Pelton Crane Omni Cleaner Instructions

image

Contents

1. Cleaning Door Gasket D Clean door gasket with Omni Cleaner Plus or non chlorinated detergent and water Note If residue is allowed to accumulate the seal could be affected and leaks may occur Cleaning Disinfecting Exterior Surfaces Clean all exterior surfaces with mild detergent and water using a sponge or cloth Disinfect exterior surfaces using an iodophor Biocide Biotrol Inc N Salt Lake City Utah or equivalent CAUTION Do Not use any disinfectant on interior stainless steel surface Damage to the chamber and or trays may result
2. 5 How to clean autoclave MAINTENANCE DELTA STERILIZER CAUTION Units is in preheat mode between 0600 1800 hours So that units will be cool to the touch turn off unit at main power switch back of unit 30 minutes before performing maintenance Maintenance and Performance Check Schedule CHECK FREQUENCY PROCEDURE ACTION Clean chamber After every 4 ozs Omni Cleaner Plus per 64 ozs distilled water See Cleaning Procedures 25 cycles Door gasket Weekly Inspect and clean using Omni Cleaner Plus or mild detergent Call authorized service representative for and water Check for leaks and have leaking gasket replaced replacement gaskets WARNING Leaking gasket could result in harm to NOTE For proper operation use only the operator because of escaping steam our replacement parts available through authorized service dealers Door latch Weekly Inspect latch mechanism for signs of wear or improper closure Call authorized Service representalive ttii proper closure or signs of wear are noted Bio Indicator Weekly Conduct testing using biological spore test indicator as de Hag ie not ee en ae scribed in Operation Sterilization Assurance ti atoa nsure mal mportant terria tion Practices and Preparation and Load ing sections in the User Manual have been followed Retest using an empty chamber then re test with properly packaged and loaded chamber If test still fail
3. cts WARNING When ring is pulled on safety valve with unit under pressure steam is discharged from the chamber straight down the pipe at high tem perature When you pull ring steam exits out bot tom Make sure path is clear and that your hand is out of the way Power and call authorized service rep resentative for replacement It is recommended that you annually have your sterilization system inspected by an authorized service and retain maintained records for the following 1 Hi Pot Test 2 3 4 5 6 N Terminal Inspection Internal Tubing inspection for leaks Filter replacements Ground continuity test Internal wire inspection for frayed wire ote For detailed service instructions refer to autoclave Service Manual 4204801 26 31 MAINTENANCE DELTA STERILIZER Cleaning Procedures NOTE Proper cleaning at regular intervals is essential since mineral deposits debris or trash may cause the unit to malfunction Draining Reservoir With the door open locate the quick drain connection A at the lower right corner of the face plate Press the drain tube connector B provided with the unit into the connection on the face plate This opens the drain line and draining of the reservoir will begin immediately Normal Chamber Cleaning Drain the reservoir using quick connect drain tube Mix Omni Cleaner Plus with distilled water according to di rections on bottle Pour i
4. nto reservoir Run one Un wrapped cycle Note It is not necessary to complete drying cycle Drain reservoir With the drain tube in place pour one to two gallons of water into the reser voir until the drained water is clear This will remove excess suds Wipe inside of the cham ber with soft cloth WARNING Do Not sterilize instruments while cleaning autoclave Pour half gallon distilled water into reservoir Run one Unwrapped cycle Note Itis not necessary to complete drying cycle Wipe the inside of the chamber the boiler ring C and the door gasket Fill reservoir with distilled water Autoclave is ready for use Cleaning Deposits and Discoloration on Stainless Steel For deposits not removed by detergent solution use a non chlorinated non metallic scouring pad such as Scotch Brite or a powder such as Bon Ami CAUTION Do not use ordinary steel wool or steel brushes on stainless steel Pads contain ing metal will damage chamber and cause instruments to be stained with rust CAUTION Do not use chlorinated cleaners Doing so will cause corrosion Rub in direction of pattern or grain of the metal Clean stainless steel surfaces contaminated with discolora tion with a 5 solution of warm oxalic acid MAINTENANCE DELTA STERILIZER Cleaning Procedures cont d Cleaning Boiler Ring C Clean boiler ring using a non chlorinated non metallic scouring pad such as Scotch Brite
5. s discontinue using sterilizer and contact an authorized service representitive With door open attempt to start cycle If door alarm does not i Door switch Weekly appear on alarm display door switch is defective or requires Call authorized service representative adjustment Unit should not operate Boiler ring Weekly Inspect for deposits Clean using non chlorinated pad which contains no metals Operation check Weekly Perform test described on pg 40 Call authorized service representative if fail ure occurs Water filters Monthly Remove filter in chamber Put filter in unltrasonic cleaner If filters are clogged are will not clean call Scrub with small soft brush under running water and an authorized service representative for replace filter When replacing make sure filter replacement filters in chamber touches center bottom wall of chamber to ensure proper draining Air filter Every Ser Inspect air filter for cracks and loose tubing connec Have service representive remove the vice Call tions cover and inspect Annually Replace Filter Call authorized service representive Safety valve ring Every 3 Manually pull ring located on upper rear of chamber If valve does not open or does not com months when chamber pressure reaches 121 kPa When pres pletely seal off after operation turn off Note For detailed service in structions refer to autoclave Service Manual sure is relieved valve automatically retra