Annex 1 - Detailed Technical specifications

image

Contents

1. High temperature 2 minutes Procedural silence Presilences baby temp alarm for 5 minutes Alerts Manual mode System alerts every 30 seconds gt 10 minutes for 15 minutes Apgar timer Alerts at 1 5 and 10 minutes f Physical properties Total height maximum 200 cm Mattress height at approximate 100 cm Weight maximum 100 Kg excluding accessories Power requirements 220 240 V 50 60 Hz 5 4 Warranty and Service 24 months from the moment of installation mie Conditions Maximum response time at beneficiary site 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language ini 2 technical personnel on si neficiar 5 5 Incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 6 SYRINGE INFUSION PUMP 4 6 1 Standard requirements CE Mark ISO 9001 2000 certified companies 6 2 Basic Structure Heated respiratory humidifier Humidifier pole with mobile stand and pole mount bracket 6 3 Technical Characteristics Syringe type infusion pump for constant drug administration Flow rate min range 0 1 ml 999 ml h max 0 1 ml increments Possibility of changing the flow rate whilst infusing Syringe loading sensor Th
2. ITEM 8 PORTABLE PULSE OXYMETER 2 Installation and commissioning User manual in English or translated into Russian language Item Quantity ITEM 9 SUCTION UNIT 3 9 1 Standard requirements CE Mark ISO 9001 2000 certified companies 9 2 Basic Structure Monitor for SpO2 and pulse rate Portable pulse oxymeter Universal connection for sensors 9 3 Technical Characteristics Portable suction unit Suitable for adult and pediatric use Vacuum range between around 20 and 300 mmHg adjustable from a button Vacuum indicator Fast aspiration at around 10 30 lpm flow Sealed water resistant unit Control panel tilted for easy viewing of the controls Main body made of durable and resistant material LED indication for bad battery Indication for battery charge status Remote internal AC charger for battery Charging circuit capable of running on low battery Power supply 220VAC 50Hz Rechargeable battery 12 volt DC rechargeable battery Battery rechargeable Long operating time at full vacuum uninterrupted min 40 60 min Low charge time max 6 8 hours Fast charging runs at least 1 hour after it is charged for 2 hours only 9 4 Accessories and consumables Reusable collecting container min 1500 ml Disposable hydrophobic bacteria filters 3pcs 2 auto clavable silicon suction tubes Carrying case
3. 6 5 Warranty and Service 24 months from the moment of installation ai Conditions Maximum response time at beneficiary site 48 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language ining for at 2 technical personnel on si neficiar 6 6 incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Equipment assembly and installation on site User manual in English or translated into Russian language ITEM 7 PHOTOTHERAPY LAMP S A peer i CE Mark sie otandara requirements SO 9001 2000 certified companies j Phototherapy Unit 7 2 Basic Structure Mobile Stand a Phototherapy Unit composed of minimum 4 fluorescent tubes with white or blue color therapeutically wave length 420 480 nm tube average life time minimum 1 500 hours indications when and what tubes need changing Technical light intensity minimum 40 uW cm2 nm at a distance of 40 7 3 Characteristics cm from the lamp Supplementary 1 set with white or blue color reserve b Mobile Stand height adjustable between minimum 45 150 cm on 3 4 castors with locks 7 4 Consumables Blue white light fluorescent tube Warranty and Service 24 months from the moment of installation 7 5 oe z z Conditions Maximum response time at beneficiary site 72 hours Training for at least 2 medi
4. maximum 200 cm Mattress height at approximate 100 cm Weight maximum 100 Kg excluding accessories Power requirements 220 240 V 50 60 Hz 20 3 Accessories Front drawers for storage X Ray tray Rails for accessories Patient skin probe 20 4 Warranty and Service Conditions Period 24 months from the moment of installation of the equipment Maximum response time 72 hours 20 5 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 21 MEDICATION INSTRUMENT TROLLEY 2 21 1 Standard requirements CE Mark ISO 9001 2000 certified companies 21 2 Basic Structure Stainless steel emergency trolley Item Quantity ITEM 21 MEDICATION INSTRUMENT TROLLEY 2 Work top Double hook IV pole Defibrillator platform with straps Antistatic ball bearing castors System waste container Rubber bumpers Minimum 2 two drawers Stainless steel structure Maximum height 1600 mm 21 3 Physical properties Maximum length 700 mm Maximum width 650 mm Castors diameter min 120 mm Technical Wor
