Equinox II Manual Rev 9 - O-Two Systems International, Inc.
Contents
1. C Anti entrainment Ring Fig 1 CLEANING DISINFECTION PROCEDURE 1 Ensure that the cylinder is turned off completely clockwise and bleed any remaining gas out of the gas supply hose 2 Remove the facemask from the Equinox II Demand Valve and then disconnect the demand valve from the regulator The regulator can be cleaned with a clean soft cloth 3 Unscrew patient valve A from the unit body being careful to ensure that the diaphragm B is retained Fig1 4 Remove the diaphragm and anti entrainment ring C from the patient valve swivel housing and wash these components thoroughly in a mild soap solution 5 Disinfect the face mask patient valve and diaphragm using commonly available cold chemical disinfecting solutions Rinse thoroughly with water after disinfection 6 Sterilization if required may be achieved by other methods compatible with the materials of manufacture of these components including Ethylene Oxide Gas and Cidex 7 The demand valve body D and gas supply hose can be wiped over with a soft cloth and mild soap solution and may be disinfected using a hard surface disinfectant that is compatible with aluminium 8 Dry all components thoroughly 9 Re assemble and check performance WARNING Do not immerse the demand valve unit or the pressure regulator into cleaning solution or disinfectant Discontinue to use and discard the face mask the patient valve and the diaphragm should disc
2. 77 2035 Website www otwo com E Mail resuscitation otwo com For your nearest Authorized O Two Medical Tech Distributor In North America Toll Free 1 800 387 3405
3. AINTENANCE WARNING The Equinox II Administration System is designed to provide years of reliable service in all emergency situations Failure to follow the maintenance and inspection routines properly could result in incorrect operation of the demand valve To ensure proper operation of the Equinox II Administration System regular inspection and checking of the demand valve and accessories for correct function should be undertaken by a responsible member of staff on a regular basis This check is to ensure that all of the accessories and demand valve components are present the gas mixture cylinder is full and that the demand valve is in working order MAINTENANCE Function parameters should be checked at least every six months and more frequently in high use applications Units with test pressures outside of the ranges listed in the product specifications should not be used O Two Medical products are not designed for field disassembly or service outside that indicated in this manual Any malfunctioning units should be returned to the manufacturer or an Authorized O Two Medical Technologies Dealer Unauthorized repairs will nullify the product warranty 3 2 CLEANING AND DISINFECTION Routine cleaning and or disinfection of the O Two Medical Equinox II Administration System should be undertaken after each patient use to maintain the equipment in a clean condition D Demand Valve Body B Diaphragm A Patient Valve
4. E TO SALE BY OR ON THE ORDER OF A LICENSED PRACTITIONER 2 WHEN THE UNIT IS IN USE DO NOT SMOKE OR USE NEAR OPEN FLAME EITHER DURING USE OR WHEN CHANGING THE CYLINDER 3 WHEN NOT IN USE ALWAYS TURN OFF THE CYLINDER 4 NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY PART OF THE CYLINDER REGULATOR OR DEMAND VALVE 5 DO NOT DISASSEMBLE ANY PART OF THE DEMAND VALVE EXCEPT WHERE DESCRIBED IN THIS MANUAL AS ANY UNAUTHORIZED DISASSEMBLY WILL INVALIDATE THE WARRANTY 6 AFTER USE ALWAYS ENSURE THAT ALL COMPONENTS ARE CLEANED IN ACCORDANCE WITH THE INSTRUCTIONS PROVIDED IN THIS MANUAL 7 ALWAYS USE THE CHECK LIST TO ENSURE THAT ALL COMPONENTS ARE REASSEMBLED CORRECTLY AND READY FOR USE 8 AFTER USE ALWAYS ENSURE THAT