1. ill LAIM points J D15 Do Tx Data J Timing Resolution 40 0ns J a goons Bus fd i on Capture Triggering on a specific transmit data packet going across an RS 232 bus A complete set of triggers including triggers for specific serial packet content ensures you quickly capture your event of interest Capture Discovering a device fault is only the first step Next you must capture the event of interest to identify root cause Accurately capturing any signal of interest begins with proper probing The MSO DPO4000B Series includes four low capacitance probes for accurate signal capture These industry first high impedance passive voltage probes have less than 4 pF of capacitive loading to minimize the affect of the probe on your circuit s operation offering the performance of an active probe with the flexibility of a passive probe The MSO DPO4000B Series provides a complete set of triggers including runt timeout logic pulse width glitch setup hold violation serial packet and parallel data to help quickly find your event With up to a 20M point record length you can capture many events of interest even thousands of serial packets in a single acquisition for further analysis while maintaining high resolution to zoom in on fine signal details From triggering on specific packet content to automatic decode in multiple data formats the MSO DPO4000B Series provides integrated support for the in
2. MSO04104B L 1 GHz 5 5 2 5 GS s on 1 2 4 channels 5M record length 4 16 channel mixed signal oscilloscope MS04104B 1 GHz 5 5 5 GS s on 1 2 4 channels 20M record length 4 16 channel mixed signal oscilloscope All Models Include One passive voltage probe per analog channel TPP0500 500 MHz 10X 3 9 pF for 500 MHz and 350 MHz models TPP1000 1 GHz 10X 3 9 pF for 1 GHz models Front Cover 200 5130 xx User Manual 071 2810 xx Documentation CD 063 4300 xx OpenChoice Desktop Software NI LabVIEW SignalExpress Tektronix Edition Software Calibration Certificates document measurement traceability to National Metrology Institute s the Quality System this product is manufactured in is IS09001 registered power cord accessory bag 016 2030 xx and a three year warranty Please specify power plug and manual language version when ordering MSO Models also Include One P6616 16 channel logic probe and a logic probe accessory kit 020 2662 xx Application Modules Application modules have licenses which can be transferred between an application module and an oscilloscope The license may be contained in the module allowing the module to be moved from one instrument to another Or the license can be contained in the oscilloscope allowing the module to be removed and stored for safekeeping Transferring the license to an oscilloscope and removing the module permits the use of more than 4 applications simultaneously Module D
3. Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Data Sheet g f i s Cpp a m O e S E aD 4 d sets Cr Ac a ajes h i gt D paol 7 re Shir er Features amp Benefits Key Performance Specifications 1 GHz 500 MHz 350 MHz Bandwidth Models 2and 4 Analog Channel Models 16 Digital Channels MSO Series Up to 5 GS s Sample Rate on All Channels Up to 20 Megapoint Record Length on All Channels gt 90 000 wim s Maximum Waveform Capture Rate Standard Passive Voltage Probes with Less than 4 pF Capacitive Loading and 500 MHz or 1 GHz Analog Bandwidth Suite of Advanced Triggers Ease of Use Features Wave Inspector Controls provide Easy Navigation and Automated search of Waveform Data 41 Automated Measurements Waveform Histograms and FFT Analysis for Simplified Waveform Analysis TekVPI Probe Interface supports Active Differential and Current Probes for Automatic Scaling and Units 10 4 in 264 mm Bright XGA Color Display Small Footprint and Lightweight Only 5 8 in 147 mm deep and 11 Ib 9 kg Connectivity Two USB 2 0 Host Ports on the Front Panel and Two on the Rear Panel for Quick and Easy Data Storage Printing and Connecting a USB Keyboard USB 2 0 Device Port on Rear Panel for Easy Connection to a PC or Direct Printing to a PictBridge compatible Printer Integrated 10 100 1000BASE T Ethernet Po
4. Both main and MagniVu waveforms are acquired on every trigger and can be switched between in the display at any time running or stopped MagniVu provides significantly finer timing resolution than comparable MSOs on the market instilling confidence when making critical timing measurements on digital waveforms P6616 MSO Probe This unique probe design offers two eight channel pods Each channel ends with a probe tip featuring a recessed ground for simplified connection to the device under test The coax on the first channel of each pod is colored blue making it easy to identify The common ground uses an automotive style connector making it easy to create custom grounds for connecting to the device under test When connecting to square pins the P6616 has an adapter that attaches to the probe head extending the probe ground flush with the probe tip so you can attach to a header The P6616 offers outstanding electrical characteristics having only 3 pF of capacitive loading a 100 kQ input resistance and capable of acquiring toggle rates gt 500 MHz and pulses as short as 1 ns in duration _ PreVu _ Trigger On sync Out HAd 2H 2 01 sync Datal DO 2E 20 20 D3 20 20 20 at 1 00ys R 50GS s eas 1M points iia poine ina ese ae eet D15 Do Token Packet Timing Resolution 2 00ns i Perin F ss ea ae SSS Bus mee Packet OUT XXh Normal Triggering on a specific OUT Token packet on a USB full speed serial bus The yello
5. performance The MSO DPO4000B Series offers a comprehensive set of integrated analysis tools including waveform and screen based cursors 41 automated measurements advanced waveform math including arbitrary equation editing waveform histograms FFT analysis and trend plots for visually determining how a measurement is changing over time Specialized application support for serial bus analysis power supply design limit and mask testing and video design and development is also available For extended analysis National Instrument s LabVIEW SignalExpress Tektronix Edition provides over 200 built in functions including time and frequency domain analysis data logging and customizable reports Mixed Domain Analysis Working with RF signals Be sure to check out the MDO4000 Series the world s first Mixed Domain Oscilloscope Built on the MSO4000B oscilloscope platform the MDO4000 Series offers a built in spectrum analyzer up to 6 GHz This combination offers you the ability to capture time correlated analog digital and RF signals in a single instrument For more information on the MDO4000 Series please visit www tektronix com mdo4000 www tektronix com 3 Data Sheet oF Oe gt z Thresholds Trigger Mode Type Source Polarity 2 00 V when Normal kun o Positive 1 00 V Runt Occurs amp Holdoff Wave Inspector controls provide unprecedented efficiency in viewing navigating and analyzing waveform data Zip through
6. 1080p 24sF 1080p 25 1080p 30 1080p 50 1080p 60 and custom bi level and tri level sync video standards Trigger on Start Repeated Start Stop Missing ACK Address 7 or 10 bit Data or Address and Data on 12C buses up to 10 Mb s Trigger on SS active MOSI MISO or MOSI and MISO on SPI buses up to 50 Mb s Note Only 2 wire SPI is supported on the DPO4102B and DPO4102B L models 12 www tektronix com Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Mode Description USB Optional Low speed Trigger on Sync Reset Suspend Resume End of Packet Token Address Packet Data Packet Handshake Packet Special Packet Error Token packet trigger Any token type SOF OUT IN SETUP Address can be specified for Any Token OUT IN and SETUP token types Address can be further specified to trigger on lt lt gt 2 a particular value or inside or outside of a range Frame number can be specified for SOF token using binary hex unsigned decimal and don t care digits Data packet trigger Any data type DATAO DATA Data can be further specified to trigger on lt lt gt 2 a particular data value or inside or outside of a range Handshake packet trigger Any handshake type ACK NAK STALL Special packet trigger Any special type Reserved Error trigger PID Check CRC5 or CRC16 Bit Stuffing Full speed Trigger on Sync Reset Suspend Resume End of Packet Token Address Packet Data P
