Home

CL-200 Manual - Warner Instruments

image

Contents

1. 65 C While specifically designed for use with Warner s Heating Cooling Platforms and In Line Solution Heater Coolers this instrument can be easily adapted for use with custom built equipment according to the requirements of the user When coupled with the SC 20 Dual In line Solution Heater Cooler the CL 200 provides efficient control of perfusion solution temperatures The unit is simple to use with a single control for temperature adjustment Total automatic control is provided in automatic mode while manual control is available in manual mode A loop speed selector is used to optimize the response of the system to accommodate thermal delay characteristics intrinsic to the setup Features of the CL 200 include Quiet operation suitable for use in sensitive electrophysiology applications Automatic and manual modes Single control temperature adjustment Freeze alert Built in protection for Peltier devices Selectable loop speed optimizes system speed and stability Simultaneous monitoring of system temperature and a separate point of interest Heat Loss compensation mode Open thermistor fault protection External inputs for computer control SN YN 088 amp SN NX NS NU 0 Compatible with all Warner Peltier driven Heater Cooler devices Warner Instruments A Harvard Apparatus Company NOMENCLATURE Text conventions This manual refers to instrument controls at two functional levels specific controls and settings of these c
2. to internal and the COOLING MODE SWITCH Item C 1s set to auto the SET TEMPERATURE CONTROL Item H will be active Setting the COOLING MODE SWITCH Item C to manual disables the SET TEMPERATURE CONTROL Item H and activates the MANUAL VOLTAGE CONTROL Item K Manual Voltage Control Item K This dial is used to manually adjust the voltage output to the heater cooler device The associated GREEN LED 1s on when the INTERNAL EXTERNAL SELECT CONTROL Item J 1s set to Warner Instruments A Harvard Apparatus Company internal and the COOLING MODE SWITCH Item C is set to manual Set value can be read from the METER Item B in the output volts position Peltier Limit LED Item L This option functions onlv when the CL 200 is used with an SC 20 Dual In line Solution Heater Cooler The RED LED is lit when the temperature of the Peltier device within the SC 20 exceeds its operational limits Activation of this circuit automaticallv sets the output voltage of the CL 200 to 0 V Freeze Alert LED Item M This option functions only when the CL 200 is used with an SC 20 Dual In line Solution Heater Cooler The BLUE LED is lit when the Peltier device within the SC 20 approaches 0 C Activation of this circuit is informative only and will not alter any instrument setting Heat Cool LED s Item N RED and BLUE LED S indicate heating or cooling power applied to the Peltier device Current Limit LED Item O This indicator light info
3. CL 200 Manual Rev 097014 Warner Instruments Dual Channel Bipolar Temperature Controller Model CL 200 Warner Instruments 1125 Dixwell Avenue Hamden CT 06514 800 599 4203 203 776 0664 203 776 1278 fax CE CL 200 Manual Rev 097014 INTRODUCTION iii i E ta data 4 NOMENCLATURE ii neonato 5 SAI AAA 5 CONTROL DESCRIPTION sa ii i A dicas 6 Front pane i i i A a a 6 POW LSW I ATI CHUA catia a DL eee hak ando dei ta 6 Meter Selector Switch and Meter Item B isis su L di ta da aa 6 COOL Mode SWwIHOM AICI OO 7 Ihermistor Recorder Outputs Items EV Nas aid 7 Monitor Temp In BING Nm F locen ene a a a a oa des 7 Loop SPCCA Switch CH CHD O a aa a ta A 7 Sel Lemperdinre Control eniT iaa 7 External Internal Select Control Tien is 7 Manual Voltage Control CHem K Jaren tii pe i mauus a tpe ra az 7 Feler emi LEDS AA tuo avete adea e ae RA us uod isa o e do Breeze Alert LE OIE E a Heat COOL AGED STII ANY a i i A a dia Current Eun LED OM Oo torpet a citet iia Thermistor Impedance Selector Switch Item Ps Heat Loss Compensation Dial Item Ola ER He I he heo VA RE NS A adt S La EE ptus Feedback Thermistor Selector Switch Item R eese nennen nennen nnne nennen nennen THEMES OO DELE ICIS as ili io cime 9 EXIemiut Volts Set Ito D soto att obe Da e Rt dons pp EL at Mi pe ds 9 External Temp Sem b