5. Russian language Item Quantity ITEM 4 BASIC INCUBATOR 1 4 1 Standard requirements CE Mark ISO 9001 2000 certified companies 4 2 Basic Structure Incubator hood Temperature controller Standard humidity module Mobile stand a Incubator hood and mattress Double wall and other parts of incubator easily removable for cleaning At least two opposite doors for quick access Small range temperature drop with door down Front door access One iris port on each lateral side two irises on front and back or doors for access Hood can be raised separately from front door Mattress with minimum surface of 2 000 cm2 Tilting of the mattress Tubing access minimum 4 b Temperature controller Microprocessor controlled Item Quantity ITEM 4 BASIC INCUBATOR 1 4 3 Characteristics Automatic test on start up Heat up time to 39 C to be 30 minute or less Removable control unit for easy service Air flow pattern directed from front to rear of incubator across entire width Centralized display for patient air temperatures control with alarms display for function settings parameters and trending easy to operate and clean Accurate air temperature control using two air probes Air temperature range between 20 and 39 C Baby temperature control using reusable skin probe Baby sk
6. 13 OXYGEN HOOD 2 13 1 Standard requirements CE Mark ISO 9001 2000 certified companies 13 2 13 3 Basic Structure Technical Characteristics Transparent oxygen hood for neonates with head waist aperture 2 two sliding doors Silicone flap in the head waist aperture Suitable for neonates and infants Fits into all incubators and open care beds tables Allows consistent and even oxygen flow to the baby Manufactured from food grade polycarbonate Easy access on each side for I V lines oxygen analyzers or ECG wires Big clear sliding doors Raked head aperture Soft material in the aperture for maintaining oxygen and humidity Interior gas deflection system for preventing baby s cooling Item Quantity ITEM 13 OXYGEN HOOD 2 Possibility to measure the interior temperature 13 4 13 5 Warranty and Service Conditions Incidental Services Period 24 months from the moment of installation of the equipment Maximum response time 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 14 RESUSCITATION BALOON 2 14 1 Technical Characteristi
7. Annex 1 Technical Specifications of required equipment for Rybnitsa Maternity Unit and Intensive Therapy Care Unit Item _ i Quantity ITEM 1 C P A P SYSTEM 1 1 1 Standard requirements CE Mark ISO 9001 2000 certified companies 1 2 Basic Structure Heated respiratory humidifier Humidifier pole with mobile stand and pole mount bracket 1 3 Characteristics Heated respiratory humidifier Microprocessor controlled Dual servo control Continuous display of saturated gas temperature Temperature probe Water out alarm Heated wire adaptor for single patient heated breathing circuits for newborns Single use delivery system Input flow range 5 15 liters minute Humidification chamber compressible volume min 250 ml Pressure manifold with different ports oxygen analyzer port pressure port inlet and outlet connectors CPAP generator mean pressure 5 10 cm H20 CPAP generator water container volume 300mI 500 ml Single patient interface Nasal tubing with collapsible extension Nasal tubing flow resistance 0 50 cm H20 at 6 l min 0 55 cm H20 at 6 l min Latex free nasal prongs with multiple sizes for perfect fit to different sizes of babies Comfortable fit single use nasal prongs Multiple sizes top opened bonnets for perfect fit to different sizes of babies 1 4 Consumables Nasal CPA
