FreeZone Freeze Dry Systems - Installation Qualification
Contents
1. IQ OQ Protocol Installation Qualification Operation Qualification FreeZone Freeze Dry Systems To be used with some FreeZone Systems manufactured prior to August 2004 see model number chart inside Labconco No 1059500 Rev A ECO D903 Available at www labconco com or by e mail in Word 2000 document Protecting your y aN laboratory environment LABCONCO Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A Purpose and Scope IQ and OQ This Qualification Protocol is solely intended to be used with Labconco FreeZone Freeze Dry Systems which are new or relocated FreeZone Tray Dryers are covered in separate documents 1058901 and 1059501 Design changes occurred in 2004 which impact which IQ OQ protocol is to be used with your Freeze Dryer Check the model number listed on the serial number tag located on the rear of the equipment Models using this protocol can be visually identified by the blue control panel label and stainless steel front Prior models have a dark gray almost black control panel and a painted front panel Use document 1058900 to perform the IQ OQ on previous models Models FreeZone Freeze Dry Systems 7420020 7740020 7752020 7753522 7753020 7934020 7960030 7420021 7740021 7752021 7753524 7753021 7934021 7960031 7420030 7740030 7752030 7753532 7753022 7934022
2. 7960032 7420031 7740031 7752031 7753534 7753024 7934024 7960034 7420040 7740040 7752040 7753542 7753026 7934026 7960036 7420041 7740041 7752041 7753544 7753027 7934027 7960037 7753030 7934030 7960040 7670020 7750020 7753031 7934031 7960041 7670021 7750021 7754030 7755030 7753032 7934032 7960042 7670030 7750030 7754031 7755031 7753034 7934034 7960044 7670031 7750031 7754032 7755032 7753036 7934036 7960046 7670040 7750040 7754034 7755034 7753037 7934037 7960047 7670041 7750041 7754036 7755036 7753040 7934040 7754037 7755037 7753041 7934041 7670520 7751020 7754040 7755040 7753042 7934042 7670521 7751021 7754041 7755041 7753044 7934044 7670530 7751030 7754042 7755042 7753046 7934046 7670531 7751031 7754044 7755044 7753047 7934047 7670540 7751040 7754046 7755046 7670541 7751041 7754047 7755047 It is written to assist the end user in validation of predetermined specifications The protocol begins with planning the site for the piece of equipment and therefore is of value prior to receipt of delivery The use of this document does not replace the need for the FreeZone Freeze Dry Systems User s Manuals Information within the User s Manual is required to complete this IQ OQ Protocol If the manual has been misplaced copies can be obtained from the manufacturer or down loaded from their website www labconco com 1 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A Re
3. electrical requirements Electrical Has the ground on the electrical service been Y N Grounding verified B Operational Qualification Step Description Specification or Acceptance Criteria Result YES NO 1 Freeze Dryer la Automatic With the freeze dryer system at ambient Y N Mode temperature turn the Main Power Switch ON Press the button labeled Refrigeration Auto Does the refrigeration system start Record the time it started lb Power to For systems without the Purge Valve option Y N Vacuum Pump when the freeze dryer collector temperature reaches 40 C the vacuum pump should start When the display shows 40 C did the pump start Or for systems with the Purge Valve option the vacuum pump should start 2 seconds after pressing the Refrigeration Auto switch Did the pump start lc Purge Valve If equipped with the Purge Valve option the Y N Option Purge Valve should remain closed until the collector temperature reaches 40 C Does the N A system switch the purge valve at 40 C 5 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A ld Refrigeration Effectiveness For FreeZone Freeze Dryers does the collector temperature reach 47 C in less than 40 minutes With the system under vacuum and 21 C ambient temperature Or for FreeZone Plus Cas
4. cade Freeze Dryers does the collector temperature reach 81 C in less than 40 minutes With the system under vacuum and 21 C ambient temperature NOTE Freeze Dryers are tested to 50 C or 84 C in the factory Conditions may vary in the field resulting in warmer acceptance temperatures Y N le Verify Displayed Temperature The temperature indicated on the LCD display is measured at the outlet of the collector coil The value was calibrated at the factory by correlating its reading with that of a reference gauge attached to a thermocouple The T type red amp blue thermocouple can be accessed outside the bottom of the insulated collector chamber Does the display correlate to the reference gauge meter 2 C Ref Instrument If the temperature does not correlate contact Labconco Product Service for calibration procedure N A If Temperature Wave Display The temperature wave graph of indicator lights on the control panel is a quick reference of the collector temperature Verify these lights are operating properly Note 1L models are not equipped with indicator lights Does the last green light illuminate when the system reaches lt 39 C N A 6 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A lg Vacuum Leaks Verify that the system is leak free by continuously run
5. ference gauge The calibration was performed at a very low level approximately 0 010 mBar NOTICE Factory calibration was performed using a precision Active Piranni Gauge calibrated to a Capacitance Manometer standard Despite the system s calibration and repeatability the readings taken at such a low level of vacuum should only be considered as a verification of a leak free system Vacuum swings can be attributed to contamination of surfaces which could take days to outgas Adjustments are discouraged Before any adjustments are made to the factory calibration of the vacuum measurement answer positively to these questions 1 Is the vacuum standard being used to verify the freeze dryer accepted by the organization to be precise and has it been calibrated recently 2 Is the level of accuracy we are attempting to reach pertinent to the freeze drying applications Does the vacuum display correlate to the reference gauge Ref Instrument If vacuum is to be calibrated contact Labconco Product Service for calibration procedure N A lj Defrost All 6 12 and 18L console models Check defroster operation With the refrigeration switches OFF press the defrost button ON Does the collector coil become warm to the touch N A 8 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A 1k Shell Free