4. CT 06514 and we can be reached by phone at 800 599 4203 or 203 776 0664 Our fax number is 203 776 1278 In addition we can be reached by e mail at support warneronline com or through our Web page at http www warneronline com Service notes Please refer all questions regarding service to our Engineering Department A Ifthe instrument POWER light fails to light check the fuse at the rear panel located in the black POWER INPUT MODULE If the fuse is found to be defective replace it with a 5x20 mm 0 8 A slow blow fuse 0 4 A for facilities using 220 240 V line voltages If the replacement fuse also fails please call Warner Instruments for assistance B Occasionally a knob on the front panel will loosen after long use These are collet style knobs and are tightened with a screw located under the knob cap To gain access to the adjustment screw pry the cap off with a thin bladed screwdriver or similar tool C Should service be required please contact the factory The problem may often be corrected by our shipping a replacement part Factory service if required will be expedited to minimize the customer inconvenience D Instruments are inspected immediately upon receipt and the customer is notified if the repair is not covered by the warranty Repairs can often be completed in 1 2 days from our receipt of the instrument E If factory service is required please observe the following instructions 1 Package the instr
5. T LOSS COMPENSATION function is disabled The standard CONTROL THERMISTOR is the sensor located in the heating cooling device This sensor is placed at the optimal point to control the temperature of the heating cooling device The MONITOR THERMISTOR is the peripheral thermistor that can be used to monitor the temperature at another location in the experimental configuration This thermistor monitors the temperature at the selected point in the experiment and is more accurate but is less stable in the control loop because of the delay between heating or cooling and sensing To protect the experiment if the monitor thermistor is removed the limit of the increase in temperature between the control and monitor thermistors in the feedback loop is 8 C This lowers the probability of overheating the experiment to the point of damage There is no protection from overcooling caused by a misplaced monitor thermistor A switch above the T2 INPUT CONNECTOR provides selection of the thermistor impedance for both the control thermistor T1 and the monitor thermistor T2 for that channel The thermistor impedance for Warner Instruments equipment is 10kQ 25 C while the thermistor impedance for Harvard equipment is 100kO 25 C An indicator on the front panel indicates an open thermistor fault of the selected feedback thermistor Normally if a thermistor becomes open as in a broken wire the controller sees this as a lower temperature The controll
6. amden CT 06514 Tel 203 776 0664 Power Controller Safety requirements for electrical equipment for measurement and laboratory use Class I CL 200 I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standards Place Hamden Connecticut USA Signature Full Name Burton J Warner Position President Warner Instruments A Harvard Apparatus Company WEEE RoHS Compliance Statement EU Directives WEEE and RoHS To Our Valued Customers Harvard Apparatus is committed to being a good corporate citizen As part of that commitment we strive to maintain an environmentally conscious manufacturing operation The European Union EU has enacted two Directives the first on product recycling Waste Electrical and Electronic Equipment WEEE and the second limiting the use of certain substances Restriction on the use of Hazardous Substances RoHS Over time these Directives will be implemented in the national laws of each EU Member State Once the final national regulations have been put into place recycling will be offered for those Harvard Apparatus products which are within the scope of the WEEE Directive Products falling under the scope of the WEEE Directive available for sale after August 13 2005 will be identified with a wheelie bin symbol Two Categories of products covered by the WEEE Directive are currently exempt from the RoHS Directive Category 8 medi
