Coagulometer Koag-o-test Vet User Manual

image

Contents

1. gt keys B Coagulation Quantitative C Coagulation 0 Results get from other items Then press V or ENTER key to turn to next step Now the cursor stays at the back of RESULT METHOD it will show S R I E U press V key it will select successively S R I E U Among them S second method R short for PTR I short for INR percentage method E absorbance value U consistence method press gt key you can select or hide the result method in the cursor Press V key or ENTER to turn to next item Input normal value s high limit in HIGH REF the value of reference reagent list is 1 9999 Input normal value s low limit in LOW REF the value of reference reagent list is 1 9998 and the low value must be less then high value Input the item s normal contrast value in NORMAL CONTRAST VALUE BGT BioGenTechnologies GmbH Koag o test Vet user manual Input ISI value in ISI O Percentage method If you select in result method you must input S standard curve Consistence curve If you select in result method you must input C S standard curve Absorbance curve If you select 0 in result method you must input C E standard curve Max Test time Time to test the given item If the code you input in is not existed in internal memory it will show THE CODE
2. This function is close to the interface of PRINT ALL PATIENTS REPORTS Chapter 5 Utility There are 3 submenus for you to setting you can modify and add test items in PROGRAM or test and print Q C report in QC or setting system utilities in UTILITY 5 1 Program This section mainly introduces how to setting the reagent test parameters It suggests you to setting parameters according to the reagent manual for every time to change the reagent 1 Enter program submenu Press 4 key to enter program submenu in main menu it will show 1 ITEM DEFINITION 2 SHOW ITEMS SUM 3 DELETE TEST 2 Item definition Main function input operation parameters of all test items into the instrument follow the below procedures to finish it BGT BioGenTechnologies GmbH Koag o test Vet user manual D Press 1 it will show ITEM No XX NAME Input a code which is 1 bit or 2 bits in XX it can t be 0 press ENTER key if it is less than 2 bits If the code which you input has been existed in system then search the code and show the parameters on the screen the cursor will stop at the back of NAME You can input or modify the test name here please input the abbreviate name which is 1 4 bits such as PT APTT TT ETC ITEM CODE XX NAME XXXXXX Press V key or ENTRE key to enter next step Select one in B C and 0 with lt and
3. instrument 3 3 5 Press OPTIC key add TT reagent after the screen show Add reagent Press it to the BGT BioGenTechnologies GmbH Koag o test Vet user manual bottom until hearing the stridulation the instrument starts to test 3 3 6 The screen show RESULT XX XS after about 40 seconds 3 3 7 Test continually with repeating step 4 to step 6 Press MENU key back to main menu 3 4 Common use Clotting and reagent dosage TEST ITEM SAMPLE ul REAGENT 1 ul REAGENT 2 ul PT 20 40 APTT 20 20 CaCl 20 TT 30 30 4 4 Result Report Enter REPORT submenu to edit and print Press 3 in main menu it will show 9 1 EDIT SAMPLE 2 DEL RESULT Edit modify all patients info by the first item including name gender age and blood group Dele patient s given result by the second item Add patient s given result by the third item Print Dele patient s all results by the fourth item 4 4 1 Edit sample info You can input every patient s name gender age and blood group according to this function Firstly it will show SAMPLES SUM XXX SAMPLE No __ MENU RETURN Input ID 1 59999 or search with 1 key press enter key to enter next interface after found needed ID If the ID you input is not existed in internal memory it will show SAMPLES SUM XXX SAMPLE No XXXXX THE SAMPLE IS NOT EXISTED CREAT NEW ITEM Y Creat new
