User Manual washwell PLATE - 5.31.DOC

image

Contents

1. 13 14 4 2 INSTRUMENT TRANSPORTATION uu ccccesesseesenseecccesesesesnsneeceeesesesnsnsceeceeeensneeens 14 4 3 STORAGE OF THE 14 5 INSTRUMENT DESCRI nnmnnn 15 5 1 PERSPECTIVE 16 5 2 REAR PANEL sa sega ga naga anana adana a isane a M na abadan genge a ma k mmi 16 5 3 TUBING DIAGRAM aire disefgexad vedono SUN CX Ra cred ua riu dur a kd Dg an 17 5 4 18 5 5 MICRO PLATE 18 5 6 PRIMING AND RINSING AREAS eene nnn nnne aa ua rna 19 5 7 19 6 INSTALLATION AND START UP INSTRUCTIONG 20 6 1 PLACING 21 6 2 POWER SUPPLE aaa a sa cea cen esed ep uv cia o dE Duke ai sai aped a EH aaa Roa EE 21 6 3 PROTECTIVE GROUNDING cawenanewunnnnnunun un uwnn ouu aua aw DNO ue du V a e 21 6 4 START UP INSTRUCTIONS 1 21 6 5 SETTING DATE AND TIME gau ai dit cf aa and Ge ck ea ed Ck d d e CE nia a 22 6 6 WASTE BOTTLE CONNECTIONS eene nnne nnam nans ananas asas asas aannn 22 6 7 22 7 PRECAUTION S pari die RE anxii gu nana nuda aC QU a eaaa aaa aaan daaarna daa an 23 7 1 5 24 Version 5 31 220 Effective Date Nov
2. Press YES Press YES Key to Enter Name Name Tests Y N Test Number will be printed on the paper Test Saves without name ABCDEFGHIJKL Select Arrow Keys to select Alphabets Yes Press Enter after selection and proceed Press Enter twice after completing i selection Instrument aspirates the waste solution from the well the programmed volume to the micro strip well Aspiration is continuously kept ON to prevent the over flow After completing dispensing cycle depending on soak time it starts aspirating the solution After this the instrument transports the micro plate back to home position Version 5 31 30 40 Effective Date Nov 2011 washwell PLATE 9 2 Strip Wash Display How to operate NO Soak Time Y N Enter soak time in sec should be min 3 to max 3540 secs and Press ENTER No of Cycles nee Enter no of wash cycles required min 1 max 9 and press ENTER WASH VOLUME 50 Y N UU 777 Select required volume of wash buffer 50 to 500 ul Soak Time in Seconds Select the type of the well to be washed Enter number of strips to be washed min 1 8 max 12 8 press ENTER Load Strips Plate and press V ENTER p Press FUNC Key Load Strips Plate and press To Exit ENTER Press Clear Twice to exit the Test To Save Test Press FUNC Key Press YES Press YES Key to Enter Name
3. gt An advisory that the electromagnetic environment should be evaluated prior to operation of the device gt Warning Do not use this device in close proximity to sources of strong electromagnetic radiation e g unshielded international RF sources as these may interfere with the proper operation Version 5 31 9 AO Effective Date Nov 2011 washwell PLATE 3 INTRODUCTION 3 1 Description 3 2 Special Features 3 3 Specifications Version 5 31 10 40 Effective Date Nov 2011 washwell PLATE 3 1 Description washwell PLATE is a versatile user friendly amp rugged instrument designed by keeping both guality and price in view The instrument is simple yet versatile as it uses the state of art hardware for its components It is intended for in vitro diagnostic use 3 2 Special Features e Built in two plate ELISA incubator e Userprogrammable soak time wash cycles and dispensing volume e Battery backed up memory for 35 tests e Waste bottle full indicator with audible alarm e Built in stabilizer e State of art equipment with versatile software e Realtime clock e Continuous aspiration to prevent overflow e Loosely held Manifold to prevent scratching of well bottoms e Uses special software to aspirate U shaped V shaped and wide wells e On switching OFF the instrument it performs RINSE operation first and then instrument automatically gets off Version 5 31
