Home

6300 laser manual H3.qxd

image

Contents

1. Use IBWRT to send the WAVELENGTH command to the controller Example IBWRI nflaser WAVE LENGTH 772 0 The STR function converts numbers to strings CALL IBWRT nflaser WAVELENGTH STRS wave length Read the output buffer CALL IBRD nflaser buff Print the contents of the output buffer OK if the operation was successful Out of Range if wavelength is not valid etc PRINT PRINT buff Return to local control CALL IBONL nflaser 0 END 33 34 Example 2 This is the same as the last program except that this time we use RS 232 communications instead of GPIB Ask user for a wavelength in nanometers wave length is a single precision floating point number 775 1 for example INPUT What wavelength would you like nm wave length RS 232 commands should be preceded by e MESSAGES WAVE STRS wavelength Open the serial communications RS 232 port with baud rate 9600 8 data bits no parity and 1 stop bit OPEN COM2 9600 N 8 1 ASC CD0O CS0 DS0 0P500 RS TB512 RB512 FOR RANDOM AS 1 Send the message to the controller PRINT 1 MESSAGES Read and print the response from the controller LINE INPUT 1 RESPONSES PRINT RESPONSES Close the port CLOSE 1 END Example 3 This program sets up and executes a wavelength scan using GPIB B
2. Specifications are typical for each model Contact New Focus for details 88
3. WARNING THIS IS A DIRECT RF CONNECTION TO THE DIODE LASER AND IMPROPER USE COULD DESTROY THE DIODE LASER DO NOT EXCEED 1 VOLT PEAK TO PEAK INPUT AVOID STATIC DISCHARGE Since this is a direct connection to the diode laser there is no current limiter that prevents too much current from passing through the diode Thus there is a danger of destroying the diode laser if the maximum current is exceeded Be sure that the current modulation applied to the diode laser will not result in driving the diode laser above its maximum current See included Data Sheet The high speed input has a protective shorting cap which should remain in place in normal operation When removing the shorting cap to connect a high frequency driver be sure to wear a ground strap to prevent static discharge Please contact New Focus if you have any questions on how to use the high speed modulation input Make sure to turn off power to the diode before con necting any device to the high speed input Wavelength Output The Wavelength Output BNC jack provides a signal from 0 to 10 V that corresponds linearly to the laser output wavelength This connection has 3 kQ output impedance Zero volts are output at the minimum operating wavelength for the laser head connected to the controller The voltage increases with wavelength according to the equation V f AA where AA is in nanometers and the parameter f depends on the particular laser center wavelength Typica
4. 120 V AC 2 0 A 239 002 220 V AC 1 6A 239 01 6 240 V AC 1 6A 239 01 6 X Service and Support Warranty New Focus Inc guarantees its lasers to be free of defects for one year from the date of shipment or for 3000 hours of operation whichever comes first This is in lieu of all other guarantees expressed or implied and does not cover incidental or consequential loss Service and Calibration Your 6300 series laser has been designed to provide years of trouble free operation Virtually no maintenance is required except for ensuring that the unit is not damaged contaminated or used in an unsafe manner To ensure lt 1 nanometer accuracy in wavelength readout the system should be recalibrated every 6 months Any questions regarding the operation or performance of the laser will be gladly answered by New Focus engineers Engineers are on duty from 8 00 5 00 PST Monday through Friday exclud ing holidays For service repair or calibration please call for a return autho rization number before shipping the unit to New Focus You can reach us at NEW FOCUS Inc 5215 Hellyer Avenue San Jose CA 951385 USA Phone 408 284 6808 Fax 408 284 4827 Email techsupport newfocus com Internet www newfocus com 87 XI Specifications Model 6304 6308 6312 6316 6320 uning Range 632 637 nm 668 678 nm 171 789 nm 834 848 nm
5. Note that an unconnected wavelength input corresponds to zero volts and the minimum wavelength If the maximum voltage is exceeded the diode laser will tune to its maximum wavelength and stop Then the wavelength readout will flash indicating you have exceeded the maximum input voltage The laser will then be kicked out of Wavelength Input Mode into Track Mode The same thing happens if you input a voltage less than the voltage corresponding to the minimum wavelength To stop the flashing either hit the local button or use the wavelength control knob to shift the wavelength a few angstroms off the edge To exit Wavelength Input Mode push the two paddle switches down at the same time The LED will stop flashing the wavelength will remain where it is and the controller will return to Track Mode Note that the wavelength control knob is disabled while the laser is in Wavelength Input Mode Also if you exit Track Mode and then re enter the laser will still be in Wavelength Input Mode 27 28 Detector Input The Detector Input is a 10 bit analog to digital converter input to the micro processor Analog signals that you connect here can be read out by your comput er when you use remote control This is a general purpose input that allows you to collect data from a photodetector during a wavelength scan or for example to create your own constant power mode with an external detector The Detector Input will accept signals from 0 to 5
6. Part of the light 10 Figure 1 New Focus laser cavity DC Motor Angle Sensor gt Output Diffraction Beam AR Grating Control In _ gt Coating L from the diode is reflected not diffracted by the grating This portion forms the output beam The grating functions as a narrow spectral filter Its passband is only a few gigahertz wide The high wavelength selectivity results because many lines of the grating are illuminated by the grazing incidence beam and because the beam is diffracted by the grating twice in each round trip through the cavity The grating spectral filter is narrow enough to force the laser to operate in a single longitudinal mode Different wavelengths diffract off the grating at different angles However only one wavelength leaves the grating in a direction that is exactly perpendicular to the surface of the tuning mirror forming the resonant laser cavity This is the lasing wavelength because it s the only one that will survive for many cavi ty round trips It follows then that we can tune the laser by changing the angle of the tuning mirror There is one very important innovation that gives us true continuous tuning with a minimum of mode hops In order to stay in one mode as we tune the laser the number of waves in the cavity must be kept constant even though the wavelength of the light in the cavity is changing The number of waves in the cavity is maintained by
7. C Controller Revision Model indicates the type of laser 6304 6328 etc Head Revision and Controller Revision indicate versions of software They will not necessarily be the same IDN gt NEW FOCUS 6304 H2 00 C2 01 Command Description Explanation Example OPC Operation Complete Query Returns 0 if a long term operation is ongoing Returns 1 if no long term operation is ongoing The laser has three long term operations Scanning see OUTPut SGAN STARt Resetting see OUTPut SCAN RESEt Wavelength Set see SOURce WAVElength Starting any of the above three operations clears the OPC status bit When the laser reaches the target wavelength the OPC status bit is set to 1 The bit is also set to 1 if the movement is interrupted either with SCAN STOP which interrupts a SCAN or RESEt or OUTPut TRACK OFF which interrupts a Wavelength Set During a wavelength scan OPC gt 0 and when the scan is completed XOPC gt 1 45 Output Commands Command Description Explanation Example 46 OUTPut SCAN RESEt Return to start wavelength The wavelength is reset to the start wavelength at the return slew rate If a scan is in progress it will be interrupted and the wavelength reset to the start wavelength Works the same whether laser is in Remote Mode or Local Mode The OPC command will return 0 until the start wavelength is reached The current w
8. FOCUS 6304 H2 00 C2 01 The IDN is the command that is sent to the controller either over GPIB or RS 232 The arrow symbol gt indicates the response from the controller which in this example is NEW FOCUS 6304 H2 00 C2 01 Types of Commands There are six general types of commands Standard Commands Output Commands Sense Commands Source Controls Source Queries and System Commands Standard Commands allow you to get information about the laser head and to query whether an operation has been completed Output Commands tell the laser to start some action such as starting or stopping a scan or reset System Commands set the operating mode of the controller Source Controls set laser operating parameters These parameters fall into two categories Those that use the internal digital to analog converters DACs and those that do not The piezo voltage the diode temperature and the laser diode current can be controlled either from the front panel with potentiometers or from voltages generated by internal DACs The other parameters wavelength scan speeds and scan start and stop wavelengths are independent of whether the potentiometers or the DACs have control When the front panel potentiometers have control the laser is considered to be in Local Mode When the internal DACs have control the laser is considered to be in Remote Mode Setting any one of the three DAC parameters forces the other two under DAC control as well
9. LED comes on e Control of the laser is given to the DACs e The diode temperature set point control DAC is driven to a voltage which produces a current within 3 of the value it had under front panel control e The current control DAC and the piezo control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control The laser is now in Remote Mode e Returns the target diode temperature set point which is the front panel temperature set point at the moment the query occurred 65 66 Example Front panel shows the following e Remote LED off Local Mode e Piezo voltage reads 30 1 e Diode temperature set point reads 20 0 e Diode current reads 59 2 TEMP 22 1 gt OK Front panel now shows e Remote LED on Remote Mode e Piezo Voltage reads 29 6 e Diode Temp set point reads 21 8 e Diode Current reads 58 4 SENSE TEMP SETP DIODE gt 21 8 returns the actual temperature set point TEMP 2e returns the target set point The diode temperature is changed until it becomes 21 8 C Command Control Form Query Form Description Explanation Control Form SOURce VOLTage LEVel PIEZo lt wsp gt lt value gt SOURce VOLTage LEVel PIEZo Write and read the voltage level of the piezoelectric transducer 0 lt value lt 100 units are in percentage of full scale This command forces the laser int
