MORPHEUS E Workstation for gaseous anaesthesia
Contents
1. 15 MONITORS type selection optional e Software for analysis of CO2 O2 N20 AG automatic identification MAC based on weight e Mainstream external device e Sidestream built in device ELECTRIC POWER SUPPLY Electric power supply 100 240Vac 45 60Hz Maximum power 120 Watt Back up battery 12Vdc 3 Ah pb battery which guarantees an autonomy of around 120 minutes Charging time Around 10 hours Morpheus_E_te_rev0 7 8 CONFORMITY TO DIRECTIVES Class and type according with IEC 601 1 Class Type B Class according with 93 42 CEE Class IIb IEC 601 1 IEC 601 1 1 IEC 601 1 2 IEC 601 1 4 EN 1281 1 UNI EN 740 UNI EN ISO 9703 3 EN 4135 93 42 EEC DIR ACCESSORIES Standard accessories User s Manual O2 supply hose N20 supply hose Air supply hose O cylinder supply hose N20 cylinder supply hose IS PAC box of 0 5 kg 10 pcs Oo cell Adult silicone patient circuit Adult Mapleson C adult patient circuit Manual ventilation KIT SHUKO VDE electric power supply cable Other optional accessories See current export price list SIARE applies the UNI EN ISO 13485 2004 Quality System and 93 42 EEC Dir ve SIARE ENGINEERING INTERNATIONAL GROUP s r l Via Pastore 18 40056 Crespellano Bologna ITALY Tel 39 051 969802 Fax 39 051 969366 E mail mail siare it http www siare it Morpheus_E_te_rev0 8 8
2. Inspiratory Pause From 0 2 to 5 sec From 0 to 60 of inspiratory time SIMV rate From 0 to 119 bpm Tidal Volume From 10 to 1500 ml Accuracy on Tidal Volume measurement 20 of real reading over 100ml 20ml of real reading under 100ml Minute Volume From 1 to 30 liters PEEP OFF 3 30 cmH20 Inspiratory Flow From 1 to 80 liters min Minimum resolution 1 Oximeter Automatic calibration procedure Bronchomanometer 20 80 cmH20 Flow trigger From 1 to 15 It min Safety Electronic and mechanical limit of airways pressure Self diagnosis system Alarms e Low High airways pressure Low High breathing rate Low High O2 concentration Low High Tidal Volume e Disconnected Patient Circuit Apnoea Apnoea Back Up Low Battery Electric Power Supply Medical Gases Flow sensor Internal to the group by magnetic perturbance reusable Morpheus_E_te_rev0 6 8 Measured parameters e PAW PEEP Vte ExpMV Rate I E FiOo e Pmax Pmean Pause Vti FLOWi FLOWe Tinsp Tesp Tpause Cs Ri Ventilation Curves CURVES Pressure Flow Volume LOOPs Volume Pressure and Flow Volume Measurement RANGE automatic Trend Scale and 72 hours trend period setting Foreseen Trends PAW PEEP VTe ExpMV Rate Events Memory storage up to 100 events per machine including the alarms Configuration and gas THIS FUCTION CAN BE AVAILABLE ON NEPTUNE 12
3. MORPHEUS E Workstation for gaseous anaesthesia code OM3 SE rev dated 02 05 2011 INTENDED USE The MORPHEUS E is a workstation for gaseous anaesthesia and it can be used on adult children and newborn patients The MORPHEUS E is suitable for administration of Oxygen Air Nitrous Oxide Halothane Enflurane Isoflurane Sevoflurane Desflurane mixtures GENERAL DESCRIPTION The unit for anaesthesia MORPHEUS E is completed with Morpheus_E_ te_rev0 Electronic gas mixing system Electronic lung ventilator with 12 TFT colour display Valves group open semi closed closed heated with soda lime absorber 0 5 Kg SIARETEX rapid connection device Selectatec compatible for 2 vaporizers Gas supply group TECHNICAL DATA Structure Wheels Drawer Cylinder support Support for 2 vaporizers Auxiliary power supply outlets Work shelf lighting Dimensions Weight Environmental conditions Light aluminium alloy and plastic moulds Pivoting antistatic wheels diameter 100 mm 2 with brakes No 3 full extension drawers No 2 vertical cylinders supports on the back side for cylinders up to 10 litres capacity and round rubber pads On horizontal guide SIARETEX rapid connection device Selectatec compatible for 2 vaporizers No 1 SCHUKO 220 Vac outlet max 6 A 12Vdc by led 71 x 77 x 138 L x P x H cm without monitor 72 kg without accessories e Temperature from 10 to 40 C e Re
4. atic deviation of fresh gas to exit of fresh gas for the manual system of anaesthesia unit valves group or to a to and fro circuit manual ventilations with visual indicator Automatic deactivation of manual ventilation systems directly by ventilator control Os emergency by pass By push button max flow 35 l min IN gas sockets on gas e No 3 sockets for distribution system O2 N20 AIR supply group e No 2 sockets for cylinder O2 N20 OUT gas sockets on gas e No 1 sockets for Oo supply group e No 1 sockets O2 AIR for active scavenger feeding e No 1 fresh gas connector for external use for ex TO AND FRO selectable by apposite membrane key on the front shelf Other e Socket for recycle of exhaust monitor gas e Connection for anaesthetic gas scavenging optional device active type or passive type Morpheus_E_te_rev0 4 8 BREATHING SYSTEM General description Compact system with automatic connections easy dismountable and autoclavable It allows the ventilation in modality real open circuit semi closed circuit closed circuit at low flows The system also allows the spontaneous and manual ventilation in case of machine breakdown or machine off The CO2 absorber canister ISI PAC of 0 5 Kg whose rapid disconnection system allows its replacement also during operation The recycling system is a selective type hence the soda lime and fresh gas consumption are reduced to the minimum The heated circ