4. 1 3 Disposal instruction In case of removal or disposal of instrument following instructions need to be followed Do not dispose in municipal waste follow local regulations for instrument disposal Plastic parts Electronic PCBs and components can be recycled so return back the instrument to manufacturer 1 4 Contacts Manufacturer European Authorized Representative Obelis s a 34 Av de Tervuren bte 44 ROBONIK INDIA PVT LTD A 374 TTC MIDC Industrial Area MAHAPE NAVI MUMBAI 400710 INDIA B 1040 Brussels BELGIUM Tel 32 2 732 59 54 Fax 32 2 732 60 03 E Mail mail obelis net Tel 91 22 67829700 Fax 491 22 67829701 Email sales robonikindia com Website www robonik in Version 5 31 640 Effective Date Nov 2011 washwell PLATE 2 GENERAL SAFETY WARNINGS 2 1 Danger warnings symbols 2 2 Use of the instrument Version 5 31 7 AO Effective Date Nov 2011 washwell PLATE 2 1 Danger Warning symbols The following symbols are used to inform the user of the safety rules Pm 9 S gt lt General Symbols js j8 Version 5 31 This symbol indicates generic danger It means that serious damage can occur to the operator if described precautions are not observed This symbol indicates HIGH ELECTRIC VOLTAGE It is dangerous to touch any part having this label Only qualified operators can access these components after unplugging
5. Name Tests Y N Yes Test Saves without name ABCDEFGHIJKL Test Number will be printed on the paper Select Arrow Keys to select Alphabets Press Enter after selection and proceed Press Enter twice after completing selection Instrument aspirates the waste solution from the well Dispenses the programmed volume in to the micro strip well Aspiration is continuously kept on to prevent the over flow After completing dispensing cycle depending on soak time it starts aspirating the solution After this the instrument transports the micro plate back to home position Version 5 31 31 AO Effective Date Nov 2011 washwell PLATE 9 3 Incubation washwell PLATE consist of a two plate incubator which can be programmed individually After the set time elapses the instrument gives an audible alarm as well as notification on display indicating the elapse of time Both the timers can be programmed separately Incubator is activated when the instrument is in power on condition It takes 10 15 minutes to attain 37 C Temperature This temperature is factory set Procedure to program incubator time is as follows Please refer point No 8 5 Version 5 31 32 AO Effective Date Nov 2011 washwell PLATE 9 4 Bottom Wash Display How to operate Press B wash key Soak Time Y N Enter soak time in sec should be min 3 to max 3540 secs and Press ENTER Soak Time in Seconds
6. No of Cycles Enter no of wash cycles required min 1 max 9 and press ENTER e LL Select required volume of wash buffer 50 to 500 ul WASH VOLUME 50 Y N Flat Well Y N NO Cross Wash Y N Round Well Y N No of Strips Load Strips Plate and press ENTER Select the type of the well to be washed Select the type of the well to be washed Enter number of strips to be washed min 1 amp max 12 amp press ENTER Load Strips Plate and press ENTER Version 5 31 33 40 Effective Date Nov 2011 washwell PLATE 9 5 Overflow Wash Top Wash Display How to operate Press OF wash key Enter soak time in sec should be min 3 to max 3540 secs and Press ENTER Enter no of wash cycles required min 1 max 9 and press ENTER Select required volume of wash buffer 50 to 500 ul NO Cross Wash Y N Round Well Y N No of Strips Load Strips Plate and press ENTER 9 6 Priming Select the type of the well to be washed Enter number of strips to be washed min 1 8 max 12 8 press ENTER Load Strips Plate and press ENTER n Please refer point No 8 3 9 7 Rinsing Please refer point No 8 4 Version 5 31 34 0 Effective Date Nov 2011 washwell PLATE 9 8 Shaking EN Display How to operate Press 9 Key to check shake mode Shake Plate Y N Set Speed SP 1 Y N Select reguired Speed SP 1 to SP 8
7. YE Time in Secs Enter Time in secs to 59 sec amp Press ENTER Key 9 9 Disinfection EN Display How to operate Pes FUNC Ke YES Washer will flush the manifold Disinfection Y N 111nn effectively disinfecting the same Version 5 31 35 40 Effective Date Nov 2011 washwell PLATE 10 TROUBLESHOOTING Version 5 31 36 40 Effective Date Nov 2011 washwell PLATE 10 1 Troubleshooting PROBLEM SOLUTION The Manifold does not dispense Clean the path of the steel tubes using needle cleaner provided Check Wash bottle tubing connections Check valve direction The Manifold does not aspirate Clean the path of the steel tubes using needle cleaner provided Check the Waste bottle cap and tighten them properly Check waste bottle tubing connections The Instrument beeps Check the Sensor cable continuously connection Check whether waste bottle is full Check the waste bottle cap and clean the cap Movement Jam Reset the instrument by pressing the CLEAR key twice or Switch OFF and Switch ON the instrument Insert the Plate properly and return Memory Full If Number of saved tests exceeds 35 tests then delete the unwanted test and save the test Version 5 31 37 AO Effective Date Nov 2011 washwell PLATE 11 DECONTAMINATION 11 11 1 amp 11 2 amp 11 3 Version 5 31 Effective Date Nov 2011 Decontamination Purpose of Decontamination General