8. Cuff set reusable all dimensions newborn pediatric and adult 1set SpO gt multisensor sensor pediatric adult with cable with approximately 5 foots 2 pcs SpO gt sensor newborn with cable with approximately 8 foots consumables for 500 tests Item Quantity ITEM 3 INTENSIVE CARE MONITOR 2 ECG electrodes newborn infant 200 pcs ECG electrodes pediatric adult 500 pcs Blood pressure cable 3 ft 2 pcs Temperature cable 3ft 2 pcs Temperature probe covers 500 pcs Blood pressure cuffs Sizes 1 2 3 4 5 6 pcs each Power requirements 220 VAC 50 Hz Autonomy with battery at least 2 hours Automatically display of battery status fuel gage while operating on battery power Automatic internal recharge when the monitor is plugged into an AC power supply Low battery audio and visual alarm Capable of recording all numeric and waveform information displayed on the screen for min 24 hours 3 4 Warranty and Service Conditions 24 months from the moment of installation Maximum response time at beneficiary site 72 hours 3 5 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation and commissioning User manual in English or translated into
9. P kit a Single use delivery system composed from single use patient humidifier chamber single use patient single heated breathing circuits for newborns single use patient pressure manifold single use patient CPAP generator with pressure setting ITEM 1 C P A P SYSTEM 1 b Single use patient interface composed from nasal Tubing nasal prongs fixation bonnets Starter kit 25 pcs complete single use nasal CPAP kits each kit consisting of delivery system 1 pc nasal tubing 1 pc Adifferent sizes of bonnets 2 pc each 8 different sizes of nasal prongs 3 pc each Warranty and Service 24 mantis from the moment of installation ls Conditions Maximum response time at beneficiary site 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language P ini 2 technical personnel on si neficiar 1 6 Incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language ITEM 2 COMPRESSOR FOR VENTILATOR CPAP 1 CE Mark ISO 9001 2000 certified companies Compatible with the ventilator or CPAP system specifications Stand alone medical air compressor Snap fit with the ventilator and CPAP system module to provide 2 1 Standard requirements med
10. VF Optional V1 V2 V3 V4 V5 V6 Possibility to adjust the amplitude of the signal Possibility to choose the lead to be displayed Cable with 3 wires l Il Ill aVL aVR aVF Cable with 10 wires Il Ill aVL aVR aVF Optional V1 V2 V3 V4 V5 V6 Respiratory rate Limits low limit at least 0 breaths per minute high limit at least 150 breaths per minute Oxygen saturation of hemoglobin O saturation low limit at least 30 high limit at least 100 O saturation low limit at least 30 high limit at least 100 Pulse rate low limit at least 30 bpm beats per minute to high limit at least 250 bpm Adjustable averaging time Signal quality meter on screen Blood pressure Oscilometry measurement Measures and displays systolic and diastolic blood pressures Automatically adjustable intervals of approximately 1 min 8hours Discriminates between pressure signals and patient movement Display current and previous pressure values systolic diastolic Automatically zeroes prior to each reading Blood pressure cable and cuffs for all age mentioned above Temperature Measurement low limit at least 28 C high limit at least 42 C Cable for temperature monitoring to be connected with probe and temperature probe or probe incorporated in the cable assembly Mandatory rectal and skin probe Extra battery 1 piece ECG cables with 3 leads 2 pieces
11. anslated into Russian language ITEM 16 BILIRUBINOMETE CUTANEOS 16 1 16 2 Standard requirements Technical Characteristics CE Mark ISO 9001 2000 certified companies Non invasive fast measurement Measurement position forehead or sternum Easy to read display Specify type and range Printer preferably integrated and paper rolls Alarms Measurements error Low battery Others specify Item Quantity ITEM 16 BILIRUBINOMETE CUTANEOS 3 Technical Requirements Power Requirements VAC 220 50Hz Plug Type F Battery rechargeable and mains operation Charging device 220V Carrying case Consumables for 200 patients 16 3 Warranty and Service Conditions Period 24 months from the moment of installation of the equipment Maximum response time 72 hours 16 4 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 17 NEBULIZER FOR DRUGS 3 17 1 17 2 Standard requirements Technical Characteristics CE Mark ISO 9001 2000 certified companies Equipment for broncho pulmonary pathologies asthma pneumopath
