Home
Lab User`s Manual (v. 4.0)
Contents
1. CHI Temperature sensor 1 2067 20 66 719 2068 20 67 71 9 2069 20 68 72 0 Junior student Own lab Activity Coach v6 1 2004 2006 CMA 008 Lab User s Guide 9 III TECHNICAL SPECIFICATIONS OF LAB 1 Connection to the computer LAB is connected to the PC via a USB port compatibility USB 1 0 USB 1 1 and USB 2 0 Max transfer rate 64 kB s 2 Power Supply LAB is powered from the USB power supply Because of this the quality of the 5V power supply for the sensors depends on the quality of the USB power supply and may differ from one computer to another 5V sensor 4 35 5 25V power supply quality dependent on host computer hardware Maximum At least 40 mA is available for sensor power current Up to 440 mA is normally available but dependent on host computer hardware and other USB devices connected to the host computer Power consumption with no sensors connected is max 60 mA 3 Analog to Digital Conversion LAB uses a 12 bit ADC The 12 bit resolution results in a voltage resolution of 1 2 mV for the 0 5 V range and 5 mV for the 10 10 V range The maximum sample rate depends on the number of used channels and the way the channels are used In normal mode the interface can take one sample every 100 us This means that one channel can be sampled at 10 kHz and two channels can be sampled at 5 kHz Sample intervals can be chosen in units of 100 us so possib
2. LAB INTERFACE 008 USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl For our international customers Check the CMA International website www cma science nl english for the latest version of this manual and updates of Coach software products Aan onze Nederlandse gebruikers Lab wordt standaard uitgeleverd met Engelse handleiding U kunt de laatste versie van de Nederlandse handleiding en de nieuwste update van Coach downloaden van de CMA website www cma science nl LAB User s Guide ver 4 0 December 2014 Author Ewa Kedzierska CMA Amsterdam 2014 2 008 LAB User s Guide TABLE OF CONTENTS I GETTING ACQUAINTED 4 1 Introduction 4 2 LAB components 4 3 LAB operating system 4 4 Computer and software requirements 5 5 Attaching LAB to the computer 5 6 Troubleshooting tips 5 7 Connecting sensors 5 Ps Vas OP sensor 5 Leh TUCO AUC SENSO VECO CIMCON lle ai 6 De OUNNUCT S Lenten eternet han nn deden 6 II USING LAB WITH THE COMPUTER 7 1 Working with Coach 6 7 2 Automatic sensor detection for Coach Activities with LAB 7 3 Example of a measurement with LAB and Coach 8 III TECHNICAL SPECIFICATIONS OF LAB 9 1 Connection to the computer 10 2 Power Supply 10 3 Analog to Digital Conversion 10 4 Memory 10 5 Inputs 11 6 Sensors which can be used with LAB 11 Ost Maximum power regu ements aranana eaa A E E E 11
3. OA USB Po werden 12 008 Lab User s Guide 3 I GETTING ACQUAINTED 1 Introduction The CMA LAB is a simple USB lab interface which can be used to introduce students to measurements with the computer The interface has two analog inputs to connect sensors with BT plugs A large number of sensors is available for LAB and can be used for measuring different physical quantities It is connected directly to the computer through a USB port and no extra power supply is needed To use LAB you need the Coach 6 Lite or Coach 6 software running on your computer 2 LAB components LAB is provided with A Analog BT input 1 B Analog BT input 2 C LED indicating interface initialization D USB cable LAB contains an embedded processor and built in memory enabling measurements at high sample rates up to 40 000 measurements per second with accurate timing independent of the host computer The FLASH memory allows easy upgrade of the internal system software 3 LAB operating system LAB is equipped with an internal operating software firmware which is stored in its FLASH memory The firmware update can be performed manually by using CMA 4 008 LAB User s Guide Firmware Update program or automatically via Coach 6 and Coach 6 Lite Coach recognizes automatically when a connected LAB needs to be updated and offers direct update 4 Computer and software requirements To use LAB with the com
4. 4 mm plugs via adapter art nr 0519 the standard range 0 to 5 V 6 should be selected pin 2 3 Auto ID I2C clock In this adapter there is no connection between 5 V DC the yellow lead and pin 1 Vin Vino Vres Output reference voltage pull up resistor 15 kQ for resistance measurements Auto ID pull up resistor 10 kQ Auto ID sensor detection input The BT inputs have the possibility for automatic sensor recognition For FC communication between LAB and intelligent sensors 6 Sensors which can be used with LAB In principle all sensors from CMA Vernier and TI can work with LAB but for a few sensors the accuracy might be degraded due to the quality of the USB power supply see details below Note that this does not depend on the LAB interface but on the host computer USB implementation and thus may differ from one computer to another 6 1 Maximum power requirements For some sensors complications can arise with the sensor power supply Computer USB ports are guaranteed to supply at least 100 mA per USB port most computers will supply up to 500 mA without problems Because the interface uses maximum 60 mA at least 40 mA is available to the sensors CMA sensors which continuously 008 Lab User s Guide 11 draw more than 40 mA are CO Sensor 0660i and CO Sensor 0661i For most computers this will not cause any problems but some computers may not provide enough power If this probl
