remote scanning electron microscopy

image

Contents

1. N 3 t wv HV Spot WD Mag Det Pressure 7 22 2005 em 20V 40 307mm ie Mic ES EERI Fig 1 A typical user interface of a modern Scanning Electron Microscope 1 When the microscopist plays with the commands the displayed images change in real time Specimen 3 axis shift and rotation magnification focus astigmatism brightness and contrast resolution and depth of focus all affect the images that separately come from different detectors working simultaneously It is common practise to use more controls at a time shift rotation and magnification for instance to individuate centre and enlarge a detail from a low magnification image in a visual sequence that closely resembles the point of view of a landing parachutist As for the latter the microscopist too needs to see and to see in real time every detail of his navigation That need of interaction the high complexity of the microscope and also its great success and diffusion in both the scientific and industrial domains made the Scanning Electron Microscope an evolved instrument highly automated and nowadays completely computer controlled It is the ideal platform to develop and test the remote operation Chapter 1 The remote control of a scientific instrument Even a test for the level of achievement of the results may be individuated for the SEM the complete remote control would NOT allow even an expert electron microscopist to distinguish betwee
2. Temperature Stage Control Temperature Advanced Temperature Actual Target Ramp E C C Min 10 50 Temp Ramp Soak C C Min hh mm ss af ax 17 This frame switches between the Heating Stage and the Peltier Cooling Stage via the two buttons at the top of the frame These functions are used just during particular functioning mode of the Quanta ESEM 200 2 5 6 The Menu bar The menu bar contains the classic menu bar commands like File Window and Help and other particular commands that are used to preset the following microscope parameters magnification accelerating voltage and spot size scan functions and detector functions Figure 5 shows the Menu bar while table 6 explains its functions File Magnificaion Beam Scan Detector Stage Tools Window Help Fig 5 The Menu bar Table 6 A detailed description of each function of the Menu bar File Administrative file functions Magnification Choice of magnification presets Beam Choice of accelerating voltage and spot size presets Scan Associated scan functions 26 Chapter 2 The Scanning Electron Microscope Detector List of detectors and control functions Stage Associated scan functions Tools Image auto functions and useful items Window The image display functions Help About Quanta UI and On line help 2 5 7 The Tool bar Finally there is the
3. UNI VERSI T DEGLI STUDI DI CAGLIARI FACOLT DI I NGEGNERI A DI PARTI MENTO DI NGEGNERIA ELETTRICA ED ELETTRONI CA DOTTORATO IN I NGEGNERI A ELETTRONI CA ED NFORMATI CA XVI II CICLO REMOTE SCANNI NG ELECTRON MICROSCOPY TESI DI DOTTORATO DI FRANCESCA MI GHELA 2006 To my family This thesis is the result of three years of work whereby I have been accompanied and supported by many people I want to express my gratitude to all of them for the help they have given me The first person I would like to thank is Professor Massimo anzi who supervised the research and helping me during the writing of this thesis My gratitude also goes to Professor Daniele Giusto who was always helpful when I needed his advises I shall always be indebted to Cristian Perra and Giovanna Mura for being always willing to discuss with me new ideas about research I would like to thank my colleagues at the Department of Electrical and Electronic Engineering especially the colleagues of the Electronic Group and the MCLab Group for their friendship and support during this period of study Finally I would like to thank my parents my sisters and my friends for the support they have given me during this journey Special thanks go to Alberto CONTENTS INTRODUCTION mcan aid unm eu e Hanae A aa Era emia tod uM DL 1 CHAPTER 1 THE REMOTE CONTROL OF SCIENTIFIC INSTRUMENT 1 1 WHY REMOTE CONTROLLING SCIENTIFIC INSTRUMENTS 3 1 2 THE REMOTE MIC
4. arr 1 Case ZPos txtZpos Text arr 1 Case RPos txtRpos Text arr 1 Case TPos txtTpos Text arr 1 To change these values the remote operator writes the new values in the text box in the Stage frame figure 1 and sends those to the server the instructions are Private Sub cmdSetX Click TCP SendData 32 amp Separatore amp txtXpos Text amp Separatore End Sub Private Sub cmdSetY Click TCP SendData 34 amp Separatore amp txtYpos Text amp Separatore End Sub Private Sub cmdSetZ Click TCP SendData 36 amp Separatore amp txtZpos Text amp Separatore End Sub Private Sub cmdSetR Click TCP SendData 38 amp Separatore amp txtRpos Text amp Separatore 102 Appendix F The implemented code End Sub Stage Current position x 0 765625 SetX Y 32836354 Set Z 26 373694 SetZ R 621 71614 Seth T 0 3609968 mm Fig 1 The Stage frame the text boxes are the same where are displayed the values sent by the server It is not possible to set remotely the T coordinate because its modification is not driven via software When the server receives the new values it forwards them to the microscope changing the stage position Case SetXPos Dim xdXpos As Double Dim xdYpos As Double Dim xdZpos As Double Dim xdRpos As Double Dim xdTpos As Double xdXpos arr 1 xdYpos txtYpos Text xdZpos txtZpos Text xdRpos txtRpos Text xdTpos txtTpos Text objSta
5. See figure 3 VIDEO CODING LAYER Codec Macroblock DATA PARTITI ONING Codec Slice Partition NETWORK ABSTRACTION LAYER a HL H 323 1P C O N T R O L D A T A Fig 3 Structure of H 264 AVC video encoder 58 Chapter 4 The video transmission 4 3 4 The microscope video encoder design As said above two ways are possible to achieve the microscope encoder to use a commercial video encoder developed using the H 264 standard or to develop an encoder ad hoc for the frame grabber video output using the same compression standard 4 The second solution has been chosen for two main reasons 1 The idea is to develop an encoder for a particular video stream knowing all the characteristics of it it will be then possible to develop an encoder extremely efficient and functional 2 It is possible to combine the H 264 features with the microscope video parameters to increase the quality of the video transmission The main condition to develop this encoder is to know the characteristics of the frame grabber video output that is the video that the encoder will elaborate This information has been asked to the instrument Manufacturer but still now no answer has been received For this reason this approach has been currently put on stand by So a new approach has been followed in order to temporary include the video in the remote application 4 4 THE STOPGAP VI DEO SOLUTI ON The principal idea of this new a
6. efficient remote control for a scanning electron microscope moreover it is also possible to extend this technology to other instruments but not only 5 3 1 The improvement of the application Several aspects of the developed application can be improved It is possible to improve the characteristics both the client and server applications The security aspect has not yet implemented as well as the implementation of different versions of the client application on the base of the client 68 Chapter 5 Discussion perspectives and conclusions kind In fact an observer could not manage the instrument while an expert microscopist must to have access all the microscope commands If the Manufacturer will give the information about the output video of the frame grabber it will be possible to develop and implement the encoder described in the fourth chapter This will increase the video quality and the video streaming performances reducing the delay transmission On the other side the networks become faster and faster and the available bandwidth larger and larger then it will be possible to have a better connection with a real time transmission and a very good video quality 5 3 2 The extension of the remote microscopy One of the goals of this study is to propose a method for remote microscopy that may aim to build up a reference for the scientific community As explained there is no standard tool developed for the microscope but th
7. has also to introduce the specimen in the chamber and to start the vacuum pumps to evacuate the air from the chamber When the chamber has reached the desired pressure the remote operator may turn on the high voltage and he can see the image of the specimen The next operation as described in the previous chapter is the calibration of the z axis This operation is not performable as a remote operation then the local operator has to do it After the calibration the remote operator will be able to focus the image only focusing the electron beam The other commands to improve the image quality are enabled since the beginning and the remote operator can use it as if it is in front of the instrument It is simple to note from this description that the use of the remote microscope is not so much different from the local use The commands included into the user interface are standard commands and each microscopist may use it without knowing the kind of the instrument 3 8 COMMENTS The project as described up to now achieved the target of developing a dedicated application for remote controlling a SEM The important chapter of the video handling has yet to be presented Anyway it may be time to compare the 48 of the project achievements here reported with the examples reported at the beginning of the chapter The first example University of Illinois is similar to the first solution here studied it is a web based applicatio
8. a EAT Hd Fa HE p EX ne 68 5 3 1 The improvement of the application 68 5 3 2 The extension of the remote microscopy 68 FINAL CONCLUSIONS an err cett da site a bos p E abe e d PEOR 70 Ux nrc ee e o E rt Pe E Rd annealed Rane du duane dead AE a det 71 APPENDIX A A 1 REFERENCES APPENDIX B B l B 2 B 3 REFERENCES APPENDIX C Gal REFERENCES APPENDIX D D 1 D 2 REFERENCES APPENDIX E ET REFERENCES APPENDIX F F 1 Contents THE ORIGIN OF THE ELECTRON MICROSCOPY THE HISTORY OF THE MICROSCOPY a aosectthspetrbateeomer tend ao tod Roof fes tiui edi dormis 72 a hue dh teet p UM Rt ERU EOD duet Spotted ter E Red E 75 THE ELECRON MICROSCOPES THEEEECTRONOGUM 2nd oid rete oh Conde dulce nian comae t cmo Lac ee 76 THE ELECTROMAGNETIC EENSES 25i sscss hire nea od aret ore d o UH RO dar 78 THE EVERHART THORNLEY DETECTOR agas bct teo taque ePaper Era iaa 80 TT M 82 ELECTRON INTERACTIONS IN A SEM SPECIMEN INTERACTION Ss esistenti inanan Publ von booa GR rund eae Matas 83 Tom TEE E 86 THE ENVIRONMENTAL SEM THESENVIRONMENTAE SEM udo abd cs ea oec ie EUR rado tt tots 87 THEJXOUANTA E SEM sasietan Ges quein eue opu apu qic etes Fr appeals 89 D 2 1 The low vacuum and ESEM working modes 90 T ETE 92 THE TRANSMISSION ELECTRON MICROSCOPE THE TRANSMISSION ELECTRON MICROSCOPE STRUCTURE eene 93 E 1 1 The functioning of the TEM column 94 E 1 2 Specimen Interactlolls uu ois ara erdt abt Mrs hast 95 E 1
9. and Source Tilt it is possible to set the effective angle of illumination of the beam coming from the gun area of the electron column The last section of the frame has a timer to monitor the filament life When a new filament is placed the timer starts to count the hours when the filament is switched on Reset Timer is the button for resetting the timer value Status Vacuum Status Vacuum a High Voltage 20 0 kv Pressure 3 1e 6 mbar Filament Current 2 24 A Emission Current 101 pA 2 5 2 The Work page The frame Status shows the value of some parameters during the microscope working process to monitor the state of the instrument The parameters shown are the Vacuum state of the specimen chamber the high voltage value the pressure value of the specimen chamber the filament current value and the emission current of the filament The second page is the Work page and is different from the first because instead of the Electron Column commands has the Stage Location commands which are the commands to move the stage and then the specimen The following table shows in detail these commands 22 C h apter 2 The Scanning Electron Microscope Table 2 A detailed description of the Work page Stage Map Coordinates n mmm m Location Open Save __Clear_ Current Position CK IGBT Clear This frame allows to change the coordinate of
10. application will be installed into the remote operator s computer and will connect to the server through the customer LAN The first step has been the classification of pro and cons of this solution just to test the convenience and the efficiency The pro identified are l Interactivity 36 of the project 2 Friendly interface 3 Reliability 4 Security The only negative aspect is the portability The interactivity and the friendly interface are due to the programming language chosen The Visual Basic language in fact is specified for developing applications that need a user interface and then it is simple to implement user interfaces easy to use The relevance of this possibility is that a friendly interface enables the remote user to understand immediately the application use without reading long manual or complex instructions facilitating the interactivity between operator and microscope Developing two applications a server and a client able connecting between them it is easier to control the reliability and security of the application First of all only the allowed users can access to the instrument through a login with user ID and password assigned by the microscope administrator Second an encrypted communication to send the information will be used in order to mask the sender and the receiver The portability is seen as a disadvantage because the applications will be developed for the Windows Operating System
11. been chosen and as a consequence of its performance it has not been changed This is the Windows Media Encoder WME To use efficiently this encoder a number of parameters have to be set the picture showed in figure 5 is the setting window of WME Some of the indicated parameters have been fixed to a particular value some others have been changed in order to find the best combination with the video player characteristics The fixed parameters are the frame rate 5 frames per second the key frame interval 30 seconds the video smoothness 85 and the decoder complexity The other parameters are the video size and the video bit rate and are the same properties indicated in the table above For these parameters a number of values have been defined and the tests have been carried out changing those The values sole Chapter 4 The video transmission showed in the table are 160X100 320x200 and 640x480 for the video size 20 Mbps and 4 Mbps for the video bit rate and 1 second or 2 seconds for the buffer size Custom Encoding Settings General 20 Mbps Video size 320 x 240 C Same as video input Frame rate 5 fps Key frame interval Video bit rate bps Buffer size seconds C Use default Video smoothness Ea Sharpest Decoder complexity Simple Total 20 01 Mbps 20009000 000000 bps Video bit rate 20 Mbps Audio bit rate Script bit rate Overhead 9 Kbps Fig 5 The WME sett
12. evacuated the system allows switching on the high Pressure t n ion m 0 60 mbar Selecting the Vent button the pump down process is stopped and the specimen chamber is ventilated Before to do this the high tension has to be switched off m Electron Column This frame regulates the electron beam Spot Size parameters The HV button is used to switch on off Ww el 40 the filament current the High Voltage scroll bar allows to set the accelerating voltage of the High Voltage electron beam from 0 to 30 KeV the Spot Size 20 00 kV scroll bar allows setting the beam size from 1 to 10 m Detectors The frame Detectors contains the scroll bar for Contrast 35 8 adjusting the contrast and brightness values to optimize the displayed image The minimum value is 1 and the maximum is 100 for both the Brightness 17 2 parameters 21 Chapter 2 The Scanning Electron Microscope m Source Control Wehnelt Changed M Limit Voltage Filament Voltage 2 03 Y M Autobias On Bias 57 gt Source Tilt d Cross Over Store Reset Timer Filament 8814 Hours In this frame there are all the commands to saturate the filament after its substitution After finding the new saturation condition the voltage of the filament may be limited for the filament security After the saturation with the command Cross Over
13. from the aperture column crossed by the electron beam For this reason it is advisable to perform a preliminary calibration of the specimen height moving it mechanically in order to perform the other focus operations only through the electromagnetic lenses After this calibration it is possible to start the real analysis changing the specimen height the image pan scroll and magnification setting the contrast and brightness value to obtain relevant images where scientific information is often coupled with photographic quality Other operations can be performed to improve the image quality In fact the image quality is related to the resolution property and on the noise level as said before the resolution in a SEM is determined to a large extent by the beam diameter on the surface of the specimen the smaller the beam the sharper the image Anyway the beam size also defines the beam intensity the original stimulus for any emission signal In that case the smaller the beam the lower the intensity that is decoupled with a worse signal to noise ratio It follows that a continuously adapting procedure drives the microscopist s job beautiful nearly noiseless images may be obtained at low magnification when low resolution is sufficient which correspond to the exploration phase of the whole specimen while imaging details at higher magnification requires higher resolution smaller beams lower intensities and high noise In this case the scanni
14. functioning of the TEM column The TEM column can be considered divided into two separated systems the system upper the specimen and the system lower it The first works as explained in the Appendix B and is the same as a SEM The second consists into an objective lens an objective aperture a selected area aperture two intermediate lenses a projector lens and a phosphor screen The objective lens forms an inverted initial image which is subsequently magnified In the back focal plane of the objective lens a diffraction pattern is formed The objective aperture can be inserted here in the black focal plane of the image Its function is to e Select those electrons which will contribute to the image and thereby affect the appearance of the image e Improve the contrast of the final image By inserting the aperture or tilting the beam different types of images can be formed 94 Appendix E The Transmission Electron Microscope The most common conditions are e No aperture the diffraction pattern is centred on the optical axis e Aperture is centred on the optical axis e Aperture displaced selecting a diffracted beam e Beam is tilted so that the diffracted beam is on the optical axis The first intermediate lens magnifies the initial image that is formed by the objective lens The lens can be focused on the initial image formed by the objective lens or on the diffraction pattern formed in the back focal plane of the objecti