7. apter cable When the CL 200 is connected to a resistive device all command inputs auto and manual modes for both internal and external commands are functional with the exception that cooling commands are not executed Third party Equipment The CL 200 heater controller will work with many other heating devices if used in conjunction with the AC 100 adapter cable To obtain maximum heating power the resistance of the third party heating element should be between 8 12 However any element that works at a maximum of 15 V and 2 4 A will also work with the CL 200 Thermistor Considerations The CL 200 heater controller is designed to accommodate Unical thermistors from Thermometrics Edison New Jersey This family of thermistors can be interchanged without recalibration of the instrument The nominal resistance of a Unical thermistor is 10 0 kQ at 25 C Other thermistors may be used with the CL 200 if the nominal resistance is also 10 0 kO at 25 C Warner Instruments A Harvard Apparatus Company OPERATION The main use of a heater cooler control device such as the CL 200 is to maintain a constant bath temperature with minimum deviation from a set temperature This is usuallv achieved bv usinga thermally controlled chamber platform or in line solution heater or a combination of both AUTOMATIC MODE In Automatic Mode the CL 200 maintains the temperature of the connected heater cooler at the value set by the user Operation
8. ature will be compensated Changes in ambient temperature or set temperature will also be compensated With the METER SELECT switch set to Set Temp the desired temperature is displayed on the panel meter Changing the meter select switch to Compensated Set Temp shows the temperature at the heating device needed to heat the point of interest to the Set Temp temperature Adding heat loss compensation is similar to moving the control thermistor away from the heating or cooling source This slows the forward loop speed in the feedback system reducing the stability of the overall system This makes it desirable to keep the heat path as short as possible If the perfusion path is too long or the chamber losses to the atmosphere are too great then the compensation will not be stable The time needed to stabilize this temperature will be longer than the time to stabilize at a temperature without the compensation Also if changes in conditions occur quickly then the system will need time to make the corrections Feedback Thermistor Selection A FEEDBACK THERMISTOR switch allows the user to select which thermistor is used for the control loop Choosing T1 control thermistor sets the peripheral s internal thermistor to control the feedback loop Selecting 72 monitor thermistor allows feedback control at the location of the monitor thermistor as selected by the user Note that when T2 is selected as the feedback thermistor the HEA
9. cal devices with the exception of implanted or infected products and Category 9 monitoring E and control instruments Most of Harvard Apparatus products fall into either Category 8 or 9 and are currently exempt from the RoHS Directive Do Not Dispose Product with Harvard Apparatus will continue to monitor the application of the RoHS Municipal Waste Directive to its products and will comply with any changes as they apply Special Collection Disposal Required Warner Instruments A Harvard Apparatus Company
10. e to be applied to the heater cooler being controlled Input functions only when the INTERNAL EXTERNAL SELECT CONTROL Item J is set to external and the COOLING MODE SWITCH Item C is set to manual Units are 1 V V input This input can be used for computer control purposes External Temp Set Item U Input BNC for adjusting the SET TEMPERATURE of the CL 200 Input functions only when the INTERNAL EXTERNAL SELECT CONTROL Item J is set to external and the COOLING MODE SWITCH Item C is set to auto Units are 100 mV C This input can be used for computer control purposes Heater Cooler input Item W This is the connection port for Warner Heater Cooler devices such as the SC 20 Dual In line Heater Cooler and QE 1HC Heating Cooling Platform Power Entry Module Item Y Comprised of the line cord attachment point line voltage selector and fuse buss Power entry is selectable between 115 VAC 60 Hz or 230 VAC 50 Hz The replaceable fuse is 0 4 amp slow blow for 230 V applications or 0 8 amp slow blow for 115 V applications Ground Terminals Item X Provides separate connections for chassis and circuit ground Banana Jacks are bridged when shipped from the factory Warner Instruments A Harvard Apparatus Company SETUP The CL 200 has been designed primarilv to drive Warner s expanding line of Peltier driven heating and cooling devices However the instrument is also capable of supplving power to Warner s line of resisti