7. Series from Tektronix provides the feature rich tools you need to simplify and speed debug of your complex design Tektronix Data Sheet Ses Pas OT ay l 1000 points Discover Fast waveform capture rate over 50 000 wfm s maximizes the probability of capturing elusive glitches and other infrequent events Comprehensive Features Speed Every Stage of Debug The MSO DPO4000B Series offers a robust set of features to speed every stage of debugging your design from quickly discovering an anomaly and capturing it to searching your waveform record for the event and analyzing its characteristics and your device s behavior Discover To debug a design problem first you must know it exists Every design engineer spends time looking for problems in their design a time consuming and frustrating task without the right debug tools The MSO DPO4000B Series offers the industry s most complete visualization of signals providing fast insight into the real operation of your device A fast waveform capture rate greater than 50 000 waveforms per second enables you to see glitches and other infrequent transients within seconds revealing the true nature of device faults A digital phosphor display with intensity grading shows the history of a signal s activity by intensifying areas of the signal that occur more frequently providing a visual display of just how often anomalies occur 2 www tektronix com Trigger On aN
8. analog channel bandwidth Power Analysis Application Module Enables quick and accurate analysis of power quality switching loss harmonics safe operating area SOA modulation ripple and slew rate dl dt dV dt DPO4ENET DPO4USB DPO4PWR DPO4LMT Limit and Mask Testing Application Module Enables testing against limit templates generated from golden waveforms and mask testing using custom or standard telecommunications or computer masks DPO4VID HDTV and Custom nonstandard Video Triggering Module www tektronix com 17 Data Sheet Instrument Options Recommended Probes Tektronix offers over 100 different probes to meet your application needs For a comprehensive listing of available probes please visit www tektronix com probes Probe Description TPP0500 500 MHz 10X TekVPI passive voltage probe with 3 9 pF input capacitance Power Plug Options Option Description Opt AO North America Opt A1 Universal Euro Opt A2 United Kingdom Opt A3 Australia Opt A5 Switzerland Opt A6 Japan Opt A10 China Opt A11 India Opt A12 Brazil Opt A99 No power cord or AC adapter Language Options 2 Option Description Opt LO English manual Opt L1 French manual Opt L2 Italian manual Opt L3 German manual Opt L4 Spanish manual Opt L5 Japanese manual Opt L6 Portuguese manual Opt L7 Simplified Chinese manual Opt L8 Traditional Chinese manual Opt L9 Korean manual Opt L10 Russian manual Opt L
9. and direction are controlled using the intuitive pan control Once again turning the control further makes the waveform scroll faster and changing direction is as simple as turning the control the other way 4 www tektronix com Search For e SS Number of Bytes Binary 0001 0001 C E I0 0ms7s 7 3 01V Bus Search events found 33 1M points Search Search Type Source Bus Search For Identifier Data on Bus BR CAN Id amp Data OXXh 11h Search step 1 You define what you would like to find Id 013 ED CED TES ECAN CED A 10 0MS s f 3 01V Bus Search events found 33 1M points Search step 2 Wave Inspector automatically searches through the record and marks each event with a hollow white triangle You can then use the Previous and Next buttons to jump from one event to the next User Marks Press the Set Mark front panel button to place one or more marks on the waveform Navigating between marks is as simple as pressing the Previous lt and Next buttons on the front panel Search Marks The Search button allows you to automatically search through your long acquisition looking for user defined events All occurrences of the event are highlighted with search marks and are easily navigated to using the front panel Previous lt and Next buttons Search types include edge pulse width glitch timeout runt logic setup and hold rise fall time parallel bus and I2
10. hardware of the MSO4000B Series will show you a white edge on the display when the system detects multiple transitions White edges indicate that more information is available by zooming in or acquiring at faster sampling rates In most cases zooming in will reveal the pulse that was not viewable with the previous settings If the white edge is still present after zooming in as far as possible this indicates that increasing the sample rate on the next acquisition will reveal higher frequency information than the previous settings could acquire 6 www tektronix com C aD 50 5 00Gs s_ f 1 46 10k points iming Re 60 The P6616 MSO probe offers two eight channel pods to simplify connecting to your device The MSO4000B Series simplifies channel setup by allowing you to group digital waveforms and enter waveform labels by using a USB keyboard By simply placing digital waveforms next to each other they form a group Once a group is formed you can position all the channels contained in that group collectively This greatly reduces the normal setup time associated with positioning channels individually MagniVu High speed Acquisition The main digital acquisition mode on the MSO4000B Series will capture up to 20M points at 500 MS s 2 ns resolution In addition to the main record the MSO4000B Series provides an ultra high resolution record called MagniVu which acquires 10 000 points at up to 16 5 GS s 60 6 ps resolution
11. optional DPO4VID video application module DPO4VID provides the industry s most complete suite of HDTV and custom nonstandard video triggers The MSO DPO4000B Series is designed to make your work easier The large high resolution display shows intricate signal details Dedicated front panel controls simplify operation Two USB host ports on the front panel allow you to easily transfer screenshots instrument settings and waveform data to a USB thumb drive Designed to Make Your Work Easier Large High resolution Display The MSO DPO4000B Series features a 10 4 in 264 mm bright LED backlit XGA color display for seeing intricate signal details Dedicated Front panel Controls Per channel vertical controls provide simple and intuitive operation No longer do you need to share one set of vertical controls across all four channels Connectivity Two USB host ports on the front panel enable easy transfer of screenshots instrument settings and waveform data to a USB thumb drive The rear panel contains two additional USB host ports and a USB device port for controlling the oscilloscope remotely from a PC or for connecting a USB keyboard The USB device port can also be used to print directly to a PictBridge compatible printer An integrated 10 100 1000BASE T Ethernet port enables easy connection to networks and a Video Out port allows the oscilloscope display to be exported to an external monitor or projector The instrument c
12. 2 675 3777 Mexico Central South America amp Caribbean 52 55 56 04 50 90 Middle East Asia and North Africa 41 52 675 3777 The Netherlands 00800 2255 4835 Norway 800 16098 People s Republic of China 400 820 5835 Poland 41 52 675 3777 Portugal 80 08 12370 Republic of Korea 001 800 8255 2835 Russia amp CIS 7 495 7484900 South Africa 41 52 675 3777 Spain 00800 2255 4835 Sweden 00800 2255 4835 Switzerland 00800 2255 4835 Taiwan 886 2 2722 9622 United Kingdom amp Ireland 00800 2255 4835 USA 1 800 833 9200 European toll free number If not accessible call 41 52 675 3777 Updated 10 February 2011 For Further Information Tektronix maintains a comprehensive constantly expanding collection of application notes technical briefs and other resources to help engineers working on the cutting edge of technology Please visit www tektronix com Copyright Tektronix Inc All rights reserved Tektronix products are covered by U S and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specification and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc All other trade names referenced are the service marks trademarks or registered trademarks of their respective companies 09 Feb 2012 3GW 20156 14 Tektronix
