Home
MetPak
Contents
1. CR10X Example program for two CS215 sensors on control port 1 Measures every minute and stores some summary statistics once per hour Table 1 Program 01 60 Execution Interval seconds Measure the two sensors with separate commands but with the Same port number Note you have to manually assign the labels first in the Inloc Editor to make sure two locations are free for each sensor 1 SDI 12 Recorder P105 1 0000 SDI 12 Address 2 00 SDI 12 Command 3 1 Port 4 1 Loc CS215T1 J 5 1 0 Mult 6 0 0 Offset 2 SDI 12 Recorder P105 1 0001 SDI 12 Address 2 00 SDI 12 Command 3 1 Port 47 3 Loc CS215T2 5 1 0 Mult 6 0 0 Offset Now store some statistics 3 If time is P92 1 0 Minutes Seconds into a 2 60 Interval same units as above 3 10 Set Output Flag High Flag 0 4 Real Time P77 1 111 Day Hour Minute Seconds midnight 0000 5 Sample P70 1 4 Reps 2 1 Loc CS215T1 J 6 Average P71 1 4 Reps 2 1 Loc CS215T1 7 Maximum P73 1 4 Reps 2 00 Time Option 3 1 Loc CS215T1 J 8 Minimum P74 1 4 Reps 2 00 Time Option 3 4 Loc CS215T1 Table 2 Program 02 0 0000 Execution Interval seconds Table 3 Subroutines End Program User Guide 5 1 2 Example CRBasic Program Measuring One Sensor Every 30 Seconds The following example is written for the CR200 datalogger Programs for the CR1000 and CR5000 would be similar although many
2. To activate the SDI 12 transparent mode on Control Port p enter 8 The TD datalogger will respond with a prompt At the prompt enter The TD datalogger will respond with 150000 Finally enter p Control Port p and press the lt enter gt key The TD datalogger will respond with entering SDI 12 Verify the new SDI 12 address by entering the command The CS215 will respond with the new address 9 To exit the SDI 12 Transparent Mode type in 08 150000 entering SDI 12 210 OAllexiting SDI 12 150000 Sentering SDI 12 exiting SDI 12 Select Device Baud Rate JV All Caps Mode Close Terminal Clear Help CR1OXTD y 3600 y FIGURE 2 Screen Capture of SDI 12 Transparent Mode on Edlog Table Based Datalogger using Control Port 8 and Changing SDI 12 Address from 0 to 1 11 CS215 Temperature and Relative Humidity Probe 6 3 CR200 Series Datalogger 1 Connect a single CS215 to the datalogger using Control Port C1 SDI12 as described in Section 4 and download a datalogger program that does not contain the SDI12Recorder instruction In the LoggerNet Toolbar navigate to and activate the TestITerminal Emulator menu The Terminal Emulator window will open In the Select Device menu located in the lower left hand side of the window select the station Click on the Open Terminal button If communications between the datalogger and PC are successful t
3. 1 Measurements at Slow Scan Rates nnessssssssssnsnsnennnnnnnnnnnnnnnnnnnnnnnnnnnnenn 5 5 1 1 Example Program for CR10X Measuring Two Sensors Every Minute 5 5 1 2 Example CRBasic Program Measuring One Sensor Every 30 Seconds 7 5 2 Measurements at Fast Scan Rates 0 0 0 ceeeeesseceseeecsseceeneecsseceeeeecaeceeneeseeseneees 7 3 2 1 Edl08 D tal sgers 2er 7 5 2 2 CRBasic Dataloggers iii ns 8 6 Changing the SDI 12 Address Using LoggerNet and a Datalogger 2 He ee 9 6 1 Array Based Edlog Dataloggers 2042200224000ennenneensennneennennnennennnennnn 9 6 2 Table Based Edlog Dataloggers 22 020422042 2000 snennensnennennnennnnnnnnnnnennn 10 6 3 CR200 Series D taldggein nn n nase nenn east 12 6 4 CR1000 D tal gger 2 8 2 8 82 25 ee decanta cba ies 13 7 Maintenance and Calibration uureesssnnnneennnnnnnnnnnnnnnnn 13 7 1 Sensor Element Replacement 224022402s00nseensnnnnennesnnesnnesnennnnsnen sone 13 Appendix A Environmental Performance A 1 A 1 Tests to Defined Standards 00 0 ec eeesesecseeeeceseeecesecseesecseeeecaeeeesaeeaeeeeenees A 1 A 2 Exposure to Pollutants neneen nennen A 1 A 3 Operating Range of RH Element ooonnonncnocnnononononoconnonncnnonncononononcnnorononccnno A 1 Appendix B Sensor Element Calibration B 1 Figures Tables 1 Screen Capture of
4. Continuous mode From Polled mode lt N gt u where lt N gt is the Unit identifier The WindSonic responds with a CONFIGURATION MODE message stops reporting wind measurements and waits for a command as detailed below 9 2 Returning to Measurement mode Q ENTER The WindSonic responds with wind measurements immediately when in Continuous mode or on receipt of a Poll signal when in Polled mode 9 3 Checking the configuration We recommend that as a standard procedure you use this command prior to and after changing any settings It shows the current settings for all the alterable settings Ik D 3 ENTER The WindSonic responds with the default settings shown below M2 U1 O1 L1 P1 B3 F1 H1 NQ El T1 S3 C2 We suggest you note down the settings so that you can easily return to them Return to measurement mode Q ENTER All of these settings are explained in the following sections 9 4 Changing a setting To change a setting refer to the sections below and enter the command of the new setting required followed by IENTER If successful the new setting will be sent as a message by the WindSonic For example to change the message format to NMEA enter M 5 ENTER The WindSonic will reply M5 When the unit is returned to the Measurement mode it wil
5. NIST and NPL traceable NPLEI Standards and Technology National Physical Laboratory Date of Certification February 2003 Valid till March 2005 Zurich March 2004 AD CL AA CEO Sensirion AG Quality Assurance Manager Sensirion AG B 1 Campbell Scientific Companies Campbell Scientific Inc CSI 815 West 1800 North Logan Utah 84321 UNITED STATES www campbellsci com info campbellsci com Campbell Scientific Africa Pty Ltd CSAf PO Box 2450 Somerset West 7129 SOUTH AFRICA www csafrica co za sales O csafrica co za Campbell Scientific Australia Pty Ltd CSA PO Box 444 Thuringowa Central QLD 4812 AUSTRALIA www campbellsci com au info Ecampbellsci com au Campbell Scientific do Brazil Ltda CSB Rua Luisa Crapsi Orsi 15 Butant CEP 005543 000 S o Paulo SP BRAZIL www campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta TSM 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Terre Neuve Les Ulis 91967 COURTABOEUF CEDEX FRANCE www campbellsci fr campbell scientific wanadoo fr Campbell Scientific Spain S L Psg Font 14 local 8 08013 Barcelona SPAIN ww
6. WindSonic sensor mounted on a six plate aspiration shield fitted with Rotronic HygroClip S3 Temperature and Humidity probe Complete with mounting bracket and handbook MetPak White WindSonic Windspeed amp direction sensor with analogue output mounted on a six plate aspiration shield Customer to fit temperature and humidity sensor of their choice up to 15mm dia Complete with mounting bracket and handbook 1575 PK 004 Option 5 WindSonic for MetPak White WindSonicWindspeed amp Direction sensor fitted with 1 meter of cable and aspiration shield mounting adaptor and SDI 12 output Distributor to mount on their own aspiration shield and fit their own probe 1575 PK 005 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 Option 6 WindSonic for MetPak White WindSonicWindspeed amp Direction sensor fitted with cable and aspiration shield mounting adaptor and analogue output Distributor to mount on their own aspiration shield and fit their own probe 1575 PK 006 4 2 Wind Sensor Specification Output Units of measure Output frequency Metres second m s Knots Miles per hour mph Kilometres per hour kph Feet per minute fpm 1 2 or 4 outputs per second Digital Analogue Parameters Polar Speed and Direction Polar Speed and Direction or UV 2 axis signed Speed Tunnel U Speed and U Direction Wind Speed Range 0 60 m s 0 30m s 0 50m s O
7. WindSonic wind sensor fitted to MetPak Opening the WindSonic unit or breaking the security seal will void the warranty and the calibration In the event of failure prior to returning the unit to your authorised Gill distributor it is recommended that 1 All cables and connectors are checked for continuity bad contacts corrosion etc 2 A bench test is carried out as described in Section 12 1 3 You contact your supplier for advice For future service or maintenance of the temperature and humidity sensors please refer to appandix 13 5 or 13 6 of this manual 11 3 Fault finding Applicable to WindSonic only Symptom Solution Check DC power to WindSonic cable and connections Check comms settings of WindSonic as detailed in Section 10 and host system match including correct Com port No output Check unit is in Continuous mode Check that in line communication devices are wired correctly NOTE It is usual for Anemometer TX to be connected to converter device RX Check comms settings of WindSonic and host system match Corrupted output Try a slower baud rate Check cable lengths and type of cable Check WindSonic and host system are both set to the same One way protocol RS232 RS422 or RS485 communication Check wiring is in accordance with the manual Failed Incorrect Check that transducer path is not blocked WindSonic ouput data invalid flag 25 MetPak Doc No 1575 PS 0001 Is
8. figure 4b shows the incorrect orientation CAUTION If the sensor is inserted in one of the wrong orientations it will not work excessive power will be drawn from the supply and the element may be damaged if left powered in this state for more than a few seconds Once the correct orientation of the probe is identified first make sure that your hands are clean to avoid getting dirt or grease on the element Grasp the body of the sensor this also ensures you are at the same electrical potential as the element and holding the black tip of the element between your finger tips pull the element out of the socket Store the old element in electrostatic protective packaging if you wish to retain it With the element removed check for dirt and or corrosion around the socket Clean any dirt away using a damp cloth to remove any salts that might be there Unpack the replacement element making sure that you touch the packaging rather than the element first to avoid static discharges directly to the element Either hold the element by the black top of the package the other end to the gold plated pins or use a pair of fine nosed pliers or tweezers to grip the sensor by the pins Carefully offer up the pins to the socket in the end of probe Confirm the correct orientation and then gently push the pins into the socket until they will not go in any further Before replacing the filter element and turning on the power to the sensor double ch
9. ground G To use more than one probe per datalogger you can either connect the different probes to different SDI 12 compatible ports on the datalogger or change the SDI 12 addresses of the probes and let them share the same connection Using the SDI 12 address minimizes the use of ports on the datalogger and also allows probes to be connected in a daisy chain fashion which can minimize cable runs in some applications However see below for limits on the total cable length There are two ways to set the SDI 12 address of the CS215 e By sending the required commands to the sensors via an SDI 12 recorder datalogger that allows talk through to the sensor e By loading a program into the recorder that sends the required commands see section 6 below As the measurement data is transferred between the probe and datalogger digitally there are no offset errors incurred with increasing cable length as seen with analogue sensors However with increasing cable length there is still a point when the digital communications will break down resulting in either no response from the sensor or corrupted readings The original SDI 12 standard specifies the maximum total cable length for the cable as being 200 feet 61 metres In practice even longer cables than this can be used but to ensure proper operation with long cables follow these guidelines e Use low capacitance low resistance screened cable as fitted by Campbell Scientific Wit