5. 5 V power supply is not exactly 5 V but a bit lower or higher it will only influence the maximum measurement range of connected sensors For example if the 5 V supply is a bit lower than 5 V a Force Sensor that can normally measure forces up to 60 N will then only measure forces up to e g 57 N Although the maximum values of the range will be affected for most sensors measurements that are not at the maximum of the range will not be affected the output of most sensors is not dependent on the exact value of the 5 V supply in the normal measurement range and the calibrations will not be affected For some sensors this does not hold true and the output of these sensors IS affected by the exact value of the 5 V power supply These sensors will seem to function normally but the accuracy of these sensors may be degraded and or possible noise on the USB power supply can be propagated to the sensor output These are the following CMA sensors Angle Sensor 013i Charge Sensor 0361i Colorimeter 03581 Conductivity Sensor 0382 Dissolved Oxygen Sensor 0376 Gas Pressure Sensor 0341 only a very small influence Heart Rate Sensor 027i needs to readjust after a sudden fluctuation in the power supply but is not sensitive to the exact value of the power supply Magnetic Field Sensor 024 and Relative Humidity Sensor 0381 12 008 LAB User s Guide Warranty The LAB is warranted to be free from defects in materials and workmanship for a period
6. B screen panel e Select the Reset Hardware option This option is only available if the LAB panel is active in Coach 7 Connecting sensors 7 1 Analog sensors A wide range of sensors can be connected to LAB Most CMA Vernier and TI sensors are recognized by the interface Both LAB inputs have two measurement ranges 0 5 V and 10 10 V and can be used simultaneously Please note that due to USB power limitations some sensors 008 Lab User s Guide 5 will not work properly for details see 6 Sensors which can be used with LAB 7 2 Automatic sensor recognition LAB automatically identifies most CMA sensors when they are connected to LAB inputs LAB recognizes Intelligent and Auto id sensors e An Intelligent sensor has a memory chip with information about the sensor Through a simple protocol 1 C the sensor communicates with LAB and transfers its data name quantity unit and calibration to the interface and the Coach program e AnAuto id sensor is equipped with a specific sensor resistor which allows automatic identification of the sensor and using proper sensor data from the Coach Sensor Library When a sensor is physically plugged into an input of LAB then LAB automatically detects the sensor When LAB recognizes the connected sensor as an Intelligent sensor or Auto id sensor then the Coach program displays calibrated signal values measured by the sensor User can overrul
7. For detailed information consult Guide to Coach 6 Lite or Guide to Coach 6 manuals which can be downloaded at CMA website 6 008 LAB User s Guide II USING LAB WITH THE COMPUTER To use LAB with the computer you need a PC computer with the Coach 6 Lite or the Coach 6 program see 4 Computer and software requirements During the measurements LAB stays connected to the computer Sampled data are transferred in real time to the computer The course of the measurement can be followed directly on the computer screen in the Coach program 1 Working with Coach 6 When you work with Coach 6 Lite 6 2 or higher you can start the program by automatic interface recognition Plug your LAB into the USB port after the interface is detected Coach 6 Lite will start and the LAB project will open When you work with Coach 6 6 2 or higher then you have to open Coach via Windows Start menu or via double clicking a Coach Activity file cma a Coach Result file cmr or a Windows shortcut to such file Coach 6 Lite and Coach 6 offers a set of elementary Exploring Science Activities for LAB and its temperature light and sound sensors These are ready to go Activities in which students solve the given problems by taking measurements with sensors The Own LAB Activity included in the Coach project Own LAB allows students to setup and perform their own experiments with any supported sensors Extensive information on wor
8. e this setting by selecting another sensor from the Coach Sensor Library When LAB recognizes the connected sensor as another type of sensor then the Unknown sensor voltage icon appears and displays signal voltage values measured by the sensor To calibrate the measured signal the user has to select manually the proper sensor from the Coach Sensor Library See also Il 2 Automatic sensor detection for Coach Activities with LAB 7 3 Counters One input at a time of the LABAB interface can be used as a counter input If sensors which generate pulses such as the Geiger Muller radiation sensor CMA art nr 0661 or the Photogate with Pulley attachment CMA art nr 0662i are connected to these inputs pulse counting is automatically performed LAB has the possibility to adjust via Coach software sensor settings the counter threshold If a counter signal crosses this threshold in the predefined direction the counter is incremented by one Sometimes it is useful to count signal changes of an analog sensor as pulses e g to measure how often the screen is refreshed using a light sensor as counter or to measure the rate with which your heart beats using a heart beat sensor as a counter With LAB it is possible via Coach software sensor settings to define analog sensors as counters If the analog signal of such a sensor crosses a specified value threshold in a predefined direction the counter is incremented by one