13. 2 70 168 Microsoft Internet Explorer D x A Fie Edit View Favorites Tools Help ay Screen Waveform Get amp Send ca ra Capture m Data Capture Settings Prererences O O HAG P ka BS amp Wh B Address http 134 62 70 168 x Bo Select Instrument Google _ visah O o o Bshare gt Ay chek G Translate Wf AutoFill P Aer o Sigin a Tektronix Tektronix com Products E ee i TT MSO4104B Zl Home Welcome MSO4104B Web Enabled User Select Channels en Instrument Information AS e Scope Instrument Model MS04104B Manufacturer Tektronix Network Configuration Serial Number cCQ000016 m n eri Data Description Oscilloscope MSO4104B b par ak r SUPPORT Firmware Version 1 16007 E LA gt z 42 HE Datasheets LEA Gi Manuals Network Informa tion i roere e Driver Download Hostname MS04104B 053CXN local eer en GGaad Help MAC Address 08 00 11 1E 77 FC c gt TCP IP Address 134 62 70 168 Instrument Address String TCPIP 134 62 70 168 INSTR LXI Identification on Instrument OOn Off Save As LXI LXI Information LXI Class c i LXI Version 13 ral Copy to Clipboard pa MS04104B Crosstalk on data lines Modify Note Tektronix OpenChoice Desktop software enables seamless connection between the oscilloscope LXI Tektronix MSO41048 MSO4104B 053CXN 134 62 70 168 Microsoft Done Internet Lo Local in
14. 20M point records by turning the outer pan control 1 Get from the beginning to end in seconds See something of interest and want to see more details Just turn the inner zoom control 2 Wave Inspector Navigation and Search A 20M point record length represents thousands of screens of information The MSO DPO4000B Series enables you to find your event in seconds with Wave Inspector the industry s best tool for navigation and search Wave Inspector offers the following innovative controls Zoom Pan A dedicated two tier front panel control provides intuitive control of both zooming and panning The inner control adjusts the zoom factor or zoom scale turning it clockwise activates zoom and goes to progressively higher zoom factors while turning it counterclockwise results in lower zoom factors and eventually turning zoom off No longer do you need to navigate through multiple menus to adjust your zoom view The outer control pans the zoom box across the waveform to quickly get to the portion of waveform you are interested in The outer control also utilizes force feedback to determine how fast to pan on the waveform The farther you turn the outer control the faster the zoom box moves Pan direction is changed by simply turning the control the other way Play Pause A dedicated Play Pause front panel button scrolls the waveform across the display automatically while you look for anomalies or an event of interest Playback speed
15. 81818 0 1122 33 1122 3344 55 FIF2 F3F4 FSF6 F7 0000 0000 0000 0000 Remote Frame Remote Frame 4568 6C65 7273 DESS CBFA 5D45 AD8 amp C 757 1A55AA55 1597EEA3 13 1122 3344 1122 3344 5566 C1C2 C3C4 B7B6 B4B4 0 2 3 8 8 4 8 1 3 5 7 8 0 0 6 8 2 4 6 8 Ca ERTA 20 0ys 100MS s TZE 1M Em Event table showing decoded Identifier DLC DATA and CRC N every CAN packet in a long acquisition 500000 Bus Decoding Tired of having to visually inspect the waveform to count clocks determine if each bit is a 1 or a 0 combine bits into bytes and determine the hex value Let the oscilloscope do it for you Once you ve set up a bus the MSO DPO4000B Series will decode each packet on the bus and display the value in hex binary decimal USB Ethernet MIL STD 1553 LIN and FlexRay only signed decimal 12S LU RJ TDM only or ASCII USB Ethernet and RS 232 422 485 UART only in the bus waveform Event Table In addition to seeing decoded packet data on the bus waveform itself you can view captured packets in a tabular view much like you would see in a software listing Packets are time stamped and listed consecutively with columns for each component Address Data etc Event table data can be saved to csv format Search Serial triggering is very useful for isolating the event of interest but once you ve captured it and need to analyze the surrounding data what do you do In the past users ha
16. 99 No manual 2 Language options include a translated front panel overlay for the selected language s Service Options Option Description Opt C3 Calibration Service 3 years Opt C5 Calibration Service 5 years Opt D1 Calibration Data Report Opt D3 Calibration Data Report 3 years with Opt C3 Opt D5 Calibration Data Report 5 Years with Opt C5 Opt R5 Repair Service 5 years including warranty TPP0502 500 MHz 2X TekVPI passive voltage probe with 3 9 pF input capacitance TPP0850 2 5 kV 800 MHz 50X TekVPI passive high voltage probe TPP1000 1 GHz 10X TekVPI passive voltage probe with 3 9 pF input capacitance TAP1500 1 5 GHz TekVPI active single ended voltage probe TDP1500 1 5 GHz TekVPI differential voltage probe with 8 5 V differential input voltage TDP1000 1 GHz TekVPI differential voltage probe with 42 V differential input voltage TDP0500 500 MHz TekVPI differential voltage probe with 42 V differential input voltage TCP0150 20 MHz TekVPI 150 Ampere AC DC current probe TCP0030 120 MHz TekVPI 30 Ampere AC DC current probe TMDP0200 750 V 200 MHz high voltage differential probe THDP0200 1 5 kV 200 MHz high voltage differential probe THDP0100 6 kV 100 MHz high voltage differential probe P5100A 2 5 kV 500 MHz 100X high voltage passive probe Recommended Accessories Accessory Description 077 0512 xx Service Manual English only SIGEXPTE NI LabVIEW SignalExpress Tektronix Edition Software Fu
17. C SPI USB Ethernet CAN LIN FlexRay RS 232 422 485 UART MIL STD 1553 and 2S LJ RJ TDM packet content 2 BY 3 eT 4 00ns 5 00GS s in 10k points Digital phosphor technology enables greater than 50 000 wfm s waveform capture rate and real time intensity grading on the MSO DPO4000B Series Digital Phosphor Technology The MSO DPO4000B Series digital phosphor technology provides you with fast insight into the real operation of your device Its fast waveform capture rate greater than 50 000 wfm s gives you a high probability of quickly seeing the infrequent problems common in digital systems runt pulses glitches timing issues and more Waveforms are superimposed with one another and waveform points that occur more frequently are intensified This quickly highlights the events that over time occur more often or in the case of infrequent anomalies occur less often With the MSO DPO4000B Series you can choose infinite persistence or variable persistence including zero persistence determining how long the previous waveform acquisitions stay on screen This allows you to determine how often an anomaly is occurring Accurate High speed Probing The TPP Series probes included standard with every MSO DPO4000B Series oscilloscope provide up to 1 GHz of analog bandwidth and less than 4 pF of capacitive loading The extremely low capacitive loading minimizes adverse affects on your circuits and is more forgiving of lo
18. USB Test Source Limit Test Any Ch1 Ch4 or any R1 R4 Mask Test Any Ch1 Ch4 Limit test vertical tolerance from 0 to 1 division in 1m division increments Limit test horizontal tolerance from 0 to 500 m division in 1 m division increments Load standard mask from internal memory Load custom mask from text file with up to 8 segments Lock to Source ON mask automatically re scales with source channel settings changes Lock to Source OFF mask does not re scale with source channel settings changes Mask Creation Mask Scaling Test Criteria Run Until Minimum number of waveforms from 1 to 1 000 000 Infinity Minimum elapsed time from 1 second to 48 hours Infinity Violation Threshold From 1 to 1 000 000 Actions on Test Failure Stop acquisition save screen image to file save waveform to file print screen image trigger out pulse set remote interface SRQ Actions on Test Complete Trigger out pulse set remote interface SRQ Results Display Test status total waveforms number of violations violation rate total tests failed tests test failure rate elapsed time total hits for each mask segment Software Software Description NI LabVIEW A fully interactive measurement software environment SignalExpress Tektronix optimized for the MSO DPO4000B Series enables Edition you to instantly acquire generate analyze compare import and save measurement data and signals using an intuitive drag and drop user interface t