12. artments a Heat source controller and functions Mean irradiance at mattress level minimum 10 mW cm2 The central Control panel will display all instructions in English or Russian language Control panel located centrally at eye level Microprocessor controller with control modes Manual mode in minimum 5 increments from 0 to 100 Automatic pre warm mode Servo baby mode Servo control temperature range between approximate 35 37 C Display temperature range between 30 and 42 C Examination lamp of minimum 50 W and minimum 1 000 Lux Item Quantity ITEM 20 RADIANT TABLE FOR NEWBORN 1 The heat source head with possibility of moving left or right 90 Patient skin probe reusable 2 pieces b Bassinet and mattress Mattress area minimum 3 000 cm2 Hydraulic tilt mechanism for the bassinet at least 10 minimum Mattress thickness minimum 2 cm Ray cassette tray to fit under mattress 1 piece Four foldable bassinet walls c Mounting column Fixed height column on mobile castors Built in rails for accessories 2 pieces d Storage compartments Front storage drawers 2 pieces e Alarms Progressive audio alarms Check patient Baby temperature deviation 1 C High temperature with heater shut off Probe fail Power fail System fail f Physical properties Total height
13. cal personnel on site at beneficiary location in Russian language 7 6 incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language ITEM 8 PORTABLE PULSE OXYMETER eS ee 8 1 Standard requirements CE Mark ISO 9001 2000 certified companies 8 2 Basic Structure Monitor for SpO2 and pulse rate Portable pulse oxymeter Universal connection for sensors Item Quantity ITEM 8 PORTABLE PULSE OXYMETER 2 8 3 Technical Characteristics a Monitor Removable handheld display for SpO2 and pulse rate ensures continuous monitoring during transport Oximetry technology clinically proven to work under motion and poor perfusion conditions Patients perfusion level display Sensor signal quality indicator to ensure best sensor position placement Full patient alarms including high and low saturation and pulse rate Downloadable 72 hour memory at 2 seconds resolution with the ability to review trend data on screen Fast time to initial reading after sensor placement Probe off detection system to alert user of probe displacement High sensitivity mode for improved performance on poorly perfused patients LCD waveform display User selectable averaging time SpO2 range 1 100 Pulse rate a
14. cs Silicone resuscitation balloons for providing fast simple and effective ventilation to non breathing patients self inflating bags in three sizes adult pediatric neonatal System assists the rescuer through Pop off pressure release valve which opens at pressures more than 20 40 cm H20 Pressure manometer Valve system respond to the rescuer and the patient Visual alert of the pressure applied or for incorrect operation Audible overpressure alert Auto clavable Oxygen reservoir system for each of the resuscitators Inspiratory resistance 3 3 5 cm H20 Bag volume between 500 750 ml Expiratory resistance 2 2 5 cm H20 Dead space max 8 ml Reservoir volume 500 750 ML Manual flow rate min 35 45 l min Demand flow rate min 0 150 l min Input pressure 50 PSI Pressure relief valve 40 cm H20 Patient connection 15mm newborns and 22 mm Low weight for easy operation Operating temperature range min 10 C to 50 C Storage temperature range min 40 C to 60 C 14 2 14 3 14 4 Accessories Warranty and Service Conditions Incidental Services Neonatal use masks 6 pcs 2 of each size Children use masks 6 pcs 2 of each size Adult use masks 6 pcs 2 of each size 24 months from the moment of installation Maximum response time at beneficiary site 48 hours Training for at least 2 medical personnel on si
15. e at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation and commissioning User manual in English or translated into Russian language Item Quantity ITEM 5 TABLE FOR NEWBORN RESUSCITATION l 1 5 1 Standard requirements CE Mark ISO 9001 2000 certified companies 5 2 5 3 Basic Structure Accessories Height adjustable table Resuscitation Radiant Warmer consisting of Heat source Bassinet and mattress Mounting column with fixed height System for resuscitation Storage compartments Front drawers for storage X Ray tray Rails for accessories Patient skin probe a Heat source controller and functions Mean irradiance at mattress level minimum 10 mW cm2 The central Control panel will display all instructions in English or Romanian language Control panel located centrally at eye level Microprocessor controller with control modes Manual mode in minimum 5 increments from 0 to 100 Automatic pre warm mode Servo baby mode Servo control temperature range between approximate 35 37 C Display temperature range between 30 and 42 C Examination lamp of minimum 50 W and minimum 1 000 Lux Apgar timer integrated The heat source head with possibility of moving left or right 90 Pat
