Contents
1. Remove the water tub Discard any excess water from the water tub Unclip all four side latches Pull apart the tub lid and base od a Daily cleaning 1 Remove the air tubing by pulling off the finger grips on the cuff Hang It in a clean dry place until next use 2 Wash the disassembled tub lid and base in warm water using a mild detergent 3 Rinse thoroughly in clean water and allow them to dry away from direct sunlight Notes e Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack e Do not wash the air tubing in a washing machine or dishwasher Weekly 1 Remove the air tubing from the S9 device and the mask 2 Wash the air tubing in warm water using mild detergent 3 Rinse thoroughly hang and allow to dry 4 Before next use reconnect the air tubing to the air outlet and mask Monthly 1 Wipe the exterior of the S9 and H5i with a damp cloth and mild detergent 2 Check the air filter for holes and blockage by dirt or dust Replace the air filter if necessary 3 Peel the flip lid seal from the flip lid and wash it in warm water using a mild detergent Ma Ry BAN intenance checklist Inspect the water tub and flip lid seal for wear and deterioration Replace the water tub if any component Is leaking or has become cracked cloudy or pitted Replace the flip lid seal if cracked or torn Clean white powder deposits in the wat
2. e pneumothorax e pathologically low blood pressure e dehydration e cerebrospinal fluid leak recent cranial surgery or trauma S9 adverse effects Patients should report unusual chest pain severe headache or increased breathlessness to their prescribing physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during the course of therapy with these devices e drying of the nose mouth or throat e nosebleed e bloating e ear or sinus discomfort e eye irritation e skin rashes H5i indications for use The H5i is indicated for the humidification of the air delivered from a CPAP or bilevel device The H51 is for use only as recommended by a physician The H5i is intended for single patient re use in the home environment and re use in a hospital institutional environment H5i for Her indications for use The Hbi for Her is indicated for the humidification of the air delivered from a CPAP or bilevel device The H5i is for use only as recommended by a physician The H5i for Her is intended for single patient re use in the home environment and re use in a hospital institutional environment H5i contraindications The H5i is contraindicated for use with patients whose upper Supraglottic airway has been bypassed Troubleshooting If there is a problem try the following suggestions If the problem cannot be solved contact your equipment supplier or
3. Catalogue er RoHS sN Serial number Direct current Q Not for use on more than one patient Me Maximum water level iss Disinfectable up to 200 F 93 C WATER Use distilled or deionized water only s m Lock unlock lad S Remove tub to fill Rx Only Prescription only In the US Federal law restricts these devices to sale by or on the order of a physician A oop dry amp China pollution control logo 1 Ky Gy China pollution control logo 2 Servicing The S9 and H5i devices are intended to provide safe and reliable operation when operated in accordance with the instructions provided by ResMed ResMed recommends that the S9 and H5i be inspected and serviced by an authorized ResMed Service Center if there is any sign of wear or concern with device function Otherwise service and inspection of the devices generally should not be required during the five year design life of the device Limited warranty ResMed Ltd hereafter ResMed warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period e Mask systems including mask frame cushion headgear and tubing 90 days excluding single use devices e Accessories excluding single use devices e Flextype finger pulse sensors e Humidifier water tubs Batteries for use in ResMed internal and external battery systems 6 months Clip
4. ResMed Do not attempt to open these devices Problem possible cause Solution No display Power is not connected Ensure the power cable is connected and the power outlet if available is on The DC plug is partially inserted into the Fully insert the DC plug back of the device The S9 and H5i are not connected Ensure that the S9 and H5i are securely attached correctly Insufficient air delivered from the S9 device Ramp time is in use Wait for air pressure to build up or change ramp time Air filter is dirty Replace air filter Air tubing is not connected properly Check air tubing Air tubing is blocked pinched or Unblock or free the air tubing Check the air tubing for punctured punctures Mask and headgear are not positioned Adjust position of mask and headgear correctly The H5i flip lid is not latched correctly Close the flip lid ensuring that it clicks into place The H5i flip lid seal is not fitted correctly Make sure the flip lid seal is facing the right way up and fitted securely Incorrect air tubing selected If you are using the SlimLine or Standard air tubing ensure that you have the correct air tubing selected via the menu Device does not start when you breathe into the mask Breath is not deep enough to trigger Take a deep breath in and out through the mask SmartStart Stop There is excessive leak Adjust position of mask and headgear Air tubing not connected properly Connect firmly at both end
