ART-SP1 Scaler /Polisher Combo Unit Instruction

image

Contents

1. Before leaving the operation room make sure that the ART SPI is turned off and the water faucet 1s tightly closed D If you find that the power output to the hand piece 1s insufficient the insert 1s probably worn out Inserts should typically be replaced after six months or when worn out E Please contact a local authorized agent for technical support Section IX Troubleshooting While Bonart offers repair service the following trouble shooting procedures can help you save time 9 A Troubleshooting Guide Basic procedures 1 Check all lines and connections to and from the system A loose plug or connection is the most common cause of problems 2 Check the setting of all control knobs System will not operate LED power indicator is not lit 1 Check that the three prong power plug is fully seated in the AC receptacle on the back of the ART SPI and that the voltage is supported 2 Check that the power plug is plugged into the wall outlet LED power indicator is lit Check that the footswitch connector 1s fully seated in the foot control receptacle on the back 91 of the system System operates No water reaches the tip scaling mode 1 Assure that the water control knob is turned on or adjusted to allow water to flow turning counter clockwise allows for more water 2 Check that the water supply control valve dental office water supply is open 3 Check that the external water tube is properly connected 9 B
2. SP1 299 2
3. WARNING Alerts users of danger that may cause severe injury when proper procedures are not followed e SYMBOLS Aa Compliance with MDD 93 42 EEC 0434 BF equipment Grounding terminal Attention Please read instructions gt OP Warnings Important Only trained and qualified doctors dentist and hygienists should operate the ART SP1 Patients with pacemakers cannot be treated with the ART SPI gt Donotimmerse the ART SPI in water gt If the patient or operator is pregnant consult a physician prior to performing Y dental scaling for safety purposes The ART SPI requires a shielded AC power cord Plug the power cord into a well grounded outlet Y N Attention to users with cardiac pacemakers Patients with pacemakers should avoid treatment with the ART SPI Electronic appliances including razors hair dryers microwave ovens TV receptors and other medical equipment such as the ART SPI may interfere with the performance of pacemakers For more information on this subject please refer to the following articles Advances in Cardiac Pacemaker The New York Academy of Sciences Vol 167 Article 2 pp 515 1075 Electromagnetic Radiation Interference with Cardiac Pacemaker U S Department of Health Education and Welfare The Individual with a Pacemaker in the Dental Environment Journal of the American Dental Association Vol 91 No 6 pp 1224 1229 Please Read Before Oper
4. indicators or chemical indicators for ensuring the efficiency of the sterilization cycle are recommended NOTE Routine spore testing will determine the sterile conditions of your office NOTE The Sterilization Assurance Level SAL of steam autoclave should be 10 According to ISO 13683 Sterilization of health Care products WARNING High room temperature conditions improper dilutions or extended immersion time in chemical sterilizers can result in damage to the plastic and other materials of the insert tips CAUTION The use of a dry heat oven incompatible chemical vapor type sterilizers or quaternary ammonium compounds must be avoided as damage can result to the plastic and elastomeric materials Main Unit To clean the main unit wipe with a nonabrasive disinfectant 230 2 Micro motor Remove all attachments to the micro motor prior to cleaning The outer surface of the micro motor should be cleaned with an antiseptic solution Never oil the micro motor Make sure no moisture gets inside the micro motor Warning Do not dip the micro motor and extension wire directly into the sterilization chemical The sterilization chemical will interfere with normal operations if it reaches the inside of the system Warning Do not leave the disinfectant on parts longer than recommended by the disinfectant manufacturer to avoid damaging materials Note The hand piece should be cleaned after each use between patients Straig