16. e equipment will include KVO function High accuracy over the entire delivery range min 5 Accept all standard type syringe from 50 60 ml 30 35 or 20 ml 10 ml or 5ml Set up fast and simple Bolus mode programmable Antibolus system Warns of pressure variation when there is a risk of occlusion or a possible leak in the infusion line Occlusion in preselected mode three pre selectable pressure alarm limits occlusion Infusion continuity protection Automatic internal battery operation during patient transfer AC power failure Infusion data memorization flow rate bolus rate volume volume limit KVO rate Event logging Electronic pressure management Infusion alarm Pre alarm end of infusion Pre alarm volume limit Occlusion alarm Alarm for the end of infusion Alarm for volume limit Technical alarms Disengaged driving mechanism alarm Low battery pre alarm Discharged battery alarm Battery capacity display Unconfirmed programming Main malfunction alarms Syringe position control Syringe barrel clasp check Plunger head detection Occlusion pressure Locking syringe Fixing accessory Power 220 V AC 50 Hz Battery life min 3h charge 6 4 Consumables 200 connectors from the syringe to the patient for each syringe ITEM 6 SYRINGE INFUSION PUMP A infusion pump
17. ical air 22 Characteristics Air quality should comply with European Pharmacopoeia cae compressed air purity class Replacement of internal filters should be performed without removing the compressor Provide at least 5 sets of air filters Power Supply 220 VAC 50 Hz 23 Warranty and Service 24 months from the moment of installation 3 Conditions Maximum response time at beneficiary site 48 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language 2 4 Incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language ITEM 3 INTENSIVE CARE MONITOR 2 Item Quantity ITEM 3 INTENSIVE CARE MONITOR 2 3 1 Standard requirements CE Mark ISO 9001 2000 certified companies 3 2 Characteristics Color Screen LCD monitor Minimum screen resolution 800 x 600 pixels Screen dimensions Color screen with at least 300 mm diagonal Speed of recording 6 25 12 5 or 25 mm second Visible in the same time on the screen One lead ECG wave form and heart rate value respiration wave form and respiratory rate Sp0 waveform and value non invasive blood pressure values Systolic diastolic and mean and temperature value Central Control panel will display all instructions in English or Russian
18. ient skin probe reusable 2 pieces b Bassinet and mattress Mattress area minimum 3 000 cm Hydraulic tilt mechanism for the bassinet at least 10 minimum Mattress thickness minimum 2 cm Ray cassette tray to fit under mattress 1 piece Four foldable bassinet walls Item Quantity ITEM 5 TABLE FOR NEWBORN RESUSCITATION 1 Ventilator tube supports c Mounting column Fixed height column on mobile castors Built in rails for accessories 2 pieces Resuscitation module Wall supply pressure 40 75 psi 275 517 kPa Cylinder pressure 2 900 psi max 19 994 kPa Auto Breath infant resuscitator Adjustable breath rate range 18 60 BPM 10 of setting I E ratio Fixed at 1 2 Nominal Adjustable positive end expiratory Pressure PEEP 0 18 H20 0 1 77 kPa Gas bleed 5 LPM Max Precision blender optional 21 100 O2 3 02 d Storage compartments Front storage drawers 2 pieces e Alarms Audio alarms Controller heating alarms Check patient 15 minutes in Manual Mode Baby temperature 1 C from Set Point High temperature Skin Temp 39 0 C 0 2 C Probe Short or open circuit No probe System fail Indicates System Fail Power fail AC Power Interruption Alarm silence reset intervals Check patient Resets clock for 15 minutes manual mode Baby temperature 10 minutes