5. device and air tubing during turbulence Always place the H5i on a level surface below the level of the user to prevent the mask and tubing trom filling with water If liquids are inadvertently spilled into or on the H5i unplug the device from the power outlet Disconnect the H5i from the device and allow it to drain and dry before re using ual Manufacturer ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Distributed by ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA ResMed UK Ltd 96 Milton Park Abingdon Oxfordshire 0X14 4RY UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip S9 H5i SlimLine ClimateLine SmartStart are trademarks of ResMed Ltd S9 and SmartStart are registered in U S Patent and Trademark Office 2012 ResMed Ltd 368875 1 2012 10 Global leaders in sleep and respiratory medicine www resmed com RESMED AutoSet AutoSet for Her Elite Escape Auto Escape POSITIVE AIRWAY PRESSURE DEVICES we E f F H5i l a H5i for Her meh HEATED HUMIDIFIER Welcome Guide WIStcliles English S9 More Comfort s9 Series Welcome Thank you for choosing the S9 Series or H5i Before operating these devices please read the entire Welcome and Information Guides NE ETAT ce OEP me j x A a Z DC power inlet Module Adapter port Air filter SD card slot Air outlet S9
6. reduce the life of these products Incorrect system setup may result in incorrect mask pressure reading Ensure the system is correctly set up Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord Make sure the area around the device is dry and clean and clear of bedding clothes or other objects that could block the air inlet or cover the power supply unit Ensure that the device is protected against water if used outdoors Enclose the device in the S9 travel bag for transport The H5i should only be used with tubing or accessories recommended by ResMed Connection of other delivery tubes or accessories could result in injury or damage to the device Do not open the H5i enclosure There are no user serviceable parts inside Repairs and servicing should only be performed by an authorized ResMed service agent Do not overfill the water tub as water may enter the device and air tubing Do not use any additives eg scented oils and perfumes These may reduce the humidification output of the H5i and or cause deterioration of the water tub materials Take care when handling the H5i as the water water tub may be hot Allow 10 minutes for the heater plate and any excess water to cool The H5i should only be connected or disconnected when the water tub is empty Make sure that the water tub is empty before transporting the H5i Do not operate the H5i on an aircraft as water may enter the
7. 15 mm inner diameter ClimateLine heated air tubing Flexible plastic and electrical components 6 3 1 9 m 19 mm inner diameter SlimLine air tubing Flexible plastic 6 1 8 m 15 mm inner diameter Standard air tubing Flexible plastic 6 6 2 m 19 mm inner diameter Heated tubing temperature cut out lt 106 F lt 41 C Notes e The manufacturer reserves the right to change these specifications without notice e The temperature and relative humidity settings displayed for ClimateLine or ClimateLine are not measured values e Check with your clinician service provider before using the SlimLine air tubing with devices other than the S9 or Hoi Humidifier performance The following settings have been tested at 22 C ambient temperature CPAP mask RH output Nominal system output AH BTPS pressure cm H20 Setting 3 Setting 6 Setting 3 Setting 6 4 90 100 10 18 10 95 100 115 21 20 95 100 11 18 a AH Absolute Humidity in mg L b BTPS Body Temperature Pressure Saturated Symbols The following symbols may appear on your S9 H5i power supply unit air tubing or packaging AN Caution Read instructions before use Cli Follow instructions for use P21 Protection against Insertion of fingers and against vertically dripping water P20 Not drip proof Type BF equipment ol Class Il equipment C Start Stop onal Manufacturer European Authorised Representative uw European RoHS LoT Batch code REF
8. 5VA 41 F to 95 F 5 C to 35 C Note The air flow for breathing produced by this therapy device can be higher than the temperature of the room Under extreme ambient temperature conditions 104 F 40 C the device remains safe 10 95 non condensing Sea level to 8 500 2 591 m 4 F to 140 F 20 C to 60 C 10 95 non condensing Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 for residential commercial and light industry environments It is recommended that mobile communication devices are kept at least 1 m away from the device Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found on www resmed com on the Products page under Service and Support Click on the PDF file for your language ResMed confirms that the S9 Series meets the Federal Aviation Administration FAA requirements RTCA DO 160 for all phases of air travel Class Il double insulation Type BF S9 technical specifications Operating pressure range Maximum single fault steady state pressure Pressure measurement tolerance Flow measurement tolerance 4 to 20 cm H20 30 cm H20 0 5 cm H20 4 of the measured reading 6 L min or 10 of reading whichever is greater DECLARED DUALNUMBER NOISE EMISSION VALUES in accordance with ISO 4871 Sound pressure level Sound power level Nominal dimensions L x W x H Weight Housing c