5. straight hand piece there may be a metal rod at the top that serves as a placeholder Remove the rod after you open the lock B Match the groove at the base of the prophy head with the top of the straight hand piece and connect them together C Close the lock by turning the locking ring counter clockwise The prophy head should now be locked onto the straight hand piece and should not fall off during operation D If you wish to remove the prophy head follow step A to release the lock on the prophy head WARNING The prophy head needs lubrication and cleaning on a regular basis When you are ready to begin polishing follow the steps below Place a rubber cup on the prophy head and put paste in the cup to begin polishing Press the Micro motor mode button to switch from scaling to polishing mode Adjust the Oscillating dial to a low speed 2 000 rpm Press on the foot switch to engage the polishing hand piece and begin polishing pa qn De prob Add paste as needed 15 5 J Assembling the Scaler Unit 1 Switch the power of the ART SPI off The power indicator should not be lit if the machine 1s off Plug the power cord of the main unit into a grounded AC power outlet 3 Install an insert into the hand piece pushing it into the hand piece until it 1s firmly seated 4 Check the water tube connection to the water supply source and main unit Section VI General Description and Information of Parts Main unit Water contr
6. water when the office 1s unoccupied 5 C Electrical Requirements Refer to Section X Specifications 25 5 D Unpacking the Unit Carefully unpack your ART SPI unit and verify that all components and accessories are included Figure 1 All components and accessories 6 ART IF 50 Streamline Universal Plastic Handle Internal Flow Insert Optional ART P 10 Straight Angle Universal Metal Handle External Flow Insert Optional NOTE While unpacking check your ART SPI unit for any damages If any damages are found please contact your dealer immediately Enter the unit serial number on your warranty card and mail it within 14 days from the date of purchase a 5 E Power Cord Power Connection e Always make sure that the power switch is set to the OFF position before performing the following tasks o Plug the detachable AC cord into the back of the unit o Plug the 3 prong plug into a grounded outlet N Safety Instructions A Grounding Prior to connecting accessories to the unit check that the main unit is plugged into a grounded wall outlet B Main voltage range and fuse Prior to plugging the AC adapter into the power outlet check that the voltage 1s supported 5 F Foot Control Cable Assembly Connection Align the pins of the foot control plug with the receptacle on the back of the device and push in until the plug is firmly seated 5 G Water Supply Line Connection Push the blue water tube ho
7. will cause damage to teeth enamel For optimum performance we suggest practicing on a model or aluminum plates to become more familiar with the scaling tips before performing on patients Table of Contents COPDES 5 exibuesmetim tani cet m VUE OperatorSalety enedauusesssiesesvuveses desee lve Wammes agase ed aid odinate Pickice OO L C pompe MM EM OO Section I Indications for USE 9 ttis Section II Contraindications and Warnings in cece cece cece eee eeees Section III PrecautionS 2 24 Section IV Infection Control 411144 44440 SECHONEN InstalldtiOf aie vest ia se tea d xeu ede POPE dau rad Section VI General Description and Information of Parts SECHON VIL Techiqu S aysudnbsiicuinidivsbustepemesbstas Section VIII System Maintenance esee Section IX Troubleshooting 2 44 4444 DecHOn X Specifications 1 teste nesese Section XL Disposal eske ESSE Mts Section I Indications for Use Ultrasonic procedures Removal of calculus and plaque during dental prophylaxis General supra and sub gingival scaling applications Periodontal debridement for all types of periodontal diseases Endodontic procedures Polishing procedures Section II Contraindications and Warnings Do not use the ART SPI f
8. ART SPI Scaler Polisher Combo Unit Instruction Manual anal Bonart Co Ltd Jorgen Kruuse A S AF 11 No 3 Wuquan 1st Rd Xinzhuang Havretoiten 4 DK 5550 Langeskov Denmark Dist New Taipei City 24892 Taiwan R O C Tel 886 2 22983980 Fax 886 2 22983981 C 434 Email sales 9bonartmed com 043 COPYRIGHTO 2012 BONART CO LTD BONART Co Ltd owns all rights to this publication This publication shall be used solely as a reference for operation maintenance and repair on BONART equipment No part of this document may be reproduced or distributed in any form for any other purpose In the event of inadvertent or deliberate publication BONART intends to enforce its rights to this manual under the copyright law Reproduction of the materials presented in this manual without the express written permission of BONART Co Ltd is prohibited The contents of this manual are subject to change without prior notice PROPERTY OF BONART CO LTD ALL RIGHTS RESERVED ART is a registered trademark of BONART Co Ltd amp its affiliates Document No RD 02 13 02 Rev 1 0 08 12 Printed in Taiwan R O C Operator Safety Please read this manual before operating unit The ART SPI Scaler Polisher Combo Unit should be operated maintained and repaired by qualified and properly trained personnel 6 NOTE CAUTION AND WARNING STATEMENTS NOTE Provides tips and advice CAUTION Provides correct operating or maintenance procedures