10. instructions have extra parameters required to control the extra capability in those dataloggers The SDI 2recorder instruction also varies somewhat between the dataloggers see the CRBasic help for the respective logger for further details CR200 Series Datalogger Example program showing measurement of a single CS215 sensor every 30 sec Declare the variables we are going to use Public CS215meas 2 An array suitable to read the SDI 12 data into Alias CS215meas 1 Temperature Alias CS215meas 2 RH Define Data Table to hold the stored data DataTable CS215data 1 1 DataInterval 0 15 min As an example store the average every 15 mins Average 1 Temperature False Average 1 RH False EndTable Main Program BeginProg Scan 30 Sec Scan every 30 seconds SDIl2Recorder CS215meas 0R 1 0 0 Read the last measured values into the array CallTable CS215data Call the table instruction to calculate and store the averages NextScan EndProg 5 2 Measurements at Fast Scan Rates Before considering making higher speed measurements with the CS215 it is important to understand one possible source of error This is that the sensor element in the CS215 contains active elements so is subject to errors caused by self heating This is where power dissipated in the element raises its temperature leading to a slight over reading of temperature and under reading of relative humidity These errors will only be significant in condi
11. minimise the effects of dust and dirt on the sensor The filter is lightweight and hydrophobic thereby minimising the effect of the time response of the sensor The probe housing is designed to withstand permanent exposure to all weathers and to fit into a range of radiation shields including lower cost compact shields Rotronic HygroClip The HygroClip modules are complete instruments with integrated temperature compensation and highly flexible power requirements Calibration data is maintained within the integrated electronics It provides linear analogue output signals for rh and C Fig 3 Position of Temperature and Relative Humidity Probe in the Radiation Shield MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 4 OPTIONS AND SPECIFICATION 4 1 Options and Part Numbers Available Product and Part Description Option 1 MetPak Wind speed wind direction temperature amp humidity with SDI 12 output White WindSonic sensor mounted on a six plate aspiration shield fitted with Campbell Scientific CS215 Temperature and Humidity probe Complete with mounting bracket and handbook MetPak White WindSonic Windspeed amp direction sensor with SDI 12 output mounted on a six plate aspiration shield Customer to fit temperature and humidity sensor of their choice up to 15mm dia Complete with mounting bracket and handbook MetPak Wind speed wind direction temperature amp humidity with analogue output White
12. of gas flow T1 Transit time of ultrasound T2 Transit time of ultrasound Figure 1 Time of Flight details MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 NORTH MARKER ARROWS Figure 2 Compass Points 3 2 Radiation Shield The Multi Plate Radiation Shield protects temperature and relative humidity sensors from error producing solar radiation and precipitation This shield relies on a combination of plate geometry material and natural ventilation to provide effective shielding MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 3 3 Temperature and Relative Humidity Sensor Two types of temperature and humidity probes are available dependant on the option ordered Campbell CS215 temperature and relative humidity probe The CS215 probe utilises a novel single chip element that incorporates both a temperature and RH sensor Each element is individually calibrated with the calibration corrections stored on chip The element is easily changed in the field Replacement of the element effectively fully recalibrates the probe both for temperature and humidity thus minimising downtime and calibration costs Low power electronics within the CS215 controls the measurement made by the sensor element applies temperature and linearization corrections to the readings and presents the data via an SDI 12 compatible interface to a datalogger The probe is fitted with a custom made yet low cost sintered plastic filter to
13. output Output type Polar mode direction calculation The system will not calculate a wind direction when the measured velocity falls below 0 05m s When configured for fixed field mode the direction output will hold the last setting until a new direction can be calculated If the unit is in CSV mode the output will switch between 0 and 5V or 4 to 20mA at the output rate selected Magnitude scaling The magnitude output can be scaled such that the full scale output represents either 30 50 or 60ms Error conditions If a valid wind measurement cannot be calculated for example all paths blocked the outputs 1 and 2 will switch between 0 and 5V or 4 to 20mA at the output rate selected Disabled outputs If the analogue outputs are not required they can be disabled Command S9 22 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 9 7 Configuration Diagnostic Information Each of these commands causes a response from the WindSonic Ttem Command Response D1 TYPE Y Type and serial No SERIAL NO 012345 Software version D2 FIRMWARE 2231 102 Unit configuration D3 M2 U1 01 L1 P1 B3 F1 H1 NA E3 D4 EXP MOD ID 0 Analogue ID if present D4 0 No Analogue outputs D4 1 Analogue outputs enabled D6 See Section 0 DELLEN 122 Self Test Still Air 10 METPAK SDI 12 WindSonic Refer to SDI 12 Commands amp Message Format CS215 Refer to Campbell Scientific User Guid
14. 082A PASS CHECKSUM CAL A9C1 A9C1 PASS For each of the Alignment and Gain tests a Pass or Refer to Manual message is generated For each of the Checksum tests a Pass or Fail message is generated 4 If any of the tests fail contact your supplier If a refer to manual message appears please see Section 12 3 Fault Finding Note that it will only pass if the specified temperature and zero wind conditions are met Check that there are no visible obstructions or damage to the unit before contacting Gill or your authorised distributor for further advice 12 3 Calibration A Custom calibration of the wind sensor is available from Gill Instruments This must be specified at the time of order or the unit must be returned to Gill Instruments for retrospective calibration Calibration of the temperature and humidity probe should be out in accordance with section 13 5 or 13 6 27 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 13 APPENDICES 13 1 Glossary amp Abbreviations Item Meaning CAL Calibration CR Carriage Return CSV Comma Separated Variable ENG Engineering FAC Factory KPH Kilometres per Hour LF Line Feed MPH Miles per Hour NMEA 0183 National Marine Electronics Association standard for interfacing version 3 marine electronic navigational devices NVM Non Volatile Memory PC IBM compatible Personal Computer ROM Read Only Memory RS232 Communications standard RS422 Communications standard RS485 Communic
15. 60m s Accuracy 2 12m s 2 12m s Resolution 0 01 m s 0 01 m s Wind Direction Range 0 359 0 359 or 0 539 Wraparound mode Accuracy 3 20m s 3 20m s Resolution 1 1 Output formats 0 5V 1 of full scale N B Analogue output impedance 1KQ Data Logger SDI 12 V1 3 Anemometer status Status OK and Error codes included as part of output message from RS232 programming port Environmental Moisture protection IP65 Temperature Operating 35 C to 70 C Storage 40 C to 80 C Humidity Operating lt 5 to 100 EMC EN 61000 6 3 Emissions EN 61000 6 2 Immunity Standards Manufactured within ISO9001 2000 quality system MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 4 3 Aspiration Shield Secification for model 41003 is contained in appendix 13 4 4 4 Temperature and Humidity Sensor Specification for the CS215 and HygroClip temperature and relative humidity are contained in appendices 13 5 and 13 6 4 5 Complete Assmebly Material White thermoplastic UV stabalized for long term weather resistance Gloss white painted aluminium mounting bracket with moulded plastic V block and satinless steel U bolt which allows it to be attached to a vertical pipe of any diameter between 25mm and 50mm Power Requirement Option 1 6 16Vdc 22mA typical Option 3 10 28Vdc 28mA typical Power requirements for the other options are dependant on temperature and relative humi
16. Beachtung der unten aufgef hrten Montage und Bedienungshinweise funktioniert der HygroClip ber Jahre hinweg perfekt Diese Anleitung gilt f r folgende F hler Typen HygroClip S S3 C C3 R R3 SC04 05 SPOS HP28 HS28 HK25 HK40 Technische Daten Einsatzbereich Feuchte 0 100 rF alle Typen Einsatzbereich Temperatur je nach F hlertyp Messbreich Feuchte 0 100 rF Temperatur je nach F hlertyp Ausg nge F hler Typ 40 85 C 0 4 0 85 V S C R SC04 SCO5 SP05 HP28 HS28 40 60 C 0 1V 3 C3 R3 One Wire DIO Alle Typen 1 Messwert nach 2 35 Messintervall 07s Versorgung 3 5 24 VDC ca 3mA Genauigkeit bei 23 C 1 5 rF 0 3K Pt100 1 3 DIN Ausgangsbirde gt 10 kOhm Schutzart IP65 ausser Sensor Filterbereich Hinweis W hlen Sie die Versorgungsspannung immer so klein wie m glich Montage alle steckbaren HygroClip Typen mit DATOS Stecker passend zu Hand ger ten und MOK Adaptern A Die 4 Markierungspunkte des HygroClip nach Zeichnung A ausrichten Den F hler mit leichtem achsialem Druck in die Aussparungen des Sicherungsringes stecken Danach nur den Sicherungsring im Uhrzeigersinn drehen Warnung Die HygroClip F hler sind steckbar Sie d rfen unter keinen Umst nden verdreht werden Das Verdrehen der Stecker zer st rt F hler und Gegenstecker und f hrt zu Garantieverlust Entfernen des HygroClip Zeichnung B durch Drehen des Sicherrungsringes im Geg
17. INSTRUMENTS MetPak User Manual Doc No 1575 PS 0001 Issue 2 Gill Instruments Limited Saltmarsh Park 67 Gosport Street Lymington Hampshire S041 9EG UK Tel 44 0 1590 613500 Fax 44 0 1590 613501 E mail anem gill co uk Website www gillco uk MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 Contents A A NN 1 FOREWORD Z INTRODUCTION AAA RO 3 3 PRINCIPLE OF OPERA TION becssccsscctsscsscnssevnsdetaccedessccssxstesdsbunstecssdetacsecosteadsatsvestssensvesasdesedsecosteavectsnestsbeasvesssessees 4 3 1 WIND SPEED AND DIRECTION SENSOR sssssccscccsssssseceeccecesssssceeeccessesaeeeccecessssseeseceeseessaeeeceeeeesssaeeeeeceseensaeeeeseeess 4 32 RADIATION SHEL Duoen e Senn Adi 5 3 3 TEMPERATURE AND RELATIVE HUMIDITY SENSOR ccccccessssssccccccessessceecccceesssseeeeecceseesseeeeceeeessseeececeseessaeeeeeeeeees 6 4 OPTIONS AND SPECIFICATION osoensis onosai iE TEE r N E E E T E E 7 4 1 OPTIONS AND PART NUMBERS AVAILABLE eienenn ieee ee eie aes eire ei eria ee iee e tes 7 4 2 SPECIFICATION 2 82 ini ne A RAR RARE RARE E 8 4 3 ASPIRATION SHIELD 85 55 sc2snwscec ves een censedse donee NEE PENE EEANN sv oussuwedanesendeenscenedsedasveccus cossesavansedcessevesuceosseseseouee 9 4 4 TEMPERATURE AND HUMIDITY SENSOR cccccccssssssecesccecessssseeececeseessaceccceeeesssaeeeeeceseessaeescceeeessaeeec