8. 1140 Effective Date Nov 2011 washwell PLATE 3 3 Specifications Manifold 8 Way Manifold autoclavable Dispensing Method Specially designed Peristaltic Pump Volume 50 500 ul Aspirating Method Continuous through diaphragm pump while dispensing hence preventing overflow Residual Volume lt 2 ul Waste Bottle One Capacity 2 liters with audible alarm when bottle is completely filled in with waste solution Wash Bottle One Capacity 2 liters Rinse Bottle One Capacity 2 liters Display 16 digit alphanumeric fluorescent Liquid contact materials Silicon stainless steel derline Memory 8 KB Non volatile RAM Battery backup supporting 35 Open channels Programming Modes Plate Wash Strip Wash Bottom Wash Overflow Wash Rinsing Priming Disinfect 1 to 59 Seconds 8 Steps For two ELISA Plates 37 C Individually Programmable 2 Nos Power Wattage 50 Watts Voltage 115 240 Volts 10 50 60 Hz Operating Conditions Temperature 18 C to 35 C Humidity Up to 85 Storage Conditions Temperature 10 C to 60 C Humidity Up to 85 ABS Fire retardant I3 1X 5X 6 Ke Approx Version 5 31 1240 Effective Date Nov 2011 washwell PLATE 4 PACKING TRANSPORT AND STORAGE 4 1 Packing 4 2 Instrument Transportation 4 3 Storage of the Instrument Version 5 31 13 40 Effective Date Nov 2011 washwell PLATE 4 1 Packing Packaging of the instrument is needed whenever to be
9. 2011 washwell PLATE 7 PRECAUTIONS Version 5 31 23 AO Effective Date Nov 2011 7 1 Version 5 31 washwell PLATE Precautions Always check for proper grounding during installation Never operate the instrument with ground wire removed Do not attempt to open the instrument and make repair without proper technical training Do not allow unauthorized persons to operate or repair the instrument Place the ELISA Plate carefully on the tray Ensure that the main power switch is in OFF position before connecting Prime the manifold before and after use Always Prime immediately after switching ON the instrument Always rinse before switching OFF Keep wash waste and rinse bottles clean Disconnect the tubing before opening wash waste and rinse bottle caps Do not operate any key after switch OFF the instrument since instrument performs RINSE operation first and then get switched OFF 2200 Effective Date Nov 2011 washwell PLATE 8 GENERAL KEY OPERATIONS 8 1 FUNC Key 8 1 1 Deleting the test 8 1 2 Setting Date and Time 8 1 3 Storing the Test 4 8 2 Menu Key 8 2 1 Recalling the saved test 8 3 PRIME Key 8 4 RINSE Key 8 5 TIMER Key Version 5 31 25 AO Effective Date Nov 2011 washwell PLATE Check whether all the tubing is properly connected Switch on the instrument 8 1 FUNC KEY ES Idle Mode Disinfection Y N NO Load wash S
10. 2011 washwell PLATE 8 GENERAL KEY 7 25 8 1 FUNG KEY Mcr U 26 8 1 1 Deleting the Test ana assi amma ma emon Eat Reb pcena 27 8 1 2 Setting Date 27 81 3 DANG Test eura Re epa i 27 8 2 iuc aana gae viis cipi Vai e RE nean KING LE DIVA Ua cU ecd EA I RC 28 8 2 1 Recalling the saved test essere 28 8 3 PRIME 8 8 4 RINSE KEY aine saga da mind idend ere cci eroe Ld enu bain co a ENS 28 8 5 TIMER KEY mm 28 9 PROGRAMMING 2 29 9 1 30 9 2 STAP WASEE adana a bre KN m nga A NG GRE cU uu RHEIN du Ma Di uci 31 9 3 INCUBATION 815 ana nange ndak fein kei Kak MO Re E cda De Dunn Dd aa Panak 32 9 4 BOTTOM WAS PR asi dva date erra ivi poi Den d cuc iva nce 33 9 5 OVERFLOW WASH TOP WASH 7 34 9 6 PRIMING e X 34 97 34 9 8 35 9 9 DISINFECTION Met 35 10 TROUBLESHROOTI Dara Dura ba 36 10 1 TROUBLESHOOTING aei dia cx EC una aka 37 11 DECONTAMINATION anu diua uds 38 11 DECONTAMINATION nagane uid tosta UE K o mi BR 39 11 1 PURPOSE OF DECONTAMINATION cerne enne nnnm nnnm nnne tnnt annan 39 11 2 GENERAL C