9. Technical Support and Repairs For technical support and assistance call 1 888 526 6278 or 626 600 5330 Monday through Friday 9 00 AM to 5 00 PM Pacific Time For other areas contact your local Bonart representative X Specifications Classification Protection against electric shock Class I Degree of protection against electric shock Type BF Mode of operation Continuous o Medical device directive IIa Standards Followed 18013485 Medical Devices Quality Management Systems Requirements for regulatory purposes 93 42 EEC MDD CE Directive EN60601 1 Medical Electrical Equipment General Requirement for Safety EN60601 1 2 Medical Electrical Equipment Electromagnetic Compatibility Requirements and Tests 20D e Specification e Power supply e Working frequency e Polisher speed e Water supply Dimension Weight Hand piece Cable Footswitch Cable Operation environment e Temperature e Relative Humidity Transport and storage conditions e Temperature e Relative Humidity e Atmospheric pressure Section XI Disposal 115V 45 50 60Hz 160VA 230V 5 50 60Hz 160VA 24 5KHz 5 for 25 KHz Insert 2000 30000 R P M 25 60 PSIG 172 414KPa 24 5cm L x 22 5cm W x 8 5cm H 4 Kg include hand piece 250 cm 250 cm 0 230 C 10 90 non condensing 0 C 60 C 10 90 non condensing 860 1060 hPa Please follow county and state regulations for disposal of the ART
10. ating Make sure the device is connected to a grounded outlet Failure to meet this requirement may severely damage the user and equipment Place the device on a flat and stable surface Placing the device on a tilted or unstable surface may degrade performance and also cause damage to the device Do not modify this device Modifications to the device will invalidate the safety codes and warranty while endangering the patient and operator Do not place heavy objects on top of the power cord and keep the power cord away from intense heat Unplug the device and call Bonart for service and instructions if you observe any abnormalities while operating the device Preface About the ART SPI Scaler Polisher Comb Unit The ART SPI is CE 0434 certified in compliance with the applicable requirements of the Council Directive 93 42 EEC The reviewing council also classifies the ART SPI in accordance with EN60601 1 and EN60601 1 2 safety requirements How it works The ART SPI consists of two parts The main electronic system and the scaling hand piece The main electronic system utilizes two closed loops One loop provides automatic tuning to match the resonance of the insert The other loop will control the tip stroke based on various working conditions The scaling hand piece utilizes a pulsing magnetic field This magnetic field 1s applied to a metal stack which flexes and moves an i
11. hand piece to clear the hand piece of any residue accumulated during treatment Then clean the surface of the hand piece with antiseptic soap or solution and rinse thoroughly with water Afterwards wipe or spray a chemical disinfectant on the hand piece Please try to use a nonabrasive disinfectant to avoid damaging the hand piece After disinfecting the hand piece a sterile insert may be re inserted into the hand piece in preparation for the next patient At the end of the day remove the insert from the hand piece and scrub both the hand piece and cable with antiseptic soap Rinse thoroughly with water NOTE Cleaning the hand piece after every patient is advised WARNING Do not dip the hand piece and extension cable directly into sterilization fluid Doing so will cause the ART SPI to malfunction WARNING Do not leave the disinfectant on the surface of the hand piece and extension cable any longer than instructed by the manufacturer of the disinfectant Insert Saliva blood and other debris will leave a residue on the insert after each use Therefore it is necessary to clean inserts after each use Cleaning can be done manually by scrubbing the insert with a brush water and detergent Be sure to rinse thoroughly after scrubbing Let the insert dry Then place the insert in a bag and autoclave at 275 F 135 C for 15 minutes or as recommended by the manufacturer of your sterilization unit to sterilize the insert Bio
