TaqMan hPSC Scorecard Panel
Contents
1. o o o 8 8 o oO Figure 2 The top panel shows the pass fail results in the main interface of the hPSC Scorecard Analysis Software The graphs on the bottom show the scores for each of the four sub panels tested These scores measure how closely the sample aligns to the expression profile of the nine undifferentiated pluripotent stem cell lines in the reference set In general scores close to zero indicate similarity to the reference set and scores greater than about 1 indicate upregulation relative to the reference set and scores less than 1 indicate downregulation relative to the reference set Notice that after seven days of EB formation the pluripotency or self renewal markers are significantly downregulated and ecto meso and endo factors are upregulated Also note that the scores for each germ layer are independent of the others For example the magnitude of the scores for ectoderm cannot be compared to the magnitude of the scores for endoderm Scores can only be compared among different samples for the same germ layer Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 4 Thermo Fish o m SCIEN TIFVE TaqMan hPSC Scorecard Panel FAQs 18 Can I analyze somatic non pluripotent primary cells with the TaqMan hPSC Scorecard Panel Somatic non pluripotent primary cell
2. sel Se renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mesoderm Endoderm o o 8 o o gt o Hg ESC Hg TRIzol 0 25 0 69 0 52 0 86 Hg Cellstoct 0 04 0 45 0 34 0 62 Hg CellsDirect 0 52 0 91 0 45 0 94 Figure 2 All samples regardless of sample size pass the test for undifferentiated cells Cells were cultured in KSR based media on irradiated MEFs in the culture dish formats shown prior to Scorecard analysis Note that negative scores for germ layer markers in undifferentiated samples just indicates lower levels of these factors relative to the undifferentiated reference standard lines and is a perfectly acceptable result The difference between scores of the self renewal pluripotency factors in these samples is unlikely to be statistically significant A minimum of 15 000 undifferentiated cells or 6 embryoid bodies is required to obtain an answer with the hPSC Scorecard analysis using the Cells to Ct or CellsDirect protocols When purifying RNA a minimum of 200 000 cells is recommended Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 6 0 Fisher Therm e m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs 17 Do I need to perform DNase treatment We recommend that you include DNase treatment of all samples as a good practice despite the fact that the majority of primers
3. ThermoFisher m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs TaqMan hPSC Scorecard Kits TaqMan hPSC Scorecard Panels E Sf ft Be i pr old f A15871 A15872 A15876 A15870 Product Qty Cat No TaqMan hPSC Scorecard Kit 2 x 96w FAST 1kit A15871 TaqMan hPSC Scorecard Kit 384w 1kit A15872 TaqMan hPSC Scorecard Panel 2x 96w FAST _ 2 plates A15876 TaqMan hPSC Scorecard Panel 384w 1 plate A15870 1 What is included with each kit or panel Panel configurations e Two 96 well 2 x 96w FAST plates with optical plate covers One 384 well 384w plate with optical plate covers Kit configurations Two 96 well 2 x 96w FAST plates with optical plate covers amp TaqMan Gene Expression Master Mix e One 384 well 384w plate with optical plate covers amp TaqMan Gene Expression Master Mix Part 1 Sample Generation 2 How many cells are required to run one experiment While we recommend that you use about 0 5 million cells per experiment or the number of cells equivalent to 1 well of a 6 well dish you can use as few as 100 000 cells when performing RNA purification If you perform a lysis protocol using Cells to CT or CellsDirect One Step qPCR kits as few as 15 000 cells is sufficient Please note however that reducing cell number can compromise the quality of the results s Can I use cells that are on feeders or in feeder free conditions Yes
4. The hPSC Scorecard Panel measures the potential for self renewal and trilineage differentiation of PSCs grown on feeders or in feeder free conditions Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support n Thermo Fish o a SCIEN TIFIE TaqMan hPSC Scorecard Panel FAQs 4 How do I harvest human PSCs cultured using feeder dependent culture systems We recommend removing feeders from your culture before proceeding with Scorecard analysis To remove feeders harvest your PSCs with collagenase allow the colonies to settle by gravity sedimentation to reduce feeder carryover and re seed cells in feeder free conditions on Geltrex matrix coated dishes and MEF conditioned medium for a 1 to 2 passages The presence of feeders can contribute to gene expression measured in the assay thus altering the gene signature pattern 5 Can I use cells on a feeder free system cultured in novel media The hPSC Scorecard Panel measures self renewal and trilineage differentiation potential of PSCs and is not restricted by culture conditions You should be aware that using novel media systems may have particular effects on the cells in terms of pluripotency and or their differentiation We recommend designing experiments that use novel media by including a control condition that utilizes traditional media or media in which expression patterns have been tested