15. of the lens See figure 2 Electron Source Soft iron pole piece Rays path Copper coils Image is inverted and rotated Fig 2 Functioning of the electromagnetic lenses 78 Appendix B The hardware of electron microscopy Focal length can be altered by changing the strength of the current All electromagnetic lenses act like thin convex lenses their thickness can be ignored for most purposes The most important property of a thin lens is its focal length f Parallel rays entering the lens are brought to focus at the distance f Simple geometry shows that the magnification M of the lens is M u The parameters u v and f are related by the thin lens equation 1 1 F ze In microscopy lenses are often used to reduce the diameter of the beam Figure 3 shows how the electromagnetic lenses are placed into the electron microscopes both SEM and TEM The double condenser system consists of two or more lenses and an aperture Its function is to control spot size and beam convergence Virtual Source First Condenser Lens Second Condenser Lens Condenser Aperture Sample Fig 3 The electromagnetic system placed in the electron microscopes Two or more lenses can act together and their ray diagrams can be constructed using the thin lens approximation for each of them The first condenser lens has to e Create a de magnified image of the gun crossover e Control the minimum s
16. processed by a Variable Data Rate Video VDRV processor This kind of processors is able to vary the amount of data processed per frame to match image quality and transmission bandwidth requirements The output of this processor is the input of the frame grabber card The role of the frame grabber is to decouple the video input and output formats allowing to read the frame buffer according any supported video standard although its live update usually proceeds at highly non standard rate number of pixels per line and per frame pixel line and frame times etc from the VDRV processor The frame grabber installed in the 52 Chapter 4 The video transmission microscope used in this project has four inputs and each input may be independently used In the QUANTA platform three of these usually have as input the output of the multi channel VDRV processor In this way different live images from different detectors may be simultaneously displayed The fourth input is usually the video output of the digital camera that looks at the specimen chamber from inside as shown in the images in the previous chapters The frame grabber video output is then sent through the PCI bus to the video card and it will be overlaid onto the windows background The LCD monitor of the microscope controller is connected to the output of the video card The following picture sketches a LCD monitor view here it is possible to see the four quads that reproduce the four v
17. program is Chapter 2 The Scanning Electron Microscope Selecting this button the recording of a video from the three quads with the specimen images will start at the same moment Clicking this button the On line help window will be opened e 7 2 6 WORKI NG WI TH A QUANTA MI CROSCOPE As described above a Quanta microscope has several functions Some of those are standard functions of a scanning electron microscope others are related to the particular model of instrument Quanta ESEM 200 This research work aims to give the origin to the remote microscopy developing a remote application independent from the microscope used as test bed instrument With regard this aspect it is interesting to see how the normal procedures to carry out an analysis with a SEM are performed with a Quanta microscope The first operation is to ventilate the chamber to insert in the specimen To do this it is sufficient push the button Vent contained in the Startup page a dialogue box will ask Really vent the chamber choosing Yes the pumps are automatically stopped and the valve to introduce the gas into the chamber is opened This dialogue box is a secure system which should avoid ventilating the chamber accidentally During the ventilation phase in the box Status included into each page the Vacuum Status is Venting the quad is not green but yellow and the value of pressure increases r
18. shadows considering a normal photograph The different shadows position can be obtained by rotating the specimen under the same detector and beam scanning condition This operation is performed by sequentially rotating the stage on which the specimen is acquiring a corresponding set of images shifting the specimen again to keep the relevant details into the imaging field by automated image driven navigation The Library used to implement the remote application contains also the routines to remote control the stage movement included in the remote application Then using the same routines it will be possible to implement this as a console application and to make it a remote application The utility of the images obtained is in the topography study of the specimen surface in fact displaying those with particular visualizing programs it will be possible to change the point of view of the image and to see the specimen surface from each side it gives a 360 vision Relevant results obtained by the complete automation of this application have been already obtained and published the research continues developing each potential of the application and probably the remote version 5 3 WORKING PROSPECTS The research work has carried out its plan the remote microscope Nevertheless the entire research project could not be considered finished There are several aspects that can be improved implemented or modified in order to realize a complete and
19. the actual platform to use for developing remote control mainly depend on non technical conditions as the availability of the instrument at lab or starting with the acquisition of a new machine its cost completeness suitability for different kind of users and so on Not the least for the specific application it is crucial the availability of programming libraries and the possibility to have by the Manufacturer the access to the proprietary codes of the software The chosen instrument is a Quanta ESEM 200 made by FEI It is an Environmental Scanning Electron Microscope ESEM that is a SEM able to introduce some partial pressure air water vapour or other gases in the specimen chamber preserving the SEM operation The instrument belongs to the latest evolution of the Scanning Electron Microscopy that brooks the constraint of observing dry conductive specimens which heavily affect the direct observation i e of biological specimens The key point is that the residual gas around the specimen from one side neutralizes the beam induced charging effects on dielectric specimens and on the other allows the observation even of wet specimens reaching the fascinating limit of observing water droplets by an electron beam Anyway the kernel of the operation is the same as for a standard SEM and for the sake of generality the project focused on standard high vacuum operation of the Quanta platform The tele microscopy project is started tog
20. the same Operating System of the PC microscope controller Then to use the application with another Operating System another dedicated application has to be developed The video elaborating system is realized using a video encoder together with the server application which takes the video card output and a video player together with the client application which is able to play the video sent by the encoder A more detailed explanation of this aspect will be given in the next chapter 3 4 2 The web based approach This alternative solution aims to achieve a web site using the Active Server Page ASP technology in conjunction with the HTML language The reason of these choices is that the ASP technology supports the Visual Basic Script language that is a programming language similar to the Visual Basic language but easier the HTML language will be used to develop the user interface V g of the project The project establishes that the web server is the same computer that should support the server application of the first solution Then in this computer an Internet Information Service IIS of Windows will be activated The access to the web site allows the access to the microscope the web site access will controlled giving a user name and a password to the remote operation and using the HTTPS protocol to transfer the microscope web pages This protocol is the protocol HTTP Hyper Test Transfer Protocol used in conjunction with the Sec
21. tool bar at the top that contains some new commands like the Auto Focus function or the Auto Contrast and Brightness function Some other commands present also into the menu bar are available this is for giving a faster access to those commands Figure 6 shows the tool bar while table 7 gives a description of each command a Tm A gla cx 9 4x4 153 Fig 6 The Tool bar Table 7 A detailed description of each function of the Tool bar oO Y Auto functions these functions allow the auto contrast and a ae brightness and the auto focus regulation a i These functions give some more useful tools to adjust the contrast and brightness values and the astigmatism correction A These commands concern the Magnification correction and the Z axis to Free Working Distance correction 4 P These functions allow changing the scan velocity 1024x884 This section contains the possible value of the pixel resolution of images displayed or recorded id ii These three functions are the Beam Blanker which allows to deflect the beam off axis and protect the specimen the Snapshot to acquire the images the Pause to stop the beam scan on the specimen When the scan is paused a symbol of pause is showed onto the image quad gi afl These three symbols represent the possible filter functions ead applicable to the scanned image E This command allows sharing the desktop with another PC 3 using the VNC
22. useful analogy with the optical microscopy the concepts of source lenses optical axis moves to describe a SEM structure by its optical analogous as for the SEM column in figure 1 the real structure of the electron gun and the electromagnetic lenses is summarized in Appendix B At the top there is the electron beam source Three condenser lenses focus the beam electrons into a fine spot less that 4 nm in diameter on the specimen and two apertures are also introduced to cut the off axis electron trajectories It should be noticed as the beam divergence is here by far exaggerate the real divergence at the final part usually not exceeding 10 rad irtual Source zs First Gendenser Leas Condenser Aperture Second Condenser Leas Objective Aperture xq Scam Ceals Objective Lens Sssaple Fig 1 The SEM column The name of the lenses and of their related apertures is surely misleading because mixes the analogous optical terms of the light projection system condenser lens and of the imaging tool objective but is the currently accepted convention The scan coils are situated between the objective aperture and the objective lens in order to induce wide deflection at the specimen level by imposing only small movements to the beam that are then amplified by the lens 13 Chapter 2 The Scanning Electron Microscope The electron energies in a SEM given by the gun accelerating voltages range from 200 to 300000 v
23. very good signal is acquired from the sample and the intensity of that signal is converted into a brighter or darker portion of the image at a given point x y on the specimen as the electron beam moves across it D 2 THEQUANTA ESEM The Quanta ESEM is a microscope of new generation because it has the environmental working mode in addition to the high vacuum mode and the low vacuum mode Then the vacuum system is particular because it has to guarantee the possibility to have the specimen chamber and the gun electron column at different pressure values In High Vacuum mode both sections are at high vacuum pressure in the other working modes Low Vacuum and ESEM the gun electron column is always at high vacuum pressure while the specimen chamber is at the required pressure 2 Figure 2 shows the entire vacuum system All valves are automatically controlled except the environmental backing valve EBV This valve must be opened for Low Vacuum and ESEM modes and closed for High Vacuum mode Depending on the mode selected a different detector is used the large field detector LFD for the Low Vacuum mode and the gaseous secondary electron detector GSED for the ESEM mode In the first chapter a description how the SEM normally works High Vacuum mode has been given now a brief description of Low Vacuum mode and ESEM will follows 89 Appendix D The Environmental SEM EBV manual Fig 2 The Quanta ESEM 200 vacuum syst
24. 1 The SEM specimen interaction classification Backscattered electrons The production of backscattered electrons varies directly with the specimen atomic number This differing production rates causes higher atomic number elements to appear brighter than lower atomic number elements This interaction is utilized to differentiate parts of the specimen that have different average atomic number Secondary electrons The production of secondary electrons is very topographic related Due to their low energy 5 eV only secondary electrons that are very near the surface 10 nm can exit the sample and be examined Any changes in topography in the sample that are larger than this sampling depth will change the yield of secondary electrons due to collection efficiencies Collection of these electrons is aided by using a collector in conjunction with the secondary electrons detector The collector is a grid or mesh with a 100 V potential applied to it which is placed in front of the detector attracting the negatively charged secondary electrons to it which then pass through the grid holes and into the detector to be counted Auger electrons They are due to the de energization of the specimen atom after a secondary electron is produced Since a lower usually K shell electrons was emitted from the 84 Appendix C Electron interactions in a SEM atom during the secondary electron process in inner lower energy shell now has a vacancy A higher ene
25. 3 Observation and recording of images 98 C RT p 99 THE IMPLEMENTED CODE AN EXAMPLE THE STAGE FUNCTIONS nicae dida vk tnde dia ird een 100 NTRODUCTI ON The first remote controlled object was a German boat during the World War Until the World War Il the remote control had just military employment only at the end of the WWII the researchers used this technique for civil applications In the late 1940 s automatic garage door openers were invented and in the 1950 s the first TV remote controls were used Beginning from these primary studies and applications the remote controlled instruments are today diffused into several fields of applications with different kinds of remote control radio control ultrasonic control laser control mechanical control and infrared control which remote control is used depends on the application For example the infrared control uses a photodetector in most televisions VCR s stereos and car audio system radio control is used in controlling remote control vehicles like planes boats cars Another kind of remote control may be now considered coming from the development of network technology the web control or internet control it is used to control some instrumentation like personal computers or other instruments through the internet network It is from these new technologies and the results obtained with the other remote instruments that the remote microscopy has been considered feasible In t
26. M Working at high vacuum pressure before to introduce a biological specimen into the chamber it has to be dessicated and coated with a layer of conductive material On the other side this process could change its original characteristics The microscope name expresses the fundamental difference between the new instrument and the classic SEM an ESEM may be operated with a poor vacuum up to 10 Torr of vapour pressure or one seventy sixth of an atmosphere in the specimen chamber In such wet mode imaging the specimen chamber is isolated by valves pressure limiting apertures and a large diameter bypass tube from the rest of the vacuum system 1 Water is the most common imaging gas and a separated vacuum pump permits fine control of its vapour pressure in the specimen chamber When the electron beam primary electrons ejects secondary electrons from the surface of the sample the secondary electrons collide with water molecules which in turn function as a cascade amplifier delivering the secondary electron signal to the positively biased Appendix D The Environmental SEM gaseous secondary electron detector GSED Because they have lost electrons in this exchange the water molecules are positively ionized and thus they are forced attracted toward the specimen which may be nonconductive and uncoated serving to neutralize the negative charge produced by the primary electron beam The field emission gun produces a brighter filament i
27. ROSCOPY aii keep but cs a a rd 5 lidjcsid src tanas tats EAn E EE E EA A A A a a 9 CHAPTER2 THE SCANNING ELECTRON MICROSCOPE 2 1 THE OPERATING PRINCIPLES OF A SCANNING ELECTRON MICROSCOPE 10 2 2 THE STRUCTURE OF THE SCANNING ELECTRON MICROSCOPE 12 2 3 WORKING WITH A SCANNING ELECTRON MICROSCOBPE seem 14 2 4 THE EXPERIMENTAL PLATFORM snide rn rn tara pd eo a on drei etr 16 2 5 THE MICROSCOPE SORT WARE cicouceath ek irte rite ebat nd Sut xd 19 2 5 1 The Startup Dae oi tossed ahi tls Ma nan phate tu ede 21 25 2 The Worte Dade noia tod terti etc UD dai qnte ead 22 2 5 3 The Option page ci heated Sa Ruch ca sah ral Sie oe ee 23 2 5 4 The Alignment Dage ucau mec eed eite toe but ictus todius 24 2 5 5 The temperature pages sepes ertt rop pct Seton 25 25 0 The Menu Dar dudo otc ier sebo m ettari a P meti 26 25 7 Jhe Tol Haiya iaioa idebsaiedutadtdale ota 27 2 6 WORKING WITH A QUANTA MICROSCOPE sse 28 PREP ER ER snc M 31 CHAPTER 3 THE DEVELOPMNT OF THE PROJ ECT 3 1 THE REMOTE ELECTRON MICROSCOBY 15i iade et atcur Robur dba id 32 320 DBETREASIBILEPISTUDY coats oco oe teni pea aine uode do O 34 3 3 3 4 3 5 3 6 3 7 3 8 REFERENCES CHAPTER 4 4 1 4 2 4 3 4 4 REFERENCES CHAPTER 5 5 1 5 2 5 3 5 4 REFERENCES Contents DEFINITION OF SPECIFICATIONS ves sisecscacsaa dpa pr pot et pot matin uh 35 IHE DESIGN cinta asd onini a aede Fiteo hee
28. SCOPE STRUCTURE There are four main components to a Transmission Electron Microscope an electron optical column a vacuum system the necessary electronics and software The column is the crucial item It comprises the same elements as the light microscope as can be seen from the ray paths of light and electrons sketched in figure 1 The light source of the light microscope is replaced by an electron gun which is built into the column The glass lenses are replaced by electromagnetic lenses and the eyepiece or ocular is replaced by a fluorescent screen The entire electron path from gun to screen has to be under vacuum otherwise the electrons would collide with air molecules and be absorbed so the final image has to be viewed through a window in the projector chamber Another important difference is that unlike glass lenses electromagnetic lenses are variable by varying the current through the lens coil the focal length which determines the magnification can be varied In the light microscope variation in magnification is obtained by changing the lens or by mechanically moving the lens Appendix E The Transmission Electron Microscope Virtual Source First Condenser Lens Seeond Candeaser Legs Condenser Aperture Sam ple Objective Lens Objective Aperture Selected Area Aperture First Inter mediate Lens Second Infermediate Lens Projector Leas P ELE g bein Screen phespher Fig 1 The TEM column E 1 1 The