10. a program To read the controller s response to a command use the IBRD function call For RS 232 operation use to get the attention of the controller before starting command entry and end each command with a carriage return lt cr gt The controller s response is automatically sent The portions of commands in square brackets are optional and can be omitted without affecting the command Also the lower case parts of commands may be omitted lt wsp gt means white space either a tab or a space lt value gt is a floating point number OFFIONIOl1 means OFF or ON or 1 or 0 is legal Multiple commands on a line are not allowed For instance if the controller receives WAVE 670 22 IDN it will change the wavelength to 670 22 but the Identification Query will be ignored Numbers may contain at most 15 characters In most cases however this constraint will not be an issue The number 670 22 will be read correctly but the number 0000000000000670 22 will be read as 67 The examples given on the following pages show the text to be included in the command and the response from the controller With RS 232 serial communications the response is sent immediately and the user may process or ignore the data With GPIB communications the response is loaded into the output buffer and the user may read it with a GPIB read or else ignore it For instance the example for IDN appears as follows 37 38 TDN gt NEW
11. e Wavelength movement stops at current value e Encoder can now change wavelength e Scan button is active Command Description Explanation SYSTem MCONtrol lt wsp gt lt INT EXT gt Puts controller into Local Mode external or Remote Mode internal Forces the laser to use internal DACs or front panel potentiometers for control of piezo voltage diode temperature set point and diode current EXT If the laser is already in Local Mode e Command has no effect If the laser is not in Local Mode e Control of diode temperature set point diode current and piezo voltage is returned to the front panel potentiometers e NOTE The pots are absolute not relative so if they have changed the values will reflect that change e Constant Power Mode is disabled INT If the laser is already in Remote Mode e Command has no effect If the laser is not in Remote Mode e Diode temperature set point diode current and piezo voltage are sensed and used as targets for the DACs e Front panel displays will change as sensed and target values can differ by up to 3 due to component variations such as resistor tolerances e Note that in Remote Mode the front panel potentiometers are ignored but they are still active If a potentiometer is changed to an extreme setting and the controller is then switched back to external control the corresponding parameter will then move to the extreme setting 83 84 Vill Error Codes
12. is now in Remote Mode e Returns the target piezo control percentage which is the front panel percentage at the moment the query occurred Example Front panel shows the following e Remote LED off Local Mode e Piezo voltage reads 30 1 e Diode temperature set point reads 20 0 e Diode current reads 59 2 VOLT 66 8 gt OK Front panel now shows e Remote LED on Remote Mode e Piezo Voltage reads 64 9 e Diode temp set point reads 19 7 e Diode current reads 58 1 SENSE VOLT PIEZO gt 64 9 returns the actual percentage VOLT gt 66 8 returns the target percentage 69 Command Control Form Query Form Description Explanation Control Form 70 SOURce WAVElength lt wsp gt lt value MIN MAX gt NM ANG SOURce WAVELength lt wsp gt lt MIN MAX gt Write and read the laser wavelength Drives the laser to the wavelength specified by lt valuelMINIMAX gt MIN lt VALUE lt MAX MIN and MAX are laser dependent and may be queried see below NM indicates nanometers this is the default ANG indicates angstroms MIN Drives the laser to its minimum wavelength MAX Drives the laser to its maximum wavelength This command does not affect and is not affected by whether the laser is in Local Mode or Remote Mode If the value is out of range e Returns Out of Range e No other action is taken If the value is within the allowed range Returns OK e Wavele
13. plus or minus one millikelvin while the temperature of the surrounding environment stays constant to within plus or minus 10 millikelvin The PZT driver board supplies 0 120 Volts to a piezoelectric fine tuning element in the laser head Using the PZT system you can tune the wavelength smoothly with sub angstrom precision The Motion Control board controls the DC motor which provides coarse wavelength control and scanning A dedicated motion control microprocessor provides a PID feedback control loop for smooth tuning The microprocessor board tells all the other boards what to do runs the digital displays and provides RS 232 and GPIB interfacing capability It also communicates with the circuit board in the laser head to determine what kind of laser head it is and to upload the wavelength calibration table Finally the mother board is the interface backplane for all the boards and also drives the LED displays 15 16 Laser Head The laser head embodies a simple ultra stable design which is shown in Figure 4 Everything is mounted on a solid metal base and enclosed in a sealed package You will never need to open the laser head to operate the laser Unauthorized opening of the laser head will void the warranty Figure 4 Laser head mechanical schematic Current Modulation Input Frequency Modulation Input Wavelength Sensor Output GPIB RS 232 Detector Input Controller Wavelength Inpu
14. the Local button while holding up the Scan Speed paddle switch on the front panel Then turn the Wavelength Adjust knob until the number of the address you want appears in the Wavelength readout dis play You can set the address to be anything from 1 to 31 In the examples that follow we assume that you set your laser controller to address 1 Note also that the GPIB address is stored in the laser head so that if you switch heads you should check to make sure that the second head has the same GPIB address as the first RS 232 RS 232 is a widely available standard for communication via your computer s serial port You might want to use RS 232 to save yourself the expense of a GPIB card The one drawback is you can only talk to one instrument at a time with serial communication You will need to know how to send commands to your computer s RS 232 serial port Any terminal emulation program will let you send and receive via the serial port interactively RS 232 ports can be configured for operation in DTE or DCE mode The laser controller is configured as a DCE port This means that the laser controller receives data on pin 2 and transmits data on pin 3 For RS 232 operations use to get attention before starting command entry and end the command with a carriage return lt cr gt is the attention signal which must be the first character of each command string RS 232 commands are identical to the GPIB commands RS 232 command
15. to sub angstrom wavelength tuning precision The laser cavity is carefully adjusted to give the best tuning performance when tuned from short to long wavelength Therefore when scanning the laser it is best to approach the desired wavelength from the short wave length side Asmall fraction of the output beam is directed to a power monitor The reading from this monitor is displayed on the front panel of the controller On the outside of the head enclosure you will see an SMA connector for high speed current modulation The use of this feature is described in Back Panel and Laser Head Connections in the next section Finally there is a monitor strip that indicates how many hours the diode laser has operated A new laser head will typically show 100 150 hours due to factory burn in The red monitor bar will slowly make its way across the window It will reach the other side when 5000 hours have elapsed 17 18 V Manual Operation Using the Front Panel Before you start make sure the controller is plugged into a wall socket and your laser head cable is plugged into the back of the controller The cable has two plugs on it which fit into the sockets marked Cable A and Cable B Make sure you have the head pointed in a safe direction and check that the AC power input connector is configured for the correct line voltage See Section VIII page 85 for information about the AC power input to the controller Turn on th
16. volts and presents a 51 kQ load Constant Power Mode Constant power mode provides you an easy way to remove the power modulation due to the etalon effect caused by the residual facet reflectivity when scanning wavelength or to stabilize the laser power over long term power drift A digital feedback loop is implemented in the firmware to keep the laser power at a set level by controlling the laser current Because of the relatively long loop time 20ms we recommend using this mode with a scan speed under 1 of the maximum speed To enable constant power Mode hold the speed adjust paddle switch up and the scan temperature paddle switch down simultaneously While both switches are held the wavelength display shows the laser power setting and you can use the wavelength adjust knob to adjust it A letter C appears in the power display indicating the controller is in constant power mode If the controller cannot reach the power setting within the mulitmode current the letter C will flash to indicate the feedback loop is open The laser current stays at the multimode current Please note that the controller is set to remote mode when constant power mode is enabled This means that in constant power mode you can only adjust the diode temperature and the PZT voltage through either GPIB or RS 232 commands There are two ways to disable constant power mode You can hold the speed adjust paddle switch down and the scan temperature paddle switc
17. wavelength and the following commands are executed WAVE SLEW RET MAX gt OK The wavelength now changes at its maximum rate WAVE SLEW RET 0 05 gt OK The wavelength now changes slowly at 0 05nm s WAVE SLEW RET gt 0 05 75 Command Control Form Query Form Description Explanation Control Form Explanation Query Form Example 76 SOURce WAVElength STARt lt wsp gt lt value gt NM ANG SOURce WAVElength STARt Read and write the start wavelength for a scan This command does not affect and is not affected by whether the laser is in Local Mode or Remote Mode MIN lt VALUE lt MAX e MIN and MAX are laser dependent See SOURCE WAVELENGTH MINIMAX for details e The default units are nanometers NM ANG indicates angstroms If the value is out of range e Returns Out of Range e No other action is taken If the value is in range Returns OK e Start wavelength is given a new value Returns the current start wavelength The start wavelength is 772 nm and the following commands are executed WAVE START 770 gt OK The start wavelength is now 770 nm WAVE START gt 770 0 Command Control Form Query Form Description Explanation Control Form Explanation Query Form Example SOURce WAVElength STOP lt wsp gt lt value gt NM ANG SOURce WAVElength STOP Writes and reads the stop wavelength set point for wavel