5. stemoell help lifetech com for support 6 Thermo Fish e m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs Part 6 Examples of Results The hPSC Scorecard Analysis Software algorithm compares the data in the user s file to the data in the reference cloud and assigns a score based on how closely the data align with the reference set of nine undifferentiated pluripotent stem cell lines The plus or minus sign in the user interface indicates whether the sample s score met the criteria for either self renewal pluri or trilineage differentiation ecto endo meso For more information about hPSC Scorecard assay results and how to Interpret results refer to the RPSC Scorecard Analysis Software Help system EB time course H9 ESCs randomly differentiated using EB formation show the appearance of markers that are representative of the 3 germ layers as early as Day 4 of differentiation a Wo EB O0y2 1s HOEB Dayd Wetih Eetogerm Mesodeim Endodemm renewal Ectoderm Mssodsrm Endoderm renew Ecioderm Mesocerm Endoderm renewal Ectoderm Mesoderm Endederm o o o gt 8 oo o o Oa U oa W ae ake WEB Daria H EB Oay9 renal Esto Mesodem Engoderm rental Eetagenm Wesossrm Endod renewal Ectoderm Mesoders Endoder oo O O 8 O O amp O O 6 ake ake ake Hg ESCs were culture in KSR based media on irradiated MEFs prior to removal of FGF Hg EBs were cultured in suspension over
6. F Oa F BssciPse Se Sel Set renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mesoderm Endoderm renews Ectoderm Mesoderm Endoderm o o o o o o o SIC St Oa E Oa F oi W Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 20 ThermoFisher SCIENTIFIC TaqMan hPSC Scorecard Panel Part 7 Troubleshooting Information FAQs Troubleshooting Guide Result Probable Cause Remedy Pluripotent Failure My cells are pluripotent but are called out Pluri Fail Pluri scores are high Lineage scores are high Pluripotent lines may have marginal differentiation Ensure that the cells are high quality with little to no differentiation at the time of harvest for gene expression analysis Cells may be contaminated with MEF feeders While most primers are specific for human cells a few lineage primers cross react with mouse cells and may skew results especially for the ectoderm layer We recommend that you eliminate contamination from the cells cultured under feeder dependent conditions by performing at least one feeder free passage Pluri scores are low Lineage scores are high There may have been extensive differentiation during culture or clones may have reprogrammed partially or incompletely We recommend using stable iPSC clones that sho
7. and confirmed The most recent version of the hPSC Scorecard Analysis application offers a differentiation index plot for viewing gene expression in EBs relative to the undifferentiated state 6 What is the minimum time required to form EBs and is there a specific protocol You can differentiate cells using any of the established methods When using suspension embryoid bodies we recommend that you allow at least 7 days of differentiation prior to analysis 7 Iam only interested in directed differentiated cells Do I need to use a particular method You can perform directed differentiation according to your own methods However the time point when expression is noticeable will largely depend on the robustness of the methods We recommend testing a few time points to monitor differentiation with time 8 Have you tested cells grown in NIM with Scorecard analysis In the data shown for directed differentiation of cells into ectoderm in monolayer culture the 3 day time point is too early to detect ectoderm expression but the cells become positive for ectoderm by day 5 H9 cells were differentiated into NSCs using Gibco PSC Neural Induction Medium denoted as NIM and hPSC Scorecard analysis was performed at various time points The control sample is undifferentiated H9 ESC Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 2 Th
8. other countries Safari is a trademark of trademarks of Apple Inc registered in the U S and other countries Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 2
