Contents

1. S9 technical specifications Operating pressure range Maximum single fault steady state pressure Pressure measurement tolerance Flow measurement tolerance 4 to 20 cm H20 30 cm H20 0 5 cm H20 4 of the measured reading 6 L min or 10 of reading whichever is greater DECLARED DUALNUMBER NOISE EMISSION VALUES in accordance with ISO 4871 Sound pressure level Sound power level Nominal dimensions L x W x H Weight Housing construction Supplemental oxygen Hypoallergenic air filter Standard air filter Air outlet H5i technical specifications Maximum heater plate temperature Temperature cut out Maximum gas temperature Nominal dimensions L x W x H Weight standard water tub Weight cleanable water tub Water capacity Docking station 24 dBA as measured according to ISO 17510 1 2002 26 dBA with uncertainty of 2 dBA as measured according o ISO 17510 1 2007 34 dBA with uncertainty of 2 dBA as measured according o ISO 17510 1 2007 6 0 x 5 5 x 3 4 153 mm x 140 mm x 86 mm 1 8 Ib 835 g Flame retardant engineering thermoplastic Recommended maximum supplemental oxygen flow 4 L min on woven acrylic and polypropylene fibers with polypropylene carrier Polyester fiber The 22 mm conical air outlet complies with ISO 5356 1 150 F 65 C 165 F 74 C lt 106 F lt 41 C Docking station and water tub 6 0 x 5 7 x 3 4 153 mm x 145 mm x 86 mm Docking station and unfilled water tub
2. Blocking the air tubing and or air inlet of the device while in operation could lead to overheating of the device Do not operate the H5i if it is not working properly or if any part of the device or H5i has been dropped or damaged A cautions Do not open the device enclosure There are no user serviceable parts inside Repairs and servicing should only be performed by an authorized ResMed service agent Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the device humidifier or air tubing These solutions may cause damage and reduce the life of these products ncorrect system setup may result in incorrect mask pressure reading Ensure the system is correctly set up Be careful not to place the device where it can be bumped or where someone is likely to trip over he power cord Make sure the area around the device is dry and clean and clear of bedding clothes or other objects hat could block the air inlet or cover the power supply unit Ensure that the device is protected against water if used outdoors Enclose the device in the S9 travel bag for transport The H5i should only be used with tubing or accessories recommended by ResMed Connection of other delivery tubes or accessories could result in injury or damage to the device Do not open the H5i enclosure There are no user serviceable parts inside Repairs and servicing should only be performed by an author
3. 1 52 Ib 0 69 kg Docking station and unfilled water tub 1 74 Ib 0 79 kg To maximum fill line 380 mL Flame retardant engineering thermoplastic aluminium uC Ai C C S H eanable water tub andard water tub r tubing technical specifications imateLine heated air tubing imateLineMAX heated air tubing SlimLine air tubing andard air tubing eated tubing temperature cut out Notes The manufacturer reserves the right to change these specifications without notice The temperature and relative humidity settings displayed for ClimateLine or ClimateLine measured values njection mo njection mo elastomer Flexible plas 6 6 2m 1 Flexible plas 6 3 1 9 m Flexible plas Flexible plas lt 106 F lt 4 ded plastic stainless steel and silicone seal ded plastic aluminium and thermoplastic ic and electrical components 5 mm inner diameter ic and electrical components 19 mm inner diameter ic 6 1 8 m 15 mm inner diameter ic 6 6 2 m 19 mm inner diameter G MAX are not Check with your clinician service provider before using the SlimLine air tubing with devices other than the S9 or H5i Humidifier performance The following settings have been tested at 22 C ambient temperature CPAP mask RH output Nominal system output AH BTPS pressure cm H20 Setting 3 Setting 6 Setting 3 Setting 6 4 90 100 10 18 10 95 100 11 5 21 20 95 100 11 18 a AH Abs
4. have removed the SD card before settings were copied to the CPAP device Ensure that the SD card is inserted correctly Reinsert the SD card and wait for the home screen or the Settings updated successfully press any key message to appear on the LCD Note This message only appears once If you re insert the SD card after you have updated your settings the message will not be re displayed The following message is NOT displayed on the LCD after you try to update the settings using the SD card Settings updated successfully press any key The settings were not updated Contact your clinician service provider immediately Water splashing on your face from the H5i The water tub is overfilled Condensation is forming in the air tube and mask Leaking water tub The water tub may be damaged or cracked The cleanable water tub is not assembled correctly Air feels too warm cold in the mask The temperature of the ClimateLine or ClimateLine air tubing is set too high low Check that the water level is below the maximum water level mark Turn the H5i or relative humidity setting down via the menu Contact your service provider for a replacement Check for damage and reassemble the cleanable water tub correctly Turn up down the heated tubing temperature via the menu General technical specifications 90W power supply unit 30W power supply unit Operating temperature Operating humidity Operating altit