18. 10 nm 30 mW 6312 710 800 nm 70 mW 6316 760 905 nm 100 mW 6320 830 1180 nm 150 mW 6324 1100 1480 30 mW 6326 1350 1650 50 mW 6327 1350 1710 50 mW 6328 1380 1710 50 mW 6330 1550 1900 25 mW Table of maximum possible wavelengths and powers accessible inside the laser head listed by model number The actual wavelength range and output power of your laser will be a frac tion of the values shown bere Note The user will NEVER need to open the laser head Contact New Focus if for some reason you want to open the laser head Unauthorized opening of the laser head will void the warranty and may result in misalignment of the laser cavity and or irreparable damage to the internal components Diode laser power at the wavelengths shown in the table above could be accessible in the interior of the laser head The laser radiation emitted from this unit may be harmful Avoid direct expo sure to the beam Avoid looking at the laser beam directly The safety labels shown on the following page are attached to this product Safety Warning Labels Aperture Label ID Certification Label Il Quick Start This section outlines the quickest path to using your 6300 Series External Gavity Tunable Diode Laser For more detail on how to operate the laser please refer to Section VI for computer control Setting Up The laser is configured at the factory for the line voltage and frequency appropriate for your country If you are
19. 32 while Remote indicates that the laser is actually in remote mode with temperature piezo voltage and laser current set by the computer If you need to cancel remote mode at any time pushing the Local button will restore local mode with full front panel control Fine Frequency Control The Piezo Voltage knob and readout are used for fine sub angstrom wave length tuning A piezoelectric transducer PZT is used to make adjustments in the tuning mirror angle that are too small to be made the by DC motor The readout is in percent of the maximum PZT voltage from 0 to 100 The next block contains the wavelength scanning and temperature controls The Wavelength Adjust knob is a multipurpose control What it does depends on which other buttons you press at the same time Similarly the Wavelength readout can display operating wavelength scan speed scan start and stop wavelengths GPIB address and serial baud rate The Temperature Adjust knob and Temperature readout control the desired operating temperature and display the actual temperature respectively Wavelength Track Mode and Ready Mode Push the Track button above the Wavelength Adjust knob to make the button light up If the button is already lit you can try turning it off and on again When the Track light is on the laser is in the Track Mode and you can use the Wavelength Adjust knob to change the wavelength When the Track light is off the Wavelength Adjust knob is disa
20. 6316 The actual modulation you will observe depends on the wavelength range of the laser head see the Acceptance Test Data sheet of your particular laser and on the rate and amplitude of the incoming signal for example full 30 GHz modulation is only available at rates up to 200 Hz The Frequency Modulation input will accept signals from DC to 2 kHz 3 dB rolloff point Increasing voltage at the Frequency Modulation input corresponds to a decrease in piezo voltage and an increase in laser frequency or a decrease in laser wavelength This decrease in piezo voltage is effectively subtracted from the front panel knob setting and the front panel display reads the result Current Modulation Input The Current Modulation input allows you to modulate the diode current as fast as I MHz This input accepts 10 to 10 volts into a DC coupled 5 kQ resistive load and provides 0 2 mA V modulation This input is summed with the front panel setting Note that the front panel current read out does not reflect the modulation input High Speed Current Modulation For high speed current modulation up to 100 MHz an SMA jack on the laser head is provided This input is AC coupled the low and high frequency roll off points are 50 kHz and about 100 MHz depending on the model The current modulation provided by this input is approximately 20 mA V To pre vent damage to the diode laser the voltage swing on this input must be less than 1 Vp p 25 26
21. 960 995 nm Minimum Power 3mW 2mW 5mW 5mW 5 5 mW ypical Max Power 5mW 6 mW 9 mW 7mW 8mW Fine Freq Tuning Range 70 GHz 0 09 nm 70 GHz 0 10 nm 75 GHz 0 15 nm 60 GHz 0 14 nm 50 GHz 0 16 nm Current Modulation BW 100 MHz 100 MHz 100 MHz 100 MHz 100 MHz R 6 nm sec 6 nm sec 8 nm sec 8 nm sec 12 nm sec Linewidth 50ms lt 300 KHz lt 300 KHz lt 300 KHz lt 300 KHz lt 300 KHz Linewidth 5 s lt 5 MHz lt 5 MHz lt 5 MHz lt 5 MHz lt 5 MHz Er ak 2kHz 2kHz 2kHz 2 kHz 2kHz Coarse Tuning Resolution 0 02 nm 0 02 nm 0 02 nm 0 02 nm 0 02 nm Tuning Repeatability 0 1 nm 0 1 nm 0 1 nm 0 1 nm 0 1 nm Operating temperatures 15 35 C 15 35 C 15 35 C 15 35 C 15 35 C odel 6324 6326 6 6328 6330 uning Range 1280 1330 nm 1470 1545 nm 1415 1480 nm 1520 1570 nm 1550 1630 nm Minimum Power 4mW 4mW 3mW 20mW 4mW ypical Max Power 7mW 10 mw 8mW 24 mW 8 mW Fine Freq Tuning Range 50 GHz 0 29 nm 50 GHz 0 23 nm 30 GHz 0 21 nm 30 GHz 0 24 nm 30 GHz 0 25 nm Current Modulation BW 100 MHz 100 MHz 100 MHz 100 MHz 100 MHz ee 15 nm sec 20 nm sec 20 nm sec 20 nm sec 20 nm sec Linewidth 50ms lt 300 KHz lt 300 KHz lt 300 KHz lt 300 KHz lt 300 KHz Linewidth 5 s lt 5 MHz lt 5 MHz lt 5 MHz lt 5 MHz lt 5 MHz RE OE 2kHz 2kHz 2 kHz 2 kHz 2kHz Coarse Tuning Resolution 0 02 nm 0 02 nm 0 02 nm 0 02 nm 0 02 nm Tuning Repeatability 0 1 nm 0 1 nm 0 1 nm 0 1 nm 0 1 nm Operating temperatures 15 35 C 15 35 C 15 35 C 15 35 C 15 35 C
22. Example OUTPut TRACk lt wsp gt OFF Exit Track Mode to Ready Mode Takes the laser out of Track Mode and places it in Ready Mode the quiet mode of operation Note that issuing a SOURce WAVElength command for instance will place the laser in Track Mode and OUTPut TRACk OFF is the way to get back into Ready Mode Works the same whether the laser is under front panel or DAC control The laser wavelength has been set to 672 4 nm with a WAVE 672 4 nm com mand The Track LED is on OUTP TRACK OFF gt OK The laser is still at 672 4 nm but is now in Ready Mode the quiet mode of operation The Track LED is off 49 Command Description Explanation Query Example 50 OUTPut STATe lt wsp gt lt OFF ON 0 1 gt OUTPut STATe Turns the laser on and off reads whether the laser is on or off ON or 1 Ifthe laser is ON no action is taken If the laser is OFF the Laser Power switch LED blinks and after 5 seconds the laser is turned on and the LED is steadily on OFF or O If the laser is ON or in the intermediate blinking state the laser turns off and the Laser Power switch LED turns off If the laser is already OFF no action is taken Returns 1 if the laser is on and 0 if it is off Works the same whether the laser is in Remote Mode or Local Mode The front panel Laser Power switch is off OUTP gt 0 OUTP ON gt OK The Laser Power switch LED begins to blink OU
23. Model 6300 User s Manual The Velocity Tunable Diode Laser U S Patent 5 319 668 European Patent EP 0 693 231 B1 CAUTION Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure A NEW FOCUS 5215 Hellyer Ave e San Jose CA 95138 1001 e USA phone 408 284 6808 e fax 408 284 4827 e mail contact techsupport com e www newfocus com Contents I User Safety 3 II Quick Start 5 III Introduction 8 IV Description of Laser System 14 V Manual Operation 18 VI Computer Control 29 VII Command Summary 37 VIII Error Codes 84 IX AC Operating Voltages 85 X Service and Support 87 XI Specifications 88 Products described in this document are covered by U S Patent 5 319 668 and European Patent EP 0 693 231 B1 Information in this document is subject to change without notice Copyright 2002 New Focus Inc All rights reserved The a NEW FOCUS and v logos NEW FOCUS Inc and Velocity are registered trademarks of New Focus Inc Document Number 628019 Rev H User Safety Laser Safety Warnings Your safe and effective use of this product is of utmost importance to us at New Focus Please read the Laser Safety Warnings before attempting to operate the laser Note To completely shut off power to the unit unplug it from the wall Model Wavelength Range Max Power 6304 600 645 nm 15 mW 6308 630 7
24. SOURce POWer LEVel lt wsp gt lt value gt Set the constant power level lt value mw gt VOLTage LEVel PIEZo CURRent LEVel DIODe Read diode current set point lt value mA gt SOURce TEMPerature LEVel DIODe Read diode temperature set point lt value deg C gt Read piezoelectric voltage set point lt value 0 100 Flength lt wsp gt lt MIN MAX gt Read output wavelength set point lt value nm gt Elength SLEWrate FORWard Read forward slew rate set point lt value nm s gt SOURce WAVELength SLEWrate RETurn Read return slew rate set point lt value nm s gt SOURce WAVElength SLEWrate MAX Read Maximum slewrate LT SOURce WAVElength STARt Read scan start wavelength set point lt value nm gt SOURce WAVElength STOP Read scan stop wavelength set point lt value nm gt System Commands SYSTem WINPut lt wsp gt lt 0 1 OFF ON gt Enables and disables Wavelength Input Mode EXT gt SYSTem MCONtrol lt wsp gt lt INT Mode control Remote Mode INT DACs with Encoder Track Scan disabled Local Mode EXT Front panel pots with all enabled 43 Standard Commands Command Description Explanation Example 44 IDN Identification Query Returns the following string identifying the instrument NEW FOCUS Model H Head Revision
25. TP gt 0 After 5 seconds have elapsed the LED is solidly on and the laser is on OUTP gt 1 Sense Commands Command Description Explanation Example SENSe CURRent LEVel DIODe Read diode laser current level Returns the magnitude of the laser diode current in milliamps The value returned is the same as that seen on the front panel of the controller at the time the command is executed Front panel current reading is 59 2 SENSE CURR DIOD gt 59 2 51 Command Description Explanation Example 52 SENSe POWer LEVel FRONt Read front facet power Returns the laser output power in milliwatts The value returned is exactly the same as that seen on the front panel of the controller at the time the command is executed Note that there are two power readings One is derived from a front facet measurement and the other from a back facet measurement The front facet power measurement looks at light sampled with a beam splitter at the laser head output The back facet measurement is merely proportional to the opti cal power emitted from the rear facet of the laser gain element It is provided for backward compatibility with 6100 series laser heads or as a diagnostic The output power indicator on the front panel reads 2 6 mW SENS POW FRON gt 2 6 Command Description Explanation Example SENSe POWer LEVel REAR Read rear f