19. ies cystic fibrosis Dedicated for all medicines bronchodilators antibiotics corticoids and mucolytic Running mode the aerosol is produced continuous Auto clavable up to 134 C Preferable reusable accessories tubes masks Nebulizer Container capacity approximately 5 ml Equipped with tube with length approximately 1m Multi position mask for adult and pediatric Equipped with mouth end piece and flask for dosage Compressor Oil free Maintenance free Small dimensions portable device Low weight portable device Low noise device approximate 30 dB Pressure adjusted to 0 5 bar Possibility to be fed by an oxygen central source compressor flow in charge with minimum 5l min Medical device from electrical class Level of protection IP 30 17 3 Accessories and Consumables Mouth end piece 5pcs Nose plug 2pcs Nose piece 2pcs Masks adult 2pcs ITEM 17 NEBULIZER FOR DRUGS 3 Masks pediatric 5pcs 17 4 17 5 Warranty and Service Conditions Incidental Services Period 24 months from the moment of installation of the equipment Maximum response time 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installatio
20. in temperature range between 35 and 37 C Keypad lock function for protection Override function at air temperature over 38 C c Standard humidity module Internal reservoir included Front mounted for easy accessibility All humidifier parts can be completely removed for filling and cleaning Humidity range up to 70 RH d Mobile stand Stand on four anti static wheels from which two locking Height variable 90 110cm or more Two front drawers e Alarms audio alarms Baby set temperature 1 C High Low air temperature 3 1 C or better Baby skin temperature probe fail Air temperature probe fail Power fail Air flow fail System fail f Accessories included Front drawers 2 pieces Mattress Ipiece Rails for accessories 2 pieces Patient skin probe reusable 2 pcs g Physical properties Power requirements 220 240 V 50 60 Hz Ambient operating temperature minimum 20 30 C Ambient humidity 0 90 RH 4 4 Accessories Front storage drawers Mattress Rails for accessories Patient skin probe 4 5 Warranty and Service Conditions 24 months from the moment of installation Maximum response time at beneficiary site 72 hours Item Quantity ITEM 4 BASIC INCUBATOR 1 4 6 Incidental Services Training for at least 2 medical personnel on sit
21. k top in plastic ABS 21 4 fem Colored strips for drawers Characteristics a Effective security system of the emergency medication Catheter basket 21 5 Accessories Internal drawer organizer Stainless steel holder for oxygen bottles with capacity 3 5 liters Period 24 months from the moment of installation of the Warranty and Service 21 6 He equipment Conditions Maximum response time 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language 21 7 incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language
22. language Measures heart rate beats per minute respirations breaths per minute SpO2 percent blood pressure mmHg temperature C Low weight approximately 6 kg Built in handle for transportation Structure Monitor Battery Patient cables in various lengths e g 3 amp 8 foots or similar range Cable for ECG electrodes Cable for blood pressure cuffs Cable for temperature sensor Cable for pulse oximeter sensors Control buttons On off Alarms standard Stand by Alarms System failure Battery low Apnea Bradycardia Tachycardia Low respiratory rate High respiratory rate Saturation low Saturation high Blood pressure low systolic diastolic mean Blood pressure high systolic diastolic mean Temperature low Temperature high Item Quantity ITEM 3 INTENSIVE CARE MONITOR 2 3 3 Accessories All alarms mentioned above are adjustable by the clinician at the bedside All alarms are audio and visual alarms The alarm tones can be regulated high to low tonality to alarm off at least Heart rate and EGK One derivation visible all the time on the screen There could be two channels visible optional Limits low limit at least 20 beats per minute high limit at least 300 beats per minute Derivations Il Ill aVL aVR a