18. SDI 12 Transparent Mode on Edlog Array Based Datalogger CR10X using Control Port 8 and Changing SDI 12 Address from 0 to 1 10 2 Screen Capture of SDI 12 Transparent Mode on Edlog Table Based Datalogger using Control Port 8 and Changing SDI 12 Address from 0 to 1 11 3 Screen Capture of SDI 12 Transparent Mode on CRBasic CR200 Series Datalogger using Control Port C1 SDI12 and Changing SDI 12 Address from O to eines ana map ein ao av endear 12 4a Correct Fitment of Sensor Element Side View ooooooccoocccooncconnnonnncconaninnnonno 15 4b Incorrect Fitment of Sensor Element Side View 15 A 1 Normal Operating Conditions of RH Element un nennen A 2 1 Cable Connect Onis range ts han bniribiprenasin 3 A 1 Environmental TestS oocococonnconnonacnnononoonaananononnonononnonennononononnonencononononannnos A 1 CS215 Temperature and Relative Humidity Probe The CS215 Temperature and Relative Humidity probe is designed as a reasonably priced probe for general meteorological and other datalogging applications It utilises the SDI 12 communications protocol to communicate with any SDI 12 recorder simplifying installation and programming 1 Introduction The CS215 probe utilises a novel single chip element that incorporates both a temperature and RH sensor Each element is individually calibrated with the calibration corrections stored on the chip The element is easily changed in the field Rep
19. The WindSonic in MetPak is default Voltage 0 5 V representing wind speeds between 0 60m sec The HygroClip S3 humiduty and temperature sensor applied with MetPak option 3 only has a voltage output If the customer fit his own temperature and humidity probe with a current output the the mode of the Windsonic output can be changed to suit Important 1 When using the current outputs the load resistance between the Analogue outputs TB3 8 and 9 and Signal Ground must be lt 250 ohms including cable resistance This is to ensure that the voltage levels do not exceed SV 2 The current outputs must be returned to Signal Ground TB3 3 Polar and Tunnel modes If the WindSonic is configured in Gill Polar mode then the analogue outputs will be in Polar mode Polar mode direction wraparound The Wind Direction in Polar mode can be configured for either Standard 0 359 or Wraparound 0 539 options If the Standard mode is used with a chart recorder large swings of the recorder pen will be experienced each time the wind direction passes between 0 and 359 Using the Wraparound mode when the wind first changes from 0 to 359 there will be a step change on the output but after this first time the output will change smoothly each time the wind passes through 360 Similarly the first time the wind veers from 539 to 180 there will be a step change after which the output will change smoothly Wind direction Vol
20. al MODEL 41003 MULTI PLATE RADIATION SHIELD YOUNG INTRODUCTION The Multi Plate Radiation Shield protects temperature and rela tive humidity sensors from error producing solar radiation and precipitation This shield relies on a combination of plate geom etry material and natural ventilation to provide effective shielding INSTALLATION The Multi Plate Radiation Shield is designed to be mounted ona relatively flatopen area For best performance the shield should be placed in a location with the following characteristics Good air circulation around shield Away from large masses asphalt masts solar panels especially metal items Away from building exhaust vents electrical machiery and motors Away from standing water water fountains and sprinklers MAINTENANCE The Multi Plate Radiation Shield requires no regular maintenance innormal use Dirtor dust accumulation on the plates and bracket can be cleaned with soap and water Do not use solvents INSTRUCTION SHEET 41003 90 REV01 03 SPECIFICATIONS Sensor Mounting Mode 41003 includes universal adapter for sensors to 12 5mm diameter Model 41003P includes custom sensor adapter that can be machined for sensors to 26mm diameter Radiation Error 1080 W m intensity 0 4 C 0 7 F RMS 3 m s 6 7 mph 0 7 C 1 3 F RMS 2 m s 4 5 mph 1 5 C 2 7 F RMS 1 m s 2 2 mph Construction UV stabilized white thermoplastic plates Aluminum moun
21. ates to the MetPak specification at all times Avoid turbulence caused by surrounding structures that will affect the accuracy of the sensors such as trees masts and buildings The WMO make the following recommendations e The standard exposure of wind instruments over level open terrain is 10m above the ground Open terrain is defined as an area where the distance between the sensor and any obstruction is at least 10 times the height of the obstruction 6 2 Bench system test Note Prior to physically mounting the MetPak in its final location we strongly recommend that a bench system test is carried out to confirm the system is configured correctly is fully functional and electrically compatible with the selected host system and cabling preferably utilising the final cable length The required data format units output rate and other options should also all be set up at this stage 11 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 6 3 Electrical Cable Cable type Generic description Twisted pairs with drain wire screened with aluminised tape with an overall PVC sheath Wire size 7 0 2mm 24 AWG The table shows some suitable manufacturers references other manufacturers equivalents can be used ae gt Batt Application No of pairs Gill ref Belden ref oi ref Analogue 4 026 03156 9728 91199 SDI 12 2 9729 Cable length The maximum cable length is dependent on the ch
22. ations standard RXD Received Data TXD Transmitted Data SDI 12 Serial Data Interface standard for microprocessor based sensors WMO World Meteorological Organisation 13 2 Guarantee For terms of guarantee contact your supplier Warranty is void if the coloured security seal is damaged or broken or the transducer caps have been damaged 28 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 13 3 Electrical Conformity EC DECLARATION OF CONFORMITY INSTRUMENTS ACCORDING TO COUNCIL DIRECTIVE 89 336 EEC We Gill Instruments Ltd declare under our sole responsibility that the products WindSonic Ultrasonic Anemometer Manufactured by Gill Instruments Ltd to which this declaration relates are in conformity with the protection requirements of Council Directive 89 336 EEC on the approximation of the laws relating to electromagnetic compatibility This Declaration of Conformity is based upon compliance of the product with the following harmonised standards Emissions EN61000 6 3 Immunity EN61000 6 2 Excluding 150kHz to 250kHz Conducted Immunity ACA Sto Signed by A C R Stickland Director Date of issue 30 5 2003 Place of issue Gill Instruments Ltd Saltmarsh Park 67 Gosport Street Lymington SO41 9EG UK i oy i 509001 AENSTERES COMPANY 29 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 13 4 Model 41003 Multiplate Radiation Shield Manu
23. dity probes fitted Refere to manuals in appendices 13 5 and 13 6 Size and Weight Size 142mm dia x 380mm height instrument Weight 1 1kg including bracket For delivery the MetPak is contained in custom designed packaging All the packaging should be retained for use if the unit has to be returned at any time MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 5 INSTALLATION REQUIREMENTS Host system gt Access to the RS232 output from the MetPak is provided for programming and fault detection via a socket in the junction box This can be connected to a PC fitted with a RS232 interface and a suitable Terminal Emulation software package For example Hyperterminal for Windows 9x Windows 2000 and XP or Terminal for Windows 3 n will normally be available on your PC WindCom is available from the Gill website gt Equipment with input output compatibility to the WindSonic Option selected For example Chart recorder or Data logger using the WindSonic Analogue output or SDI 12 output Cable To connect between the MetPak and the host system See Section 7 3 Cable type for cable specification gt There are restrictions on the maximum cable lengths for correct operation gt The cable should be routed up the inside of the mounting tube Mounting Pipe gt Mounting fits vertical pipe 25 50mm 1 2in diameter See Figure 3 in section 7 4 Alignment amp Mounting Details gt For non hostile environments Alum
24. e It is only possible to use SDI 12 commands to program the wind temperature and humidity sensors Typical output for wind data is shown in Gill output message format WindSonic default mode identity 0 CS215 default mode identity 1 Note Campbell probe factory default is normally set to 0 to avoid conflict default has been set to 1 this applies to configurations preset by Gill and also by customers who fit their own Campbell probe 23 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 10 1 SDI 12 COMMANDS SDI 12 configured units comply with SDI 12 Interface Standard V1 3 We advise that you download the full SDI 12 specification from www sdi 12 org All other commands not defined below respond with a zero data values ready response All units of measurement are in metres per second This cannot be re scaled Name Command Gill Output Continuous Measurement aRO a lt dir gt lt mag gt lt status gt lt CR gt lt LF gt Polar Continuous Measurement aRl a lt U gt lt V gt lt status gt lt CR gt lt LF gt UV Continuous Measurement aRCO a lt dir gt lt mag gt lt status gt lt CRC gt lt CR gt lt LF gt with CRC Polar Continuous Measurement aRCI a lt U gt lt V gt lt status gt lt CRC gt lt CR gt lt LF gt with CRC UV Start Measurement aM a0003 Polar Send Data aDO a lt dir gt lt mag gt lt status gt lt CR gt lt LF gt Start Measurement UV aM1 a0003 Send Data aDO a
25. ecesensaeeeceeeenes 9 4 5 COMPLETE ASSMEBLY 5 INSTALLATION REQUIREMENTS sssssscscccecctelaccstessscsssaetoevesensVevesteaccstessocssbasensvesens eoecdesssddesscvcssbacenseseusstecoosetes 10 6 A Y AAA RO 11 6 1 INSTALLATION GUIDELINES sas 11 6 2 BENCH SYSTEM TEST wll 6 3 ELECTRICAL aan 2 CAD C 56 soci hides See eA hehe pew tN E AANE EE EEE EAA E NEN E E E E 12 Connecting the Analogue Output Option 3 and 4 uessensesersensennennennenennennensennennennenensensensennennennennnensensensennennenannn 13 Connecting to an SDI 12 Interface Option 1 and 2 uesussersersensenennenesnnenennenennennenennenensennennennennnennensensennennenannn 14 Connections to WindSonic cable for MetPak options 5 and 6 unussenseseesensennennennennennenennensennennennnensensennennennenann 15 T MECHANICAL etsesehsegessnnnstendeesssdneet nsesnnssenssennsnndenesnsdnerhsneegehesentsenntunsnenssss nettdneerensennsseneteundenssnednestensnneneennee 16 Orientation Gearan A aaa a a eve TON 16 PALI GHIMEN A EEEE ETA EA E EE EAE EEE 16 MONNE ae ann ii E RO ANSERA ENEE ARNE ER AA AEE OE 16 35 10H NINO AENA 18 8 1 GILL FORMAT POLAR CONTINUOUS DEFAULT FORMAT AVAILABLE FOR RS232 PROGRAMMING PORT 18 9 ANALOGUE OUTPUTS AND OPTIONS vsssesscscesisesesscsesoarssconssnssseontenteesoesasdsbanssenesensssnnsessesescvsasasvanssensssoesedien 20 9 1 ENTERING CONFIGURATION MODE SEE SECTION 8 1 FOR MESSAGE FORMAT cccssccceesseeeeseeee