12. ht Hand Piece Attachment The straight hand piece should be lubricated at the end of each day of use Drop 2 3 drops of mineral oil into the top and allow it to flow down through the hand piece Prophy Head The prophy angle should be cleaned to remove hair and debris after each patient At the end of the day it should be cleaned and lubricated To clean the Prophy Head follow the steps below 1 Remove the rubber cup 2 Remove the cup holder with pliers by turning the knurled knob right the knurled knob has a reversed thread Wash away debris from the inside Then put 1 drop of oil onto the gears 3 1n multipurpose oil works well 3 Make sure all hair and fur if any is removed from the end of the prophy angle 4 Replace the knurled knob Section V Installation 5 A General Information If the installation of your ART SPI system is performed by non Bonart distributor personnel check to see that the requirements below are followed 5 B Water Line Requirements e Incoming water supply line pressure to the Scaler must be 25 psi 172 kPa minimum to 60 psi 414Kpa maximum If your dental water supply line pressure is above 60 psi install a water regulator on the water supply line to your ART SPI unit e After the above installation requirement is fulfilled thoroughly flush the water prior to connecting to the scaler 2 g e Please use the manual shut off valve on the dental water system supply line to shut off
13. in unit It is located on the back of the main unit at the bottom left Footswitch Socket The foot switch socket is where you connect the foot switch to the main unit It is located next to the water hose input on the back of the main unit 9 Water hose input connector The water hose input connector is the contact point for the water tube and the main water supply line Use either a male or female quick connect to connect the water tube to the water supply The ART SPI comes with a male quick connect attached to the water tube Power cord socket The power cord socket is where you connect the power cord to the main unit Forward Reverse switch The Forward Reverse switch allows you to control the direction of the micro motor It 1s located on the back of the main unit 18 Section VII Techniques 7 A Performing Ultrasonic Scaling Procedures l Use purified or distilled water to prevent infection when patients experience tissue laceration during treatment Keep the power cord tidy to avoid tripping and other accidents Position the footswitch in an easily accessible spot for the user Keep the footswitch cord tidy to avoid tripping and other accidents Hold the empty hand piece in an up right position when testing to see if the scaler is working properly Tap the footswitch until water appears Do not repeatedly tap or hold the footswitch down The empty hand piece will melt Lubricate the rubber O ring on
14. it A Do not depress the foot switch while turning on the ART SPI Micro motor Mode button The micro motor Mode button is used to switch between polishing and scaling mode When pushed the LED indicator should be lit indicating that the system is operating in polishing mode Under polishing mode the micro motor is engaged To return to scaling mode push the button again Water control knob The water control knob allows you to control the amount of water that flows through the hand piece If turned clockwise the amount of water will decrease If turned counter clockwise the amount of water will increase Please note that water is necessary to cool the tip and prevent overheating so it is necessary to ensure that an adequate amount of water flows through the hand piece while operating the device The larger the water stream the lower the temperature and vice versa WARNING Do not turn the water knob counter clockwise more than three full turns Doing so will damage the internal water line needle WARNING If needed gently adjust the water control knob to prevent any serious damages to the water valve Power control knob The Power Control Knob allows you to control the power intensity When turned clockwise the power intensity increases When turned counter clockwise the power intensity decreases xq us Te 8 10 11 Micro Motor Socket The micro motor socket is where you connect the micro motor to the ma
