1. All items returned for service should be sent postage prepaid in the original packaging or other suitable carton padded to avoid damage Labnet International Inc will not be responsible for damage incurred by improper packaging Labnet International Inc may elect for onsite service for larger equipment This warranty does not cover damage caused by accident neglect misuse improper service natural forces or other causes not arising from defects in original material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED IN DURATION OF 12 MONTHS FROM THE ORIGINAL DATE OF PURCHASE LABNET INTERNATIONAL INC S SOLE OBLIGATION UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT AT LABNET INTERNATIONAL INC DISCRETION OF A DEFECTIVE PRODUCT LABNET INTERNATIONAL INC IS NOT LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGE COMMERCIAL LOSS OR ANY OTHER DAMAGES RESULTING FROM THE USE OF THIS PRODUCT Some states do not allow limitation on the length of implied warranties or the exclusion or limitation of incidental or consequential damages This warranty gives you specific legal rights You may have other rights which vary from state to state No individual may accept for or on behalf of
2. manufacturer that the proposed method will not damage the equipment To clean the centrifuge use a damp cloth and a mild noncorrosive detergent ph lt 8 After cleaning ensure all parts are dried thoroughly before attempting to operate the unit Do not immerse the centrifuge in liquid or pour liquids over it Questions and Service Should you have a question about the centrifuge or require service for the unit please use the contact numbers listed below Do not send in a unit for service without first calling to obtain a repair authorization number and a decontamination form Should the centrifuge require service be sure to decontaminate it and enclose the decontamination form The unit should be properly packed to avoid damage Any damage resulting from improper packaging shall be the responsibility of the user Specifications Dimensions 6 x 6 153 x 153mm Electrical 115V 50 60Hz 130mA PUSS en 0 5A 250V F Speed 6000rpm ROP mmm 2000 x g Ambient Operating Range 0 to 40 C lt 85 Spare Parts Replacement 1 5ml rotor C 1300 RT Replacement strip rotor C 1300 RTS Replacement allen key C 1300 RW Replacement fuses C 1200 F Appendix Symbols and Conventions The following chart is an illustrated glossary of the symbols that may be used in this manual or on the product The electrical warning indicates the presence of a potential hazard which could result in electrical shock CAU
3. Labnet International Inc any other obligation of liability or extend the period of this warranty Mail Warranty Registration to Register online at Labnet International Inc www labnetinternational com 31 Mayfield Ave Edison NJ 08837 lt cut along the dotted line To validate the warranty complete and return this card within 10 days Model Serial No Date Tested Date Rec d PO Name Title Phone Institution Address City State____ Zip Postal Code Country Purchased from distributor How would you rate the quality of this product I Excellent Good O Fair L Poor What feature s on this product made you purchase it What feature s would you change to improve the performance of this product Labnet 31 Mayfield Avenue Edison NJ 08837 P 732 417 0700 N 732 417 1750 Labnet International Inc Labnet Labnet International Inc
4. Spectrafuge Mini Centrifuge ser Manua CE va LIT M00402 C1301 C1301 230V Labnet International Inc July 2012 Rev 2 General Information The Spectrafuge Mini Centrifuge is supplied with two rotors one for microtubes and one for 0 2ml PCR strip tubes The 6 x 1 5ml rotor is designed to centrifuge up to six individual 1 5ml plastic microcentrifuge tubes It will also accept 0 5ml and 0 4ml tubes with the adapters supplied with the unit Adapters are available separately for 0 2ml tubes The strip tube rotor is designed to centrifuge two 8 x 0 2ml strip tubes It will also accept up to 16 individual 0 2ml tubes Both rotors are designed for applications requiring relatively low g forces such as microfiltration cell separation and quick spin downs from the walls of tubes Please read this manual thoroughly before attempting to operate the centrifuge If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Warranty The centrifuge and rotor are covered by a one year warranty This warranty covers the centrifuge and rotor This warranty is void in the event of incorrect operation use of nonstandard spare parts or accessories and unauthorized modification of the rotor or centrifuge The rotor warranty does not cover normal wear and tear or damage and or weakening caused by radioactive contamination cleaning agents or over tightening of the