23. mentary table completely recessable underneath the Item Quantity ITEM 19 CLASSICAL TABLE FOR GYNECOLOGY AND 2 OBSTETRICS main bed Standard side rails to fix accessories Detachable adjustable height and width leg support Detachable adjustable height and width knee crutches with pads cover and strap Easy to clean water and stain proof mattress Specify material and cleaning detergent used Drain pan to collect physiologic and irrigating fluids Waste receptable with drainage hose Auxiliary side arm board with pad swiveling Auxiliary height adjustable infusion pole 19 3 Warranty and Service Conditions Period 24 months from the moment of installation of the equipment Maximum response time 72 hours 19 4 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 20 RADIANT TABLE FOR NEWBORN 1 20 1 Standard requirements CE Mark ISO 9001 2000 certified companies 20 2 Basic Structure Height adjustable table Radiant Warmer consisting of Heat source Bassinet and mattress Mounting column with fixed height Storage comp
24. n User manual in English or translated into Russian language ITEM 18 GLUCOMETER 1 6 4 Standurdvequirements 2E 7 ents EES EEEE ISO 9001 2000 certified companies Whole blood measurements Technical Measurement range min 10 700 mg dl 18 2 Characteristics No light influence Compensated with temperature influence Serial imprecision max 4 18 3 Consumables Strips for 500 of tests open market not limited to one company Period 24 months from the moment of installation of the Warranty and Service 18 4 pe equipment Conditions i Maximum response time 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language 18 5 Incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item ITEM 19 CLASSICAL TABLE FOR GYNECOLOGY AND OBSTETRICS 19 1 Standard requirements CE Mark ISO 9001 2000 certified companies 19 2 Characteristics Non electric delivery bed Robust mobile delivery bed on castors with breaks Easy adjustable sections with handle Specify number of sections Height adjustable with handle Specify range Trendlenburg and reverse Trendelenburg position easy adjustable with handle Specify range Shoulder and head rest Supple
25. nical Requirements 10 4 He A as 0 Conditions Maximum response time at beneficiary site 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language 10 5 Incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 11 ELECTROCARDIOGRAPH ECG 1 CE Mark ISO 9001 2000 certified companies 12 channel ECG with 12 leads Minimum 6 Number of traces to be displayed simultaneously Preferable touch screen Resolution at least 640 480 pixels 11 2 Characteristics Possibility of introducing for each patient the following data name date of birth date of recording weight age diagnosis and comments The central Control panel will display all instructions in English or Romanian language 11 1 Standard requirements Item Quantity ITEM 11 ELECTROCARDIOGRAPH ECG 1 2 leads ECG limb derivations Il Ill aVL aVF aVR precordial derivations V1 V2 V3 V4 V5 V6 Manual selection for any 12 lead Graphic LCD viewing area for life 3 channel display with patient information Indicator for faulty leads Digital filters for all possible interferences Please specify which types of filters Calibration 1mV test Sensitivity and Fre
26. pproximately 0 250 bpm Perfusion optional 0 02 10 Saturation accuracy in motion for infant 3 digits 8 hour battery life for the portable display Configurable display including big numbers option On screen alarm values Alimentation with power cord for 220 V b Sensors Availability of complete range of adhesive and reusable sensors for all patient weights Sensors with recessed photo detector to reduce ambient light and electromagnetic interference Extended adhesive sensor life by use of rejuvenatable adhesive Low weight adhesive sensors lt 5gms Single patient use with proven clinical lifetime of minimum 8 days Adhesive sensors manufactured from durable moisture resistant material 8 4 Accessories and Consumables Interconnection cable 1 pcs 1 SpO2 sensor reusable for adults finger 1 SpO2 sensor reusable for children over 10 kg 50 single patient use SpO2 sensors for infant 30 single patient use SpO2 sensors for neonates lower than 2 Kg 8 5 Warranty and Service Conditions 24 months from the moment of installation Maximum response time at beneficiary site 72 hours 8 6 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Item Quantity