26. The laser controller will indicate an error code in the wavelength display if it is unable to perform its proper functions Some of these errors can be cleared by switching the laser controller off and then back on If the errors persist please contact New Focus for assistance Errors 1 through 7 could indicate a broken or stuck switch Check the appropriate switch and if the errors persist contact New Focus Error 1 Error 2 Error 3 Error 4 Error 5 Error 6 Error 7 Error 8 Error 9 Error 10 Laser Power switch depressed during power up Scan button depressed during power up Track button depressed during power up Multifunction switch depressed during power up Scan Speed switch depressed during power up Local button depressed during power up Multiple stuck buttons during power up No head is connected to the controller or compatibility plug for 6100 series lasers is missing Cavity temperature at or greater than 35 C Make sure the laser head is heat sinked Diode temperature at or greater than 35 C Make sure the laser head is heat sinked These errors indicate a hardware problem If these errors persist contact New Focus Error 11 Error 12 Error 13 Error 14 Error 15 Error 16 Error 17 Microprocessor RAM error EPROM revision number mismatch Current board revision wrong Temperature board revision wrong Laser head EEPROM stability error or there is a compatibility conflict between control
27. acet power Returns a number proportional to the power emitted from the rear facet of the laser gain element This is provided for backward compatibility with 6100 series laser heads or as a diagnostic Use the front facet power measurement to determine the actual optical power output from the laser SENS POW REAR gt 2 9 53 Command Description Explanation Example 54 SENSe TEMPerature LEVel DIODe Read diode temperature Returns the temperature of the laser diode gain element in degrees Celsius This value is exactly the same as that which appears on the front panel of the laser The diode temperature is actively controlled If the room temperature ever gets so high that the coolers have insufficient capacity to reduce the temperature the control loop can become unstable and destroy the cooling elements To prevent damage the laser is automatically shut down if the laser diode gain element temperature exceeds 35 C The diode temperature is 21 1 C SENS TEMP LEV O DE gt 21 1 Command Description Explanation Example SENSe TEMPerature LEVel CAVity Read cavity temperature Returns the temperature of the laser cavity in degrees Celsius This value is exactly the same as that which appears on the front panel of the laser The laser cavity temperature is actively maintained by the controller at about 30 C There is no command to set or chang
28. and the controller will be placed into Remote Mode Each Source Control has an accompanying Query Source Queries return the value of various controller set points A Source Query of diode current diode temperature or piezo voltage is only meaningful if the laser is in Remote Mode For this reason if any of those three parameters are queried while the laser is in Local Mode the laser is placed in Remote Mode with the current temperature set point and piezo voltage taking on the last front panel values they had before the switch Sense Commands return the value of various laser operating conditions Note that there is an important difference between Sense Commands and Source Queries Sense Commands read actual voltages currents temperatures or wavelengths the value returned is the same as that seen on the controller front panel at the time the command is executed In contrast Source Queries read set points values that are theoretical goals These values may not necessarily match the values displayed on the front panel For instance the queried diode current and piezo voltages can differ from the actual values by up to 3 due to component variations such as resistor tolerances However in the case of wavelength queries the wavelength set by a Source Command will match the wavelength returned by a Source Query because each laser is factory calibrated and has its specific wavelength calibration table loaded into the laser head The n
29. avelength is 672 nm the Start Wavelength is 670 0 nm and the Return Slew Rate is 6 nm s OUTP SCAN RESET gt OK The laser wavelength is changed at 6 nm s until the wavelength 670 0 nm is reached Command Description Explanation Example OUTPut SCAN STARt Start restart a wavelength scan Begins scanning from the current wavelength to the stop wavelength at the forward slew rate Works the same whether the laser is in Remote Mode or Local Mode The OPC command will return 0 until the stop wavelength is reached The current wavelength is 670 0 nm the Stop Wavelength is 672 0 nm and the Forward Slew Rate is set at 5 nm s OUTP SCAN START gt OK The laser wavelength is tuned at 5 nm s until 672 0 nm is reached 47 48 Command Description Explanation Example OUTPut SCAN STOP Stop pause a wavelength scan Stops a scan or reset if one is going on leaving the laser in Ready Mode Works the same whether the laser is in Remote Mode or Local Mode The OPC flag is set to 1 showing that the SCAN or RESET is stopped The laser is performing a scan and then the following commands are issued to the controller OPC gt 0 OUTP SCAN STOP gt OK XOBCZ gt l The scan is stopped the controller is in Ready Mode and the wavelength is set to the wavelength the laser was at when the OUTP SCAN STOP command was issued Command Description Explanation
30. bled and the laser is in the Ready Mode In Track Mode the laser wavelength is actively controlled using the tuning 19 20 motor that turns the end mirror In Ready Mode the laser runs open loop without active wavelength control Because of the nature of the wavelength control scheme however note that the wavelength of the laser is significantly more stable in Ready Mode than it is in Track Mode With the laser in Track Mode you can set the wavelength by rotating the Wavelength Adjust knob Note that the display shows 0 01 nanometer resolution for setting wavelength but only 0 1 nanometer resolution once the operating wavelength is established This is because the precision with which you can set the wavelength is greater than the accuracy of the wavelength readout When you select a new wavelength with the Wavelength Adjust knob the DC motor moves the end mirror until the new wavelength is reached Setting the Temperature To set the temperature hold the multipurpose scanning and temperature paddle switch to the left of the Temperature Adjust knob up toward Diode While you are still holding the switch up turn the Temperature Adjust knob until the Temperature readout shows the number you want to set After you let the multipurpose switch go the Temperature readout will show the actual temperature which should soon become the set temperature Normally you will leave the temperature at the recommended operating tem pera
31. by turning the wavelength adjust knob For other functions such as setting scanning speed and wavelength range enable wavelength input mode and enable constant power mode please refer to the switch knob combinations listed on the next page Switch key Combinations Set diode temperature Set scanning start wavelength Set scanning stop wavelength Set start to stop scan speed Set stop to start scan speed Enable External Wavelength Input Mode Track LED ON Disable External Wavelength Input Mode Enable Constant Power Mode and Set Power Level Disable constant Power Mode Set GPIB adress Set RS 232 baud rate Loca Loca Y Scan Speed A Scan Speed y Scan Speed a Scan Speed Y Scan Speed D Scan Speed Y Scan Speed R Scan Speed Wavelength Adjust Wavelength Adjust Q Wavelength Adjust Q Wavelength Adjust Wavelength Adjust Wavelength Adjust Q Wavelength Adjust Start Vv Diode A Stop Cavity Start w Diode A c Stop Cavity Start Diode A c Stop D Cavity Start Y Diode A c Stop Cavity Start Diode A H Stop D Cavity Start Diode A c Stop D Cavity Start w Diode A c Stop Cavity Q Temperature Adjust II Introduction The New Focus 6300 series External Cavity Tunable Diode Laser is a stable narrow linewidth source of tunable light The 6300 series laser can be operated manually from the
32. ctly Each laser is calibrated with respect to wavelength at the factory and a calibration table loaded into the laser head The wavelength set by a Source Command will match the sensed wavelength to the number of decimal points returned The number of decimal points depends on the accuracy of the angle measurement technique used to measure the wavelength Currently the accuracy of this measurement is 0 1 nm The wavelength is 679 6 nm SENS WAVE gt 679 6 Source Controls and Source Queries Command Control Form Query Form Description Explanation Control Form SOURce CURRent LEVel1 DIODe lt wsp gt lt value gt SOURce CURRent LEVel DIODe Write and read the diode current set point 0 lt value lt 152 The units are milliamps This command sets the diode laser current This command always forces the laser into Remote Mode If the value sent is out of range e The controller returns Out of Range e No other action is taken If the value is in range e The diode laser current is changed e The controller returns OK If the laser was in Remote Mode e The Remote LED is already on e The current control DAC is driven to a voltage which produces the requested current within 3 e The piezo voltage and the diode temperature are unaffected 61 62 Explanation Query Form If the laser was in Local Mode e The Remote LED comes on e Control of the laser is given t