26. eck the sensor is inserted in the correct orientation referring to Figure 4 Screw the filter back onto the end of the probe making sure it clears the sensor element If the element appears too close to the filter there is a fair chance that it has been inserted in the incorrect orientation or that the legs of the element have been bent Screw the filter onto the thread and tighten gently with your fingers CAUTION Only tighten the filter approximately Y of a turn by hand when the filter is fully screwed onto the thread Over tightening the filter will damage it and cause problems in inserting and removing the probe from some shields NOTE It is virtually impossible to touch and damage the sensing elements which are enclosed within the black moulded plastic at the tip However if during the process of handling the element dirt salt or grease are left on the plastic it may influence the measurements made User Guide Gold coloured y side of the tip Senking Element Sensing part of the element Printing on this side Gold pins Sensor ee connector sticking out of Tag the end of the tube Thread for the filter i Centre line of the sensor body and socket Figure 4a Correct Fitment of Sensor Element Side View Sensing part of element NOT on 7 x the centre line 7 Centre line of the sensor body Figure 4b Incorrect Fitment of Sensor Element Side View 15 Ap
27. econds to execute User Guide NOTE The sensor will automatically switch to making measurements every 11 seconds 2 second the first time it receives the aR command It makes the measurements based on its own clock which is not synchronised to the datalogger clock This means measurements made with the aR command can be up to 11 seconds old This is not normally an issue for environmental measurements due to slow rate of change of RH and temperature If measurements are requested more frequently than every 11 seconds the aR option will return repeated values up until the point the sensor takes its next measurement If measurements are requested at 2 sec or faster the sensor will increase its measurement rate to approximately every 5 seconds The automatic measurement mode can only be cancelled by powering down the sensor to reset it e aMC aCC aRC which are the same as the instructions above but where the C at the end of the instruction forces a validation for the data received from the sensor using a checksum If the checksum is invalid the logger will re request the data up to three times The checksum validation increases the measurement time by about 40 milliseconds if there are no errors Retries will increase the measurement time in proportion to the number of retries Use of the checksum option is only normally necessary for very long cable runs Where in all cases a is the address of the sensor and is the comma
28. enuhrzeigersinn bis zum Anschlag dann HygroClip abziehen Das Montagest ck MOC wird mit der mitgelieferten Mutter befestigt Bohrdurchmesser 12 5 mm Nur HygroClip C und C3 Elektrischer Anschluss nach Zeichnung C Justierung Kalibrierung Die HygroClip F hler sind ab Werk justiert Wir empfehlen je nach Einsatz eine Kalib ration nach ca einem Jahr Dies kann vor Ort mit ROTRONIC SCS Feuchte Standards durch Sie erfolgen Die Handhabung entnehmen Sie der den Feuchte Standards beiliegenden Anleitung Betreffend Justierung kontaktieren Sie ROTRONIC oder unsere Vertretung A 4 Chee TI menes EEE Pe gt verriegelt verrouill locked ay TECHNIK FOR Proris messger te gmbh www rotronic de Einsteinstrasse 17 23 D 76275 Ettlingen Telefon 49 7243 383 250 Fax 49 7243 383 260 instruments uk Itd www rotronic co uk Unit IA Crompton Fields Manor Royal Crawley West Sussex RH10 9EE Phone 44 1293 57 10 00 Fax 44 1293 57 10 08 Grindelstrasse 6 CH 8303 Bassersdorf Telefon 41 44 838 11 11 Fax 41 44 837 00 73 ROTRONIC HyGrRoCuip Your ROTRONIC HygroClip is a product designed according to ISO 9001 and meeting EN50082 2 EMC requirements The HygroClip works perfectly for many years when you follow the assembly and operating guidelines These instructions are valid for the following probe types HygroClip S S3 C C3 R R3 SC04 05 SP05 HP28 HS28 HK25 HK40 Technical Data Operating range humidi
29. eseeeeeeeeeeseeeeeaee 21 9 2 RETURNING TO MEASUREMENT MODE ssssssccccceessssseeecceeeesssseceeceeseesseeeceeseesas ev 9 3 CHECKING THE CONFIGURATION 0 5 21 9 4 CHANGING A SETTING 21 9 5 POWER UP MESSAGE 21 9 6 ANALOGUE SETTINGS lt td aladdin ti deTeedabebocatience ae a eara a 22 Polar mode direction caldul tion sa aarussssesuniiinasunsnsnsaunnbsksannunansannehsinnauneineinseite 22 Magnit de scaling sau nenn Run A EAE EEE REEE EERO ERE a aA 22 Error CONAN E AEE E AEEA E E E EEA NR A A E E N AE E A 22 Disabled outputs uueseesersersennennennenersennennennennennenennn 9 7 CONFIGURATION DIAGNOSTIC INFORMATION ccccccssssssseeeceeesessceecceceesssaeeeccesessseeecceseesssaeeeeceesesssseeesceseesaeees 10 AS AAA A O 10 1 SDI 12 COMMANDS 5 9056 ocen5 seca 205s assdead tse dta sin asa ae Sao eH aA aed R A 11 MAINTENANCE FAULT FINDING cccscsssssssscsssscsssscnccssccssssssnsscssccssscssnsscnscsssssssssscnsscnesssscsosssscnssssses 11 1 CLEANING edad ido e a29 11 2 SERVICING 006 28 11 3 FAULT FINDING 5229 11 4 RETURNING UNIT 26 11 5 STATUS ERROR CODES 26 12 PETS ON 27 12 1 BENCH DES Ts u ON 27 12 2 SELF TEST STILL AIR FOR WINDSONIC WIND SENSOR csscccesssceeesseeeeeeeeeeeneeeceeaeeceeaeeceeaeeeeeseeeceeaeeeeeeneeenss 27 12 3 CATIBR TION 2 53 si aan RoR Ma oh AA vals RSC SA sea Raa Bebe teas 27 13 APPENDICES 3isssecassscsdessccescesbessodeetetedennsss cosede
30. h such cable distances of several hundred metres to the sensor are possible e Ensure that the power ground cable has low resistance and is connected to the same ground reference as the datalogger control ports e Be aware that daisy chaining sensors reduces the maximum cable length roughly in proportion to the number of sensors connected in parallel CS215 Temperature and Relative Humidity Probe e If long cabling or interference appear to be causing spurious readings then use the checksummed versions of the SD1 12 instructions see section 5 4 2 Power Conservation The CS215 draws about 120u A of current between measurements from its 12V supply When used with the aR command the average consumption increases by 50uA over the quiescent as the sensor automatically takes a measurement every 10 seconds In most applications these levels of current consumption are insignificant compared to the datalogger and other power uses so the sensor can be permanently connected In very low power applications battery power can be conserved by turning the 12V supply to the CS215 on just before the measurement allowing a warm up time of at least 100 ms and then turning it off afterwards This switching can be achieved in different ways depending on the type and model of your datalogger If available the switched 12V output of the datalogger can be used or a small power switch or solid state relay such as the PSW12 5 Reading
31. haracter for the new address sent as extended commands using P68 Refer to the datalogger manual for further details CRBasic programs can simply have the aAb string embedded as the string command However changing the address using a program is not the simplest way as most Campbell Scientific dataloggers with the exception of the CR5000 at present support a method of directly interacting with SDI 12 sensors via a terminal emulator This allows you to get confirmation that the change of address has worked using the command This can be done using a computer running LoggerNet to issue any valid SDI 12 command through the datalogger to the CS215 as described in the following sections 6 1 Array Based Edlog Dataloggers 1 Connect a single CS215 to the datalogger using Control Port p as described in Section 4 and download a datalogger program that contains the SDI 12 Recorder Instruction 105 instruction with valid entries for each parameter 2 Inthe LoggerNet Toolbar navigate to and activate the TestITerminal Emulator menu The Terminal Emulator window opens In the Select Device menu located in the lower left hand side of the window select the station 3 Click on the Open Terminal button If communications between the datalogger and PC are successful the red bar located in the upper left hand side of the window will turn green 4 Click inside the Terminal Emulator window and press the lt enter gt key un
32. he red bar located in the upper left hand side of the window will turn green Press the lt enter gt key until the datalogger responds with the CR200 gt prompt To query the CS215 for its current SDI 12 address press the lt enter gt key at the CR200 gt prompt make sure you are in Caps Lock mode and enter the command SDI 12 gt and press the lt enter gt key The CS215 will respond with the SDI 12 address To change the SDI 12 address press the lt enter gt key At the CR200 gt prompt enter the command SDI12 gt aAb Where a is the current address shown in the above step and b is the new address The CS215 will change its address and the datalogger will exit the SDI 12 Transparent Mode and respond with Fail Verify the new SDI 12 address Press the lt enter gt key at the CR200 gt prompt enter the command SDI12 gt and press the lt enter gt key The CS215 will respond with the new address Invalid CR200 gt SDI12 gt 0 CR200 gt SDI12 gt 2 1 2_ Terminal Emulator E xj CR200 gt SDI12 gt 0A1 Fail Select Device Baud Rate CR200Series y I All Caps Mode 115200 zi Close Terminal Clear Help FIGURE 3 Screen Capture of SDI 12 Transparent Mode on CRBasic CR200 Series Datalogger using Control Port C1 SDI12 and Changing SDI 12 Address from 0 to 1 12 User Guide 6 4 CR1000 Datalogger The CR1000 datalogger supports changing the address
33. in a similar way to the CR200 In terminal mode issue the SDI12 command then give the control port number in response to the prompt The normal SDI12 command can then be entered there is no need to enter the SDI12 prefix as with the CR200 If changing the address you will initially get a fail message as with the CR200 To terminate the SDI12 terminal mode press lt enter gt with no command 7 Maintenance and Calibration The CS215 probe requires minimal maintenance e Check monthly to make sure that the radiation shield is free from dust and debris e Clear the white filter on the end of the sensor of similar debris If dirt or salt is engrained into the filter it should be cleaned with clean water or more simply replaced Make sure the filter is done up firmly with your fingers do not over tighten The life of the humidity chip element is quoted as many years with a typical drift of less than 1 per year when used in clean environments However as it can be difficult to define what the sensor has been exposed to and as the element of the CS215 is relatively low cost we recommend that you replace the sensor element at the normal interval you would recalibrate similar probes e g annually Replacing the element effectively brings the probe back to a factory calibration state both for temperature and RH without incurring a costly return for factory calibration If you wish to have an old element s calibration chec