19. acket Handshake Packet Special Packet Error Token packet trigger Any token type SOF OUT IN SETUP Address can be specified for Any Token OUT IN and SETUP token types Address can be further specified to trigger on lt lt gt 2 a particular value or inside or outside of a range Frame number can be specified for SOF token using binary hex unsigned decimal and don t care digits Data packet trigger Any data type DATAO DATA Data can be further specified to trigger on lt lt gt 2 a particular data value or inside or outside of a range Handshake packet trigger Any handshake type ACK NAK STALL Special packet trigger Any special type PRE Reserved Error trigger PID Check CRC5 or CRC16 Bit Stuffing High speed Trigger on Sync Reset Suspend Resume End of Packet Token Address Packet Data Packet Handshake Packet Special Packet Error Token packet trigger Any token type SOF OUT IN SETUP Address can be specified for Any Token OUT IN and SETUP token types Address can be further specified to trigger on lt lt gt 2 a particular value or inside or outside of a range Frame number can be specified for SOF token using binary hex unsigned decimal and don t care digits Data packet trigger Any data type DATAO DATA1 DATA2 MDATA Data can be further specified to trigger on lt lt gt 2 a particular data value or inside or outside of a range H
20. an mount external network drives for easy storage of screen images setup files or data files Setup or data files can then be directly recalled and loaded into the oscilloscope from the network drive location The MSO DPO4000B Series is LXI Class C compliant Compact Form Factor A compact portable form factor allows the MSO DPO4000B Series to be easily moved between labs and with a depth of just 5 8 in 147 mm it saves you valuable space on your test bench Mixed Signal Oscilloscopes MSO4000B DPO4000B Series The MSO DPO4000B Series compact form factor frees up valuable space on your bench or desktop 00 VRMS SAT IA TekVPI probe interface simplifies connecting your probes to the oscilloscope TekVPI Probe Interface The TekVPI probe interface sets the standard for ease of use in probing In addition to the secure reliable connection that the interface provides TekVPI probes feature status indicators and controls as well as a probe menu button right on the compensation box itself This button brings up a probe menu on the oscilloscope display with all relevant settings and controls for the probe The TekVPI interface enables direct attachment of current probes without requiring a separate power supply TekVPI probes can be controlled remotely through USB GPIB or LAN enabling more versatile solutions in ATE environments www tektronix com 9 Data Sheet LXI Tektronix MSO4104B MSO4104B 053CXN 134 6
21. andshake packet trigger Any handshake type ACK NAK STALL NYET Special packet trigger Any special type ERR SPLIT PING Reserved SPLIT packet components that can be specified include Hub Address Start Complete Don t Care Start SSPLIT Complete CSPLIT Port Address Start and End bits Don t Care Control Bulk Interrupt Full soeed Device Low speed Device Isochronous Data is Middle Data is End Data is Start Data is All Endpoint Type Don t Care Control Isochronous Bulk Interrupt Error trigger PID Check CRC5 or CRC16 High speed support only available on models with 1 GHz analog channel bandwidth Ethernet Optional 10BASE T Trigger on Start Frame Delimiter MAC Addresses MAC Q Tag Control Information MAC Length Type IP Header TCP Header TCP IPv4 MAC Client Data End of Packet FCS CRC Error MAC Addresses Trigger on Source and Destination 48 bit address values MAC Q Tag Control Information Trigger on Q Tag 32 bit value MAC Length Type Trigger on lt lt gt 2 a particular 16 bit value or inside or outside of a range IP Header Trigger on IP Protocol 8 bit value Source Address Destination Address TCP Header Trigger on Source Port Destination Port Sequence Number and Ack Number TCP IPv4 MAC Client Data Trigger on lt lt gt 2 a particular data value or inside or outside of a range Selectable number of bytes to trigger on from 1 16 Byte offs
22. ased remote control All web interaction conforms to LXI Class C specification OpenChoice Desktop IVI Driver e Scope Web based Remote Control LXI Class C Web Interface Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Display Characteristics Description 10 4 in 264 mm liquid crystal TFT color display 1 024 horizontal x 768 vertical pixels XGA Vectors Dots Variable Persistence Infinite Characteristic Display Type Display Resolution Waveform Styles Persistence Graticules Full Grid Solid Cross Hair Frame IRE and mV Format YT and simultaneous XY YT Waveform Capture Rate gt 50 000 wim s maximum Input Output Ports Port Description USB 2 0 High speed Host Supports USB mass storage devices printers and Port keyboard Two ports on front and two ports on rear of instrument Rear panel connector allows for communication control of oscilloscope through USBTMC or GPIB with a TEK USB 488 and direct printing to all PictBridge compatible printers USB 2 0 Device Port LAN Port RJ 45 connector supports 10 100 1000 Mb s XGA Video Port DB 15 female connector connect to show the oscilloscope display on an external monitor or projector Auxiliary Input Front panel BNC connector Input Impedance 1 MQ Max input 300 Vams CAT Il with peaks lt 425 V Probe Compensator Front panel pins Output Amplitude 2 5 V Frequency 1 kHz Auxiliary Out Rear panel BNC connector Vour Hi 22 5 V open circ
23. ation Signal Inputs 12C Any Ch1 Ch4 and any DO D15 on MSO models SPI Any Ch1 Ch4 and any DO D15 on MSO models Recommended Probing 2C SPI Single ended Note Only 2 wire SPI is supported on DPO4102B and DPO4102B L models Ethernet Serial Triggering and Analysis Module Enables triggering on packet level information on 10BASE T and 100BASE TX buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information Signal Inputs Any Ch1 Ch 4 for single ended probing Any Ch1 Ch4 Math Ref1 Ref4 for differential probing Recommended Probing 10BASE T Single ended or differential 1OOBASE TX Differential USB Serial Triggering and Analysis Module Enables triggering on packet level content for low speed full speed and high speed USB serial buses Also enables analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information for low speed full speed and high speed USB serial buses Signal Inputs Low speed and Full speed Any Ch1 Ch4 and any DO D15 on MSO models for single ending probing only Low speed Full speed and High speed Any Ch1 Ch4 Math Ref1 Ref4 Recommended Probing Low speed and Full speed Single ended or differential High speed Differential USB high speed supported only on models with 1 GHz
24. cy Readou DC AC HF reject attenuates gt 50 kHz LF reject attenuates lt 50 kHz noise reject reduces sensitivity 20 ns to 8 s Provides 6 digit frequency readout of triggerable events Description Positive or negative slope on any channel or front panel auxiliary input Coupling includes DC AC HF reject LF reject and noise reject Trigger Delay by Time 4 ns to 8 s Or Trigger Delay by Events 1 to 4 000 000 events Trigger on width of positive or negative pulses that are gt lt or a specified period of time 4 ns to 8 s Trigger when no pulse is detected within a specified time 4 ns to 8 s Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again Trigger when any logical pattern of channels goes false or stays true for specified period of time 4 ns to 8 s Any input can be used as a clock to look for the pattern on a clock edge Pattern AND OR NAND NOR specified for all analog and digital input channels defined as High Low or Don t Care Trigger on violations of both setup time and hold time between clock and data present on one or more of the input channels Trigger on pulse edge rates that are faster or slower than specified Slope may be positive negative or either Trigger on all lines odd even or all fields on NTSC PAL and SECAM video signals Trigger on 480p 60 576p 50 720p 30 720p 50 720p 60 8751 60 1080i 50 1080i 60 1080p 24
