1. Check to make sure the fan is operating properly and the device is being used within its environmental temperature Illuminator has shut range see Installation Recommendations down and is hot Make sure there is nothing obstructing the airflow to the Microscope Illuminator It is required that a minimum of eight cubic feet of airspace is provided around the device SCOPELED Page 15 of 18 Copyright2015 Trouble Shooting SCOPELED G150 SERIES ILLUMINATOR SPECIFICATIONS Absolute Maximum Ratings Power Adapter Voltage 10 8V DC to 13 2V DC Input Current Max 3 5A Operating Temp 0 C to 40 C 32 F to 104 C Storage Temp 20 C to 75 C 4 F to 167 F Humidity 10 to 90 CCT Range 3000K to 5600K Table Top Model Weight 0 771KG 1 7Ibs Pole Mount Model Weight 0 635KG 1 4Ibs Optical Specifications Typical Optical Characteristics Tunable White Output Type G150T TW or G150P TW 8mm Light Guide Typical CCT Radiant Flux 7 Luminous Flux Color Settings mW Lm 3000 K 510 135 4300 K 930 260 5600 K 740 205 Typical Optical Characteristics Single Color Output Type G150T 5600K or G150P 5600K 8mm Light Guide Typical CCT Radiant Flux Luminous Flux Color Settings mW Lm 5600 K 1280 400 SCOPELED ee Copyright2015 Specifications Mechanical Dimensions SCOPELED G150 SERIES ILLUMINATOR Mech
2. DC Power Jack SCOPELED Page 9 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR Pole Mount Model Installation 1 Set the G150 Series ON OFF switch to OFF 2 Mount the illuminator to the pole 3 Connect the 12V AC DC power adapter to the illuminator at the DC power port shown on the model images below and then plug the 12V AC DC power adapter to wall socket ON OFF Switch Pole Mount Model SCOPELED Page 10 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR 4 Pole Adapter Optional Pole Adapter will be necessary for microscope stand with vertical or horizontal pole diameters less than 40mm Insert the Pole Adapter into the G150P 40mm opening be sure the openings on the Pole Adapter are facing the post thumb screw Control Knobs ON OFF Switch Light Output Port SCOPELED Page 11 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR Installation Recommendations Mounting The ScopeLED G150 Series Illuminator is available as a Table Top unit or as a Pole Mount unit that is intended for mounting on typical microscope stands from 32mm to 40mm in diameter The following considerations should be made before choosing a mounting option Environment The illuminator is not suited for wet or damp locations Be sure to protect all electronics from moisture Do not mount or locate the unit in an explosive environment near combustibl
3. 1 rd 0 8 COLOR INTENSITY Single Color G150T 5600K or G150P 5600K 2 6 1 7 0 8 INTENSITY Preprogrammed Modes Preprogrammed Modes for Tunable White Output Type G150T TW or G150P TW Mode White Light CCT Value 1 3000 K 2 4300 K 3 5600 K Preprogrammed Modes for Single Color Output Type G150T 5600K or G150P 5600K Mode White Light CCT Value 1 5600 K SCOPELED Page 13 of 18 Copyright2015 Manual Operation SCOPELED G150 SERIES ILLUMINATOR Basic Operation 1 Follow HARDWARE SETUP to properly install the G150 Series Illuminator onto your application system 2 Press the ON FF switch on the control panel to the ON position to turn the illuminator on The illuminator will operate at the setting indicated by the control knobs 3 Use the COLOR tuning color mode knob to select the desired preprogrammed colors white light CCT values or custom settings These knob settings are also referred to in this manual as preprogrammed modes See Preprogrammed Modes for mode descriptions Rotating the Color tuning knob clockwise will progress through these modes in the order listed 4 The INTENSITY knob is used to manually adjust the intensity or brightness of the modes without changing its color parameters Turning the knob fully clockwise results in maximum brightness of the selected mode Turning the knob fully counter clockwise results in dimming the selected mode to zero out SCOP
4. 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR 3 Light Guide Installation Place the illuminator on a desk Connect the single or bifurcated gooseneck to the light guide adapter and secure it with the set screw Insert the gooseneck or ring light into the LED illuminator light output port and secure it with the tightening screw SCOPELED Page 6 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR 4 Lead Base Support Stand Optional If using a rigid gooseneck light guide with the G150T or G180T table top models the optional SLAG BASE lead base is recommended Remove the four screws on the bottom of the LED illuminator Then place the base over the bottom lining up the screw holes and re insert the screws to attach the base to the bottom of the illuminator Lead Base SCOPELED Page 7 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR The optional SLAG SS support stand may also be used but will not provide as much stability as the lead base Snap the U shaped opening of the support stand into the indent ring located behind the LED illuminator light output port Push down hard to fully seat the stand in the ring If needed extend out the foldable arms for additional lateral support Support Stand SCOPELED Page 8 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR Table Top Controls ON OFF Switch Control Knobs
