdX Possible Problem and or Remedy Power is OFF Table is unplugged from the power source Interference with scale platform Ensure there is nothing in contact with the sides of the scale platform Ensure the correct UNIT of measure is selected Low or intermittent voltage Unplug for one 1 minute to reset system then reconnect to original power source Disconnect power cord and use different power source Motion or activity on the scale Remove source of motion Ensure platform is level and supported evenly at corners Weight on scale may exceed the zero limit Unplug for one 1 minute to reset system System Error Unplug the unit for one 1 minute to reset system then reconnect to original power source Weight on scale exceeds capacity Remove excess load The load placed on the scale is too heavy on a corner shown as X Move load to center of the scale platform
10. e on a regular basis Calling For Service Note Model Serial Number information is required when calling for service If service is required contact your Midmark dealer To contact Midmark directly 1 800 Midmark 1 800 643 6275 or 937 526 3662 8 00 am until 5 00 pm Monday through Friday EST excluding standard U S holidays 10 Specifications Weights Dimensions Electrical Specifications Maximum Patient Weight AAA 300 Ibs 136 kg Weight of Table 45 tabletop eie GAA Rete 366 Ibs 166 kg 45 tabletop with packaging amp Skid 435 Ibs 197 kg 60 tabletop iconic tet iiec i 395 Ibs 179 kg 60 tabletop with packaging amp Skid 465 Ibs 211 kg Add 5 Ibs 2 2 kg for units equipped with a scale Range of Motion amp Dimensions sss see illustration below Electrical Requirements AA Dedicated Circuit 115 VAC 60Hz 10A Power Cord acciona e a a E 4 ft 121 cm long Duty Cycle CO Intermittent Operation 1 minute ON 9 minutes OFF pump run time coococcnncccnccccancnananonno nana non c conc cnn cnnnncnnncns Intermittent Operation thermal protection 90 minute ON 230 case temp GFI Receptacle cccccccecesceeeeeneeeseneeeeeneeeesseeeesnees Class A 2P 125V 60HZ