5. RESMED S9 AutoSet S9 AutoSet for Her S9 Elite S9 Escape S9 Escape Auto POSITIVE AIRWAY PRESSURE DEVICES H5i H5i for Her HEATED HUMIDIFIER ClimateLine ClimateLine SlimLine Standard AIR TUBING Information Guide English Please read the entire Information and Welcome Guides before using your device A caution In the US Federal law restricts this device to sale by or on the order of a physician S9 AutoSet indications for use The S9 AutoSet self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 AutoSet self adjusting system is intended for home and hospital use S9 AutoSet for Her indications for use The S9 AutoSet for Her self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 AutoSet for Her self adjusting system is intended for home and hospital use S9 Elite indications for use The S9 Elite CPAP system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Elite CPAP system is intended for home and hospital use S9 Escape Auto indications for use The S9 Escape Auto self adjusting system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Escape Auto self adjusting s
6. ce The H5i is for use only as recommended by a physician The H5i for Her is intended for single patient re use in the home environment and re use in a hospital institutional environment H5i contraindications The H5i is contraindicated for use with patients whose upper Supraglottic airway has been bypassed Troubleshooting If there is a problem try the following suggestions If the problem cannot be solved contact your equipment supplier or ResMed Do not attempt to open these devices Problem possible cause Solution No display Power is not connected Ensure the power cable is connected and the power outlet if available is on The DC plug is partially inserted into the Fully insert the DC plug back of the device The S9 and H5i are not connected Ensure that the S9 and H5i are securely attached correctly Insufficient air delivered from the S9 device Ramp time is in use Wait for air pressure to build up or change ramp time Air filter is dirty Replace air filter Air tubing is not connected properly Check air tubing Air tubing is blocked pinched or Unblock or free the air tubing Check the air tubing for punctured punctures Mask and headgear are not positioned Adjust position of mask and headgear correctly The H5i flip lid is not latched correctly Close the flip lid ensuring that it clicks into place The H5i flip lid seal is not fitted correctly Make sure the flip lid seal is facing the right way up and fi
7. d Ltd 368875 1 2012 10 Global leaders in sleep and respiratory medicine www resmed com
8. ed air for longer than several minutes can in some circumstances lead to suffocation This applies to most models of CPAP or bilevel devices Oxygen supports combustion Oxygen must not be used while smoking or in the presence of an open flame 1 Ports may be incorporated into the mask or in connectors that are near the mask Always ensure that the device is turned on and airflow generated before the oxygen supply is turned on Always turn the oxygen supply off before the device is turned off so that unused oxygen does not accumulate within the device enclosure and create a risk of fire Do not leave long lengths of air tubing around the top of your bed It could twist around your head or neck while you are sleeping Do not use electrically conductive or antistatic air tubings Do not use the air tubing if there are any visible signs of damage Only ResMed air tubing and accessories should be used with the device A different type of air tubing or accessory may alter the pressure you actually receive reducing the effectiveness of the treatment Only use the ResMed 90W or 30W power supply unit Use the 90W power supply unit to power the system comprising the device H5i air tubing DC DC converter and battery pack The 30W power supply unit is designed to power the device only and recommended for traveling Only ResMed products are designed to be connected to the module connector port Connecting other devices could damage the device
9. f the device or the power supply are dropped or mishandled if water is spilled into the enclosure or if the enclosure is broken discontinue use and contact your ResMed Service Center Beware of electrocution Do not immerse the device humidifier power supply or power cord in water In the event of a spill disconnect the device from the power supply and let the parts dry Always unplug the device before cleaning and make sure that all parts are dry before plugging in the device Explosion hazard do not use in the vicinity of flammable anesthetics Make sure the power cord and plug are in good condition and the equipment is not damaged Keep the power cord away from hot surfaces The device should only be used with masks and connectors recommended by ResMed or by a physician or respiratory therapist A mask should not be used unless the device is turned on Once the mask is fitted ensure that the device is blowing air The vent hole or holes associated with the mask should never be blocked Explanation The device is intended to be used with special masks or connectors which have vent holes to allow continuous flow of air out of the mask When the device is turned on and functioning properly new air from the device flushes the exhaled air out through the mask vent holes However when the device is not operating insufficient fresh air will be provided through the mask and the exhaled air may be rebreathed Rebreathing of exhal