9. em occurs a powered USB hub can be used Some sensors draw more than 40 MA for a very short time after plug in inrush current In most cases this will not cause any problems but with some sensors a reset of the interface will occur due to a voltage decrease or temporarily shut off of the USB power supply After the interface has reset the sensor and interface will again function normally The Coach software will automatically re establish the connection to the interface If this happens you may notice a delay of a number of seconds and the LED on the LAB interface will switch on for a short time when the interface resets The CMA sensors that are likely to cause a reset of the interface are Blood Pressure Sensor 0377i Colorimeter 03581 Force sensors 06631 Ion Specific Sensors 039x ORP Sensor 0313i and Spirometer 03781 Please note that this depends on the USB power supply and may differ from one computer to another on some computers a reset will not occur when one of the sensors from the list below is connected on some computers a reset may also occur when other sensors are connected 6 2 USB power quality Computer USB ports are only guaranteed to have the 5 V supply between 4 35 V and 5 25 V so the power supply does not always provide exactly 5 V Because the power supply for the sensors is directly derived from the USB power supply the sensor power supply quality will be comparable to the USB power supply quality Normally if the
10. he sensor are displayed on its icon 2 if there is already a normal sensor icon on this Panel input then nothing happens the normal sensor icon remains on the screen panel and calibrated signal values measured by the sensor are displayed on the icon When a sensor is physically disconnected from one of the inputs of LAB then depending on the current settings on the LAB screen panel Coach acts as follows usually within few seconds 1 if there is a detected sensor icon on this Panel input then the detected sensor icon disappears from the Panel 2 if there is a normal sensor icon on this Panel input then the normal sensor icon remains on the Panel without any change however the measured values are not displayed on the sensor icon anymore When the user tries to replace a detected sensor icon by dragging a normal sensor icon on it then the question Replace the currently connected sensor appears on the computer screen Clicking Yes replaces the detected sensor icon by the normal sensor icon 3 Example of a measurement with LAB and Coach Guided by this step by step tutorial you can perform your first temperature measurements with LAB Before you start make sure that the Coach 6 2 Lite software is installed on your computer In this experiment you are going to investigate temperature variations of water ina beaker During this measurement LAB collects temperature data and the measured data are displayed on the scree
11. king with Coach 6 Lite and Coach 6 programs can be found respectively in the Guide to Coach 6 Lite and Guide to Coach 6 manuals and in the on line Coach Help System 2 Automatic sensor detection for Coach Activities with LAB In Coach Activities with LAB two kinds of sensor icons can be used normal green and detected orange Normal sensor icons can be selected from the Coach Sensor library and placed on the sensor palette or on the LAB screen panel by the user Detected sensor icons are used for sensors automatically recognized by LAB These sensor icons exist only on the LAB screen panel and cannot be moved by the user When the mouse cursor is placed above a detected sensor icon then the tool tip hints that it is a detected sensor appears see the figure Temperature sensor 0511 CMA 25 150 C detected Temperature sensor 0511 CMA 25 150 C When a sensor is physically connected to one of the inputs of LAB then depending 008 Lab User s Guide 7 on the current settings on the LAB screen panel Coach acts as follows usually within few seconds 1 if there is no sensor icon on this Panel input then the sensor icon corresponding to the sensor detected by LAB appears on that Panel position For Intelligent and Auto id sensors calibrated signal values measured by the sensor are displayed on its icon for another type of sensors signal voltage values measured by t