4. Wide X 11cm in Height Environment Temperature 15 C 30 C Humidity lt 90 Weight 8kg 1 3 Structure Introduction 1 Main components view Display Thermal Printer Operation Keyboard Reagent incubation Sample incubation position position Test channel BGT BioGenTechnologies GmbH Koag o test Vet user manual 2 Operation Panel view F Se Temperature Digit keys input indication light digits and characters Menu key back to last menu Cursor keys Enter key enter the operation Feed key handwork feed paper Software timer pap startup key Optical test channel activation key 3 Back view Syringe socket fixation seat RS 232 computer port Three core Earth pole power socket 4 Menu l MEASURE 5 QC 2 PRINT 6 CLOCK The menu has 7submenus after power is on the operator press the cursor keys 1 7 to enter into the corresponding menu directly In daily work the operator do the first item to measure samples do the second item to print out the result report BGT BioGenTechnologies GmbH Koag o test Vet user manual Chapter 2 Installation 2 1 Save and Transportation Condition Environment Temp 0 C 40 C Relative Humidity lt 90 2 2 Work Condition Koag o test Vet Coagulator is a precise electrical instrument for the instrument can be used normally and get an accurate result the environment must acquire cer
5. and you will get D Decio D Press 2 key twice quickly and you will get E O eee aed E Press 2 key three times quickly and you will get F 2 2 2 F You can get the others alphabet and mark with the analogy 2 Numeric input method Input numeric directly under the Numeric input interface with pressing 3 You can select capital letters or miniscule by using OPTIC key 24 BGT BioGenTechnologies GmbH Koag o test Vet user manual Appendix 2 Install printer paper 1 Open the printer cover interface press FEED press it again to stop it after there is no residual paper get the surplus paper out 2 Put new paper in paper groove the thermal printer paper are different in two sides so pull the end of paper out when the thermal side faces downwards Paper Cover The non therma paper groove output paper roll axis 3 Take the roll axis out upwards put the thermal side under the roll axis as figure 2 the direction of paper should be vertically with the one of the roll axis it avoids the printer paper to be locked so that it can t print normally 4 Install the roll axis put the gear in the right side of the printer then install another side and press it to lock it tight 5 Press FEED adjust it if the paper can t get out or its direction is not vertically with the one of roll axis or it can t work press FEED to stop it after printing a segment When paper get out
6. of the paper cover 6 Pull out of the printer paper from cover and then close it Attention FEED function consists in Main menu Measure sample interface Waiting keyboard input interface PRINT DELETE RESULT REPORT interface and RESULT REPORT of Q C report interface close FEED function in other interfaces 25
7. print Q C curve and result in QC QC menu 1 SELECT QC TEST 2 SELECT QC NO 3 PRINT QC CURVE 4 PRINT QC RESULT 1 Preparation 1 1 Select Q C material 1 The Q C material dry powder produced by professional supplier 2 Healthy person s blood plasma 3 Animals blood plasma 1 2 Prepare Q C material Dilute the Q C blood plasma dry powder according to the manual lay it aside for 20 minutes use it after it becomes stable 2 Test Q C material 2 1 Enter REPORT submenu in main menu creat a patient No such as 59000 It suggests to input 3 to avoid to confuse with other normal patients 2 2 Enter MEASURE submenu press 1 to input the item No such as TT input 3 press ENTER to assure and quit item installation then press 2 to input the patient No such as 59000 press ENTER key to assure and quit patient No installation 2 3 Measure the prepared Q C materials with the method to measure sample 2 4 You needn t do the first step operation in testing the item Q C result repeat 2 and 3 steps directly item No and patient No should be the same to the ones in first time to test You can enter QC to print the Q C curve and result if the test results over 5 units 3 Print Q C result and curve 3 1 Enter QC select the item No of the Q C in the operation I SELECT QC TEST and setting the target value You can eva