11. eni U seen im ivit A a UM IS 9 Heater Cooler input Item Mais 9 Power Entry Module C Hent Lo acest see tht ette A A AA ens Mo a 9 Ground Lerman SALTA ssa timeo vete EE tM A E 9 SETUP 10 Warner Peltier Driven SYSCCMS iii G A 10 Operating With Other Equipment iii iii iii E dua Fu do EE n xe OA 10 Warner Resistive Hearne Sy SIOembss 4 ease dix at a C aca 10 T OA AVL EQUINE ics restate staal eh ia utu eii teat uiuos tie bites dete 10 Thermistor ConsiuderallOWy sau ees dese on aa dia aa a pue vos idas 10 OPERATION c a a a M 11 Table of Contents AUTOMA TICMODE Dose 11 CL 200 Manual Rev 097014 Heat L ss Compensaciones 11 Feedback Thermistor Selec ici iia ica a ta da a b dad 12 MANDAL MODE bi iii ig a a bla b A a L 13 APPENDIX anni tas 14 SDE ITI CA cOn aida inosd 14 Accessories and replacement parts eee tete yet tasa eer aveva vo nu eee ee venas ua ei eun een Rae Na EE eaa VN E SER TEUER Prae 14 Warranty and E quatur 15 WATAN Vi e a Sa A lt Mali 15 SAUCIUS ER At E ele 15 COFDBCUOOEHS o A a A RR CE eT NA aa 16 The CL 200 Dual Channel Bipolar Temperature Controller from Warner Instruments is a versatile and user friendly thermal control device This dual channel automatic temperature controller is capable of accurately maintaining two Peltier devices between 6 C and 65 C The unit will also maintain a resistive heater from ambient to
12. er will then increase the output power to try to raise the temperature This can cause overheating of the experiment if the fault 1s not noticed In the CL 200 the selected feedback thermistor is monitored for an open condition If an open fault is detected the output power is shut off preventing overheating A red LED indicates the presence of an open thermistor Warner Instruments A Harvard Apparatus Company MANUAL MODE In manual mode the MANUAL VOLTAGE control Item K is used to set the output voltage to a fixed value In manual mode the SET TEMPERATURE control Item H is disabled as is the feedback svstem of the CL 200 Operation is again straightforward Set the CL 200 to accept internal commands by setting the INTERNAL EXTERNAL SELECT control Item J to internal Place the CL 200 into manual mode by switching the COOLING MODE SWITCH Item C to manual Switch the METER Item B to Output Volts and adjust to the desired set point using the MANUAL VOLTAGE control Item K Alternatively the CL 200 can be commanded using external inputs Set the CL 200 to external commands by setting the INTERNAL EXTERNAL SELECT control Item J to external In this mode the CL 200 will accept external voltage commands at the EXT VOLT SET BNC Item T located on the instrument rear panel In this mode the instrument operates as a 0 15 VDC power supply that can provide up to 2 4 A of current to a device Use this mode with heater systems that w
13. ge Set BNC Heater Cooler Device 15 pin D connector Peltier 6 C to 65 C Resistive Ambient to 65 C 3 digit LED display of C or V 0 1 C 0 01 Volt 0 01 Ampere Set temperature C Control Temperature C Monitor Temperature C Output Voltage V Output Current A Heat Loss Compensation Temp C 8 9 X 42 6 X 29 2 cm W x H x D 10 Ib 4 54 Kg Ambient temperature 0 to 40 C Altitude sea level to 2000 m Relative humidity 0 95 Equipment is intended to be operated in a controlled laboratory environment Accessories and replacement parts ACC 1 64 1427 Adapter cable for heater only components Warner Instruments A Harvard Apparatus Company Warrantv and service Warrantv The model CL 200 is covered by our Warranty to be free from defects in materials and workmanship for a period of two years from the date of shipment If a failure occurs within this time we will either repair or replace the faulty component s This warranty does not cover instrument failure or damage caused by physical abuse or electrical stress inputs exceeding specified limits In the event that instrument repairs are necessary shipping charges to the factory are the customer s responsibility Return charges will be paid by Warner Instruments Inc Normal business hours are 8 30 AM to 5 30 PM EST Monday through Thursday and 8 30 AM to 5 00 PM on Friday Our offices are located at 1125 Dixwell Avenue Hamden
