1. angle for radiation 1 1 Intended Use The TLC 2H Phototherapy unit is 1ntended for the treatment of infant with hyperbilirubinemia jaundice The unit can be used for infants in a baby cot bassinet bed incubator or under radiant warmer The unit is equipped with castors to provide mobility 1 2 Specifications TLC 2H TLC 2H 1H Power Source Power Consumption Lamps Blue light Intensity Tube life time Hour meter Height adjustment Housing of lamps Mobile stand dimension Weight Castor AC 220V AC 50Hz 72W 4 lamps 90W 5 lamps 108W 6 lamps 4 5 6 blue light Fluorescent tubes 400 500 nm with electric ballast 62uW cm nm 4 Lamps at distance of 30 cm 1500 hours 0 99999 9 hours App 120 165 cm App 65W x 43Hx 13D cm App 65W x 43D x 110 H cm knock down App 11 kg 2 inches TLC 1H 2H Instruction for Use 1 3 Drawing and Components Lamp enclosure Fuse box Power cord attachment Height Adjustment with locking knob Roll Stand A Figure 1 Main components of TLC 2H Difuser Castor two with brake Pl SCR th Lamp enclosure is the housing of the lamps made of ABS material It can be tilted in three different positions horizontal 45 and vertical 90 Hour meter shows the working time of the phototherapy unit from 0 to 9999 hours It helps user to control how long the therapy has been performed and the lifetime of lam
2. meter to control the duration of the therapy Switch on power Warning Never look directly into the phototherapy lamps Make sure that the power indicator is ON as well as the phototherapy lamps The hour meter starts counting Monitor the patient The patient shall be regularly monitored during treatment Measure the patient s bilirubin level periodically Turn off the unit during checking Make sure that the baby s eyes are protected When the therapy is finish turn off the power N Warning always turn off the power switch before remove the power cord TLC 1H 2H Instruction for Use 5 Troubleshooting Guide Warning always disconnect the power cord before opening the phototherapy unit for repair The unit does not turn on No power Verify that the power cord in The power indicator is off plugged in Defective fuse Check the fuse Repeated defective fuse Defective circuitry Contact Tesena technical service or authorized distributor Not all lamps are on Loosely lamp starter Verify that the lamps are installed appropriately Defective lamp Change the defective lamp The unit will not move Casters are locked Unlock the casters 6 Maintenance Cleaning Warning always disconnect the power cord before cleaning Clean the unit as well as the lamp enclosure with soft cloth dampened with water to remove dust and debris Use commercial cleaner or hospital disinfectant if water alone is ineffective Checking the radiation sou
3. TLC 2H INFANT PHOTOTHERAPY UNIT Instruction for Use Tesena is a registered trademark of PT Tesena Inovindo O 2008 PT Tesena Inovindo All rights reserved TLC 2H Instruction for Use Rev 01 PT Tesena Inovindo Jl Jusin 43 Susukan Ciracas Jakarta Timur 13750 Indonesia Telephone 62 21 8412414 Fax 62 21 87792140 Email support Q tesena co id WWW tesena co id TLC 1H 2H Instruction for Use TLC 1H 2H Instruction for Use Contents 1 Product Description 55 seis cei avo uve scl ei 4 M G 4 Speci DCA ONS cisci eee vue ETE PURI 4 Drawing and components eee eee eee een 5 2 Safety Information Pm 6 3 Assembly Inst FUctIOD See RS e PESE KM UPPER ESO TEE OAM PEE SU MM sens 7 4 Operating InstFUuCcl0l oos xo e vus o UE NUT 8 Se Troubleshooting c ooo ete ea eas ern hernede a aa 9 Gx Maintendlioe T DAN IU UE AUTOR VIEN SEER 9 Te TOD VICE sais r cT 10 1 Product Description TLC 1H 2H Instruction for Use The Tesena TLC 2H is an overhead phototherapy unit with roll stand The unit contains four fluorescent lamps emitting blue light with certain wavelength useful for jaundice therapy The lamp s enclosure height can be adjusted vertically along the roll stand rod to have the effective distance to the patient It can also be tilted from horizontal position to have the effective
4. o the lamps Always protect the baby s eyes with the protecting shields during treatment The patient adjacent to the phototherapy unit may also need to be protected N Warning It is recommended to monitor the baby s skin temperature during treatment Incorrect use of phototherapy unit may cause hazardous body temperatures 25 Warning Maintain the distance between lamps diffuser and the patient not closer that 30 cm N Warning If the baby lies in a thermotherapy device incubator radiant warmer or heated mattress the use of phototherapy unit could impact the heat supply of those thermotherapy device N Warning Always switch off the power after finishing the therapy Switch off the power and disconnect the power cord during cleaning and maintenance 3 TLC 1H 2H Instruction for Use Assembly Instructions The phototherapy unit 1s shipped in five separated components head part lamps compartment foot part outer and inner rods and castors It is recommended to utilize appropriate tolls to install the unit Assembly the unit with the following steps see figure 2 Unpack the shipping boxes carefully and check whether the component is complete 2 Insert the inner rod into the outer rod Tighten the locking part with the provided screw 3 Install castors to foot part Press and rotate each castor counter clockwise to fasten it to the foot part Tighten with appropriate tools Castors with break are installed in diagonal