8. Coagulometer Koag o test Vet User Manual a lest OOOOO 0000 BGT BioGenTechnologies BGT BioGenTechnologies GmbH Von Langen Weg 10 D 48565 Steinfurt Tel 02551 4090 Fax 02551 1298 Web www biogentechnologies de E Mail info biogentechnologies de BGT BioGenTechnologies GmbH Koag o test Vet user manual Content Chapter T IOAHELON ra maren estes dabhed saeta eend 2 11 Test PANCI ic 92 ted dd id 2 1 2 Technical Criterions iccc tccciccctccsdessdccsdestecstscvdessdsevdesstsevlesudsevdestscvlesulsevdesstscvlesniderdesetsentesune 3 1 3 Structure Introduction nononono eneeeneeenenenerenenenereneeeneeeneeeneeeneeenereneeeneeeneeeneeenvenee 3 Chapter 2 stat icons 5 2 1 Save and Transportation Condition ns ennn sernerrnserensereneeersnerrnserenneeenneersneesnneeennn 5 22 Works CONO DUI 5 2 3 Instrument Installation aenar ni sous A A E A A EA 5 Chapter 3 Test Flow Chatero s 6 Chapter 4 General Operation Sa 7 4 1 Sample Preparation A Se NN 7 4 2 Testing Ml mossi r a rias 8 4 3 Basic Operation Rules ns nnsennserenserenneeerneerrneerseersnserenneeenneeennneersnerenneeenseeenen 10 4 4 Result Reportan ta da dao 11 Chapter a UI is 15 O EN PEI A e es EA 15 52 0 Cxand Reference Value miii iii in A 18 O Kool E PE E EEE A A E E A A A E EE E 19 SAUTE Yi A a aia 19 Chapteto UV alien ane ey orr ae Ea E hae 22 SPAR EET aL Ce PPAT E A E EAEI AAN E ENE E O E EE N 22 Trouble and Handle sic ccscccc
9. ELE The position of 0 is the result of the given patient APTT shows the item name position of 35 8 shows the test result S shows the unit You can check them with A V press ENTER key to delete if there is any results you needn t This step can t be recovered so be careful Press MENU to return to last interface 4 4 3 Add result Some tests can t be finished in this analyzer so you must use other instruments while the results must be sent out concurrently define this test before using this function SAMPLE SUM XXXXX SAMPLE No MENU RETURN ID which you input here can be existed No or new No use search function at the same time Then it will show SAMPLE No XXXXX NAME XXXXXXXX ITEM No __ Now you can use search function V or input ID directly but it must be defined ID It will show ID XXXXXX NAME XXXXXXX ITEM No XX NAME XXXXX RESULT _ Input the result get from other methods press MENU key to return to last interface BGT BioGenTechnologies GmbH Koag o test Vet user manual 4 4 4 Print Dele result report 1 PRINT APPOINTED SAMPLE 2 PRINT ALL SAMPLES 3 DEL SELECTED SAMPLE 4 DELE ALL SAMPLES 1 Print Report 1 Print the appointed patient If you want to print the appointed patient select this function it will show PRINT SAMPLE No MENU RETURN You can input ID 1 59999 to print directly or search wit
10. IN NOT EXISTED CREAT NEW ITEM Y N If you answer Y it will save the code and turn to step If you answer N it will give up the code and return to input CODE state CURVE EXAMPLE S CURVE C S CURVE C E CURVE 0 0 40 0S 575 0mg dL 6 0S 575 0mg dL 0 080E 25 0 30 0S 287 0mg dL 11 6s 287 0mg dL 0 055E 50 0 22 08 143 5mg dL 19 8S 143 5mg dL 0 039E 75 0 16 08 71 8mg dL 34 4S 71 8mg dL 0 020E 100 0 12 0S 0 0mg dL 60 0S 0 0mg dL 0 0E 3 Item sum Press 2 ITEM SUM to see saved test item sum in internal memory press 1 the system will recover factory installation any modification saved in memory will lost it will suggest after press BGT BioGenTechnologies GmbH Koag o test Vet user manual 1 once press 1 twice if you want to continue or return to last interface with pressing any key There are solidified 15 test items in factory 4 Dele test items This function will delete unnecessary items delete them first in internal memory or 1t will show wrong suggestion 5 2 Q C and Reference Value The aim of Q C is to test control the precision degree of the test supervise the change of stability improve the conformity of sample test result about one group or between two groups in common test work it will assure the reliability of the test result for every patient sample You need to test Q C material in MEASURE then