25. d Additionally custom masks can be created and used for characterizing signals Tailor a test to your specific requirements by defining test duration in number of waveforms or time a violation threshold that must be met before considering a test a failure counting hits along with statistical information and actions upon violations test failure and test complete Whether specifying a mask from a known good signal or from a custom or standard mask conducting pass fail tests in search of waveform anomalies such as glitches has never been easier Video Design and Development Many video engineers have remained loyal to analog oscilloscopes believing the intensity gradations on an analog display are the only way to see certain video waveform details The MSO DPO4000B Series fast waveform capture rate coupled with its intensity graded view of the signal provides the same information rich display as an analog oscilloscope but with much more detail and all the benefits of digital oscilloscopes Standard features such as IRE and mV graticules holdoff by fields video polarity and an Autoset smart enough to detect video signals make the MSO DPO4000B Series the easiest to use oscilloscope on the market for video applications And with up to a 1 GHz bandwidth and up to four analog inputs the MSO DPO4000B Series provides ample performance for analog and digital video use The MSO DPO4000B Series video functionality is further extended with the
26. d to manually scroll through the waveform counting and converting bits and looking for what caused the event With the MSO DPO4000B Series you can have the oscilloscope automatically search through the acquired data for user defined criteria including serial packet content Each occurrence is highlighted by a search mark Rapid navigation between marks is as simple as pressing the Previous lt and Next buttons on the front panel www tektronix com 7 Data Sheet 360ys 800mv ie 110mA xxy 52 5mW 88 3mW SOA Mask Test Passed 889 Waveforms Acquired Failing Samples 0 2 WAT res 00ms E SOMS s l fF 15 07 100k points Safe operating area measurement Automated power measurements enable quick and accurate analysis of common power parameters Show Results Limit Mask Test Results Current Totals Test Status Failed n a Total Waveforms 39 669k 39 669k Violations 1 1 oes Hits ae 100ns EF 00GS7S f 2 06 A 10k points Show Results a mg um off Limit Test saan a mask created from a i waveform and compared against a live signal Results showing statistical information about the test are displayed Power Analysis Optional Ever increasing consumer demand for longer battery life devices and for green solutions that consume less power require power supply designers to characterize and minimize switching losses to improve efficiency In addition the supply s power levels output p
27. dustry s broadest range of serial buses 12C SPI USB Ethernet CAN LIN FlexRay RS 232 422 485 UART MIL STD 1553 and I2S LJU RJ TDM The ability to decode up to four serial and or parallel buses simultaneously means you gain insight into system level problems quickly To further help troubleshoot system level interactions in complex embedded systems the MSO4000B Series offers 16 digital channels in addition to its analog channels Since the digital channels are fully integrated into the oscilloscope you can trigger across all input channels automatically time correlating all analog digital and serial signals The MagniVu high speed acquisition enables you to acquire fine signal detail up to 60 6 ps resolution around the trigger point for precision measurements MagniVu is essential for making accurate timing measurements for setup and hold clock delay signal skew and glitch characterization Ea TTT mT TT Tear Zoominhactorc100 k St pe et Se a a a a a rr oae a a a a a aa a ea a a a a a a a a at Bus Search events found 118 Search 2C decode showing results from a Wave Inspector search for Address value 50 Wave Inspector controls provide unprecedented efficiency in viewing and navigating waveform data Search Finding your event of interest in a long waveform record can be time consuming without the right search tools With today s record lengths pushing beyond a million data points locating your event can
28. e Positive Pulse Width Negative Pulse Width Burst Width Phase Positive Overshoot Negative Overshoot Peak to Peak Amplitude High Low Max Min Mean Cycle Mean RMS Cycle RMS Positive Pulse Count Negative Pulse Count Rising Edge Count Falling Edge Count Area and Cycle Area Mean Min Max Standard Deviation User definable reference levels for automatic measurements can be specified in either percent or units Isolate the specific occurrence within an acquisition to take measurements on using either the screen or waveform cursors A waveform histogram provides an array of data values representing the total number of hits inside of a user defined region of the display A waveform histogram is both a visual graph of the hit distribution as well as a numeric array of values that can be measured Sources Channel 1 Channel 2 Channel 3 Channel 4 Ref 1 Ref 2 Ref 3 Ref 4 Math Types Vertical Horizontal Waveform Count Hits in Box Peak Hits Median Max Min Peak to Peak Mean Standard Deviation Sigma 1 Sigma 2 Sigma 3 14 www tektronix com Waveform Math Characteristic Arithmetic Math Functions FFT Advanced Math Description Add subtract multiply and divide waveforms Integrate Differentiate FFT Spectral magnitude FFT Vertical Scale Linear RMS or dBV RMS FFT Window Settings Rectangular Hamming Hanning Blackman Harris Define extensive algebraic expressions including waveform
29. escription DPO4AERO Aerospace Serial Triggering and Analysis Module Enables triggering on packet level information on MIL STD 1553 buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information Signal Inputs Any Ch1 Ch4 Math Ref1 Ref4 Recommended Probing Differential or single ended only one single ended signal required Audio Serial Triggering and Analysis Module Enables triggering on packet level information on 12S LJ RJ and TDM audio buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information Signal Inputs Any Ch1 Ch4 and any DO D15 on MSO models Recommended Probing 12S LJ RJ TDM Single ended Automotive Serial Triggering and Analysis Module Enables triggering on packet level information on CAN and LIN buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information Signal Inputs LIN Any Ch1 Ch4 and any DO D15 on MSO models CAN Any Ch1 Ch4 and any DO D15 on MSO models single ended probing only Recommended Probing LIN Single ended CAN single ended or differential DPO4AUTOMAX Extended Automotive Serial Triggering and Analysis Module Enables triggering on packet
30. et options of Don t Care 0 1499 100BASE TX Trigger on Start Frame Delimiter MAC Addresses MAC Q Tag Control Information MAC Length Type IP Header TCP Header TCP IPv4 MAC Client Data End of Packet FCS CRC Error Idle MAC Addresses Trigger on Source and Destination 48 bit address values MAC Q Tag Control Information Trigger on Q Tag 32 bit value MAC Length Type Trigger on lt lt gt 2 a particular 16 bit value or inside or outside of a range IP Header Trigger on IP Protocol 8 bit value Source Address Destination Address TCP Header Trigger on Source Port Destination Port Sequence Number and Ack Number TCP IPv4 MAC Client Data Trigger on lt lt gt 2 a particular data value or inside or outside of a range Selectable number of bytes to trigger on from 1 16 Byte offset options of Don t Care 0 1499 CAN Optional Trigger on Start of Frame Frame Type data remote error overload Identifier standard or extended Data Identifier and Data End of Frame Missing ACK or Bit Stuffing Error on CAN signals up to 1 Mb s Data can be further specified to trigger on lt lt gt 2 or a specific data value User adjustable sample point is set to 50 by default LIN Optional Trigger on Sync Identifier Data Identifier and Data Wakeup Frame Sleep Frame Errors such as Sync Parity or Checksum Errors up to 100 Kb s by LIN definition 20 Kb s FlexRay Optional Trigge