5. position 4 Install rods into foot part Fix it with the provided screw 5 Install the head part to stand Fix it with the provided screw o gt Figure 2 Assembly process TLC 1H 2H Instruction for Use Operating Instructions AN Warning This is a medical equipment Only qualified personnel may operate this unit People without authority may not operate this unit N Warning Please read all the precautions and the operating instructions below before operating the unit N Warning The phototherapy unit could influence the body temperature of the patient It is necessary to monitor the patient s temperature during treatment Incorrect use of phototherapy unit may cause hazardous body temperatures Operating instructions l Check the total irradiance Check the total irradiance for bilirubin of the lamps before using the unit This phototherapy unit provides an average irradiance of 405 uW cm at 40 cm distance Prepare the infant Warning Protect infant eyes with the eyes protecting shields appropriate for the phototherapy treatment Infant may lie in a baby bassinet baby cot incubator or under infant radiant warmer Free patient s skin from any cover to maximize the phototherapy effect Position the unit over infant Adjust the height of the unit The distance between patient and lamp s enclosure shall not closer than 30 cm Check the hour meter Before use note the working time shown by the hour
6. ps Height adjustment the height of the roll stand is adjustable by free the locking knob Diffuser The diffuser provides more uniform illumination to the effective area of treatment 2 TLC 1H 2H Instruction for Use Safety information Please read all sections of this manual carefully before using the phototherapy unit Follow all precautions to ensure the safety of the patient user and those near to the instrument IN Warning This 15 a medical equipment Only qualified personnel may operate this unit People without authority may not operate this unit N Warning Incorrect use of this phototherapy unit or the use of parts that are not manufactured supplied by PT Tesena Inovindo can damage the unit and may cause hazardous situation to patient and or user Use only lamps which are recommended by the manufacturer N Warning Do not use the unit if the unit or its part 1s defect or suspicious as defect Contact the manufacturer or distribution for technical support N Warning Operate this unit only in the treatment room The light radiation may affect other people near to the equipment The presence of other radiation sources may adversely affect the patient N Warning Do not use the unit in combination with other equipment Incompatibility with other equipment could lead to incorrect function of the unit Do not put things on the lamps discloser This unit 1s not designed to endure load N Warning Do not look directly int
7. rce The lifetime of the radiation source blue light fluorescent lamp is limited It is necessary to periodically check their total irradiance for bilirubin The lamp shall be changed after a decrease of 25 of its total irradiance for bilirubin All lamps have to be changed when the expected lifetime has been exceeded N Warning Use only lamps which are recommended by the manufacturer The use of other lamps which are not approved by the manufacturer can influence the safety and effectivity of the phototherapy N Warning only qualified personnel should perform this service TLC 1H 2H Instruction for Use 7 Service For product service and spare parts please contact your distributor or directly contact us at PT Tesena Inovindo Jusin 43 Susukan Ciracas Jakarta Timur 13750 Indonesia Telephone 62 21 8412414 Fax 62 21 87792140 Email support tesena co id WWW tesena co i1d 10

Download Pdf Manuals

image

Related Search

Related Contents

GTX/LTS2000 Handheld Radio Service Manual Part 2 of 3  MANUAL DE USO E INSTALACIÓN  Fuji Frenic Mini Compact Drives Instruction  取扱説明書(PDF)  GammaRAE II R Manual  Manual De Informes de Profesores De Posgrado (Edición: Profesores)    Toshiba Tecra A8 Laptop User Manual  document étudié n°1 - Archives départementales de la Côte-d`Or  