11. RT 3 Press MENU back to last menu 6 SET CLOCK Enter SET CLOCK submenu it will show SET CLOCK 1 TIME FORMAT 2 DATE AND TIME MENU RETURN Set the time with pressing 1 or 2 then it will return to last interface You can enter CLOCK interface to observe it 7 COM SETUP Press 7 enter the set COM interface COM SETUP SERIAL ON PARITY EVEN BAUD RATE 19200 If you want to transmit the sample data measured just now to the program of the computer you can enter this function to setup the RS 232 port 8 System Reposition When the system is disturbed by other reasons this function can help you to reposit the system be back to out firm installation Input XXXXX in UTILITY submenu to startup the function it 21 BGT BioGenTechnologies GmbH Koag o test Vet user manual will ask you 1 RESTORE SYSTEM 2 RESTORE DATABASE Press 1 to reposit the system then press Y twice press 2 to clean the patients file away then press Y twice press MENU back all patients information and results will be lost after reposition in internal memory so be careful Chapter 6 Maintenance 6 1 Maintenance Maintenance of Koag o test Vet Coagulator is as follows 1 Maintenance for every day 1 Put blank flow cell to protect light source when test channel is not in using 2 Check if there is dust in test channel and e
12. ccssccssecesecsvccesccsvecesccevecesccevecescceveces dcevecesecevecesccevecesccevecesccevaceseceveseses 23 Appendix 1 Input MEDE ca aaa 24 Appendix 2 Install punterpaPrlaai aii 25 BGT BioGenTechnologies GmbH Koag o test Vet user manual Chapter 1 Introduction 1 1 Test Principle Koag o test Vet coagulator is a single channel Hemostasis Analyzer designed to test manifold coagulate parameters Its test principle is optical colorimetric method After mixing reagent and blood plasma Fibrin protein change to Fibrin protein solidification so that the light consistency to test sample is changed then the instrument test the solidification end It can test all parameters in coagulating process in a large range including ITEM No ITEM 1 PT 2 APTT 3 TT 4 FIB 5 VT 6 PS 7 PC 8 F2 9 F5 10 F7 11 F8 12 F9 13 F10 14 F11 15 F12 16 XA 17 LA BGT BioGenTechnologies GmbH Koag o test Vet user manual 1 2 Technical Criterions Sample incubation position 5 Reagent incubation position 2 Software Timer 1 Test Channel Number 1 Saved Items Sum 1000 Wavelength 470nm Temperature Control 3740 5 C Reagent Consumption Low to 20 uL Sample Consumption 20 30 uL Coherence Error lt 5 Calculation S PTR INR g L etc Power Supply A C 220V 15 50 HZ 3HZ Efficiency 120W Inner Printer 24 bits Thermal Printer Wide 58mm Size 34cm in Length X 30 5cm in
13. e y pol 1 ml CHART 1 CHART 2 CHART 3 CHART 4 Add reagent operation Put the pipette in the cuvette s cliff as chart 5 add reagent Press the pipette to the bottom the reagent must be discharged with certain speed in order that the reagent and plasma mixing equably the pipette should be over the liquor surface without touching it as chart 6 CHART 5 CHART 6 BGT BioGenTechnologies GmbH Koag o test Vet user manual 4 3 Basic Operation Rules 1 Power on Switch the electrical power on the system starts to self test the constant temperature instrument heat the test channel concurrently TEMP indicator light will be on after waiting for 10 20 minutes the tests can be done 2 Power off Press MENU key to return to the main menu after finishing operation then switch the electricity power off remember it can not be power off in the process of testing or printing 3 Basic Operation Rules of Parameters 3 1 PT TEST SHANGHAI TAIYANG REAGENT 3 1 1 Switch the power on wait for the TEMP indicator light on Sample 3500 centrifugal for 10 minutes 3 1 2 Incubating the PT applied reagent Get proper quantity according to the tests for 10 minutes at 37 C 3 1 3 Add plasma 20ul incubating for 2 minutes Add the sample to the bottom of the cup without any bubble 3 1 4 Put the incubated sample in the test channel add 40ul PT applied reagent with online sampling instrument 3 1 5 Press OPTIC key add PT reagen
14. ed because of some reason then it will save wrong data it will effect the result report if without modification this function is as follows 1 MODIFY SUM 2 DELETE THE APPOINTED SAMPLE C D E A A The amount of patients saved in the system range 1 255 20 BGT BioGenTechnologies GmbH Koag o test Vet user manual B The actual amount of patients detected by the instrument C Current position of patients file range 0 249 D Patient No range 1 9999 E Saved patient s sample range 0 30 1 In normal condition A should be equal to B now 1 key is useless In actual application because there are too many disturbance factors especially the moment of turning on and turning