11. Considerations Decontamination procedure 38 40 washwell PLATE 11 Decontamination If the instrument is to be shipped after being exposed to potentially hazardous material it should be decontaminated The following procedure outlines the method to decontaminate the instrument before packaging and shipment Purpose of Decontamination Decontamination minimizes the risk to all who come in contact with the instrument during shipping handling and servicing Version 5 31 General Considerations Any laboratory instrument that has been used for clinical analysis is considered a bio hazard and should be decontaminated prior to handling Intact skin is generally considered an effective barrier against infectious organisms however small abrasions and cuts may not be always visible Prophylactic gloves must be worn when handling instruments that have not been decontaminated Gloved hands should be considered contaminated and must be kept away from eyes mouth and nose at all times Mucous membranes are considered as the prime entry routes for infectious agents Wear eye protection and a surgical mask when there is a possibility of aerosols Eating and or drinking while decontaminating instruments is not advisable Decontamination procedure A solution of 0 596 Sodium Hypo Chlorite NaOCL solution Bleach is used Commercial bleach is 5 NaOCL household bleach is 3 NaOCL When using commercial bleach use a 10 1 mi
12. by the locally available approved plug in cable Power cord should be CE CSA and UL marked Voltage 115 240 Volts 10 50 60 Hz 6 3 Protective Grounding Warning Please make sure that electrical power source is properly grounded 6 4 Start up Instructions Instrument Starting Procedure 1 Switch on the instrument The instrument will display WASH WELL PLATE 2 The instrument carries out Self Test to check all the internal parameters Itdisplays WP V5 31 and the time indicating that initialization is complete 4 If the date and time are not correctly displayed switch off the instrument and switch it ON once again The instrument is now ready for use Version 5 31 2140 Effective Date Nov 2011 washwell PLATE 6 5 Setting date and time Setting of date and time can be done with the help of FUNC key Display How to operate Set TIME Y N Press YES Key Enter Date in Same format Press ENTER key Date DD MM YY Enter Time in same format Press ENTER key Time HH MM SS 6 6 Waste Bottle connections Please refer point No 5 3 Tubing Diagram 6 7 Rinsing Strip Load rinsing strip in the rinsing slot of plate holder Press FUNC key Load wash STR Y N Place strip and press enter Place the rinsing strip on the slot and press enter key Direction of Plate RINSING STRIP SLOT gt SA PLATE AREA Fig a Version 5 31 22 AO Effective Date Nov
13. diagram of point No 5 3 e Ensure that the main power switch is in OFF position before connecting e Plug the instrument to the A C mains Confirm proper grounding for trouble free operation The washwell PLATE is provided with an inbuilt voltage stabilizer to take care of voltage fluctuations Version 5 31 16 40 Effective Date Nov 2011 washwell PLATE 5 3 Tubing Diagram REAR PANEL TUBING CONNECTIONS Waste o o AR Red Bottle Contrast Oo Manifold Sensor Blue UN Green NG Wash Bottle Washer Waste Bottle a Bottle Washer Cs Sensor A Green Red Orange Prime Motor Rinse Motor To Waste bottle To wash bottle Note Load ELISA plate carefully in the slot provided for correct washin Version 5 31 17 AO Effective Date Nov 2011 washwell PLATE 5 4 Keyboard jJ 5 5 Micro plate Carriage Movement of plate is precisely controlled by stepper motor and timing belt to position the plate exactly below the manifold for washing Ref Fig a Direction of Plate Loading Priming amp Rinsing Area Plate Loading Area Fig a Version 5 31 18 40 Effective Date Nov 2011 5 6 5 7 Version 5 31 washwell PLATE Priming and Rinsing Area Carriage plate has built in priming and rinsing area priming and rinsing is essential immediate