5. lative humidity from 10 to 90 non condensing ELECTRONIC GAS MIXING SYSTEM General description Auto test Morpheus_E_te_rev0 It has the function to regulate the capacity and the concentration of gas mixture Air O2 N20 by displaying them on the TFT colour monitor and deliver it to the anaesthetic gas vaporizer It allows to select the mixture to be delivered Air O9 or No0 O9 and the Oo enrichment for delivered mixture in case of emergency The anaesthesia module includes a device which guarantees a minimum concentration of 25 oxygen in all gas erogating conditions MIX LIFE device Through the three pressure gauges on the front panel it allows the continuous control of medical gas feeding pressure coming from the gas pipelines system The electronically controlled flowmeter box is equipped with a safety traditional calibrated flowmeter indicating the total delivered flow The electronic flowmeter box provides the option on demand to use an alternative anesthetic gas in spite of N2O the Xenon At electronic flowmeter start up a control test is automatically performed 2 8 Oxygen rotameter Nitrous oxide rotameter Air rotameter Medical gas supply Gauges Electric power supply Back up battery Alarms Morpheus_E_te_rev0 Scale 0 1 15 l min Resolution 0 1 l min Accuracy 7 of read value Scale 0 1 9 l min Resolution 0 1 l min Accuracy 7 of read value Scale 0 1 15 l min Re
6. solution 0 1 l min Accuracy 7 of read value OXYGEN e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 90 l min NITROUS OXIDE e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 15 l min MEDICAL COMPRESSED AIR e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 90 l min No 3 on front panel O2 N20 AIR scale 0 10 bar 100 240Vac 45 60Hz 12Vdc 3 0 Ah Pb battery it grants 120 minutes operation e Lack or low oxygen pressure with consequent cut off of nitrous oxide delivery e Lack or low nitrous oxide pressure e Lack or low air pressure e O02 concentration lower than 21 e Power failure e Low battery e Gas leakage 3 8 Safety devices AGAINST THE ADMINISTRATION OF HYPOXIC MIXTURES MIX LIFE it always guarantees a minimum concentration of 25 oxygen on mixtures which includes nitrous oxide IN CASE OF LACK OR LOW OXYGEN PRESSURE CUT OFF audible alarm with immediate cut off of nitrous oxide delivery AGAINST HIGH PRESSURE IN SUPPLY Safety valve calibrated at 0 8 bar IN CASE OF LACK OR COMPRESSED AIR LOW PRESSURE All the devices gas feeding supplied by compressed air are automatically supplied by oxygen AGAINST THE SIMULTANEOUS DELIVERY OF AIR AND N20 e Selector button on the flowmeter front panel Control for activation of Setting of manual modality on ventilator with autom
7. uit reduces the condensation and heats the fresh gas The transition from one modality to another is completely controlled by the ventilator without any user s action on valves group LUNG VENTILATOR User s interface 12 TFT high resolution colour display with membrane keyboard and encoder Control modality Electronic by microprocessor Dead space compensation system Automatic Flow generation Electronic system Gas feeding Medical compressed air or Oxygen with pressure included between 280 kPa and 600 kPa 2 8 6 bar Option Turbine for gas feeding independent from gas pipelines system Morpheus_E_te_rev0 5 8 Autotest Primary test At device s start up a control test of medical gas supply circuit losses INSP and EXP flow sensors operation pressure sensor patient circuit losses back up battery state oxygen cell integrity of audible alarm is automatically performed This test takes around 30 seconds Subtest the device has a subtest which is activated by the user in the ventilator menu This subtest permits to verify the dead space and losses or to perform the oxygen cell calibration Ventilation modalities Ventilation modalities optional Breathing rate 1 E Ratio MANUAL VC VAV APCV SPONT SIMV PS volumetric PSV APNOEA BACK UP PCV TV CPAP SIMV PS by pressure BILEVEL S ST From 5 to 120 bpm 1 4 4 1 Inspiratory time