off it is instability in the twinkle of power on and power off now it is very easy disturbed to make mistake It will suggest wrong information after detecting the mistake now you need adjust the patients file with handwork modification or it will show wrong result Method Press 1 the cursor will appear in A modify the value in A equal to the one in B then press ENTER key to correct the mistake 2 If you delete the patient s information in patients file search the patient with ft the patient No will be shown in D press 2 to delete the patient Note This operation can not be recovered so be careful If you delete the appointed result of the patient operate in RESULT REPO
15. h V key press ENTER key to print patient s report or press MENU to return 2 Print all samples a Startup this function it will show START FROM _ Now you can input the startup date of the patient then it will show BEGINDATE XX XX END DATE _ It will look for all files in internal memory after you input an end date starts to print if you find the result fit to the condition print for every result until finish printing if you want to stop printing in midway only method is to close the system b When the system suggests you to input the BEGIN DATE press ENTER key to print all results in internal memory then it will show PRINT ALL SAMPLES Y BGT BioGenTechnologies GmbH Koag o test Vet user manual Now you can press Y N MENU Press Y to print all results instantly Press N to return to BEGIN DATE state Press MENU to return to last interface 2 Dele patient report 1 Delete appointed patient report Select this function to delete the appointed patient s record and all saved results of the patient in internal memory Firstly it will show DEL SELECTED SAMPLE Sample No _ MENU RETURN You can input with digit keys or use search function then press ENTER key to delete Press ENTER key after you find the appointed ID the system will delete it directly without suggestion so be careful 2 Dele all patients
16. luate or consult other materials 3 2 Select the corresponding patient No such as 59000 in the operation 2 SELECT QC No BGT BioGenTechnologies GmbH Koag o test Vet user manual 3 3 Carry out the Operation 3 4 to print the Q C curve and result Including MEAN and SD after carrying out operation 1 2 3 4 When you test and print the Q C curve and result later you should define the target value according to the last Q C result Note 1 The operation for all patients in every test are the same except the test items Select the patient No and the same test items for each test you can not test other items for the appointed patient No or it will not print the correct Q C curve and result 2 If there is not target value for the Q C material you can test it for 10 times continually to assure all items target values then calculate the MEAN and SD as the target value and deviation The target value should be ascertained when you change new reagent 3 Because the normal value range for different areas and different people are different also the instruments are different other range only can be reference Please creat the normal range fit to this area s people 5 3 Clock You can enter CLOCK submenu it will show current time You can modify time and date in MAINTAIN Koag o test Vet COAGULATION ANALYZER DATE 01 01 2001 TIME 12 00 00 MENU RETURN 5 4 UTILITY This menu mainly consi
17. ment will give off stridulation after testing the corresponding test result will be shown on the line of XXXXX Y Y Y Y Y j Cuvette Lean on the j Cuvette s cliff i 5 Press 1 and 2 to set new items and patient No if the test is unfinished Do with the tested cuvette get a new incubated sample in the test channel test the new one 6 Press MENU key back to the main menu after finishing last test Then enter the Result to do the corresponding operation and to print the tested patient s total result report Addition 1 Other channels can be startup when one channel stars to test these channels test concurrently without disturbance BGT BioGenTechnologies GmbH Koag o test Vet user manual 2 If the patient s ID is 0 the instrument will not save or print the channel s result after testing 3 Operate carefully when add samples into sample instrument do not let bubble enter into the liquor The correct operation to add plasma and add reagent in as follows Add plasma operation Put the pipette s tip to the bottom of the cuvette Press the pipette to the first stop position the state is as chart 2 there is some plasma in the needle Put the pipette to the position as chart 3 press the pipette to the bottom to add residual plasma The pipette should be over the liquor surface when add all plasma as chart 4 it will effect the test result if it is underneath or touching the liquor surfac