15. maged bent or reshaped Discard immediately Use the long axis of the scaling tip insert to wipe accretions from the tooth Do not gouge 8 the tooth with the tip 3 C Precautions for hand piece In order for the system s hand piece to function consistently and effectively Accumulated air pressure from the handle or hand piece must be released every 10 to 15 minutes on every procedure by doing the following steps 1 Put the insert into the hand piece and make sure it 1s fully seated 2 Hold the hand piece vertically with the insert tip on top depress the foot Switch control 3 Let the water spray out for a minute or so then start scaling procedure Section IV Infection Control 4 A General Infection Control Recommendations As with all dental procedures use standard protective equipment such as face masks eyewear face shields gloves and protective gowns To ensure safety to the operator and patient carefully follow the Infection Control Procedures detailed in this section 4 B Water Supply Recommendations All dental water supply systems should conform to applicable Centers for Disease Control and Prevention CDC and the American Dental Association ADA standards These standards should apply to flushing chemical flushing and general infection control procedures 4 C Cleaning and Sterilization Hand piece Prior to cleaning remove the insert from the hand piece Next drain any water left in the
16. nstalled scaling tip insert approximately 25 000 cycles per second in an elliptical pattern based on the amount of power used Thus more power results in greater tip displacement which allows for more robust scaling However the vibrating metal stack generates heat and requires adequate amounts of fluid to cool the hand piece and tissues during treatment When the two parts of the ART SPI are placed together the properties of the main unit and scaling hand piece enable the ART SPI to power away the heaviest calculus deposits In addition tips may be rotated 360 degrees within the hand piece to relieve cord tension and alignment Asa result the ART SPI provides maximum comfort to the patient and ease of use to the operator Features e Scaler and polisher in one system e Automatic phase lock and power control e Automatic feedback control e Micro motor speed control of 2 000 to 30 000 RPMS Technical Support For technical support and repairs in the U S call 1 888 526 6278 or 626 600 5330 Monday through Friday 9 00 AM to 5 00 PM Pacific Time For other areas contact your local Bonart representative Supplies amp Replacement Parts To order supplies or replacement parts 1n the U S contact your Bonart Distributor or Dealer A list of Bonart distributors can be found on our website at www bonartmed com CAUTION Since the scaling tip insert is made of stainless steel avoid direct contact with teeth Direct contact
17. ol knob 9 Water hose input Hand piece Power control knob 10 Power cord socket Power button Micro motor socket 11 Forward Reverse switch Micro motor Mode button Foot switch socket Figure 2 ART SP1 diagram 1 Main Unit The main unit generates power and produces a signal that 1s passed to the hand piece The hand piece is then powered up and vibrates the installed insert 2 Hand piece The hand piece is the housing for scaling inserts Inserts are interchangeable and can be installed with a firm push To remove the insert hold the insert at the grip and pull from the hand piece Be careful not to cut yourself in the process Inserts that are compatible with the ART SPI are made by Bonart Dentsply and Hu Friedy 2462 The hand piece will heat up when used due to the power generated by the main unit Therefore it is necessary to cool the hand piece by ensuring that an adequate amount of water flows through the hand piece CAUTION Make sure that new insert is firmly seated CAUTION Do not depress the footswitch when there is no insert installed Doing so will cause the hand piece to melt NOTE The ART SPI is a stand alone unit and is now only available in 25Khz Please do not use 30Khz inserts with the hand piece Power button ON OFF The power button is used to turn the ART SPI main unit on and off When the power 1s on the LED indicator should be lit When turned off the LED power indicator should not be l
18. or amalgam restorative dental procedures Do not use the ART SPI if the patient or operator is wearing a pacemaker Do not immerse the ART SPI in water or liquid If the ART SPI has water damage return the machine to Bonart for servicing Do not modify the ART SPI Modifications will invalidate the warranty on the machine as well as invalidate safety codes and endanger the patient and operator Section III Precautions 3 A Precautions for all Ultrasonic Scaler Units and Systems Ensure sufficient water flow to the scaler tip during use to cool the hand piece and insert Take precaution as all ultrasonic scalers produce aerosols that may transmit contagious diseases Keep the unit away from intense heat Excessive heat may damage the electronic components Switch the water valve off when the ART SPI is not in use Avoid treating patients with pacemakers due to the magnetic field produced by the ART SPI The magnetic field may cause monitoring devices to malfunction and provide inaccurate readings 3 B Precautions for Ultrasonic Prophylaxis Procedures Scaling tips inserts will wear with use Inserts with 2mm of wear will lose approximately 50 of their scaling efficiency In general it is recommended that inserts be discarded and replaced after 2mm of wear to maintain optimal efficiency and avoid breakage Avoid direct contact of inserts with the patient s lips cheeks and tongue Do not reuse scaling tips inserts that are da