14. the instrument from the Supply This symbol indicates that the instrument makes use of chemical reagents and other dangerous Corrosive irritant or harmful CHEMICAL SUBSTANCES which can cause damage to people or materials When this label is found pay attention to the manufacturer s recommendations This symbol indicates that the instrument involves the handling of samples which can be infected urine or human serum In this condition infection or contamination might occur Pay attention to the general safety warnings when in presence of such biological substances Use Protective clothes gloves and glasses This symbol in the user manual indicates that damages to the instrument or erroneous results could occur if the given warnings are not followed This symbol indicates a portion which is particularly important and should be studied carefully This symbol indicates a Protective Earth or Ground terminal Symbol for Manufacturer Symbol for IN VITRO DIAGNOSTIC MEDICAL DEVICE Symbol for AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY 840 Effective Date Nov 2011 washwell PLATE 2 2 Use of the instrument gt The instrument should be used for the designed purposes only and under specified conditions by qualified personnel following proper procedures and safety rules provided in this manual gt THIS MANUAL CONTAINS INSTRUCTIONS FOR OPERATION BY QUALIFIED PERSONNEL ONLY gt A qualified user
15. transported or shipped by courier or other means To pack the instrument please follow the instructions as below described e Decontaminate the instrument as explained in chapter 11 of this manual e Put the instrument into the original packaging box Instrument has to be properly protected by plastic protective material Put copy of safety clearance certificate copy of Safety clearance certificate is attached at the end of this manual e Mark the package with address instrument identification and warning labels 4 2 Instrument transportation The transportation of the instrument in unpacked condition must be limited within the room where it is used to avoid damage 4 3 Storage of the Instrument Before storing the instrument for a long period please pack it carefully as described above and store indoors Version 5 31 14A0 Effective Date Nov 2011 washwell PLATE 5 INSTRUMENT DESCRIPTION 5 1 Perspective view 5 2 Rear panel view 5 3 Tubing diagram 5 4 Keyboard 5 5 Micro plate carriage 5 6 Priming and rinsing area amp 6 5 7 Manifold Version 5 31 15 40 Effective Date Nov 2011 washwell PLATE Different views of the instrument in picture below 5 1 Perspective View Front view Display Alphanumerical key pad Wash Bottle Waste Bottle Rinse Bottle Wash Bottle Rinse Bottle Keyboard amp Display Micro Plate 5 2 Rear Panel View Please refer
16. ONSIDERATIONS eren eere rennen maan ntn tnnt natn asma naa tana 39 11 3 DECONTAMINATION PROCEDURE 7 aa a aan 39 12 SAFETY CLEARANCE CERTIFI 40 Version 5 31 Effective Date Nov 2011 washwell PLATE The contents of this manual with all figures tables and graphics are intellectual property of ROBONIK INDIA PVT LTD Unauthorized commercial non commercial excerption or copying of contents and use of this manual in total or in parts are strictly forbidden unless the editor gives written permission for it washwell PLATE manual was written and produced with the utmost care However errors cannot be fully excluded Robonik does not take any responsibility and accept no liabilities of any kind that may occur because of errors in the manual Manufacturer ROBONIK INDIA PVT LTD Plot No A 374 TTC MIDC Industrial Area Mahape Navi Mumbai 400 710 INDIA Tel 91 22 67829700 Fax 91 22 67829701 Email sales robonikindia com Website www robonik in washwell PLATE Version 5 31 Version 5 31 4A0 Effective Date Nov 2011 washwell PLATE 1 GENERAL INFORMATION 1 1 Warranty information 1 2 Technical service 1 3 Disposal instruction 1 4 Contacts Version 5 31 50 Effective Date Nov 2011 washwell PLATE 1 1 Warranty Information Each Instrument is comp