12. le rates are 10 kHz 5 kHz 3 3 kHz 2 5 kHz 2 0 kHz 1 67 kHz etc In fast mode the interface takes one sample every 25 us sample rate 40 kHz During fast mode only one sensor actuator can be connected Counter operation is possible for one channel at a time The channel is monitored every 200 us to determine if the signal has passed the set threshold value up or down This means that the minimum width of a pulse that must be counted is 200 us The maximum frequency of a signal to be counted is 2 5 kHz in case of a symmetric signal Counter operation is not available in fast mode 4 Memory LAB is equipped with Flash non volatile memory which contains the internal operating system firmware Flash memory can be rewritten to update the firmware For buffering of sampled data LAB is equipped with 32 kB of RAM memory which can contain maximal 16 000 samples 10 008 LAB User s Guide 5 Sensor inputs Inputs 1 and 2 are used for analog sensors with right hand BT connectors with the following pin outs Analog Input 1 and 2 AI Note Te The BT plug has two signal pins Vin pin 1 and a 6 Vin low pin 6 The input range of pin 1 is 10V to a 10V and pin 6 is 0 to 5V oo In the Coach software it is possible to select the Right hand connector range i e select the pin of which signal will be taken see the Coach help keyword Input range Standard range is 0 to 5 V pin 6 For connected sensors with
13. measuring time is reached or by clicking the red Stop button Analyze the measured data All processing and analysis tools of Coach are available from the Diagram and Table Tool menus right click the diagram table window or click its Menu button f Coach 6 Lite Lite old Own Lab test cma N Program Window Help ELN EN NN CHI Temperature sensor mx 71 9 C Own Lab mx In this activity you can perform experiments with Lab You decide what kind of experiment you would like to set up and which sensor you will use 1 Connect a sensor to Lab Lab detects the sensor and its icon appears on the Lab screen panel Lab automatically recognizes most sensors if the sensor is not recognized the Unknown sensor voltage icon appears 2 You can overrule these settings by dragging a sensor icon from the sensor palette to one of input of the screen panel 3 Four sensor icons are already placed on the sensor palette To add a sensor icon to the palette click an empty place on the sensor palette click Select from Library select the desired sensor from the list and click OK 4 To present the measured data in graph table on meter or as digital value right click the sensor icon on the panel and select the way of presenting data and place it in one of the empty panes 5 Use the Seffings button to change the measuring time and trigger conditions 6 Jse the areen Sfar button to start a measurement Mi
14. n in a graph a table and as a digital value The measured value is also displayed on the sensor icon on the LAB screen panel Materials 1 LAB 2 PC computer with the Coach 6 2 Lite installed 3 CMA Temperature sensor art nr 0511 4 250 mL beaker and 100 mL hot water Measurement procedure 1 Connect LAB to the computer 2 After LAB is detected Coach 6 Lite will start automatically and the LAB 8 008 LAB User s Guide gt 10 11 12 13 14 project will open Open the project Own LAB and select the activity Own LAB Plug the CMA Temperature sensor CMA art Nr 0511 into input 1 of LAB LAB detects the temperature sensor and its icon appears on the LAB screen panel You can overrule these settings by dragging a sensor icon from the sensor palette to input 1 of the LAB screen panel Right click the icon and select Display as Diagram Place the diagram in one of the empty panes Right click the icon and select Display as Table Place the table in one of the empty panes Right click the icon and select Display as Value Place the value in one of the empty panes Determine the room temperature Filla beaker with hot water and place the temperature sensor in the water Click the Measurement settings button and set the measuring time to 3 minutes and measurement frequency to 50 per second Click the green Start button to start a measurement The measurement stops after the
15. of 12 months from the date of purchase provided that it has been used under normal laboratory conditions This warranty does not apply if LAB has been damaged by accident or misuse Note This product is to be used for educational purposes only It is not appropriate for industrial medical research or commercial applications Rev 10 12 2014 008 Lab User s Guide 13
16. puter you need Coach 6 Lite free or Coach 6 licensed version 6 2 or higher running on a Pentium IIl computer with Windows with at least 192 MB RAM recommended 256 MB and unused USB port Both programs can be downloaded via CMA website at www cma science nl 5 Attaching LAB to the computer To connect LAB to your computer simply plug its USB plug into a free USB port of the computer The LED of the interface will switch on for a short period of time This means that the interface initializes Windows might show a notice that a new USB device is found After LAB is detected and you have Coach 6 Lite installed on your computer then Coach 6 Lite will start automatically and the LAB project will open Automatic detection of LAB can be switched off by right clicking the Coach 6 icon in the taskbar and selecting Stop 6 Troubleshooting tips lt may happen that the communication between LAB and the computer is lost If this happens you have to reset LAB This can be done in two ways in the Coach software 1 When there is no communication on opening of a Coach Activity Results and the program displays the message that the panel LAB can not be initialized then unplug the interface and plug it in again after a few seconds or plug it into another USB port of your computer Press Retry in Coach 2 When the communication with the interface is lost while working in a Coach Activity Result then e Right click the LA