29. Sub IbIZ Click TCP SendData 35 amp Separatore End Sub Appendix F The implemented code Private Sub IbIR Click TCP SendData 37 amp Separatore End Sub Private Sub IbIT Click TCP SendData 39 amp Separatore End Sub The server receives the requests and forwards it to the microscope controller when the server receives the answer displays the values and sends these to the client Instructions to get and display the coordinate values Case GetCurrentPosition Dim getXPos As Double Dim getYPos As Double Dim getZPos As Double Dim getTPos As Double Dim getRPos As Double objStageControl GetPosition getXPos getYPos getZPos getTPos getRPos txtXpos Text getXPos txtYpos Text getYPos txtZpos Text getZPos ixtRpos Text getRPos txtTpos Text getTPos Instructions to send the coordinate values to the client Case XPos TCP1 SendData 31 amp Separatore amp txtXpos 1000 amp Separatore Case YPos TCP1 SendData 33 amp Separatore amp txtYpos 1000 amp Separatore Case ZPos TCP1 SendData 35 amp Separatore amp txtZpos 1000 amp Separatore Case RPos TCP1 SendData 37 amp Separatore amp txtRpos 1000 amp Separatore 101 Appendix F The implemented code Case TPos TCP1 SendData 39 amp Separatore amp txtTpos 1000 amp Separatore When the client receives the values displays them using the following instructions Case XPos txtXpos Text arr 1 Case YPos txtYpos Text
30. a computer and accessing the web In the same way many more or less recently developed microscopes belonging to the scanning probe family may be technically ready for remote control because of the significant automation of their functions but on one side require very special manual setup of the same machine to perform different kinds of investigations Atomic Force Microscopy Scanning Probe Microscopy Scanning Capacitance Microscopy Scanning Near field Optical Microscopy etc and on the other hand they have an image acquisition time that is rather slow reducing the effective possibility for visual navigation Of course one technique must exist to make the remote microscopy challenging but feasible and attractive This is the case of Electron Microscopy Figure 1 illustrates a typical user interface of a Scanning Electron Microscope SEM of new generation As showed this is a software window this means that a modern SEM is completely except for some particular functions driven via software The window of the software that controls the microscope is divided into two parts one is that with the commands to control the instrument at the upper and right edges of the frame and the other is that with the images of the specimen observed by the microscopist Chapter 1 The remote control of a scientific instrument Moussa tye 7 oly O60 mbar Electron Columa EM sew 2 HV eral wo Ma e 200K 4 0 207mm 0 880
31. adio links or even cellular phone lines Even video based devices are not demanding under the remote control point of view when they observe slow events the acquisition of a chemical maps by several machines X ray Fluorescence Auger Electron Spectroscopy Secondary lon Mass Spectrometry etc lasts several minutes and may be set up and its results mailed on the current networks by simple operations The challenging case for which only partial achievements have been reported is that of an instrument where the operator must interact with a live image to reach the proper setup of the instrument and to navigate into the field of the experiment The previous example of remote driving is perfect nobody thought in designing a remotely controlled vehicle for Mars because any command raised by the appearance of a sudden obstacle would have a reaction time not shorter than 8 minutes Inside a scientific laboratory a whole class of instruments belongs to the challenging level the microscopes 1 2 THE REMOTE MICROSCOPY Different from many other instruments a microscope requires the live interaction of the operator with the instrument to adjust magnification focus and other optical parameters and then to explore the surface of the specimen sometimes running across wide not interesting areas sometimes dwelling upon a region changing the previously set focus magnification and optical corrections looking for minute details that may be cruc
32. ain and firm Once the choice of the instrument for the test bed brought the project on an experimental ground it has been possible to identify the development of the video architecture as the most critical aspect because of the required features real time and full resolution transmission Video handling has proven indeed troublesome but mainly for some unexpected situations than for the strictly technical challenge The instrument was purchased under the constraint that everything necessary for remote controlling it were or available or made available by the Manufacturer The technical specifications of the machine duly listed a sequence of computer controlled functions the illustrated internal LAN architecture the availability of pertinent and tested interfacing libraries and implied the full suitability of the hardware for executing the required operations Chapter 5 Discussion perspectives and conclusions The technical bad surprise was the architecture of the video The composition of the image on the LCD screen as reported in chapter 4 was made by simply and roughly overlaying the output of the Viper Quad frame grabber onto the graphical output of the video board of the SEM Controller PC Even worst while the standard SVGA output of the PC video board was regularly accessible the video stream from the frame grabber was only available as an analog composite video No hardware access or software possibility to read the video memory
33. al operator without any data loss and without any delay This simple statement makes the real challenge of the whole project and for this reason a significant part of the research work has been dedicated to it The crucial point is that a video stream needs a compromise between bandwidth and data compression or even data loss Moreover any video encoding introduces some time delay Both things keep the remote user actually far from the instrument In order to find a solution the microscope video architecture has been studied 4 2 THE MICROSCOPE VI DEO ARCHI TECTURE The image of a Scanning Electron Microscope is created measuring the intensity of the current of secondary electrons or backscattered electrons emitted from the specimen under the stimulus of the electron beam Several different detectors have been developed for the electron microscopes and the most celebrated one in Scanning Electron Microscopy the Everhart Thornley scintillator photomultiplier is described in Appendix B In any case the detector outputs an electric signal proportional to the collected electron intensity as the beam scans the specimen surface the signal changes according to the variation of the electron emission In the last generation instruments the analog response of the detector is sent to an analog to digital converter whose output is used to write the video memory according with the scanning sequence of the image field The signal thus generated is
34. apidly When the chamber is totally ventilated the Vacuum Status shows Vented and the quad is red Now it is possible to open the chamber pushing the frontal side and to place the specimen on the stage and close again the chamber The next phase is to activate the vacuum pumps pressing the button Pump on the Startup page Then the pumps will activated automatically and the Vacuum Status will be Pumping while the pressure value on the Status box will decrease rapidly When the vacuum pressure value is reached the Vacuum Status will be Vacuum and the button to turn on off the High Voltage will be enabled At this point it is possible to turn on the high voltage pressing the button HV on the user interface and the first images of the specimen will be displayed on the monitor of the microscope controller computer Also in this case the image usually is not well focused and properly balanced 28 Chapter 2 The Scanning Electron Microscope To focus the image it is necessary to calibrate the height of the specimen that is the z axis with the electron beam focused before to perform this operation the automatic movement of the stage along the z axis is forbidden while the movement is allowed along the other axes The Stage Location box into the Work page shows the value of each coordinate then reading here the z coordinate value it is possible to focus the electron beam at this point To complete the calibration oper
35. apter 2 The Scanning Electron Microscope Fig 2 The Quanta ESEM 200 Microscope e co ae Keyboard Mouse rr Support Network Microscope Controller HHEEEFHENIN Minolta color Hub 1 Part of EDonedte 6100 DeskLasor Fig 3 The architecture of the Quanta local network 2 18 Chapter 2 The Scanning Electron Microscope The architecture of the machine aims to allow any microscopist to perform his analyses storing and processing his images possibly executing other software applications if required but preserving the instrument from possible operation conflicts To this purpose a local area network LAN is made of two computers one the Microscope Controller where the operating Quanta software is installed and the other the Support PC dedicated to the operator s utilities see figure 3 2 5 THE MICROSCOPE SOFTWARE On a practical ground and beyond the commercial celebrations the instrument is indeed an excellent SEM and at least under the technical point of view an ideal platform for the proposed experimentation The software interface is friendly and immediate the commands are distributed following their function and their relevance Figure 4 shows a snapshot of it The four quads represent the image taken from different sources into the specimen chamber The upper image on the left represents the image acquired from the secondary electrons detector the upper image on the right side comes from the bac
36. ation it is necessary to push the button for the Z axis to Free Working Distance Correction included into the Tool bar Now the z axis and the focused electron beam are calibrated and the focusing operation of the image will be performed just operating on the lenses the automatic movement along the z axis is enabled too The Quanta software allows focusing the image also pressing the right mouse button and moving the mouse to the left or right on the image This is a specific Quanta operation that simplifies a lot the focusing operation The other operations to optimize the quality image are performed using the buttons in the user interface To set the contrast and brightness values the scroll bars in the Detectors box have to be used The high voltage value and the spot size values may be set from the Electron Column box The magnification can be changed into three modes clicking the Magnification menu in the Menu bar and choosing a value scrolling the scroll bar in the Option page or pressing the and keys on the keyboard Also this last way is a specific Quanta solution The system of images acquisition consists in a digital device which acquire the images as if it is a digital photo camera Before to acquire the image it is possible to set the pixel resolution choosing the size of the acquired image and to set the scan speed slower is the scan speed higher will be the resolution of acquired image The buttons to set th
37. ba id asdb A 36 3 4 1 The server client approach sene 36 3 4 2 The web based apprOOEtissu Renee tease etg t bora declan 37 THE IMPLEMENTATION OF THE WEB BASED APPLICATION 38 3 51 The debugging plase dest orn eoi a ebbe tatam rdg 40 THE IMPLEMENTATION OF THE SERVER CLIENT APPLICATION sisses 40 3 6 1 The debugging phase sod ca enia Dre tho o Mah 47 WORKING WITH A REMOTE SCANNING ELECTRON MICROSCOPE ssese 48 COMMENTS Sining ne ure aaepe dtl edrdbd od cut del ced 48 Et na ro b ree RE Re FS Ee Ro Re et RR UTR re ERA vada AT ERES 50 THE VIDEO TRANSMISSION THE MICROSCOPE VIDEO s xil hod rubent cr ation tai E cab Ri 51 THE MICROSCOPE VIDEO ARCHITECTURE 25 2 t 0 2 4 2 GAL BREUI i Ud de 52 THE VIDEO ARCHITECTURE DEVELOPMENT 2 5 tuti tarte ca ad 54 4 3 1 The system properties odds deo P Happen pe P in utes 54 4 3 2 The video architecture design esses 55 4 3 3 The H 264 AVC video coding standard sss 57 4 3 4 The microscope video encoder design sss 59 THESSTOPGAP VIDEO SOLUTION az aevitirivt i torbetande qid dut td bornes 59 4 4 1 The video application tests cesses eee 60 phe 64 DISCUSSION PERSPECTIVES AND CONCLUSI ONS ABOUT SOME UNEXPECTED TECHNICAL AND NON TECHNICAL DIFFICULTIES 65 APPLICATION Siea aa bulb oe Ert edt dole t CR a NR D rte 67 VOR KING PROSPECT Sissdicel resort val i rger hype caer
38. banks at high speed was available The problem was technically huge a composite video may only stream under the control of some codec and stated that the available standard codec introduce some transmission delay a dedicated one should be developed for the specific application This seemed for a time the greatest difficulty to be overcome and a study was started of the video characteristics But the video generator looked as a perfect black box without any possibility to study and manage the data that is inhibiting the creation of the fast dedicated codec for SEM data streaming The interaction with the Manufacturer on this point continuously oscillated from enthusiasm to frustration A meeting at the main plants in Eindhoven was not able to solve the problem The last comment of the designer of the Quanta was about the known impossibility to have real remote control of that instrument because it was never considered as a requirement since the very first phases of the design The Manufacturer announced a deep radical revision of the video chain in the next Quanta series and promised an alpha release for our project by the end of 2005 and after many solicitations at the end decided to keep silent And still keeps In the same way the surprisingly difficult relationship with the FEI headquarter was responsible for aborting the web application solution as described in chapter 3 because of the impossibility to debug those commands of the sup
39. cope is performed using a modified version of the Virtual Network Computing VNC software developed to remote control the computers through the network In this case the local operator gives the control to the desktop to the remote operator The cons of this solution are the low quality of the image transmitted and the not real time interaction between the remote operation and the instrument A similar solution has been applied into another laboratory Centro Grandi Strumenti University of Modena and Reggio Emilia Italy 4 Here the tool used has been NetMeeting but only some microscope commands are available like magnification and brightness Like the previous example the interactivity between instrument and remote operator and the very bad image quality are the weak spots of this solution To summarize very local networking and the low quality and high delay of the standard video streams along the net are the main reason for achieving at most a demonstrative level but not real remote operative conditions The concept of remote microscope used in this project is different the remote control is not an existing software tool adapted to the microscope but is a new application developed for this instrument and that gives the possibility to use it as if the operator was in front of it 5 The complexity of this project is important and relevant and its development has been long and challenging but the work is justified from the importa
40. cted Fig 5 User interface developed for the client application The following table explains each frame in detail Table 2 The microscope commands included in the client user interface grouped in frames on the base of their function m Electron Column ew Ww BET 45 The frame Electron Column contains the commands to turn on off the filament voltage to set its value and to set the spot size value Spot Size Ei J8 e 83 Focus Focus 25 Bn The frame Focus contains the commands to focus the image using the free working distance 44 of the project Detectors Contrast Li a Brightness ET gt The frame Detectors contains the commands to set the contrast and the brightness values Detectors Name EsemList Detector List Bsed Active Detector Ltd This frame gives the possibility to know which kind of detector is installed and which detector is already active Current position x 0 765625 Set Y 3 2836354 SetY Z 26 373694 Setz Rotation The frame Rotation gives the possibility to Scan 0 OK rotate the image changing the scan direction of Rotation the electron beam degrees DwellTime The frame Dwell Time allows the user to change DwellTime 1 OK the scan speed of the electron beam microsec Magnification The frame Magnification contains the command ScanFiedl rea
41. d and black zones where there are no radiating materials The objects in the image appear as 2 dimensional 2D objects The electrons originated by the interactions above described are then detected by dedicated devices which collect it The detectors for backscattered electrons and secondary electrons are usually either a scintillation detector or a solid state detector In the former case electrons strike a fluorescent screen which emits light which is amplified and converted into an electrical signal by a photomultiplier tube The latter detector works by amplifying the minute signal produced by the incoming electrons in a semiconductor device The energy intensity of the collected electrons will be used from the detectors to modulate the brightness and the contrast of the monitor which will display the specimen images 85 Appendix C Electron interactions in a SEM REFERENCES 1 Ludwig Reimer SEM Physics of Image Formation and Microanalysis Springer Verlag Berlin Heidelberg New York Tokyo 1985 2 http www unl edu CMRAcfem em htm 86 APPENDI X D THE ENVI RONMENTAL SEM The SEM has overcome the limit of the TEM to not see the specimen surface Although it maintains an important heavy limit it is not possible to observe a biological specimen unless it is dried To overcome this limit a new instrument was developed 15 years ago the Environmental Scanning Electron Microscope ESEM D 1 THE ENVIRONMENTAL SE
42. d perform sensing and or acting operations according to defined theories and protocols Chapter 1 The remote control of a scientific instrument Precision in terms of sensitivity and accuracy calls for automation and the more sophisticated is an instrument the more automated are its operations Nowadays automation is nearly a synonym of computer control and computers are the nodes of the communication networks 2 Scientific investigation is made by the scientists whose planning of the experiments and whose reading of the results makes the difference also when the instrumentation is the same Few scientists have all instruments at their disposal Often several different equipments are dispersed throughout many different laboratories maybe in different countries Remote control would make closer the scientific community by effectively promoting collaborative approaches as the often announced distributed laboratories 3 Not all instruments in any laboratory are full time employed For the most expensive machines a wider access would be beneficial 4 Training requires to move the students to the laboratory Remote control would invert the situation 5 Consulting and expertises could be offered by a much wider set of scientists and laboratories all around the world The above mentioned reasons also indicate a specific meaning that will be here adopted for the term remote that is at long distance It is an important point becau