17. d manifold needles 8 4 RINSE KEY Pressing Rinse Key peristaltic Pump Rinse Motor operates continuously for few seconds Distilled water from rinse bottle rinses the tubing and manifold needles an 8 5 TIMER KEY men NO Set timer Y N YES NO YES Time in seconds NO YES YES Both the timers get reset Version 5 31 28 AO Effective Date Nov 2011 washwell PLATE 9 PROGRAMMING 9 1 Plate Wash 9 2 Strip Wash 9 3 Incubation 9 4 Bottom Wash 9 5 Overflow Wash Top Wash 9 6 Priming 9 7 Rinsing 9 8 Shaking 8 9 9 9 9 Disinfection Version 5 31 29 AO Effective Date Nov 2011 washwell PLATE 9 1 Plate Wash Display How to operate PLATE WASH o Press P wash key NO Soak Time Y N Soak Time in Seconds Enter soak time in sec should be min 3 to max 3540 secs and Press ENTER No of Cycles LL V V Enter no of wash cycles required min 1 max 99 and press ENTER WASH VOLUME 50 Y N Select required volume of wash buffer 50 to 100 150 200 250 300 350 400 YES Flat Well Y N 8 500 ul NO Cross Wash Y N Round WELL Y N Select the type of the well to be YES NO washed Enter strip Pos Enter the strip position No No of Strips Enter number of strips to be washed min 1 amp max 12 amp n press ENTER po 4 gonane Press FUNC Key Press Clear Twice to exit the Test To Save Test Press FUNC Key
18. has to make sure that environmental condition is suitable the installation is correct the use and maintenance is proper according to the general safety rules as well as to the particular precautions described in the manual However the user is not entitled to repair the instrument gt qualified technician only is entitled to maintain and fix the instrument according to the instructions given in this manual using the original spare parts only gt room temperature and humidity should be maintained as specified in the manual gt The instrument has to be used as described in this manual If it is not use the protection provided by the instrument may be impaired gt Alterations to the instrument by unauthorized personnel are strictly prohibited The user is liable amp solely responsible for any improper modification to the instrument and for the consequences derived as a result gt Should the instrument need extraordinary maintenance please contact Robonik service or an authorized service center Specialized technicians equipped to repair the instrument using original spare parts will carry out the maintenance gt This IVD equipment complies with the emission ad immunity requirements as per IEC61326 series gt Warning This equipment has been designed and tested to CISPER11 Class A Ina domestic environment it may cause radio interference in which case you may need to take measures to mitigate the interference
19. letely tested and guaranteed for twelve months from delivery The warranty applies to all the mechanical and electrical parts It is valid only for proper installation use and maintenance in compliance with the instructions given in this manual ROBONIK will at its discretion repair or replace parts which may be found defective in the warranty period The warranty does not include any responsibility for direct or indirect personal and material damages caused by improper use or maintenance of the instrument Parts that are inherently subject to deterioration are excluded from the warranty In case of defects due to misuse of the instrument any travel and man hour expenses will be charged extra In case of Tenders warranty would be as per tender terms and conditions 1 2 Technical Service ROBONIK is always accessible to the customers for any kind of information about installation use maintenance etc When asking for service please refer to this manual and report the data reported on the identification label Serial No and Model Name Only qualified technicians are entitled to fix the instrument the user as described in this manual should carry out ordinary maintenance ROBONIK s technical service or an authorized service center with specialized technicians with suitable instrumentation and original spare parts are always available for extraordinary maintenance repair under an annual maintenance contract or on specific demand