9. samples types for analysis 40 Do I need to run a new sample if I want to look at a different sample type No The software analyzes and retains all of the qPCR data from the plate upon import If you want to look at a different sample type simply select the desired sample type from the drop down list and the analysis software will automatically update the results Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 4 Fisher Therm e m SClENTIFIE TaqMan hPSC Scorecard Panel FAQs 41 My sample has a red flag on it or is not showing any results What does that mean The hPSC Scorecard Analysis Software performs a set of quality controls checks upon data import to ensure the quality of the results For more information about hPSC Scorecard assay results and how to Interpret results refer to the hPSC Scorecard Analysis Software Help system You may see one of the following red flag warnings Internal positive control The expression levels of the housekeeping genes were low signifying that there was an issue with the qRT PCR The sample is excluded from analysis and must be repeated e Bad Rox detected it is suggestive that some of the assay wells may not have received the proper amount he ROX dye is used as a passive reference during the qPCR If the ROX dye is not of Master Mix This may explain
10. span across an intron and do not amplify or in some cases minimally amplify contaminating genomic DNA See http products invitrogen com ivgn product AM1906 for more information 18 How long can I store RNA after it is prepared Properly prepared RNA will last for years when stored properly at 80 C 19 How much RNA is required for the RT reaction We recommend using 1 ug of RNA per sample Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 7 Fisher Therm e a SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs Part 3 Sample Loading 20 Why should I set up cDNA synthesis in 8 wells of a 96 well plate or in 8 well PCR strips Setting up your cDNA synthesis in 8 wells of a 96 well plate or in 8 well PCR strips facilitates sample loading of the 96 well and 384 well hPSC Scorecard Panel plates Follow the schematic below RNA RT Master Mix 8 x 50ul Reaction Dilute with 20 ul H2O 70 ul 2X TagMan Gene Expression Master Mix 21 How do I load the samples on to my 384 well plate if I don t have a multichannel pipette You can load samples using a single channel pipette but this method is time consuming and may increase pipetting error We strongly recommend that you use an 8 or 16 channel multichannel pipette You also can dispense samples using automated systems with the understanding that additional
11. the course of 14 days prior to scorecard analysis The summary shows that the cells are positive for pluripotency EB formation at day 2 is positive for ectoderm but negative for meso and endoderm As the EB formation progresses from day 4 to day 14 the scores for all three germ layers continue to increase whereas the scores for the self renewal factors decrease The table with the scores also reflect the same pattern You may notice a tiny red flag M in one of boxes which appears because the software includes quality control checks and lets you know that feedback and additional information is waiting for you Random differentiation via EB formation versus monolayer H9 ESCs differentiated for four days either using EB formation or monolayer differentiation in EB media show appearance of markers for the 3 germ layers by Day 4 Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support a ThermoFisher o m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs When a culture is allowed to differentiated by removing FGF or not changing media the expression of SSEA4 drops and the scorecard fails pluripotency but shows trilineage differentiation Early versus late passage iPSC BS3C iPSC an established iPSC clone is scored as pluripotent and the parental fibroblast displays mesoderm positive expression The 5 clones derived from BJ fibroblasts usin
12. thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support TaqMan hPSC Scorecard Panel FAQs Value Range Definitions High Zero Low positive Scorecard value Scorecard value close to o negative Scorecard value Pluripotent When compared to the Reference standard pluripotent markers are Upregulated Comparable Downregulated Depending on germ layer markers sample may be pluripotent Sample is not pluripotent Germ Layers Ectoderm Mesoderm Endoderm When compared to the Reference standard germ layer markers are Upregulated Comparable Downregulated Sample is differentiated not pluripotent Depending on pluripotent markers sample may be pluripotent For Research Use Only Not for use in diagnostic procedures Last update 15 Oct 15 2015 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified Essential 8 is a trademark of Cellular Dynamics International Inc TaqMan is a registered trademark of Roche Molecular Systems Inc used under permission and license TRIzol is a registered trademark of Molecular Research Center Inc Google and Chrome are trademarks or registered trademarks of Google Inc Internet Explorer is a registered trademark of Microsoft Corporation in the United States and or