43. d Electrons image or someone else The above explained frames are all related to the use of microscope The other frames located in the user interface contain specific commands to use the client application in conjunction with the server application Table 3 explains them The connection is based on the TCP IP protocol the client searches the server by its IP address The transmission occurs by two separated channels a channel from the client to the server and a channel from the server to the client in order to decrease the probability to loose some data At the moment the connection it is possible just for one client at a time but it is possible to enlarge the authorization for more clients In order to respect the security conditions each client will connect to the server by a particular user name and password and the server has to recognize it before to allow the connection Table 3 The commands concerning the connection between the client application and the server application included in the client user interface Server Connection The Server Connection frame is the frame where there are the commands to connect Close 192 84 153 107 A en the client application to the server application In the text box the client inserts the IP address of the server and then clicks on the button to request the connection When the connection is allowed the button label becomes Close the connection and the other labels i
44. d video acquisition device for the PCI bus 1 It features four independent acquisition channels each comprised of an 8 bit analog to digital converter synchronization circuitry anti aliasing filter optional and input lookup tables The four channels can acquire images from four non genlocked monochrome cameras Each channel also supports separate horizontal and vertical sync signals input or output a trigger input optically isolated for protection from potentially hazardous voltages a trigger output exposure control output and a strobe output The features described permit each channel to acquire images from four completely independent cameras The Viper Quad installed in the Quanta microscope acquires the image taken with the electron detectors and converts it into a standard video format Local video this block represents the video card of the microscope controller computer which receives from the PCI bus the Viper Quad output and overlays it with the window background Frame analysis this block takes the output of the frame grabber and analyses it before to decode it with the H 264 encoder Modified H 264 this block decodes the output of the frame grabber and compresses it on the base of the H 264 international standard Stream the stream block allows the transmission on the network of the coded video Decoder this block is at the end of the network transmission on the remote client side and decompresses the received video s
45. e considering a frame rate value equal to 10 frames per seconds Looking at the table the worst case has a very low signal to noise ratio and the images are not significant and comprehensible while the best case does not allow useful analysis the refresh time is too long These cases are extreme situations in the table there are middle resolution values with a good compression ratio Normally 54 Chapter 4 The video transmission the resolution used is 1024x884 pixels with a frame rate of 10 frames per seconds because it reaches a good compromise between noise reduction and scan speed Table 1 Trade off between video resolution and compression ratio Microscope Frame rate Compression resolution pixel frame per second Ratio 4 3 2 The video architecture design Together with the video architecture design also the technology has been decided The video architecture designed is a video streaming architecture The technology chosen is based on the H 264 AVC compression standard Figure 2 shows a block diagram of it MODIFIED Fig 2 The block diagram of the architecture designed for the video application In order to give a more detailed explanation a description of each block follows 55 Chapter 4 The video transmission Frame grabber this block is the frame grabber card In the used instrument it is a Viper Quad card made by the Dalsa Coreco Society The Viper Quad is a single slot high spee
46. e electrons that pass the 80 Appendix B The hardware of electron microscopy Faraday cage are accelerated into the scintillator When the electrons strike the scintillator they produce light which is in turn directed to the photomultiplier by the light pipe The photomultiplier produces an output signal that is then related to the total number of electrons collected Because the scintillator is typically in direct line of sight with the sample backscattered electrons which usually are too energetic to be deflected much by a 200 V potential will also produce a signal in the detector even if a negative potential is applied to the Faraday cage In this case it is a pure backscattered electron detector By adjusting the cage potential it is possible to tweak the topographic contrast given by the detector Most SEMs will have an E T style detector although not all will allow adjustment of the collection potential Some E T detectors are designed to be mobile so that the geometrical collection efficiency can be adjusted in addition to the collection potential In a SEM of the past generation this electric signal was used to manage the input intensity of a cathode ray tube CRT 81 Appendix B The hardware of electron microscopy REFERENCES 1 Ludwig Reimer SEM Physics of Image Formation and Microanalysis Springer Verlag Berlin Heidelberg New York Tokyo 1985 2 http www matter org uk tem electron gun electron sourc
47. e of the most useful inquiring techniques for many science field applications A 1 THE HISTORY OF THE MICROSCOPY The earliest instrument for seeing very small object was made by the Dutchman Antony van Leeuwenhock 1632 1723 and consisted of a powerful convex lens and adjustable holder for the object being studied specimen With this remarkably simple microscope van Leeuwenhock could magnify objects up to 500x and with it he discovered protozoa spermatozoa and bacteria and was able to classify red blood cells by shape The limiting factor in van Leeuwenhock s microscope was the quality of the convex lens Later the idea was conceived to add another lens to magnify the image produced by the first lens This compound microscope consisting of an objective lens and an eyepiece together with a means focusing a mirror or a source of light and a specimen table for holding and positioning the specimen is the basis of light microscope today Appendix A The origin of electron microscopy A modern light microscope often abbreviated LM has a magnification of about 1000x and enables the eye to resolve objects separated by 0 0002 m In the continuous struggle for better resolution it was found that the resolving power of the microscope was not only limited by the number and quality of lens but also by the wavelength of the light used for illumination It was impossible to resolve points in the object which were closer together than the wavelength o
48. e scan speed are into the Tool bar and are the and the near to the rabbit faster and to the turtle slower To do this there is also another procedure finer because it allows setting the time to perform a line scan and a frame scan Also Clicking the Scan menu in the Menu bar it is possible to change the scan speed The commands to set the pixel resolution are included in the Tool bar too After the setting of these parameters clicking the Snapshot button on the Tool bar it is possible to acquire the image The images are stored in the Support computer this is a secure system specific of the Quanta instrument Another particular aspect of the Quanta interface is that each of 29 Chapter 2 The Scanning Electron Microscope the most used commands can be activated with different modes some faster and some finer more precise 30 Chapter 2 The Scanning Electron Microscope REFERENCES 1 http www feicompany com 2 The standard user manual of the FEI Quanta microscope ES s CHAPTER 3 THE DEVELOPMENT OF THE PROJ ECT The focus of this chapter is to describe all the phases of the remote microscopy project from the starting plan to the ultimate version of the remote application This Will include the dramatic program change that forced to abort the first idea because of technical difficulties unknown even to the Manufacturer 3 1 THE REMOTE ELECTRON MI CROSCOPY The remote electron microscopy is n
49. e scientific world has started to look at it The remote application here developed aims to be included as a tool of each scanning electron microscope Moreover the transmission electron microscopes become more and more completely driven via software and then it will be possible to think to adapt it also for the remote control of the TEM Moreover a sort of partial remote control has been already implemented for it for the only sake of removing the operator from the TEM room and avoiding human perturbation during critical analyses then it will be not so much complex to develop the remote control as that developed during this research An example of the attention to those possibilities is a project carried out by the FEI Laboratory and some European Universities 2 where the remote control can be included and it should be a standard application This project is called Collaborative Environment and it is a solution implemented to overcome geographical distances between collaborating laboratories and to facilitate data sharing The job consists of the development of a new solution for sharing data and instruments The sharing of electron microscopes could be easily included in this Collaborative Environment 69 Chapter 5 Discussion perspectives and conclusions 5 4 FINAL CONCLUSI ONS This research work has demonstrated that remote microscopy is feasible and works Also if it is not yet a standard widespread application the bigger inter
50. ed and streamed with appropriate coding standard 5 The last aspect considered is the application security The scanning electron microscope is a very expensive instrument and although it has several systems of protection against not allowed operations it is necessary to provide also the remote application with a system security This system should monitor the access to the microscope allows the enabled remote operators and blocks the not allowed clients which try to enter to the microscope 3 4 DESI GN The purpose of this phase was to identify the solution after the features description Two solutions able to satisfy all the characteristics defined have been found The first is a server client application consisting of the development of two different applications a server application and a client application able to communicate between them The second solution is a web based application consisting of a particular web site that allows the access to the microscope A detailed study of pros and cons for each solution has been carried out 3 4 1 The Server client approach This approach consists of the implementation of two different applications a server application and a client application able to communicate between them using the TCP IP protocol The server application must be installed into a third computer near the microscope and connected to the microscope controller PC through the Support Network figure 1 The client
51. efines the spatial resolution of the imaging system Depending on the selected emission detector the map will tell something different on the observed specimen Two accidents made the previous things the basement of such a successful machine as a SEM e The interaction volume of the previous point iii resulted much smaller than the smallest detail observable by optical microscopes For one of the emissions in particular the emission of secondary electrons the area of emission was roughly 1000 times smaller this means 1000 times more magnification available than in optical microscopy The electron scanning imager was a Microscope e The emission detection of electrons in a SEM surprisingly resembled reflection detection of light from illuminated solids the electron images appeared not only detailed but also shaded according to the curvature of the surfaces and even shadowed according to geometrical laws for propagation of straight rays the e beam trajectories As a result a SEM image looks natural beyond any expectation different from the great majority of analytical techniques i e clinical echography including those based on imaging even a child may read a SEM picture in terms of geometric evidence To be precise another accident contributes to the natural appearance of the SEM images the electron beams are thin This means that their impact area onto the specimen and then the resolution in the final image is nearly
52. em D 2 1 The Low Vacuum and the ESEM working modes Working at Low Vacuum mode means working with the LFD detector while working at ESEM mode means working with the GSE detector or with the gaseous back scattered GBS detector The vacuum system instead works at the same manner for both cases The bottom of the specimen chamber is closed off by the main valve and a large bore pipe allows the oil diffusion pump that would normally pump on the chamber from below to bypass the chamber and pump off the upper portion of the column instead The top of the specimen chamber cannot be completely closed off because the electron beam must be able to enter it So the pole piece insert a bullet shaped device that contains the final aperture and is located where the electron beam enters the specimen chamber is replaced with one specifically designed for ESEM and Low Vacuum This wet bullet insert has the pressure limiting apertures PLA in it in figure 2 PLA1 and PLA2 The apertures are simple discs with small holes bored through the centre The principle at work here is that if there is a small enough pinhole between two different vacuum levels and the difference between levels of vacuum is that great vacuum will not diffuse from one level to another through the pinhole Then there is a very good 90 Appendix D The Environmental SEM vacuum at the electron column at the top of the column and at the mid portion of the column instead
53. ent a standard for the remote microscope Then the user interface developed 41 of the project should be not related to the instrument used for the experiment but standard and useful at the same time so that an expert microscopist can use it immediately while a not expert user can learn easily the instrument working process For these reasons the most important is the user interface of the client application the server application in fact will be installed onto a PC sited in the same room as the microscope then nobody will use it to control the microscope but just to control the remote connection The only active commands are those related to the connection with the client application table 1 All the other commands are not enabled while into the text boxes included in the user interface the server writes the values received form the microscope controller application Figure 4 sketches this user interface Table 1 The commands related to the connection included in the server user interface When the client requests the connection to the server to allow the connection the server has to push on the Connection button Connect while Connect Disconnect the button Disconnect interrupts the Disconnected ra connection The button Run Microscope can be i used to connect the server application to the microscope without any client application connected to it The label Disconnected reports the c
54. es htm 3 http www matter org uk tem lenses electromagnetic lenses htm 82 APPENDIX C ELECTRON I NTERACTI ONS IN A SEM In this appendix a description of the interaction between the bream electrons and the specimen electrons will be given together with the description of the signal originated from these C 1 SPECIMEN I NTERACTIONS The interactions between the beam electrons and the specimen electrons are the origin of the SEM images Several physical phenomena occur but just five of these are used in an ordinary SEM see Fig 1 1 The specimen itself emits secondary electrons 2 Some of the primary electrons are reflected backscattered electrons 3 Electrons are absorbed by the specimen 4 The specimen emits X rays 5 The specimen sometimes emits photons These phenomena are interrelated and all depend on the topography the atomic number and the chemical state of the specimen The number of secondary and absorbed electrons at each point of the specimen is more influenced from the topography of the specimen rather than the other properties It is for this reason that these three phenomena are exploited primarily to image the specimen surface In fact Appendix C Electron interactions in a SEM depending on the phenomena analysed it is possible to acquire information on different specimen aspects 1 2 Backscattered e Incident e X rays Cathodoluminescence Secondary e Interaction volume Fig
55. est of the scientific community in this argument is manifested more and more by the laboratories where the scientists are carrying out some researches in this direction To highlight this comment it is sufficient to check as this argument has been included as a specific topic called Remote Microscopy in the digital era in the scientific programme of the worldwide most important conference on microscopy The 16 International Microscopy Congress 3 to be held in Sapporo Japan on the next September 2006 It is the place where also this PhD thesis job will be presented 4 70 Chapter 5 Discussion perspectives and conclusions REFERENCES 1 R Pintus S Podda and M Vanzi mage alignment for 3D reconstruction in a SEM Microelectronics Reliability 45 2005 pp 1581 1584 2 A van Balen et a A Collaborative Environment for Users of Advanced Scientific Instruments Third International Conference on Creating Connecting and Collaborating through Computing C5 05 pp 83 90 3 http www imc16 jp 4 F Mighela C Perra M Vanzi Long Distance SEM Remote Control submitted to the 16 International Microscopy Congress Sapporo Japan September 2006 71 APPENDI X A THE ORIGIN OF THE ELECTRON MI CROSCOPY The word is derived from the Greek micros small and skopeo look at Ever since the dawn of science there has been interest in being able to look at smaller and smaller details The electron microscopy is on
56. ether with a bigger project that has regarded the realization of a Microscopy and Nano Analysis Laboratory Several 16 Chapter 2 The Scanning Electron Microscope instruments had been acquired for this project including the scanning electron microscope and it has been decided to use it as the test bed for our remote control studies Figure 2 shows the Quanta instrument In the picture there is the scanning electron microscope and two computers to control it the microscope controller PC which operates directly on the instrument and the support PC The electron column containing the beam source the specimen chamber and the block with the pumps to create the high vacuum into the chamber have been highlighted to be easily identifiable The Manufacturer introduces his product saying that 1 The Quanta ESEM 200 is a versatile high performance low vacuum scanning electron microscope with a tungsten electron source with three imaging modes high vacuum low vacuum and ESEM to accommodate the widest range of samples of any SEM system The key features are e Seamless point and click transition between imaging modes e Superior low vacuum low kV imaging e Simultaneous secondary electron SE and back scattered electron BSE imaging in low vacuum mode e Allows for in situ dynamic experiments e True surface SE imaging in all vacuum modes and voltages e Easy to use four quadrant single quadrant user interface 17 Ch