19. reating sensitive areas e reat less sensitive area first and then return to sensitive areas e If the problem persists reduce the power output intensity of the hand piece Section VIII System Maintenance Daily start up I Open the water supply valve 2 Turn the ART SPI on using the main power switch Verify that the LED power indicator LED is lit 3 Set the power control knob to full power 4 Turn the water control knob counter clockwise to the maximum setting 5 Hold the empty hand piece upright over a sink or drain 6 Depress the footswitch and flush the water line for at least 2 minutes 7 Place a scaling insert or a compatible insert into the hand piece and adjust the water level as desired NOTE If no water is detected check the water supply and the water control knob NOTE If the power LED is not lit contact a local authorized agent so Preparations for treating each patient Remove clean and sterilize all used ultrasonic inserts 2 Make sure the scaling insert tip has been sterilized before treating the next patient 3 Clean and disinfect the surface of the power cord hand piece and cable assembly Daily shutdown 1 Use the power switch to turn the ART SPI off 2 Close the water supply valve Daily Maintenance The suggestions below will help maximize the lifespan of the scaler A The ART SPI should be used in a well ventilated room B Handle the ART SPI with care when moving it C
20. se into the stainless steel receptacle until the hose cannot be pushed any further Then tighten the screw Connect the quick connect to the water supply line The ART SPI should be prepackaged with a male quick connect If a female quick connect is necessary please contact a Bonart distributor to purchase it Inspect all connections for leaks To remove the water line from the ART SPI scaler first turn off the water supply or disconnect the water supply line Then loosen the screw on the water tube from the receptacle of the unit and gently pull the water hose out 5 H Hand piece Cable Assembly Connection The ART SPI scaling hand piece is designed only for inserts with a frequency of 25khz The ART SPI is no long available in 30khz so please do not try using a 30khz insert with the ART SPI 14 5 I Polishing Hand Piece Assembly and Start Up Prophy Angle Straight Hand Piece Micro motor Knurled Knob Locking Ring To assemble the polishing hand piece connect the Prophy Head Straight Hand Piece and Micro motor by following the steps below 1 Connect the straight hand piece and the micro motor by pushing the straight hand piece over the rod of the micro motor It should snap into place 2 Attach the prophy angle to the straight hand piece by following the steps below A Unlock the prophy head from the straight hand piece by turning the locking ring clockwise This should open the lock Note When you first unpack the
21. the insert with water before placing it into the hand piece Install the insert by firmly pushing until seated Hold the hand piece over a sink or drain with an insert installed Verify that water is reaching the tip Check your scaling inserts for wear and replace as needed Use your foot to press down on the foot switch to allow water to reach the hand piece Release the foot switch to stop the water Control the amount of water that flows through the hand piece by using the water control knob Turn counter clockwise to increase water flow and clockwise to decrease water flow NOTE To lower the temperature of the hand piece increase the water flow CAUTION A continuous flow of water is required to keep the hand piece cool during use 219 7 B Patient Comfort Considerations 1 Position the patient comfortably on the patient chair Adjust the chair angle and position to access the patient s oral cavity with ease and comfort 2 Put the saliva ejector into the patient s mouth to remove excess saliva and debris 3 Rotate the patient s head gently so that the oral cavity can be easily accessed without causing discomfort to the patient see figure below 4 During treatment try to keep the angle between the surface of the tooth and scaling insert at 15 degrees If the patient experiences any discomfort during treatment follow the suggestions below e Increase the hand piece movement speed on the surface when t