5. ELED Page 14 of 18 Copyright2015 Manual Operation SCOPELED G150 SERIES ILLUMINATOR TROUBLE SHOOTING Problem Solution Check that the green light on the power adapter is on and verify that all of the connections to the adapter and the Microscope Illuminator are properly connected Turn the intensity knob to maximum intensity and cycle through the different modes using the color tuning knob The color tuning knob can be in a custom mode position where the custom mode has not been programmed Light output only Make sure that the protective dust cover on the end cap has appears red been removed The fan should be on constantly If the fan fails to engage turn off the unit immediately and contact the manufacturer Turn the intensity knob to maximum intensity then cycle through each preset mode using the color tuning knob If any of the colors appear unusually dim please contact the manufacturer Flickering is often a result of the driver not receiving a high enough DC voltage Verify that the 12V DC output from the power adapter is correct Shorten the line from the power supply to the Microscope Illuminator Never exceed more than a 15 ft length from the power supply Power cycle the Microscope Illuminator by shutting off the power to the power adapter and restarting Illuminator does not turn on Fan does not turn on Illuminator is not emitting all of the colors Illuminator is flickering
6. Pole Mount Both the Table Top and Pole Mount Illuminators come in Tunable White or Single Color output types Please reference the label at the bottom of the unit to confirm that you are using one of these models If not contact the manufacturer for the proper user manual Tunable White G150T TW or G150P TW The tunable white light illuminator has three preprogrammed selectable CCTs of 3000K 4300K and 5600K Single Color G150T 5600K or G150P 5600K The single color illuminator provides single white color at 5600K only A customized color model may be available upon special request SCOPELED Page 2 of 18 Copyright2015 Product Overview SCOPELED G150 SERIES ILLUMINATOR Naming Convention for ScopeLED G150 Series product s q so __ _ Device SLG150 Light Guide Illuminator Mounting T Table Top P Pole Mount 5600K 5600K Fixed CCT TW Tunable White Light Included Hardware amp Accessories Illuminator x 1 Tightening Set Screw x 1 AC DC Power Adapter x 1 AC Power Cord x 1 Quick Start Guide x 1 Pole Adapter 40 to 32mm Optional PN SLAG PA32 Pole Adapter 40 to Specify ID Optional PN SLAG PACU G Series Support Stand for Table Top Models Using Ring Light Guides Optional PN SLAG SS G Series Lead Base for Table Top Models Using Rigid Light Guides Optional PN BASE Rigid Bifurcated Gooseneck Light Guide Optional PN SLAG BIFR Ring Light Guide Optional PN SLAG RING Light Gui
7. SCOPELED G150 SERIES ILLUMINATOR OPERATION MANUAL Table Top Pole Mount S SCOPELED illuminating your view COPYRIGHT 2015 2016 ScopeLED All rights reserved Printed in the United States of America This manual may not be reproduced in whole or in part in any form or by any means without the express written permission of ScopeLED NO LIABILITY FOR ERRORS ScopeLED reserves the right to correct technical and typographical errors in this manual at any time without prior notice In no event shall ScopeLED be liable for errors in this manual or for any damages arising out of or relating to this manual PRODUCT WARRANTY LIMITATION OF REMEDITS ScopeLED warrants to the original Buyer all of its products to be free from defects in both workmanship and material for a period of one year from the date of shipment This warranty extends to all products which have proved defective through normal use but excludes products that have been damaged mishandled disassembled modified or misused by Buyer or any other person This warranty is in lieu of all other warranties and ScopeLED disclaims all other warranties express or implied including any warranty of merchantability fitness for a particular purpose or arising from the course of dealing between the parties or usage of trade ScopeLED does not extend any warranty of any kind whatsoever to any purchaser of the products from Buyer or to any end user of the products Sco