14. ill operate properly with a fixed voltage and lack of feedback thermistors The CL 200 has a convenient feature for switching modes In the CL 200 the control loop integrator is held at its last value when switching to off allowing the integrator to resume when the Automatic mode is again selected When the CL 200 is switched to manual mode the integrator tracks the manual voltage When the mode is set from manual to off this value is held Switching to automatic mode allows the control loop integrator to begin at the last voltage setting of the manual mode This allows the user to set the integrator to any desired initial value for the automatic mode Warner Instruments A Harvard Apparatus Company APPENDIX Specifications Maximum Output Voltage Maximum Output Current Manual Voltage Range Maximum Output Power Power Requirements Power Fuse 5 x 20 mm Front Panel Input Recorder Outputs Rear Panel Inputs Temperature Range Meter Display Resolution Meter Readouts Enclosure Weight Operating conditions 15 V Heat Cool 0 to 15 V Heat Only 5 0 Amperes DC per channel 0 to 15 VDC 75 Watts 30 Load Each Channel 100 130 or 200 260 VAC 50 60 Hz 80 VA 0 8 A Slow Blow for 100 130 VAC 0 4 A Slow Blow for 200 260 VAC External thermistor BNC calibrated to 10 0 kQ at 25 C Monitor Temp Out BNC 100 mV C Control Temp Out BNC 100 mV C External Temperature Set BNC 100 mV C External Volta
15. is straightforward Connect your heater cooler device and sensor thermistor as described on page 10 Set the CL 200 to internal commands by setting the INTERNAL EXTERNAL SELECT control Item J to internal Place the CL 200 into auto mode by switching the COOLING MODE SWITCH Item C to auto Switch the METER Item B to Set Temp and adjust to the desired set point using the SET TEMPERATURE control Item H Alternatively the CL 200 can be commanded using external inputs Set the CL 200 to external commands by setting the INTERNAL EXTERNAL SELECT control Item J to external In this mode the CL 200 will accept external temperature settings from the EXT TEMP SET BNC Item U on the instrument rear panel This input is calibrated to 100 mV C The selectable LOOP SPEED control Item G is used to adjust the speed of the feedback loop in the heater cooler thermistor system which controls the rate of change of the CL 200 output voltage Non Warner heater systems with feedback thermistors can be used in automatic mode provided they are compatible with the CL 200 requirements LOOP SPEED is normally set to fast to provide the shortest cycle time between the application of power to the heater element and the sensing of temperature at the thermistor For heater cooler thermistor systems with long response times such as with Warner s SC 20 in line solution heater cooler the fast setting will cause the temperature to overshoot the target resulting in
16. ontrols To increase readabilitv we emplov the following text conventions Since our goal is to provide claritv rather than complexitv we welcome anv feedback vou mav wish to provide Warner Instrument product model numbers are presented using bold tvpe References to controls are specified using SMALL CAPS References to control settings are specified using italic tvpe Special comments and warnings are presented in highlighted text Any other formatting should be apparent from context VN Y WV Warner Instruments A Harvard Apparatus Company CONTROL DESCRIPTION Front panel A labeled image of the face panel for one channel of the CL 200 is shown below Operating components are identified bv letters Refer to this diagram as an aid in identification of the described components Q G H R S K DO Me AI J L AAA PA e HEAT LOSS LOOP SPEED SET TEMP C HAN N E L 2 MANUAL Le COMPENSATION VOLTAGE LIMIT FAST 2 x e HEAT FEEDBACK Qe MONITOR TEMP 2 on seas OUTPUT VOLTS OUTPUT AMPS Le 100K 25 C CONTROL TEMP 9 E SET TEMP HLC TEMP CURRENT LIMIT Power Switch Item A Supplies power to the CL 200 Meter Selector Switch and Meter Item B THE SELECTOR SWITCH selects the parameter displayed on the associated METER Included options are HLC Temp Set Temp Control Temp T1 Monitor Temp T2 Output Volts and Output Amps and are described below Set Temp