10. hen reconnect it to restart the device There is a fault in your ClimateLine or Discontinue using your ClimateLine or ClimateLine air ClimateLineMAX air tubing tubing and contact your clinician service provider Use Standard or SlimLine tubing in the interim Displays error message Humidifier fault replace humidifier Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and then reconnect it to restart the device There is a fault in your H5i heated Discontinue using your H5i and contact your clinician humidifier service provider Refilling the humidifier with cold water Allow the humidifier to cool before re filling while humidifier is still hot after therapy Problem possible cause Solution Filling the humidifier with ice cold water Use room temperature water on a warm day or with hot water Displays error message High temperature fault refer to user manual Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and then reconnect it to restart the device Air filter is blocked Replace your air filter Disconnect the power cord and then reconnect it to restart the device Air tubing is blocked Check your air tubing and remove any blockages Disconnect the power cord and then reconnect it to restart the device Humidifier setting is too high resulting Turn the humidifier setting down and empty the water from in accumulation of wa
11. ized ResMed service agent Do not overfill the water tub as water may enter the device and air tubing Do not use any additives eg scented oils and perfumes These may reduce the humidification output of the H5i and or cause deterioration of the water tub materials Take care when handling the H5i as the water water tub may be hot Allow 10 minutes for the heater plate and any excess water to cool The H5i should only be connected or disconnected when the water tub is empty Make sure that the water tub is empty before transporting the H5i Do not operate the H5i on an aircraft as water may enter the device and air tubing during turbulence Always place the H5i on a level surface below the level of the user to prevent the mask and tubing from filling with water If liquids are inadvertently spilled into or on the H5i unplug the device from the power outlet Disconnect the H5i from the device and allow it to drain and dry before re using ual Manufacturer ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Distributed by ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA EC REP ResMed UK Ltd 96 Milton Park Abingdon Oxfordshire 0X14 4RY UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip S9 H5i SlimLine ClimateLine SmartStart are trademarks of ResMed Ltd S9 and SmartStart are registered in U S Patent and Trademark Office 2012 ResMe
12. olute Humidity in mg L b BTPS Body Temperature Pressure Saturated Symbols The following symbols may appear on your S9 H5i power supply unit air tubing or packaging A Caution Read instructions before use AR Follow instructions for use IP21 Protection against insertion of fingers and against vertically dripping water P20 Not drip proof A Type BF equipment Class I equipment b Start Stop wal Manufacturer EC REP European Authorised Representative we European RoHS Batch code REF Catalogue number m gt M SN Serial number Direct current Not for use on more than one patient At Maximum water level Disinfectable up to 200 F 93 C Pater Use distilled or deionized water only a Yh Lock unlock Remove tub to fill Rx Only Prescription only In the US Federal law restricts these devices to sale by or on the order of a physician A Keep dry China pollution control logo 1 ORO China pollution control logo 2 Servicing The S9 and H5i devices are intended to provide safe and reliable operation when operated in accordance with the instructions provided by ResMed ResMed recommends that the S9 and H5i be inspected and serviced by an authorized ResMed Service Center if there is any sign of wear or concern with device function Otherwise service and inspection of the devices generally should not be required during the fi
13. ter in the air the air tubing tubing Displays error message Check ResMed 30 90W Power Supply Unit and fully insert the connector The DC plug is partially inserted into the Fully insert the DC plug back of the device or inserted too slowly A non ResMed power supply unit is Remove the power supply unit and replace with a ResMed connected to the device power supply unit The power supply unit is being covered Make sure that the power supply unit is free from bedding by bedding clothes or other objects that could cover it Displays error message Tube blocked please check your tube Air tubing is blocked Check your air tubing and remove any blockages Displays error message High leak please check system setup and all connections There is excessive leak Adjust position of mask and headgear Air tubing is not connected properly Connect firmly at both ends The H5i flip lid is not latched correctly Close the flip lid ensuring that it clicks into place Displays error message Please close H5i flip lid attach tube and press any key H5i flip lid is not closed Close the flip lid ensuring that it clicks into place Problem possible cause Air tubing is not connected properly Solution Connect firmly at both ends The following message is displayed on the LCD after you try to update settings or copy data to the SD card Card error please remove SD card and contact service provider SD card is not inserted correctly You may