13. why a particular assay is not showing the anticipated gene expression level Sendai Virus Detected SEV The hPSC Scorecard Panel includes a control to look for Sendai virus expression A flag will appear if a Ct value of greater than 30 is observed for this control On some rare occasions the SEV oligonucleotides will cross react with other genes resulting in a false SEV flag in differentiated cells If you see this flag and did not anticipate Sendai virus detection you may want to repeat the sample or look at the amplification plot In the 96 well plate well B9 in the 384 well plate wells D3 D9 D15 and D21 to see if the automated Ct value call is real or just an error e Insufficient Data The analysis software requires a minimum number of genes to be amplified in each the four categories i e self renewal pluri ectoderm mesoderm and endoderm to accurately analyze your data If the minimum number of genes in each category does not amplify or if data has been omitted by the user the Insufficient Data flag will appear and the sample is excluded from analysis The sample should be repeated 42 What lines were used to generate the reference standard Thirteen pluripotent stem cell lines were used to generate the reference standard including e H9ESC P28 e HUES13 P56 e HUES48 P21 e HUES63 P46 e iPS BS3C P35 e HUES28 P28 e HUES49 P21 e HUES64 P30 e iPS18C P29 e HUES44 P25 e HUES53 P24 e HUES65 P29 e HUES9 P28 43 How ar
14. Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support a Fisher Therm e m SCIENTIFIE TaqMan hPSC Scorecard Panel FAQs 31 How many genes are tested per sample 93 There are 9 self renewal genes 74 lineage specific genes and 10 housekeeping and control genes s2 I accidentally ran a comparative Ct experiment type instead of a standard curve experiment type Can I still use my data with the hPSC Scorecard Analysis Software The hPSC Scorecard Analysis Software currently only accepts data collected with the Standard Curve experiment type To change your experiment type open your data in the instrument software In the Set up menu change your experiment type to Standard Curve You may receive a warning message but that is okay After you change the experiment type return to the Analysis menu and reanalyze the data by pressing the green Analyze button After you save your data you can import the file into the hPSC Scorecard Analysis Software Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 2 isher ThermofFi o a SeIENTIFIE TaqMan hPSC Scorecard Panel FAQs Part 5 Data Analysis 33 What web browsers are compatible with the analysis software The hPSC Scorecard Analysis Software is compa
15. e the scores calculated A proprietary algorithm compares the Ct values for each marker set to the values in the reference database and calculates the score based on how well the expression correlates In general scores close to 0 indicate comparable expression to that of the reference standard using undifferentiated cells Scores higher than 1 indicates up regulation relative to undifferentiated cells and less than 1 indicate down regulation relative to undifferentiated cells Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 3 ThermoFish o m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs 44 How do I export Ct values in the software The only way to view the Ct values in the current version is through the expression plot We expect the software s Excel export in include the gene names and Ct values in a future version of the software 45 My results don t look right how do I get help Please send an email to our stem cell technical support specialists at stemcell help lifetech com 46 Will I need to update my software No software update is ever required Because the hPSC Scorecard Analysis Software is cloud based you will always have access to the most up to date version of the software every time you log in Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact
16. ermoFisher o a SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs fp SC RE 3Days SDays 7 Days Neural Induction H9 ESC Media A1647801 mple Name Ectoderm Hg CM Ctrl 0 53 H9 NIM Day3 3 62 012 H9 NIM Day5 H9 NIM Day7 9 What is the earliest passage that I can use when I generate iPSCs We recommend that you culture iPSC clones to at least passage 8 10 until they are stable and homogeneous prior to analyzing the hPSC Scorecard Panel Early passage iPSC clones may give low self renewal pluri in v1 1 of the analysis software scores or show higher expression of lineage genes See Part 6 Examples of results for more information on how to interpret the hPSC Scorecard Analysis Software output SIPS E tbroblast Table 1 Comparison of hPSC Scorecard Panel results at different passage numbers At passage 5 Ps the expression profile of cells does T not correlate to that of a truly pluripotent cell line At passage 8 P8 the BSe PSC1PS F expression profile matches the reference standard for the four good BSE IPSC3PS x clones BS4 iPSC2 BS4 iPSC3 BS4 iPSC7 BS4 iPSC8 but not for the BSE SCS Corer meso and ecto represent qualities of a good clone those shaded green BSEPSC7 PS indicate upregulation gt 1 and shaded red indicate downregulation lt 1 one poor clone B54 iPSCs Scores close to zero 0 for pluri endo 10 When I use Sendai