8. a KAKA KAKA KK KAK H HAAA KAWAR 2 Included Hardware amp Accessories sscesseeeeeeeeeeeeeeeeeeneeeeeeeeeeeaaeeeeeeeeeeeeeeeseeeeeeneeees 3 Safety ln SIEUGtl eairt aaraa i aa aa a aaae saaa ae naa Aa r aa ara aAa a ai aaraa sdn aneian Erini 4 HARDWARE SETUP a kaba r k kaba nak ka ba ek NANANA 5 Table Top Model Installation eeeeeeeeeeeeeeeeeeeeeeeeeeaeeeeeneeeeeeeeeeeaseeeeeeeeeeeeeeenneees 5 Pole Mount Model Installation mm 10 Installation Recommendations kk kaka kk kk AK KAK KK K KAK HHH AKA WAA 12 MANUAL OPERATION uma mnman nananana nanan Ea aa Aaaa kana k n 13 Control PAn ls lid sal k dikakan ak kER RE RRA AER kK TE 13 Basic Opeha Nona ANAL asa kalangan lk dka kes a ka W rek ke adie 14 TROUBLE SHOOTING o zay xer daka da keka k y kayan k xeya kek kaka kek aya kev naka nek xeva kek aka nekan 15 SPECIFICATIONS nkutu xx sti NAAN GG ya kaya kaya zila kaya kuya kuy kur ka ya ku aka ya sa aka ku su akuya s d ANO kada 16 Absolute Maximum Ratings kk KAKA KAKA KHK KA HAAA KAK HKH WAA 16 Optical SpeclfiGatlons sy ii li lik Aa AA es kul k K k Ku l aub H W Kak e 16 Mechanical Dimensions a AARAL 17 Mechanical Drawings eiii aha ALA 17 SCOPELED G150 SERIES ILLUMINATOR PRODUCT OVERVIEW This product must be installed in accordance with the applicable installation code by a person familiar wi
9. anical Dimensions for all Microscope Illuminator models Model Dimensions Illuminator only unit in mm Table Top 167 L x 103 W x 103 H Pole Mount 164 L x 98 W x 109 H Mechanical Drawings G150 Table Top Series SCOPELED Copyright2015 Specifications Page 17 of 18 SCOPELED G150 SERIES ILLUMINATOR G150 Pole Mount Series SCOPELED Page 18 of 18 Copyright2015 Specifications
10. de Adapter Optional Custom Design SCOPELED Page 3 of 18 Copyright2015 Product Overview SCOPELED G150 SERIES ILLUMINATOR Safety Instruction Warnings e Never look directly at the light output port e Inappropriate usage of ScopeLED G Series products could result in eye damage e Installation of the product onto readily available microscopes should always be performed according to the written guidelines Attention also see Installation Recommendations amp Absolute Maximum Ratings e The statements regarding safety of operation of the G Series will apply only when the product is operated correctly e Any installation or attachment of light guide to the G Series Illuminator should always be performed with power adapter and other cables disconnected e When installing or removing a light guide the G Series Illuminator ON OFF switch should always be set to OFF e The G Series must not be operated in explosion prone environments The G Series products must always be operated with shielded cables SCOPELED Page 4 of 18 Copyright2015 Product Overview SCOPELED G150 SERIES ILLUMINATOR HARDWARE SETUP Table Top Model Installation 1 Set the G150 Series ON OFF switch to OFF 2 Connect the 12V AC DC power adapter to the illuminator at the DC power port shown on the model images below and then plug the 12V AC DC power adapter to wall socket ON OFF Switch Table Top Model SCOPELED Page 5 of
11. e material Air Volume Do not encase or bury the unit in any way that restricts airflow Orientation The post mount version can be oriented vertically horizontally or inverted Be sure that the mounting method and location is capable of supporting the weight of the illuminator Power Adapter The output connector of the power adapter must be a female barrel connector with dimensions of 2 1mm x 5 5mm for direct compatibility with the G150 Series Illuminator Though the fitting between the output connector of the power supply and input connector of the Microscope Illuminator may be snug the connection will not be able to support the weight of the Illuminator Do not use the power supply as its own hanging device Light Guides The G150 Series Illuminator is most efficient for light guides in the range of 4 to 25 4 mm outer diameter The Microscope Illuminator is designed to accept a variety of light guides Custom size adapters can be made upon request SCOPELED Page 12 of 18 Copyright2015 Hardware setup SCOPELED G150 SERIES ILLUMINATOR MANUAL OPERATION Control Panel To manually operate the G150 Series Illuminator two knobs are available for color tuning Not applicable for single color model and intensity levels The intensity knob is identical for all three light output types of the G150 Series Illuminator tunable white and single color Tunable White G150T TW or G150P TW 3000K ker ak 5600K k 2 6
12. peLED at its sole choosing will replace or repair to proper working condition any products under warranty that are returned Products repaired or replaced under warranty are only warranted for the remaining unexpired period of time of the original warranty ScopeLED reserves the right to issue a credit memo for any defective product as an alternative to product replacement or repair ScopeLED will not accept Buyer generated debit memos Buyer may not set off or withhold payment because any product is defective In no event shall ScopeLED liability under this warranty and this contract exceed the purchase price of the products In no event shall ScopeLED be liable under this warranty or this contract for consequential incidental or special damages RETURN MATERIAL AUTHORIZATION TERMS ScopeLED will only accept a return of products for which a ScopeLED Return Material Authorization RMA Number has been issued to Buyer prior to the shipment of the return products to ScopeLED This RMA Number must be displayed on all return shipment documents ScopeLED will refuse all returns that are not accompanied by an RMA Number All risks of any such refused shipment are the sole responsibility of Buyer For warranty returns ScopeLED will only accept return products accompanied by a statement of defects ScopeLED will not evaluate returns not including this information and such returns will be returned to Buyer at Buyer s expense Warranty retu