14. tted securely Incorrect air tubing selected If you are using the SlimLine or Standard air tubing ensure that you have the correct air tubing selected via the menu Device does not start when you breathe into the mask Breath is not deep enough to trigger Take a deep breath in and out through the mask SmartStart Stop There is excessive leak Adjust position of mask and headgear Air tubing not connected properly Connect firmly at both ends Problem possible cause Solution SmartStart Stop is disabled Enable SmartStart Stop Note SmartStart Stop is not available in S9 Escape and S9 Escape Auto Device does not stop when you remove your mask SmartStart Stop is disabled Enable SmartStart Stop Note SmartStart Stop is not available in S9 Escape and S9 Escape Auto SmartStart Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system being used Only use equipment recommended by ResMed The patient is using a nasal pillows Disable SmartStart Stop mask with a set pressure less than 7 cm H20 Pressure rises inappropriately Talking coughing or breathing in an Avoid talking with a nasal mask on and breathe as normally unusual manner as possible Mask cushion is buzzing against the Adjust the headgear skin Displays error message Heated tube fault replace tube Device has been left in a hot Allow to cool before re use Disconnect the power cord environment and t
15. ude Storage and transport temperature Storage and transport humidity Electromagnetic compatibility Aircraft use IEC 60601 1 classification Input range 100 240V 50 60HZz Nominal for aircraft use 115V 400Hz Typical power consumption 70W 80VA Maximum power consumption 110W 120VA Input range 100 240V 50 60HZz Nominal for aircraft use 115V 400Hz Typical power consumption 20W 40VA Maximum power consumption 36W 75VA 41 F to 95 F 5 C to 35 C Note The air flow for breathing produced by this therapy device can be higher than the temperature of the room Under extreme ambient temperature conditions 104 F 40 C the device remains safe 10 95 non condensing Sea level to 8 500 2 591 m 4 F to 140 F 20 C to 60 C 10 95 non condensing Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 for residential commercial and light industry environments It is recommended that mobile communication devices are kept at least 1 m away from the device Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found on www resmed com on the Products page under Service and Support Click on the PDF file for your language ResMed confirms that the S9 Series meets the Federal Aviation Administration FAA requirements RTCA DO 160 for all phases of air travel Class II double insulation Type BF
16. ue to cigarette pipe cigar or other smoke Warranty is void on product sold or resold outside the region of original purchase Warranty claims on defective product must be made by the initial consumer at the point of purchase This warranty replaces all other expressed or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from region to region For further information on your warranty rights contact your local ResMed dealer or ResMed office A WARNINGS Read the entire manual before using the device Use the device only as directed by your physician or healthcare provider e Use the device only for the intended use as described in this manual Advice contained in this manual should not supersede instructions given by the prescribing physician If you notice any unexplained changes in the performance of the device if it is making unusual or harsh sounds i
17. ve year design life of the device Limited warranty ResMed Ltd hereafter ResMed warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period e Mask systems including mask frame cushion headgear and tubing 90 days excluding single use devices e Accessories excluding single use devices e Flextype finger pulse sensors umidifier water tubs w IT atteries for use in ResMed internal and external battery systems 6 months ip type finger pulse sensors 1 year PAP and bilevel device data modules ximeters and CPAP and bilevel device oximeter adapters umidifier cleanable water tubs e Titration control devices TOG e CPAP bilevel and ventilation devices including external power supply units 2 years e Battery accessories e Portable diagnostic screening devices e Humidifiers This warranty is only available to the initial consumer It is not transferable If the product fails under conditions of normal use ResMed will repair or replace at its option the defective product or any of its components This limited warranty does not cover a any damage caused as a result of improper use abuse modification or alteration of the product b repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs and c any damage or contamination d
18. ystem is intended for home and hospital use S9 Escape indications for use The S9 Escape CPAP system is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 66 Ib gt 30 kg The S9 Escape CPAP system is intended for home and hospital use S9 contraindications Positive airway pressure therapy may be contraindicated in some patients with the following pre existing conditions severe bullous lung disease pneumothorax pathologically low blood pressure dehydration cerebrospinal fluid leak recent cranial surgery or trauma S9 adverse effects Patients should report unusual chest pain severe headache or increased breathlessness to their prescribing physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during the course of therapy with these devices drying of the nose mouth or throat nosebleed bloating ear or sinus discomfort eye irritation e skin rashes H5i indications for use The H5i is indicated for the humidification of the air delivered from a CPAP or bilevel device The H5i is for use only as recommended by a physician The H5i is intended for single patient re use in the home environment and re use in a hospital institutional environment H5i for Her indications for use The H5i for Her is indicated for the humidification of the air delivered from a CPAP or bilevel devi