57. f the light a few hundred nanometres Using light with a shorter wavelength blue or ultraviolet gave a smaller improvement immersing the specimen and the front of the objective lens in a medium with a high refractive index oil gave another small improvement but these measures together only brought the resolving power of the microscope just less than 100 nm In the 1920s wave mechanics indicated as electron propagation in vacuum may be considered quite similar to the scalar diffraction theory of light Furthermore it was found that electric and magnetic fields have the same effect on electrons as a transparent media have on light refraction which enables the design and construction of lenses Dr Ernest Ruska at the University of Berlin combined these characteristics and built the first Transmission Electron Microscope abbreviated TEM in 1931 For this and subsequent work on the subject he is considered the electron microscope inventor and was awarded the Nobel Prize in 1986 The first electron microscope used two magnetic lenses and three years later he added a third lens and demonstrates a resolution of 100 nm twice as good as that of the light microscope Today using more magnetic lenses in the imaging system a resolving power between 0 2 nm and 0 5 nm at magnification of over 1 million times can be achieved 1 Using electron propagation instead of light requires operating in vacuum which makes an electron microscope a more demandin
58. g instrument than a standard optical device Moreover specimen thinning down to submicrometric thickness renders the specimen themselves quite fragile and not all materials may afford such preparation Additionally the transverse information supplied by the TEM nearly completely disregards any information of the specimen surface Early attempts at producing images from the surface of a specimen involved mounting the specimen nearly parallel to the electron beam which then strikes the surface at a very small angle Only a very narrow region of the specimen appears focus in the image and there is considerable distortion The technique has not found wide application in the study of surfaces 13 Appendix A The origin of electron microscopy To overcome this TEM limit it was proposed the principle of scanning the surface of a specimen with a finely focused electron beam to produce electron emission by a bulk specimen and from the collection of that emission an image of the scanned surface The first published description appeared in 1935 in a paper by German physicist Dr Max Knoll Although von Ardenne performed some experiments with what could be called a Scanning Electron Microscope usually abbreviated SEM in 1937 it was not until 1942 that three Americans Dr Zworukin Dr Hillier and Dr Snijder first described a true SEM with a resolving power that should be compared with the 1 mm resolving power of the best optical microscope that is hu
59. geControl Move STAGE AXIS X xdXpos 1000 xdYpos xdZpos xdTpos xdRpos Case SetYPos Dim ydXpos As Double Dim ydYpos As Double Dim ydZpos As Double Dim ydRpos As Double Dim ydTpos As Double ydXpos txtXpos Text ydYpos arr 1 103 Appendix F The implemented code ydZpos txtZpos Text ydRpos txtRpos Text ydTpos txtTpos Text objStageControl Move STAGE AXIS Y ydXpos ydYpos 1000 ydZpos ydTpos ydRpos Case SetZPos Dim zdXpos As Double Dim zdYpos As Double Dim zdZpos As Double Dim zdRpos As Double Dim zdTpos As Double zdXpos txtXpos Text zdYpos txtYpos Text zdZpos arr 1 zdRpos txtRpos Text zdTpos txtTpos Text objStageControl Move STAGE AXIS Z zdXpos zdYpos zdZpos 1000 zdTpos zdRpos Case SetRPos Dim rdXpos As Double Dim rdYpos As Double Dim rdZpos As Double Dim rdRpos As Double Dim rdTpos As Double rdXpos txtXpos Text rdYpos txtYpos Text rdZpos txtZpos Text rdRpos arr 1 rdTpos txtTpos Text objStageControl Move STAGE AXIS R rdXpos rdYpos rdZpos 1000 rdTpos rdRpos 104
60. he debugging phase The testing phase has been carried out after each modification of the software In fact the functionality of different microscope functions was already tested at the beginning of the research work in this phase it was necessary to debug the Visual Basic code Different errors have been found into the code and have been removed following the debug indications The testing phase has been very useful to realize a good user interface and especially to establish a correct connection between server and client 47 of the project 3 7 WORKI NG WI TH A REMOTE SCANNI NG ELECTRON MI CROSCOPE In order to give a comparison between to use the microscope locally and remotely a sequence of the actions performed to carry out a remote analysis will follow When the remote operator starts to work with the remote microscope the first operations that he has to perform are related to the connection from the client application to the server application He requires the connection to the server and if the server allows it he connects the client application to the microscope and can start to use it The starting operation of a microscope analysis is to introduce the specimen into the chamber If the chamber is on the vacuum state it is necessary to stop the vacuum pumps and to ventilate it These operations are not performable on remote then the local operator always present in front of the microscope has to do it Obviously he
61. his research the application of the remote control to the scanning electron microscopy SEM will be described This project is born from the consideration that the scanning electron microscopy is a widely diffused inquiring technique but at the same time expensive and not affordable for every research laboratory Introduction The description of the study presented in this thesis is divided into six chapters In the first chapter a short preface to the remote control for the scientific instrument is given to introduce the research work presented in this thesis The second chapter contains the description of the scanning electron microscope used as test bed of the project It is analyzed on the side both hardware and software aspect The development of the remote application has been divided into two main parts the software to achieve to remote control and the video transmission application The first part is described in the third chapter which contains a detailed description of every phase of it It starts form the feasibility study and continues with the software features definition the possible solutions the design of the chosen solution and the software application implementation The fourth chapter is dedicated to the second part the development of the video transmission architecture This part has been classified as the most delicate part on the base of the features required The fifth chapter reports the research result the work
62. hitecture Even if the encoder is not dedicated and the used applications are commercial some characteristics of the video system indicated for the original solution have to been reached also in this situation Then the real time transmission and the video quality are fundamental parameters for the application In order to find the best solution several commercial encoder and video player have been tested following a particular procedure 4 4 1 The video application tests The first results obtained using the commercial applications have shown that the video quality available is not so bad in relation of the video quality of the original video a PAL standard video On the side of the delay transmission the results are definitely worst In fact the minimum delay registered is about 2 seconds the worst value is about 10 seconds This alteration depends on the network traffic if there is a lot of traffic obviously the bandwidth available is reduced and then the delay increases These values depend also on the player use on the side of the client to 60 Chapter 4 The video transmission reach the minimum delay value more than one video player has been tested The results of these experiments are showed in table 2 Table 2 Performances comparison of different video player tested WM Player VLC Resolution Bit rate buffer size Medium Medium Mbps sec Delay sec Delay sec 320x200 First of all the encoder has
63. ial Space Charge A Electron Beam EE node Plate posikive potential Fig 1 The Electron GUN The thermionic filament is made from high melting point material of relatively low working function in order to emit many electrons Common choices are Tungsten 24 5 eV Lanthanum hexaboride 3 0 eV The Wehnelt Cylinder is a biased grid which is a few hundred volts different in potential from the filament cathode This causes the electron paths to cross over from the anode hole The diameter of the beam at this cross over is the effective source illumination which can be only a few tens of micrometers The anode is a positively metal plate i e at earth potential which has a hole in it Its function is to accelerate the electron beam to the high voltage potential A thermionic electron gun is operated as follows e Select the accelerating potential kV e Increase the current passing through the filament until the knee of the emission curve is reached saturation giving the best compromise between the beam current emitted as high as possible and the filament lifetime as long as possible e Adjust the bias to give the desired combination of source size and beam current The filament current controls the temperature of the filament and then the number of electrons emitted or beam current Generally it is desired a large number of 1 The working function is the energy in electron volts needed by an electron
64. ial for the analysis In other words the microscopist needs full resolution full rate not delayed images simply to do his job Here is the bottleneck of the long distance Remote Control full resolution and full rate mean real and guaranteed wide band over a long distance connection not delayed calls for extremely efficient and fast coding or possibly unencoded video streaming Chapter 1 The remote control of a scientific instrument Compared to the video transmission the remaining data to exchange is minimal and it refers to the remote commands that read some measurement and set some parameter Anyway the different automation of the various kinds of microscopes leads to the funny situation that the simplest instrument the optical microscope is also the less suitable to be remotely driven It is common indeed to have manual mechanical controls for the specimen movements for the choice of the objective among a set in a carrousel for the focus for the inclination or filtering of the light etc It is of course possible to look for completely motor driven microscopes but this brings the instrument outside the widespread standard makes it a rare and much more expensive device and leaves the doubt that of the many prompts to remote control the only one serious for the optical microscope is the possibility to have a consulting expert able to see the live images which may be achieved by simply connecting the microscope camera to
65. ideo output of the frame grabber figure 1 EE File Magnification Beam Scan Detector Stage Tools Window Help Pages Tm AR ga a ees Ot 13 e E 3 m Vacuum Mode 4 High Vacuum C Low Vacuum water m C ESEM Water Pressure oly 0 60 mbar p Electron Column 3 ot Size E o UST 38 High Voltage ef 200v p Detectors Contrast Brightness m n ees Pressure EN ERR l 20 0 kV kil 2 mm sty SSD m Source Control F wWehnelt Changed Limit Voltage THIR RD QUAD E f A d i 4 pm Dom COMPOSITIONOF Fou A x P Source Tilt kaid NTER DI GTA EE Reset Timer Filament 881 Hours rSttu Vacuum Status Vacuum High Voltage 20 0 kV Pressure 31e8 mbar Filament Current 224 A HV me WD ae Pressure 7 22 2005 ee s 111938AM SMCI Emission Cure aMNTO HA Fig 1 The visualization of the four video sources elaborated by the frame grabber 53 Chapter 4 The video transmission 4 3 THE VIDEO ARCHI TECTURE DEVELOPMENT Starting from this architecture the planning of the architecture to perform the video streaming is started This phase has divided into more parts first the study of the system properties then the planning of the architecture solution and last the choice of the technology 4 3 1 The system properties First of all the re
66. in the specimen chamber there is a relatively poor vacuum without endangering the electron gun At the bottom of the column the oil diffusion pump is bypassing the mid portion and contributing to the better vacuum in the upper column this arrangement also helps scavenge any water vapour that rises through the pressure limiting apertures The hole in the centre of the gaseous secondary electron detector GSED functions as the final aperture through which the primary electron beam passes and its bore size determines how poor the vacuum can be in the specimen chamber The GSED in most cases forms a fitted seal over the pole piece insert the wet bullet If the GSED has a 500 um aperture in it the pressure in the chamber can be increased to as high as 10 Torr if the GSED has a 1mm aperture in it the pressure the chamber can be only as high as 5 Torr Using the LFD detector it is not fitted over the pole piece insert then the wet bullet itself provides the final aperture and the chamber pressure can be only as high as 1 Torr 91 Appendix D The Environmental SEM REFERENCES 1 http www itg uiuc edu ms equipment microscopes esem how it works htm 2 The standard user manual of the FEI Quanta microscope 92 APPENDIX E THE TRANSMI SSI ON ELECTRON MI CROSCOPE This section is dedicated to explain the Transmission Electron Microscope which Is its working principle and how it works E 1 THE TRANSMI SSI ON ELECTRON MI CRO
67. ing electrons cause the specimen to emit X rays whose energy and wavelength are related to the specimen elemental composition 7 The impinging electrons cause the specimen to emit photons or light this is called cathodoluminescence 8 Finally electrons which have lost amount of energy because of interaction with the sample can be detected by Energy Loss Spectrometer which is the equivalent of a prism in light optics Incident e Interaction volume Inelastically Scattered e Unscattered e Elastically Scattered e Fig 3 The TEM specimen interaction classification In a standard TEM the first two phenomena contribute to the formation of the normal TEM image for non crystalline biological specimens while for crystalline specimens most non biological materials phase contrast and diffraction contrast are the most important factors in image formation It is necessary to add accessories or peripheral equipment to the basic microscope in order to exploit the additional information which can be obtained by studying the last five interactions listed above The electrons that pass through the specimen may have different interaction with it in particular there are Unscattered Electrons They are the incident electrons transmitted through the thin specimen without any interaction occurring inside the specimen The transmission of unscattered 96 Appendix E The Transmission Electron Microscope electrons is inversely propo
68. ing perspectives and the final conclusions The thesis continues with six appendixes where some arguments related to the main research theme are described in order to better complete the manuscript content CHAPTER 1 THE REMOTE CONTROL OF A SCIENTIFIC NSTRUMENT The first remote instruments have been used during the Wold War I Starting from this moment an increasing number of instruments has become remote controlled Despite the many announcements real Remote Control has not been achieved for a vast class of instruments 1 1 WHY REMOTE CONTROLLING SCIENTIFIC NSTRUMENTS The intuitive significance of remote control is the technical possibility to have a device whose operations are driven by an operator far from it Although not strictly necessary remote control is usually the ultimate companion of automation you can remotely play an electronic keyboard but not a classic piano and telecommunication you can drive an electric vehicle looking at the images of an on board camera in the next room or in the next town but not on Mars because of the delay in communications that inhibits the real time feedback of the operator Not all devices are suitable for remote control and not all operations of a device are possible or even advisable to be controlled from elsewhere Nevertheless thinking in remote control of scientific instruments makes sense under a number of aspects l Scientific Instruments are suitable for it They indee
69. ing window On the side of the video players tested some of those can be integrated in the remote client application some others no The results showed in the table concern only two video players which summarize all the experiments carried out These players are Windows Media Player WMP that can be integrated and VideoLAN Client VLC which can not be integrated in the remote client application As reported in the table the minimum value of the delay has been obtained using the VLC player and is equal to 2 seconds 62 Chapter 4 The video transmission In these experimental tests it should be pointed out the mismatch between the video input resolution PAL and the streaming resolution This mismatch requires that the encoder has also to process a re sampling for each frame In theory the video streaming should be realized at the source resolution or integer scaled resolutions in order to minimize the complexity and hence the computational load for the re sampling operation This delay value is a very good result if it is compared with the commercial product available for the streaming In fact from state of the art of the video streaming transmission it is possible to reach as the minimum delay about 0 5 seconds and difference of 1 5 seconds could be considered not significant Even if this result is globally appreciable a delay of 2 seconds makes longer and heavier working at the microscope increasing the length to improve the
70. inocular viewer As in other branch of science a permanent record of what has been observed with the eye is desired Antony von Leeuwenhoek see Appendix A painstakingly enabled microscopists to photograph what they saw Electrons have the same influence on photographic material as light Therefore it is only necessary to replace the fluorescent screen with a photographic film in order to record the image Today the image acquisition is allowed by a digital system in fact all new TEM have a digital monitor on which see the specimen then with particular software made by the microscope Manufacturer it is also possible to take a digital photograph of the specimen observed 98 Appendix E The Transmission Electron Microscope REFERENCES 1 David B Williams and C Barry Carter Transmission Electron Microscopy Basics Plenum Press New York and London 1996 99 APPENDIX F THE MPLEMENTED CODE In this section an example of the code implemented for achieving the remote application will be given F 1 AN EXAMPLE THE STAGE FUNCTIONS As example of the developed software the code that achieves the stage movement is here reported To move the stage the remote operator before ask the actual stage position to the server here the code that executes this instruction Private Sub IbIX Click TCP SendData 31 amp Separatore End Sub Private Sub IbIY Click TCP SendData 33 amp Separatore End Sub Private
71. ked also integrated into the web site To resolve this problem different implementation solutions have been tried without achieve relevant results The reason of those problems is that not all the functions of the XTLibInterface are compatible with VBScript and then it was normal that some of these gave error when implemented The root cause was clearly inside of the Manufacturer s proprietary code and then the Manufacturer was asked for assistance They suggested to use only Visual Basic programming language because it was the first edition of the XTLibInterface and they tested successfully only the Visual Basic language The other languages like C or also VBScript were not tested and then not reliable At this time the entire web based project was aborted and the research work was focused on the other solution the server client application 3 6 THEIMPLEMENTATI ON OF THE SERVER CLIENT APPLICATION The server client applications are particular kind of applications which can be implemented following two predefined architectures a two level architecture and a three level architecture The number of levels depends on the number of the breaks 40 of the project between the data origin and the client application In this case there are three levels because the server application to satisfy the client requests has to forward these requests to the microscope Then the server application is a client for the microscope as sho