13. rns proved defective through damage mishandling disassembly modification or misuse by Buyer or any other person and warranty returns found non defective will be subject to evaluation and processing fees and repair costs if applicable Non warranty returns will be evaluated and will be subject to evaluation and processing fees If non warranty repair work is necessary Buyer will be notified of repair costs before a repair work order is initiated Confirming POs are required for non warranty repair work For warranty returns Buyer is responsible for one way freight costs to ScopeLED including any duty and taxes ScopeLED will cover freight costs for return shipment to Buyer Shipment charges billed to ScopeLED without prior approval from ScopeLED will be re invoiced to Buyer For non warranty returns Buyer is responsible for two way freight costs including any duty and taxes If shipment consists of returns that are both warranty and non warranty the shipment will be considered as non warranty ScopeLED will not accept Buyer generated debit memos All international return shipments to ScopeLED including packaging and airway bill must be marked Goods made in the United States enter as American Goods Returned AGR and state the reason for the return to the United States ScopeLED will refuse all returns that are not properly documented All risks of any such refused shipment are the sole responsibility of Buyer Interna
14. th the construction and operation of the product and hazards involved ScopeLED G150 Series Illuminator Overview ScopeLED G150 Series offers high brightness tunable color and adjustable intensity in a compact flexible and convenient package Different colored LED chips packed into DiCon s Dense Matrix LED Array couple light efficiently into a light guide illuminating your sample with selectable colors to enhance contrast protect wavelength sensitive samples and optimize viewing comfort The LEDs do not emit in the UV and IR spectral regions eliminating the need for special filters Undesirable light and sample interactions such as photobleaching in fluorescence and unintended curing in photosensitive adhesives can be avoided by selecting a color channel that doesn t interact with the sample The G150 Series combines optical power flexibility and usability in a compact form and is a great choice for your application ScopeLED G Series Family Product Feature Matrix Features Type ScopeLED G Series Ae G150 G180 G250 i Table Top x x ro muning Pole Mount x x x Controls Manual Control x x X USB Control x x x Tunable White 3 modes 6 modes LED Color RGB Single Color x x a San Page 1 of 18 Copyright2015 Product Overview SCOPELED G150 SERIES ILLUMINATOR Models amp Naming Convention The ScopeLED G150 Series has two different mounting types Table Top
15. tional returns should be sent via Federal Express UPS or DHL International returns may be processed using ScopeLED s brokerage EWI Inc 305 Harbor Way South San Francisco CA 94080 Contact David Li at TEL 650 794 1388 FAX 650 794 1389 If one of these carriers or ScopeLED s broker is not used ScopeLED may invoice Buyer for any additional costs including duty and taxes REVERSE ENGINEERING CONFIDENTIALITY Buyer shall not reverse engineer decompile disassemble modify reproduce or copy any products or any software within any products Buyer shall not analyze or identify the chemical composition or the physical characteristics of any products Buyer shall not furnish ScopeLED specifications to any other person SOFTWARE LICENSE ScopeLED does not transfer ownership of software contained in any products ScopeLED grants to Buyer a perpetual non exclusive license to use software in the operation of the product in which it is contained This license is transferable only with the transfer of ownership of the product SCOPELED G150 SERIES ILLUMINATOR TABLE OF CONTENTS PRODUCT OVERVIEW aan bela yake babek ka kaba k na kula ka kaka k RS kek u W R bu u kU R aku W waa W da 1 ScopeLED G150 Series Illuminator Overview cccccsceesseeeeeeeeeeeeeeeeeeeeeeeeeeeeeneeees 1 ScopeLED G Series Family Product Feature Matrix cccccceeeeeeeeeeeeeeeeeeeeeeeeeeees 1 Models amp Naming Convention X kak