31. hat does not require any programming Standard MSO DPO4000B Series support for acquiring controlling viewing and exporting your live signal data is permanently available through the software The full version SIGEXPTE adds additional signal processing advanced analysis mixed signal sweeping limit testing and user defined step Capabilities and is available for a 30 day trial period standard with each instrument Enables fast and easy communication between a Windows PC and the MSO DPO4000B Series using USB or LAN Transfer and save settings waveforms measurements and screen images Included Word and Excel toolbars automate the transfer of acquisition data and screen images from the oscilloscope into Word and Excel for quick reporting or further analysis Provides a standard instrument programming interface for common applications such as LabVIEW LabWindows CVI Microsoft NET and MATLAB Enables control of the MSO DPO4000B Series over a network connection through a standard web browser Simply enter the IP address or network name of the oscilloscope and a web page will be served to the browser Connect to the MSO DPO4000B Series through a standard web browser by simply entering the oscilloscope s IP address or network name in the address bar of the browser The web interface enables viewing of instrument status and configuration status and modification of network settings and instrument control through the e scope web b
32. he color coded digital waveform display groups are created by simply placing digital channels together on the screen allowing the digital channels to be moved as a group You can set threshold values for each channel enabling support for up to 16 different logic families oe Vi E Mixed Signal Design and Analysis MSO Series The MSO4000B Series Mixed Signal Oscilloscopes provide 16 digital channels These channels are tightly integrated into the oscilloscopes user interface simplifying operation and making it possible to solve mixed signal issues easily Color coded Digital Waveform Display The MSO4000B Series has redefined the way you view digital waveforms One common problem shared by both logic analyzers and mixed signal oscilloscopes is determining if data is a one or a zero when zoomed in far enough that the digital trace stays flat all the way across the display The MSO4000B Series has color coded digital traces displaying ones in green and zeros in blue www tektronix com 5 Data Sheet D15 D0 gt f Timing Resolution 2 00ns J White edges indicate additional information is available by zooming in As shown here Zooming in on the white edge reveals a hidden glitch Tek D15 D0 f Timing Resolution 60 6ps The MagniVu high resolution record provides 60 6 ps timing resolution enabling you to make critical timing measurements on your digital waveforms The multiple transition detection
33. level information on CAN LIN and FlexRay buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools packet decode tables with time stamp information and eye diagram analysis software Signal Inputs LIN Any Ch1 Ch4 and any DO D15 on MSO models CAN Any Ch1 Ch4 and any DO D15 on MSO models single ended probing only FlexRay Any Ch1 Ch4 and any DO D15 on MSO models single ended probing only Recommended Probing LIN Single ended CAN FlexRay Single ended or differential Computer Serial Triggering and Analysis Module Enables triggering on packet level information on RS 232 422 485 UART buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp information Signal Inputs Any Ch1 Ch4 and any DO D15 on MSO models single ending probing only Recommended Probing RS 232 UART Single ended RS 422 485 Differential DPO4AUDIO not available on DPO4102B or DPO4102B L models DPO4AUTO DPO4COMP Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Module DPO4EMBD Description Embedded Serial Triggering and Analysis Module Enables triggering on packet level information on 2C and SPI buses as well as analytical tools such as digital views of the signal bus views packet decoding search tools and packet decode tables with time stamp inform
34. ll Version FPGAView xx MSO Support for Altera and Xilinx FPGAs TPA BNC TekVPI to TekProbe BNC Adapter TEK USB 488 GPIB to USB Adapter ACD4000B Soft Transit Case HCTEK54 Hard Transit Case Requires ACD4000B RMD5000 Rackmount Kit TEK DPG Deskew Pulse Generator 067 1686 xx Deskew Fixture Warranty Three year warranty covering all parts and labor excluding probes 3 Probes and accessories are not covered by the oscilloscope warranty and service offerings Refer to the datasheet of each probe and accessory model for its unique warranty and calibration terms 18 www tektronix com Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar GPIB Product s complies with IEEE Standard 488 1 1987 RS 232 C and with Tektronix IEEE 488 Standard Codes and Formats Mixed Signal Oscilloscopes MSO4000B DPO4000B Series www tektronix com 19 Data Sheet www tektronix com Contact Tektronix ASEAN Australasia 65 6356 3900 Austria 00800 2255 4835 Balkans Israel South Africa and other ISE Countries 41 52 675 3777 Belgium 00800 2255 4835 Brazil 55 11 3759 7627 Canada 1 800 833 9200 Central East Europe and the Baltics 41 52 675 3777 Central Europe amp Greece 41 52 675 3777 Denmark 45 80 88 1401 Finland 41 52 675 3777 France 00800 2255 4835 Germany 00800 2255 4835 Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835 Japan 81 3 6714 3010 Luxembourg 41 5
35. mean scrolling through thousands of screens of signal activity The MSO DPO4000B Series offers the industry s most comprehensive search and waveform navigation with its innovative Wave Inspector controls These controls speed panning and zooming through your record With a unique force feedback system you can move from one end of your record to the other in just seconds User marks allow you to mark any location that you may want to reference later for further investigation Or automatically search your record for criteria you define Wave Inspector will instantly search your entire record including analog digital and serial bus data Along the way it will automatically mark every occurrence of your defined event so you can quickly move between events Analyze Verifying that your prototype s performance matches simulations and meets the project s design goals requires analyzing its behavior Tasks can range from simple checks of rise times and pulse widths to sophisticated power loss analysis and investigation of noise sources Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Value Mean Min Max Std Dev Tio ons Ws o0cs7s ep 7 20v I 0k points Analyze Waveform histogram of a falling edge showing the distribution of edge position jitter over time Included are numeric measurements made on the waveform histogram data A comprehensive set of integrated analysis tools speeds verification of your design s
36. nger ground leads And since the probe bandwidth matches your oscilloscope bandwidth you can see the high frequency components in your signal which is critical for high speed applications The TPP Series passive voltage probes offer all the benefits of general purpose probes like high dynamic range flexible connection options and robust mechanical design while providing the performance of active probes In addition a low attenuation 2X version of the TPP probes is available for measuring low voltages Unlike other low attenuation passive probes the TPP0502 has high bandwidth 500 MHz as well as low capacitive loading 12 7 pF Mixed Signal Oscilloscopes MSO4000B DPO4000B Series 1420MH2_CLOCK nt Tahi itn iw Faia a a da hha aha gaa thas aa a a a 2 STROBE F COUNTS F COUNTS F counta Ey COUNTS BICIE L couNTI fy COUNTO o CLOCK Parallel a a a T a a Ty A ST SS ST YS ST a a a 5 00 V D15 D0 Timing Resolution 2 00ns_ gt The MSO Series provides 16 integrated digital channels enabling you to view and analyze time correlated analog and digital signals K Sa Se Digital Channel PD 7D 0 Vertical Position 560mdiv Select 1 600 V _ eeu 1 400 V 1 400 V Choose Preset 1 600 V on 1 800 V 2 000 V 2 200 V 2 400 V yJ AEEGREEEE kez various 1 00GS s 10k points Timing Resoluti D15 D0 Edit MagniVu On Off Thresholds Labels On Off With t