72. kscattered electrons detector Downside there are on the left a combination of the first two images and on the right a lateral view of the inside specimen chamber taken by an internal camera used to check the mechanical movements of the specimen in order to avoid collisions and damages The distribution of the images along the four quadrants may be set in many different ways enabling other detectors when available or switching to a single wide frame It is interesting to observe as really different emissions give different responses the upper images are in the screen live pictures simultaneously updated as the beam scans the specimen surface and refer to exactly the same area On the right side of the image area of the user interface there are the commands to control the microscope parameters The commands are organized in five pages and grouped on the base of the managed parameters even if the most important commands are present into each page MEE Chapter 2 The Scanning Electron Microscope pages v HV Spot wD Det Pressure 200 v 49 307men a ssp Reset Timer Filament 6814 Hours Status Vacuum Status Vacuum i High Voltage mo Ww Filament Current 224 A Emission Curent 101 Fig 4 The Quanta user interface The pages are named Startup page Work page Option page Adjustment page Temperature page The other main parts of the user interface are the Menu bar and the Tool bar
73. lephony video conferencing TV storage DVD and or hard disk based especially high definition DVD streaming video and many others H 264 AVC was developed over a period about four years The root of this standard lie in the ITU T s H 26L project initiated by the VCEG which issue a Call for Proposals CfP in early 1998 and created a first draft design for its new standard in August of 1999 In 2001 when ISO IEC s MPEG had finished development of its most recent video coding standard known as MPEG 4 Part 2 it issued a similar CfP to invite new contributions to further improve the coding efficiency beyond what was achieved on that project VCEG chose to provide its draft design in response to MPEG s CfP and proposed joining forces to complete the work Several other proposals were also submitted and were tested by MPEG as well Therefore to allow speed progress ITU T and ISO IEC agreed to join forces together to jointly develop the next generation of video coding standard In order to satisfy the wide areas of applications considered by the two organizations three basic features sets called profiles were established to address these application areas the Baseline Main and Extended profiles The Baseline profile was designed to minimize complexity and provide high robustness and flexibility for use over a broad range of network environments and conditions the Main profile was designed with an emphasis on compression coding efficiency capabili
74. mage primary electron beam than both tungsten and lanthanum hexaboride sources permit non destructive imaging of fragile specimens and its accelerating voltage may be lowered significantly Figure 1 shows a diagram of an ESEM column Here it is possible to see the pressure valves that isolate the specimen chamber from the other parts of the microscope at high pressure The water vapour is introduced to the specimen chamber via a separate dedicated vacuum pump that can control the chamber pressure with great accuracy Gun Chamber 10 Torr 1055 Torr Auxiliary gas Water vapour Fig 2 The vacuum system of an Environmental SEM 88 Appendix D The Environmental SEM The primary electron beam is very energetic and it penetrates the water vapour with little apparent scatter scanning across the surface of the sample Secondary electrons are released from the surface of the sample as they are in a normal SEM but they encounter water vapour molecules once they exit the surface The water vapour molecules when they are struck by the secondary electrons produce secondary electrons themselves which in turn produce secondary electrons from adjacent water vapour molecules Thus the water vapour functions as a cascade amplifier amplifying the original secondary electron signal from the sample The amplified secondary electron signal is collected at the Gaseous Secondary Electron Detector GSED with its strong and positive charge So a
75. n but its user interface includes only few commands not sufficient to obtain exhaustive analyses and high quality images The second example University of Michigan has a complete different approach because it does not implement any user interface but it sends the user interface proper of the instrument to the remote operator This is different form the main goal of the reported research work the standardization of the remote microscopy Sending the proper interface of the instrument the remote user will be locked up of learning the use of that particular microscope The last example reported University of Modena differs into the approach also in this case it uses an already existing tool NetMeeting and no much commands are available for the remote control Moreover the image quality is very bad and the real time interactivity is not reached 49 of the project REFERENCES 1 http www itg uiuc edu technology remote microscopy 2 http www itg uiuc edu 3 http emalwww engin umich edu people jfmjfm jfmjfm html 4 http www cigs unimore it 5 F Mighela C Perra Remote Control for Microscopy Applications submitted to Instrumentation and Measurement technology Conference Sorrento Italy April 2006 50 CHAPTER 4 THE VI DEO TRANSMI SSI ON Vision is the base of microscopy In this chapter an insight on the creation and handling of the images in a SEM Will be given and then how these image
76. n by user name and password and then the loading of the page with microscope commands Figure 2 shows this page 38 of the project lt q Documento senza titolo Microsoft Internet Explorer 0 x File Modifica Visualizza Preferiti Strumenti H Indietro gt A Acerca XjPreferii Cronologia E Indirizzo lei http localhost Sito Quanta homepage mio asp vai Collegamenti User ID Password Enter Microscope E Operazione completata eu Intranet locale a Fig 2 The home page of web based application After some tests carried out in order to find the best implementation of the commands for managing these independently it was decided to use the HTML frameset The microscope commands have been grouped on the base of their function and a frame for each group has been projected This work has been carried out developing the frames separately and testing it The implemented frames are Status here it is possible to read the pressure value the specimen chamber state and the form functioning of the microscope The test is positive High Voltage this frame gives the high voltage control here it should be possible to read the voltage value to turn on off the beam voltage and to set the value The result of testing this frame is that it is possible to read the value but not to set it or turn on off the voltage Magnification in this frame
77. n the local and the remote operation which emphasizes another relevant point of such a project the need to release a product the Remote Scanning Electron Microscope that is not for a couple of specialists but is friendly for any user of the standard microscopy community which means a careful design of a Remote User Interface able to operate and also easy to be operated In the family of the Electron Microscopes another instrument seems rather close to achieve remote control the Transmission Electron Microscope TEM All the same the research on this microscope has developed a particular type of remote application in order to remove the operator from the instrument This is not exactly the remote control as said before but makes a long walk in the direction of complete long distance applications The research on the TEM is working indeed to realize the complete software control of the instrument then this should be the next remote control project Chapter 1 The remote control of a scientific instrument REFERENCES 1 The standard user manual of the FEI Quanta microscope CHAPTER 2 THE SCANNI NG ELECTRON MI CROSCOPE In this chapter a detailed description of the Scanning Electron Microscope and its functions will be given After the SEM used during the project will be presented The study of its characteristics and features has been a useful phase to recognize and select which functions include in the remote application Mo
78. nce of the instrument into each laboratory both those of the Universities and those of the private societies The application will work at the same mode for all possible remote users the difference will be in the operations allowed to them An expert user will be enabled to perform a complete SEM session a student will have only some commands allowed like brightness contrast magnification but he will not manage for example the high tension or the pressure chamber a customer not microscopist will be just a spectator This project is started from the feasibility study because it was necessary to develop a software application After this the work has continued determining the characteristics and the features that the application should satisfy and after the 33 of the project possible solutions have been identified The last step has been the choice of the kind of solution and its implementation 3 2 THE FEASIBILI TY STUDY As illustrated in the previous chapter the Quanta ESEM 200 has a lot of technical features that make it an instrument unique and distinctive Anyway for this research work the fundamental characteristic is a peculiarity of its architecture the internal LAN that connects the Support PC and the Microscope Controller PC Accessing that LAN will open the door to the remote control This has been guaranteed from the Manufacturer which supplied a Dynamic Link Library DLL developed exactly to interface an exte
79. ndicate the channels transmission state Dati inviati con successo Connected 46 of the project The Connection frame is the frame to connect the client application to the Connection microscope Before to establish the D7769 Connect Disconnect connection all the buttons are red as first the client clicks on the button Connect Microscope the series number of the microscope will sent from the server and the Button Run Microscope is enabled after the remote operator clicks this button and the microscope commands of the client user interface will be enabled These two buttons become then green while the button to disconnect the client from the microscope is enabled and becomes yellow Chat The Chat frame gives the possibility to the remote and the local operator to communicate using the chat The client writes its messages into the white text box and then clicks Send to send the massages to the server In the upper text box will be displayed all the messages written by the client and the server Besides the frames in the user interface there are the menu bar this is similar to that of the original Quanta software the reason is that the functions present are typical functions for every scanning electron microscope Then the reproduction of this menu bar does not represent a loss of standard characteristics for the remote application here developed 3 6 1 T
80. ndred times better 74 Appendix A The origin of electron microscopy REFERENCES 1 Simon amp Schuster Macmillan Encyclopedia of Physics New York pp 452 454 1996 75 APPENDIX B THE HARDWARE OF ELECTRON MI CROSCOPY The electron microscopes have some common hardware components as the electron gun and the electromagnetic lenses and other dedicated components as the Everhart Thornley detector B 1 THE ELECTRON GUN The electron gun represents the source of an electron microscope and is the equivalent of the light for the optical microscopes 1 The function of the electron gun is to provide an intense beam of high energy electrons 2 There are two main types of gun Figure 1 shows the thermionic electron gun which is the most commonly used It is robust relatively cheap and does not require an ultra high vacuum In the thermionic electron gun electrons are emitted from a heated filament about 2700 C and then accelerated towards an anode A divergent beam of electrons emerges from the anode hole The field emission gun may also be viewed In the field emission gun a very strong electric field 10 Vm is used to extract electron from a metal filament Temperatures are lower than that needed for thermionic emission This gives much higher source brightness than in thermionic guns but requires a very good vacuum Appendix B The hardware of electron microscopy Filament Wehnelt Cap negative potent
81. ng speed of the beam must be lowered or the area repeatedly scanned in order to reduce noise by time integration The operations then continuously change between a live visual navigation at medium low magnification and several inspections of single details where magnification is followed by beam adjustment speed reduction and recalibration of brightness and contrast The practical resolution however depends also on the properties of the specimen on the specimen preparation technique and on many instrumental parameters such as beam intensity accelerating voltage proper correction of possible beam distortion in 15 Chapter 2 The Scanning Electron Microscope particular astigmatism distance from the last lens to the specimen usually referred to the working distance and the angle of the specimen surface with respect to the detector Then managing these parameters and under optimum conditions a resolution better than 5 nm can currently be attained The obtained images may also been stored A modern SEM is usually equipped with a digital acquisition and storage system The images are taken as if there is a digital photo camera and stored as an ordinary image file Longer exposure times may then be allowed than for visual inspection in order to have full resolution pictures at suitable resolution and acceptable noise 2 4 THE EXPERI MENTAL PLATFORM Any up to date SEM is currently completely computer controlled and the choice of
82. olts they are not high as in a TEM see Appendix E because it is not necessary to penetrate a thin specimen but on the contrary a small interaction volume should be obtained in the specimen which would indicate the lowest possible energy for the electron beam It was a series of technical limitations in electron optics that at the begging of SEM prevented from using low acceleration voltages it was more difficult to obtain good thin undistorted beams at low than at high energy as well as the rise of many electrostatic interaction effect charge accumulation and coulombian perturbation that disappeared for many materials for beam energies higher than 10 keV Currently many of those problems have been overcome but the SEMs still keep the 20 30 keV upper limit mainly because of its suitability to excite other important ancillary emissions x rays described in Appendix C while extend their operation range for imaging to lower and lower energies 2 3 WORKI NG WI TH A SCANNI NG ELECTRON MI CROSCOPE When a microscopist starts an analysis with a SEM he has to perform a series of fixed operations First of all he has to introduce the specimen into the chamber that obviously has to be on air state if not it is necessary to select the command to stop the pumps and to introduce air filtered or nitrogen into the chamber This operation that first insulates the whole column by shutting some valves is usually completely automated A
83. onnection state Chat This is the chat frame the server writes the message into the box below and pushes Send to send them to the client In the upper zi test box al sent A messages are displayed 42 of the project TE uield File Magnification Beam Scan Detector Stage Tools Window Help Client Connection Focus Listen Free Working ien Distance Server State Server State Detectors r Electron Column Contrast el Run HV ky Brightness Spot Size Rotation Scan Rotation r Vacuum degrees o OK DwellTime r Stigmator ea DK L Y e ra r Detectors Stage Current position Stage Type x v pi mm Stop Stage R GoTo T S03 kv S Pascal 1 x Connection Disconnected Fig 4 The server application user interface The other characteristics considered during the development of the user interface have been the functionality the layout and the connection security In fact it is important for example to keep the interrelated commands close each other as the high voltage turn on off and the high voltage setting At the same time it is important to have commands simple to use and giving a real time reaction as example a scroll bar to change the contrast and brightness value instead of a text box where put
84. or axially annular detectors or more usually off axis The position of the detector is important for the natural look of the images It may be demonstrated indeed a reciprocity theorem that states as a SEM image looks like a normal optical photograph where the optical axis of the camera is the column axis the camera looks at the specimen from the point of view of the electron source and the light source has the position and the extension of the entrance of the detector The column the specimen and the detectors operate at high vacuum which calls for a complex vacuum pumping system able to preserve the hot region the electron source and the lenses and coils at high vacuum also during the specimen insertion removal when the whole specimen chamber is ventilated at atmospheric pressure Outside the column chamber the command console includes all the electric nowadays electronic controls for all SEM functions and the observation screen The first SEMs mounted analogical screens CRTs whose deflection coils were driven by the same waveform that drove the electron beam inside the column and whose beam intensity the brightness onto the phosphor screen was modulated by the output signal of the electron detector 12 Chapter 2 The Scanning Electron Microscope Current SEMs use the monitor of a driving computer and the output of the detectors is sent to continuously update some part of memory banks of the video board The
85. ored using a Gamma function The Reset button Digital Brightness 0 00 brings the Gamma function back to default b OL Gamma 1 0 Qe 2 5 4 The Adjustment page The fourth page is dedicated to the beam alignment Table 4 shows these commands this page contains also the Status frame present in the Start up page and shown in table 1 24 Chapter 2 The Scanning Electron Microscope Table 4 A detailed description of the Adjustment page Alignments 1 Tetrode Alignment 2 Gun Alignment 3 Condenser Alignment 4 Final Lens Alignment 5 Stigmator Alignment 6 Stage Rotation Centre 7 PLA Centering These commands are used to align the column and determine fine tuning for the electromagnetic system As shown various adjustments are available 2 5 5 The Temperature page The last page has the temperature control These commands are used with a particular stage that allows carrying out experiments at high temperature directly into the specimen chamber to observe live the effect of temperature on the specimen In table 5 these commands are shown The other commands present in this page are present in the Startup page too and are the Vacuum frame the ectro Column frame the Detectors frame and the Status frame 25 Chapter 2 The Scanning Electron Microscope Table 5 A detailed description of the Temperature page