17. g CytoTune reprogramming at passage 5 and 8 are marked pluri negative at early passage but marked pluri positive by P8 This may be due to heterogeneity of the cultures at early stages and therefore iPSC at closer to passage 10 is recommended for the hPSC Scorecard Panel analysis Red flag indicates the presence of SEV that is detected at P5 in some clones but is diluted and not detected at P8 4 of the 5 clones pass the self renewal test pluri at P8 while 1 clone fails at P8 and hence is not ideal to be chosen for further culturing o o o 0 ojo 0 0 OjO o 6 0 O aje T RNA was isolated from early and late passage iPS clones that were cultured in KSR based media on irradiated MEFs then analyzed using the hPSC Scorecard CytoTune va is detected at passage 5 in these clones They do not clear the Sendai virus the tiny red flags indicate SeV and pass pluripotency until passage 8 Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support a8 Thermo Fishe e m SCLENTIFIG TaqMan hPSC Scorecard Panel FAQs Impact of SEV presence on pluripotency Clones derived using CytoTune reprogramming kits that have residual presence of SEV do not affect pluripotency or trilineage differentiation m DFIsta PIs DFA4S1 4 EBDay7 m HO derived NSC BJ HDFn PS Set Sal Sat reneial Ectoderm Mesocerm Endoderm
18. le or down regulated germ layer factors Non human cells and genomic DNA Because the primers are human specific and most assays are not amplified murine embryonic feeders show up as insufficient data left most data However cells harvested along with MEFs may result in an incorrect self renewal read out since the 10 genes that are amplified can impact the results It is therefore recommended to harvest ESCs with minimal MEF contamination Genomic DNA harvested from cells also are called out as insufficient data because most Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 39 ThermoFishe o m SCIEN TIFIE TaqMan hPSC Scorecard Panel FAQs primers span an intron to prevent amplification of contaminating gDNA However as best practice DNase treatment is recommended MEFs H9 ESC MEFs H9 ESC H9 ESC Genomic DM renateel Ectoderm Mesoderm Endoderm renewal Ectoderm Mesoderm Endodem o o o 0o O 6 6 6 Cells in feeder free media system H9 ESC and CytoTune kit derived iPSC cultured in feeder free media system such as StemPro SFM on Geltrex or Essential 8 Medium on Vitronectin are scored positive for self renewal pluri HO Esc P50 HO Stom HO EBVO Ser el Set renewal Ectoderm Mesodarm Endoderm renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mesoderm Endod o o o o o o o O E Oa
19. m e m SCIENTIFIE TaqMan hPSC Scorecard Panel FAQs Part 4 Gene Expression Assay 24 What qRT PCR instruments are currently supported ViiA 7 QuantStudio 12K Flex QuantStudio 7 Flex QuantStudio 6 Flex StepOnePlus 7500 Fast and 7900HT Fast Real Time PCR systems are compatible with the hPSC Scorecard Panel 25 What if I don t have access to a compatible qRT PCR system The hPSC Scorecard Panel analysis is also available as a service You can send Life Technologies isolated RNA samples cDNA samples or even cells in TRIzol reagent See lifetechnologies com cellmodel for more information 26 Where can I find the run template files for a compatible qRT PCR system Template files are available for download from the product catalog pages and at lifetechnologies com scorecardhelp 27 How do I set up my experimental run The template files contain all of the necessary experimental details for you to run your instrument There is no need to modify any of the experimental details if you are using the template file edt Step by step guides on how to run the qPCR instruments are available at lifetechnologies com scorecardinstrument 28 Do I need to use the template files We highly recommend that you use the template file because it contains all of the necessary experimental details for your qPCR run If the template file is not working for you please send an email to our stem cell technical sup
20. port specialists at stemcell help lifetech com 29 How do I access the hPSC Scorecard Analysis Software Please visit lifetechnologies com scorecarddata for access to the analysis software so T want to know the list of genes before I run the samples Where I can find it To see the gene names you ll need to create a demo project a Log in to the hPSC Scorecard Analysis Software at lifetechnologies com scorecarddata with your Life Technologies username and password Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 10 ThermoFishe o m SCLENTIFIG TaqMan hPSC Scorecard Panel FAQs b Click on the encircled link to create your demo project c Click the Demo Project name in the Project list For npormens Desenption d Click on the checkbox next to the name of the project above the results summary panel Results Summary Demo project Pates maven Eo Eo L e Click the dark gray Reports Button at the upper right of the software f Click the Excel XLS tab select the results to include Scorecard results Scorecard values CT values Fold change values and the Excel file type as either XLS or XLSX and then click Download Data g Open the spreadsheet file and click on the CT Values worksheet tab to see the list of gene targets
21. renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mosoderm Endoderm o J o o o o o o o o t Oa E DKF DHT oF Examples of results obtained primary cells can be highly variable since the panel was not designed for use with terminally differentiated cells In addition donor to donor variation is to be expected Skewed lineage and non pluripotent somatic cell NSCs derived from H9 ESCs show expression of ecto without endo or meso genes The efficiency and specificity of the differentiation method largely dictates the expression pattern Non pluripotent stem cells such as BJ fibroblast derived human foreskin cells show expression in ecto However prolonger culture of these cells can create a drift and BJs with ecto and meso positive pattern have also been observed Ho ESC P50 HO StomPre HOEDMtro Ser Sert Ser renewal Ectoderm Mesodarm Endoderm renewal Ectoderm Mesoderm Endoderm renewal Ectoderm Mexoderm Endod o o o o t La t T ve BEE Hg ESC O W eo i OA F m essc PSC 8590 St Pr BSC Ebvo ental Eetodarm Mesoderm Endoderm renewal Ectoderm Mesoderm Endoderm renewal Eclodemn Mesoderm Endoder o o o o oO o o o 80808060 o ra ee ea e Sae Bs3ciPS OA E OA W OA k Note that all media conditions show expression profiles similar to the reference set for the self renewal factors and comparab
22. roject window and verify that you have used one of the following TaqMan hPSC Scorecard catalog products A15870 A15876 A15872 or A15871 The analysis software is compatible only with data obtained with those specific panels If you see an error message that identifies that the experiment was not analyzed by the Ct method follow the steps outlined below a Open the eds file in your data collection software b Click to open Analysis Settings dialog from the menu go to Analyze gt Analysis Settings Check to see if Ct algorithm is selected i Tf Ct is selected close Analysis Settings dialog ii If Crtis selected change it to Ct and then save and close the Analysis Settings dialog Click the Analyze button e Save it f Click Upload Data to import the new file 39 What sample type should I select The software allows you to easily evaluate pluripotency and or inductive differentiation one germ layer at a time To make it a little easier a set of common sample types have been already built into the software to allow you to get to your results faster e All Pluripotency Quickly confirm the self renewal capacity of undifferentiated samples All Ectoderm All Endoderm All Mesoderm Evaluate the inductive differentiation specified to one or more germ layers e All Trilineage Evaluate all three germ layers simultaneously e g of EB samples e Multiple Sample Types Allows you the flexibility to mix and match
23. s such as the parental lines used for iPSC generation are not pluripotent in nature and their scores will be low However the expression of lineage markers will largely rely on homogeneity of the cells Note that the markers in the panel are designed to evaluate early germ layer specification and not any particular terminally differentiated state Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 5 ThermoFishe e m SCLENTIFIG TaqMan hPSC Scorecard Panel FAQs Part 2 Sample Preparation 14 How long can I store cells harvested in TRIzol reagent You can store homogenized samples at room temperature for a few hours or at 60 to 70 C for at least one month See http tools invitrogen com content sfs manuals trizol_reagent pdf for more information 15 Can I prepare RNA using the RNA kit of my choice Yes but we recommend using Life Technologies products because the hPSC Scorecard Panel protocol was developed and optimized using Life Technologies reagents and kits 16 Can I skip the RNA purification step and use a lysis protocol instead Yes while RNA purification is recommended it is not strictly required TRizol H RNA Isolation H DNase Treatment H RT H PCR Ceatis to ct ase Treatment H RT H Por eetsDirect DNase Treatment RT POR E H9 TRIzol M H9 CelistoCt