9. Exhibit 43 RESMED S9 and H5i Series SLEEP THERAPY SYSTEM Accessories Guide English De o Filter cover rT 36859 9 filters k S9 SD card Standard 1 pack 36850 2 pack 36851 12 pack 36852 50 pack 36853 Hypo allergenic 1 pack 36855 2 pack 36856 12 pack 36857 50 pack 36858 H5i heated humidifier 9 PRODUCT CODES S9 AutoSet 36005 S9 AutoSet and H5i 36015 S9 AutoSet and Climate Control Kit 36025 S9 Elite 36003 S9 Flite and H5i 36013 REIMBURSEMENT CODES S9 Elite and Climate Control Kit 36023 Filter disposable 2 per month A7038 S9 Escape 36001 Tubing 1 per 3 months A7037 S9 Escape and Hi 36011 Tubing with integrated heating 1 per 3 months A4604 S9 Escape and Climate Control Kit 36021 Humidifier water tub 1 per 6 months A7046 59 Escape Auto 36002 S9 Escape Auto and Hd 36012 S9 Escape Auto and Climate Control Kit 36022 a eee TM L Booo wwo Series HS with cleanable water tub 36901 More Comfort Includes H5i and ClimateLine Tubing RESMED S9 ACCESSORIES K i j SlimLine tubing SlimLine Standard tubing Standard tubing S9 90W power S9 travel bag DC DC converter 36810 tubing wrap e 6 6 replacement wrap supply unit 36860 24V 90W output 36811 coop PS 33963 36821 36970 e 6 6 replacement SO 30W power autoclavable standard tubing supply unit 14986 36920 6 replacement standard tubing 14994 H5i ACCESSORIES ClimateLine H5i water tub H5i c
10. at a glance The S9 system comprises the following elements e S9 device SlimLine air tubing 90W power supply unit S9 Travel bag SD card e S9 SD card protective folder Optional components include e H5i heated humidifier Standard air tubing 30W power supply unit does not support H5i e Power Station battery pack DC DC Converter 24V 90W Traveling tips When traveling with your S9 only e Ensure you pack the SlimLine or Standard air tubing as the ClimateLine or ClimateLine heated air tubing is not designed to connect directly to the S9 device e Please ensure you purchase and travel with the approved power cord for the region where you will be using the S9 device l English ClimateLine ClimateLine connection port H5i at a glance H5i standard The H5i system comprises the following elements water tub e H5i heated humidifier e H5i standard water tub Optional components include e ClimateLine heated air tubing ClimateLine heated air tubing H5i cleanable water tub Traveling tips When moving or traveling with your H5i e Ensure that the water tub is empty e Disconnect the H5i from the S9 by using the release button Setup 1 Align your H5i with your S9 and push them together until they click into place 2 Connect the DC plug of the power supply unit to the rear of the S9 3 Connect the power cord to the power supply unit 4 Plug the other end of t
11. ave long lengths of air tubing around the top of your bed It could twist around your head or neck while you are sleeping Do not use electrically conductive or antistatic air tubings Do not use the air tubing if there are any visible signs of damage Only ResMed air tubing and accessories should be used with the device A different type of air tubing or accessory may alter the pressure you actually receive reducing the effectiveness of the treatment Only use the ResMed 90W or 30W power supply unit Use the 90W power supply unit to power the system comprising the device H5i air tubing DC DC converter and battery pack The 30W power supply unit is designed to power the device only and recommended for traveling Only ResMed products are designed to be connected to the module connector port Connecting other devices could damage the device Blocking the air tubing and or air inlet of the device while in operation could lead to overheating of the device Do not operate the H5i if it is not working properly or if any part of the device or H5i has been dropped or damaged Z cautions Do not open the device enclosure There are no user serviceable parts inside Repairs and servicing should only be performed by an authorized ResMed service agent Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the device humidifier or air tubing These solutions may cause damage and