86. ot a recent idea and along the years numerous solutions have been attempted to realize it The relevant difference between them is the concept of remote microscopy that they represent In fact the remote microscopy could be achieved in more than one way depending on the purpose If the function is to satisfy some training requests the remote microscope is just a demonstrative application as proposed in the Virtual SEM Project 1 by the Imaging Technology Group 2 Beckman Institute University of Illinois USA The remote application developed by this laboratory is a web based application and the test bed instrument is a Philips XL 30 ESEM FEG The remote application is developed for educational reasons and the available operations for the remote control are contrast and brightness setting image focusing magnification regulation status bar and more an interactive chat system between the remote operator and the local operator The limits of this application are the restricted access to the instrument justified maybe by of the project the educational goal but that inhibits the possible applications of it by several different users Another example developed for educational reasons is located at the University of Michigan John F Mansfield North Campus Electron Microbeam Analysis Lab University of Michigan USA 3 In this case to achieve the remote control an already existing tool has been modified The remote access to the micros
87. plied libraries that proved wrong or ineffective in a web environment Looking at that problematic support by the supplying Company it seems an even greater success to have experimented real remote control on the QUANTA It remains as a stimulating perspective the plan to complete the original programme maybe on a different platform to achieve the ultimate goal of a practically non delayed full resolution video transmission 66 Chapter 5 Discussion perspectives and conclusions 5 2 APPLICATIONS The results obtained show not only that the remote scanning electron microscopy is possible but that is useful and immediately practical The first application fields are e The video conferencing analyses e The educational field e The remote development and testing of new applications The v deo conferencing analyses The other immediate opportunity given form the remote microscopy includes not only to carry out the analyses remotely but also to ask the advice of an expert microscopist or an expert on the category of the analysed specimens This last possibility will make quicker performing the analyses because it is easier to send a specimen than organizing the travel for a person and on the other side it will improve quality of the analyses having an expert at disposal for any doubts The educational field In the courses of microscopy training the more effective teaching is to train using the instrument Normally i e at ou
88. pot size obtainable in the rest of the condenser system 79 Appendix B The hardware of electron microscopy The second condenser lens affects e The convergence of the beam at the specimen e Diameter of the illuminated area of the specimen The state of focus makes a big difference to the beam convergence at any point on the specimen The convergence is defined as the angular range from which the incident electrons on a point appear to come The condenser aperture controls the fraction of the beam which is allowed to hit the specimen It therefore helps to control the intensity of illumination and in the SEM the depth of field B 3 THE EVERHART THORNLEY DETECTOR The Everhart Thornley detector E T for short is the common used device to detect the secondary electrons in a SEM It is sketched in figure 4 Collector Gnd amd screen 200v Fhaternultipiier Pireetocathade Laght pipe oe 0 kV Fig 4 The secondary electrons detector A typical E T detector consists of a Faraday cage the collector grid in front of a scintillator in turn coupled to a light pipe leading to a photomultiplier tube The Faraday cage is typically kept at a positive potential on the order of a few hundred volts so as to efficiently collect most of the secondary electrons emitted from the sample The scintillator typically has a thin coating of some conductor sufficient to maintain a positive voltage of several kilovolts so that th
89. pproach is the same of the first solution to elaborate a video and to send it to the client application The problem is to get the original video For this solution a particular device has been used To obtain the video the output of the VGA card of the microscope controller PC has been taken with a particular device This instrument is a televiewer it takes a digital video signal and translates it into an analog video signal The video input is the output of the VGA card of the microscope controller PC this video enters into the device which transforms it into a Phase Alternation by Line PAL standard video This video is then acquired by a particular video card installed into the server computer and that gives available the video for the server application This video card is an Osprey 100 card The PAL standard is a colour television standard the images are reproduced by 625 lines and with a frequency of 50 Hz 59 Chapter 4 The video transmission On the server computer the video encoder takes the Osprey card output and encodes it for the client application This encoder is a commercial encoder not a dedicated encoder The reason is that the video PAL is a low quality video and the increase of efficiency and quality achievable with an ad hoc encoder do not justify the work to implement it moreover this is a provisional solution Figure 4 shows the architecture of this temporary solution CLI ENT Fig 4 The stopgap arc
90. quirements for the video transmission have to be identified this should be a real time and full resolution transmission The real time is justified for the necessity of a microscopist to see immediately the alteration made on the image like the variation of the contrast or the variation of the magnification while the full resolution is necessary considering the size of the objects analyzed The factor that has the main influence on the availability of these properties is the relation between the bandwidth available and the bandwidth required for the transmission The currently available bandwidth has been estimated with several tests at the SEM laboratory the result was that the network average bandwidth available for the transmission is about 10 Mbps This encouraged to proceed because the typical byte rate in a SEM that is simply the pixel dwell time is 1 MBps A more detailed analysis showed that the video microscope size is not always the same because it is possible to choose the resolution and the pixel dwell time that ultimately define the frame rate The worst case considering a dwell time of 0 1us the faster available is with a resolution of 512x442 pixels while the best case is a resolution of 3584x3094 pixels This means that the video microscope has to be manipulated to be sent on the network in particular it has to be compressed Table 1 shows the compression ratio calculated for every resolution value available at the microscop
91. r University these courses are attended by more then five people this makes impossible to give the chance to each student to use the instrument Generally one of them uses the instrument and the others look at the demonstrations which is an obvious limitation for a good and successful training Using the remote application for the SEM this limit is overcome The remote microscope could simplify these courses giving the possibility to show the use of instrument by a maxi screen Moreover the remote microscopy allows each student to use the microscope and to practise in performing analyses with a SEM without overcrowding the laboratory rooms The experimentation of particular applications A Scanning Electron Microscope performs a series of specific operations and analyses as it is a SEM and other operations and analyses defined by the Manufacturer This does not avoid the possibility to study and carry out research about new SEM applications that now may be developed and tested on a specific instrument from remote places This is the situation in our same laboratory where this study has been carried out 67 Chapter 5 Discussion perspectives and conclusions Another research line for instance consists of an application that reconstructs a three dimensional vision of the specimen surface starting from four standard 2D images 1 The characteristic of these images used to make the reconstruction is the position of the black zones or
92. reover this analysis has contributed to evaluate the real research difficulties 2 1 THE OPERATING PRINCIPLES OF A SCANNING ELECTRON MICROSCOPE The operation of a SEM stands on two pillars 1 The capability of accelerated electrons launched from outside onto any material body to cause a variety of emissions of electrons and photons from the surface of the body itself 2 The scanning method that enables a small probe to sequentially explore a wide domain to produce a complete image of that domain The first point is then emphasized by the experimental findings that i it is possible to produce very narrow electron beams ii it is easy to electrically control their parameters energy of the electrons divergence of the beam axial symmetry lateral deflection etc iii all emissions are to some extent quite closely limited to a narrow volume around the impact of the electron beam and iv each emission is Chapter 2 The Scanning Electron Microscope sensitive to both the composition and the topography of the specimen Finally v each emission has a suitable detector able to detect it and its specimen dependent variations The second point takes advantage of the above findings and allowing the thin beam to raster scan the surface of a specimen transforms point measurements into maps The extension of the interaction volume indicates how far the beam should be moved to have not interfering signal from two adjacent points that is d
93. rgy electron from the same atom can fall to a lower energy filling the vacancy This creates an energy surplus in the atom which can be corrected by emitting an outer lower energy electron an Auger Electron Auger Electrons have a characteristic energy unique to each element from which it was emitted from These electrons are collected and sorted according to energy to give compositional information about the specimen Since the Auger Electrons have relatively low energy they are only emitted from the bulk specimen from a depth less than 3 nm X rays The X rays are caused by the de energization of the specimen atom after a secondary electron is produced Since a lower usually K shell electron was emitted from the atom during the secondary electron process an inner lower energy shell now has a vacancy As the electron falls it emits energy usually X rays to balance the total energy of the atom so it X rays or light emitted from the atom will have a characteristic energy which is unique to the element from which it originated These signals are collected and sorted according to energy to yield micrometer diameter of bulk specimens limiting the point to point comparisons available Cathodoluminescence The cathodoluminescence CL technique uses the phenomenon of emission of electromagnetic radiation from a material exposed to an electrical bombing In the image obtained there are a number of white zones where the radiating material is situate
94. rnal application with the microscope software This Library consists of a series of Component Object Model COM Together with this the Manufacturer has given another application a Distributed Component Object Model DCOM client This object is an extension of the Component Object Model COM that allows COM components to communicate across network boundaries This application installed on a computer connected with the microscope internal network allows an application developed with the library to interface itself with the microscope Figure 1 shows the internal network of the Quanta instrument ELECTRICAL MICROSCOPE CONSOLE CONSOLE MICROSCOPE CONTROLLER PC MI NE CROSCOPE TWORK SUPPORT PC SUPPORT NETWORK CUSTOMER LAN Fig 1 The block diagram of the instrument connections 34 of the project The following point has been the test of these libraries named XTLibInterface These libraries consist of a series of COM object directly instantiable and their COM interface retrievable by a client application through COM The testing phase has concerned the main libraries function to verify the working system and the code implementation The programming languages supported are Visual Basic e C the testing tools have been developed using the Visual Basic language 3 3 DEFI NI TI ON OF SPECI FI CATI ONS In this phase the main attention was paid to establish what the final application sho
95. rtional to the specimen thickness Areas of the specimen that are thicker will have fewer transmitted unscattered electrons and so will appear darker conversely the thinner areas will have more transmitted and thus will appear lighter Elastic Scattered Electrons They are the incident electrons scattered deflected from their original path by atoms in the specimen in an elastic fashion no loss energy These scattered electrons are then transmitted through the remaining portions of the specimen All electrons follow Bragg s Law and thus are scattered according to Wavelength 2 Space between atoms in specimen sin angle of scattering All incident electrons have the same energy thus wavelength and enter the specimen normal to its surface All incidents that are scattered by the same atomic spacing will be scattered by the same angle These similar angle scattered electrons can be collected using magnetic lens to form a pattern of spots each spot corresponding to a specific atomic spacing a plane This pattern can then yield information about the orientation atomic arrangements and phases present in the area being examined Inelastic Scattered Electrons They are the incident electrons that interact with specimen atoms in an inelastic fashion loosing energy during the interaction These electrons are then transmitted through the rest of the specimen Inelastic scattered electrons can be utilized in two ways 1 Electron Energy Los
96. s Spectroscopy the inelastic loss of energy by the incident electrons is characteristic of the elements that were interacted with These energies are unique to each bonding state of each element and thus can be used to extract both compositional and bonding state i e oxidation state information on the specimen region being examined 2 Kikuchi Bands bands of alternating light and dark lines that are formed by inelastic scattering interactions that are related to atomic spacing in the specimen These bands can be either measured their width is inversely proportional to atomic spacing or followed like a roadmap to the real elasticity scattered electron pattern 97 Appendix E The Transmission Electron Microscope The main condition to obtain the specimen interactions is the size of the specimen It has to be thin otherwise the electrons are stopped and no image would be performed In fact there is an axiom in TEM that thinner is better and specimens below 100 nm should be used wherever possible and in extreme cases specimen thicknesses 50nm are essential 1 E 1 3 Observation and recording of images The image on the fluorescent screen can be observed through a large window in the projection chamber some models even have two windows In order to examine fine detail or to assist correct focusing of the image a special fine grain focusing screen can be inserted into the beam and observed through a high quality 12x b
97. s are made available to the remote operator 4 1 THE MICROSCOPE VI DEO Different from other remote applications where the eyes and the hands of the operator are moved off the operation field and primarily ask for instrumental substitutes to emulate the human presence on site the SEM interfaces even with the local operator by a video screen and a set of commands that are linked to the internal sensors and actuators by some software interface The possibility to pipe the command set into a long distance net has been illustrated in the previous chapter and the tests showed the capability to drive a SEM through a loop connection a keyboard in front of the SEM screen was connected to the instrument via a server client connection running on the general LAN of the University or even on external public networks The SEM reacted as under local commands without errors or delays In other words the remote control was fully demonstrated while the operator microscope interaction was obtained by direct observation of the local non remote LCD screen Chapter 4 The video transmission Now a microscopist works looking at a live specimen that may also be the most unanimated crystal but continuously moves enlarges sharpens lightens under the solicitation of the instrument controls Then also a remote microscopist should look at the specimen in the same way The simple idea is to send to the remote operator exactly the same images observed by the loc
98. s the chamber is on air it is possible to open it and to position the specimen on the specific carrier called stage At this point the microscopist mechanically closes the chamber and enables the vacuum system to start air evacuation The pumps lead the pressure into the chamber to the operating value 10 torr in normal conditions And then continue working to maintain stable this value The pressure of the chamber is continuously monitored by dedicated sensors When the pressure value is correct a specific indicator enables the command to turn on off the high voltage Turning on the high voltage and then allowing an electrical current to heat the filament electrons are emitted The electron beam is now on the lenses are active and the detectors enabled The first rough images are displayed on the monitor 14 Chapter 2 The Scanning Electron Microscope To obtain a sharp image the electron beam has to be focused onto the specimen surface This means that the point of the beam with the minimum diameter has to strike the specimen Two movements into the SEM lead to this result there is the possibility i to bring up and down the specimen by mechanically moving the stage and also ii to focus the beam at different point through the electric regulation of the objective lens and then it is possible to reach the focus at different height This allows on one side to achieve optimal focus and also to know the distance of the specimen
99. se many instruments need to be remote in a different way operation in hazardous environments would require to have the operator remote in the sense of outside the analytical room as well as in the opposite case when the operator itself may be dangerous for the experiment as for High Resolution Electron Microscopy In those cases wide band local networking easily enables today to move the instrument consoles a few meters away from the experiment preserving the full control by the operator and the complete security of the closed LAN On the other side it is clear as those kinds of instruments are also the most immediate candidate to long distance control The concept of interaction implied in the term control is another split point There are instruments where the operator launches the experiment and waits for the results in seconds minutes or longer times as for instance in several methods for chemical analysis In that case the remote control is of limited requirements and also Chapter 1 The remote control of a scientific instrument of limited usefulness loading the specimen is of course a necessarily local operation and the remote control would only automate the setup of the analysis parameters and then the distribution of the output results another example a weather control station may be controlled by sending and receiving a very small quantity of data at quite relaxed time intervals which is currently performed by r