33. e AC power with the key switch on the left hand side of the front panel The calibration table GPIB address baud rate laser head serial num bers and other pertinent operating information stored in an EEPROM in the laser head are uploaded to the controller The LED displays will show the model number head software revision number and controller software revi sion number After a few seconds the display will become active About 30 minutes of warm up time is recommended for best performance Figure 5 Front panel of laser controller A NEW FOCUS VELOCITY _ Tuneble Diode Laser Controller The top half of the front panel contains all the indicators and readouts while the lower half has control knobs to adjust operating parameters The AC Power key switch is the on off for the whole unit including the laser head When you push the Laser Power button it flashes for a 5 second safety delay and then lights up to indicate that current is flowing through the diode laser in the laser head unit This allows you to set up all the desired operating parameters with the AC Power on but before you actually start generating light The Addressed and Remote indicator LEDs are status indicators Addressed lets you know if your computer has established contact with the laser via GPIB or RS 2
34. e sure to look at Example 1 before reading this program INCLUDE gbdecl bas CLS buff SPACES 32 Find the laser controller at GPIB address tl CALL IBDEV 0 1 0 T10s 1 0 nflaser CALL IBCLR nflaser Set up start and stop wavelengths INPUT What start wavelength would you like nm wavelength CALL IBWRT nflaser WAVELENGTH START STRS wavelength INPUT What stop wavelength would you like nm wavelength CALL IBWRT nflaser WAVELENGTH STOP STRS wavelength Set forward and return scan speeds to their max imum values CALL IBWRT nflaser WAVELENGTH SLEW FORW 100 CALL IBWRT nflaser WAVELENGTH SLEW RET 100 Get ready to scan CALL IBWRT nflaser OUTPUT SCAN RESET PRINT Tuning to the start wavelength Wait until laser gets to start wavelength Here we use the OPC operation complete query to see if OUTPUT SCAN RESET has finished ready 0 35 36 WHILE ready 0 CALL IBWRT nflaser OPC CALL IBRD nflaser buf ready VAL buff WEND Start scan PRINT Push any key to begin the scan WHILE INKEYS WEND CALL IBWRT nflaser OUTPUT SCAN START END Vil Command Summary With GPIB and for programs written in QuickBAsic all commands are issued using the IBWRT function call in
35. e the cavity temperature set point If the room temperature gets high enough that the coolers have insufficient capacity to reduce the temperature the control loop can become unstable and destroy the cooling elements To prevent damage the laser is automatically shut down if the laser diode gain element temperature exceeds 35 C The laser cavity temperature is 30 0 C SENS TEMP LEV CAV gt 30 0 55 56 Command Description Explanation Example SENSe TEMPerature SETPoint DIODe Read diode temperature set point Returns the current temperature set point for the laser diode gain element in degrees Celsius This value is exactly the same as the set point that appears on the front panel of the laser The diode temperature is maintained by an analog control loop This command returns the measured set point of the control loop This value can be changed either from the front panel or with the SOURce TEMPerature command The diode temperature set point is 21 1 C SENS TEMP SETP DIODE gt 21 1 Command Description Explanation Example SENSe TEMPerature SETPoint CAVity Read cavity temperature set point Returns the current temperature set point for the laser cavity in degrees Celsius This is the only way this value can be read it cannot be read from the front panel The cavity temperature is actively controlled at about 30 C This command returns t
36. ect Otherwise it disables Constant Power Mode and leaves the laser in Remote Mode The constant power level is saved in memory Example The laser has not been in Constant Power Mode after power up and it is in Local Mode with the front panel power display reading 1 5 mw SOUR CPOW ON gt OK e The remote LED is ON e C appears in power display e The Constant Power Mode is enabled and the laser power is kept at 1 5m SOUR CPOW OFF gt OK e The Constant Power Mode is disabled and C disappears in power display e The remote LED remains On e The laser is in remote mode 79 Command Description Explanation Example 80 SOURce POWer LEVel1 lt wsp gt lt value gt Set the constant power level This command does not affect if the laser power is not ON This command will force the laser to Remote Mode if it is previously in Local Mode 0 lt Value lt 99 9 the units are in mw Although this command accepts a value up to 99 9 mw the actual set level is limited by an internal maximum value which is system dependent See the Acceptance Test Data sheet for this number If the laser is already in Constant Power Mode this command sets the constant power level to the specified value or internal maximum limit whichever is lower If the laser is not in Constant Power Mode this command enables Constant Power Mode and sets power level to the specified value or the internal maximum
37. ength scans This command does not affect and is not affected by whether the laser is in Local Mode or Remote Mode MIN lt VALUE lt MAX e The MIN and MAX values are laser dependent See SOURCE WAVELENGTH MINIMAX for details e The default units are nanometers NM ANG indicates angstroms If the value is out of range e Returns Out of Range e No other action is taken If the value is in range e Returns OK e Stop wavelength is set to the new value Returns the current stop wavelength The stop wavelength is 776 nm and the following commands are executed WAVE STOP 774 gt OK The stop wavelength is now 774 nm WAVE STOP gt 774 0 77 78 Command Description Explanation SOURce CPOWer lt wsp gt lt 0 1 OFF ON gt Enables and disables Constant Power Mode This command does not affect if the laser power is not ON This command will force the laser to Remote Mode if it is previously in Local Mode ON or 1 If the laser is already in Constant Power Mode this command has no effect Otherwise it turns on Constant Power Mode If the laser is in Local mode it is forced to Remote mode If the laser has not been in his mode after power up the constant power level is set to the power reading at the time this command is executed If it has the constant power level is set to the previous setting OFF or 0 If the laser is not in Constant Power Mode this command has no eff
38. ew rate only affects scanning The slew rate during a wavelength change from a SOURCE WAVELENGTH command is always at the maximum value Returns the current forward slew rate Example The laser is scanning from the start to the stop wavelength and the following commands are executed WAVE SLEW FORW MAX gt OK The wavelength now changes at its maximum rate WAVE SLEW FORW 0 05 gt OK The wavelength now changes slowly at 0 05nm s WAVE SLEW FORW gt 0 05 73 Command Control Form Query Form Description Explanation Control Form Explanation Query Form 14 SOURce WAVELength SLEWrate RETurn lt wsp gt lt value MAX gt SOURce WAVELength SLEWrate RETurn Write and read the forward slew rate for a wavelength scan 0 01 lt VALUE lt MAX The units are in nm s MAX is the maximum scan rate which is laser dependent This command does not affect and is not affected by whether the laser is in local mode or remote mode If the value is out of range e Returns Out of Range e No other action is taken If the value is in range e Returns OK e The return wavelength slew rate is set to the desired value e The slew rate only affects scanning The slew rate during a wavelength change from a SOURCE WAVELENGTH command is always at the maximum value Returns the current return slew rate Example The laser is resetting from the stop to the start
39. ext four pages have an index of all the possible computer control commands Then the pages that follow the index of commands give detailed information about each command including a description of the command and examples of how the command is used 39 40 Index of Commands Standard Commands Output Commands Sense Commands IDN OPC OUTPut OUTPut SOUTP Ut OUTPut Identification Query Operation Complete Query lt Value 0 Not complete 1 Complete SCAN RESEt Stop and return to start wavelength SCAN STARt Start Restart scan SCAN STOP Stop Pause scan TRACk lt wsp gt OFF Exit Track Mode to Ready Mode OUTPut STATe lt wsp gt lt OFF ON 0 1 gt Turn laser on off OUTPut STATe u u Read laser on off lt value 0 Off 1 On gt ENSe CURRent LEVel DIODe Read current level lt value mA gt ENSe POWer LEVel FRONt Read front facet power lt value mW gt SENSe POWer LEVel REAR Read rear facet power lt value proportional to power arbitrary units gt SENSe TEMPerature LEVel DIODe Read diode temperature lt value deg C gt SENSe HW EMPerature LEVel CAVity Read cavity temperature lt value deg C gt EMPerature SETPoint DIODe Read diode temperature set point lt value deg C gt SENSe H SENSe TEMPerature SETP
40. front panel of the controller or remotely using computer control In this section we ll describe the key features of the 6300 series and explain some of the theory of external cavity diode laser design In Section IV page 14 we give a thorough description of each part of the 6300 system Section V page 18 covers manual operation while Section VI page29 goes over computer control of the laser system Section VII page 37 has detailed explanations of the computer control commands and error codes are listed in Section VIII page 84 The AC operating voltages of the controller are detailed in Section IX page 85 and service and support are described in Section X page 87 Finally a table of specifications is given in Section XI page 88 Features The 6300 series laser incorporates a simple stable mechanical design with a minimum number of optical components Low noise analog circuits precisely set critical operating parameters such as diode temperature and current Digital control facilitates remote operation and computer interfacing The 6300 is a modular system The same control unit will work with any 6300 series laser head The control units work in both manual and remotely programmed modes and are compatible with RS 232 and GPIB addressing Microprocessor based motion control allows you hands free wavelength scan ning capability You set the start and stop wavelengths and scanning speed and the microprocessor controlled DC m
41. h up simultaneously This will leave the controller in remote mode Or you can press the Local button This disables both remote and constant power modes After leaving this mode the power setting is stored in memory VI Computer Control Anything you can adjust from the front panel you can also adjust by computer control The 6300 Series system gives you an unprecedented level of external control over a high precision narrow bandwidth laser source Computer control lets your computer talk to the laser controller to set or read a variety of operating parameters and perform certain actions The controller is always in one of two operating modes Local Mode or Remote Mode The operating mode is independent of computer control or manual control The wavelength can be set or scanned in either operating mode The piezo voltage the diode temperature and the diode current on the other hand are all either controlled by the external knobs Local Mode as discussed earlier or all by internal digital to analog converters DACs that can be set remotely Remote Mode When any one of these DACs is set by computer control all three are then forced into Remote Mode The unspecified values default to their front panel settings at the moment of the switch to Remote Mode Pressing the Local button returns the laser to Local Mode Back panel inputs and outputs are still active in Remote Mode Note that the controller will allow communication over the co