34. inium tube can be used gt For hostile environments you should select a material suitable for the intended environment For example stainless steel 316 for marine use 10 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 6 INSTALLATION Do NOT remove the black rubber transducer caps Warranty is void if the coloured security seal is damaged or removed 6 1 Installation Guidelines The MetPak has been designed to meet and exceed the stringent standards listed in its specification As with any sophisticated electronics good engineering practice should be followed to ensure correct operation e Always check the installation to ensure the MetPak is not affected by other equipment operating locally which may not conform to current standards e g radio radar transmitters boat engines generators etc Guidelines o Avoid mounting in the plane of any radar scanner a vertical separation of at least 2m should be achieved o Radio transmitting antennas the following minimum separations all round are suggested VHF IMM Im MF HF 5m Satcom 5m avoid likely lines of sight e Use cables recommended by Gill If cables are cut and re connected incorrectly perhaps in a junction box then EMC performance may be compromised if cable screen integrity is not maintained e Earth loops should not be created wire the system in accordance with the installation guidelines e _ Ensure the power supply oper
35. ion Response Time with Filter Environmental Performance Calibration Traceability Supply Voltage Current Consumption Diameter Length Housing Material Filter Material Standard Cable Length EMC Compliance lt 15s 63 response time in air moving at 1 m s at humidities lt 85 See Appendix A See Appendix B 6 16V DC Typically 120uA quiescent 1 7mA during measurement takes 0 7 sec 12mm at sensor tip maximum 18mm at cable end 180mm including cable strain relief Anodised aluminium Sintered high density polyethylene average pore size 10 13 um 3 metres Tested and conforms to IEC61326 The CS215 is normally housed inside a model 41303 5 Solar Radiation Shield when used in the field To fit the CS215 in the Radiation Shield 1 Loosen the plastic split collar at the base of the shield reversing the removable portion if necessary and gently insert the probe 2 Tighten the collar until it lightly grips the probe body 3 Continue to push the probe up into the body of the shield until the step in the tube stops it going any further 4 Tighten the collar until it securely grips the probe 5 Loop the cable up and fix it to the underside of the radiation shield arm or to the mounting tower using a cable tie CAUTION Failure to secure the cable can lead to breakage of the wires due to fatigue if the cable is allowed to blow to and fro in the wind The probe will also fit int
36. ked so that you can formally record the probe s pre calibration state pre sensor replacement you can measure its performance by plugging it into another sensor For a fee Campbell Scientific can provide a certificate of performance for sensor elements returned to them 7 1 Sensor Element Replacement To replace the element 1 Disconnect the sensor from the 12 V power supply 2 Remove the filter by unscrewing it in an anti clockwise direction when looking towards the tip of the sensor CAUTION The filter cap unscrews from the probe Attempting to pull if off will destroy it 3 Identify the sensor element referring to figure 4a below which shows a side on view of the end of the probe and sensor element Before removing the element carefully study the probe and note its orientation after reading the following description e The element plugs into the black plastic socket that protrudes by about 1mm from the end of the metal body of the sensor e There are eight holes in the socket whilst the element only has four pins in line e The element will work when fitted into either side of socket but must be installed in one of the two possible orientations to work 13 CS215 Temperature and Relative Humidity Probe 14 e The correct orientation is with the black moulded tip of the element that contains the sensing components mounted directly above the centre of the socket e Figure 4a shows the correct orientation whilst
37. l be in NMEA format Note The factory set default settings are shown in bold in the following sections 9 5 Power up Message Item Options Command ON H1 OFF H2 Power up message If the power up message is On then a message similar to that below is output whenever the unit goes into Measurement mode ie when power is first applied to the unit or unit returns to Measurement mode from Configuration mode WINDSONIC Gill Instruments Ltd FIRMWARE 2231 102 RS232 CFG RS 232 Auto CHECKSUM ROM 9F5E 9F5E PASS CHECKSUM FAC 04F4 04F4 PASS CHECKSUM ENG 0824 0824 PASS CHECKSUM CAL DFBF DFBF PASS If any of these checksums fail the word PASS will be replaced by FAIL and the unit will use its default settings If the checksum repeatedly fails after power on contact your supplier NOTE The exact message above will vary 21 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 9 6 Analogue settings The following commands are only valid for the MetPak Options 4 and 6 Item Options Command Voltage or Current 0 5 Volt T1 output 4 20 mA T2 0 30 metres per second S3 Analogue 0 50 metres per second S5 range 0 60 metres per second S6 Analogue output disabled S9 Analogue Wrap around 0 539 Cl Wraparound Normal 0 359 C2 Where Analogue range Customer selected range Analogue wraparound Chart recorder facility Voltage or current
38. lacement of the element effectively fully recalibrates the probe both for temperature and humidity thus minimising downtime and calibration costs Low power electronics within the CS215 controls the measurement made by the sensor element applies temperature and linearization corrections to the readings and presents the data via an SDI 12 compatible interface to a datalogger The probe is fitted with a custom made yet low cost sintered plastic filter to minimise the effects of dust and dirt on the sensor The filter is lightweight and hydrophobic thereby minimising the effect of the time response of the sensor The probe housing is designed to withstand permanent exposure to all weathers and to fit into a range of radiation shields including lower cost compact shields 2 Specifications Sensing Element Sensirion SHT75 Temperature Measurement Operating Range 40 C to 70 C Accuracy 0 3 C at 25 C 0 4 C over 5 C to 40 C 0 9 C over 40 C to 70 C Response Time With 120 s 63 response time in air moving at m s Filter Default units Degrees celsius Relative Humidity Operating Range 0 to 100 RH 20 to 60 C see Appendix A Accuracy at 25 C 2 over 10 90 4 over 0 100 Short Term Hysteresis lt 1 RH Temperature Compensated to better than 2 over 20 to 60 C Dependence Typical Long Term Better than 1 0 per year Stability CS215 Temperature and Relative Humidity Probe 3 Installat
39. late changing wind direction and to check that both axes are functioning The fan will also give output changes from the temperature and relative humidity sensor Note that this a quick functional test 7 Use the office fan to check the temperature and humidity sensor output 12 2 Self Test Still Air for WindSonic wind sensor This test checks Alignment Gain Checksums and can be performed from the RS232 port inside of the junction box Alignment tests The unit performs a transducer geometry check and compares the result with its factory setting Gain tests The unit performs a check of its operating gain against its factory setting Checksum tests The unit performs a check of its program and data memory Important This test is a stringent laboratory test which will only be passed if carried out under still air conditions at room temperature 17 23 C NOTE Option 4 SDI 12 units must be in Gill mode for the self test to function see Section 11 1 Place a cardboard box over the WindSonic sensor 2 Go into Configuration Mode ENTER 3 Carry out the Self test by entering D 6 ENTER A message similar to that shown below will be generated ALIGNMENT LIMITS 0D59 0CF5 ALIGNMENT U OD15 PASS ALIGNMENT V OD16 PASS GAIN 0 0001 PASS GAIN 1 0001 PASS GAIN 2 0001 PASS GAIN 3 0001 PASS CHECKSUM ROM AB7D AB7D PASS CHECKSUM FAC 04F4 04F4 PASS CHECKSUM ENG 082A
40. logue Channel 1 Analogue Channel 2 Signal Ground Chassis ground 9 DC Power supply 10 28 Vdc Data logger or Chart recorder Signal names Temp R H MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 TBI TB3 1X 1 Tol v RX 2 9 va F f sic GND 3 8 WIND DR Connections from WindSonic N C 4 7 WIND SP SCREEN 5 6 c N C 6 5 SCREEN x 17 RS 232 4 6 lt I H 18 3 I SIG CND Outputs to Customer Equipment v d9 2 X V 10 X ve 1 c SIG GND 1 2 5 R H Connections from Rotronic plo 3 4 EMP Temperature and Relative Humidity fa vn Y Sensor scREEN l 6 SW OFF 1l n c TB2 TB4 Connecting to an SDI 12 Interface Option 1 and 2 TB 3 Junction Box Damla 10 Way Terminal Block Chart recorder Signal names Signal names Signal GND 3 V 10 GND V 9 1 V 9 16 V dc TXD 1 RXD 2 Serial Data Line Note SW1 must be in closed position for SDI 12 option If you wish to use RS232 programming port then SW1 must be in open position and only external links between TXD and RXD must be temporarily removed TB1 TB3 TX 1 oj yv RX 2 91 Vi SIG GND 3 8 n c N c 4 7 N C Connections from WindSonic sc A 6 N C a c 5 SCR lt y 7 RS 232 4 N C D 8 3 SIG GND Ouputs to Customer V 9 2 H SDI12 SIG Pannen v 10 INK quip ES
41. lt U gt lt V gt lt status gt lt CR gt lt LF gt Start Measurement with aMC a0003 CRC Polar Send Data aDO a lt dir gt lt mag gt lt status gt lt CRC gt lt CR gt lt LF gt Start Measurement with aMCl a0003 CRC UV Send Data aDO a lt U gt lt V gt lt status gt lt CRC gt lt CR gt lt LF gt Acknowledge Active a a lt CR gt lt LF gt Send Identification al al3GillInst 1405 SDI v101 lt sn gt lt CR gt lt LF gt Change Address aAb b lt CR gt lt LF gt Address Query a lt CR gt lt LF gt Measurement on demand aXLPM OLP lt CR gt lt LF gt Continuous Reading aXHPM OHP lt CR gt lt LF gt Default operating mode aXHPM Gill Output Message Format al3GillInst 1405 SDIv A E SDI 12 SDI 12 Gill Gill WindSonic address version Model No Software Version Software Version may change 24 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 11 MAINTENANCE amp FAULT FINDING 11 1 Cleaning If there is any build up of deposit on the unit it should be gently cleaned with a cloth moistened with soft detergent Solvents should not be used and care should be taken to avoid scratching any surfaces The unit must be allowed to defrost naturally after being exposed to snow or icy conditions do NOT attempt to remove ice or snow with a tool Do NOT remove black rubber transducer caps 11 2 Servicing There are no moving parts or user serviceable parts requiring routine maintenance in the
42. n a naturally aspirated radiation shield The instrument uses a rugged U bolt mounting clamp that attaches to any vertical pipe up to 50mm diameter The instrument can be supplied with either a digital SDI 12 output or a four channel analogue output option suitable to interface to industry standard data loggers Low power consumption enables the instrument to be used in remote locations where power is at a premium An electrical junction box is fitted to the mounting bracket allowing convenient termination of all electrical cables MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 3 PRINCIPLE OF OPERATION 3 1 Wind Speed and Direction Sensor The MetPak uses the WindSonic wind speed and direction sensor WindSonic measures the times taken for an ultrasonic pulse of sound to travel from the North transducer to the South transducer and compares it with the time for a pulse to travel from S to N transducer Likewise times are compared between West and East and E and W transducer If for example a North wind is blowing then the time taken for the pulse to travel from N to S will be faster than from S to N whereas the W to E and E to W times will be the same The wind speed and direction can then be calculated from the differences in the times of flight on each axis This calculation is independent of factors such as temperature TIME OF FLIGHT THEORY Therefore L Distance between transducer faces C Speed of sound V Velocity