27. quency response test Sensitivity adjustment 5 10 and 20 mm mV Gain adjustment x 2 x 1 and x 2 Full screen preview to determine quality of ECG ECG interpretation with interpretation statements Low battery check and alarm Optional indication of bradycardia and tachycardia Printer Printing speed 25 mm s or 50 mm s Recording paper thermal A4 size Printout of patient ECG with patient information Storing at least 100 records at least 10 seconds long strips Defibrillation protection Pacemaker detection Built in accumulator At least 2 hours autonomy The accumulator is re charging automatically Thermal paper rolls 50 pcs 11 3 Warranty and Service 24 months from the moment of installation Conditions Maximum response time at beneficiary site 72 hours Training for at least 2 medical personnel on site at beneficiary location in Russian language 11 4 incidental Services Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM 12 HOT AIR STERILIZIATION 2 12 1 Standard r quiremente ws andard requirements ISO 9001 2000 certified companies Hot air sterilizer with chamber volume min 50 liters Chamber made from stainless steel Door with lateral opening with handling for safety against
28. te at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation ITEM 14 RESUSCITATION BALOON 2 ITEM 15 LARINGOSCOPE 15 1 Standard requirements CE Mark ISO 9001 2000 certified companies 15 2 Technical Characteristics Fiber optic laryngoscope Laryngoscope handle Fiber optic laryngoscope blades 2 pcs each size adult pediatric and neonatal Rechargeable battery Knurled finish for sure grip Blade with fiber optic for cool light Halogen or xenon light for true tissue color Long lasting illumination Removable fiber optic light pipe for instant replacement Fiber optic resistant about 1000 cycles of sterilizations Blades McIntosh type size 1 2 3 Miller type size 1 2 3 One piece stainless steel Wireless blades eliminate electrical contact Charger for battery 220VAC 50 Hz 15 3 Warranty and Service Conditions Period 24 months from the moment of installation of the equipment Maximum response time 48 hours 15 4 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or tr
29. the _ accidental opening a Charactenistigs Selected programs for sterilization heating or drying of materials Min 3 working programs Ventilation air system in spiral way to assure the temperature in Item Quantity ITEM 12 HOT AIR STERILIZIATION 2 chamber Control panel and LED display Display of temperature and time Indication of the already done and current sterilization phase RS 232 interface for printer or PC communication Acoustic alarm for overtake of programed temperature or other error Precise control of the sterilization cycle parameters with microprocessor Delayed heating start and stop function Temperature sensor Chrome plated trays 2 pcs Working temperature at least from 10 C above ambient to 250 C Time required to reach 250 C max 60 min Weight max 75 kg Power Maxim power consumption 1 5 kW Power supply 220 VAC 50 Hz 12 3 Warranty and Service Conditions 24 months from the moment of installation Maximum response time at beneficiary site 48 hours 12 4 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation User manual in English or translated into Russian language Item Quantity ITEM
30. with shoulder strap and handle 1pc 9 5 Warranty and Service Conditions Period 24 months from the moment of installation of the equipment Battery warranty at least 6 months Maximum response time 72 hours 9 6 Incidental Services Training for at least 2 medical personnel on site at beneficiary location in Russian language Training for at least 2 technical personnel on site at beneficiary location in Russian language Installation and commissioning User manual in English or translated into Russian language Item Quantity ITEM 10 FETAL HEART DETECTORS 1 CE Mark ISO 9001 2000 certified companies Compact lightweight and portable device complete with carrying case rechargeable batteries gel and optional head set if any Specify measurement method Audible and visible heart beat indication Fetal Heart rate calculation and display Specify FHR range bpm Display size Indicate displayed parameters Speaker yes no Power Requirements VAC 220 50Hz Plug Type F CONTROLS ON OFF Volume Audio mute Calibration Easy to clean Waterproof fluid resistant Battery operation Rechargeable standard batteries With integrated battery charger or ext charging station Consumables if any covering 1200 patients Warranty and Service 24 months from the moment of installation 10 1 Standard requirements 10 2 Characteristics 10 3 Tech