42. having the tuning mirror rotate around a specific pivot point The pivot point creates a relationship between the cavity length and the laser wavelength The laser wavelength is set by the standard law for diffraction of light off a grating A A sin8 sin A is the spacing between grooves in the grating while and O refer to the incident and diffracted angles of the laser beam measured from a line normal to the surface of the grating The length D of the cavity can be broken into two parts and See Figure 2 Ifwe call the distance from the pivot point to the place where the beam strikes the grating L we can see from the figure that l Lsin and Lsin Therefore the total cavity length is Z sin6 sin Dividing the total cavity length by the wavelength shows that the total number of waves in the cavity is L A which is a constant This translates into a tunable laser with no mode hops That s the basic theory behind the cavity design In the next section we ll show how the theory is put into practice by describing the components of the New Focus 6300 series laser system 11 12 Figure 2 Drawing of important angles and optimum pivot point Grating Pivot Point References 1 2 M G Littman and H J Metcalf Spectrally narrow pulsed dye laser with out beam expander Applied Optics vol 17 p 2224 2227 1978 M G Littman Single mode operation of grazi
43. he set point for the control loop inside the controller This quantity is set at the factory and cannot be changed by the user SENS TEMP SETP CAV F 30 20 57 58 Command Description Explanation Example SENSe VOLTage LEVel PIEZo Read piezoelectric voltage setting lt value 0 100 gt Returns the current piezo voltage setting in percentage of maximum from 0 to 100 percent This value is the same as that which appears on the front panel of the laser Note that as the piezo voltage increases the lasing frequen cy increases and the wavelength decreases The piezo voltage is 66 8 percent of its maximum SENS VOLT PI gt 66 8 EZ Command Description Explanation Example SENSe VOLTage LEVel AUXiliary Read voltage at the Detector Input lt value 0 5 volts gt Returns the voltage at the auxiliary Detector Input on back panel of the laser controller The value returned is from 0 to 5 volts This is the only way this value can be read it cannot be read on the front panel The resolution for the Detector Input is 10 bits so the step size is 5 1024 4 88 mV The Detector Input sees 1 22 volts SENS VOLT AUX gt 1 22 59 Command Description Explanation Example 60 SENSe WAVElength Read laser wavelength lt value nm gt Returns the current wavelength of the laser in nanometers This value will match the front panel display exa
44. lding down the Scan Speed switch Performing a Scan To start a scan push the Scan button If the laser was in Track Mode the Track LED will turn off If the laser is at the start wavelength it will begin scanning at the scan speed Otherwise it will go to the Start wavelength at the return scan speed and wait Push the Scan button again and the laser will begin scanning When the laser arrives at the Stop wavelength it will stop and wait there If you push the Scan button again the laser will reset to the Start wavelength There is a Trigger jack SMB type connector an SMB to BNC adapter is supplied with the laser under the Scan Start Stop switch Voltage at this input rising above 3 V and returning to zero is equivalent to depressing and releasing the Scan button Stopping a Scan If you push the Scan button in the middle of a scan or a reset it will stop leaving you in Ready Mode If you push the Track button in the middle of a scan or a reset the scan will be halted and the controller will be in Track Mode 23 24 Back Panel and Laser Head Connections There are several input and output connectors on the back panel of the controller They allow you to operate your laser with external analog or digital signals There are two connectors labeled Cable A and Cable B for the cables that go from the controller to the laser head Cable A carries analog signals to the laser head while Cable B carries digital information You ca
45. ler and laser head Motion control board is not functioning Communication error between microprocessor board and motion control board IX AC Operating Voltages The Model 6300 Controller can operate at 100 120 220 or 240 V AC at AC frequencies of 47 63 Hz The unit is configured at the factory for the standard AC voltage in the owner s country To select a different operating voltage please refer to Fig 8 and follow the directions below 1 Disconnect the power cord 2 Open the cover of the power module on the rear panel using a small blade screwdriver or similar tool 3 Insert the tool into the voltage selection slot and remove the wheel from the unit Figure 8 AG line power connector AC voltage selection wheel IEEE 468 Frequency roulat ol Toser ove eng 4 Turn the wheel and re insert into the module so the desired voltage is shown Do not attempt to rotate the wheel while it is still in the power entry module the wheel must be removed turned and then re inserted 85 5 Close the cover 6 Verify the proper voltage is showing through the module s window The power entry module requires two 5x20 mm slow blow fuses such as Littelfuse s Slo Blo 239 series one for the hot line and the other for the neutral line Replacement fuses should be as follows AC Voltage Fuse Rating Littelfuse 100 V AC 2 0 A 239 002
46. limit whichever is lower If the laser is in Local Mode it is forced to Remote Mode The laser is in Local Mode with the front panel power display reading 1 5 mw SOUR CPOW 2 5 gt OK e The laser power is set to 2 5mW System Commands Command Description Explanation SYSTem WINPut lt wsp gt lt 0 1 OFF ON gt Enables and disables Wavelength Input Mode This command works the same in both Local Mode and Remote Mode ON or 1 e All wavelength activity such as scanning or even moving to a newly sourced wavelength is halted e The laser will immediately begin moving to the wavelength corresponding to the voltage at the Wavelength Input jack e The controller does not have to be in Track Mode to execute this command e Track Mode is entered on execution of this command e The Track LED will flash on and off signifying that the Wavelength Input is active e Encoder and Scan button are disabled OFF or 0 e The laser remains in Track Mode with the encoder active e Wavelength motion stops e Track LED stops flashing and remains on signifying Track Mode e Encoder can now change wavelength e The Scan button is active again 81 82 Example SYST WINP ON gt OK e Track LED flashes e Wavelength changes to voltage specified by value at Wavelength Input jack on back panel of controller e Encoder and Scan button are disabled SYST WINP OFF gt OK e Track LED stops flashing
47. lly fis between 0 2 and 1 V nm See the Acceptance Test Data Sheet for the particular Wavelength Output characteristics of your laser Wavelength Input The Wavelength Input jack is for coarse analog control of the wavelength For example you could use it for your own analog wavelength scanning control The DC signal that you apply to the Wavelength Input is digitized and sent to motion control board for wavelength control You can scan it as fast as the fastest internally controlled scan If your analog control signal is not too noisy you should be able to control the wavelength in angstrom or even sub angstrom increments Note however that if you don t have a quiet analog control voltage this mode of operation has less wavelength stability than Track Mode If you want to use the Wavelength Input jack you must first place the laser in Wavelength Input Mode To do this you must first be in the Track Mode Then push both front panel paddle switches scan speed and wavelength tempera ture multifunction up at the same time The wavelength display will show all ones 111111 and the Track light will begin to flash on and off The laser will immediately begin changing to the wavelength which corresponds to the voltage at the input The Wavelength Input jack accepts a DC signal between 0 and 10 V that corresponds linearly to the laser output wavelength in the same way as the Wavelength Output jack The input impedance of this control is 51 kQ
48. mputer interface without switching to Remote Mode Only commands which alter the state of the instrument will send it into Remote Mode For example you can use a computer to monitor all laser operating parameters while manually changing the wavelength from the front panel Please note that due to noises from DACs the intensity noise of the laser is higher in remote mode than in local mode for noise frequency up to a few kHz For noise sensitive applications we recommend you to operate in local mode for the best result GPIB GPIB stands for general purpose interface bus It is also known as the IEEE 488 standard GPIB is a standard protocol for personal computers to communicate with laboratory instruments and several manufacturers make 29 30 printed circuit board cards that plug into your computer and allow it to speak on the GPIB You will have to learn from the manufacturer of your GPIB card how to configure it to talk to an instrument at a given address and how to issue commands to it from your favorite programming language In QuickBasic for example commands are sent to the laser controller through the GPIB instructions IBWRT and IBRD IBWRT sends an instruction to the controller IBRD reads the controller s response to the command Examples of controller responses are OK or 772 3 Setting the GPIB Address To set the laser controller s GPIB address the controller must be in Ready Mode Press and hold