43. nd terminator These two characters are normally sent implicitly by Campbell Scientific dataloggers 5 1 Measurements at Slow Scan Rates If the scan interval of your measurement sequence is very long ten seconds or more use the aM command to read the data from the sensor in the main table or scan for CRBasic as this minimises power consumption If the scan rate has to be faster please use the aR command or refer to section 5 2 below With Campbell Scientific dataloggers the aM command involves just one program instruction The datalogger sends the command waits and then automatically sends the aDO get data command and collects the measurements from the sensor The sensor returns two values for all measurement instructions the temperature in degrees Celsius and the relative humidity as a percentage 0 100 When using Shortcut for Windows if your version does not include specific support for the CS215 use the generic SDI12 Sensor instruction setting the first parameter label to represent temperature and the second for relative humidity 5 1 1 Example Program for CR10X Measuring Two Sensors Every Minute The program below shows a simple example reading two sensors which have been set up with different addresses one address 0 the default and the other with address 1 both are connected to control port 1 This example uses the aM command for minimum power consumption CS215 Temperature and Relative Humidity Probe
44. nd identify a suitable landmark and its bearing Mounting Pass the cable through the pole Note the customer must fit appropriate strain relief to the cable It is the responsibility of the customer to ensure that the MetPak is mounted in a position clear of any structure which may obstruct the airflow or induce turbulence Do NOT mount the MetPak in close proximity of high powered radar or radio transmitters A site survey may be required if there is any doubt about the strength of external electrical noise Temperature ans Humidity Probes should be mounted inside of the Radiation Shield as shown in Fig 3 section 3 3 16 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 142mm dia Measurement Plane Figure 4 Alignment amp Mounting details 17 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 8 OUTPUT FORMATS On applying power to the MetPak it will provide measurements in one of the following formats gt SDI 12 Polled See section 10 gt Analogue Voltage output see section 9 An RS232 port is located inside of the junction box to enable changes to be made to the factory settings or for fault diagnostics You will need to connect a PC to the RS232 programming port You can communicate with the anemometer using the free Gill WindCom software or using Hyperterminal Instructions for using Hyperterminal are given below Configuring Hyperterminal Note Other terminal emulato
45. o the larger shields designed to take the HMP45C probe e g the URS1 or 41003 shields The 41003 shield requires either the 18 5 or 12 mm adaptors parts 41046 DS 18 and 41046 DS 12 respectively For those shields tighten the sensor grip around the probe where it best matches the size of the grip The probe will also fit directly into most of the R M Young shields where the probe enters the shield from the top pointing downwards CAUTION When removing the probe from the shield withdraw it from shield carefully as the foam filter is easily broken if bent or knocked 4 1 Long Cables User Guide For other applications the CS215 can be installed in any orientation Install it away from obvious sources of heat including solar radiation NOTE It is best to protect the filter at the top of the sensor from direct exposure to liquid water as whilst the hydrophobic nature of the filter will repel light rain driving rain can force itself into the pore structure of the filter and then take some time to dry out The CS215 can be used with all Campbell Scientific and most other dataloggers that support the SDI 12 protocol Wiring colours and connections between the sensor and a datalogger are shown in Table 1 Table 1 Cable Connections Cable Description Connection to Datalogger Red 12V supply 12V White amp Black Power ground G Green SDI 12 signal Control port or dedicated SDI 12 port CR5000 Screen Power
46. onds PanelTemp PTemp 250 Battery Batt_volt NextScan SlowSequence Define Data Table to hold the stored data for the CS215 DataTable CS215dat 1 1 DataInterval 0 15 min 1 As an example store the average every 15 mins Average 1 Temperature IEEE4 False Average 1 RH IEEE4 False EndTable Scan 30 sec 0 0 SDI12Recorder CS215meas 1 0M 1 0 0 Read the values into the array NOTE we had to use a control port for SDI 12 as this is in the slow sequence User Guide CallTable CS215dat Call the table instruction to calculate and store the Avs NextScan EndProg 6 Changing the SDI 12 Address Using LoggerNet and a Datalogger Up to ten CS215s or other SDI 12 sensors can be connected to a single datalogger control port Each SDI 12 device must have a unique SDI 12 address between 0 and 9 The CS215 also supports the extended range up to 126 The factory set SDI 12 address for the CS215 is 0 The CS215 SDI 12 address is changed in software by issuing the aAb command to the CS215 over the SDI 12 interface where a is the current address and b is the new address The current address can be found by issuing the command The address can be changed by loading a program into the datalogger that includes an instruction to send the aAb command to change the sensor from its old address a to new address b For an Edlog datalogger this involves using instruction 105 with the character A followed by the c
47. osen communication The table shows the typical maximum lengths at the given baud rates using the recommended cable If any problems of data corruption etc are experienced then a slower baud rate should be used Alternatively a thicker or higher specification cable can be tried WindSonic Option Communication format Baud rate Max cable length Option 3 and 4 Analogue Voltage o p N A 6 5 m 20 ft Option 1 and 2 SDI 12 1200 90m 300ft Power supply The MetPak requires a DC supply of between 9 16vdc E 22ma typical if fitted with a Campbell CS215 temperature and relative humidity probe or 10 28vdc E 28ma typical if fitted with a Rotronic HygroClip S3 probe 12 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 Connecting the Analogue Output Option 3 and 4 Remove junction box lid Analogue output wind TB3 Junction Box 10 Way Terminal Block Termi Signal names nal nos Analogue Channel 1 Wind Speed 7 Analogue Channel 2 8 Wind Direction Signal Ground Do NOT connect N A at this end V supply V supply 10 Cable 5 twisted pairs C OOO gt Z H Screen and drain wires Note SW1 must be in open position Analogue outputs for temperature and relative humidity TB4 Junction Box 6 Way Terminal Block Termi Signal names nal nos Temp 4 R H 5 13 Data logger or Chart recorder Signal names Ana
48. ou can do this yourself on site with the ROTRONIC SCS humidity standards Please refer to the instructions included with the humidity standards Contact ROTRONIC or our representatives concerning adjustment yellow white Colour brown green black yellow white retome sarl www rotronic fr 56 Bld de Courcerin F 77183 Croissy Beaubourg Tel 33 1 60 95 07 10 Fax 33 160 17 12 56 perironie instrument corp www rotronic usa com 160 East Main Street Huntington N Y 11743 USA Phone 1 631 427 38 98 Fax 1 631 427 39 02 brown Function Temperature VDC GND DIO rel humidity brun vert noir jaune blanc www rotronic com 13 6 Campbell CS215 Temperature and Humidity PROBE manual 12 0655 0007 CS215 Temperature and Relative Humidity Probe User Guide Issued 8 6 06 Copyright 2004 Campbell Scientific Ltd CSL 324 Guarantee This equipment is guaranteed against defects in materials and workmanship This guarantee applies for twelve months from date of delivery We will repair or replace products which prove to be defective during the guarantee period provided they are returned to us prepaid The guarantee will not apply to e Equipment which has been modified or altered in any way without the written permission of Campbell Scientific e Batteries e Any product which has been subjected to misuse neglect acts of God or damage in transit Campbell Scientific will return guaranteed equipmen
49. over High levels of pollutants may cause permanent damage to the sensing polymer As a general rule the sensor will not be damaged by levels of chemicals which are not too dangerous to human health see table A 1 so damage is not normally a problem in outdoor applications Avoid exposing the sensor to chemicals at higher concentrations A 3 Operating Range of RH Element The RH sensor is specified to work over the entire humidity range of 0 100 RH for the temperature range 20 to 60 C It will give readings over an extended range as shown in the figure below although the electronics of the CS215 probe are not specified to operate beyond 70 C CS215 Temperature and Relative Humidity Probe 100 zu When used outside the range of normal conditions or when subject to prolonged periods of condensation or freezing the sensor calibration may be temporarily altered normally resulting in a change of lt 3 RH Once the sensor returns to normal conditions the calibration will settle back over the course of several days to the standard calibration In laboratory conditions it is possible to speed up this process by a reconditioning process as follows 80 90 C at lt 5 RH for 24h baking followed by 20 30 C at gt 74 RH for 48h re hydration N Maximum operating pa conditions Normal operating conditions 40 20 O 20 40 60 80 100 120 Figure A 1 Normal Operating Conditions of RH Element A 4 Measurement belo
50. pendix A Environmental Performance This Appendix details tests and limitations of the sensor when exposed to extremes of the environment A 1 Tests to Defined Standards The sensor element has been tested by the manufacturer and found to comply with various environmental test standards as shown in the table below Table A 1 Environmental Tests Environment Norm Results Temperature JESD22 A104 B Within Specifications Cycles 40 125 C 1000 cycles HAST Pressure JESD22 A110 B 2 3bar Reversible shift by 2 RH Cooker 125 C 85 RH Salt Atmosphere DIN 50021SS Within Specifications Condensing Air Within Specifications Freezing cycles 20 90 C 100cycles Reversible shift by 2 RH fully submerged 30min dwell time Various DIN 72300 5 Within Specifications Automotive Chemicals Cigarette smoke Equivalent to 15years in a Within Specifications mid size car N B The temperature sensor passed all tests without any detectable drift Package and electronics also passed 100 A 2 Exposure to Pollutants All capacitative sensors are susceptible to pollutants to some degree The vapours may interfere with the polymer layers used in the structure of the sensing element The diffusion of chemicals into the polymer may cause temporary or even permanent shifts in both offset and sensitivity After low levels of exposure in a clean environment the contaminants will slowly outgas and the sensor rec