A FULL GAS CYLINDER IS ATTACHED BEFORE RETURNING THE UNIT TO ITS NORMAL STORAGE POSITION 1 4 PERFORMANCE SPECIFICATION Demand Valve Minimum Peak Flow rate ccccccesesseeseseeseneees 140 LPM Exhalation Resistance 0 0 to 6 cmH20 60 LPM Inhalation Resistance c0ccce 0 to 6 cmH 0 60 LPM Input PreSSUre cee ceceecseeeceeeeeteeeeeeetaeeeaeaeeeeeeeess 50 PSI Initiation PreSSUre ccccccccccsccccteceeeeeeeeees 2 to 2 5 cmH O Operating Temp 18 C to 50 C 0 F to 122 F Storage Temp 40 C to 60 C 40 F to 140 F Input CONNECION ccccccccceeeeeesesseeeeeeeeeeeaees 5 8 18 TPI Patient COMM CtOM cccccceecccceeceeeseeeeeeeene
5. O O Equinox II NITROUS OXIDE OXYGEN 50 50 ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E USER MANUAL Made in Canada by O Two Medical Technologies Inc Part Number 17MP1342 Rev 9 Apr 2015 CHAPTER 1 1 1 2 1 3 1 4 2 1 2 2 2 3 3 1 3 2 TABLE OF CONTENTS TITLE PAGE INTRODUCTION Equinox II Administration System 2 Warranty Information 2 Safety Precautions 3 Performance Specification 4 OPERATING PROCEDURE Demand Breathing 5 Operating Instructions 5 Scavenging of Waste Nitrous Oxide 6 Oxygen Using the Scavenger Adapter SERVICING Routine Maintenance Cleaning and Disinfection REPLACEMENT PARTS ACCESSORIES 10 1 INTRODUCTION 1 1 Equinox II ADMINISTRATION SYSTEM The Equinox II Administration System is intended for administration of 50 50 N20 O2 analgesic gas mixture and is specifically designed for use in emergency and obstetric analgesia The Equinox II Administration System consists of a demand valve a single stage pressure regulator a supply hose and a universal patient face mask It can be operated anywhere 50 Nitrous Oxide 50 Oxygen mixture cylinders are available The Equinox II Demand Valve is lightweight portable and extremely durable and is intended to administer 50 50 N20 02 analgesic gas on patient s demand The pressure regulator is intended to lower the high variable cylinder pressure to a lower constant working pressure for the Demand Valve Not
6. e Only those individuals trained in the administration of Nitrous Oxide Oxygen mixtures should use this equipment Thoroughly review the instruction manual before use 1 2 WARRANTY Equinox II Administration System is manufactured from the finest quality i materials Each individual part is subject to strict quality control tests to ensure 4 exceptionally high standards The manufacturer warrants to the purchaser of i Equinox II Administration System that its component parts are free from 4 defects in material and workmanship for a period of two years from the date of purchase The manufacturer will replace and or repair all parts of the device at its option for two years from the date of purchase at no cost to the purchaser upon the notification of the defects in writing by the purchaser All shipping i costs shall be borne by the purchaser The manufacturer shall be liable under 4 this warranty only if the device and its parts have been used and serviced in the a1 normal manner described in the instruction manual There are no other 4 expressed or implied warranties This warranty gives no specific legal rights You may also have other rights which may vary according to local regulations 1 3 SAFETY PRECAUTIONS The Equinox II Administration System is intended for use by suitably trained and qualified personnel The following precautions should always be observed 1 CAUTION FEDERAL LAW IN THE U S A RESTRICTS THIS DEVIC