37. o 50 mV div 1 V 1 V Input Channels 16 Digital D15 DO 50 5 mV div to 0 5 V 0 5 V Thresholds Per channel Thresholds 99 5 mVidiv Threshold Selections TTL CMOS ECL PECL User Defined 100 mV div to 500 mVidv i SOV User defined Threshold 40 V 505 mV div to 995 mV div 5 V 5 V Range 1 Vidiv to 5 V div 100 V 5 V Maximum Input Voltage 42 Voeax 5 05 V div to 10 V div 50 V NA Threshold Accuracy 100 mV 3 of threshold setting Input Dynamic Range 30 Vp p 200 MHz 10 Vp p gt 200 MHz Minimum Voltage Swing 400 mV Input Impedance 100 kQ Probe Loading 3 pF Vertical Resolution 1 bit Horizontal System Analog Channels Characteristic DPO4034B DPO4054B DPO4102B DPO4104B DPO4102B L DPO4104B L MSO04034B MSO04054B MSO04102B MSO4104B MS0O4102B L MS0O4102B L Maximum Sample Rate 2 0 GS s 2 5 GS s 5 GS s 2 5 GS s All channels Maximum Record Length 20M points 5M points All channels Maximum Duration at Highest 8 ms 8 ms 4 ms 2 ms Sample Rate All channels Time Base Range 1 ns to 1 000 s 400 ps to 1 000 s 400 ps to 1 000 s Time Base Delay Time Range 10 divisions to 5000 s Channel to Channel Deskew 125 ns Range Time Base Accuracy 5 ppm over any 21 ms interval www tektronix com 11 Data Sheet Horizontal System Digital Channels Characteristic Trigger Sensitivity All MSO4000B Models Characteristic Description a Sample Rate 500 MS s 2 ns resolution Internal DC Coupled Maximum Record Length 20M points 5M points
38. on L models 1 MQ Path For 1 mV div to 4 98 mVidiv 0 75 div from DC to Main All models 50 MHz increasing to 1 3 div at rated bandwidth eae Bale IO 3 Ol DE ESEUN 50 Q Path For 25 mVidiv 0 4 div from DC to 50 MHz 9 350 MHz and increasing to 1 div at rated bandwidth Maximum Record Length 10k points centered around the trigger 500 MHz models MagniVu ni 50 Q Path 0 4 div from DC to 50 MHz increasing to 1 div at Minimum Detectable 1 ns Pulse Width Typical 1 GHz models rated bandwidth Channel to Channel 200 ps External et Typical MH Auxiliary Input 200 mV from DC to 50 MHZ increasing to 500 mV He Input Toggle 500 MHz at rated bandwidth Maximum frequency sine wave that can accurately be reproduced as a logic square wave Requires the use Trigger Level Range of a short ground extender on each channel Characteristic Description This is the maximum frequency at the minimum swing Any Channel 8 divisions from center of screen amplitude Higher toggle rates can be achieved with External 48V higher amplitudes KEMA Auxiliary Input Trigger System Line Fixed at about 50 of line voltage Characteristic Main Trigger Modes Trigger Coupling Trigger Holdoff Range Trigger Modes Mode Edge Sequence B trigger Pulse Width Timeout Runt Logic Setup and Hold Rise Fall Time Video Extended Video Optional I2C Optional SPI Optional Description Tri F ency Readout Auto Normal and Single rigger Frequen
39. r control instrument settings save screen images and instrument data and 200 built in functions that provide additional signal processing advanced savel load instrument setups of your MSO DPO4000B Series oscilloscope analysis sweeping and user defined step capabilities directly from the web interface through a password protected web page 10 www tektronix com Mixed Signal Oscilloscopes MSO4000B DPO4000B Series Characteristics Vertical System Analog Channels Characteristic DPO4034B DPO4054B DPO4102B DPO4102B L DPO4104B DP0O4104B L MSO04034B MSO4054B MSO04102B MSO4102B L _MS0O4104B MSO4104B L Input Channels 4 2 4 Analog Bandwidth 3 dB 350 MHz 500 MHz 1 GHz 1 GHz 5 mV div 1 V div Calculated Rise Time 5 mV div 1 ns 700 ps 350 ps 350 ps typical Hardware Bandwidth Limits 20 MHz or 250 MHz Input Coupling AC DC Input Impedance 1 MQ 1 50 Q 1 Input Sensitivity 1 MQ 1 mV div to 10 V div Input Sensitivity 50 Q 1 mV div to 1 V div Vertical Resolution 8 bits 11 bits with Hi Res Max Input Voltage 1 MQ 300 Vrus CAT II with peaks lt 425 V Max Input Voltage 50 Q 5 Vrms With peaks lt 20 V DC Gain Accuracy 1 5 derated at 0 10 C above 30 C Channel to Channel Isolation 2100 1 at lt 100 MHz and 230 1 at gt 100 MHZ up to the rated bandwidth for any two channels having equal volts div settings Offset Range Vertical System Digital Channels Range 1 MQ 50 Q Characteristic All MSO4000B Models 1 mV div t
40. r on Start of Frame Type of Frame Normal Payload Null Sync Startup Identifier Cycle Count Complete Header Field Data Identifier and Data End of Frame or Errors such as Header CRC Trailer CRC Null Frame Sync Frame or Startup Frame Errors up to 100 Mb s RS 232 422 485 UART Trigger on Tx Start Bit Rx Start Bit Tx End of Packet Rx End of Packet Tx Data Rx Data Tx Parity Error and Rx Parity Error up to 10 Mb s Optional www tektronix com 13 Data Sheet Mode MIL STD 1553 Optional I2S LJ RJ TDM Optional not available on DPO4102B or DPO4102B L models Parallel Available on MSO models only Description Trigger on Sync Word Type t Command Status Data Command Word set RT Address T R Sub address Mode Data Word Count Mode Code and Parity individually Status Word set RT Address Message Error Instrumentation Service Request Bit Broadcast Command Received Busy Subsystem Flag Dynamic Bus Control Acceptance DBCA Terminal Flag and Parity individually Data Word user specified 16 bit data value Error Sync Parity Manchester Non contiguous data Idle Time minimum time selectable from 4 us to 100 us maximum time selectable from 12 us to 100 us trigger on lt minimum gt maximum inside range outside range RT Address can be further specified to trigger on lt gt lt 2 a particular value or inside or outside of a range Trigger on Word Select Frame Sync or Da
41. rt for Network Connection and Video Out Port to Export the Oscilloscope Display to a Monitor or Projector Optional Serial Triggering and Analysis Automated Serial Triggering Decode and Search Options for 12C SPI USB Ethernet CAN LIN FlexRay RS 232 422 485 UART MIL STD 1553 and I2S LJ RJ TDM Mixed Signal Design and Analysis MSO Series Automated Triggering Decode and Search on Parallel Buses Per channel Threshold Settings Multichannel Setup and Hold Triggering MagniVu High speed Acquisition Provides 60 6 ps Fine Timing Resolution on Digital Channels Optional Application Support Power Analysis Limit and Mask Testing HDTV and Custom Video Analysis Feature rich Tools for Debugging Mixed Signal Designs With the MSO DPO4000B Mixed Signal Oscilloscope Series you can analyze up to 20 analog and digital signals with a single instrument to quickly find and diagnose problems in complex designs Bandwidths up to 1 GHz and up to 5X oversampling on all channels ensure you have the performance you need to see fast changing signal details To capture long windows of signal activity while maintaining fine timing resolution the MSO DPO4000B Series offers deep record length of up to 20M points standard on all channels With Wave Inspector controls for rapid waveform navigation automated serial and parallel bus analysis limit and mask testing and automated power analysis the MSO DPO4000B Oscilloscope
42. s reference waveforms math functions Perform math on math using complex equations FFT Intg Diff Log Exp Sart Abs Sine Cosine Tangent Rad Deg scalars up to two user adjustable variables and results of parametric measurements Period Freq Delay Rise Fall PosWidth NegWidth BurstWidth Phase PosDutyCycle NegDutyCycle PosOverShoot NegOverShoot PeakPeak Amplitude RMS CycleRMS High Low Max Min Mean CycleMean Area CycleArea and trend plots e g Intg Ch1 Mean Ch1 x1 414xVAR1 Power Measurements Optional Measurement Power Quality Measurements Switching Loss Measurements Harmonics Ripple Measurements Modulation Analysis Safe Operating Area dV dt and di dt Measurements Description Ves V crest Factors Frequency lau crest Factor True Power Apparent Power Reactive Power Power Factor Phase Angle Power Loss Tan Tor Conduction Total Energy Loss Ton Tor Conduction Total THD F THD R RMS measurements Graphical and table displays of harmonics Test to IEC61000 3 2 Class A and MIL STD 1399 Section 300A V ripple and lripple Graphical display of Pulse Width Pulse Width Period Frequency Duty Cycle and Duty Cycle modulation types Graphical display and mask testing of switching device safe operating area measurements Cursor measurements of slew rate Limit Mask Testing Optional Characteristic Description Included Standard Masks ITU T ANSI T1 102
