1. BOND icon WINDOW Home screen SELECT from the function bar Slide setup icon CLICK 4 CLICK in the field and type the accession or what how you want to label the slide OK to close the Add case dialog Select the case ID in the case list on the left of the screen CLICK Add slide tab CLICK on the arrow down from the Marker tab to select antibody CLICK i Repeat from the list per antibody ordered After all slides have been added click Close to close the Add slide dialog Review the details in the slide list found on the right side of the screen If you need to change details for a slide right click on the slide Select Slide properties from the submenu then click OK CLICK from the Slide setup screen CLICK then CLICK Print Apply each printed BOND bar code label to its corresponding slide Apply the label over the painted section of the slide without extension over any slide edge Install an individual slide in an empty position in the Slide tray with the label end on the opposite side of the Place the slide down so it sits in the recessed position in the tray Hold a covertile by the head and position it over the slide with the key in the neck aligned with the recess in the tray Lay the covertile on the side fitting the key into the slot Tally the covertile with a statmark pen Toss the covertile when the tally reaches 10 or when the covertile is scratched worn or cracked Lift the tray and rest the end on the e
2. patient tissue sample on a 25 x 75 mm positively charged microslide 2 Using a permanent marking pen label slide with e antibody name e procedure date e accession number at bottom edge PATIENT 3 Routinely dry slides at 60 C in convection oven for 30 min 4 Remove from IHC refrigerator the reagent tray s containing any and all detection kits and primary antibody containers required for protocols being performed Prepare for operation by uncapping and loading tray on the instrument 5 Allow antibodies reagents to reach room temperature prior to staining 6 Preliminary check before start up Automated Immunohistochemical Staining Bond Page 2 of 5 10 11 12 13 14 15 16 17 Ensure the processing module is clean and up to date by checking the Cleaning and Maintenance Schedule Daily prerun tasks are checking the bulk waste containers are empty and the bulk reagents have sufficient correct reagent in them Check wash blocks and mixing station clean or replace Check that the slide labeler has enough supply of labels Turn ON BOND III instrument at the operational PC CPU and at also at the red switch located facing the right side of the instrument Sign on using name bonduser password bond12345 Check the Status screen to ensure there are no processing module notifications Rectify before attempting to run any slides Prepare slide barcode labels From the PC desktop menu select the
3. Ip Pacific Pathology Associates Inc Clinical Histology Procedure histo13 01 Automated Immunohistochemical Staining Utilizing the Bond Ill Instrument Final Approval October 2010 Effective October 2010 Next Review Date February 2013 List all stakeholder s and dates of approval Stakeholder Name s Shelly Siegel HT Date 9 21 10 Reviewed _ Revised x Stakeholder Name s Mark Magilner MD Date 10 4 2010 Reviewed x Revised _ Stakeholder Name s Shelly Siegel Date 10 11 11 Reviewed X Revised _ Stakeholder Name s Shelly Siegel Date 2 14 12 Reviewed X Revised _ Stakeholder Name s Date Reviewed _ Revised _ Describe briefly the most recent revision made to this policy procedure or protocol amp why New Procedure Purpose Policy Statement The demonstration of antigens in tissue and cells by immunostaining is a process involving first the binding of an antibody to an antigen of interest and second visualization of the bound antibody by a biotin free polymeric horseradish peroxidase HRP The Bond Polymer Refine Detection system detects tissue bound mouse and rabbit IgG and some mouse IgM primary antibodies The detection kit is intended for staining sections of formalin fixed paraffin embedded tissue on the Bond automated system It utilizes a novel controlled polymerization technology to prepare polymeric HRP linker antibody and avoids the use of streptavidin and biotin Therefore e
4. ime a five gallon carboy of bulk ancillary reagent is emptied it shall be decontaminated then rinsed thoroughly with DH20 wiped dry and or drained Routine instrument maintenance is performed daily monthly and quarterly by a staff Histologist and documented New lots of antibody reagents are compared to the previous lot using an appropriate panel of control tissue Documentation of validated new antibodies are maintained on the Antibody Quality Control Sheet Control tissue blocks are validated prior to patient testing It is documented and maintained on the Tissue Block Quality Control Sheet The director of Pathology validates all antibodies and tissue control blocks before patient use P M is performed annually by a Bond representative under a purchased manufacturer service maintenance contract PROFICIENCY TESTING CAP Proficiency Survey Immunohistochemistry CAP HistoQip HER2 and ER PR RESULTS Ti 2 Positive results will be demonstrated with variations of colored staining based upon use of specific detection kit systems negative results will be demonstrated by the presence of counterstain only Positive result on negative control may be an indicator of endogenous biotin in certain types of tissue sample brain breast kidney or most frequently liver in some patient cases See QUALITY CONTROL PROCEDURE NOTES In the event of an equipment downtime arrangements have been made with Northwest Pathology in Eugene OR for testi