12. d on the LCD after you try to update the settings using the SD card Settings updated successfully press any key The settings were not updated Contact your clinician service provider immediately Water splashing on your face from the Hbi The water tub is overfilled Condensation is forming in the air tube and mask Leaking water tub The water tub may be damaged or cracked The cleanable water tub is not assembled correctly Air feels too warm cold in the mask The temperature of the ClimateLine or ClimateLine air tubing Is set too high low Check that the water level is below the maximum water level mark Turn the H5i or relative humidity setting down via the menu Contact your service provider for a replacement Check for damage and reassemble the cleanable water tub correctly Turn up down the heated tubing temperature via the menu General technical specifications 90W power supply unit 30W power supply unit Operating temperature Operating humidity Operating altitude Storage and transport temperature Storage and transport humidity Electromagnetic compatibility Aircraft use IEC 60601 1 classification Input range 100 240V 50 60Hz Nominal for aircraft use 115V 400Hz Typical power consumption 70W 80VA Maximum power consumption 110W 120VA Input range 100 240V 50 60 z Nominal for aircraft use 115V 400Hz Typical power consumption 20W 40VA Maximum power consumption 36W 7
13. displayed displayed in blue highlighted in orange 2 CaS amp EE nd Tube SlimLine it Tube SlimLine lt HOME zg HOME Push Dial 4 Turn amp until the mask 5 Press to confirm your EENS G to return to the setting that you require is choice HOME screen displayed Note Mask Fit is only available in S9 AutoSet and S9 Elite Selecting the mask type is not possible if S9 Essentials has been enabled by your clinician G pu 7 Getting started Notes Make sure the power is connected Adjust the ramp time or humidification level if required Fit your mask as described in your mask user guide To start therapy simply breathe into the mask and or press Lie down and arrange the air tubing so that it is free to move if you turn in your sleep To stop treatment at any time remove your mask and or press SmartStart is only available in S9 AutoSet and S9 Elite If your clinician has enabled SmartStart your device will start automatically when you breathe into the mask and stop automatically when you remove your mask Once therapy has started a treatment screen is displayed If power is interrupted during treatment the device automatically restarts therapy when power is restored g English Cleaning and maintenance You should regularly carry out cleaning and maintenance as described in this section Disassembling the water tub Slide the latch Lift open the flip lid
14. er tub by using a solution of one part household vinegar to 10 parts water Reassembling and filling the water tub aRwWN gt Place the tub lid back onto the base Clip all four side latches Fill the water tub with distilled or deionized water up to the maximum water level mark Return the water tub to the H5i Close the flip lid ensuring that it clicks into place Replacing the air filter Replace the air filter every six months or more often If necessary 1 2 3 4 No Remove the air filter cover from the back of the S9 device Remove and discard the old air filter Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover Replace the air filter cover tes Ensure the air filter and air filter cover are fitted at all times Do not wash the air filter The air filter is not washable or reusable TU Air filter cover A Air filter 11 English SD card An SD card has been supplied to gather therapy data from your S9 device and provide settings updates from your clinician When instructed to do so disconnect the S9 device from the power outlet remove your SD card insert it in the protective folder and send it to your clinician Removing the card 1 Push in the SD card to release it 2 Remove the card 3 Insert the card into the protective folder 4 Send the protective folder back to your clinician as instructed For more information on remov
15. f a physician S9 AutoSet indications for use The S9 AutoSet self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 AutoSet self adjusting system is intended for home and hospital use S9 AutoSet for Her indications for use The S9 AutoSet for Her self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 AutoSet for Her self adjusting system is intended for home and hospital use S9 Elite indications for use The S9 Elite CPAP system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Elite CPAP system is intended for home and hospital use S9 Escape Auto indications for use The S9 Escape Auto self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Escape Auto self adjusting system is intended for home and hospital use S9 Escape indications for use The S9 Escape CPAP system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Escape CPAP system is intended for home and hospital use S9 contraindications Positive airway pressure therapy may be contraindicated in some patients with the following pre existing conditions e severe bullous lung disease