7. etaees 15 22 mm Weight an i ania iai aani 210 grams 7 4 oz Dimensions inches Dia 51mmx81mm 2 0 x3 2 Pressure Regulator Inlet Connection CGA 965 N2O0 O2 Pin Indexed Yoke Outlet CONNECCOM ccccceecceessesssseeeeeneeeesesneaenens 5 8 Thread Operating Pressure eseseeseeeeeereeserrreeerreeeee 500 3000 PSI Q tp t PRESSURE sirenen a duaaareanenae 50 60 PSI Weight eren e a A 450 grams 1 0 Ib 4 2 OPERATING PROCEDURE 2 1 DEMAND BREATHING The O Two Medical Equinox II Demand Valve is designed to operate on compressed Nitrous Oxide Oxygen gas It is intended to provide 50 50 Nitrous Oxide Oxygen mixture to the patient at a safe delivery pressure and adequate flow rate upon demand An inspiratory effort by the patient will activate the demand valve and Nitrous Oxide Oxygen will flow to the patient at a rate and volume in line with their inspiratory effort 2 2 OPERATING INSTRUCTIONS 1 Ensure that the Equinox II Demand _ Valve Regulator connection is secured and no leaks present Turn the cylinder hand wheel counter clockwise to the ON position at least one full turn Ensure that the regulator contents gauge reads at least 500 psi If not replace with a full cylinder Attach the Universal Facemask to the Equinox II Demand Valve outlet Instruct the patient in its operation Most protocols require the gas mixture be self administered by the pa
8. oloration or deterioration be observed 4 REPLACEMENT PARTS ACCESSORIES 01EQ7001E Equinox II Demand Valve 01RE05S10 B Equinox Pressure Regulator 02FM4999 Cs Universal Face Mask case 12 17MP1329 Equinox II 7 Supply Hose 01EQ7006 Equinox II Demand Valve Bleed Pin with Chain 17MP1528 Silicone Diaphragm 17MP1320 Equinox Supply Fitting assembly 17MP9039 Non rebreathing Valve Assembly Other sizes and seal types are available Contact your nearest O Two Medical authorized distributor for a complete catalogue on all O Two Medical manufactured products 10 MANUFACTURERS AND DISTRIBUTORS OF CPR Bio Barrier Face Shields Universal Pocket Ventilators Rescue Breather CPR Protective Barriers Demand Valve Resuscitators CAREvent Automatic Transport Ventilators O Two eSeries Electronic Automatic Transport Ventilators O Two Pressure Regulators for Use with Medical Gases SMART Bag Bag Valve Mask Resuscitators EasyGrip Bag Valve Mask Resuscitators Burn Relief Burn Kits and Dressings Equinox II Nitrous Oxide Oxygen Administration Systems Equinox Relieve and Equinox Advantage Nitrous Oxide Oxygen Gas Mixing and Delivery System Your Representative is O TWO MEDICAL TECHNOLOGIES INC Innovation in Resuscitation C E 0120 EU Representative Marcel Houben 7575 Kimbel Street Rue Vin ve 32 ae x 4030 LIEGE Mississauga Ontario Belgique Canada L5S 1C8 Tel 905 677 9410 Fax 905 6
9. tient to prevent inadvertent excessive administration by an attendant To trigger the demand valve and obtain an adequate flow of gas it is necessary to make a good mask to face seal Simply place the mask on the patient s face covering both the nose and mouth Instruct the patient to hold the mask on his her face with a light pressure and breath normally The sound of gas flowing should be heard with each inspiration If not adjust the patient s hold on the mask for a better seal and ask the patient to breath more deeply 5 7 Continue administration as per local management protocol 8 Ensure that the cylinder is turned OFF after each use and remaining gas released from the system by placing the bleed pin attached to the regulator into the center hole on top of the Equinox II Demand Valve and depress until gas flow ceases 2 3 SCAVENGING OF WASTE NITROUS OXIDE OXYGEN USING THE SCAVENGER ADAPTER The optional Scavenger Adapter is designed to replace the Non rebreathing Patient Valve assembly to scavenge exhaled air from the patient during inhalation analgesia The 19 mm exhalation port meets the requirements of the International Standards for Conical Connectors for Low Pressure Medical Gas Systems and will allow the attachment of corrugated tubing so that the patient s exhaled breath can be scavenged from the exhalation valve to the exhaust method chosen by the facility using the device 3 SERVICING 3 1 ROUTINE M