18. ndicate light is on it suggests the temp of instrument s two test channels have added to the constant temperature of 37 0 5 C sample testing can be started In the main menu press 1 to enter into the MEASURE interface it will show SAMPLE No 4096 T1 00 00 TEST No 2 NAME APTT 1 Press TIMER key to startup the software timer when incubate the sample and reagent the incubation time will be shown on the screen to help the operator to finish incubation correctly It will give an alarm when the time is enough press TIMER key to stop the alarm 2 Press 1 A appear the cursor input the test items input the corresponding item No directly to select the needed items press ENTER key to assure the system will ENTER automatically if input 2 numbers press MENU key to cancel 3 Press 2 B appear the cursor input patient No then press ENTER to assure or press MENU to cancel Every channel finishes per test the patient No will add 1 automatically please modify the patient No for the same patient 4 Put the incubated sample in the corresponding test channel absorbing the incubated reagent quantity with pipette press OPTIC X Please add reagent will be twinkled at XXXXX YYYYY row on the screen add reagent instantly along the cuvette cliff press the pipette to the bottom Move away the pipette instrument starts self test automatically The instru
19. o the three core socket which has installed earth line BGT BioGenTechnologies GmbH Koag o test Vet user manual Chapter 3 Test Flow Chart Switch power on Open instrument to incubate until the TEMP light is on Press 1 to enter MEASURE submenu Put the sample or reagent in the incubation position press TIMER to time Press 1 to input test items in all channels Press 2 to input test patient No in corresponding channel Put the incubated sample in corresponding test channel Press OPTIC PLEASE ADD REAGENT twinkle Add reagent along the cuvette without bubble Self test automatically Show the result on the screen YES Prepare next sample and reagent No Press MENU to return Press 2 to enter PRINT Edit and Print result Press MENU to return Switch power off BGT BioGenTechnologies GmbH Koag o test Vet user manual Chapter 4 General Operation 4 1 Sample Preparation Two methods to collect samples the flow chart is as follows Collocate anticoagulant reagent Vein Sample Blood Mix with anticoagulant instantly Anticoagulant tube sample blood vein directly 3600r min centrifugal 10min get the upper liquor blood plasma 1 Use 0 109mol L sodium citrate liquor as anticoagulant reagent the ratio of blood sample to anticoagulant is 9 1 2 The hemostatic belt must not be tied too tight and the time must not past 5 minutes in vein
20. patient record if you answer Y or press ENTER then turn to next step answer N or press MENU to return BGT BioGenTechnologies GmbH Koag o test Vet user manual SAMPLE S SUM SAMPLE No __ You can input new patient ID Turn to next step after finding patient or creating new patient it will show at the second row SAMPLE No XXXXX NAME You can input 12 characters here at most press ENTER key if it is less than 12 characters then it will show the patient ID and name at the first row ID XXXXX NAME XXXXXXXX GENDER X AGE XXX BLOOD GROUP XX Gender input M or F M is male F is female Age input 0 999 unit is Y year Blood group A B O AB Note Only input corresponding content in above items the system will accept If the ID you input is existed in internal memory it will show the saved info Directly you can modify them then press MENU to return you can t input the patient s TEL and ADD but there 1s room in result report for you to add 4 4 2 DEL Result You can delete wrong result in test result according to this function Firstly input patient ID which has been existed or use search function SAMPLE SUM XXXXX SAMPLE No NATINITT DETI IDNI Next it will show BGT BioGenTechnologies GmbH Koag o test Vet user manual SAMPLE No XXXXX NAME XXXXX ITEM SUM XX 0 APTT 35 8 S PRESS ENTER TO D