16. he power cord into the power outlet 5 Connect one end of the air tubing firmly onto the air outlet 6 Connect the assembled mask system to the free end of the air tubing Notes e Place the power supply unit away trom the H5i to allow for adequate ventilation e For more information on assembling your mask see your mask user guide e Recommended masks are available on www resmed com on the Products page under Service amp Support 3 English Filling the water tub 1 Slide the latch and lift open the flip lid 2 Remove the water tub 3 Fill the water tub through the center hole with distilled or deionized water up to the maximum water level mark 380 mL Return the water tub to the H5i Close the flip lid ensuring that it clicks into place oo S9 basics Start Stop button Starts or stops treatment Home A Humidity level Ramp all Setup menu Allows you to Heated tube Rese make changes to settings Climate Control Info menu D Allows you to 3 view your sleep Start Stop statistics Push dial Push dial Turning the dial is allows you to amp Setup menu amp scroll through the menu and change settings Pushing the Info menu dial allows you to enter into a menu and The Info and Setup menus are disabled if confirm your choice S9 Essentials has been enabled by your clinician English S9 Essentials S9 Essentials i
17. in good condition and the equipment is not damaged Keep the power cord away from hot surfaces The device should only be used with masks and connectors recommended by ResMed or by a physician or respiratory therapist A mask should not be used unless the device is turned on Once the mask is fitted ensure that the device is blowing air The vent hole or holes associated with the mask should never be blocked Explanation The device is intended to be used with special masks or connectors which have vent holes to allow continuous flow of air out of the mask When the device is turned on and functioning properly new air from the device flushes the exhaled air out through the mask vent holes However when the device is not operating insufficient fresh air will be provided through the mask and the exhaled air may be rebreathed Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation This applies to most models of CPAP or bilevel devices Oxygen supports combustion Oxygen must not be used while smoking or in the presence of an open flame 1 Ports may be incorporated into the mask or in connectors that are near the mask Always ensure that the device is turned on and airflow generated before the oxygen supply is turned on Always turn the oxygen supply off before the device is turned off so that unused oxygen does not accumulate within the device enclosure and create a risk of fire Do not le
18. ing and inserting your card refer to your S9 SD Card Protective Folder provided with your device Please retain the S9 SD Card Protective Folder for future use a ya NN Ur e ig iT MM Notes e For Troubleshooting information refer to the Information Guide e For more product information see www resmed com s9 Series 368656 2 2013 01 S9 Series mie WELCOME AMER wal Manufacturer ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Distributed by ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA ResMed UK Ltd 96 Milton Park Abingdon Oxfordshire 0X14 4RY UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip S9 AutoSet S9 AutoSet S9 Escape S9 Elite S9 Escape Auto H5i SlimLine and ClimateLine are trademarks of ResMed Ltd and S9 AutoSet Escape SlimLine and ClimateLine are registered in U S Patent and Trademark Office 2013 ResMed Ltd Global leaders in sleep and respiratory medicine www resmed com
19. iod during which the pressure increases from a low start pressure to the treatment pressure To adjust your ramp time 1 From your HOME 2 Press E 3 Turn 4 untilthe 4 Press 4 to screen turn lt j The selection is ramp time you confirm your choice until the RAMP highlighted in require is displayed screen is displayed orange in blue Using mask fit Available in S9 AutoSet and S9 Elite mask fit helps you fit your mask properly This feature delivers treatment pressure for a three minute period prior to starting treatment during which you can check and adjust your mask fit to minimize leaks 1 Fit your mask as described in the mask user guide 2 From your HOME screen push and hold 4 for three seconds One of the following MASK FIT screens is displayed MASK FIT Y MASK FIT 3 If necessary adjust your mask mask cushion and headgear until you have a Good mask fit After three minutes the pressure reverts to the set pressure and treatment will begin You can end mask fit at any time by pressing 4 7 English Selecting the mask type To select your type of mask SETTINGS GJ Tube SlimLine Mask Pillows SETTINGS het Tube SlimLine wM SETTINGS i Tube SlimLine Mask Fit gt Mask Fit gt g HOME sk Fit lt lt HOME z a ay i Mask Full Face Mask Fit gt Mask Fit gt 1 2 tees amp The SETTINGS 2 Turn amp until Mask is 3 Press The selection is screen Is
20. laimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from region to region For further information on your warranty rights contact your local ResMed dealer or ResMed office Z WARNINGS Read the entire manual before using the device Use the device only as directed by your physician or healthcare provider Use the device only for the intended use as described in this manual Advice contained in this manual should not supersede instructions given by the prescribing physician If you notice any unexplained changes in the performance of the device if it is making unusual or harsh sounds if the device or the power supply are dropped or mishandled if water is spilled into the enclosure or if the enclosure is broken discontinue use and contact your ResMed Service Center Beware of electrocution Do not immerse the device humidifier power supply or power cord in water In the event of a spill disconnect the device from the power supply and let the parts dry Always unplug the device before cleaning and make sure that all parts are dry before plugging in the device Explosion hazard do not use in the vicinity of flammable anesthetics Make sure the power cord and plug are
21. leanable H5i flip lid H5i flip lid seal H5i air outlet tubing 36802 water tub 36891 36892 and outlet clip 36995 36800 36894 DATA MANAGEMENT ACCESSORIES ResScan SD card SD card reader S9 oximeter adapter S9 wireless module Version 3 11 protective folder 36931 36940 36930 31309 including SD Card e USB extension cable 7071015 36841 10 pack Only compatible with the S9 AutoSet amp S9 Elite Series ResMed Corp San Diego CA USA 1 858 836 5000 or 1 800 424 0737 toll free ResMed Ltd Bella Vista NSW Australia 61 2 8884 1000 or 1 800 658 189 toll free See ResMed com for other ResMed locations worldwide H5i ResScan S9 AutoSet S9 Elite S9 Escape S9 Escape Auto M ore Comfo rt and SlimLine are trademarks of ResMed Ltd AutoSet ClimateLine Escape and S9 are trademarks of ResMed Ltd and are registered in the U S Patent and Trademark Office 2011 ResMed Specifications may change without notice 1014082 2 11 01 Global leaders in sleep and respiratory medicine www resmed com RESMED S9 AutoSet S9 AutoSet for Her S9 Elite S9 Escape S9 Escape Auto POSITIVE AIRWAY PRESSURE DEVICES aTM aTM H5i H5i for Her HEATED HUMIDIFIER ClimateLine ClimateLine 4 SlimLine Standard AIR TUBING Information Guide English Please read the entire Information and Welcome Guides before using your device N CAUTION In the US Federal law restricts this device to sale by or on the order o
22. onstruction Supplemental oxygen Hypoallergenic air filter Standard air filter Air outlet H5i technical specifications Maximum heater plate temperature Temperature cut out Maximum gas temperature Nominal dimensions L x W x H Weight standard water tub Weight cleanable water tub Water capacity Docking station 24 dBA as measured according to ISO 17510 1 2002 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510 1 2007 34 dBA with uncertainty of 2 dBA as measured according to ISO 17510 1 2007 6 0 x 5 5 x 3 4 153 mm x 140 mm x 86 mm 1 8 Ib 835 g Flame retardant engineering thermoplastic Recommended maximum supplemental oxygen flow 4 L min Non woven acrylic and polypropylene fibers with polypropylene carrier Polyester fiber The 22 mm conical air outlet complies with ISO 5356 1 150 F 65 C 165 F 74 C lt 106 F lt 41 C Docking station and water tub 6 0 x 5 7 x 3 4 153 mm x 145 mm x 86 mm Docking station and unfilled water tub 1 52 Ib 0 69 kg Docking station and unfilled water tub 1 74 Ib 0 79 kg To maximum fill line 380 mL Flame retardant engineering thermoplastic aluminium Cleanable water tub Injection molded plastic stainless steel and silicone seal Standard water tub Injection molded plastic aluminium and thermoplastic elastomer Air tubing technical specifications ClimateLine heated air tubing Flexible plastic and electrical components 6 6 2 m
23. r supply unit is free from bedding by bedding clothes or other objects that could cover it Displays error message Tube blocked please check your tube Air tubing is blocked Check your air tubing and remove any blockages Displays error message High leak please check system setup and all connections There is excessive leak Adjust position of mask and headgear Air tubing is not connected properly Connect firmly at both ends The HB5i flip lid is not latched correctly Close the flip lid ensuring that it clicks into place Displays error message Please close H5i flip lid attach tube and press any key H5i flip lid is not closed Close the flip lid ensuring that it clicks into place Problem possible cause Air tubing is not connected properly Solution Connect firmly at both ends The following message is displayed on the LCD after you try to update settings or copy data to the SD card Card error please remove SD card and contact service provider SD card is not inserted correctly You may have removed the SD card before settings were copied to the CPAP device Ensure that the SD card is inserted correctly Reinsert the SD card and wait for the home screen or the Settings updated successfully press any key message to appear on the LCD Note This message only appears once If you re insert the SD card after you have updated your settings the message will not be re displayed The following message is NOT displaye