100. specimen images These results have strengthened the idea that implementing a dedicated encoder the video transmission will reach the best possible features Moreover an increase of the video quality will be certainly achieved because the original wont be a PAL video standard but a digital video created for the Super VGA system 63 Chapter 4 The video transmission REFERENCES 1 Viper Quad User s Manual 2 Gary Sullivan Pankaj Topiwala Ajay Luthra 7he H 264 AVC Advanced Video Coding Standard Overview and Introduction to the Fidelity Range Extensions SPIE Conference on Applications of Digital Image Processing XXVII August 2004 3 Thomas Wiegand Gary Sullivan Gisle Bjontegaard Ajay Luthra Overview of the H 264 AVC Video Coding Standard IEEE Transactions on Circuits and Systems for Video Technology J uly 2003 4 2005 F Mighela C Perra D Giusto M Vanzi Video Streaming in Electron Microscopy Applications nternational Workshop VLBVO05 Cagliari Italy September 2005 LNCS 3893 Springer 64 CHAPTER 5 Di SCUSSI ON PERSPECTI VES AND CONCLUSI ONS 5 1 ABOUT SOME UNEXPECTED TECHNICAL AND NON TECHNICAL DIFFICULTIES The results obtained by the Remote SEM Project are significant and confirm the possibility to fulfil the requirements for an efficient remote microscopy Starting the project a lot of uncertainty aspects were as obvious present but the idea to achieve the remote microscopy was cert
101. the commands to read and set the magnification value are located Developing it some errors have been found In fact when it has been loaded for testing a signal error has been given from the system Contrast and Brightness in this frame there are the commands to read and set the contrast and the brightness value The value is correctly changed but to see the new value it is necessary to reload the frame 39 of the project Image this frame is dedicated to display the images of the microscope testing the frame there is the same error as the magnification frame when it is loaded At this point before to continue with the framesets implementation considering the number of errors of the already done implementations a series of tests was programmed to debug them 3 5 1 The debugging phase The debugging phase has been carried out first frameset by frameset Then the framesets were integrated into the web site and tested again one at a time The web site test was performed into two different moments before it was carried out trying the connection to the site from the same web server computer in the second moment the connection to the web site was tried from another computer different from the server web and different from the microscope controller The results were that the framesets that worked independently and in the local server worked also in remote server the framesets that did not worked independently did not wor
102. the same even on non flat surfaces It follows that details at different depth in the specimen result equally sharp in the image which is quite uncommon in optical microscopy where the larger the magnification the smaller or even null the focus depth The human brain perceives any high focus depth as associated to not small objects those that do not 11 Chapter 2 The Scanning Electron Microscope require to be brought near the nose to be observed and then reads the SEM images as pictures of familiar macroscopic objects the eye of a fly recalls a soccer ball a tuft of hair a stack of logs a porous surface a Karst mountain The role of those accidents is often underestimated but makes the difference with many other scanning techniques and gives account of the wide diffusion of SEM images also outside the laboratories in scientific divulgation and even in commercial advertisements 2 2 THE STRUCTURE OF THE SCANNING ELECTRON MICROSCOPE A Scanning Electron Microscope consists of an electron optical column which contains an electron gun the electron source a series of electromagnetic lenses that produce several subsequent smaller and smaller images of the source and cast the last and smallest one onto the specimen and the deflection coils that move the beam to scan a defined area of the specimen The specimen is mounted under the electron column and a set of detectors suitable for the various emissions are mounted
103. the stage and then to move the specimen In this page it is possible to move the specimen changing the position of the red dot that represents the specimen on the stage The other signs presents are the stored positions Using the button Open Save and Clear it is possible to store recall and then clear some specimen positions in order to easy find again particular zones of the specimen Stage Map Coordinates Changed Locked LI 0 000 mm Li m 0 000 mm D Tz 9999 mm F rR oo EE rT oo ROB Absolute C Relative Add This page allows the same functions of the previews the difference is that to move the stage it is necessary to write the new coordinate values into the texts boxes 2 5 3 The Option page The Option page is the third it contains the commands to enhance the image displayed into the selected quadrant do Chapter 2 The Scanning Electron Microscope This frame shown in table 3 replaces the Stage frame of the Work page Table 3 A detailed description of the Option page Imaging In this frame it is possible to set the Magnification 500 x magnification value and to correct the 4 stigmatism value via the 2D box X and Y p coordinates Stigmator Enhance Image The digital brightness and digital contrast Digital Contrast 1 0 correction may be used to correct the contrast gt and the brightness of an image already st
104. the value The security is decisive to preserve the microscope from incorrect use and then possible damages The server client application realized until now satisfies almost these features In fact it is simple to use and it is not related to the original microscope interface As it is possible to see from figure 5 the commands have been grouped on the base of their function Moreover relating the user interface developed with the Quanta user interface it is immediately to note that are not corresponding Looking the picture it is possible to distinguee the different frame created to group the analogous commands 43 of the project Client i VLC ee yuv JPEE DirectX output Hex Vacuum OK 801497144 Electron Column E Detectors Spot Size 45 EM Rotation Scan Stigmator Ratton 30 3498 Mult 5 T DwellTime DwellTime 1 218052 OK microsec EE m Magnification ScanFiedl rea peo 338 Se Name EserList he Zl Detector List Gsed E Current position Stage Type X 0 765625 7 Y 3 2836354 SetY Active Detector Lfd z top Stage Z 26 373694 SetZ R 621 71614 Seth T 0 3609968 mm Channels Set Active Detector CCDChannel State Channel State Status Server Connection E Pw r Connection T pares EE eee Cose 18284153107 Addess Pressure BOTAS TAAE2TOBAE Pascal diu Run Microscope Dati inviati con successo Mode Conne
105. to change the magnification selecting directly the OK size scan area onto the specimen Increasing the 338 area selected the magnification decreases Stage The frame Stage allows the user to move the stage then to move the specimen mechanically The values of the axes are changed independently The T value is the value of the tilt rotation this value is changed manually then it is only possible to read it R 621 71614 Seth T 0 3609968 mm Status High BEDE e Pressure 8 01437144621084E Pascal ce The frame Status contains a report about the working conditions of the instrument The data reported are the High Voltage the Pressure chamber the microscope working mode and the chamber pressure state Vacuum OK 8 01 437144 The frame Vacuum gives the possibility to change the pressure value while the microscope is working in the Low Vacuum or ESEM mode 45 of the project Stigmator The Stigmator frame allows the astigmatism 30 9498 OK correction of the images following the axis of a x two dimensional Cartesian system Y 21 8052 OK The Channels frame allows making on pause state the image onto the screen the CCD channel is the image of the specimen chamber given from Channels the digital camera the other channel is the channel selected from the operator than it could CCDChannel State Channel State be the Secondary Electrons image the Backscattere
106. to overcome the natural barrier preventing it leaking out of the atom 77 Appendix B The hardware of electron microscopy electrons emitted from a small region of the filament This is done by saturating the filament that means increasing the filament current until the beam current no longer rises The bias potential controls the size of the region of filament that emits electrons and then it affects both the source size and the beam current If the bias is too high no region of the filament will emit and the beam is said to be pinched off The main reason to alter bias is to change the brightness of the beam The high voltage has a different role on the base of the instrument kind In the TEM higher energy electrons permit the examination of thicker specimens but may cause specimen damage Higher voltage microscopes are also more expensive Most TEMs have a maximum high voltage of about 200 kV In the SEM the maximum high voltage is usually from 25 kV to 30 kV and its choice is determined by a compromise between penetration which increases with high voltage and beam diameter which decreases with kV B 2 THE ELECTROMAGNETI C LENSES An electromagnetic lens is a relatively simple device By applying current to wire coiled around an iron cylindrical core a strong magnetic field is generated This field acts as a convex lens bringing off axis rays back to focus 3 The image obtained is rotated to a degree that depends on the strength
107. treamed Video Client is the application that allows the remote operator to see the video decode by the decoder It is a video player and it can be any video player H 264 compatible The more interesting and delicate aspect of this architecture project is the video encoder from it depends the satisfaction of the system properties above analysed There are more then one way to achieve a suitable encoder the first possibility is to choose a commercial encoder and testing its performance on the system the second is to realize a dedicated encoder In the last case it will be possible to adapt the encoding architecture and in particular the encoder parameters to the type of content and maximize the compression efficiency Anyway all these possible solutions have a common aspect they will be based on the H 264 AVC compression standard 56 Chapter 4 The video transmission The reason is that at the moment this is the best compression standard available Here a brief description of its main features is given while a detailed description of the encoder planning follows 4 3 3 The H 264 AVC video coding standard H 264 AVC is the latest international video coding standard 2 It was jointly developed by the Video Coding Expert Group VCEG of the ITU T and the Moving Picture Expert Group MPEG of ISO IEC It uses state of the art coding tools and provides enhanced coding efficiency for a wide range of applications including video te
108. ty and the Extended profile was designed to combine the robustness of the Baseline profile with a higher degree of coding efficiency and greater network robustness and to add enhanced modes useful for special trick uses for such applications as flexible video streaming The new standard has been designed for technical solutions including at least the following application areas 3 57 Chapter 4 The video transmission e Broadcast over cable satellite Cable Modem DSL terrestrial etc e Interactive or serial storage on optical and magnetic devices DVD etc e Conversational services over ISDN Ethernet LAN DSL wireless and mobile networks modems etc or mixtures of these e Video on demand or multimedia streaming services over ISDN Cable Modem DSL LAN wireless networks etc e Multimedia Messaging Service MMS over ISDN DSL Ethernet LAN wireless and mobile networks etc Moreover new applications may be deployed over existing and future networks This raises the question about how to handle this variety of applications and networks To address this need for flexibility and customizability the H 264 AVC design covers a Video Coding Layer VLC which is designed to efficiently represent the video content and a Network Abstraction Layer NAL which formats the VLC representation of the video and provides header information in a manner appropriate for conveyance by a variety of transport layers or storage media
109. uld do First of all the following features have been fixed 1 The user interface of the remote application should be friendly for a standard microscopist it should enable a remote operator to use the microscope as if he was in front of it 2 All the microscope commands necessary to perform an analysis with it have to be available for the remote operator These have been identified in high voltage ON OFF to set the high voltage value the chamber pressure value setting the brightness and the contrast the spot size value the magnification value to select scan speed to focus the image to move the specimen stage and to rotate the electron beam All these functions are available with the XTLibInterface libraries 3 The remote and the local operator should communicate directly through a chat for instance A system as a chat is useful to change information between the operators that are attending the analysis and also the possibility to work together on the same image for example it should be useful In order to satisfy this last point it could be added a digital blackboard to the chat 4 To carry out a microscopic analysis it is necessary to look at the specimen then it will be necessary to enable the remote operator to see the specimen sending the same video observed by the local operator The first solution consists to take the video card output to be sent on the 35 of the project network the video has to be elaborat
110. ure Socket Layer protocol SSL The web pages will contain the microscope commands and there will be a frame where play the video stream sent by the encoder In fact the video elaboration principle is the same as the first solution the difference is that the player is embedded in the web page In the other case it was a separate application The positive aspects of this way are the available friendly interface the reliability the security and the portability That negative is the nteractivity Then the characteristics are a little bit different from those of the server client application In fact a web site is accessible from every internet user independently from the Operating System used Vice versa the interactivity available with the ASP technology and the VBScript language is limited from the language and the technology itself Considering the positive and negative aspects of both selected solutions the web based solution has been preferred The decision has been made especially in view of its portability more significant than its low interactivity 3 5 THE MPLEMENTATI ON OF THE WEB APPLI CATION The web application development has been divided into two parts the implementation of the VBScript code for the microscope commands and the HTML code for the user interface and the development of the video coding system The first part is started organizing the web pages structure The home page is the access page here there is the logi
111. ve lens This determines whether the viewing screen of the microscope shows a diffraction pattern or an image The second intermediate lens together with the projector lens can be used for varying the magnification from hundreds to several hundred thousands of times This is done by varying the lenses strength Not all lenses will necessarily be used at lower magnifications E 1 2 Specimen interactions What happens in the specimen during the electron bombardment Contrary to what might be expected most specimens are not affected by the electron bombardment as long as it is kept under control When electrons impinge on the specimen a number of things happen 1 Some of the electrons are absorbed as a function of the thickness and composition of the specimen these cause what is called the amplitude contrast in the image 2 Other electrons are scattered over small angle depending on the composition of the specimen these cause what is called phase contrast in the image 3 In crystalline specimens the electrons are scattered in very distinct directions which are a function of the crystal structure these cause what is called diffraction contrast in the image 4 Some of the impinging electrons are reflected these are called backscattered electrons 5 The impinging electrons can cause the specimen itself to emit electrons these are called secondary electrons 95 Appendix E The Transmission Electron Microscope 6 The imping
112. whose buttons are located on the upper side of the frame A detailed study of the characteristics of the instrument and the operations available with the Quanta software has been carried out in order to define which characteristics and functions to include into the remote application and which not for the sake of instrument safety This has been accomplished by page by page analysis as reported in the following sections 2 20 Chapter 2 The Scanning Electron Microscope 2 5 1 The Startup page The first page named Startup page contains the commands used for changing the microscope parameters for setting the work mode for choosing the detector and for verifying the microscope status during the analysis The name used to identify these commands is Vacuum Mode Electron Column Source Control Detector and Status Table 1 gives a detailed description of the Startup page functions Table 1 A detailed description of the Startup page m Vacuum This frame allows the user to ventilate and pump Vent the system and to choose the working mode High Vacuum is the default mode If the Low Vacuum Mode mode or the ESEM mode is selected it is possible High Vacuum to adjust the pressure value into the specimen chamber using the scroll bar Moreover it is C Low Vacuum Water possible to choose the gas Selecting the Pump button it starts the pump down s ese water z of the specimen chamber When the chamber is
113. wn in figure 3 REQUEST Tl S i o ir Fig 3 The data flows from the client to the microscope To build two applications able to communicate a particular software object has to be used the socket Using Visual Basic as the programming language the object used has been Winsock The socket is a software object that connects an application to a network protocol Then using it both for the client and the server application these will be able to connect to the network and the client will connect itself to the server using the server IP address The server is then connected to the microscope controller computer through the DCOM client and then can forward the client requests to control the instrument Also this implementing work has been divided into two parts the implementation of server client applications to control the microscope and the implementation of the video application The last will be explained in detail in the third chapter To realize the first part the work started from the code written to test the XTLibInterface The already developed functions have been integrated into a unique form At the same time the not yet added functions have been first tested and then integrated The main attention during this work has been paid on the user interface configuration In fact the remote microscopes are not a commercial product but there are few prototypes into some research laboratories One of the goals of this work is to implem