51. rs are configured in a very similar way Decide on an available Com port that you want to use Usually Com1 Run Hypertrm exe Typically accessed via Start gt Programs gt Accessories gt Hyperterminal Create a New Connection File gt New Connection Entera Name eg WindSonic 1 Change Connect Using to Direct to Com 1 or other Com port as required Oy Ot E ee Adjust the Port settings to match WindSonic settings WindSonic default settings are Bits per second 9600 Data bits 8 Parity None Stop bits 1 Flow Control Handshaking None Example of message format Q 229 002 74 M 00 106 See Section 10 Configuring if you need to change WindSonic settings This RS232 programming port will default to Gill Polar Continuous Details are available in section 8 1 this allows changes to the analogue output factory setting 8 1 Gill format Polar Continuous Default format available for RS232 programming port lt STX gt Q 229 002 74 M 00 lt ETX gt 06 aaa WindSonic Wind Wind Units Status Check node address direction speed sum Where lt STX gt Start of string character ASCII value 2 WindSonic node address Unit identifier Wind direction Wind Direction Wind speed Wind Speed Units Units of measure knots m s etc Status Anemometer status code see Section 12 5 for further details lt ETX gt End of string character ASCII value 3 Checksum This i
52. s the EXCLUSIVE OR of the bytes between and not including the lt STX gt and lt ETX gt characters 18 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 WindSonic node address The default setting is Q do not change this Wind direction Indicated in degrees from 0 to 359 with respect to the WindSonic North marker In fixed field mode and when the wind speed is below 0 05 metres sec the direction will not be calculated but the last calculated direction above 0 05 m s will be output Wind speed and units Shows the Wind Speed in the Measurement Plane See Figure 3 Alignment amp Mounting details in one of the following units Units Identifier Metres per second default M Knots N Miles per hour P Kilometres per hour K Feet per minute F Status This indicates either Correct operation Code 00 or Error codes See Section 12 5 for explanation of codes Checksum This enables the host system to check that the data has been correctly received This is the EXCLUSIVE OR of the bytes between and not including the lt STX gt and lt ETX gt characters Output rate not displayed The WindSonic samples continuously and delivers wind information at 1 default setting 2 or 4 outputs second 19 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 9 ANALOGUE OUTPUTS AND OPTIONS Note Analogue outputs are only available from MetPak Option 4 and 6 Voltage output
53. senseosaes sodeovededenessecesesecendcosedssoaesveceieenssoseseseseaseesesesoasseeceiecessoessesestess 28 13 1 GLOSSARY s ABBREVIATIONS ss netten li iaalsssteseslastseeeslsetsissnseddsilusne nn 28 13 2 GUARANTEE Be 13 3 EREGTRICAL CONFORMITY een ave hata EE LAVA LAAT OE RRS 13 4 MODEL 41003 MULTIPLATE RADIATION SHIELD MANUAL ccccccessssseececeessessececceeeesssaeeeceeesssaseeeecesessssaeeeeess 13 5 ROTRONIC HYRGOCLIP S3 MANUAL ccccssseccceceessssceeceeeeessseeeceeeeees 13 6 CAMPBELL CS215 TEMPERATURE AND HUMIDITY PROBE MANUAL MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 1 FOREWORD Thank you for purchasing the MetPak manufactured by Gill Instruments Ltd To achieve optimum performance we recommend that you read the whole of this manual before proceeding with use Do NOT remove black rubber transducer caps Gill products are in continuous development and therefore specifications may be subject to change and design improvements without prior notice The information contained in this manual remains the property of Gill Instruments and should not be copied or reproduced for commercial gain 2 INTRODUCTION The Gill MetPak is a compact and lightweight multi sensor instrument that measures the most essential weather parameters Gill ultrasonic technology as used in the proven WindSonic instrument measures wind speed and direction Temperature and humidity are measured using industry standard probes housed i
54. sue 02 CN 4240 March 2006 11 4 Returning unit If the unit has to be returned it should be carefully packed in the original packaging and returned to your authorised Gill distributor with a full description of the fault condition 11 5 Status error codes The Status code is sent as part of each wind measurement message These can be observed on the RS232 programming port inside of the junction box Code Status Condition 00 OK Sufficient samples in average period 01 Axis 1 failed Insufficient samples in average period on U axis 02 Axis 2 failed Insufficient samples in average period on V axis 04 Axis and 2 failed Insufficient samples in average period on both axes 08 NVM error NVM checksum failed 09 ROM error ROM checksum failed 10 System gain at Wind velocity reported could be in error maximum A NMEA data valid Vv NMEA data invalid 26 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 12 TESTS 12 1 Bench test 1 Couple the MatPak to the host system and power supply using a known working test cable 2 Check that the unit is correctly configured by going into Configuration mode and using D3 See Section 9 3 Checking the configuration 3 Check for normal output data and that the Status Code is OK 00 Tf the status code is other than 00 refer to Section 11 5 Status error codes 5 Use an office fan or similar to check that the unit is sensing wind turning the unit to simu
55. t by surface carrier prepaid Campbell Scientific will not reimburse the claimant for costs incurred in removing and or reinstalling equipment This guarantee and the Company s obligation thereunder is in lieu of all other guarantees expressed or implied including those of suitability and fitness for a particular purpose Campbell Scientific is not liable for consequential damage Please inform us before returning equipment and obtain a Repair Refer ence Number whether the repair is under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for re pairs can be given on request When returning equipment the Repair Reference Number must be clearly marked on the outside of the package Note that goods sent air freight are subject to Customs clearance fees which Campbell Scientific will charge to customers In many cases these charges are greater than the cost of the repair ea Campbell scientific Campbell Scientific Ltd Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UK Tel 44 0 1509 601141 Fax 44 0 1509 601091 Email support campbellsci co uk www campbellsci co uk Contents 1 INtTOdU CIO naar een 1 2 Specifications 2 1 3 MUS CAN ALL o p ida 2 A NV NUON aa ENEE EEEE EEE EEEE eat 3 NS 20 aaa ans aan 3 CAN AS nee E ES 4 5 Reading the Sensor 2 2 4 5
56. tage output V Current output mA degrees Standard Wraparound Standard Wraparound 0 0 0 4 00 4 00 180 2 5 1 67 12 00 9 33 360 5 0 3 33 20 00 14 67 540 n a 5 0 n a 20 00 Low wind speeds in polar mode Whilst the wind speed is below 0 05 metres sec the wind direction will not be calculated In CSV mode the Channel 2 output will switch between 0 and 5V or 4 and 20mA at the output rate selected In fixed field mode the last valid direction will be output until a new value can be calculated Wind Speed scaling The Wind Speed output can be configured such that full scale output represents 30 50 or 60 metres sec In all cases zero Wind Speed is represented by OV or 4mA Tunnel mode When configured in Tunnel mode Channel 1 output represents the wind speed along the U axis South North scaled as described above Channel 2 gives the direction of the wind along the U axis where a U vector wind from the N direction is represented by OV or 4mA and a U vector is represented by 5V or 20mA Configuring Analogue outputs WindCom may be used to change the factory default settings to configure WindSonic alternatively the user may elect to use another terminal emulator package This section describes the commands used to change User and Communications settings 20 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 9 1 Entering Configuration mode see section 8 1 for message format From
57. the Sensor The CS215 output is measured using a standard SDI 12 instruction to read the data from an SDI 12 sensor For Campbell Scientific Edlog dataloggers instruction 105 is used For CRBasic dataloggers the SDI12Recorder instruction is used If using the sensor with other SDI 12 recorders please refer to your system s documentation Further details of the SDI 12 commands can be found on this Web site www sdi 12 org The CS215 complies with a subset of the SDI 12 1 3 instruction set Specifically it supports these SDI 12 commands e aM initiate measurement and the subsequent aDO get data command which is automatically sent by a Campbell Scientific datalogger e aC initiate concurrent measurement where the datalogger gets the data some time later using the aDO Command see the datalogger manual for a more detailed explanation e al send identification e aAb change address to b e query the address of one sensor CS215 sensors with serial numbers after E1587 also support these additional SDI 12 commands Older sensors can be upgraded to do this but as they need to be returned to the factory for an upgrade this is not normally economic e aR or aRO continuous measurements where the sensor will start to make measurements every 11 seconds automatically and return the most recent value without delaying its response to make the measurement see note below This instruction usually takes less than 300 millis
58. til the datalogger responds with the prompt 5 To activate the SDI 12 Transparent Mode on Control Port p enter pX and press the lt enter gt key The datalogger will respond with entering SDI 12 If any invalid SDI 12 command is issued the datalogger will exit the SDI 12 Transparent Mode 6 To query the CS215 for its current SDI 12 address enter the command The CS215 will respond with the current SDI 12 address CS215 Temperature and Relative Humidity Probe 7 To change the SDI 12 address enter the command aAb where a is the current address from the above step and b is the new address The CS215 will change its address and the datalogger will exit the SDI 12 Transparent Mode 8 To activate the SDI 12 Transparent Mode on Control Port p enter pX and press the lt enter gt key Verify the new SDI 12 address by entering the command The CS215 will respond with the new address 9 To exit the SDI 12 Transparent Mode enter 2_ Terminal Emulator z x 8X entering SDI 12 210 0Allexiting SDI 12 8X entering SDI 12 1 exiting SDI 12 Select Device crt OX y VW All Caps Mode Baud Rate 3600 y Close Terminal Clear Help FIGURE 1 Screen Capture of SDI 12 Transparent Mode on Edlog Array Based Datalogger CR10X using Control Port 8 and Changing SDI 12 Address from 0 to 1 6 2 Table Based Edlog Dataloggers 1 Connect a single CS215 to the datalogger Con