20. ly after starting the instrument and before shutting down the instrument It helps smooth flow of buffer solution and prevents crystallization Ref Fig a Manifold Instrument has a specially designed 8 way manifold The manifold has a provision to clean the liquid path conveniently hence avoiding common clogging problem because of crystallization etc Manifold is moved vertically up and down by well controlled stepper motor for proper placement Manifold needles are laboratory grade steel smooth ends to prevent scratching of the well bottom 19 40 Effective Date Nov 2011 washwell PLATE 6 INSTALLATION AND START UP INSTRUCTIONS 6 1 Placing the instrument 6 2 Power Supply 6 3 Protective grounding 6 4 Startup instructions 6 5 Setting time 6 6 Waste Bottle connections 6 7 Rinsing Strip Version 5 31 20 AO Effective Date Nov 2011 washwell PLATE While installing and setting up the instrument the safety warnings and general precautions described in section 7 must be observed 6 1 Placing the instrument The instrument has to be placed on a level bench Amin distance of 1m should be maintained from the rear panel to avoid damage to power cord and instrument Room temperature has to be between 18 C and 35 C with a relative humidity below 85 Protect it from direct sunshine 6 2 Power supply Once the instrument has been placed plug it into a power source
21. trip Y N NO Delete Test Y N NO Set Time Y N NO Exit Y N NO Version 5 31 Effective Date Nov 2011 YES YES YES YES Washer will flush the manifold effectively disinfecting the same Place Strip amp press Enter Ref 8 1 1 Delete Option Ref 8 1 2 Modify Time Option 26 AO washwell PLATE 8 1 1 Deleting the Test Display Delete Test Y N Test No Deleting Test Yes Test Deleted 8 1 2 Setting Date and Time Refer point No 6 5 8 1 3 Storing the Test How to operate Press FUNC Key Press YES Key Enter the Test No of the test to be Deleted To save the programmed test under any programming Mode Display Save Test Y N Yes Name Tests Y N Save Test without name ABCDEFGHIJKL Yes Version 5 31 27 AO Effective Date Nov 2011 How to operate Press FUNC Key Press YES Press YES Key to Enter Name Test Number will be printed on the paper Select Arrow Keys to select Alphabets Press Enter after selection and proceed Press Enter twice after completing selection washwell PLATE 8 2 MENU KEY EN This key is used to recall the test number to be performed 8 2 1 Recalling the saved test Enter the test No to be recalled 2127 and Press ENTER 83 PRIME KEY Pressing the Prime Key Peristaltic Pump Prime Motor operates continuously for few seconds Wash buffer from wash bottle primes the tubing an
22. washwell PLATE washwell PLATE ELISA Plate Washer User Manual Version 5 31 2 ROBONIK INDIA PVT LTD Plot No A 374 TTC MIDC Industrial Area Mahape Navi Mumbai 400 710 INDIA Tel 91 22 67829700 Fax 91 22 67829701 Email sales robonikindia com Website www robonik in Version 5 31 1 AO Effective Date Nov 2011 washwell PLATE TABLE OF CONTENTS 1 GENERAL I anan neanen nannamamma aa nananana anna muuna 5 1 1 WARRANTY INFORMATION 7 7 6 1 2 TECHNICAL 6 1 3 DISPOSAL INSTRUCTION 6 4 ZCONTAGCTSE icc divae EE dan Yu R aga Ce Va aa a d EET ak RR d D ina iud 6 2 GENERAL SAFETY WARNIN G S ee 7 2 1 DANGER WARNING 8 2 2 USE OF THE 0 nnmnnn nnmnnn nenna 9 3 INTRODUCTION e00eensneennnen sanan nen KNIN HAS NNN asanan anan ananas 10 3 14 EAE EEE E E EA 11 3 2 SPEGIALFEATURES 1 a NGU gan Ula RC rcu dude CR Ga D avaama a dC 11 3 3 SPECIFICATIONS aas EE A a ga ab Ci EE Raka aba m N EA aaa AG aa WA eag 12 4 PACKING TRANSPORT AND 5
23. xture if using household bleach a 6 1 mixture is required This is a caustic solution It is important to wear gloves and eye protection when handling it Wipe down the carrier and all exposed surfaces of the unit with the bleach solution Remove the top shroud of the instrument and wipe down the top surface of the instrument base as well as the inside of the top shroud Reassemble the unit and discard the used gloves and towels 39 Effective Date Nov 2011 washwell PLATE 12 SAFETY CLEARANCE CERTIFICATE Please complete all information requests on this form prior to returning the instrument to the manufacturer or your local distributor for servicing repairs or return Thank you for your co operation Customer Address Country Post Code Model No Accessories Returned Date of Purchase if known Complaint Has the equipment been exposed to any of the following a Blood body fluids pathological specimens If YES please specify b Other Biohazard if YES Please specify Version 5 31 40 AO Effective Date Nov 2011 Contact Position Dept Tel Fax Serial No delete as applicable YES NO YES NO