49. n use your 6300 series controller with older 6100 series laser heads by plugging the 6100 series head into the Cable A socket You must also plug a special 6100 ID plug into the Cable B socket to let the controller know that a 6100 series head is connected If you need to do this call us and we can provide you the necessary ID plug Figure 7 Back panel of laser controller An Interlock connector is provided for external safety systems The laser will not emit light unless the Interlock terminals are shorted Five BNC connectors are provided for external analog control GPIB IEEE 488 and RS 232 are available for computer interfacing For information on computer control please see the next section of the manual Computer Control Analog BNC inputs and outputs are available for Frequency Modulation Current Modulation Wavelength Input Wavelength Output and Detector Input In addition there is an SMA connector on the laser head for high speed current modulation In the rest of this section we Il describe how each of these inputs and outputs works in detail Frequency Modulation Input The Frequency Modulation input is for external analog control of the PZT voltage It is useful for making fine frequency adjustments and for FM spectroscopy Sweeping the Frequency Modulation input from 3 V to 3 V corresponds to changing the laser frequency by an amount on the order of 30 GHz to 30 GHz for the Model
50. ng incidence pulsed dye laser Optics Letters vol 3 p 138 140 1978 K Liu and M G Littman Novel geometry for single mode scanning of tunable lasers Optics Letters vol 6 p 117 118 1981 M G Littman Single mode pulsed tunable dye laser Applied Optics vol 23 p 4465 4468 1984 K C Harvey and C J Myatt External cavity diode laser using a grazing incidence diffraction grating Optics Letters vol 16 p 910 912 1991 T Day F Luecke and M Brownell Continuously tunable diode lasers Lasers and Optronics p 15 17 June 1993 13 IV Description 14 of Laser System Controller The controller s job is to provide a stable low noise power source for the diode laser set the temperature in the laser head command wavelength scanning and provide readouts of all relevant laser parameters Conceptually the circuitry inside the controller is built in two layers analog and digital The analog layer incorporates low noise design for temperature current and wavelength fine tuning The digital layer includes all the readouts and circuits to set operating points and scan parameters This layer acts as an interface between you or your computer and the analog layer Figure 3 Block diagram of circuit boards User Inputs Local Control Panel LEDs M Mother Board 1 RS 232 Microprocessor B d i T PZT Laser Head oar GPIB Board Board Board Motion Con
51. ngth is set to the desired wavelength Explanation Query Form Example Returns the latest target wavelength The actual wavelength may be changing due to a newly sourced value or because the laser is in the midst of a reset or a scan MIN Returns the Minimum wavelength achievable with the laser MAX Returns the Maximum wavelength achievable with the laser Front panel displays wavelength 679 6 nm WAVE 670 25 gt OK The following occurs e The wavelength starts changing at the maximum rate until it reaches 670 25 nm e The wavelength display will probably flicker between 670 2 and 670 3 as this is the round off point Note that you can use OPC to determine when the wavelength change has been completed 71 Command Control Form Query Form Description Explanation Control Form Explanation Query Form 72 SOURce WAVElength SLEWrate FORWard lt wsp gt lt value MAX gt SOURce WAVElength SLEWrate FORWard Write and read the forward slew rate for a wavelength scan This command does not affect and is not affected by whether the laser is in local mode or remote mode 00 1 lt VALUE lt MAX The units are in nm s MAX is the maximum scan rate which is laser dependent If the value is out of range e Returns Out of Range e No other action is taken If the value is in range e Returns OK e Forward wavelength slew rate is set to the desired value e The sl
52. o Remote Mode If the value is out of range e Returns Out of Range e No other action is taken If the value is in range Returns OK e Piezo voltage is changed If the laser is in Remote Mode e Remote LED is already on e The piezo control DAC is driven to a voltage which produces the requested percentage to within 3 e The diode current and the diode temperature are unaffected 67 68 Explanation Query Form If the laser is in Local Mode e Remote LED comes on e Control of the laser is given to the DACs e The piezo control DAC is driven to a voltage which produces the requested percentage to within 3 e The diode current control DAC and the temperature control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control e The laser is now in Remote Mode If the laser was in Remote Mode e Remote LED is already on e The current piezo voltage and diode temperature are unaffected e Returns the the last piezo control percentage sourced If the laser was in Local Mode e Remote LED comes on e Control of the laser is given to the DACs e The Piezo control DAC is driven to a voltage which produces a percentage within 3 of the value it had under front panel control e The current control DAC and the temperature control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control The laser
53. o the DACs e The current control DAC is driven to a voltage which produces the requested current within 3 e The temperature control DAC and the piezo control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control The laser is now in Remote Mode If the laser was in Remote Mode e Remote LED is already on e The current piezo voltage and diode temperature are unaffected e Returns the the last current sourced If the laser was in Local Mode e Remote LED comes on e Control of the laser is given to the DACs e The current control DAC is driven to a voltage which produces a current within 3 of the value it had under front panel control e The temperature control DAC and the piezo control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control The laser is now in Remote Mode e Returns the target current which is the front panel current at the moment the query occurred Example Front panel shows the following e Remote LED off Local Mode e Piezo Voltage reads 30 1 e Diode temp set point reads 20 0 e Diode current reads 59 2 CURR 26 4 gt OK Front panel now shows e Remote LED on Remote Mode e Piezo voltage reads 29 6 e Diode temperature set point reads 19 7 e Diode current reads 26 1 SENSE CURR DIODE gt 26 1 returns the actual current CURR g
54. oint CAVity Read cavity temperature set point lt value deg C gt SENSe VOLTage LEVel PIEZo Read piezoelectric voltage lt value 0 100 gt SENSe VOLTage LEVel AUXiliary Read user analog voltage lt value 0 5 V gt SENSe WAVElength Read output wavelength lt value nm gt Source Controls SOURce CURRent LEVel DIODe lt wsp gt lt value gt Write diode current set point SOURce TEMPerature LEVel DIODe lt wsp gt lt value gt Write diode temperature set point SOURce VOLTage LEVel PIEZo lt wsp gt lt value gt Write piezoelectric voltage set point SOURce WAVElength lt wsp gt lt value MIN MAX gt Write output wavelength set point 41 42 Source Queries SOURce WAV SOURce WAV SOU SOU SOURce SOURce SOURce WAVI SOURce WAVI Elength SLEWrate FORWard lt wsp gt lt value MAX gt Write forward slew rate set point Elength SLEWrate RETurn lt wsp gt lt value MAX gt Write return slew rate set point Rce WAVE Write scan s eng Rce WAVE eng th STARt lt wsp gt lt value gt tart wavelength set point th STOP lt wsp gt lt value gt Write scan stop wavelength set point SOURce CPOWer lt wsp gt lt 0 1 OFF ON gt Constant Power Mode Control Enables and disables Constant Power Mode
55. otor takes care of the rest The unique cav ity design assures continuous tuning with minimal mode hopping Five LED digital readouts display important operating parameters including wavelength and diode temperature Inputs are available for low speed 1 MHz and high speed up to 300 MHz diode current modulation Theory of Design The New Focus 6300 series tunable lasers take advantage of the broad gain bandwidths available in semiconductor diode lasers Most tunable lasers use liquid organic dyes for gain These dyes are notoriously messy toxic and diffi cult to use The all solid state diode laser design by comparision is compact clean and efficient In addition to being widely tunable the 6300 series lasers offer narrow linewidths This is accomplished by using a laser cavity design that began at the Massachusetts Institute of Technology see refs 1 6 The original inven tors had dye lasers in mind when they worked out their cavity we have adapt ed the design to diode laser gain media Our version of the laser cavity is shown in Figure 1 A high reflection coating on one end of the diode laser forms one end of the cavity and a high reflecting tuning mirror forms the other Starting from the diode the beam in the cavity passes through a collimating lens and then strikes a diffraction grating at near grazing incidence The beam is diffracted toward the tuning mirror which reflects the light back on itself for the reverse path
56. ptical power the laser is emitting When the current display is solidly on the laser operates in a single longi tudinal mode However above a certain current level the front panel current display will begin to flash indicating that the laser could be operating multi mode This operating regime allows you to access higher output powers at the expense of possible single mode operation If you require single mode 21 22 operation keep the current below the level where the display begins to flash The current settings for multimode operation may be wavelength dependent Please refer to the included Acceptance Test Data sheet for actual settings At this point you have a basic knowledge of how to use the laser You know how to turn it on change its wavelength and temperature and adjust the power output If you have a way to monitor the laser wavelength with sub angstrom preci sion you will be able to see the effect of the varying the Piezo Voltage control Try turning it while the laser is in Ready Mode You should be able to see the wavelength change smoothly over as much as a few angstroms If you adjust the Piezo Voltage control when the laser is in Track Mode the tuning motor will counter the piezo motion and try to keep the laser wavelength constant Scanning Once you have had a chance to test out the basic operation of the laser you are ready for wavelength scanning The laser cavity is carefully adjusted to give the bes
57. r a wavelength and then tells the laser to tune to that wavelength If you are using a National Instruments GPIB card in your computer be sure to read the file GPIBPC QBASIC README QB for instructions on using the QuickBasic inter active environment with GPIB Other manufacturer s cards will have similar instructions The next line must be included for QuickBasic to work with a National Instruments GPIB card INCLUDE qbdecl bas Clear screen allocate 32 spaces to the string variable buff CLS buff SPACES 32 Find the laser controller at GPIB address 1 CALL IBDEV 0 1 0 T10s 1 0 nflaser used to specify function calls EOS character and EOI mode of the device i I O timeout value 10 sec secondary GPIB address primary GPIB address change if laser is at a different address U GPIB board number Clear the controller s GPIB port nflaser is an integer that identifies the laser It was set by IBDEV above we never need to know its value except that it is set to 1 if IBDEV fails to find the device CALL IBCLR nflaser Ask user for a wavelength in nanometers wave length is a single precision floating point number 775 1 for example INPUT What wavelength would you like nm wave length