24. s Problem possible cause Solution SmartStart Stop is disabled Enable SmartStart Stop Note SmartStart Stop is not available in S9 Escape and S9 Escape Auto Device does not stop when you remove your mask SmartStart Stop is disabled Enable SmartStart Stop Note SmartStart Stop is not available in S9 Escape and S9 Escape Auto SmartStart Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system being used Only use equipment recommended by ResMed The patient is using a nasal pillows Disable SmartStart Stop mask with a set pressure less than 7 cm H320 Pressure rises inappropriately Talking coughing or breathing in an Avoid talking with a nasal mask on and breathe as normally unusual manner as possible Mask cushion is buzzing against the Adjust the headgear skin Displays error message Heated tube fault replace tube Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and then reconnect it to restart the device There is a fault in your ClimateLine or Discontinue using your ClimateLine or ClimateLineM X air ClimateLine air tubing tubing and contact your clinician service provider Use Standard or SlimLine tubing in the interim Displays error message Humidifier fault replace humidifier Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and then reconnect i
25. s designed to make device interaction and menu navigation easier for you If enabled by the clinician S9 Essentials disables the Info and Setup functionality so that you can simply start and stop therapy and adjust ramp humidification and Climate Control Adjusting humidity level Ranging from OFF to 6 you can adjust the humidity level at any time to find the setting that is most comfortable for you To adjust your level of humidity 1 From your HOME 2 Press jp 3 Turn amp until 4 Press amp to screen turn amp The selection is the humidity level confirm your choice until the HUMIDITY highlighted in that you require Is BEVEL screen S orange displayed displayed in blue Warming the H5i The warm up feature is used to pre heat the water before starting AQ HUMIDITY LEVEI treatment o aa To start warming 1 Navigate to the humidity level icon 2 Push and hold w for three seconds The Warming Up status A bar is displayed You can stop warm up at any time by pressing forat least three seconds You can start treatment at any time by pressing Note In order to assist the heater plate in cooling your S9 device will continue to blow air for up to an hour after treatment has stopped However you can unplug the device from the power outlet at any time and allow the heater plate to cool without air flow Adjusting ramp time Designed to make the beginning of treatment more comfortable ramp time is the per
26. t to restart the device There is a fault in your H5i heated Discontinue using your H5i and contact your clinician humidifier service provider Refilling the humidifier with cold water Allow the humidifier to cool before re filling while humidifier is still hot after therapy Problem possible cause Solution Filling the humidifier with ice cold water Use room temperature water on a warm day or with hot water Displays error message High temperature fault refer to user manual Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and then reconnect it to restart the device Air filter is blocked Replace your air filter Disconnect the power cord and then reconnect it to restart the device Air tubing is blocked Check your air tubing and remove any blockages Disconnect the power cord and then reconnect it to restart the device Humidifier setting is too high resulting Turn the humidifier setting down and empty the water from in accumulation of water in the air the air tubing tubing Displays error message Check ResMed 30 90W Power Supply Unit and fully insert the connector The DC plug is partially inserted into the Fully insert the DC plug back of the device or inserted too slowly A non ResMed power supply unit is Remove the power supply unit and replace with a ResMed connected to the device power supply unit The power supply unit is being covered Make sure that the powe
27. type finger pulse sensors 1 year CPAP and bilevel device data modules Oximeters and CPAP and bilevel device oximeter adapters Humiditier cleanable water tubs Titration control devices CPAP bilevel and ventilation devices including external power supply units 2 years Battery accessories Portable diagnostic screening devices Humidifiers This warranty is only available to the initial consumer It is not transferable If the product fails under conditions of normal use ResMed will repair or replace at its option the defective product or any of its components This limited warranty does not cover a any damage caused as a result of improper use abuse modification or alteration of the product b repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs and c any damage or contamination due to cigarette pipe cigar or other smoke Warranty is void on product sold or resold outside the region of original purchase Warranty claims on defective product must be made by the initial consumer at the point of purchase This warranty replaces all other expressed or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for any incidental or consequential damages c