43. ta Data can be further specified to trigger on a range Maximum data rate for 2S LJ RJ is 12 5 Mb s Maximum data rate for TDM is 25 Mb s lt gt 2 a specific data value or inside or outside of Trigger on a parallel bus data value Parallel bus can be from 1 to 20 bits in size Binary and Hex radices are supported 1 Trigger selection of Command Word will trigger on Command and ambiguous Command Status words Trigger selection of Status Word will trigger on Status and ambiguous Command Status words Acquisition Modes Mode Sample Peak Detect Averaging Envelope Hi Res Roll Description Acquire sampled values Captures glitches as narrow as 800 ps 1 GHz models or 1 6 ns 500 MHz and 350 MHz models at all sweep speeds From 2 to 512 waveforms included in average Min Max envelope reflecting Peak Detect data over multiple acquisitions Real time boxcar averaging reduces random noise and increases vertical resolution Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms div Waveform Measurements Measurement Cursors Automatic Measurements Measurement Statistics Reference Levels Gating Waveform Histogram Waveform Histogram Measurements Description Waveform and Screen 29 of which up to eight can be displayed on screen at any one time Measurements include Period Frequency Delay Rise Time Fall Time Positive Duty Cycle Negative Duty Cycl
44. tranet ane our The LXI web interface provides access to network settings enables remote instrument control and is accessible from any standard web browser Extended Analysis Acquiring data and measurements from the MSO DPO4000B Series is as For simple tasks the included OpenChoice Desktop enables fast and easy simple as connecting a USB cable from the oscilloscope to your PC Key communication between the oscilloscope and your PC through USB or LAN software applications NI LabVIEW SignalExpress Tektronix Edition for transferring settings waveforms and screen images OpenChoice Desktop and Microsoft Excel and Word toolbars are The MSO DPO4000B Series can also be connected to your network using included standard with each oscilloscope to enable fast and easy direct the LAN port The included LXI web interface provides information about communication with your Windows PC the current configuration of your MSO DPO4000B Series oscilloscope NI LabVIEW SignalExpress Tektronix Edition enables you to instantly including network configuration The LXI web interface also provides remote acquire generate analyze compare import and save measurement data instrument control through the popular e Scope web based instrument and signals using an intuitive drag and drop user interface that does not control capability You can make changes to the network configuration require any programming The optional Professional Version offers ove
45. uit 21 0 V 50 Q to ground Vout Lo 0 7 V into a load of lt 4 mA 0 25 V 50 Q to ground Output can be configured to provide a pulse out signal when the oscilloscope triggers the internal oscilloscope reference clock out or an event out for limit mask testing Time base system can phase lock to an external 10 MHz reference 10 MHz 1 Rear panel security slot connects to standard Kensington lock Standard MIS D 100 100 mm VESA mounting points on rear of instrument External Reference In Kensington Lock VESA Mount LAN eXtensions for Instrumentation LXI Characteristic Description Class LXI Class C Version V1 3 Power Source Characteristic Description Power Source Voltage 100 to 240 V 10 Power Source Frequency 45 to 66 Hz 85 to 264 V 360 to 440 Hz 100 to 132 V Power Consumption 225 W maximum www tektronix com 15 Data Sheet Physical Characteristics Dimensions mm in Height 229 9 0 Width 439 17 3 Depth 147 5 8 Weight kg Ib Net 5 11 Shipping 10 7 23 6 Rackmount Configuration 5U Cooling Clearance 2 in 51 mm required on left side and rear of instrument Environmental Characteristic Description Temperature Operating 0 C to 50 C Nonoperating 20 C to 60 C Humidity Operating High 40 C to 50 C 10 to 60 Relative Humidity Low 0 C to 40 C 10 to 90 Relative Humidity Nonoperating High 40 C to 60 C 5 to 60 Relative Humidity Low 0 C to 40 C 5 to 90 Relati
46. urity and harmonic feedback into the power line must be characterized to comply with national and regional power quality standards Historically making these and many other power measurements on an oscilloscope has been a long manual 8 www tektronix com and tedious process The MSO DPO4000B Series optional power analysis tools greatly simplify these tasks enabling quick and accurate analysis of power quality switching loss harmonics safe operating area SOA modulation ripple and slew rate di dt dv dt Completely integrated into the oscilloscope the power analysis tools provide automated repeatable power measurements with a touch of a button no external PC or complex software setup Is required Limit Mask Testing Optional A common task during the development process is characterizing the behavior of certain signals in a system One method called limit testing is to compare a tested signal to a known good or golden version of the same signal with user defined vertical and horizontal tolerances Another common method called mask testing is to compare a tested signal to a mask looking for where a signal under test violates the mask The MSO DPO4000B Series offers both limit and mask testing capability useful for long term signal monitoring characterizing signals during design or testing on a production line A robust set of telecommunications and computer standards are provided to test for compliance to a standar
47. ve Humidity Altitude Operating 9 843 ft 3 000 m Nonoperating 39 370 ft 12 000 m Regulatory Electromagnetic EC Council Directive 2004 108 EC Compatibility Safety UL61010 1 Second Edition CSA61010 1 Second Edition EN61010 1 2001 IEC 61010 1 2001 16 www tektronix com Ordering Information MSO DPO4000B Family Model Description DPO4000B Models DPO4034B 350 MHZ 2 5 2 5 2 5 GS s on 1 2 4 channels 20M record length 4 channel digital phosphor oscilloscope DPO4054B 500 MHz 2 5 2 5 2 5 GS s on 1 2 4 channels 20M record length 4 channel digital phosphor oscilloscope DPO4102B L 1 GHz 5 2 5 GS s on 1 2 channels 5M record length 2 channel digital phosphor oscilloscope DPO4102B 1 GHz 5 5 GS s on 1 2 channels 20M record length 2 channel digital phosphor oscilloscope DPO4104B L 1 GHz 5 5 2 5 GS s on 1 2 4 channels 5M record length 4 channel digital phosphor oscilloscope DPO4104B 1 GHz 5 5 5 GS s on 1 2 4 channels 20M record length 4 channel digital phosphor oscilloscope MSO4000B Models MS04034B 350 MHz 2 5 2 5 2 5 GS s on 1 2 4 channels 20M record length 4 16 channel mixed signal oscilloscope MS04054B 500 MHz 2 5 2 5 2 5 GS s on 1 2 4 channels 20M record length 4 16 channel mixed signal oscilloscope MS04102B L 1 GHz 5 2 5 GS s on 1 2 channels 5M record length 2 16 channel mixed signal oscilloscope MS04102B 1 GHz 5 5 GS s on 1 2 channels 20M record length 2 16 channel mixed signal oscilloscope
48. w waveform is the D and the blue waveform is the D A bus waveform provides decoded packet content including Start Sync PID Address End Point CRC Data values and Stop Serial Triggering and Analysis Optional On a serial bus a single signal often includes address control data and clock information This can make isolating events of interest difficult The MSO DPO4000B Series offers a robust set of tools for debugging serial buses with automatic trigger decode and search for 12C SPI USB Ethernet CAN LIN FlexRay RS 232 422 485 UART MIL STD 1553 and I2S LJU RJ TDM Serial Triggering Trigger on packet content such as start of packet specific addresses specific data content unique identifiers etc on popular serial interfaces such as 2C SPI USB Ethernet CAN LIN FlexRay RS 232 422 485 UART MIL STD 1553 and I2S LJ RJ TDM Bus Display Provides a higher level combined view of the individual signals clock data chip enable etc that make up your bus making it easy to identify where packets begin and end and identifying subpacket components such as address data identifier CRC etc Mixed Signal Oscilloscopes MSO4000B DPO4000B Series M 1 00ms p Vertical iv Zoom Factor 50 X Time Identifier DLC Data Event Table 0103 1122 3344 55 1122 3344 5566 7788 FFFF 0000 EEEE 1111 4269 6C6C AE4F FFF1 0272 DF6B 11 10000001 12345678 1597EEB2 1597EEB2 327DE32 140014 160016 181
Download Pdf Manuals
Related Search
Related Contents
Hygiène des mains Guide de bonnes pratiques - CLIN Paris-Nord PSR8 regulator manual EOS System Message Guide Brinkmann Candle Glow Garden Lanterns User's Manual Protocol - Hussmann 平成 22 年度 第 6 回 宇多野病院治験審査委員会 会議の記録の概要 日 Manual Allview P4 Life Analizador de gas inalámbrico Copyright © All rights reserved.