58. s are sent as a continuous string For example WAVE 671 84 lt cr gt The controller does not echo the command If you send an action command such as change the current or tune the wavelength the controller will return one of the following responses OK if the command was properly executed Out of Range if a source value is outside the allowed range or Unknown Command if the command was not recognized If the command is a Source Query or a Sense Command the controller will return a value or Unknown Command if the command was not recognized For example you send WAVELENGTH lt cr gt and you get back 671 8 Setting the RS 232 Communications RS 232 communications may be performed at 300 1200 2400 4800 9600 or 19200 baud The data format is 8 bits no parity with 1 stop bit The laser must be in Ready Mode to change the controller s baud rate Press the Local button and hold the Scan Speed paddle switch down at the same time The wavelength control knob then steps you through the baud rates Programming Examples Next we present three elementary programs written in QuickBasic These example programs demonstrate the basics of talking to your controller through GPIB or RS 232 The section that follows these programming examples details the entire command set and describes the use and syntax of all the commands 31 32 Example 1 This simple QuickBasic program asks the user fo
59. t DC Motor Front Angle LC In this design a diode laser is used as the gain medium One end of the Sensor o Remote Local Panel Controls High Speed Current Modulation Input to 300 MHz gt Control In up to z Tuning Mirror Power Monitor Diffraction AR Grating Coating Output Beam diode laser has a high reflectivity coating which acts as an end mirror of the external cavity The other end of the diode has an antireflection coating The diode laser is bonded to a temperature sensor and a thermoelectric cooling block which maintains the diode temperature constant to within 1 millikelvin The laser beam radiating from the diode is collimated by a lens before striking a high quality diffraction grating The diffraction grating is precisely aligned at New Focus and its position is fixed with respect to the diode From the diffraction grating a fraction of the beam is directed to the tuning mirror The position of this mirror determines the operating wavelength of the laser The tuning mirror is mounted on a stiff arm An angle sensor near the pivot point of the arm provides data for wavelength readout The other end of the arm is moved by a DC motor driven screw and a piezoelectric transducer PZT The DC motor makes coarse wavelength changes while the PZT is used for micron scale movements which corresponds
60. t 26 4 returns the target current 63 Command Control Form Query Form Description Explanation Control Form 64 SOURce TEMPerature IEVel DIODe lt wsp gt lt value gt SOURce TEMPerature LEVel1 DIODe Write and read the temperature set point of the diode 16 lt VALUE lt 26 The units are in C This command forces the laser into Remote Mode If the value is out of range e Returns Out of Range e No other action is taken If the value is in range e Returns OK If the laser was in Remote Mode e Remote LED is already on e The diode temperature set point control DAC is driven to a voltage which produces the requested set point within 3 e The piezo voltage and the diode current are unaffected Explanation Query Form If the laser was in Local Mode e Remote LED comes on e Control of the laser is given to the DACs e The diode temperature set point control DAC is driven to a voltage which produces the requested set point within 3 e The current control DAC and the piezo control DAC are driven to voltages which produce values within 3 of the value they had under front panel potentiometer control e The laser is now in Remote Mode If the laser was in Remote Mode e Remote LED is already on e The current piezo voltage and diode temperature are unaffected e Returns the the last diode temperature set point Sourced If the laser was in Local Mode e Remote
61. t tuning performance when tuned from short to long wavelength Therefore when scanning the laser it is best to approach the desired wavelength from the short wavelength side Coarse wavelength scanning is performed with the tuning motor inside the laser head that controls the angle of the end mirror Setting the Start and Stop Wavelengths Press the multipurpose scanning and temperature paddle switch up toward Start and hold it there The Wavelength readout changes when you do this to show you the starting wavelength for scanning You can now change the start of scan wavelength by turning the Wavelength Adjust knob Note keep holding the multipurpose switch up Setting the end of scan wavelength is just as easy as setting the starting wave length This time you hold the multipurpose switch down toward Stop while you use the Wavelength Adjust knob to change the stop wavelength The laser will scan in whichever direction you set it The stop wavelength can be larger or smaller than the start wavelength Setting the Scan Speed The next step is to set the scanning speed If you hold the Scan Speed switch up you will see a number appears on the Wavelength readout This number is the scanning speed from the start wavelength to the stop wavelength The units are in 0 01nm per second To set the scan speed turn the Wavelength Adjust knob while holding the Scan Speed switch up The return scan stop to start speed can also be set while ho
62. trol Board Laser Radiation Out The analog layer consists of four printed circuit boards the current board the temperature board the piezoelectric transducer PZT driver board which controls wavelength fine tuning and the motion control board for coarse wavelength tuning The digital layer includes the microprocessor board and the mother board There is another digital circuit board in the laser head that contains information specific to each laser head such as a wavelength calibration table and the laser head serial numbers The current board is a low noise analog DC current supply which provides up to 150 mA of current of either positive or negative polarity to the diode laser The AC ripple in the output is less than 0 4 pA RMS If desired the current supply can be modulated up to 1 MHz through a BNC jack on the rear panel of the controller For proper operation use only with a cable shorter than three meters The temperature board controls the laser temperature by supplying current to thermoelectric Peltier elements in the laser head Precise temperature control is achieved through the use of a two stage system One thermoelectric element maintains the overall temperature of the laser cavity while another is specifically dedicated to the diode temperature within the cavity In this way the laser temperature is stabilized to within
63. ture shown on the Acceptance Test Data Sheet There may be certain wavelengths near the limits of your laser s tuning range where the light out put power will drop significantly This can happen because of residual etalon effects caused by small reflections at the antireflection coated facet of the diode laser In this particular case you will want to change the diode temper ature a few degrees and the power will be restored See Figure 6 pg 21 Or use the constant power mode to keep the power stable See Constant Power Mode section Figure 6 Changing temperature to avoid power dips a I I l l I I l I I l i 2 C Diode Temperature l l Press the multipurpose scanning and temperature paddle switch down to check the cavity temperature The cavity temperature is factory set and actively regulated to about 30 C If either the diode temperature or the cavity temperature rises as high as 35 C the instrument will shut down If the laser head is exposed to a severe thermal environment you should check the temperatures to be aware of impending shut down Setting Current Power On the far right hand side of the panel you ll see the Current Adjust knob and the Power and Current readouts The current is set with the Current Adjust knob Do not worry about setting the current too high the controller knows the limits of the diode laser regardless of which head you re driving The Power readout shows how much o
64. unsure how your unit is configured check Section IX for more detail Place the laser head on a stable surface with good heat sink capability Point the laser to a safe direction Plug in the two laser head cables into the back of the controller Plug in line power Turn on AC power with the key switch The LED display will show the model number head software revision number and controller software revision number After a few seconds the display will become active Allow 30 minutes for warm up Operation Set the current adjust knob to fully counter clockwise position Press Laser Power button The LED in the button will flash for 5 seconds and then stay on Now you can increase laser current to reach the power you want by turning the current adjust knob clockwise Laser current is hard limited to a safe operation value Before reaching the maximum current the current display will start to flash at some current to indicate that the laser may enter the multimode operation regime The recommended laser operation temperature is listed on the Acceptance Test Data Sheet You can see the temperature set point by lifting the Scan Temperature paddle switch up While the switch is lifted at up position you can use the temperature adjust knob to change it For fine frequency control you can use Piezo Voltage adjust knob Press track button and the LED in the button will light up The laser is now in Track Mode You can adjust wavelength

Download Pdf Manuals

image

Related Search

Related Contents

1 603 - 2.3 Timing Tensioner Tool Instructions  Indoor Fogger Indoor Fogger Indoor Fogger3  Kenwood LZH-81TJ4 Car Video System User Manual  Philips DIS702  BISON: Bio-Interface for the Semi- global analysis  Streamline Series Surgical Tables - User Manual  Mode emploi 10.40 italien  instructions manual analog chronograph with 30 min. counter, day  Manual - Active Sales Co.  Manuel CH_version finale  