5. liminates non specific staining as a result of endogenous biotin The Bond DAB enhancer precipitates as a copper salt at the reaction site producing a darker brown staining deposit Bond Polymer Refine Detection Catalog DS9800 200 300 tests works as follows e The specimen is incubated with hydrogen peroxide to quench endogenous peroxidase activity e A post primary antibody solution enhances penetration of the subsequent polymer reagent e A poly HRP anti mouse rabbit IgG reagent localizes the primary antibody e The substrate chromogen 3 3 diaminobenzidine DAB visualizes the complex with a brown precipitate e Hematoxylin counterstaining allows the visualization of cell nuclei Bond Polymer Refine Red Detection a biotin free polymeric alkaline phosphatase AP linker antibody conjugate system Catalog DS9390 100 tests works as follows e Auser supplied specific primary antibody is applied e Post primary IgG linker reagent localizes mouse antibodies e Poly AP IgG reagent localizes rabbit antibodies e The substrate chromogen Fast Red visualizes the complex with a red precipitate e Hematoxylin counterstaining allows the visualization of cell nuclei Automated Immunohistochemical Staining Bond Page 1 of 5 Slide barcode labels operate the instruments automated sequential dispensing of appropriate reagent and antibody dilutions Definitions e na STEPS KEY POINTS SPECIMEN Paraffin sections of 3 micron or 4
6. micron thickness are collected from an 40 45 C flotation bath containing deionized water and mounted on 25 x 75 mm positive charged slides or cytological samples prepared by smear or Cytospin technique fixed in 95 alcohol MATERIALS REAGENTS Bond system ancillary hardware consumables e Bond Universal Slide Labels Printing Ribbon Slide Labeller type TLP 3842 e Handheld ID scanner e Bulk reagent waste containers e Covertiles slide trays reagent trays Ancillary Products Reagents CAUTION STRONG OXIDIZER IRRITANT ee iat A ee Wear appropriate protective equipment ear appropriate protective equipment e Bond wash solution 10X AR9590 e Bond Ready to use antibody dispenser open e Bond Epitope Retrieval Solution 1 AR9961 comer e Anti Fluorescein antibody AR0222 e Bond Epitope Retrieval Solution 2 AR 9640 Aspirating Probe Cleaning System CS9100 e Bond dewax solution AR9222 e Enzyme Pretreatment Kit AR9551 e Primary Antibody Diluent AR9352 e DAB Enhancer AR9432 INSTRUMENTATION OR EQUIPMENT Bond Ill instrument PROCEDURE 1 From a tissue flotation bath of deionized H2O retrieve the cut paraffin section of the patient case on the bottom of the microslide which contains the known positive tissue or multi tissue controls on the top of the slide for the requested antibody Only in circumstances where the patient tissue sample is larger than the provided slide space use the slide as a control only and place the
7. ndial icon Brief Slide history change to current day and beginning of run to the end Print to the HP deskjet printer Use the Brief Slide history sheet to quality control check all slides before distributing to the pathologists Any commenis staining issues need to be noted on the sheet and communicated to the pathologist for immediate repeats if necessary CALIBRATION Annual calibration is performed on each micropipette by a professional service QUALITY CONTROL 1 9 Positive result on negative control may be an indicator of endogenous biotin in certain types of tissue sample brain breast kidney or most frequently liver in some patient cases The resulting negative result on the negative control validates the prior staining reaction as endogenous biotin Each time the procedure is performed a known positively reactive tissue for the specified antigen shall be simultaneously processed with the patient tissue For each patient block on which IHC is performed a corresponding patient NEGATIVE control slide shall be simultaneously stained with the patient slide s excluding the application of the Primary Antibody NEGATIVE control receives most harsh treatment of all protocols performed on the case Ventana PATHWAY HER 2 4 in 1 Control Slide validates staining quality per each protocol run for HER2 The pH of Reaction Buffer is measured and documented on the pH Buffer form each time a working solution is prepared Each t
8. ng Equipment Supplies If Applicable Form Name amp Number or Attachment Name If Applicable Histo13 Author Position Lead Histologist Review Revision Authority Position Not Individual Name Lead Histologist Expert Consultant Position s Not Individual Name s Automated Immunohistochemical Staining Bond Page 4 of 5 References Required for Clinical Documents MANUFACTURER S PACKAGING BROCHURE INSERT Leica Bond User Manual Rev H01 Leica Biosystems Melbourne 2009 Is there a Regulatory Requirement Yes No _ If yes insert requirement information here Review History No Changes Revision History Note changes in area under header Computer Search Words Policy Procedure or Protocol Cross Reference Information Automated Immunohistochemical Staining Bond Page 5 of 5
9. ntrance to an empty slide staining assembly Position the right runner of the tray in the guides and gently slide the tray as far as it will go into the module The tray should slide in easily and audibly click when it is in place Press the Load Unload button on the front panel beneath the loaded tray lt there are reagents are not available the software displays an attention icon below the slide list Right click on the icon for more information The slides are now ready for a staining run The progress bar will be in the staining phase and the run status will be Slides ready CLICK the play to go icon to begin immediately or delay by right clicking on the slide tray The progress bar will switch to the processing phase and the run status will be Proc OK When finished the processing module tab icon flashes green Following completion of the instrument staining program e Press the Load Unload button and remove slide trays e Remove the covertiles by lifting the end off the slide Do NOT SLIDE across the surface of the slide you may damage the tissue Place covertiles in 70 alcohol container or toss at 10 uses e Place slides in rack and remove liquid coverslip by manually dipping in 3 changes of 100 alcohol and 2 changes of Xylene Coverlip on automated tissue tek and let slide dry a few minutes before distributing to the pathologists Automated Immunohistochemical Staining Bond Page 3 of 5 18 Print from the Su