6. modations been made to provide a suitable vacuum pump It must be capable of at least 144 Liters min free flow for 6 12 and 18 Liter models at least 86 Liters min free flow for 1 2 5 and 4 5 Liter models 0 0002 mBar ultimate vacuum an inlet fitting suitable for a 34 inch ID vacuum hose and the same voltage rating as the freeze dryer An oil mist eliminator exhaust filter is recommended Pumps operating with 115V Freeze Dryers should have a NEMA 5 15P plug Pumps operating with 230V Freeze Dryers should have a reverse IEC 320 plug 3 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A ld Manifolds With the exception of the 4 5 Liter Benchtop model freeze drying chambers or manifolds are not included with the freeze dryer Has a sample manifold or chamber been purchased for this application Y N N A Prior to Operation 2a Damage Claims Have the delivered products been inspected for any signs of damage that may have occurred while in transit Keep packaging materials until inspection is complete WARNING Do not attempt to pull a vacuum on a freeze dryer with any damage to the clear lids or stainless steel manifolds chambers Implosion and potential for injury can occur If damaged refer to the User s Manual for information on shipping damage claims 2b 2c Teflon Coated Handling S
7. ning the vacuum pump with the refrigeration system ON The rate the freeze dryer without samples achieves a low level of vacuum less than 0 133 mBar is dependent upon many factors Inside volume amp surface area of the system Cleanliness or cleaners used on interior Condition amp size of the vacuum pump Period of time the parts have been exposed to environmental conditions Based on the freeze dryer s displayed vacuum level the freeze dryer should reach its lowest level in less than 18 hours If not refer to Vacuum Troubleshooting Guide in the User s Manual Does the system reach a displayed vacuum level of less than 0 050 mBar in 18 hours lh Vacuum Wave Display The control panel s vacuum wave graph of indicator lights is a quick reference of vacuum level Verify these lights are operating Note 1L models are not equipped with these indicator lights Does the last green light flash when the system reaches lt 0 450 mBar And does the last green light remain illuminated when the system reaches lt 0 133 mBar N A 7 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A li Verify Displayed Vacuum The vacuum level indicated on the LCD display is measured between the collector and the vacuum pump The value was calibrated at the factory by correlating its reading with that of a re
8. nnot recommend specific procedures to do this The performance validation should be designed to meet the specifications and accuracy required of the application In general this requires establishing acceptance criteria making several runs and testing the results with calibrated equipment and qualified personnel 2 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A A Installation Qualification Step Description Specification or Acceptance Criteria Result YES NO 1 Site Planning la Space Refer to Appendix B in User s Manual for Y N Requirements dimensions of the model you have chosen Has adequate floor or counter space been provided for placement of the equipment A minimum of 3 inches is required between the back of a freeze dryer and the wall and between the sides and the walls for proper airflow through the refrigeration system lb Electrical Refer to the User s Manual for electrical Y N Service requirements Are services available for the equipment to be connected to an electrical circuit of adequate size and the proper voltage 115V models are shipped with a NEMA 5 20P Y N power cord plug 230V models are shipped with a NEMA 6 20P power cord plug If these are N A not suitable is one available to match the service outlet at the installation site Ic Vacuum Pump Refer to the User s Manual Have Y N accom
9. olvents If aggressive acids are to be used within this freeze dryer Teflon coated collector chamber is recommended Have the use of acids and potential for damage been considered Has the Safety Officer or equivalent reviewed the safe handling and disposal of solvents trapped as well as used vacuum pump oil N A N A 2d Manifold Installation Place the desired drying chamber or manifold onto the 3 inch diameter port located on the top of the freeze dryer 4 5 L models have a drying chamber permanently installed Larger drying chambers are to be held in place with a stainless steel pin provided N A 2e Vacuum Pump Installation Install the vacuum pump per the User s Manual Is the pump attached to the vacuum port on the collector chamber with the large rubber hose and clamps provided Vacuum Pump Electrical Vacuum Ballast Is the vacuum pump plugged into the rear of the freeze dryer and the pump s power switch turned to the ON position The vacuum pump s ballast should remain closed Is the pump s ballast closed 4 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A 2f Electrical Plug the freeze dryer into a dedicated electrical Y N Connection outlet Has the electrical service been verified to be adequate in size and voltage The ID plate on the rear of the freeze dryer has the