17. reports the adjustment from either the SET TEMPERATURE control Item H or the EXT TEMP SET BNC Item Displayed units are C Control Temp reports the actual temperature of the feedback thermistor situated within the heater cooler device being used Displayed data is informative only and is in units of C HLC Temp reports the heat loss compensated temperature of a phantom thermistor at the site of interest Displayed data is informative only and is in units of C Monitor Temp reports the temperature of the sensor thermistor which connects via the MONITOR TEMP IN BNC Item F This provides a convenient means to determine the temperature of any point of interest within your configuration Displayed units are C Output Volts reports the adjustment from either the MANUAL VOLTAGE control Item K or the EXT VOLT SET BNC Item T Displayed units are Volts Warner Instruments A Harvard Apparatus Company Output Amps reports the current output command associated with either the MANUAL VOLTAGE control Item K or the EXT VOLT SET BNC Item T Displayed units are Amperes The METER also provides an indication that the main POWER is on Cooling Mode Switch Item C Selects between Automatic or Manual operational modes or center select for off Thermistor Recorder Outputs Items E V BNC connectors are provided to send thermistor readings to a data acquisition system or chart recorder MONITOR TEMP OUT Item E repor
18. rms the user that the channel has reached the maximum current output of 5 Amperes This indicator is informative only and will not alter the function or settings of the device Thermistor Impedance Selector Switch Item P This switch can be set to indicate the type of thermistor in use with the peripheral device The thermistor impedance for Warner Instruments equipment is 10kO 25 C while the thermistor impedance for Harvard equipment is 100kQ 25 C Heat Loss Compensation Dial Item Q This dial can be used in auto mode as an adjustment for the temperature gradient between T1 and T2 This circuit utilizes the change in power output to adjust for the heat lost from the system and maintain a stable temperature at the preparation Feedback Thermistor Selector Switch Item R This switch can be set to select the thermistor that will be used as the control point for the feedback regulation of set temperature Thermistor Open LED Item S This LED indicator will signify an open thermistor circuit broken thermistor wire Warner Instruments A Harvard Apparatus Company Rear panel A schematic of the rear panel of the CL 200 is shown below Important components are identified bv letter T X Refer to this diagram as an aid in identification of the described components CONTROL EXTERNAL HEATER COOLER TEMP Ti TEMP SET CHANNEL 1 mmm a External Volts Set Item T Input BNC for the application of a specified voltag
19. system oscillation For this condition try medium or slow settings to find a more optimal feedback rate Heat Loss Compensation An additional control sets the amount of HEAT LOSS COMPENSATION This allows the CL 200 to control the temperature at a location downstream from the heater or cooler where there is heat transferred to the environment By setting the amount of correction for the heat loss the Control Thermistor can control the temperature as if there were a phantom thermistor at the location of interest Changes in the Temperature Setting or the thermal load will be corrected and be reflected in the Compensated Set Temperature as displayed on the front panel meter Meanwhile the temperature at the point of interest will be maintained The amount of Heat Loss Compensation needed will vary with each setup With the HEAT LOSS COMPENSATION control set fully off counterclockwise the set temperature is maintained at the control thermistor Heat loss or gain in the case of cooling at the location of the experiment will cause a temperature closer to ambient than is desired for that set temperature Adjusting the set temperature further from the ambient temperature will correct for this loss but any change in Warner Instruments A Harvard Apparatus Company conditions will cause a new error If the temperature is instead changed using the HEAT LOSS COMPENSATION control then changes in perfusion rate or input fluid temper