21. sample blood or the cruor and the fibrin dissolve gene will be activated Sampling can not be done in vein transfusion position or neighborhood position Puncture must appear blood for once avoid blood clot bubble pollution of bubble and tissue the quantity of sampling blood must be accurate it can not be used if it is less than 90 of arranged quantity Select the second one or third one if select anticoagulant tube Vacuum sample tube 3 Mixing the blood sample with anticoagulant sufficiently and instantly after sampling the blood centrifugal as soon as quickly 4 The centrifugal speed is 3500rph time is 10minutes the test operation must be done in 60minutes after sampling blood 5 The blood plasma tube must be airproof with plug to avoid the PH value to raise 6 All instruments injectors sampling tube testing cup etc which contact with sample must be plastic or silica glass metal It can t use simple glass tube or the cruor will be activated also it can t use eradicator or purge liquor 7 Several factors infect the test results 1 Na2C204 EDTA and glycogen aren t fit as anticoagulant 2 Use fibrin solve liquor can prolong the coagulate time 3 Sampling and management improperly will change the coagulate time 4 The instruments which used repeatedly must clean in time and drying BGT BioGenTechnologies GmbH Koag o test Vet user manual 4 2 Testing sample Switch the power on for 10 20 minutes the i
22. sts in UTILITY submenu Press 7 to enter it in main menu 1 TEMP ADJ 5 DATABASE 2 SET FAN 6 SET CLOCK 3 ADC 7 COM SETUP 4 PTR SETUP 1 TEMP ADJ The screen will show SET TEMPERATURE 37 00 CELL TEMPERATURE 37 02 CURSOR CHG SET MENU RETURN You can change or set temperature with lt gt keys by 1 C unit or with V A keys by 0 1 C unit 2 SET FAN It will show BGT BioGenTechnologies GmbH Koag o test Vet user manual SET FAN 1 TURN ON FAN 2 TURN OFF FAN MENU RETURN Now 1 Fan is on 2 Fan is off 3 ADC It shows the internal AD transfer parameter of the instrument service personnel use it for maintenance only 4 PTR SETUP 1 PRINT DAILY RESULT 2 INSTANT PRINT 3 FORWARD PRINT 4 LQ300K Select I PRINT DAILY RESULT then to print report in REPORT it will print all patients and results in internal memory for that day General select this mode You can select the time to print the report OIf you don t print report during test please don t select 2 INSTANT PRINT If you print a report for every sample during test please select 2 INSTANT PRINT Select 3 FORWARD PRINT it will print out results forward otherwise it will be backward Select 4 LQ300K only when you want to print by the outer printer such as EPSON LQ300K 5 DATABASE The internal memory will be effect
23. t after the screen show Add reagent Press it to the bottom until hearing the stridulation the instrument starts to self test 3 1 6 The screen show RESULT XX XS after about 30 seconds 3 1 7 Test continually with repeating step 4 to step 6 Press MENU key back to main menu 3 2 APTT TEST 3 2 1 Switch the power on wait for the TEMP indicator light on Sample 3500 centrifugal for 10 minutes 3 2 2 Get some APTT reagent Get proper quantity according to the tests without incubating 3 2 3 Add plasma 20uL mix with 20uL APTT reagent incubating for 3 minutes at 37 C 3 2 4 Put the incubated mixing liquor in the test channel absorb 20Ul CaCl incubated liquor with online sampling instrument and it s consistency is 0 025mol L 3 2 5 Press OPTIC add CaCl liquor When you hearing the stridulation after the screen shows ADD REAGENT it starts to test 3 2 6 It will show RESULT XX XS after about 100s 3 2 7 Test continually with repeating step 4 to step 6 press MENU back to main menu 3 TT TEST 3 3 1 Switch the power on wait for the TEMP indicator light on Sample 3500 centrifugal for 10 minutes 3 3 2 Incubating TT reagent Get proper quantity according to the tests for 10 minutes at 37 C 3 3 3 Add plasma 30ul incubating for 3 minutes Add the sample to the bottom of the cup without any bubble 3 3 4 Put the incubated cuvette in the test channel add 30ul TT reagent with online sampling