59. ting bracket white Stainless steel U bolt mounting clamp Dimensions 12 cm 4 7 in D x 27 cm 10 6 in H Mounting fits vertical pipe 25 50 mm 1 2 in diameter Weight Net wt 0 7 kg 1 6 Ib Ship wt 1 4 kg 3 2 Ib MULTI PLATE RADIATION SHIELD INFLUENCE OF VENTILATION RATE AND SUN ANGLE ON INTERNAL TEMPERATURE RADIATION INTENSITY 1080 W m 70 SUN ANGLE e o T E lt uj a lt mi a x lt 90 SUN ANGLE VENTILATION RATE m s DATA SUMMARIZED FROM COMPARISON TESTING OF SELECTED NATURALLY VENTILATED RADIATION SHIELDS BY GERALD C GILL SEPTEMBER 1983 R M YOUNG COMPANY 2801 Aero Park Drive Traverse City Michigan 49686 U S A TEL 231 946 3980 FAX 231 946 4772 E mail Met sales youngusa com 30 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 REFERENCES MULTI PLATE RADIATION SHIELD References containing additional information about development and applications of the Multi Plate Radiation Shield are listed below in chronological order Gill G C Development of a Small Rugged Radiation Shield for Air Temperature Measurements on Drifting Buoys Report to NOAA Data Buoy Office for Development Contract 01 7 038 827 IF 1979 23 pp 17 figs Gill G C Comparison Testing of Selected Naturally Ventilated Solar Radiation Shields Report to NOAA Data Buoy Office for Development Contract NA 82 OA A 266 1983 15 pp 15 figs Gammill B Tempera
60. tions of very low windspeed and when taking measurements very frequently For instance in still air measuring the probe at second intervals will lead to an over reading of temperature by up to 0 3 C and at 2 second intervals the error is about 0 1 C Therefore for the best accuracy it is advisable to read to sensor no more often than every 5 seconds In addition to issues with self heating the length of time a CS215 measurement takes can limit the speed at which a program table can run especially if there are other lengthy measurements too The following sections describe this issue and ways to optimise the program to avoid this 5 2 1 Edlog Dataloggers The aM and aR commands takes about 740 and 300 milliseconds respectively to make a measurement from the CS215 If the SDI 12 instruction is included in the main program table table 1 in Edlog dataloggers or the main scan in CRBasic dataloggers the program will be delayed for this interval which will limit the maximum scan rate for fast running programs CS215 Temperature and Relative Humidity Probe The aR Command would therefore be preferable in fast running programs as it takes less time to execute The aC concurrent measurement command can also be used where the sensor measurement is initiated with one command and data is collected after a minimum delay of one second or any time thereafter With Campbell Scientific dataloggers this is done by using the SDI 12 recorder instruction
61. trol Port p as described in Section 4 and download a datalogger program that contains the SDI 12 Recorder Instruction 105 instruction with valid entries for each parameter 2 Inthe LoggerNet Toolbar navigate to and activate the TestITerminal Emulator menu The Terminal Emulator window will open In the Select Device menu located in the lower left hand side of the window select the station 3 Click on the Open Terminal button If communications between the datalogger and PC are successful the red bar located in the upper left hand side of the window will turn green 4 Click inside the Terminal Emulator window and press the lt enter gt key until the TD datalogger responds with the gt prompt 10 2 Terminal Emulator u xj User Guide 5 To activate the SDI 12 transparent mode on Control Port p enter 8 The TD datalogger will respond with a prompt At the prompt enter The TD datalogger will respond with 150000 Finally enter p Control Port p and press the lt enter gt key The TD datalogger will respond with entering SDI 12 6 To query the CS215 for its current SDI 12 address enter the command The CS215 will respond with the current SDI 12 address 7 To change the SDI 12 address enter the command aAb where a is the current address from the above step and b is the new address The CS215 will change its address and the datalogger will exit the SDI 12 Transparent Mode 8
62. ture Sensor Shield Comparison Internal Test Report Physical Science Laboratory New Mexico State University 1985 Payne R E Air Temperature Shield Tests WHOI 87 40 Woods Hole Oceanographic Institution 1987 22 pp Payne R E The MR A Meteorological Data Sensing Recording and Telemetering Package for Use on Moored Buoys Journal of Atmospheric and Oceanic Technology Vol 5 No 2 1988 pp 286 297 Crescenti G H Payne R E and Weller R A Improved Meteorological Measurements from Buoys and Ships I MET Preliminary Comparison of Solar Radiation Air Temperature Shields WHOI 89 46 IMET TR 89 03 Woods Hole Oceanographic Institution 1989 53 pp Weller R A and Hosom D S Improved Meteorological Measurements from Buoys and Ships for the World Ocean Circulation Experiment Proceedings Oceans 89 Seattle IEEE 1989 pp 1410 1415 Fougere A J Brown N L and Hobart E Digital Output Temperature Sensing Module for Oceanographic amp Atmospheric Measurements Proceedings Marine Instrumentation 90 San Diego Marine Technology Society 1990 pp 46 51 Anderson T and Mattison l A Field Test of Thermometer Screens Report 900426 Swedish Meteorological and Hydrological Institute 1990 15 pp 19 figs 31 13 5 Rotronic HyrgoClip S3 manual ROTRONIC HvsroCL r Mit dem ROTRONIC HygroClip haben Sie ein Produkt erworben das nach ISO 9001 gebaut und nach EN50082 2 EMV vertr glich ist Bei
63. ty 0 100 rh all types Operating range probe dependent Measuring range Humidity 0 100 rh Temperature probe dependent Outputs Probe type 40 85 C 0 4 0 85 V S C R SC04 SCO5 SP05 HP28 HS28 40 60 C 0 1 V 3 C3 R3 One Wire DIO All probe types 1 measurment value after 235 Measurement interval 07s Supply 3 5 24 VOC ca 3 mA Precision at 23 C 1 5 rh 0 3K Pt100 1 3 DIN Load gt 10 kOhm Protection rating 1P65 except sensor filter area Note select the supply voltage as low as possible Mounting all plug in HygroClip probes with DATOS connector fitting hand helds and MOK adaptors A Align the 4 marking points according to sketch A Insert the probe with slight axial pressure into the bayonet ring Then turn the bayonet ring only clockwise Make sure not to twist the probe itself Removing of HygroClip sketch B Turn the bayonet ring only counterclockwise till the stop and pull off in axial direction Fasten the MOC mounting connector with the nut provided Bore diameter 12 5 mm HygroClip C and C3 only Warning The HygroClip probes are plug in types They must under no circumstances be twisted Twisting of the connectors will destroy probe and mating connector and void warranty Electrical connection according to sketch C Adjustment calibration The HygroClip has been adjusted at the factory We recommend calibration after about 1 year depending on use Y
64. v1 bye 2 SIG GND 2 5 j n c Zuo SD112 SIG 4 Me Connections from Campbell a ls N A eee N C 4 sw 3 ne ABLE SDI12 Temperature and Ne 5 2 n c sisi Relative Humidity Sensor SCREEN 6 Li N C TB2 TB4 14 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 Connections to WindSonic cable for MetPak options 5 and 6 MetPak Option 5 SDI 12 Colour Function Orange V k 9 16 Vdc Orange White V Green Signal Ground GND Blue White RX Blue TX pro Serial Dataline Drain Wire Screen Cable Screen MetPak Option 6 Analogue Colour Function Orange V V 10 28Vdc Orange White V V Green Signal Ground SIG Ground Brown Analogue 1 Brown White Analogue 2 Analogue Voltage Output a Programming output Blue White RX PR mm 15 MetPak Doc No 1575 PS 0001 Issue 02 CN 4240 March 2006 7 MECHANICAL Orientation Normally the MetPak is mounted on a vertical pole ensuring a horizontal Measuring Plane See Figure 3 Alignment amp Mounting details For indoor use the unit may be mounted with the Measurement Plane set to any required orientation Alignment The MetPak should be aligned to point to North or any other reference direction for example the bow of a boat There are two arrows a coloured rectangle and an alignment notch to aid alignment See Figure 3 Alignment amp Mounting Details Note It is usually simpler to work first with a compass at ground level a
65. w 0 C A 2 The CS215 provides a humidity reading that is referenced to the saturated water vapour pressure above liquid water even at temperatures below 0 C where ice might form This is the common way to express relative humidity and is as defined by the World Meteorological Organization If an RH value is required referenced to ice the CS215 readings will need to be corrected One consequence of using water as the reference is that the maximum humidity that will normally be output by the sensor for temperatures below freezing is as follows 100 RH at 0 C 95 RH at 5 C 91 RH at 10 C 87 RH at 15 C 82 RH at 20 C 78 RH at 25 C 75 RH at 30 C In practical terms this means that for instance at 20 C the air is effectively fully saturated when the sensor outputs 82 RH Appendix B Sensor Element Calibration The sensor elements are individually calibrated during manufacture The manufacturer publishes the certificate below as proof of traceability CERTIFICATION SENSIRION THE SENSOR COMPANY Product SHTxx Family Description Digital Humidity and Temperature Sensor This is to certify that above product is calibrated using dedicated equipment which is calibrated according to ISO IEC 17025 and calibrated by comparison against natural standards natural physical constants consensus standards or by ratio type measurements using self calibrating techniques Standards incorporated to establish this comparison are
66. w campbellsci es info campbellsci es Please visit www campbellsci com to obtain contact information for your local US or International representative
67. with the aC command The datalogger will return 99999 or NaN for CRBasic loggers for the temperature reading for the call of the instruction that initiates the measurement At the next call of the instruction the datalogger will request the data and record the correct temperature 5 2 2 CRBasic Dataloggers CRBasic dataloggers that support the Slowsequence function the CR1000 and CR5000 can run the SDI 12 instruction entirely as a background process causing minimum interference to other measurements that use the analogue hardware This is implemented by including the measurement in a Slowsequence section of the program thus allowing faster programs to run as the main scan NOTE For the CR5000 use a control port rather than the SDI 12 port to allow the SDI12recorder instruction to run in the slow sequence The example program below shows how the Slowsequence can be used to make measurements from the CS215 whilst the main scan can run at a much faster rate CR5000 Series Datalogger Example program showing measurement of a single CS215 sensor every 30 sec in the slow sequence Declare the variables we are going to use Public Batt_Volt PTemp CS215meas 2 An array suitable to reading the SDI12 data into Alias CS215meas 1 Temperature Alias CS215meas 2 RH Main Program in this example it is running fast just measuring the battery voltage and panel temperature BeginProg Scan 50 mSec 0 0 Scan every 50 millisec
Download Pdf Manuals
Related Search
Related Contents
User Manual LON Power Supply LPS133 Art. no 994LZ Luxury Lift Oper.qxd - La-Z-Boy F I A T P A N D A 00X39VL0 T000.book Massive Top Selection Spot light 53143/31/10 Manuel de l`utilisateur - Infosec Bianca Concolino Mancini Abram, « Métathéâtre et folie Manuel d`entretien et d`utilisation IMPORTANT! User Manual iView Copyright © All rights reserved.
Failed to retrieve file