24. sample will be required to compensate for the dead volume 22 How do I load the samples onto the 384 well plates if I only have an 8 channel pipette The tips of most 8 channel pipettes will align with every other well so you will need to pipette twice to load every well in each column of the plate The first set of 8 samples can be loaded in rows A C E G I K M and O The second set of 8 samples can then be loaded into the alternate rows B D F H J L N and P Follow the schematic below 8 Channel Pipette Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 8 Fisher Therm e a SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs 23 How do I load the samples onto the 384 well plates if I only have a 16 channel pipette The tips of most 16 channel pipettes will align with every well in each column of the plate However if your cDNA reactions were set up in 8 wells of a 96 well plate or in 8 well PCR strips additional sample will be required to compensate for the dead volume When you insert 2 tips of the 16 channel pipette into 1 well the tips can t reach the bottom of the well resulting in a need for additional dead volume 16 Channel Pipette Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 9 Fisher Ther
25. tible with the following web browsers Internet Explorer 9 Safari Google Chrome 34 How do I access the hPSC Scorecard Analysis Software Please visit lifetechnologies com scorecarddata for access to the analysis software 85 I can t sign in to my Life Technologies account what should I do Retrieve your username or password using the appropriate links on the sign in page If you don t have a username and password click Register Now in the New Customer section of the log in page 36 Where do I find more information about how to use the software The user manual is available from the help menu within the software and at lifetechnologies com scorecardsoftware 87 How do I create a project and upload my data Click Create a Project to identify its Name Instrument Type Block and Description click OK and then click Upload Data to start using the analysis software Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support 3 ThermoFisher o m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs s8 My data isn t being uploaded what should I do The analysis software validates the data file upon import If your file is not imported or you receive an error click the pencil icon Eto edit the project confirm that you have the correct Instrument Type and Block designated in the p
26. virus SEV to generate iPSCs can I check for the presence of residual virus SEV is included in the panel and can detect the presence of the Sendai virus backbone Note that this method will not distinguish between the different reprogramming factors but just the presence or absence of the residual virus in the cells If an unexpected signal is detected check the amplification curve to determine if it is a false positive If it s a false positive you may ignore the flag Visit thermofisher com scorecard for more product information data protocols and troubleshooting tips or contact stemoell help lifetech com for support ThermoFishe o m SCIENTIFIC TaqMan hPSC Scorecard Panel FAQs 11 Will the presence of Sendai virus affect my results Results indicate that the presence or absence of Sendai virus in established iPSC clones does not have an impact on pluripotency 12 Does the reprogramming method affect the Scorecard results Minor differences in gene expression profiles are sometimes observed based on the reprogramming method but in general the hPSC Scorecard results do not change significantly for lines derived using various reprogramming methods As an example ESC and iPS derived using episomal or Sendai based reprogramming systems were tested before and after seven days of spontaneous differentiation Cells were grown in KSR based media on irradiated MEFs prior to removal of FGF for EB formation o o ojo o o
27. w minimal spontaneous differentiation lt P8 If you use lower passage clones ensure the homogeneity of the culture based on morphology and growth characteristics All values are negative my cells are negative for pluri and lineage Cells may be contaminated with genomic DNA gDNA Although most primers span across introns some of the control genes cross react with gDNA This cross reactivity is factored in during normalization resulting in scores that shift toward the negative We strongly recommend using DNase treatment If you use a method of RNA isolation or sample preparation that is not recommended by the provided protocol it is best to confirm that the sample resulting from the method is of a sufficiently high quality A260 280 and 260 230 values are within range indicating high RNA quality and DNase treatment was not performed Because DNA and RNA each have absorbance at A260 the A260 280 ratio is not a good indicator of the presence of contaminating genomic DNA The ratio instead can be used to identify the absence of contaminants such as protein or organic solvents We recommend that you design your experiment with a no RT control using a recommended test primer ACTB Hs01060665_g1 to test for gDNA contamination In the absence of gDNA contamination the resulting CT value would be close to 40 If gDNA contamination is present you will observe lower CT values Visit