24. tain requirement as follows Environment Temp 15 C 30 C Relative Humidity lt 90 Pressure 86KPa 106KPa Voltage A C 220V 33V 50Hz 3Hz Earth Pole Special Medical instrument s earth pole the required resistance is less than 0 50 Addition the instrument must be put in a platform which has enough room stable level without any vibration the position should avoid the equipments such as fans air conditioners etc in order not to effect the accuracy of temperature control it is prohibited to be installed in the environment which is high temperature sunlight shines directly or has much dust also it can not be put with the equipments which has strong efficiency heavy interference such as the centrifuges refrigerators thermostats etc in the same room or use the same elec Power socket 2 3 Instrument Installation 2 3 1 Open the instrument s packing box First check carefully whether the instrument accessories in the packing box match with the packing list If there is any shortage or damage contact the suppliers quickly 2 3 2 Second get out the instrument cautiously after checking carefully then put it on a stable smooth work platform 2 3 3 Third Connect the earth pole to the earth which is on the right side of the back of instrument with medically special earth line 2 3 4 Fourth Get the elec Power line connect one line to the elec Power socket which is in the middle of the back board then connect the other line t
25. ut no doesn t connect to the printer head l 4 thermal face to printer head letters b Printer paper isn t thermal i b Install printer paper correctly paper a Invalidate reagent a Replace reagent b Incubate reagent and sample b Operate follow the guideline improperly Instable results 7 c Earth line hasn t been connected c Connect earth line again with disturbance d Contact our customers d Trouble with instrument department If there are other trouble which can t be solved please contact our after service department 23 BGT BioGenTechnologies GmbH Koag o test Vet user manual Appendix 1 Input Method The system uses input method in PATIENT NAME DOCTOR NAME REPORT HEAD ADD TEST ITEM the consent method is ENGLISH press MENU to enter INPUT METHOD interface it will show INPUT METHOD 1 ENGLISH CAPITAL 2 ENGLISH MINISCULES 3 NUMERIC 1 English input method The correspondent relation of cursor keys to ENGLISH characters are as follows 1 A B C 2 D E F 3 GHI 4 J K L 5 M N O 6 P Q R 7 S T U 8 V W X 9 Y Z 0 a B y 6 6 n 0 LK A 1 V 6 0 7 0 T 0 0 W 0 Press 1 key one time and you will get A ai tes A Press 1 key twice quickly and you will get B an eee oe BEN B Press 1 key three times quickly and you will get C 1 1 C Press 2 key one time
26. very incubation position everyday clean them with soft tissue or tampon with alcohol 3 Clean used flow cell and needle Method First clean them with tap water then immerge them with 0 5 NaClO liquor for 30 minutes clean them with tap water again at last clean them with DI water twice dry them for preparation 2 Clean instrument s appearance Wipe the instrument appearance with wet cloth Turn the electricity power off and get away the plug Note Clean the instrument with clean water neutral purge reagent or non causticity scour but you can t clean it with organic reagent or hot water or it will be distorted or damaged 22 BGT BioGenTechnologies GmbH Koag o test Vet user manual Trouble and Handle phenomena cause method a The power doesn t supply a Turn on the power b The electricity doesn t work b Check the power socket No Display a i c Signal display line is loose c Check the display cable d Matter with screen d Replace screen Adjust the contrast adjustor The contrast of the display is Dark Display changed with the environment Cont l adjustor a Replace the new paper b Check if there are any gauffer on dn a No paper Fail Printing the paper or if the paper is vertical b Printer paper is put incorrectly i aj with the roll axis If it is not uplift the roll axis and put paper even a Printer paper thermal face a Connect the printer paper Print b