20. ts the temperature of the sensor thermistor which connects via the MONITOR TEMP IN BNC Item F Output is calibrated to 100 mV C CONTROL TEMP OUT Item V back panel reports the actual temperature of the feedback thermistor situated within the heater cooler device being used Output 1s calibrated to 100 mV C Monitor Temp In BNC Item F BNC for connecting a sensor thermistor used to sample the temperature at a point of interest This input is designed for use with the TA 29 cable assembly Loop Speed Switch Item G This switch sets the feedback loop speed for the CONTROL TEMP THERMISTOR when the CL 200 is used in automatic mode Feedback options include fast medium and slow Optimally set to fast for most applications but can be set to medium or slow for systems with longer thermal delay characteristics For example use the medium speed for use with the SC 20 Bipolar In Line Solution Heater Set Temperature Control Item H This dial adjusts the set point of the automatic temperature control system The associated GREEN LED is on when the INTERNAL EXTERNAL SELECT CONTROL Item J is set to internal and the COOLING MODE SWITCH Item C 1s set to auto Set value can be read from the METER Item B in the set temp position External Internal Select Control Item J This switch selects between internal or external control inputs when the instrument is in Manual mode Item C When the INTERNAL EXTERNAL SELECT CONTROL 1s set
21. ument with at least 3 inches of cushioning on all sides Use the original shipping carton if it is available 2 Insure the shipment for its full value 3 Include with the shipment an explanation of the problem experienced Warner Instruments A Harvard Apparatus Company Certifications Declaration of Conformitv CE MARKING EMC Application of Council Directive 89 336 EEC Standards To Which Conformity Is Declared Manufacturer s Name Manufacturer s Address Equipment Description Equipment Class Model Numbers EN55022 Class A EN61000 3 2 EN61000 3 3 EN50082 1 1992 EN61000 4 2 EN61000 4 3 ENV50204 EN610000 4 4 EN610000 4 8 EN610000 4 11 Warner Instruments LLC 1125 Dixwell Avenue Hamden CT 06514 Tel 203 776 0664 Power Controller ITE Class A CL 200 I the undersigned herebv declare that the equipment specified above conforms to the above Directive s and Standard s Place Hamden Connecticut USA Pah gria Full Name Burton J Warner Signature Position President Warner Instruments A Harvard Apparatus Company CL 200 Manual Rev 090714 17 Declaration of Conformitv CE MARKING LVD Application of Council Directive 73 23 EEC Standards To Which Conformitv Is Declared Manufacturer s Name Manufacturer s Address Equipment Description Equipment Class Model Numbers EN61010 1 1993 Warner Instruments LLC 1125 Dixwell Avenue H
22. ve heaters Series 20 heater platforms in line solution heaters etc The instrument will auto detect when a resistive onlv heating device has been attached and will disable cooling commands Warner Peltier Driven Svstems The set up for the CL 200 is straightforward when used with a Peltier driven heater cooler First connect the cable from the heater cooler into the 15 pin D connector on the back of the CL 200 Then connect the TA 29 thermistor supplied with the heater cooler to the MONITOR TEMP IN BNC ltem F If using the SC 20 in line solution heater cooler set the CL 200 loop speed to medium If usinga different device then set the loop speed to the setting specified in that devices user s manual or initially set the loop speed to fast and adjust as necessary to achieve stable performance If using external control inputs make connections from vour command output e g analog out on the A D board or a power supply to either the EXT TEMP SET Item U or EXT VOLT SET Item T BNC s on the rear panel Operating With Other Equipment The CL 200 can be used to power other Warner equipment such as the SH 27B and SF 28 inline solution heaters In addition the controller can be used as an independent device to power third party equipment so long as the following considerations are met Warner Resistive Heating Systems Use of the CL 200 to power Warner resistive heaters is permitted if used in conjunction with the AC 100 ad