5. TION This symbol refers you to important operating and maintenance servicing instructions within the product Instruction Manual Failure to heed this information may present a risk of damage or injury to persons or equipment This symbol identifies a Protective Earth PE terminal which is provided for connection of the supply system s protective earth green or green yellow conductor EQUIPMENT DISPOSAL This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be disposed of with unsorted wasie Instead it s your responsibility to correctly dispose of your equipment at lifecycle end by handing it over to an authorized facility for separate collection and recycling It s also you responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the p disposal and recycling of the equipment For more information about where you can drop off your waste equipment please contact your local dealer from whom you originally purchased this equipment By doing so you will help to conserve natural and environmental resources and you will ensure that your equipment is recycled in a manner that protects human health Thank you Labnet International Inc Labnet 31 Mayfield Ave Edison NJ 08837 x P 732 417 0700 F 732 417 1750 Labnet International In
6. allen screw Fuses are not covered by the warranty Exchanging Rotors The centrifuge is supplied with two interchangeable rotors The rotors are easily exchanged using the allen wrench provided Locate the set screw on the rotor shaft below the rotor head Using the wrench loosen the set screw and pull up on the rotor to remove Place the other rotor into position on the motor shaft Take care not to press the rotor down so far as to compress the rubber collar at the base of the shaft Tighten the screw using the allen wrench until finger tight Do not over tighten as this will cause damage to the rotor and the motor shaft Installation Place the centrifuge on a level clean surface position the centrifuge near an accessible power outlet so that the cord and outlet are within easy reach of the operator Operation Before operating the centrifuge be sure that the power switch is in the on position To begin a run simply close the lid of the centrifuge No other controls are required To stop rotation press down on the lid release tab on the front of the unit This will release the lid lock and the rotor will slow gradually to a complete stop Caution Do not attempt to open the lid or remove samples until the unit has come to a complete stop After the rotor has stopped the lid may be opened by grabbing it with the thumb on the front and fingers on the back then lifting the lid back on the hinge Safe operation of the centrifuge require
7. c Email labnetinfo corning com www labnetinternational com NOTES Declaration of Conformity Number CE 00401 Labnet International Inc 31 Mayfield Ave Edison NJ 08837 USA Labnet International declares that the devices described below are in conformity with the EC directives listed In the event of unauthorized modification of any of the devices listed below this declaration becomes invalid spectrafuge Mini Centrifuge Device Name Device Models Numbers C1301 C1301 230V Relevant EC Directives Low Voltage Directive 2006 95 EC EMC Directive 2004 108 EC RoHS 2002 95 EC WEEE 2012 19 EU Harmonized Standards EN 61010 Issued 2001 EN 61010 2 020 2006 EN 61326 1 2006 May 1 2012 x F j j F F s f F E id E a fe W 0 0 j PLI f Le LP Tr Date Peter Will Product Line Manager net Labnet International Inc LIMITED WARRANTY Labnet International Inc warrants that this product will be free from defects in material and workmanship for a period of one 1 year from date of purchase This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in the supplied instruction manual Should this product require service contact Labnet International Inc s Service department at 732 417 0700 to receive a return authorization number and shipping instructions Products received without proper authorization will be returned
8. s that the rotor be loaded in a balanced fashion Failure to load the rotor properly presents a hazardous condition and may damage the centrifuge Individual tubes strip tubes and or adapters must always be loaded symmetrically to ensure proper balance Never run the centrifuge with only one strip tube in place Safety Precautions NEVER use the centrifuge in any manner not specified in these instructions NEVER put hands in the rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flammable explosive or corrosive materials NEVER centrifuge hazardous materials outside of a hood or proper containment facility NEVER continue to operate a centrifuge that is vibrating excessively ALWAYS load the rotor symmetrically Each plate should be counterbalanced by another plate of the same brand type and volume ALWAYS use plates that have been sealed in some manner to avoid spillage ALWAYS load plates so that the sealed side faces the center of the rotor and the bottom of the plate faces out ALWAYS be sure the plates are pushed all the way to the bottom of the rotor slot before operating the centrifuge ALWAYS locate the centrifuge within easy access to an electrical outlet Cleaning Before using any cleaning or decontamination methods except those recommended by the manufacturer check with the