10. r the bottom Turn refrigeration and vacuum on The refrigeration and vacuum should be inoperable Are they disabled N A N A Does the display read Moisture in Collector Y N A 9 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A Routine Maintenance Below are helpful hints to be included in the organization s preventive maintenance plan 2a 2b Vacuum Grease Vacuum Pump Oil Vacuum grease should be applied to rubber components as required In general vacuum grease should be the first step in troubleshooting vacuum leaks Thin layers of grease are adequate for all seals Only use grease specially formulated for low vacuum service Do not use grease on the flat gaskets used to seal freeze dry valves to the chamber Is vacuum grease readily available and documented Type of grease used The vacuum pump oil should be changed as needed Change oil that appears cloudy or discolored At a minimum oil should be changed every 1000 hours of service An alarm can be set on the freeze dryer to alert personnel when 1000 hours of operation has been reached Has there been a preventive maintenance plan established for the vacuum pump Type of oil to be used 2c Inspect for Wear amp Damage Is there a procedure to periodically inspect the interior metal surfaces for corro
11. s Y N Manual for the machine 11 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A C Summary Labconco FreeZone Freeze Dry System IQ OQ Document 1059500 Revision Equipment Location FreeZone Ser No Model No User Protocol Revision or Date published Contact print name Title Review the Response columns for answers of NO Use the area below to describe the deficiency or unacceptable results Those deficiencies are to be followed with an instruction for Corrective Actions Once acceptable results are obtained the deficiency is accepted by initialing the Corrective Action Step Deficiency followed by Corrective Action Initial 12 Labconco Product Service 1 800 522 7658 1 816 333 8811
12. sion due to acids 10 Labconco Product Service 1 800 522 7658 1 816 333 8811 Labconco FreeZone Freeze Dry Systems IQ OQ Protocol 1059500 Revision A Is there awareness and a maintenance procedure to check the clear acrylic and glass parts for chips cracks deep scratches or Y N chemical attack WARNING This is a safety issue Implosion can occur with damaged or corroded components Will all the rubber components be periodically Y N inspected so that they are free from drying cracks or deterioration 2d Refrigeration At least annually will the refrigeration Y N System condenser be cleaned of dust that would restrict Cleaning free airflow include in the preventive maintenance schedule 3 Personnel Training 3a User Training Have personnel that will use the FreeZone Related to Freeze Dry System been adequately trained Equipment Are personnel familiar with All the buttons and displays on the front Y N Collector capacity limits before defrosting Defrosting and draining methods Safe handling of solvents drained The various alarms on the control panel The use of vacuum grease Opening closing and venting sample valves Setting the vacuum level Setting the Vacuum Chamber temperature Using the Shell Freezer Cleaning of the freeze dryer and neutralization of acids 3b User s Manual Are the personnel who are to use or maintain the Tray Dryers able to locate the User
13. sponsibilities End User The ultimate user or otherwise appointed personnel in the lab is responsible to ensure the freeze dryer is installed and operating properly This document can assist in that validation This document cannot however anticipate every application or unique situation encountered with the installation and operation It is therefore essential that users lab managers and safety officers work together to broaden the scope of this document through careful forethought End User Employer The employer is responsible for supporting the validation through adequate resources and training The organization shall also ensure the validation process has been fully carried out prior to applying the freeze dryer Records should be stored in a safe easily retrievable location The location of the equipment and required validation should be included in the company s quality system Manufacturer Labconco Corporation certified ISO 9001 is responsible to fully test each FreeZone Freeze Dry System prior to shipment The manufacturer must retain these records Labconco s staff of Product Service Representatives and Product Specialists can assist with information on the purchase delivery and installation Labconco is not responsible for the actual installation or validation processes Performance Qualification Once the freeze dryer has been checked for proper installation and operation its performance can be validated Labconco ca
14. zer Option If the freeze dryer is equipped with a Shell Freezer check its operation Press the Shell Freezer button Do the rollers turn Y N A N Does the bath temperature shown on the display reach 40 C in 2 hours N A Verify the display reading by placing a thermocouple inside the center of the bath almost touching the bottom of the chamber Does the display correlate with the reference gauge meter 2 C Ref Instrument N A 11 Vacuum Control All models except 1L Check vacuum control operation With refrigeration running at lt 40 C and vacuum operating at lt 075 mBar set the vacuum level at 0 120 mBar Within a 5 minute period the vacuum shown on the display should not vary beyond 0 100 to 0 160 mBar Are these readings achieved N A lm In Vacuum Chamber Option Moisture Sensor Some 6 12 and 18L console freeze dryers have a Vacuum Chamber built into the top surface next to collector chamber If the freeze dryer is equipped with a Vacuum Chamber check its operation Securely tape a thermocouple to the center of the chamber Set the Vacuum Chamber temperature for 40 C Does the display correlate with the reference gauge meter 3 C over a 10 minute period Ref Instrument All models except 1L are equipped with a moisture sensor Check moisture sensor operation Turn refrigeration and vacuum pump off Pour water into the collector to cove